Tobin Design
Tobin Design
PHYSICAL BASIS
OF MECHANICAL
VENTILATION
This page intentionally left blank
CLASSIFICATION OF 2
MECHANICAL VENTILATORS
AND MODES OF VENTILATION
Robert L. Chatburn
A good ventilator classification scheme describes how venti- chapter; these topics have been treated elsewhere.7,8 The
lators work in general terms, but with enough detail so that chapter does, however, explore in detail control schemes
one particular model can be distinguished from others. It and ventilator modes because these directly affect patient
facilitates description by focusing on key attributes in a logi- management.
cal and consistent manner. A clear description allows us to
quickly assess new facts in relation to our previous knowl-
edge. Learning the operation of a new ventilator or describ- CONTROL SYSTEM
ing it to others then becomes much easier. Understanding
how the ventilator operates, we can then anticipate appropri- Models of Patient–Ventilator Interaction
ate ventilator management strategies for particular clinical
situations. The classification system described in this chapter To understand how a machine can be controlled to replace
is based on previously published work.1–7 or supplement the natural function of breathing, we need to
A ventilator is simply a machine, a system of related ele- first understand something about the mechanics of breath-
ments designed to alter, transmit, and direct energy in a pre- ing itself. The study of mechanics deals with forces, displace-
determined manner to perform useful work. We put energy ments, and the rate of change of displacement. In physiology,
into the ventilator in the form of electricity (energy = volts force is measured as pressure (pressure = force/area), dis-
× amps × time) or compressed gas (energy = pressure × placement as volume (volume = area × displacement), and
volume). That energy is transmitted or transformed (by the the relevant rate of change as flow [average flow = Δvolume/
ventilator’s drive mechanism) in a predetermined manner Δtime; instantaneous flow (Vɺ ) = dv /dt , the derivative of vol-
(by the control circuit) to augment or replace the patient’s ume with respect to time]. Specifically, we are interested in
muscles in performing the work of breathing. Thus to the pressure necessary to cause a flow of gas to enter the air-
understand mechanical ventilators in general, we must first way and increase the volume of the lungs.
understand their basic functions: (a) power input, (b) power The study of respiratory mechanics is essentially the
transmission or conversion, (c) control scheme, and (d) out- search for simple but useful models of respiratory system
put. This simple format can be expanded to add as much mechanical behavior. Figure 2-1 illustrates the process by
detail as desired (Table 2-1). which the respiratory system is represented first by a graphi-
A discussion of input power sources and power con- cal model, and then by a mathematical model based on the
version and transmission is beyond the scope of this graphical model. Pressure, volume, and flow are measurable
45
46 Part II Physical Basis of Mechanical Ventilation
Control Variables
variables in the mathematical model that change with time In the equation of motion, the mathematical form of any of
over the course of one inspiration and expiration. The rela- the three variables (i.e., pressure, volume, or flow as func-
tion among them is described by the equation of motion for tions of time) can be predetermined, making it the inde-
the respiratory system.9 The derivation of this equation stems pendent variable and making the other two the dependent
from a force-balance equation that is an expression of New- variables. We now have a theoretical basis for classifying ven-
ton’s third law of motion (for every action, there is an equal tilators as pressure, volume, or flow controllers. Thus, during
and opposite reaction): pressure-controlled ventilation, pressure is the independent
variable and may take the form of, say, a step function (i.e.,
PTR = PE + PR (1)
a rectangular pressure waveform). The shapes of the vol-
where PTR is the transrespiratory pressure (i.e., pressure ume and flow waveforms for a passive respiratory system
at the airway opening minus pressure at the body surface), (Pmus = 0) then depends on the shape of the pressure wave-
PE is the pressure caused by elastic recoil (elastic load), and form as well as the parameters of resistance and compliance.
PR is the pressure caused by flow resistance (resistive load). On the other hand, during volume-controlled ventilation, we
Transrespiratory pressure can have two components, can specify the shape of the volume waveform making flow-
one generated by the ventilator (P vent) and one generated by dependent and pressure-dependent variables. The same rea-
the respiratory muscles (Pmus). Elastic recoil pressure is the soning applies to a flow controller. Notable exceptions are
product of elastance (E = Δpressure/Δvolume) and volume. interpulmonary percussive ventilation, and high-frequency
Chapter 2 Classification of Mechanical Ventilators and Modes of Ventilation 47
Flow
Transairway
pressure
Transrespiratory
pressure
Transthoracic
pressure
Volume
FIGURE 2-1 The respiratory system is often modeled as a single flow resistance (representing the endotracheal tube and the airways) connected
to an elastic chamber (representing the lungs and chest wall). Flow through the airways is generated by transairway pressure (pressure at the airway
opening minus pressure in the lungs). Expansion of the elastic chamber is generated by transthoracic pressure (pressure in the lungs minus pressure
on the body surface). Transrespiratory pressure (pressure at the airway opening minus pressure on the body surface) is the sum of these two pressures
and is the total pressure required to generate inspiration. The “airway-pressure” gauge on a positive-pressure ventilator displays transrespiratory
pressure.
oscillatory ventilation, both of which control only the dura- A particular variable is measured and used to start, sustain,
tion of flow pulses; the resulting airway pressure pulses and end each phase. In this context, pressure, volume, flow,
along with actual inspiratory flows and volumes depend on and time are referred to as phase variables.11 Figure 2-2 shows
the instantaneous values of respiratory system impedance. the criteria for determining phase variables.
Because neither pressure, volume, nor flow in the equation
of motion are predetermined, we would classify this type of
device as a “time controller.” Trigger Variable
It follows from the preceding discussion that any conceiv-
able ventilator can control only one variable at a time: pres- All ventilators measure one or more variables associated with
sure, volume, or flow. Because volume and flow are inverse the equation of motion (i.e., pressure, volume, flow, or time).
functions of one another, we can simplify our discussion Inspiration is started when one of these variables reaches a
and consider only pressure and volume as control variables. preset value. Thus, the variable of interest is considered an
I discuss later in “Modes of Ventilation” exactly how ventila- initiating, or trigger, variable. Time is a trigger variable when
tor control systems work. We will see that it is possible for the ventilator starts a breath according to a set frequency
a ventilator to switch quickly from one control variable to independent of the patient’s spontaneous efforts. Pressure is
another, not only from breath to breath, but even during a the trigger variable when the ventilator senses a drop in base-
single inspiration. line pressure caused by the patient’s inspiratory effort and
begins a breath independent of the set frequency. Flow or
volume are the trigger variables when the ventilator senses
Phase Variables the patient’s inspiratory effort in the form of either flow of
volume into the lungs.
Because breathing is a periodic event, the ventilator must Flow triggering reduces the work the patient must per-
be able to control a number of variables during the respira- form to start inspiration.12 This is so because work is pro-
tory cycle (i.e., the time from the beginning of one breath portional to the volume the patient inspires times the
to the beginning of the next). Mushin et al10 proposed that change in baseline pressure necessary to trigger. Pressure
this time span be divided into four phases: the change from triggering requires some pressure change and hence an
expiration to inspiration, inspiration, the change from inspi- irreducible amount of work to trigger. With flow or volume
ration to expiration, and expiration. This convention is triggering, however, baseline pressure need not change, and
useful for examining how a ventilator starts, sustains, and theoretically, the patient need do no work on the ventilator
stops an inspiration and what it does between inspirations. to trigger.
48 Part II Physical Basis of Mechanical Ventilation
Does peak pressure Does peak volume Does peak flow reach No variables are
reach a preset value No reach a preset value No a preset value before No targeted during
before inspiration before inspiration inspiration ends? inspiration.
ends? ends?
Does expiration start Does expiration start Does expiration start Expiration starts
because a preset No because a preset No because a preset No because a preset
pressure is met? volume is met? flow is met? time is met.
FIGURE 2-2 Criteria for determining the phase variables during a ventilator-assisted breath.
The patient effort required to trigger inspiration is Organization’s use of the term limit as applying to alarm situ-
determined by the ventilator’s sensitivity setting. Some ven- ations only.
tilators indicate sensitivity qualitatively (“min” or “max”). Clinicians often confuse target variables with cycle vari-
Alternatively, a ventilator may specify a trigger threshold ables. To cycle means “to end inspiration.” A cycle variable
quantitatively (e.g., 5 cm H2O below baseline). Once the always ends inspiration. A target variable does not terminate
trigger variable signals the start of inspiration, there is inspiration; it only sets an upper bound for pressure, volume,
always a short delay before flow to the patient starts. This or flow (Fig. 2-3).
delay is called the response time and is secondary to the
signal-processing time and the mechanical inertia of the
drive mechanisms. It is important for the ventilator to have Cycle Variable
a short response time to maintain optimal synchrony with
patient inspiratory effort. The inspiratory phase always ends when some variable
reaches a preset value. The variable that is measured and
used to end inspiration is called the cycle variable. The cycle
Target Variable variable can be pressure, volume, flow, or time. Manual
cycling is also available on some ventilators.
Here target means restricting the magnitude of a variable When a ventilator is set to pressure cycle, it delivers flow
during inspiration. A target variable is one that can reach and until a preset pressure is reached, at which time inspiratory
maintain a preset level before inspiration ends (i.e., it does flow stops and expiratory flow begins. The most common
not end inspiration). Pressure, flow, or volume can be tar- application of pressure cycling on mechanical ventilators is
get variables and actually all can be active for a single breath for alarm settings.
(e.g., using the Pmax feature on a Dräger ventilator). Note that When a ventilator is set to volume cycle, it delivers flow
time cannot be a target variable because specifying an inspi- until a preset volume has passed through the control valve.
ratory time would cause inspiration to end, violating the pre- By definition, as soon as the set volume is met, inspiratory
ceding definition. Astute readers may notice that in the past flow stops and expiratory flow begins. If expiration does not
I have used the term limit where here I have used target. This begin immediately after inspiratory flow stops, then an inspi-
was done to be consistent with the International Standards ratory hold has been set, and the ventilator is, by definition,
Chapter 2 Classification of Mechanical Ventilators and Modes of Ventilation 49
A B C
Ventilator
pressure
Volume
Flow
FIGURE 2-3 This figure illustrates the distinction between the terms target and cycle. A. Inspiration is pressure-targeted and time-cycled. B. Flow is
targeted, but volume is not, and inspiration is volume-cycled. C. Both volume and flow are targeted, and inspiration is time-cycled. (Reproduced, with
permission, from Chatburn.6)
time cycled (see Fig. 2-3). Note that the volume that passes Time cycling means that expiratory flow starts because a
through the ventilator’s output control valve is never exactly preset inspiratory time interval has elapsed.
equal to the volume delivered to the patient because of the
volume compressed in the patient circuit. Some ventilators Baseline Variable
use a sensor at the Y-connector (such as the Dräger Evita
4 with the neonatal circuit) for more accurate tidal volume The baseline variable is the parameter controlled during expi-
measurement. Others measure volume at some point inside ration. Although pressure, volume, or flow could serve as the
the ventilator, and the operator must know whether the ven- baseline variable, pressure control is the most practical and
tilator compensates for compressed gas in its tidal volume is implemented by all modern ventilators. Baseline or expi-
readout. ratory pressure is always measured and set relative to atmo-
When a ventilator is set to flow cycle, it delivers flow spheric pressure. Thus, when we want baseline pressure to
until a preset level is met. Flow then stops, and expiration equal atmospheric pressure, we set it to zero. When we want
begins. The most frequent application of flow cycling is in baseline pressure to exceed atmospheric pressure, we set a
the pressure-support mode. In this mode, the control vari- positive value, called positive end-expiratory pressure (PEEP).
able is pressure, and the ventilator provides the flow neces-
sary to meet the inspiratory pressure target. In doing so, flow
starts out at a relatively high value and decays exponentially MODES OF VENTILATION
(assuming that the patient’s respiratory muscles are inactive
after triggering). Once flow has decreased to a relatively low The general goals of mechanical ventilation are to pro-
value (such as 25% of peak flow, typically preset by the man- mote safety, comfort, and liberation (Table 2-2).1 Specific
ufacturer), inspiration is cycled off. Manufacturers often set objectives under these goals include ensuring adequate gas
the cycle threshold slightly above zero flow to prevent inspi- exchange, avoiding ventilator induced lung injury, optimiz-
ratory times from getting so long that patient synchrony is ing patient-ventilator synchrony, and minimizing the dura-
degraded. On some ventilators, the flow-cycle threshold may tion of ventilation. The preset pattern of patient-ventilator
be adjusted by the operator to improve patient synchrony. interaction designed to achieve these objectives is referred to
Increasing the flow-cycle threshold decreases inspiratory as a mode of ventilation. Specifically, a mode can be classified
time and vice versa. according to the outline in Table 2-3.2
50 Part II Physical Basis of Mechanical Ventilation
even if the patient is passive at end-inspiration, the patient’s controlled mandatory breaths on infant ventilators. The
resistance and compliance determine the cycle point and thus key distinction between CMV and IMV is that with CMV,
the size of the breath for a given pressure-support setting. In the ventilator attempts to deliver a mandatory breath every
contrast, for a patient on continuous positive airway pressure, time the patient makes an inspiratory effort (unless a man-
each breath is spontaneous but unassisted. Breaths are spon- datory breath is already in progress). This means that
taneous because the patient determines the timing and size of during CMV, if the operator decreases the ventilator rate,
the breaths without any interference by the ventilator. Breaths the level of ventilator support is unaffected as long as the
during continuous positive airway pressure are not assisted patient continues making inspiratory efforts. With IMV,
because airway pressure is controlled by the ventilator to be the rate setting directly affects the number of mandatory
as constant as possible (i.e., Pvent = 0). Understanding the dif- breaths and hence the level of ventilator support. Thus,
ference between assisted and unassisted spontaneous breaths CMV is normally viewed as a method of “full ventilator
is very important clinically. When making measurements of support,” whereas IMV is usually viewed as a method of
tidal volume and respiratory rate for calculation of the rapid- partial ventilator support. Of course, actual “full ventila-
shallow breathing index, for example, the breaths must be tory support” can only be achieved if the patient is making
spontaneous and unassisted. If they are assisted (e.g., with no inspiratory efforts, for example, is paralyzed, but the
pressure support), an error of 25% to 50% may be introduced. term is often used loosely to mean supplying as much sup-
A mandatory breath is any breath that does not meet the port as possible for a given patient condition.
criteria of a spontaneous breath, meaning that the patient
Given the two ways to control inspiration (i.e., pressure
has lost control over the timing and/or size. Thus, a manda-
and volume) and the three breath sequences (i.e., CMV, IMV,
tory breath is one for which the start or end of inspiration
or CSV), there are five possible breathing patterns; volume
(or both) is determined by the ventilator, independent of the
control (VC)-CMV, VC-IMV, pressure control (PC)-CMV,
patient; that is, the machine triggers and/or cycles the breath.
