SDS2000X HD - ServiceManual - EN01A
SDS2000X HD - ServiceManual - EN01A
Digital Oscilloscope
Service Manual
EN01A
Contents
This manual contains information and warnings that must be followed by the user for safe operation
and to keep the product in a safe condition.
Carefully read the following safety precautions to avoid personal injury and prevent damage to the
instrument and any products connected to it. To avoid potential hazards, please use the instrument
as specified.
Only use a local/state-approved power cord for connecting the instrument to mains power sources.
The instrument grounds through the protective terra conductor of the power line. To avoid electric
shock, the ground conductor must be connected to the earth. Make sure the instrument is grounded
correctly before connecting its input or output terminals.
The potential of the signal wire is equal to the earth, so do not connect the signal wire to a high voltage.
Do not touch the exposed contacts or components.
To avoid fire or electric shock, please look over all ratings and signed instructions of the instrument.
Before connecting the instrument, please read the manual carefully to gain more information about
the ratings.
After the equipment is started, there will be no alarm information and error information at the interface
under normal conditions. The curve of the interface will scan from left to right freely. If there is a button
in the scanning process or there is an alarm or error prompt, the device may be in an abnormal state.
You need to view the specific prompt information. You can try to restart the setting. If the fault
information is still in place, do not use it for testing. Contact the manufacturer or the maintenance
department designated by the manufacturer to carry out maintenance to avoid the wrong test data
caused by the use of the fault or endanger the personal safety.
If you suspect that there is damage to the instrument, please let qualified service personnel check it.
Only probe assemblies that meet the requirement of UL61010-031 and CAN/CSA-C22.2
No.61010-031 shall be used.
Not to use the equipment for measurements on mains circuits, not to use the equipment for
measurements on voltage exceed the voltage range describe in the manual. The maximum
additional transient voltage cannot exceed 1300 V.
The responsible body or operator should refer to the instruction manual to preserve the
protection afforded by the equipment. If the equipment is used in a manner not specified by
the manufacturer, the protection provided by the equipment may be impaired.
Any parts of the device and its accessories are not allowed to be changed or replaced, other
than authorized by the manufacturer or agent.
When the following symbols or terms appear on the front or rear panel of the instrument or in this
manual, they indicate special care in terms of safety.
This symbol shows that the switch is an On/Standby switch. When it is pressed,
the oscilloscope’s state switches between Operation and Standby. This switch
does not disconnect the device's power supply. To completely power off the
oscilloscope, the power cord must be unplugged from the AC socket after the
oscilloscope is in the standby state.
The " CAUTION " symbol indicates a potential hazard. It calls attention to a
CAUTION procedure, practice, or condition which may be dangerous if not followed. Do not
proceed until its conditions are fully understood and met.
The " WARNING " symbol indicates a potential hazard. It calls attention to a
procedure, practice, or condition which, if not followed, could cause bodily injury
WARNING
or death. If a WARNING is indicated, do not proceed until the safety conditions
are fully understood and met.
The design of the instrument has been verified to conform to the EN 61010-1 safety standard per the
following limits:
Environment
The instrument is used indoors and should be operated in a clean and dry environment with an
ambient temperature range.
Note: Direct sunlight, electric heaters, and other heat sources should be considered when evaluating
the ambient temperature.
Ambient Temperature
Operating: 0℃ to +50℃
Note: Direct sunlight, radiators, and other heat sources should be taken into account when assessing
the ambient temperature.
Humidity
Non-operating: 5% ~ 95% RH
Altitude
Non-operating: ≤ 15,000 m
Installation ( overvoltage ) category II refers to the local power distribution level which applies to
equipment connected to the AC line ( AC power ).
Degree of Pollution
Note: Degree of Pollution II refers to a working environment that is dry and non-conductive pollution
occurs. Occasional temporary conductivity caused by condensation is expected.
IP Rating
This instrument relies on forced air cooling with internal fans and ventilation openings. Care must be
taken to avoid restricting the airflow around the apertures ( fan holes ) at each side of the oscilloscope.
To ensure adequate ventilation it is required to leave a 15 cm ( 6 inches ) minimum gap around the
sides of the instrument.
CAUTION: Do not block the ventilation holes located on both sides of the
oscilloscope.
CAUTION: Do not allow any foreign matter to enter the oscilloscope through the
ventilation holes, etc.
The instrument operates with a single-phase, 100 to 240 Vrms ( +/-10% ) AC power at 50/60 Hz ( +/-
5% ).
No manual voltage selection is required because the instrument automatically adapts to line voltage.
Depending on the type and number of options and accessories ( probes, PC port plug-in, etc .) , the
instrument can consume up to 120 W of power.
