0% found this document useful (0 votes)
95 views2 pages

Syringe Pump KL-6061 N: The Don Medical Devices

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
95 views2 pages

Syringe Pump KL-6061 N: The Don Medical Devices

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

The Don Medical Devices

Syringe Pump KL-6061 N

Work with
docking station

�:::a I syn "'e I


I" ii

Flow Rate 8 rrL h


VTBI 17 ml

1 rifused ml Wide range flow


rate and VTB I

Occlusion
sensitivity: 12
levels

Innovative Infusion Systems, Since 1994


: (
Syringe Pump
KL-6061N
Technical Specifications
Model Kl-6061N
Syringe Size 5, 10, 20, 30, 50/60 ml
Applicable Syringe Compatible with syringe of any standard
VTBI 0.1-99999.99 ml (in 0.01 ml increments)
Infused Volume 0-99999.99 ml (in 0.01 ml increments)
Syringe 5 ml: 0.1-100 ml/h
Syringe 10 ml: 0.1-300 ml/h
Syringe 20 ml: 0.1-600 ml/h
Syringe 30 ml: 0.1-800 ml/h
Flow Rate
Syringe 50/60 ml: 0.1-1500 ml/h
0.1-99.99 ml/h in 0.01 ml/h increments
100-999.9 ml/h in 0.1 ml/h increments
1000-1500 ml/h in 1 ml/h increments
Bolus Rate (Manual, Auto) 50 ml/h-1500 ml/h (in 0.01, 0.1, 1 ml increments)
Bolus Volume 0.1-50 ml (in 0.1 ml increments)
Anti-Bolus Automatic
Accuracy ±2% (mechanical accuracy :51%)
Rate mode, micro mode, time mode, body weight,
Infusion Mode
initial dose mode, sequential mode, ramp up/down mode
KVO Rate 0.1-10 ml/h (in 0.1 ml/h increments)
Occlusion, near empty, end program, low battery, end battery, no battery
Alarms AC power off, motor malfunction, system malfunction, standby,
syringe installation error, syringe drop off
Automatic power switching, automatic syringe identification, mute key,
Additional Features purge, bolus, anti-bolus, system memory, key locker,
change flow rate without stopping the pump
Drug Library Available
Occlusion Sensitivity 12 levels (20-130 kPa, in 10 kPa increments)
Occlusion Unit kPa, mmHg, psi, bar
History Log 100,000 events
Alarm Volume 8 levels
Auto. multi-channel relay Available
Power Supply, AC AC100-240V 50/60Hz
Battery DC14.4V, rechargeable
Battery Life 1O hours at 5 ml/h
Working Temperature 5-40 'C
Relative Humidity 15-80%
Atmospheric Pressure 860-1060 hpa
Size 290*175*84 mm
Weight 2.5 kg
Waterproof IPX3
Safety Classification Class II, type CF