PC-IMV, PC-CSV (see Table 2-2). VC-CSV is not possible
It is possible to superimpose a short mandatory breath on
because volume control implies that inspiration ends after a
top of a longer spontaneous breath, as in the case of high-
preset tidal volume is delivered, hence violating the patient
frequency oscillatory ventilation.
cycling criterion of a spontaneous breath.
Having defined spontaneous and mandatory breaths, there
are three possible breath sequences, designated as follows:
• Continuous spontaneous ventilation (CSV). All breaths are Targeting Schemes
spontaneous.
• Intermittent mandatory ventilation (IMV). Spontane- Targeting schemes are feedback control systems used by
ous breaths are permitted between mandatory breaths. mechanical ventilators to deliver specific ventilatory pat-
When the mandatory breath is triggered by the patient, terns.1 The targeting scheme is a key component of a mode
it is commonly referred to as synchronized IMV. Because classification system. Before we can describe specific target-
the trigger variable can be specified in the description of ing schemes used by ventilators, we must first appreciate the
phase variables, I will use IMV instead of synchronized basic concepts of engineering control theory.
IMV to designate general breath sequences. The term closed-loop control refers to the use of a feed-
• Continuous mandatory ventilation (CMV). Spontaneous back signal to adjust the output of a system. Ventilators use
breaths are not permitted between mandatory breaths, as closed-loop control to maintain consistent pressure and flow
the intent is to provide a mandatory breath for every patient waveforms in the face of changing patient/system condi-
inspiratory effort. CMV originally meant that every breath tions. This is accomplished by using the output as a feedback
was mandatory. The development of the “active exhalation signal that is compared to the operator-set input. The differ-
valve,” however, made it possible for the patient to breathe ence between the two is used to drive the system toward the
spontaneously during a mandatory pressure-controlled desired output. For example, pressure-control modes use
breath on some ventilators. In fact, it was always possible airway pressure as the feedback signal to control gas flow
for the patient to breathe spontaneously during pressure- from the ventilator. Figure 2-4 is a schematic of a general
Disturbances
Feedback signal
FIGURE 2-4 Generalized control circuit (see text for explanation). The “plant” in a control circuit for mechanical ventilation is the patient.
(Reproduced with permission from Chatburn RL. Mireles-Cabodevila E, Closed loop control of mechanical ventilation. Respir Care. 2011;56(1):
85–98.)
52 Part II Physical Basis of Mechanical Ventilation
control system. The input is a reference value (e.g., operator (relationship between the input and the output of the con-
preset inspiratory pressure) that is compared to the actual troller) as a targeting scheme. The history of these schemes
output value (e.g., instantaneous value of airway pressure). clearly shows an evolutionary trend toward increasing lev-
The difference between those two values is the error signal. els of automation. In fact, we can identify three groups of
The error signal is passed to the controller (e.g., the software targeting schemes based on increasing levels of autonomy:
control algorithm). The controller converts the error signal manual, servo, and automatic. Manual targeting schemes
into a signal that can drive the effector (e.g., the hardware) require the operator to adjust all the target values. Servo
to cause a change in the manipulated variable (e.g., inspira- targeting schemes are unique in that there are no static
tory flow). The relationship between the input and the out- target values; rather, the operator sets the parameters of a
put of the controller is called the transfer function in control mathematical model that drives the ventilator’s output to
theory. Engineers need to understand the transfer function follow a dynamic signal (like power steering on an auto-
in terms of complex mathematical equations. Clinicians, mobile). Automatic targeting schemes enable the ventila-
however, need only understand the general operation of tor to set some or all of the ventilatory targets, using either
the function in terms of how the mode affects the patient’s mathematical models of physiologic processes or artificial-
ventilatory pattern, and we will use that frame of reference intelligence algorithms.
in defining targeting schemes. The “plant” in Figure 2-4 The basic concept of closed-loop control has evolved into
refers to the process under control. In our case, the plant is at least six different ventilator targeting schemes (set-point,
the patient and the delivery circuit connecting the patient dual, servo, adaptive, optimal, and intelligent). These tar-
to the ventilator. The plant is the source of the “noise” that geting schemes are the foundation that makes possible sev-
causes problems with patient–ventilator synchrony. At one eral dozen apparently different modes of ventilation. Once
extreme, a paralyzed patient and an intact delivery circuit we understand how these control types work, many of the
pose little challenge for a modern ventilator to deliver a apparent differences are seen to be similarities. We then
predetermined ventilatory pattern, and thus synchrony is avoid a lot of the confusion surrounding ventilator market-
not an issue. At the opposite extreme is a patient with an ing hype and begin to appreciate the true clinical capabilities
intense, erratic respiratory drive and a delivery circuit with of different ventilators.
leaks (e.g., around an uncuffed endotracheal tube) making
patient–ventilator synchrony virtually impossible. The chal- SET-POINT
lenge for both clinicians and engineers is to develop tech-
nology and procedures for dealing with this wide range of In set-point targeting, the operator sets specific target values
circumstances. and the ventilator attempts to deliver them (Fig. 2-5). The
The plant alters the manipulated variable to generate the simplest examples for volume-control modes are tidal vol-
feedback signal of interest as the control (output) variable. ume and inspiratory flow. For pressure-control modes, the
Continuing with the example above, the manipulated vari- operator may set inspiratory pressure and inspiratory time
able is flow, but the feedback control variable is pressure (i.e., or cycle threshold.
ventilator flow times plant impedance equals airway pres-
sure), as in pressure-control modes. DUAL
Closed-loop control can also refer to the use of feedback
As it relates to mechanical ventilation, volume control means
signals to control the overall pattern of ventilation, beyond
that inspired volume, as a function of time, is predetermined
a single breath, such as the use of end-tidal carbon dioxide
by the operator before the breath begins. In contrast, pres-
tension as a feedback signal to control minute ventilation. sure control means that inspiratory pressure as a function of
The process of “setting” or adjusting a ventilation mode
time is predetermined. “Predetermined” in this sense means
can be thought of as presetting various target values, such as
that either pressure or volume is constrained to a specific
tidal volume, inspiratory flow, inspiratory pressure, inspira-
mathematical form. In the simple case where either pressure
tory time, frequency, PEEP, oxygen concentration, and end-
or flow are preset constant values (e.g., set-point targeting, as
tidal carbon dioxide concentration. The term target is used explained above), we can say that they are the independent
for two reasons. First, just like in archery, a target is aimed
at but not necessarily hit, depending on the precision of the
control system. An example is setting a target value for tidal
Disturbances
volume and allowing the ventilator to adjust the inspiratory
pressure over several breaths to finally deliver the desired Set-point
Operator
value. In this case, we could more accurately talk about deliv- Pressure Ventilator Patient
Volume
ering an average target tidal volume over time. Flow
The second reason for using target is because the term Flow or volume
control is overused and we need it to preserve some funda- Pressure
mental conventions regarding modes such as volume control FIGURE 2-5 Set-point targeting. (Reproduced, with permission, from
versus pressure control. From this use of the term target, we Chatburn RL. Computer control of mechanical ventilation. Respir Care.
can logically refer to the control system transfer function 2004;49:507–515.)
Chapter 2 Classification of Mechanical Ventilators and Modes of Ventilation 53
variables in the equation of motion. The equation of motion provides the safety of a guaranteed minimum tidal volume
for the respiratory system is a general mathematical model of with the patient comfort of flow synchrony provided by pres-
patient–ventilator interaction: sure control.
P(t ) = EV (t ) + RVɺ (t ) (4)
SERVO
where P(t) is inspiratory pressure as a function of time (t), E The term servo was coined by Joseph Farcot in 1873 to
is respiratory-system elastance, V(t) is volume as a function describe steam-powered steering systems. Later, hydraulic
of time, R is respiratory-system resistance, and Vɺ is flow as a “servos” were used to position antiaircraft guns on warships.
function of time. Thus, for example, if pressure is the inde- Servo control specifically refers to a control system that con-
pendent variable, then both volume and flow are dependent verts a small mechanical motion into one requiring much
variables, indicating pressure control. If volume is the inde- greater power, using a feedback mechanism. As such, it
pendent variable, then pressure is the dependent variable, offers a substantial advantage in terms of creating ventilation
indicating volume control. Because volume is the integral of modes capable of a high degree of synchrony with patient
flow, if Vɺ is predetermined, then so is V(t). Therefore, for breathing efforts. That is, ventilator work output can be
simplicity, we include the case of flow being the independent made to match patient work demand with a high degree of
variable as a form of volume control. fidelity. We apply the name servo control to targeting schemes
Only one variable (i.e., pressure or volume) can be inde- in which the ventilator’s output automatically follows a vary-
pendent at any moment, but a ventilator controller can ing input. This includes proportional-assist ventilation (PAV;
switch between the two during a single inspiration. When Fig. 2-6),14 automatic tube compensation (ATC),15 and neu-
this happens, the targeting scheme is called dual set-point rally adjusted ventilatory assist (NAVA),16 in which the air-
control or dual targeting. There are two basic ways that way pressure signal not only follows but amplifies signals
ventilators have implemented dual targeting. One way is to that are surrogates for patient effort (i.e., volume, flow, and
start inspiration in volume control and then switch to pres- diaphragmatic electrical signals). Note that the term servo
sure control if one or more preset thresholds are met (e.g., a control has been loosely used since it was coined to refer to
desired peak airway pressure target). An example of such a any type of general feedback control mechanism, but I am
threshold is the operator-set Pmax in volume control on the using it in a very specific manner, as it applies to ventilator
Dräger Evita XL ventilator. The other form of dual targeting targeting schemes.
is to start inspiration in pressure control and then switch to
volume control (e.g., if a preset tidal volume has not been
met when flow decays to a preset value). This was originally
described as “volume-assured pressure-support ventila- Disturbances
tion,”13 but is currently only available as a mode called “Vol- Set-point
Operator
ume Control Assist Control with Machine Volume” in the Elastic load Ventilator Patient
CareFusion Avea ventilator.
Resistive load
Dual targeting is an attempt to improve the synchrony Pressure, volume, and flow
between patient and ventilator. This can be seen in the equa-
tion of motion if a term representing the patient inspiratory
force (muscle pressure or Pmus) is added: Pmus = Loadnormal + Loaddisease
Exhaled volume
weight
Ventilator Patient
Frequency
Pressure
Volume
Flow Disturbances
Pressure
FIGURE 2-7 Adaptive targeting. Notice that the operator has stepped Ventilator Patient
back from direct control of the within-breath parameters of pressure
and flow. Examples of adaptive targeting are pressure-regulated volume Flow
control (PRVC) on the Siemens ventilator and autoflow on the Dräger
Evita 4 ventilator. (Reproduced, with permission, from Chatburn RL. Pressure
Computer control of mechanical ventilation. Respir Care. 2004;49: FIGURE 2-8 Optimal targeting. A static mathematical model is used
507–515.) to optimize some performance parameter, such as work of breath-
ing. The only commercially available form of optimal targeting is the
adaptive-support ventilation (ASV) mode on the Hamilton Galileo ven-
tilator. (Reproduced, with permission, from Chatburn RL. Computer
ADAPTIVE control of mechanical ventilation. Respir Care. 2004;49:507–515.)