Note: The instrument automatically adapts to the AC line input within the following ranges:
Frequency Range 47 - 63 Hz
The instrument includes a grounded cord set containing a molded three-terminal polarized plug and
a standard IEC 320 ( Type C13 ) connector for making line voltage and safety ground connection.
The AC inlet ground terminal is connected directly to the frame of the instrument. For adequate
protection against electrical shock hazards, the power cord plug must be inserted into a mating AC
outlet containing a safety ground contact. Use only the power cord specified for this instrument and
certified for the country of use.
The position of the oscilloscope should allow easy access to the socket. To make the oscilloscope
completely power off, unplug the instrument power cord from the AC socket.
The power cord should be unplugged from the AC outlet if the oscilloscope is not to be used for an
extended period.
1.6 Cleaning
Clean only the exterior of the instrument, using a damp, soft cloth. Do not use chemicals or abrasive
elements. Under no circumstances allow moisture to penetrate the instrument. To avoid electrical
shock, unplug the power cord from the AC outlet before cleaning.
Do not operate the oscilloscope if there is any visible sign of damage or has been subjected to severe
transport stresses.
If you suspect the oscilloscope’s protection has been impaired, disconnect the power cord and secure
the instrument against any unintended operation.
Proper use of the instrument depends upon careful reading of all instructions and labels.
This section lists the safety standards with which the product complies.
Canadian certification
CAN / CSA-C22.2 No. 61010-1:2012 / A1:2018-11. Safety Requirements for Electrical Equipment
for Measurement, Control, and Laboratory Use – Part 1: General Requirements.
CAN / CSA-C22.2 No. 61010-2-030:2018. Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use – Part 2-030: Particular requirements for testing and
measuring circuits.
First, verify that all items listed on the packing list have been delivered. If you note any omissions or
damage, please contact your nearest SIGLENT customer service center or distributor as soon as
possible. If you fail to contact us immediately in case of omission or damage, we will not be
responsible for replacement.
The oscilloscope has a 3-year warranty ( 1-year warranty for probes and accessories ) from the date
of shipment, during normal use and operation. SIGLENT can repair or replace any product that is
returned to the authorized service center during the warranty period. We must first examine the
product to make sure that the defect is caused by the process or material, not by abuse, negligence,
accident, abnormal conditions, or operation.
SIGLENT shall not be responsible for any defect, damage, or failure caused by any of the following:
c) For any damage or malfunction caused by the use of non-SIGLENT supplies. Furthermore,
SIGLENT shall not be obligated to service a product that has been modified. Spare,
replacement parts, and repairs have a 90-day warranty.
The oscilloscope's firmware has been thoroughly tested and is presumed to be functional.
Nevertheless, it is supplied without warranty of any kind covering detailed performance. Products not
made by SIGLENT are covered solely by the warranty of the original equipment manufacturer.
For convenience, text surrounded by a box border is used to represent the button of the front panel.
For example, Default represents the "Default" button on the front panel.
Text with shading is used to represent the touchable or clickable menu / button / region on the touch
screen. For example, Display represents the " Display " menu on the screen:
For the operations that contain multiple steps, the description is in the form of " Step 1 > Step 2 >...".
As an example, follow each step in the sequence to enter the upgrade interface:
Press the Measure button on the front panel as step 1, click the Tools option on the screen as
step 2, and click the Track option on the screen as step 3 to enter the Track setting menu.
Before initiating performance verification or any adjustments, it is recommended the user follow these
procedures. The following topics are discussed in this chapter.
For more detailed information about oscilloscope operation, please refer to the User Manual.
The functional check details four types of checks used to determine if the oscilloscope is operating
properly.
The standard power supply for the instrument is 100~240 V, 50/60 Hz. Please use the power cord
provided with the instrument to connect it to AC power.
Press the Power button located at the lower-left corner of the front panel to start the oscilloscope.
After the boot-up the oscilloscope will begin to perform its power-on tests automatically, after which
the Default button can be pressed to recall the factory default settings.
When a passive probe is used for the first time, you should compensate it to match the input channel
of the oscilloscope. Non-compensated or poorly compensated probes may increase measurement
inaccuracy or error. The probe compensation procedures are as follows:
1. Connect the coaxial cable interface ( BNC connector ) of the passive probe to any channel of the
oscilloscope.
2. Connect the probe to the “ Compensation Signal Output Terminal ” ( Cal ) on the front of the
oscilloscope. Connect the ground alligator clip of the probe to the “ Ground Terminal ” under the
compensation signal output terminal.
5. Use a non-metallic driver to adjust the low-frequency compensation adjustment hole on the probe
until the waveform displayed is as the “ Perfectly compensated ” in the figure above.
Press the Auto Setup button to enable the waveform auto setting function. The oscilloscope will
automatically adjust the horizontal time base, vertical scale, and trigger mode according to the input
signal to obtain an optimum waveform display.
