Vortex Flowmeter Manual

1.Tehcical data:

1. Accuracy class: Liquid-1.5,Gas-1.0,Steam-1.0;

2. Nominal pressure: 1.6MPa, 2.5MPa, 4.0MPa;


3. Nominal diameter: Pipe segment/card mounted: DN15-300mm; Insertion:
DN200-2000mm;
4. Flow rate: Liquid 0.27-9m/s, Gas: 4-40m/s, Steam: 3-50m/s;
5. Medium temperature: -40℃~350℃;
6. Pressure loss: drag coefficient cd≤2.4;
7. Power supply: pulses output: +12VDC; 4-20mA output: +24VDC;
LCD display on site: 3.6V5 No.1 lithium battery power supply, with a service
life of more than 2 years;
8. Ambient temperature: pulses output: -30~+65℃;
4-20mA output: -10~+55℃; LCD display on site: -25~+55℃;

9. Explosion proof: IaⅡCT6 (Intrinsic safety);


10. Output signal: Voltage pulse: Low level ≤ 1V, high level ≥ 12V, pulse width
0.4ms, load resistance>150 Ω,
On site LCD display: 5-digit display of instantaneous flow rate (m3/h, kg/h, t/h),
Conversion accuracy ± 0.1%; Accumulated flow display with 8 digits (m3, kg, t),
conversion accuracy ± 0.1%;
11. Applicable media: gases (air, oxygen, nitrogen, coal gas, natural gas,
chemical gases, etc.), liquids (water, high-temperature water, oil, food liquids,
chemical liquids, etc.), steam (saturated steam, superheated steam).
2. Flow range: unit:m³/h

liquid gas
Size
Standard Extended (20℃,101325Pa)

3. LUGB-Z Integrated Temperature and Pressure System

1. Double row LCD, 8-bit digital display;


2. Users can set their own small signal cutoff;
3. 4-20mA two-wire current output (equipped according to user ordering
requirements);
4. Online temperature and pressure compensation, manual constant density
compensation;
5. 12V~24V or 3.6V lithium battery power supply;
6. The temperature pressure integrated vortex flowmeter comes with
temperature and pressure compensation, making installation more convenient.
4. Installation:

The sensor should be installed on a pipeline with the same diameter as the
horizontal, vertical, and inclined (liquid flow from bottom to top) pipeline.
sensor. Upstream and downstream of the pipeline should be equipped with a
certain length of straight pipe section, which should conform to the length of
the front straight pipe section 15-20D and the rear straight pipe section 5-10D
The requirements.
The pipeline near the installation of the liquid sensor should be filled with the
measured liquid. Sensors should be avoided from being installed in pipes with
strong mechanical vibrations on the road.
The inner diameter of the straight pipe section should be as consistent as
possible with the sensor diameter. If it cannot be consistent, a pipe slightly
larger than the sensor diameter should be used, the error should be ≤ 3%
and not exceed 5mm.
When the measured medium contains a large amount of impurities, a filter
should be installed beyond the required length of the straight pipe section
upstream of the sensor.
Sensors should be avoided from being installed in places with strong
electromagnetic field interference, small space, and inconvenient
maintenance.