An adaptive targeting scheme involves modifying the func-
tion of the controller to cope with the fact that the system
parameters being controlled are time varying. As it applies breathing, mimic natural breathing, stimulate spontaneous
to mechanical ventilation, adaptive targeting schemes allow breathing, and reduce weaning time.20 The operator inputs
the ventilator to set some (or conceivably all) of the targets the patient’s weight. From that, the ventilator estimates the
in response to varying patient conditions. Modern inten- required minute alveolar ventilation, assuming a normal
sive care unit ventilators may use adaptive flow targeting as dead space fraction. Next, an optimum frequency is calcu-
a more accurate way to deliver volume control modes than lated based on work by Otis et al21 that predicts a frequency
set-point targeting. For example, the Covidien PB 840 venti- resulting in the least mechanical work rate:20
lator automatically adjusts inspiratory flow between breaths
MV − fVD
to compensate for volume compression in the patient circuit −1 + 1 + 4π 2 RC E
and thus achieving an average target tidal volume equal to VD
the operator-set value.17 Aside from this application of adap- f = (6)
2π 2 RC E
tive targeting, there are four distinct approaches to basic
adaptive targeting, which are represented by the mode names where MV is predicted minute ventilation (L/min) based on
pressure-regulated volume control (inspiratory pressure auto- patient weight and the setting for percent of predicted MV
matically adjusted to achieve an average tidal volume target, to support, VD is predicted dead space (L) based on patient
Fig. 2-7), mandatory rate ventilation (inspiratory pressure weight, RCE is the expiratory time constant calculated as the
automatically adjusted to maintain a target spontaneous slope of the expiratory flow volume curve and f is the com-
breath frequency), adaptive flow/adaptive I-time (inspiratory puted optimal frequency (breaths/min). The target tidal vol-
time and flow automatically adjusted to maintain a constant ume is calculated as MV/f. The ASV controller uses the Otis
inspiratory time-to-expiratory time ratio of 1:2), and manda- equation to set the tidal volume (Fig. 2-8). As with simple
tory minute ventilation (automatic adjustment of mandatory adaptive pressure targeting, the inspiratory pressure within a
breath frequency to maintain a target minute ventilation). breath is controlled to achieve a constant value and between
breaths the inspiratory pressure is adjusted to achieve a tar-
get tidal volume. Unlike simple adaptive pressure targeting,
OPTIMAL
however, the target is not set by the operator; instead, it is
Optimal targeting is an advanced form of adaptive target- estimated by the ventilator in response to changes in respi-
ing.18 Optimal targeting in this context means that the ven- ratory-system mechanics and patient effort. Individual pres-
tilator controller automatically adjusts the targets of the sure-targeted breaths may be mandatory (time triggered and
ventilatory pattern to either minimize or maximize some time cycled) or spontaneous (flow triggered and flow cycle).
overall performance characteristic (Fig. 2-8). Adaptive- ASV adds some expert rules that put safety limits on fre-
support ventilation (ASV) on the Hamilton ventilators is the quency and tidal volume delivery and reduce the risk of auto-
only commercially available mode to date that uses optimal PEEP. In that sense, this mode may be considered an intelligent
targeting. This targeting scheme was first described by Teh- targeting scheme, or more appropriately, a hybrid system (i.e.,
rani in 199120 and was designed to minimize the work rate of using a mathematical model and artificial intelligence).
Chapter 2 Classification of Mechanical Ventilators and Modes of Ventilation 55
Ventilator Disturbances
Patient
Weight Controller Effector
Flow Patient
Diagnosis Expert Rules IP (Hardware)
Inspired flow
Pressure
FIGURE 2-9 An intelligent targeting system for automatically adjusting pressure support levels (e.g., SmartCare/PS). IP, inspiratory pressure.
(Reproduced, with permission, from Chatburn RL, Mireles-Cabodevila E. Closed loop control of mechanical ventilation. Respir Care. 2011;56(1):85–98.
56 Part II Physical Basis of Mechanical Ventilation
Single neuron
Threshold
Inputs Weights Summation function Output
1
X Σ
0
Neural network
an overall unit activation. If this activation exceeds a cer- system learning from experience so that the control actually
tain threshold, the unit produces an output response. Large spans between patients instead of just between breaths.
numbers of neurons can be linked together in layers (see
Fig. 2-10). The nodes in the diagram represent the summa-
tion and transfer processes. Note that each node contains Mode Classification
information from all neurons. As the network learns, the
weights change, and thus the values at the nodes change, When Mushin et al wrote the classic book on automatic
affecting the final output. ventilation of the lungs,10 the emphasis was on classifying
In summary, ventilator control schemes display a definite ventilators and there were very few modes on each device.
hierarchy of evolutionary complexity. At the most basic level, These devices have undergone a tremendous technological
control is focused on what happens within a breath. We can evolution during the intervening years. As a result, there are
call this manual control, and there is a very direct need for now more than 170 names of modes on ventilators in the
operator input of static set-points. The next level up is what United States alone, with as many as two dozen available
we can call automatic control. Here, set-points are dynamic on a single device. The proliferation of names makes edu-
in that they may be adjusted automatically over time by the cation of end users very difficult, potentially compromis-
ventilator according to some model of desired performance. ing the quality of patient care. In addition, although there
The operator is somewhat removed in that inputs are entered may be more than 170 mode names, these are not uniquely
at the level of the model and take effect over several breaths different modes. Consequently, the emphasis today in
instead of at the level of individual breath control. Finally, describing ventilators must be on classifying modes, shift-
the highest level so far is what might be considered intelligent ing awareness from names to tags. Much has been written
control. Here, the operator can be eliminated altogether. Not on the subject,2,5, 29–31 and this section gives a brief overview
only dynamic set-points but also dynamic models of desired of the development and application of a ventilator mode
performance are permitted. There is the possibility of the taxonomy.
Chapter 2 Classification of Mechanical Ventilators and Modes of Ventilation 57
You can easily appreciate the motivation for classifying mode can deliver. CSV implies all spontaneous breaths;
modes, just as we do animals or plants (or cars or drugs) IMV allows spontaneous breaths to occur between
because of their large number and variety. The logical basis mandatory breaths and CMV does not.
for a mode taxonomy, however, is not apparent without some 8. There are only five basic ventilatory patterns: VC-CMV,
consideration. This basis has become a teaching system VC-IMV, PC-CMV, PC-IMV, and PC-CSV. All modes
I have developed and tested and is founded on ten simple can be categorizes by these five patterns. This provides
constructs (or aphorisms), each building on the previous one enough practical detail about a mode for most clinical
to yield a practical taxonomy. These aphorisms summarize purposes.
many of the ideas discussed previously in this chapter, and 9. Within each ventilatory pattern there are several variations
there is even some evidence that they are recognized inter- that can be distinguished by their targeting scheme(s).
nationally by clinicians.32 In simplified form, the aphorisms When comparing modes or evaluating the capability of
are as follows: a ventilator, more detail is required than just the ventila-
tory pattern. Modes with the same pattern can be dis-
1. A breath is one cycle of positive flow (inspiration) and neg-
tinguished by describing the targeting schemes they use.
ative flow (expiration). The purpose of a ventilator is to
There are at present only six basic targeting schemes: set-
assist breathing. Therefore, the logical start of a taxonomy
point, dual, servo, adaptive, optimal, and intelligent.
is to define a breath. Breaths are defined such that dur-
10. A mode of ventilation is classified according to its con-
ing mechanical ventilation, small artificial breaths may be
trol variable, breath sequence, and targeting scheme(s).
superimposed on large natural breaths or vice versa.
A practical taxonomy of ventilatory modes is based on
2. A breath is assisted if pressure rises above baseline during
just four levels of detail: the control variable (pressure or
inspiration or falls during expiration. A ventilator assists
volume), the breath sequence (CMV, IMV, or CSV), the
breathing by doing some portion of the work of breath-
targeting scheme used for primary breaths (CMV and
ing. This occurs by delivering volume under pressure.
CSV), and, if applicable, secondary breaths (IMV).
3. A ventilator assists breathing using either pressure con-
trol (PC) or volume control (VC). The equation of In teaching these constructs to respiratory therapists and
motion is the fundamental model for understanding physicians, most educators would agree that knowing a con-
patient–ventilator interaction and hence modes of ven- cept and applying it are two different skills. As with any tax-
tilation. The equation is an expression of the idea that onomy, learning the definitions and mastering the heuristic
only one variable can be predetermined at a time; pres- thinking required to actually categorize specific cases requires
sure or volume (flow control is ignored for simplicity further guidance and some practice. Say, for example, your
and for historical reasons, and because controlling flow task is to compare the capabilities of two major intensive care
directly will indirectly control volume and vice versa). unit ventilator models for a large capital purchase. Memoriz-
4. Breaths are classified according to the criteria that trig- ing the ten aphorisms may not translate into the ability to
ger (start) and cycle (stop) inspiration. A ventilator must classify the modes offered on these two ventilators as a basis
know when to start and stop flow delivery for a given for comparison. To facilitate that skill, I created the three tools
breath. Because starting and stopping inspiratory flow shown in Figures 2-11 and 2-12 and in Table 2-4. Using these
are critical events in synchronizing patient–ventilator tools you can create a simple spreadsheet that defines and
interaction, and because they involve uniquely different compares the modes on any number of ventilators. Table 2-5
operator-influenced factors, they are distinguished by is an example of such a table for the Covidien PB 840 ventila-
giving them different names. tor and the Dräger Evita XL ventilator. When implemented as
5. Trigger and cycle criteria can be either patient or machine a spreadsheet with built-in data-sorting functions, the table
initiated. A major design consideration in creating modes becomes a database with several major uses:
is the ability to synchronize breath delivery with patient
1. A “Rosetta Stone” that can be used to translate from mode
demand and at the same time to guarantee breath delivery
name to mode classification and vice versa. In this way
if the patient is apneic. Therefore, understanding patient–
modes can be identified that are functionally identical but
ventilator interaction means understanding the difference
have different proprietary names.
between machine and patient trigger and cycle events.
2. A tool for engineers to describe performance character-
6. Breaths are classified as spontaneous or mandatory based
istics of individual named modes. Information like this
on both the trigger and cycle criteria. A spontaneous breath
should be available to users in the ventilator’s manual.
arises without apparent external cause. Thus, it is patient
3. A system for clinicians to compare and contrast the capa-
triggered and patient cycled. Any machine involvement
bilities of various modes and ventilators.
in triggering or cycling leads to a mandatory breath. Note
4. A paradigm for educators to use in teaching the basic
that the definition of a spontaneous breath is independent
principles of mechanical ventilation.
of the definition of an assisted or unassisted breath.
7. Ventilators deliver only three basic breath sequences: One can imagine the utility of an expanded database con-
CMV, IMV, and CSV. The two breath classifications taining the classification of all modes on all commercially
logically lead to three possible breath sequences that a available ventilators.
58 Part II Physical Basis of Mechanical Ventilation
Review list of
operator-initiated
Evaluate ventilator
settings and
specifications
ventilator-initiated
settings
Identify what
happens
during a single
breath
VT = Tidal volume
TI = Inspiratory time
Intrabreath
VT is preset Control variable
*Examples:
directly or by flow Yes is
Volume-assist control
and TI* volume
Volume SIMV
Record control
No
variable
No
Paw is
proportional to
*Examples: Yes
inspiratory
Automatic tube compensation effort*
Proportional-assist ventilation
Neurally adjusted ventilatory assist
No
Control variable
is
time*
*Example:
Interpulmonary percussive ventilation
FIGURE 2-11 Algorithm for determining the control variable when classifying a mode. SIMV, synchronized intermittent mandatory ventilation.
(Copyright 2011 by Mandu Press Ltd. and reproduced with permission.)
VENTILATOR ALARM SYSTEMS MacIntyre33 has suggested that alarms also be catego-
rized by the events that they are designed to detect. Level
As with other components of ventilation systems, ventilator 1 events include life-threatening situations, such as loss
alarms have increased in number and complexity. Fortu- of input power or ventilator malfunction (e.g., excessive
nately, the classification system I have been describing can or no flow of gas to the patient). The alarms in this cat-
be expanded to include alarms as well (see Table 2-1). egory should be mandatory (i.e., not subject to operator
Chapter 2 Classification of Mechanical Ventilators and Modes of Ventilation 59
Yes
Patient
can cycle No No
inspiration*
*Normal operation
not alarm condition
Spontaneous
Yes breaths between Yes
mandatory*
*Example
Spontaneous - SIMV
breath is possible
No
Mandatory Breath
Machine cycle
breath is not No Yes sequence is
possible*
possible IMV
*Normal operation
not safety backup feature
Breath
sequence is
CSV
Record breath
sequence
FIGURE 2-12 Algorithm for determining the breath sequence when classifying a mode. (Copyright 2011 by Mandu Press Ltd. and reproduced with
permission.)