Auto Setup will not work on all signal types, especially time-varying bursts or slow signals ( < 20
Hz ). If Auto Setup cannot achieve desired settings, you can manually adjust the vertical,
horizontal, and trigger systems.
Select the channel with the lowest frequency when several channels are connected to signals.
4.1.4 Self-Calibration
The self-calibration program can quickly calibrate the oscilloscope to reach the best working state
and the most precise measurement. It is recommended to perform a self-calibration if the change of
ambient temperature is more than 5℃. Make sure the oscilloscope has been warmed up or operated
for more than 30 minutes before the self-calibration.
2. Operate Utility > Menu > Maintenance > Self Calibration , and the following dialog
will appear. Select Continue to start the self-calibration program.
The oscilloscope will not respond to any operations until the self-calibration is finished. After the self-
calibration is completed, touch the screen or press any button to exit.
The SDS2000X HD series oscilloscope is designed with four standard interfaces: USB Host, USB
Device, LAN, and Pass/Fail. Connecting to other instruments via these interfaces enables the
oscilloscope to achieve additional capabilities. To ensure the oscilloscope is operating properly, it is
recommended that the user first test the interfaces.
Tools:
Steps:
1. Insert a U-disk ( flash drive ) into the USB Host interface on the front panel of the oscilloscope.
2. A message prompt “ USB flash drive detected .” appears on the screen, and an icon will be
shown in the lower right corner of the screen which confirms the U-disk has been successfully
recognized.
Tools:
Steps:
3. Run NI-MAX software. Click “ Device and Interface ” at the upper left corner of the NI software
interface and immediately the “ USBTMC ” device symbol is displayed.
4. Click “ Open VISA Test Panel ” option button, and then the following Interface will appear. Next,
click the “ Input/Output ” option button and click the “ Query ” option button to view the Read
operation information.
Use to test if the LAN interface operates correctly when connected with NI Visa software.
Tools:
Steps:
1. Set up the National Instruments Measurement and Automaton Explorer ( NI-MAX ) software on
a computer and install the driver using the following instructions.
2. Perform Utility > I/O > LAN Config to open the LAN Config dialog box. And input usable
IP Address and Subnet Mask.
3. Connect the oscilloscope to the computer using a LAN cable via the LAN interface ports.
4. Run NI-MAX software. Click “ Device and Interface ” at the upper left corner of the NI software
interface and select the “ LAN ” device symbol.
5. Click “ Open VISA Test Panel ” option button, which then displays the following interface ( below ).
Then click the “ Input/Output ” option button and click the “ Query ” option button. The “ Read ”
operation information will be displayed.
To test the Mask Test function and signal output by viewing on another oscilloscope.
Tools:
A second oscilloscope
Steps:
2. Enable Channel 1.
3. Perform Analysis > Mask Test to open the Mask Test dialog box, and touch Mask
Setup > Create Mask to set the mask. Set the items as below:
Parameter Setting
Enable to Test On
Source CH1
Type All in
Display Info On
Mask Setting
Mask X 0.32
Mask Y 0.32
Note: Press Create Mask to complete Mask setting, after mask x setting and mask y setting.
After selecting the corresponding items as shown in the table above, the screen will appear as in
Figure 5:
1. Connect the Pass/Fail output signal ( on the rear panel ) to the input terminal of another
oscilloscope.
2. Move the waveform on the screen up, down, or sideways so that any portion of it falls outside of
the Pass / Fail mask. If the interface is operating normally then a pulse will appear on the second
oscilloscope. Adjust the second oscilloscope’s time base and voltage scale as needed. If there
is no viewable pulse on the second oscilloscope then there may be a problem with this function.
This chapter explains testing the oscilloscope to verify performance specifications. For accurate test
results, please let the test instruments and the oscilloscope warm-up for at least 30 minutes before
testing.
DUT 9500B
Note: The 9530 active head in Figure 7 is connected to any channel of the 9500B, and its BNC
connector is connected to the specified channel of the oscilloscope according to the test.
Note: The 9530 active heads in Figure 8 are connected to any two channels of the 9500B, and their
BNC connectors are connected to specified channels of the oscilloscope according to the test.
DUT 9500B
Note: The 9530 active heads in Figure 9 are connected to any two channels of the 9500B. The BNC
connector of one active head is connected to CH1 of the oscilloscope, and the one of another active
head is connected to the external trigger of the oscilloscope.
To verify whether or not a test passes ( i.e. whether the readings are within the appropriate limits ), it
is necessary to record the readings in the Performance Test located in the Test Record.
Self-Calibration
To calculate the DC gain error of a vertical scale, at least nine input and reading values are required
to generate a two-dimensional array.
