operations-instructions-36900
operations-instructions-36900
SITRANS RD100
SITRANS
Safety Guidelines: Warning notices must be observed to ensure personal safety as well as that of
others, and to protect the product and the connected equipment. These warning notices are
accompanied by a clarification of the level of caution to be observed.
Qualified Personnel: This device/system may only be set up and operated in conjunction with this
manual. Qualified personnel are only authorized to install and operate this equipment in accordance with
established safety practices and standards.
Unit Repair and Excluded Liability:
• The user is responsible for all changes and repairs made to the device by the user or the user’s
agent.
• All new components are to be provided by Siemens Milltronics Process Instruments.
• Restrict repair to faulty components only.
• Do not reuse faulty components.
Warning: Cardboard shipping package provides limited humidity and moisture protection. This product
can only function properly and safely if it is correctly transported, stored, installed, set up, operated, and
maintained.
This product is intended for use in industrial areas. Operation of this equipment in a residential area
may cause interference to several frequency based communications.
© Siemens AG 2011
SITRANS RD100 Loop Powered Meter
SITRANS RD100
SITRANS RD100 is a 2-wire loop powered, NEMA 4X enclosed remote digital display for
process instrumentation.
This digital meter is easy to use with a display of 3.5 digits, 1" (2.54 cm) high. It accepts 4 to
20 mA input and operates from -40 to +80 °C (-40 to +176 °F).
SITRANS RD100 is CSA and FM Approved.
Safety Notes
Special attention must be paid to warnings and notes highlighted from the rest of the text
by grey boxes.
Note: means important information about the product or that part of the operating
manual.
The Manual
This manual provides specifications and instructions for the operation of SITRANS RD100
Remote Display. The manual is designed to help you get the most out of your Remote
Display, and it provides information on the following:
• Meter specifications
• Wiring diagrams
• Installation requirements
• Setup instructions
If you have any questions, comments, or suggestions about the manual content, please
email us at [email protected]. For the complete library of Siemens Milltronics
manuals, go to www.siemens.com/processautomation.
Display
• 1.0 " (25.4 mm) high LCD, numeric display range from -1000 to 1999
Enclosure
• Impact-resistant glass-filled polycarbonate body
• Color: gray
• Impact-resistant clear polycarbonate cover: NEMA 4X, Type 4X, IP67
Mounting
• Standard:
• wall mount
• Optional:
• panel mounting kit (does not provide NEMA 4X seal to panel)
• 2" pipe mounting kit
Entity Parameters
• Vmax= 30 V, Imax = 175 mA, Ci = 0 µF, Li = 0 µH, Pi = 1.3 W
Input
• 4 to 20 mA
Connections
• Screw terminal block
Calibration Range
• 4 mA input: display of -1000 to +1000 counts
• 20 mA input: display of 4 mA count value +20 to 2000 counts, to a max. display of
1999 counts
Accuracy
• ±0.1% of span ±1 count
Conversion Rate
• 2.5 conversions per second
Approvals
• Intrinsically Safe:
• FM/CSA Class I, II, III, Division 1, Groups A to G T4
• FM/CSA Class I, Zone 0, Group IIC
• Non-incendive:
• FM/CSA Class I, Division 2, Groups A to D
• FM/CSA Class II and III, Division 2, Groups F and G
See "RD100 Remote Display - Approval Connections Control Drawing" shipped with the
unit and located on the Instruction Manual CD, for complete Intrinsically Safe installation
instructions.
Wall Mounting
wall mounting
holes beneath
D cover screws E
F
A C
Dimensions
A: 80 mm (3.15") C: 60 mm (2.36") E: 65 mm (2.56")
B: 140 mm (5.51") D: 120 mm (4.72") F: 20 mm (0.79")
Wall mounting holes: Ø 4 mm (0.16") Weight: 340 g (12 oz)
washers
panel
mounting plate
RD100
Notes
• The optional panel mounting kit does not provide a NEMA 4X seal to panel.
• Mounting brackets require 20 mm (0.8") clearance on either the top or the side
of the meter for installation.
Instructions
1. Insert the RD100 through the mounting plate.
2. Mount the complete assembly on the panel.
3. Secure with 4 brackets and 4 screws (supplied) at the corners of the meter. Use
washers as needed to allow for panel thickness.
LO calibration control
DP1
HI calibration control
DP2
balance control DP3
(factory adjust only)
S- S+
black
red
S+ S-
S- S+
loop jumper
DP1
DP2
DP3
S+ S-
calibrated
current source
The display PCB may be removed from the enclosure for bench calibration.
Loop jumper must be installed on input signal PCB to maintain loop. Refer to
Servicing display PCB outside the loop on page 9.
WARNINGS:
• To maintain hazardous area protection, the input signal must
always be connected to the input signal PCB, and not directly to
the display PCB.
• Electrostatic hazard. Clean only with a moist cloth. Protect
enclosure from exposure to chemical solvents and excessive
ultraviolet (UV) light (such as sunlight).
Calibration connections
To access the input terminals, remove the enclosure cover and the display PCB.
1. Loosen the four screws on the enclosure cover and remove the cover.
2. Completely loosen the left screw that holds the display PCB to the enclosure and
loosen the right screw about four turns so the display PCB remains attached to the
enclosure.
3. Rotate the display PCB 90° to gain access to the input signal PCB.
4. Next, connect a calibrated current source as shown in Calibrator connected to input
signal PCB on page 6.
Calibration
LO and HI calibration controls are located to the left of the display (see Calibrator
connected to input signal PCB on page 6).
1. Apply a signal equal to 4 mA and adjust the LO control to display the desired reading.
2. Apply a signal between 16 and 20 mA and adjust the HI control to display the desired
reading.
3. Complete the calibration procedure by making any minor adjustments to the LO and
HI controls.
Installation
To install the meter, remove the display PCB from its enclosure and connect a ½" conduit
fitting to the hole provided. Refer to Removing display PCB from the loop on page 9 for
Loop connections
Disconnect power to the loop and install the meter as illustrated in Control loop
connected to input signal PCB on page 6 and Removing display PCB from the loop on
page 9. Replace the enclosure cover.
www.siemens.com/weighing