60
TABLE 2-4: EXPLANATION OF HOW TARGETING SCHEMES TRANSFORM OPERATOR INPUTS INTO VENTILATOR OUTPUTS
P, pressure; V, volume; F, flow; T, time; R, resistance; E, elastance; MV, minute volume; Edi, electrical activity of diaphragm; low impedance, low resistance and/or elastance;
WB Target, within-breath preset parameters of the pressure, volume, or flow waveform; BB Target, between breath targets modify high impedance, high resistance and/or elastance;
WB targets or overal ventiltory pattern; Cycle, end of inspiration; NA, not available as operator preset, ventilator
determines value if applicable.
Source: Copyright 2011 by Mandu Press Ltd, and reproduced with permission.
Chapter 2 Classification of Mechanical Ventilators and Modes of Ventilation 61
CMV, continuous mandatory ventilation; CSV, continuous spontaneous ventilation; IMV, intermittent mandatory ventilation.
Source: Copyright 2011 by Mandu Press Ltd. and reproduced with permission.
62 Part II Physical Basis of Mechanical Ventilation
choice), redundant (i.e., multiple sensors and circuits), difficult. It is like buying a Ferrari and putting wooden
and noncanceling (i.e., alarm continues to be activated, wheels on it. In the future, water vapor should be treated
even if the event is corrected, and must be reset manu- like any other desirable inhaled gas constituent (e.g., air,
ally). Level 2 events can lead to life-threatening situations oxygen, helium, or nitric oxide) and metered from within
if not corrected in a timely fashion. These events include the ventilator. The inspiratory part of the patient circuit
such things as blender failure, high or low airway pressure, should be a sterile, insulated, permanent part of the ven-
autotriggering, and partial patient circuit occlusion. They tilator right up to the patient connection, which can be a
also may include suspicious ventilator settings such as an disposable tip for cleaning purposes. The gas should be
inspiratory-to-expiratory timing (I:E) ratio greater than delivered under high pressure as a jet to provide not only
1:1. Alarms for level 2 events may not be redundant and conventional pressure, volume, and flow waveforms but
may be self-canceling (i.e., alarm inactivated if event ceases also high-frequency ventilation. The jet also can be used
to occur). Level 3 events are those that affect the patient– to provide a counterflow PEEP effect, eliminating any
ventilator interface and may influence the level of support need for an exhalation–valve system. The disposable tip
provided. Examples of such events are changes in patient could be designed to house disposable sensors and would
compliance and resistance, changes in patient respiratory be the only part of the circuit to be exposed to the patient’s
drive, and auto-PEEP. Alarm function at this level is similar exhaled gas. If ventilator manufacturers saw themselves as
to that of level 2 alarms. Level 4 events reflect the patient providers of the entire system, instead of letting third par-
condition alone rather than ventilator function. As such, ties deal in plastic connecting tubing, I think we would see
these events usually are detected by stand-alone monitors, a huge evolutionary step in ventilator performance, better
such as oximeters, cardiac monitors, and blood-gas ana- patient outcomes, and potential savings in labor costs for
lyzers. Some ventilators, however, are able to incorporate providers.
the readings of a capnograph in their displays and alarm Third, the most exciting area for development probably
systems. is in the intelligence that will be built into future ventila-
tor control circuits. The real challenge in closed-loop con-
trol of ventilation is defining, measuring, and interpreting
the appropriate feedback signals. If we stop to consider
THE FUTURE all the variables a human operator assesses, the problem
looks insurmountable. Not only does a human consider a
Almost 20 years ago, Warren Sanborn predicted that ventilators wide range of individual physiologic variables, but there
today would “… report the patient’s metabolic state; manage are the more abstract evaluations of such things as meta-
oxygen delivery; calculate cardiac output, synchronize breath bolic, cardiovascular, and psychological states. Add to this
delivery with cardiac cycle to maximize cardiac output…and the various environmental factors that may affect opera-
perform all these functions automatically or at least presenting tor judgment, and we get a truly complex control problem
consensus-based advisory messages to the practitioner….”17 (Fig. 2-13).
Some of these ideas were never developed commercially. I would like to speculate now about a response to this
Some were tried and abandoned. Some, have evolved beyond challenge. The ideal control strategy would have to start out
Warren’s broad vision. with basic tactical control of the individual breath. Next, we
There are three basic ways to improve ventilators in the add longer-term strategic control that adapts to changing
future. First, just like computer games, ventilators need to load characteristics. Mathematical models could provide the
improve the operator interface constantly. Yet very little basic parameters of the mode, whereas expert rules would
research has been done to call attention to problems with place limits to ensure lung protection.
current displays.34,35 We have come a long way from using a Next, we sample various physiologic parameters and use
crank to adjust the stroke of a ventilator’s piston to set tidal fuzzy logic to establish the patient’s immediate condition.
volume. The operator interface must provide for three basic This information is passed on to a neural network, which
functions: allow input of control and alarm parameters, would then select the best response to the patient’s condition.
monitor the ventilator’s status, and monitor the ventilator– The neural network ideally would have access to a huge
patient interaction status. We have a long way to go before database comprised of both human expert rules and actual
the user interface provides an ideal experience with these patient responses to various ventilator strategies. This
functions. arrangement would allow the ventilator not only to learn
Second, the weak link in the patient–ventilator system from its interaction with the current patient but also to con-
is the patient circuit. We buy a $35,000 ventilator with tribute to the database.
state-of-the-art computer control, and then we connect it Finally, the database and this ventilator could be net-
to the patient (priceless) with a $1.98 piece of plastic tub- worked with other intelligent ventilators to multiply the
ing that is subject to filling with condensate from a heated learning capacity exponentially (Fig. 2-14). Whatever
humidifier whose design has not changed appreciably the future brings, it seems clear that ventilators will have
in 20 years. The resistance and compliance of the deliv- more intelligence built in to increase patient safety and
ery circuit make flow control and volume delivery more decrease the time required to provide care.
Chapter 2 Classification of Mechanical Ventilators and Modes of Ventilation 63
Set-point
Adjustment
Pressure
Volume
Flow
Pressure Resp rate
Environment
(PIP and PEEP) Heart rate
Volume PeCO2
Frequency PaO2
Time FiO2 FiO2
Cost SpO2
Triage priority P0.1
Experience Disturbances
Alarms
Operator Ventilator Patient
Flow
Pressure
Human
experts
Optimization
models
Strategic
Competitive control
Registry neural
Database network Intelligent Expert
Prior Determine control rules
experience best
rules Disturbances
Ventilator Patient
Fuzzy
logic
Determine
patient
condition
FIGURE 2-14 A potential approach to the challenge of fully automated control of mechanical ventilation. (Reproduced, with permission, from
Chatburn RL. Computer control of mechanical ventilation. Respir Care. 2004;49:507–515.)
64 Part II Physical Basis of Mechanical Ventilation
SUMMARY AND CONCLUSION 11. Desautels DA. Ventilator performance. In: Kirby RR, Smith RA,
Desautels DA, eds. Mechanical Ventilation. New York, NY: Churchill
Livingstone; 1985:120.
Mechanical ventilators have become so complex that a 12. Sassoon CSH, Girion AE, Ely EA, Light RW. Inspiratory work of
system of classification is necessary to communicate intel- breathing on flow-by and demand flow continuous positive airway
pressure. Crit Care Med. 1989;17:1108–1114.
ligently about them. The theoretical basis for this classifica- 13. Amato MB, Barbas CS, Bonassa J, et al. Volume-assured pres-
tion system is a mathematical model of patient–ventilator sure support ventilation (VAPS): a new approach for reducing
interaction known as the equation of motion for the respira- muscle workload during acute respiratory failure. Chest. 1992;102:
tory system. From this model we deduce that as far as an 1225–1234.
14. Younes M. Proportional assist ventilation, a new approach to ventila-
individual inspiration is concerned, any conceivable venti- tor support: I. Theory. Am Rev Respir Dis. 1992;145:114–120.
lator can be classified as either a pressure, volume, or flow 15. Guttmann J, Eberhard L, Fabry B, et al. Continuous calculation of
controller (and in rare cases, simply an inspiratory-expira- intratracheal pressure in tracheally intubated patients. Anesthesiology.
1993;79(3):503–513.
tory time controller). An individual breath is shaped by the 16. Sinderby C, Beck J, Spahija J, et al. Inspiratory muscle unloading by
phase variables that determine how the breath is triggered neurally adjusted ventilatory assist during maximal inspiratory efforts
(started), targeted (sustained), and cycled (stopped). in healthy subjects. Chest. 2007;131(3):711–717.
A mode of ventilation can be characterized using a four- 17. Sanborn WG. Microprocessor-based mechanical ventilation. Respir
Care. 1993;38(1):72–109.
level taxonomy: (a) control variable, that is, pressure or vol- 18. Stengel RF. Optimal Control and Estimation. Mineola, NY: Dover Pub-
ume according to the equation of motion; (b) the breath lications, 1994.
sequence, that is, CMV, IMV, or CSV; (c) targeting scheme 19. Tehrani FT. Automatic control of an artificial respirator. Conf Proc
for primary breaths; and (d) targeting scheme for second- IEEE Eng Med Biol Soc. 1991;13:1738–1739.
20. Tehrani FT. Automatic control of mechanical ventilation, Part 2:
ary breaths. The trend in ventilator targeting schemes The existing techniques and future trends. J Clin Monit Comput.
has been from basic manual control (within-breath con- 2008;22(6):417–424.
trol requiring operator input of static set-points), to more 21. Otis AB, Fenn WO, Rahn H. Mechanics of breathing in man. J Appl
Physiol. 1950;2:592–607.
advanced automatic control (between-breath control of set- 22. Intelligent control. http://en.wikipedia.org/wiki/Intelligent_control.
points that are adjusted automatically by the ventilator with Last modified October 10, 2011. Last accessed April 30, 2010.
minimal operator input), to the highest level of intelligent 23. East TD, Heermann LK, Bradshaw RL, et al. Efficacy of computerized
control (in which the operator theoretically may be elimi- decision support for mechanical ventilation: results of a prospective
multicenter randomized trial. Proc AMIA Symp. 1999;251–255.
nated altogether in favor of artificial-intelligence systems 24. Lellouche F, Mancebo J, Jolliet P, et al. A multicenter randomized trial
capable of learning). of computer-driven protocolized weaning from mechanical ventila-
tion. Am J Respir Crit Care Med. 2006;174(8):894–900.
25. Rose L, Presneill JJ, Cade JF. Update in computer-driven weaning from
mechanical ventilation. Anaesth Intensive Care. 2007;35:213–221.
26. Rose L, Presneill JL, Johnston L, Cade JF. A randomised, controlled
REFERENCES trial of conventional versus automated weaning from mechanical
ventilation using SmartCare/PS. Intensive Care Med. 2008;34(10):
1. Chatburn RL, Mireles-Cabodevila E. Closed-loop control of mechani- 1788–1795.
cal ventilation: description and classification of targeting schemes. 27. Snowden S, Brownlee KG, Smye SW, Dear PR. An advisory system for
Respir Care. 2011;56(1):85–102. artificial ventilation of the newborn utilizing a neural network. Med
2. Chatburn RL. Understanding mechanical ventilators. Expert Rev Inform (Lond). 1993;18:367–376.
Respir Med. 2010;4(6):809–819. 28. Gottschalk A, Hyzer MC, Greet RT. A comparison of human and
3. Chatburn RL, Volsko TA. Mechanical ventilators. In: Wilkins RL, machine-based predictions of successful weaning from mechanical
Stoller JK, Scanlan CL, eds. Egan’s Fundamentals of Respiratory Care. ventilation. Med Decis Making. 2000;20:243–244.
8th ed. St. Louis, MO: Mosby; 2003:929–962. 29. Chatburn RL. Classification of ventilator modes: update and proposal
4. Chatburn RL. Mechanical ventilators: classification and principles of for implementation. Respir Care. 2007;52(3):301–323.
operation. In: Hess DR, MacIntyre NR, Mishoe SC, et al, eds. Respi- 30. Chatburn RL, Volsko TA. Mechanical ventilators. In: Stoller JK, Kac-
ratory Care: Principles and Practice. Philadelphia, PA: Saunders; marek RM, eds. Egan’s Fundamentals of Respiratory Care. 10th ed. St.
2002:757–809. Louis, MO: Mosby Elsevier; 2011 (in press).
5. Chatburn RL, Primiano FP Jr. A new system for understanding modes 31. Chatburn RL, Volsko TA. Mechanical ventilators: classification and
of mechanical ventilation. Respir Care. 2001;46:604–621. principles of operation. In: Hess DR, MacIntyre NR, Mishoe SC, et
6. Chatburn RL. Fundamentals of Mechanical Ventilation. Cleveland al, eds. Respiratory Care: Principles and Practice. 2nd ed. Philadelphia,
Heights, OH: Mandu Press; 2003. PA: Saunders; 2011 (in press).