Use the least square method below to fit the DC Gain error:
Where:
“ LINEST ” is a function in Microsoft EXCEL© that uses the “ least squares ” method to calculate
a line that best fits your dataset. It creates a two-dimensional array by using the least square
method and then returns an array describing the line. The first value of the returned array
represents the slope of the fitting line.
“ Verror ” represents the difference between the voltage setting and the measurement value.
2. Connect a selected channel of the oscilloscope to the 9500B with an active head as shown in
Figure 7.
3. Set the oscilloscope into Non-interleaving mode (both CH1/CH2 and/or both CH3/CH4
activated). Set the timebase of the oscilloscope to 500 us/div, the “ Max Mem Depth ” in the
menu of Acquire to 1M, and the impedance of the selected channel to 1 MΩ.
4. Set the vertical scale of the selected channel to 10 V/div, the position to 0 V. Select the
Measure menu of the oscilloscope to display the “ Mean ” measurement of the selected
channel.
5. Set DC voltage output level of the 9500B according to Table 4, and record the “ Mean ” value
of the channel as Vmean.
6. Calculate Verror ( Verror= Vmean – Vsetting ), and set the next output level as step 5.
7. Calculate gain error with these 9 data points: Vsetting1~Vsetting9, Verror1~Verror9 by using the
‘ LINEST ’ function. Check if the slope falls within the range shown in Table 4.
8. Set the next vertical scale of the selected channel to the other settings in Table 4, and repeat
step 4 to step 7.
9. After measuring all vertical scales of the selected channel in turn, turn the 9500B output to Off.
10. Connect another channel of the oscilloscope to the 9500B with the active head. Check the other
channels in the same way as step 4 to step 9.
11. Set the oscilloscope into Interleaving mode (only one of CH1/CH2 and/or only one of CH3/CH4
activated). Set the impedance of the selected channel to 1MΩ, set the timebase to 100 us/div,
the “ Max Mem Depth ” to 2M, and repeat step 4 to step 10 to measure all channels of the
oscilloscope in turn according to Table 4 which the vertical scale less than or equal to 1 V/div.
When the vertical scale is 1 mV/div, Vsetting is calculated to follow [ -3, -2.25, -1.5,-0.888, 0, 0.888, 1.5,
2.25, 3 ] * 1 mV.
When the vertical scale is within 1 mV to 5 mV, Vsetting is calculated to follow [ -3, -2.25, -1.5, -0.75,
0, 0.75, 1.5, 2.25, 3 ] * vertical scale.
When the vertical scale is more than 5 mV, Vsetting is calculated to follow [ -3.75, -2.8125, -1.875,
-0.9375, 0, 0.9375, 1.875, 2.8125, 3.75 ] * vertical scale.
Two components must be verified for offset accuracy. The first part is the offset gain error relative to
the setting of an offset value ( Gain component ). The second part is the error only relative to the
setting of the vertical scale ( Full-scale component ).
Note: Refer to datasheet for the max offset spec at every vertical scale.
To calculate the offset gain error of a vertical scale ( i.e. Gain component ), at least five input and
reading values are required to generate a two-dimensional array.
Use the least square method below to fit the Offset Gain error:
Where:
“ LINEST ” is a function in Microsoft EXCEL© that uses the “ least squares ” method to calculate
a line that best fits your dataset. It creates a two-dimensional array by using the least square
method and then returns an array describing the line. The first value of the returned array
represents the slope of the fitting line.
Point 3 0 0 0 0 -1%
Steps:
2. Connect a selected channel of the oscilloscope to the 9500B with an active head as shown in
Figure 7.
3. Set the timebase of the oscilloscope to 5 ms/div, the “ Max Mem Depth ” in the menu of
Acquire to 2M, and the impedance of the selected channel to 1 MΩ.
4. Set the vertical scale of the channel to 2 V/div. Select the Measure menu of the oscilloscope
to display the “ Mean ” measurement of the selected channel.
5. According to Table 7, set offset of the channel to Offsetsetting, and set the 9500B output level to
Vsetting = - Offsetsetting, then record the “ Mean ” value of the oscilloscope as Vmean.
6. Calculate Offseterror ( Offseterror= Vmean - Vsetting ), and set the next Offsetsetting as step 5.
7. Calculate offset slope with these 5 data points: Vsetting1~Vsetting5, Offseterror1~Offseterror5 by using
the ‘ LINEST ’ function. Check if the slope falls within the range of slope error shown in Table 7 .
8. Set offset of the channel to 0 V, and turn the 9500B output to Off, then record the “ Mean ” value
of the oscilloscope as Vzero. Check if the Vzero falls within the range of Full-Scale Component. Turn
the 9500B output to On for the next measurement.