7. Branson RD, Hess DR, Chatburn RL. Respiratory Care Equipment. 2nd 32. Chatburn RL. Determining the basis for a taxonomy of mechanical
ed. Philadelphia, PA: Lippincott Williams & Wilkins; 1999. ventilation. Respir Care. 2009;54(11):1555.
8. Cairo JM, Pilbeam SP. Mosby’s Respiratory Care Equipment. 7th ed. St. 33. MacIntyre NR. Ventilator monitors, displays, and alarms. In:
Louis, MO: Mosby; 2004. MacIntyre NR, Branson RD, eds. Mechanical Ventilation. Philadelphia,
9. Rodarte JR, Rehder K. Dynamics of respiration. In: Macklem PT, PA: Saunders; 2001:131–144.
Mead J, eds. Handbook of Physiology. Section 3: The Respiratory System. 34. Wachter SB, Johnson K, Albert R, et al. The evaluation of a pulmo-
Volume. III: Mechanics of Breathing. Part 1. Bethesda, MD: American nary display to detect adverse respiratory events using high resolution
Physiological Society; 1986;131–144. human simulator. J Am Med Inform Assoc. 2006;13(6):635–642.
10. Mushin WW, Rendell-Baker L, Thompson PW, Mapelson WW. Auto- 35. Uzawa Y, Yamada Y, Suzukawa M. Evaluation of the user interface
matic Ventilation of the Lungs. 3rd ed. Oxford, England: Blackwell Sci- simplicity in the modern generation of mechanical ventilators. Respir
entific; 1980:62–131. Care. 2008;53(3):329–337.
BASIC PRINCIPLES 3
OF VENTILATOR DESIGN
Robert L. Chatburn
Eduardo Mireles-Cabodevila
THE VENTILATOR AS A “BLACK BOX” There was a time when you could take a handful of
simple tools and do routine maintenance on your car engine.
A mechanical ventilator is an automatic machine designed to About that time the average clinician could also completely
provide all or part of the work the body must do to move gas disassemble and reassemble a mechanical ventilator as a
into and out of the lungs. The act of moving air into and out training exercise or to perform repairs. In those days (the
of the lungs is called breathing, or, more formally, ventilation. late 1970s), textbooks1 describing ventilators understand-
The simplest mechanical device we could devise to assist ably paid much attention to the individual mechanical
a person’s breathing would be a hand-driven, syringe-type components and pneumatic schematics. In fact, this philoso-
pump that is fitted to the person’s mouth and nose using phy was reflected to some extent in previous editions of this
a mask. A variation of this is the self-inflating, elastic book. Today, both cars and ventilators are incredibly complex
resuscitation bag. Both of these require one-way valve mechanical devices controlled by multiple microprocessors
arrangements to cause air to flow from the device into running sophisticated software (Fig. 3-1). Figure 3-2 shows
the lungs when the device is compressed, and out from the pneumatic schematic of a current intensive care ventila-
the lungs to the atmosphere as the device is expanded. These tor. All but the most rudimentary maintenance of ventilators
arrangements are not automatic, requiring an operator to is now the responsibility of specially trained biomedical
supply the energy to push the gas into the lungs through engineers. Our approach to describing ventilator design has
the mouth and nose. Thus, such devices are not considered thus changed from a focus on individual components to
mechanical ventilators. a more generalized model of a ventilator as a “black box,”
Automating the ventilator so that continual opera- that is, a device for which we supply an input and expect
tor intervention is not needed for safe, desired operation a certain output and whose internal operations are largely
requires three basic components: unknowable, indeed, irrelevant, to most clinical operators.
What follows, then, is only a brief overview of the key design
1. A source of input energy to drive the device; features of mechanical ventilators with an emphasis on input
2. A means of converting input energy into output energy in power requirements, transfer functions (pneumatic and
the form of pressure and flow to regulate the timing and electronic control systems), and outputs (pressure, volume,
size of breaths; and and flow waveforms). The rest of the chapter focuses on
3. A means of monitoring the output performance of the the interactions between the operator and the ventilator
device and the condition of the patient.
65
66 Part II Physical Basis of Mechanical Ventilation
A B
C D
FIGURE 3-1 Examples of commonly used intensive care ventilators: A. Dräger Infinity V500, B. Hamilton G5, C. Maquet Servo i,
D. Covidien PB840. (Image with permission from Nellcor Puritan Bennett LLC, Boulder, Colorado, doing business with Covidien.)
Chapter 3 Basic Principles of Ventilator Design 67
A 8 B 9 10 11
3 7
5 P 18 F
17 E
27 28
4 6 P .
Co2 V
E 12
20 E E
2 19
13
23 24 G C 14
Air
26 O2
H E D . Insp. gas
25 21 O2 P V
I E E 16 15 E Exp. gas
22 Nebulizer gas
Inputs
Mechanical ventilators are typically powered by electricity
or compressed gas. Electricity, either from wall outlets
(e.g., 100 to 240 volts AC, at 50/60 Hz) or from batteries
(e.g., 10 to 30 volts DC), is used to run compressors of
various types. Batteries are commonly used as the primary
power source in the home-care environment but are
usually reserved for patient transport or emergency use
in hospitals. These sources provide compressed air for
motive power as well as air for breathing. Alternatively, the
power to expand the lungs is supplied by compressed gas
from tanks, or from wall outlets in the hospital (e.g., 30 to
80 pounds per square inch [psi]). Some transport and emer-
gency ventilators use compressed gas to power both lung
inflation and the control circuitry. For these ventilators,
knowledge of gas consumption is critical when using
cylinders of compressed gas.
The ventilator is generally connected to separate sources
of compressed air and compressed oxygen. In the United
States, hospital wall outlets supply air and oxygen at 50 psi,
although most ventilators have internal regulators to reduce
this pressure to a lower level (e.g., 20 psi). This permits the
delivery of a range of oxygen concentrations to support
FIGURE 3-3 CareFusion Infant Flow SiPAP device.
the needs of sick patients. Because compressed gas has all
moisture removed, the gas delivered to the patient must be
warmed and humidified so as to avoid drying out the lung
tissue. tubing that conducts the gas to the patient, called the patient
circuit. Such devices are used in small transport ventilators
Conversion and Control and automatic resuscitators. Manually adjusted variable-
orifice flow meters have been used in simple infant ventila-
The input power of a ventilator must be converted to a tors in the past (e.g., Bourns BP-200) and are currently used
predefined output of pressure and flow. There are several in the Infant Flow SiPAP device (CareFusion, Minneapolis,
key systems required for this process. If the only power MN), as shown in Figure 3-3. The advent of inexpensive
input is electrical, the ventilator must use a compressor microprocessors in the 1980s led to development of digital
or blower to generate the required pressure and flow. A control of flow valves that allow a great deal of flexibility in
compressor is a machine for moving a relatively low flow shaping the ventilator’s output pressure, volume, and flow
of gas to a storage container at a higher level of pressure waveforms (Fig 3-4).2 Such valves are used in most of the
(e.g., 20 psi). A blower is a machine for generating rela- current generation of intensive care ventilators.
tively larger flows of gas as the direct ventilator output Directing flow from the source gas into the patient
with a relatively moderate increase of pressure (e.g., 2 psi). requires the coordination of the output flow-control valve
Compressors are generally found on intensive care ventila- and an expiratory valve or “exhalation manifold” (Fig. 3-5).
tors whereas blowers are used on home-care and transport In the simplest case, when inspiration is triggered on, the
ventilators. Compressors are typically larger and consume output control valve opens, the expiratory valve closes, and
more electrical power than blowers, hence the use of the the only path left for gas is into the patient. When inspira-
latter on small, portable devices. tion is cycled off, the output valve closes and the exhalation
valve opens, flow from the ventilator ceases and the patient
exhales out through the expiratory valve (see Fig. 3-2). The
FLOW-CONTROL VALVES
most sophisticated ventilators employ a complex interaction
To control the flow of gas from a compressor, ventilator engi- between the output flow-control valve and the exhalation
neers use a variety of flow-control valves, from very simple valve, such that a wide variety of pressure, volume, and flow
to very complex. The simplest valve is just a fixed orifice flow waveforms may be generated to synchronize the ventilator
resistor that permits setting a constant flow to the external output with patient effort as much as possible.
Chapter 3 Basic Principles of Ventilator Design 69
Wires to
controller
Coils
Actuator
Outputs
Just as the study of cardiology involves the use of electro-
cardiograms and blood pressure waveforms, the study
of mechanical ventilation requires an understanding of
output waveforms. The waveforms of interest are the
pressure, volume, and flow.
characteristic waveform shapes, the specific baseline EFFECTS OF THE PATIENT CIRCUIT
values are irrelevant. What is important is the relative
The pressure, volume, and flow the patient actually
magnitudes of each of the variables and how the value of
receives are never precisely the same as what the clini-
one affects or is affected by the value of the others.
cian sets on the ventilator. Sometimes these differences
Figure 3-7 illustrates the typical waveforms available on
are caused by instrument inaccuracies or calibration error.
modern ventilators. These waveforms are idealized; that is,
More commonly, the patient delivery circuit contributes
they are precisely defined by mathematical equations and are
to discrepancies between the desired and actual patient
meant to characterize the operation of the ventilator’s con-
values. This is so because the patient circuit has its own
trol system. As such, they do not show the minor deviations,
compliance and resistance. Thus, the pressure measured
or “noise,” often seen in waveforms recorded during actual
inside a ventilator upstream of the patient always will be
ventilator use. This noise can be caused by a variety of
higher than the pressure at the airway opening because
extraneous factors such as vibration and flow turbulence. Of
of patient circuit resistance. In addition, the volume and
course, scaling of the horizontal and vertical axes can affect
flow coming out of the ventilator’s exhalation manifold
the appearance of actual waveforms considerably. Finally,
will exceed those delivered to the patient because of the
the waveforms in Figure 3-7 do not show the effects of the
compliance of the patient circuit.
resistance and compliance of the patient circuit.
Exactly how the mechanical properties of the patient
No ventilator is an ideal pressure, volume, or flow
circuit affect ventilator performance depends on whether
controller, and ventilators are designed to only approximate
they are connected in series or in parallel with the patient.
a particular waveform. Idealized waveforms as shown in
It turns out that the resistance of the patient circuit is
Figure 3-7 are, nevertheless, helpful because they are used
connected in series whereas the compliance is modeled as
commonly in other fields (e.g., electrical engineering),
a parallel connection. To understand this, we first make the
which makes it possible to use mathematical procedures
simplifying assumption that we can examine the patient cir-
and terminology that already have been established. For
cuit’s resistance separate from its compliance. It is intuitively
example, a standard mathematical equation is used to
obvious that the same flow of gas that comes from the ven-
describe the most common ventilator waveforms for each
tilator travels through the circuit tubing as through the
control variable. This known equation may be substituted
patient’s airway opening. We also can see that the pressure
into the equation of motion, which is then solved to get the
drop across the patient circuit will be different from that
equations for the other two variables. Once the equations
across the respiratory system because they have different
for pressure, volume, and flow are known, they are easily
resistances. By a definition we borrow from electronics,
graphed. This is the procedure that was used to generate the
when two circuit components share the same flow but have
graphs in Figure 3-7.
different pressure drops, they are connected in series. This
A B C D E
Pressure
Volume
Inspiration
Flow
Expiration
FIGURE 3-7 Idealized ventilator output waveforms. A. Pressure-controlled inspiration with a rectangular pressure waveform. B. Volume-controlled
inspiration with a rectangular flow waveform. C. Volume-controlled inspiration with an ascending-ramp flow waveform. D. Volume-controlled inspi-
ration with a descending-ramp flow waveform. E. Volume-controlled inspiration with a sinusoidal flow waveform. The short dashed lines represent
mean inspiratory pressure, and the long dashed lines represent mean pressure for the complete respiratory cycle (i.e., mean airway pressure). Note that
mean inspiratory pressure is the same as the pressure target in A. These waveforms were created as follows: (a) defining the control waveform using a
mathematical equation (e.g., an ascending-ramp flow waveform is specified as flow = constant × time), (b) specifying the tidal volume for flow-control
and volume-control waveforms, (c) specifying the resistance and compliance, (d) substituting the preceding information into the equation of motion
for the respiratory system, and (e) using a computer to solve the equation for the unknown variables and plotting the results against time. (Reproduced,
with permission, from Chatburn RL. Fundamentals of Mechanical Ventilation. Cleveland Heights, OH: Mandu Press; 2003:143.)