9. Set the vertical scale of the channel to the other settings in Table 7, and repeat step 5 to step 7.
11. Connect another channel of the oscilloscope to the 9500B with the active head. Check the other
channels in the same way as step 4 to step 10.
Gain Full-scale
Vertical Scale Offsetsetting Component Component
Error Limit Error Limit(V)
Some products do not list the DC measurement accuracy in the datasheet, but it can be calculated
by combining the DC gain Accuracy and Offset Accuracy. Verifying this specification is not necessary
if the DC gain accuracy and Offset accuracy had been verified.
Use the method below to calculate the High/Low limiting for the DC measurement results:
Upper limit = Vsetting + (DC gain accuracy x | Vsetting + Offsetsetting| + Offset accuracy)
Lower limit = Vsetting - (DC gain accuracy x | Vsetting + Offsetsetting| + Offset accuracy)
Where:
The instance above give the DC measurement limiting with two different input and offset settings.
Here the limiting calculation with no Zero offset is not listed, because the calculation process is
complicated. It is not recommended to verify the device with no zero offset setting ( offset ≠ 0 ).
Steps:
The equipment and device configuration sequence is the same as the Offset Accuracy verification
setting.
This test checks the bandwidth of all the analog channels. In the test, the impedance of 9500B should
be set to be the same as analog channels.
Steps:
1. Connect the selected channel of the oscilloscope to the 9500B via the active head, as shown in
Figure 7.
2. Set the vertical scale of the selected channel to 10 mV/div, impedance to 50 Ω, “ Max Mem Depth ”
to 2M.
3. Press the Measure button on the front panel of the oscilloscope to display “Stdev ”
measurement of the selected channel.
7. Record the “ Stdev ” measurement of the waveform as Vstdev and the first frequency point as
reference value as Vref.
8. Set the vertical scale of the oscilloscope channel under test to the next setting according to Table
10, and repeat step 3 to step 7. The amplitude of the 9500B output is always 6*vertical scale.
After the test, calculate the dB value ( regarding the amplitude at 1 MHz as 0 dB ) at every frequency
point. Check if the value is in the limited range in Table 12 Use the formula below to calculate the dB
value:
Vstdev
dB value = 20 ∗ log10 Vref
Note: 20 Hz is the initial accuracy limit of measuring the 10 MHz input. If the oscilloscope has been
used for a long time ( e.g. > 1 year ), the error caused by aging should be added to this limit. Refer to
the datasheet for the aging spec.
Steps:
1. Connect the selected channel of the oscilloscope to the 9500B, as shown in Figure 7.
5. Set waveform of the 9500B to sine, amplitude to 600 mV, and frequency to 10 MHz.
6. Check if the trigger ( hardware-measured ) frequency which is on the top right of the screen is in
the range of 9.999980 MHz to 10.00002 MHz.
Figure 10
This test checks the trigger level accuracy of the analog channels. In this test, the impedance of both
the 9500B and the oscilloscope should be set to 50 Ω.
Steps:
1. Connect the selected channel of the oscilloscope to the 9500B as shown in Figure 7.
2. Set the impedance of the 9500B to 50 Ω, frequency to 2 kHz, amplitude to 800 mV, and
waveform to sine.
4. Set the impedance of the selected channel, vertical scale to 100 mV/div, trigger source to the
selected channel, trigger slope to positive, and coupling mode to DC.
7. Record the value of “Level@x ”, calculate the level error ( error = 300 mV - Level@x ).
9. Set the trigger level to 0 mV and -300 mV, repeat step 6 to step 8 in turn.
12. Check the other analog channels in the same way as steps 1 to 11.
This test checks the trigger level accuracy of the external trigger channel. In this test, the impedance
of both the 9500B and the oscilloscope should be set to 1 MΩ.
Steps:
1. Connect channel 1 and the external trigger channel of the oscilloscope to the 9500B with two
active heads, as shown in Figure 9.
3. Set channel 1 of the oscilloscope to 150 mV/div, set the impedance of channel 1 and external
trigger to 1 MΩ, trigger source to EXT, trigger slope to positive, and coupling mode to DC.
4. Set the impedance of the 9500B to 1 MΩ, frequency to 1 kHz, amplitude to 8*vertical scale,
and waveform to sine.
7. Record the value of “ Level@x ”, calculate the level error ( error = 450 mV - Level@x ).
9. Set the trigger level to 0 mV and -450 mV, and repeat step 5 to step 8 in turn.
11. Set the oscilloscope vertical scale to 345 mV/div, trigger source to EXT/5, trigger slope to
positive, and coupling mode to DC.
12. Set the impedance of the 9500B to 1 MΩ, frequency to 1 kHz, amplitude to 8* vertical scale,
and waveform to sine.
14. Record the value of Level@x, calculate the level error ( error = 1035 mV - Level@x ).
15. Check if the trigger level is in the limited range in Table 13.
17. Set trigger slope to negative, and repeat step 13 to step 16.
This test checks trigger sensitivity at the frequency of 10 MHz and also at the bandwidth frequency.
Steps:
1. Connect the selected channel of the oscilloscope to the 9500B, as shown in Figure 7.
2. Set the selected channel vertical scale of the oscilloscope to 100 mV/div, timebase to 50 ns/div,
and coupling mode to AC, and impedance to 50 Ω.
3. Set the impedance of the 9500B to 50 Ω, frequency to 10 MHz, waveform to sine, and output
to 50 mV which is measured by the oscilloscope.
5. Press the Trigger Level knob to set the level to the center of the waveform.
6. Adjust the trigger level within the waveform range and observe the ( hardware-measured )
frequency which is on the top right of the screen ( as shown in Figure 10), and check if a level
can be found to satisfy the frequency measured in Table 14.
8. Set frequency of the 9500B to the bandwidth frequency of the oscilloscope, and adjust the
output whose “ amplitude ” measured by the oscilloscope to 50 mV. Set the timebase of the
oscilloscope to 2 ns/div, and then repeat steps 4 to 7.
Steps:
1. Connect the external trigger channel of the oscilloscope to the 9500B, as shown in Figure 7.
2. Set the impedance of the 9500B to 50 Ω, amplitude to 200 mV, frequency to 10 MHz, and
waveform to sine.
3. Set the trigger source of the oscilloscope to EXT, trigger slope to Positive, and impedance of
the external trigger channel to 50 Ω.
4. Press the Trigger Level knob to set the level to the center of the waveform.
5. Adjust the trigger level within the range of [ -100 mV, 100 mV ], observe the ( hardware-
measured ) frequency which is on the top right of the screen ( as shown in Figure 10), and
check if a level can be found to satisfy the frequency measured in Table 14.
7. Set frequency of the 9500B to the 300 MHz ( bandwidth frequency of the external trigger ) and
amplitude to 500 mV. Repeat steps 3 to 6. In step 5, change the trigger level range to
[ -250 mV, 250 mV ].
9. Set the trigger source of the oscilloscope to EXT/5, trigger slope to Positive.
10. Press the Trigger Level knob to set the level to the center of the waveform.
11. Adjust the trigger level within the range of [ -500 mV, 500 mV ], observe the ( hardware-
measured ) frequency which is on the top right of the screen ( as shown in Figure 10), and
check if a level can be found to satisfy the frequency measured in Table 14.
13. Set the frequency of the 9500B to 300 MHz ( bandwidth frequency of the external trigger ),
amplitude to 1.5 V. Repeat steps 9 to 12. In step 11, change the trigger level range to
[ -750 mV, 750 mV ].
This test checks the skew between two analog channels. In this test, the impedance of both the 9500B
and the oscilloscope should be set to 50 Ω.
Steps:
1. Connect two selected channels of the oscilloscope to the 9500B shown as in Figure 8.
2. Press the ‘ Aux ’ key at the right of the ‘ OSCILLOSCOPE CALIBRATOR ’ panel of the 9500B
and select the Skew function.
3. Set the impedance of the 9500B to 50 Ω, frequency to 1 kHz, and amplitude to 0.6 V.
5. Set the oscilloscope timebase to 2 ns/div, trigger source is the selected channel.
6. Press the Measure button on the front panel of the oscilloscope to display the “ Skew ”
measurement between two selected channels.
This test checks the input impedance of all analog channels and the Ext trigger channel with different
coupling modes and vertical scales ( 100 mV/div, 200 mV/div, and 2 V/div ).
DC( AC ) / 50 Ω 49.5~50.5 Ω
1 MΩ 980000~1020000 Ω
Ext
50 Ω 49.5~50.5 Ω
Steps:
1. Connect the selected channel of the oscilloscope to the 9500B, as shown in Figure 7.
2. Set the vertical scale of the selected channel to 100 mV/div, impedance to 1 MΩ, and coupling
mode to AC.
3. Press the ‘ Aux ’ key at the right of the ‘ OSCILLOSCOPE CALIBRATOR ’ panel of the 9500B
and select the Load Resistance Measurement Function. Record the reading displays on the
screen of the 9500B.
4. Set coupling mode of the channel to DC. Record the reading displays on the screen of the
9500B.
6. Test at 2 V/div and 200 mV/div, using the same procedure as steps 2 to 5.
8. Press the ‘ Aux ’ key at the right of the ‘ OSCILLOSCOPE CALIBRATOR ’ panel of the 9500B
and select the Capacitance Measurement Function. Record the reading of the 9500B.