Chapter 3 Basic Principles of Ventilator Design 71
means that the patient circuit resistance, however small, adds We can get a more intuitive understanding of this equa-
to the total resistive load seen by the ventilator. Thus, in a tion if we put in some values. Suppose, for example, that
volume-controlled breath, the peak inspiratory pressure is we use the perfect patient circuit that has zero compliance.
higher, and in a pressure-controlled breath, the tidal volume Substituting zero for CPC, we get
and peak flow are lower. In practice, the effect of patient
Vset Vset
circuit resistance is usually ignored because it is so much Vdelivered = =
lower than the resistance of the respiratory system. 1 + (C PC /C RS ) 1 + (0/C RS )
(2)
Now consider the patient circuit compliance. The V V
effective compliance of the patient circuit is a com- = set = set = Vset
1+ 0 1
bination of the tubing compliance and the compress-
ibility of the gas inside it. As the ventilator delivers the which shows that there is no effect on the delivered tidal vol-
breath to the patient, pressure at the airway opening ume. Suppose now that CPC is as large as CRS (i.e., CPC = CRS).
rises relative to atmospheric pressure, which is the driv- Now we have
ing force for flow into the lungs. The patient circuit is
connected between the ventilator and the airway, so the Vset V V
Vdelivered = = set = set (3)
pressure it experiences across its walls is the same as that 1 + (C PC /C RS ) 1+ 1 2
experienced by the respiratory system (remember that
we are ignoring its resistance now, so we can ignore any in which case, half the volume from the ventilator goes to
pressure drop between the ventilator outlet and the air- the patient, and the other half is compressed in the patient
way opening). The volume change of the patient circuit circuit. Some ventilators automatically compensate for gas
tubing is different from that of the respiratory system lost to the patient circuit.2
because the compliance of the circuit is different. Because The effect of the patient circuit is more troublesome
the patient circuit and the respiratory system fill with during volume-controlled modes than during pressure-
different volumes during the same inspiratory time, the controlled modes. This is so because during volume control,
flows they experience are different (remember that flow = the ventilator meters out a specific volume of gas, and unless
volume ÷ time). Again borrowing a definition from elec- it measures flow at the airway opening, it has no way of
tronics, if two circuit components share the same pressure knowing how much goes to the patient and how much goes
drop but different flows, they are connected in paral- to the patient circuit. In contrast, during pressure-controlled
lel. Because they are in parallel, the two compliances are modes, the ventilator simply meters out a set pressure change
additive, so the total compliance is greater than either no matter where the gas goes. Because the respiratory system
component. and the patient circuit compliance are in parallel, they both
Patient circuit compliance sometimes can be greater than experience the same driving pressure (peak inspiratory
respiratory system compliance and thus can have a large pressure minus end-expiratory pressure), so tidal volume
effect on ventilation. It must be accounted for either auto- delivery is affected very little. The only effect might be that
matically by the ventilator or manually by increasing the tidal the patient circuit compliance may tend to increase the pres-
volume. For example, when ventilating neonates, patient sure rise time, which would tend to decrease peak flow and
circuit compliance can be as much as three times that of the tidal volume slightly.
respiratory system, even with small-bore tubing and a small- Another area where patient circuit compliance causes
volume humidifier. Thus, when trying to deliver a preset trouble is in the determination of auto-PEEP. There are
tidal volume during volume-controlled ventilation, as little several methods for determining auto-PEEP. One method
as 25% of the set volume will be delivered to the patient, with to determine auto-PEEP during mechanical ventilation is
75% compressed in the patient circuit. The compliance of the to create an expiratory hold manually (i.e., delay the next
patient circuit can be determined by occluding the tubing at inspiration) until static conditions prevail throughout the
the patient Y, delivering a small volume under flow control lungs (i.e., no flow anywhere in the lungs). The pressure
(using zero positive end-expiratory pressure [PEEP]), and at this time (total PEEP) minus the applied PEEP is an
noting the resulting pressure. Using a short inspiratory hold estimation of global auto-PEEP. Note that auto-PEEP may
will make it easier to read the pressure. Then compliance vary throughout the lungs depending on the distribution of
is calculated as before, by dividing the volume by the pres- lung disease and may not reflect pressure behind collapsed
sure. Once the patient circuit compliance is known, the set areas in patients with severe flow limitation. Auto-PEEP is
tidal volume can be corrected using the following equation: an index of the gas trapped in the system at end expiration
secondary to an insufficient expiratory time:
Vset
Vdelivered = (1) Vtrapped
1 + (C PC / CRS ) measured auto-PEEP = (4)
C total
where Vdelivered is the tidal volume delivered to the patient,
Vset is the tidal volume setting on the ventilator, CPC is the where Vtrapped is the volume of gas trapped in the patient and
patient circuit compliance, and CRS is the respiratory system the patient circuit at end-expiration (above that associated
compliance. with applied PEEP), and Ctotal is the total compliance of the
72 Part II Physical Basis of Mechanical Ventilation
mode of ventilation (volume vs. pressure control breaths) or where O2 required is 100% oxygen flow in L/min, f is the
the breath sequence (mandatory vs. spontaneous) will help breathing frequency in breaths/min, VT is the tidal volume in
understand how the setting will affect the ventilator output liters and the FIO2 is the patient O2 concentration desired in
(see Chapter 2). decimal format (i.e., 30% = 0.3). An oxygen analyzer should
The operator input is presented below in the order be used to confirm the measurements. It must be recognized
that follows the progression of a breath; starting with the that changes in oxygen flow, breathing rate, or tidal volume
gas inhaled, to triggering, targeting, cycling, and baseline will change the FIO2.
variables. When transporting the critically ill patient, availability
of oxygen supplies for the mechanically ventilated patient
is crucial. Size and weight of cylinders makes transport dif-
Inspired Gas Concentration ficult and presents an increased risk of fire. Branson et al.
have described a solution using a portable oxygen concen-
A mechanical ventilator has the capacity of deliver- trator (SeQual Eclipse II) paired with the Impact 754 and
ing different mixtures of gas. Most ventilators allow the Pulmonetics LTV-1200 ventilators.11
administration of specific concentrations of oxygen. A few For the rest of the current mechanical ventilators, the
allow the administration of helium, nitric oxide, or anesthe- ventilator adjusts the mixture of air and oxygen to achieve
sia gases. the desired FIo2. The mixing of air is achieved by an internal
or external blender. A blender may use proportioning valves
OXYGEN that regulate the flow of air and oxygen to a mixing changer
(Fig. 3-10). It is similar to the mechanism used to mix hot
Oxygen is the most common gas administered to patients and cold water in a shower—the more oxygen needed, the
undergoing mechanical ventilation. The oxygen percentage larger the opening for oxygen and the smaller it is for air.
in the inspired gas (FIO2) can be regulated in most ventilators To work properly, the blender requires a constant pressure
by means of a direct adjustment of a specific control (21% to within the working ranges of the device.
100%). However, this is not true for all ventilators. For exam- Most current ventilators have oxygen sensors to moni-
ple, some home ventilators (e.g., LP-10 or the LTV 1150, tor the FIO2. The oxygen sensor gives feedback to the opera-
Pulmonetic, CareFusion) use a connection to a low-pressure tor to adjust the mixture, or alarms if there is a discrepancy
oxygen source to the ventilator or the patient circuit. The between the set and delivered FIO2. The oxygen sensors detect
following formula can calculate the flow of oxygen to achieve changes in electrical current, which is proportional to the
a desired oxygen concentration: oxygen concentration. The most common techniques are:
f × VT × (desired F Io2 − 0.21) (a) paramagnetic, (b) polarographic, and (c) galvanic.12
O2 required = (6)
0.79
Oxygen Air
of any signal from the patient. The operator typically sets a TIME
breath frequency for machine-triggered breaths. A patient-
Time is measured by the internal ventilator processor. The
triggered breath is one for which inspiration is started solely
next breath is time triggered (in the absence of a patient
by a signal from the patient. The key operator set variable
trigger event) when the expiratory time has reached the
for patient triggering is sensitivity, or the magnitude of
threshold to maintain a set respiratory rate (e.g., if the set
the patient signal required to initiate inspiratory flow. The
rate is 10 breaths per minute and the inspiratory time is set at
patient signal can be obtained from measuring the airway
1 second, then the expiratory time is 5 seconds). Some modes
pressure, flow, volume, electromyogram (EMG),21 abdomi-
allow the user to set the inspiratory and expiratory time [e.g.,
nal motion (Graseby capsule22), thoracic impedance,23 or any
airway pressure release ventilation (APRV) and biphasic],
other measurable signal of respiratory activity.24 Most inten-
thus fixing the inspiratory-to-expiratory timing (I:E) ratio
sive care ventilators measure pressure and flow (volume is
and respiratory rate. In an effort to improve patient–ventila-
integrated from flow) at the circuit. There are only a few
tor interactions, the ventilator may synchronize the manda-
ventilators that use other sources of signaling, diaphragmatic
tory breath with the patient’s triggering signal if it falls within
EMG (Servo i NAVA), thoracic impedance (Sechrist SAVI),
a threshold. The classic example is synchronized intermit-
and abdominal motion (Infant Star STAR SYNC, which is no
tent mandatory ventilation (SIMV). More recently APRV, as
longer commercially available).24,25
programmed in the Evita XL, delivers a machine breath if
Ventilator triggering characteristics can be evaluated using
the patient trigger signal falls within 25% of the triggering
different metrics.23,26–28 The most sophisticated device for
time.29 Time triggering is also found as a safety mechanism.
evaluating ventilator performance is the ASL lung simulator
The operator or manufacturer enters a time after which the
(IngMar Medical Ltd., Pittsburgh, PA). This device can simu-
apnea alarm will trigger the delivery of a preset breath after a
late both passive lung mechanics (e.g., resistance and compli-
preset time is reached.
ance) as well as patient inspiratory and expiratory effort. It
can display and record pressure, volume, and flow signals, and
PRESSURE
calculate a wide variety of performance metrics. Figure 3-11
shows an example of these waveforms with specific reference The patient inspiratory effort causes a drop in pressure in
points for calculating performance metrics (from opera- the airway and the circuit. Inspiration starts when pres-
tor’s manual for software version 3.2). Using these reference sure falls below the preset “sensitivity” threshold. The site
points we can define the following key trigger metrics: Pmin of measurement will have an impact on the performance
(maximum pressure drop relative to PEEP during the trigger of the device. Pressure signals travel at the speed of sound,
phase), pressure-time product (∫ Paw−PEEP dt from start of approximately 1 ft/ms.30 The farther the sensor is from the
effort to return of airway pressure [Paw] to PEEP), patient signal source, the longer the potential time delay. The closest
trigger work (∫ Paw−PEEP dv from start of effort to return of measurements can be done in the trachea. Tracheal pressure
Paw, to PEEP), and time to trigger (period from the start of measurements reflect actual airway pressure as the endotra-
effort to the return of Paw to PEEP). cheal tube resistance is bypassed. When used for ventilator
80.00
70.00
60.00
Flow
50.00 Volume
40.00
30.00 E F
Paw
20.00 0 B
10.00 4.00
0.00 2.00
–10.00 0.00
–20.00
-Pmus –2.00
D
–30.00 A –4.00 G
–40.00 5.00A
–50.00 0 C
–60.00
Time
0.05 0.1 0 (seconds)
–70.00
0.0116009 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5.0116
FIGURE 3-11 Reference points on pressure, volume, and flow waveforms recorded by the ASL 5000 (IngMar Medical Ltd, Pittsburgh, PA). A. Start
of inspiratory effort, B. beginning of inhalation as determined by the “breath start volume threshold,” C. lowest pressure during the trigger phase,
Pmin, D. return of airway pressure to baseline during the trigger phase, E. end of inspiratory time, i.e., negative-going zero flow crossing, F. begin-
ning of exhalation as determined by the “expiratory start volume threshold,” and G. end of expiratory time, i.e., positive-going zero flow crossing.
(Reproduced, with permission, from Ingmar Medical. ASL 5000 v3.2 Operator’s Manual. Pittsburgh, PA: Author.)
76 Part II Physical Basis of Mechanical Ventilation
Advantages Disadvantages
A. Exhalation port: Well protected from mechanical abuse. During Requires protection from moisture of exhaled gas. During spontaneous
mechanical inhalation, accurately reads pressure at the Y. During inspiration, underestimates pressure generated at the Y to trigger the
inhalation, increases in inspiratory or expiratory circuit resistance ventilator. During exhalation, underestimates pressure at the Y. During
do not compromise inspiratory flow output, except for manyfold exhalation, increases in expiratory circuit resistance compromise
increases. expiratory flow. Hence, system requires well-maintained expiratory
filter to ensure that expiratory circuit resistance remains low.
B. Inhalation port: Well protected from mechanical abuse. Does not During mechanical inhalation, overestimates pressure at the Y. During
require protection from moisture or additional filters. During spontaneous inspiration, underestimates pressure generated at the Y
exhalation, accurately reads pressure at the Y as long as the to trigger the ventilator. During inhalation, increases in inspiratory
inspiratory circuit remains patient. During inhalation, increases in circuit resistance compromise inspiratory flow output. For example,
expiratory circuit resistance do not compromise inspiratory-flow factors such as selection of humidifier and type of patient circuit yield
output. varying patient inspiratory efforts for fixed ventilator settings.
C. Patient Y: During inhalation and exhalation, accurately reads both Susceptible to mechanical abuse. Requires a separate pressure-sensing
inspiratory and expiratory pressures. Pressure readings reflect tube, which is prone to occlusion, blockage, and disconnection, all of
relative condition of inspiratory and expiratory circuits. which prevent sensing of patient effort.
Source: Modified, with permission, from Sassoon CSH. Mechanical ventilator design and function: the trigger variable. Respir Care. 1992;37:1056–1069.
triggering, tracheal pressure sensing results in decreased an inspiratory effort, the expiratory flow drops, creating a
work of breathing.31–33 However, tracheal pressure measure- difference between the inspiratory and expiratory flow val-
ments are not routinely done and require special equipment ues. When the difference in values reaches the preset sensi-
(endotracheal tube with monitoring port) and no current tivity threshold, a breath is delivered. Some systems (Puritan
ventilator uses it to routinely trigger the ventilator. Bennett, 7200) allow the operator to set both the bias flow
The other sites of pressure measurement are the patient and the trigger sensitivity. Newer devices set the bias flow
circuit Y or at the inspiratory or expiratory ports, each with according to the operator selected value for the triggering
its advantages and disadvantages (Table 3-2). Trigger perfor- sensitivity. For example, the Puritan Bennett 840 sets the flow
mance will also be affected by the presence of humidifiers, 1.5 L/min above the selected sensitivity, and the Hamilton
filters, water condensation, patient circuit and exhalation G5 automatically sets the bias flow equal to two times the set
valves. These will most often dampen, or rarely amplify, the sensitivity threshold. As a backup, if flow sensor is kinked or
pressure signal. Clinically, the presence of a dampened signal taken out of line, an internal pressure trigger of −2 cm H2O is
will require a larger pressure change (higher work of breath- used until the flow sensor is “online” again.
ing) to reach the trigger threshold. On the contrary, presence Flow change may be detected by placing a sensor just
of water in the pressure tubing may cause oscillation, which before the endotracheal tube. The close proximity to the
can falsely trigger mechanical breaths. patient may enhance triggering. It, however, exposes the sen-
The trigger pressure sensitivity is usually set at 0.5 to sor to secretions and moisture, which may affect its perfor-
1.5 cm H2O below the baseline pressure. Common practice mance. Flow triggering seems more efficient than pressure
is to increase the sensitivity (i.e., decrease the pressure drop) triggering in terms of work of breathing.34 This, however,
until autotriggering occurs and then reduce sensitivity until seems of no particular clinical relevance in the presence of
the autotriggering just stops.30 Note that each ventilator appropriately set pressure triggering.35 Flow sensing may
comes with predetermined manufacturer set values and can cause autotriggering secondary to noninspiratory flow
be adjusted. changes. The flow change can happen in either the ventilator
circuit (leak in the circuit or endotracheal tube) or the patient
(cardiogenic oscillations or bronchopleural fistula).36,37
FLOW
A novel approach to flow triggering is offered on the
Flow triggering is based on the detection of a change in a Dräger Infinity V500 ventilator in the APRV mode. Rather
constant, small, baseline (bias) flow through the patient cir- than setting a T-low time to determine the time triggering
cuit. The operator sets a flow sensitivity threshold. When the of each mandatory breath, the operator may set a percent of
change in flow reaches the threshold, a breath is delivered. peak expiratory flow as the trigger threshold.
The changes in flow are detected at the expiratory valves or
by a flow sensor in the patient circuit. The ventilator mea- VOLUME
sures the flow from the ventilator and from the patient. In
a closed circuit, the two flow values should remain equal A breath may be triggered when a preset volume is detected
in the absence of patient effort. When the patient makes as the result of a patient inspiratory effort. This is similar to
Chapter 3 Basic Principles of Ventilator Design 77
flow triggering but using volume has the theoretical advan- adjusted. The cardiac cycle may also cause interference with
tage of being less susceptible to signal noise (i.e., integrat- the signal.22,25
ing flow to get volume cancels out some noise because of
flow oscillations). Volume triggering is rare in ventilators
but can be found on the Dräger Babylog VN500 infant Target Variables
ventilator.
During inspiration, the variable limiting the magnitude
of any parameter is called the target variable (previously
DIAPHRAGMATIC SIGNAL known as the limit of the control variable, but the term
limit is now reserved for alarm and safety conditions rather
The ideal approach to coordinate a mechanical ventilator
than control settings).42A target is a predetermined goal of
with the patient inspiratory effort would be to use the neu-
ventilator output. Targets can be viewed as the parameters
ral output of the respiratory center. Direct measurement of
of the targeting scheme (see Chapter 2). Within-breath tar-
the respiratory center output is currently not possible. The
gets are the parameters of the pressure, volume, or flow
phrenic nerve has been used as a trigger signal in animal
waveform. Examples of within-breath targets include
models,38,39 but not in humans. The only available clinical
inspiratory flow or pressure rise time (set-point targeting),
approach is measurement of the diaphragmatic electrical
inspiratory pressure and tidal volume (dual targeting), and
activity (Edi). Because the Edi is an electric signal, it easily
constant of proportionality between inspiratory pressure
becomes contaminated by the electrical activity of the heart,
and patient effort (servo targeting). Between-breath targets
the esophagus, and other muscles.21 More importantly, the
serve to modify the within-breath targets and/or the over-
Edi requires an intact respiratory center, phrenic nerve,
all ventilatory pattern. Between-breath targets are used with
neuromuscular junction, and assumes that the diaphragm is
more advanced targeting schemes, where targets act over
the primary inspiratory muscle (e.g., rather than accessory
multiple breaths. A simple example of a between-breath
muscles of ventilation).
target is to compare actual exhaled volume to a preset
The only clinically available system that uses diaphrag-
between-breath tidal volume so as to automatically adjust
matic signal trigging is the neurally adjusted ventila-
the within-breath constant pressure or flow target for the
tory assistance (NAVA) system. An esophageal catheter is
next breath. Examples of between-breath targets and tar-
used to measure the Edi. The sensitivity is set by entering
geting schemes include average tidal volume (for adap-
a value above the background electrical noise. The trigger
tive targeting), percent minute ventilation (for optimal
value is set in microvolts and represents the change in the
targeting), and combined partial pressure of carbon dioxide,
electrical signal rather than an absolute value.40 The default
volume, and frequency values describing a “zone of comfort”
setting is 0.5 microvolts, but it can be adjusted from 0 to
(for intelligent targeting).
2 microvolts. As a backup trigger signal in the absence of
a measurable Edi, NAVA uses flow or pressure triggering,
whichever happens first. PRESSURE
The ventilator uses microprocessors to control the delivery
of pressure. The pressure can be delivered with any pres-
OTHER SIGNALS
sure profile and in response to many signals. Currently,
The BiPAP Vision (Respironics Inc., Murrysville, PA) uses most modes of ventilation in which inspiratory pressure is
a triggering mechanism called shape-signal. The ventila- targeted deliver the pressure rapidly and attempt to main-
tor microprocessor generates a new flow signal, which tain the pressure constant throughout the inspiratory phase
is offset from the actual flow by 0.25 L/s and delays it for (square waveform). This means that the performance of the
300 milli seconds. The delay causes the flow shape signal ventilator depends on the delivery of the pressure wave-
to be slightly behind the patient’s flow rate. The mechani- form and any departure from the ideal waveform leads to
cal breath is triggered when a sudden decrease in expiratory differences in performance between ventilators.43,44
flow from an inspiratory effort crosses the shape signal.41
The Sechrist SAVI system (Sechrist Industries, Anaheim, Inspiratory Pressure. The pressure rise during inspiration
CA) is the only mode available that uses transthoracic associated with volume and flow delivery is set by
electrical impedance to trigger the ventilator.25 The tho- the operator (pressure control–continuous mandatory
racic impedance is obtained by placing two chest leads, ventilation) or closed-loop algorithms (e.g., pressure-
one in the anterior axillary line on the right and the other regulated volume control). Care should be exercised while
in the posterior axillary line on the left. The sensors are setting the ventilator or reading the literature as there is
placed high enough to avoid costal and subcostal retrac- significant variability between ventilator manufacturers and
tions. The chest sensors measure the electrical impedance peer-reviewed literature in the definitions and nomenclature
across the human body. As a breath occurs, the transtho- related to inspiratory pressures.43 The main problem
racic impedance changes as a result of a different ratio of stems from what historically has been used to define the
air-to-fluid in the thorax. The triggering threshold can be inspiratory pressure. For example, in the same ventilator, for
78 Part II Physical Basis of Mechanical Ventilation
Spontaneous breath
Pressure
support
Inspiratory pressure Pressure
Inspiratory pressure support
Pressure
support
PEEP
CPAP
P-low
pressure control–continuous mandatory ventilation breaths and peak airway pressure are synonymous. For example,
the peak inspiratory pressure is stated in reference to the measured peak airway pressure is often higher than set peak
set end-expiratory pressure (PEEP), but for APRV the peak inspiratory pressure because of pressure transients from an
inspiratory pressure is stated in reference to the atmospheric underdamped pressure-control system or noise from patient
pressure. To compound the confusion, on some ventilators movement. The introduction of the so-called active exhala-
the value of pressure support is set relative to PEEP (e.g., tion valve made possible unrestricted spontaneous breaths
Drager Evita XL, Puritan Bennett 840), on others (LTV 950) during the inspiratory phase of a mandatory pressure-
pressure support is set relative to the atmospheric pressure control breath. New modes brought new terms. For
(i.e., atmospheric pressure = zero airway pressure), and on at example, P-high or PEEP high refers to the peak inspiratory
least one ventilator (BiVent in Servo i) pressure support may pressure above atmospheric pressure in APRV (again, there
be set relative to inspiratory pressure (P-high). Figure 3-12 is no standardization of either terminology or symbology
illustrates the two different ways used to define inspiratory in this mode).
pressure and the four different ways to define pressure
support. Figure 3-13 illustrates the proposed solution to this Pmax. The Drager Evita XL, when set in volume-control
problem.43 In this proposal, the term inspiratory pressure modes, allows the operator to set the maximum pressure
is defined as the set change in airway pressure during (Pmax) that can be achieved during the delivery of a
inspiration relative to set end-expiratory airway pressure mandatory breath. The goal is to prevent pressure peaks
during pressure-control modes. while maintaining the set tidal volume. When the Pmax is
On some ventilators, inspiratory pressure rise is set rela- reached during a given inspiration, the ventilator switches
tive to atmospheric pressure rather than set end-expiratory from volume control to pressure control (dual targeting)
pressure. To distinguish this from inspiratory pressure as using the Pmax setting as the inspiratory pressure target. If
defined relative to PEEP, the term peak inspiratory pressure the set tidal volume cannot be reached in the set inspiratory
has been proposed.43 In contrast “peak airway pressure” is time, the ventilator will alarm.45
the measured peak airway pressure relative to atmospheric
pressure. Often, for a good pressure-control system, there Rise Time. The speed with which the airway pressure
is seemingly no difference between set peak inspiratory reaches the set inspiratory pressure is called the rise time.
pressure and measured peak airway pressure on the airway- (Rise time for flow can be set in the Maquet Servo i, but
pressure waveform during pressure-control modes. And this feature is rare on ventilators.) The rise time may be
even if the operator sees a transient small difference, this is set by the operator or automatically adjusted based on a
not considered clinically important in most nonalarm cases. computer algorithm (e500, Newport Medical Instruments
This leads clinicians to conceptually oversimplify what they Inc, Newport Beach, California). The name used to indicate
see and make the mistake of assuming inspiratory pressure pressure rise time varies by ventilator brand (e.g., inspiratory
Chapter 3 Basic Principles of Ventilator Design 79
Spontaneous breath
PS-PIP
Airway pressure
Airway pressure
IP PS-PEEP Pplt PS-PEEP
PIP PS-Patm Ppeak PS-Patm
PEEP PEEP
Spontaneous
Volume
Mandatory
Mandatory
Spontaneous
Flow
Flow
FIGURE 3-13 Idealized pressure, volume, and flow waveforms for pressure control and volume control illustrating the use of proposed conventions
for both set and measured airway pressures. IP, inspiratory pressure; PEEP, positive end-expiratory pressure; PIP, peak inspiratory pressure; Ppeak,
peak pressure; Pplt, plateau pressure; PS-Patm, pressure support relative to atmospheric pressure; PS-PEEP, pressure support relative to positive end
expiratory pressure; PS-PIP, pressure support relative to peak inspiratory pressure. (Reproduced, with permission, from Chatburn RL, Volsko TA.