9. Set the vertical scale of the selected channel to 100 mV/div, impedance to 1 MΩ, and coupling
mode to AC.
10. Check if the capacitance is within the specified range in Table 15.
12. Check the other analog channels in the same way as steps 8 to 11.
When checking the EXT trigger channel, it is the same as checking analog channels. Record the
reading displays on the screen of the 9500B, and check to verify it is within the specified range in
Table 15.
This chapter describes how to remove major parts from the SDS2000X HD series oscilloscope.
Only qualified personnel should perform the disassembly procedures. Disconnect the power before
you begin to remove or replace the parts. Otherwise, potential personal injuries or damages to the
components may occur.
Hazardous voltages exist on the power supply and LCD modules. To avoid
electrical shock, disconnect the power cord and wait at least three minutes for
the capacitors to discharge before you begin disassembly.
Multifunction screwdriver
Antistatic gloves
This section describes how to remove the modules in the oscilloscope in detail. To install the removed
parts or replace a new part, follow the instructions in reverse order.
1. Lift the handle and remove the two PM4*8 screws on the rear-cover of the handle. Then push
the two part of handle to different direction to remove the handle.
2. Remove the four PWM3*8 screws from the bottom of each foot, then remove the two feet.
Steps:
Steps:
2. Remove the eight screws from the outside of the rear metal cover.
Steps:
1. Remove the eight PM3*6 screws outside the back panel chassis.
2. Lift the back panel chassis module slowly and unplug the connecting cables from the mainboard.
Steps:
1. Remove the four PM3*6 screws on the power supply and remove the power supply.
2. Remove the four PB5*30 screws bottom of the fan and remove the fan.
3. Remove the three PM3*6 screws on the I/O board remove the I/O board.
Steps:
1. Remove the channel overlay sticker from the front panel and remove the four PM3*6 screws.
4. Lift the front chassis module slowly and unplug the middle connecting cable from the front panel.
Steps:
1. Remove the one PM3*6 screws on the probe interface board, unplug the connecting cable and
remove the probe interface board.
3. Remove the thirteen PM3*6 screws on the mainboard and remove the mainboard.
Steps:
1. Remove each front-panel knob by firmly grasping the knob ( with pliers protected by a soft cloth
to prevent scratches, if necessary ) and pull it away from the front panel.
This section describes how to remove the modules in the oscilloscope in detail. To install the removed
parts or replace a new part, follow the instructions in reverse order.
Check if the power button is illuminated. Orange indicates the oscilloscope is in standby
status and it needs a press to power on. White indicates that the oscilloscope is in the power-
on status.
If the power button is not lit, check whether the power is correctly connected.
Check whether the fuse is burned out. If the fuse needs to be changed, please contact
SIGLENT timely and return the instrument to the factory for replacement by the maintenance
personnel authorized by SIGLENT.
Go to the Troubleshooting chapter and check the power supply modules and acquisition
board.
2. The screen displays only the logo. The instrument is frozen or unresponsive:
Remove all the USB devices plugged into the oscilloscope and power cycle the oscilloscope
to try again.
If not working, try to power cycle the oscilloscope and press the math button on the front
panel multiple times when the oscilloscope is booting up.
If not working, wait ten minutes to see if the oscilloscope can boot successfully. After the
successful boot, try to upgrade to the latest firmware to solve the problem.
Try to recover the system using a U-disk including the recovery package. Contact SIGLENT
for the recovery package and further instructions.
If the display is still stuck at the logo screen, contact SIGLENT for further instructions.
Check whether the probe is correctly connected to the signal connecting wire.
Check whether the signal connecting wire is correctly connected to the BNC ( namely
Check whether there are signals generated from the item to be tested.
Check the Trace option of the associated channel is Visible, not Hidden.
Check the horizontal time/division setting is reasonable and the acquisition time is short
enough for fast display.
Disconnect associated channel input and recall the default setup following the procedures
in the User Manual.
Turn on the associated channel again and press the Auto Setup key.
6. The tested voltage amplitude is greater or lower than the actual value ( Note that this problem
usually occurs when the probe is used ):
Check whether the attenuation coefficient of the channel complies with the attenuation ratio
of the physical probe.
Check whether it is a " false wave ”: When the signal frequency is very large ( more than half
of the sample rate ), it is easy to appear as a " false wave " because of the violation of the
Nyquist theorem. At this point, a smaller time base should be set to make the sample rate is
more than 2 times the signal frequency.
Check the trigger type: General signals should use the “ Edge ” trigger and video signals
should use the “ Video ” trigger. Only when the proper trigger type is used, can the waveform
be displayed stably.
Check the trigger mode in the trigger menu ( TRIGGER ). Set to “ Normal ” or “ Single ” and
ensure that the trigger level is set near the middle of the expected waveform amplitude or
set the mode to “ Auto ”.
Note: Using Auto Setup can automatically finish the above setting.
Check whether the Touch button backlight on the front panel is lighted up. If not, the touch
screen is not turned on. Press the button to enable the touch screen
Check whether the USB storage device is operating correctly on another piece of equipment
or computer.
Make sure that the USB storage device being used is flash storage type. This oscilloscope
does not support hardware storage types.
Restart the instrument and then insert the USB storage device to check it.
If the USB storage device still cannot be used normally, please contact SIGLENT.
This chapter contains information and procedures to troubleshoot general hardware failures.
Only qualified personnel should perform troubleshooting procedures. Disconnect the power cord
whenever possible. DO NOT try to troubleshoot if visible damage is detected on the power supply
module and the acquisition board, return it to SIGLENT for further repair.
The following flowchart describes how to troubleshoot the oscilloscope in the most general case. This
does not guarantee a 100% recovery of all possible hardware failures. Contact SIGLENT if you cannot
solve the problem.
Begin
Test Power
Supply Module
Replace Power
Good? NO
Supply Module
YES
Replace Main
Good? NO
Board
YES
Test LCD
YES
End
Disconnect all the external cables and devices from the front and back panel.
Disconnect the power cord from the oscilloscope, and wait until the power button light is off.
Remove the rear plastic cover and metal cover following the instructions in the chapter
Disassembly Procedures.
Note that the power supply module is on the metal frame. It can be tested while attached to the
frame.
Disconnect the low-voltage side power supply cable from the power supply module labeled CN2.
Reconnect the power cord to the receptacle of the power supply module. Be cautious when
working near the power supply module without the protective cover, high voltage may exist on
the heat sink and other exposed conductors.
If the voltage is slightly out of specification, try to adjust SVR1 on the power supply with a properly
insulated screwdriver.
This guide is based on the acquisition board revision C (labeled SDY8.007.278C). Future revisions
will be compatible unless described differently.
Most of the test points are located on the top side of the board, it is not necessary to completely
remove the board from the chassis to do the troubleshooting unless noted. It is recommended to
re-assemble the power supply module to the metal frame for safety considerations. Follow the
instructions in the chapter Disassembly Procedures.
Reconnect the power cable to the acquisition board after you have confirmed that the power
supply module is in good condition.
In most cases, you can hear the relays click if the acquisition board power up successfully. If you
cannot hear the click or there is no sign that the acquisition board is powering up, go to To Check
the Power Supply Module section to check the standby power and the main power.
If any power rail or the signal under test is abnormal, replace the acquisition board.
Check the board-level power supplies. Find and measure those test points using a multi-meter.
Figure 21 Test points for the Acquisition board power supplies (continued)
Figure 23 Test points for the Acquisition board power supplies (continued)
There are 3 clock oscillators and associated fanout buffers on board to source different circuits. Each
of them serves one or more functions. The voltage compliance of all the clocks under test is
LVCMOS33 or LVCMOS25.
Clock
Test Point Compliance Served Function
Frequency
33.3333 MHz
TP95 LVCMOS33 Reference Clock for the SMP CPU subsystem
± 25 ppm
± 25 ppm
TP102 LVCMOS25 Reference Clock for the SMP FPGA subsystem
Reconnect the power cable to the acquisition board after you have confirmed that the power
supply module is in good condition.
Power up the acquisition board by pushing the front panel power button.
These power rails are shared with the power supply network described in the Check the board-level
power supplies section. If they are within specification without the LCD cable connected, but out of
specification when connected, there may be something wrong with the LCD panel or driver board.
Contact SIGLENT support for further instruction.
Find connector J8, use an oscilloscope to measure the signal using the method below. Be aware that
a 200 MHz oscilloscope may not be suitable for signal quality check, but good enough for signal
presence detection.
If the signal is stuck, double-check with the LCD cable disconnected. If the signal is still stuck,
replace the acquisition board. If not, replace the LCD panel or the LCDI board.
Headquarters:
SIGLENT Technologies Co., Ltd
Add: Bldg No.4 & No.5, Antongda Industrial
Zone, 3rd Liuxian Road, Bao'an District,
Shenzhen, 518101, China
Tel: + 86 755 3688 7876
Fax: + 86 755 3359 1582
Email: [email protected]
Website: int.siglent.com
North America:
SIGLENT Technologies America, Inc
6557 Cochran Rd Solon, Ohio 44139
Tel: 440-398-5800
Toll Free: 877-515-5551
Fax: 440-399-1211
Email: [email protected]
Website: www.siglentna.com
Follow us on
Facebook: SiglentTech
Europe:
SIGLENT Technologies Germany GmbH
Add: Staetzlinger Str. 70