Documentation issues for mechanical ventilation in pressure-control modes. Respir Care. 2010;55(12):1705–1716.)
slope, P-ramp, plateau%, and slope rise time). Adjusting slow rise time should be avoided. A more gradual rise may be
the rise time influences the synchronization between the needed in awake patients (for comfort) or patients with low
patient and the ventilator secondary to changes in the initial compliance to prevent pressure overshoot and premature
inspiratory flow rate. The lower the rise time, the faster the cycling of inspiration (Fig. 3-14).
pressurization rate46 and the higher the peak inspiratory
flow.47 A higher initial inspiratory flow rate may decrease
TIDAL VOLUME
the work of breathing but can lead to patient discomfort and
worse patient–ventilator synchrony. Conversely, too slow The operator is required to enter a tidal volume in any
a rise time may result in increased work of breathing and volume-control mode. This may be a direct setting or an indi-
longer mechanical inspiratory time, leading to a dissociation rect one by setting frequency or minute ventilation. The ven-
between patient breathing effort and the mechanical breath. tilator will control the tidal volume and the pressure will be
That is, the relation between work of breathing, respiratory the dependent variable. A tidal volume target, however, may
drive, and comfort with the duration of the rise time is not also be set when the mode uses adaptive targeting in pressure
proportional.46,48 Because rules for setting an optimal rise control (e.g., pressure-regulated volume control [PRVC] on
time are lacking, based on these studies, both very rapid and the Maquet ventilators).49 In such a case, inspiratory pressure
80 Part II Physical Basis of Mechanical Ventilation
A B C
2
1
(L/S)
Flow
0
–1
–2
2.0
1.5
1.0
Volume
(L)
0.5
–0.5
60
40
(cm H2O)
pressure
Airway
20
–20
60
Esophageal
40
(cm H2O)
pressure
20
–20
FIGURE 3-14 Examples of different pressure rise times in three breaths in pressure-support mode. A. Rise time is set very low, resulting in a lower
peak inspiratory flow. B. Rise time is set higher, resulting in a higher peak flow and shorter inspiratory time. C. Rise time is set very high, resulting
in “ringing” of airway pressure signal and peak flow that is uncomfortable to the patient, who exerts an expiratory effort and prematurely terminates
inspiration (indicated by the positive deflection of esophageal pressure). (Reproduced, with permission, from Macintyre NR. Patient-ventilator inter-
actions: optimizing conventional modes. Respir Care. 2011;56(1):73–81.)
is automatically adjusted between breaths by the ventilator to mode, when, by the equation of motion, what is being
achieve an average measured tidal volume equal to the oper- controlled is pressure during a breath. A caveat with this
ator set target. There are four basic ways ventilators deliver targeting scheme is that in the presence of the patient’s
a preset tidal volume (from least used to most commonly inspiratory efforts, the tidal volume may be higher than
used): set, and the support provided by the ventilator may be
inappropriately low.50,51
1. By measuring the volume delivered and using the signal
4. By controlling flow, the volume delivered is indirectly
in a feedback control loop to manipulate the volume
controlled. Because flow and volume are inverse functions
waveform.
of time (i.e., volume is the integral of flow and flow is the
2. By the displacement of a piston or bellows. An example
derivative of volume), controlling one controls the other.
of this is the Puritan Bennett LP10 home-care ventilator
In simple ventilators, there is no feedback signal for flow,
(piston) or some anesthesia ventilators (bellows).
just a known flow for an adjustable amount of inspira-
3. By controlling the inspiratory pressure within a breath
tory time. On more sophisticated ventilators, the operator
and automatically adjusting it between breaths to deliver
can regulate the shape of the inspiratory flow waveform.
a minimum set tidal volume. The volume delivered is
A square waveform will create higher peak airway pres-
targeted by a closed-loop algorithm, known as adaptive
sures and will require less time to deliver the set volume
pressure control (see Chapter 2). This targeting scheme
(which may result in lower mean airway pressures) than
is available in most modern critical care ventilators under
a descending ramp pattern.52–54 Some ventilators offer
multiple names (e.g., PRVC, autoFlow, VC+, APV).
one waveform (e.g., the Dräger Evita XL offers only the
A common confusion is that this is a volume-control
Chapter 3 Basic Principles of Ventilator Design 81
Flow Patient
PC (Adaptive targeting) Target tidal volume
set by operator
Volume
Airway Pressure
pressure
square waveform) others have more (e.g., the Hamilton tidal volume and frequency. The ventilator then monitors
Veolar offers 50% or 100% descending ramps, sinusoidal, the total minute ventilation as the sum of the minute ven-
and square).55 Most current ventilators only provide the tilations generated by mandatory and spontaneous breaths.
square waveform or a descending ramp profile. If the total minute ventilation is below the target value, the
mandatory breath frequency will increase. As long, however,
Figure 3-15 compares volume delivery between standard
as the spontaneous minute ventilation is at least equal to the
volume and pressure control modes versus modes using
target value, mandatory breaths will be suppressed. In this
adaptive pressure control.
way, the proportion of the total minute ventilation gener-
ated by spontaneous breaths can range from 0% to 100%.
As a result, MMV may be considered a mode of automatic
MINUTE VENTILATION
weaning.
In volume-control modes, the minimum minute ventilation Another version of MMV was used on the Hamilton
is set by entering the tidal volume and respiratory rate. This Veolar ventilator (now obsolete); the target minute venti-
assures that the patient will receive a minimum amount of lation was maintained by automatic adjustment of inspira-
ventilatory support. Some modes provide the option to enter tory pressure (adaptive pressure support). That mode was
a target minute ventilation (as a percent of the calculated replaced by ASV on newer Hamilton ventilators.49 ASV is the
minute ventilation for a given ideal body weight, adaptive- only commercially available mode to date that uses optimal
support ventilation [ASV]; e.g., Hamilton G5), while others targeting. It was first described by Tehrani in 1991.57 The
will calculate it from the entered tidal volume and respira- operator inputs the patient’s height and percent of minute
tory rate (mandatory minute volume [MMV]; e.g., Dräger ventilation to be supported (25% to 350%). The ventila-
Evita XL). The concept of automatically adjusting the venti- tor then calculates the ideal body weight and estimates the
lator settings to maintain a constant minute volume was first required minute alveolar ventilation assuming a normal
described by Hewlett and Plat in 1977.56 As implemented, dead space fraction. Next, an optimum frequency is calcu-
for example, on the Dräger Evita XL ventilator, MMV is a lated based on work by Otis et al9 that predicts a frequency
form of volume control–intermittent mandatory ventilation. resulting in the least mechanical work rate. The target
The operator presets the target minute ventilation by setting tidal volume is calculated as minute ventilation divided by
82 Part II Physical Basis of Mechanical Ventilation
Operation Operator enters a Operator enters Operator enters a set Adaptive pressure Pressure support
set rate and tidal a set rate and rate and tidal volume. control breaths is titrated based
volume. Patient tidal volume. Patient may breath with target tidal volume on expert rules to
may trigger breaths Patient may or without assistance. If and rate according achieve the range
above set rate. breath in between his minute ventilation to mathematical etPCO2.
mandatory breaths falls below minimum, model.
with or without then mandatory breaths
assistance. initiate at a set rate.
Control variable Volume Volume Volume Pressure Pressure
Breath sequence CMV IMV IMV IMV CSV
Minimum minute set VT × set f set VT × set f set VT × set f Targeted by ventilator Targeted by ventilator
ventilation based on operator- to maintain
entered body “comfort zone”
weight. based on VT, f, and
etPCO2.
Maximum Variable: VT × total f Variable: VT × total f Variable: VT × total f Variable but ventilator Variable but ventilator
minute will reduce support will reduce support
ventilation if patient attempts if patient attempts
to increase above to increase above
estimated minute estimated minute
ventilation ventilation
requirement. requirement.
Abbreviations: A/C, assist/control; ASV, adaptive support ventilation; etPCO2, end-tidal pressure of carbon dioxide; f, ventilatory frequency—total f reflects the sum of
machine- and patient-triggered breaths; MMV, mandatory minute volume; SIMV, synchronized intermittent mandatory ventilation; VT, tidal volume.
respiratory frequency (MV/f). In ASV, there are two breath mode.27 Second, the interface may add confusion. For exam-
patterns based on the patient’s respiratory effort. If there is ple, in the Dräger Evita XL, while on volume control, the
no patient effort, the ventilator delivers adaptive pressure- operator will need to set the inspiratory flow, the inspiratory
control ventilation; if there is patient effort, the patient time, and tidal volume, whereas on the Hamilton G5, the
receives adaptive pressure support. In both instances, the options are customizable in three different ways! (Hopefully,
inspiratory pressure within a breath is controlled to achieve all conducive to the same output.) The operator can enter
a target tidal volume.49 (a) the I:E and the percent pause in inspiration, (b) the peak
Table 3-3 summarizes the determinants of minimum and inspiratory flow and inspiratory time, or (c) the percent
maximum minute ventilation for some common modes. inspiratory time and plateau pause time. Underscoring that
knowledge of the device used is essential. Finally, to add to
the confusion, there are incorrect conclusions that some-
INSPIRATORY FLOW times permeate practice:
The inspiratory flow can be adjusted by the operator on 1. In pressure-control mode, the flow is controlled as a
most ventilators that provide volume-control modes (see descending ramp. In a pressure-controlled breath, the vol-
“Tidal Volume” above). In general, the ventilator operator ume and the flow are the manifestation of the respiratory
will choose a peak flow and may have some waveform pat- system characteristics (resistance and compliance) and
tern options (e.g., square waveform or descending ramp). the patient’s respiratory effort. If the patient is passive (no
Although these settings appear simple, there are several respiratory effort), the flow will decay exponentially (see
points that may cause differences in performance and inter- Fig. 3-7, A). If the patient has a respiratory effort, the flow
pretation of data. First, the ventilator uses a microprocessor pattern will be variable, according to the characteristics of
to control the delivery according to the preset tidal volume, the patient effort, the ventilator settings (inspiratory pres-
inspiratory time, flow pattern, pressure limits, and ventila- sure, pressurization algorithm, triggering, etc.), and the
tor-specific algorithms. During the breath, the flow delivery respiratory system characteristic. The only way to have a
is adjusted according to a closed-loop feedback mechanism standard descending ramp is to select that waveform and
and proprietary software.2 The consequence is a difference have the computer control the flow delivery in volume
in performance among ventilator brands, even in the same control.
Chapter 3 Basic Principles of Ventilator Design 83
2. The “autoflow” function adjusts the flow in a volume- In 1996, Davis et al52 tested the hypothesis that a descend-
controlled breath to the patient’s demand. Autoflow is ing ramp flow waveform is responsible for improvements in
available in Dräger Evita ventilators. It appears as an gas exchange during pressure-control ventilation for acute
add-on for three modes of volume-control ventilation lung injury. They compared volume control with a square or
(controlled mechanical ventilation [CMV] or intermittent descending ramp waveform to pressure control with a square
positive-pressure ventilation [IPPV], SIMV, and MMV). pressure waveform. Both pressure control and volume con-
This “add on” is defined in the manual as automatic trol with a ramp waveform provided better oxygenation at
regulation of the inspiratory flow adjusted to the changes lower peak airway pressure and higher mean airway pres-
in lung conditions and to the spontaneous breathing sure compared to volume control with the square-flow
demands.58,59 What this “add on” does is turn the mode waveform.
from a volume-control mode to an adaptive pressure- Polese et al61 compared square, sinusoidal, and descend-
control mode. This is the same as being on PRVC on ing ramp flow waveforms in patients after open heart
the Maquet ventilators. They all automatically adjust the surgery. They found that PaO 2 and PaCO 2 were not affected
inspiratory pressure to achieve a target tidal volume and by changes in waveform. Peak airway pressure was high-
because this is a pressure-controlled breath, the flow will est with the sinusoidal waveform while mean airway pres-
be variable (see “Tidal Volume” above). sure and total work of breathing were least with the square
waveform. Yang et al53 applied square, sine, and descending
The inspiratory flow setting has importance at differ-
ramp flow waveforms to patients with chronic obstructive
ent levels. The work of breathing is related to the peak flow
pulmonary disease (COPD) and found that the descend-
and the pressurization rate. The balance between patient
ing ramp reduced inspiratory pressure, dead space-to-tidal
and ventilator work of breathing will be affected by the
volume ratio, and PaCO 2, but increased alveolar–arterial
inspiratory flow setting. In regards to cycling, high flows
oxygen tension difference with no change in arterial oxygen-
can lead to high peak inspiratory pressures (peak inspira-
ation or hemodynamic variables.
tory pressure [PIP] is directly proportional to resistance,
Our own experience is that many clinicians prefer the
the higher the flow, the higher the PIP), which may lead to
descending ramp flow waveform when using volume control
reaching the pressure or flow-cycling threshold and ending
modes, with the observation that patients tend to be more
the breath prematurely.59 But a more practical issue is this:
comfortable, perhaps because of the higher flow earlier in
does the flow-wave shape itself have any effect on patient
the inspiratory phase.
outcome? Like most other questions about ventilator set-
Figure 3-16 illustrates an algorithm that can be used
tings affecting patient outcome, after more than 30 years
to adjust inspiratory flow to improve patient–ventilator
of research on this particular subject we still do not know
synchrony.62
the answer.
Studies from the early 1960s to early 1980s produced
conflicting results, prompting Al-Saady and Bennett to PERCENT SUPPORT
design a better-controlled study, keeping tidal volume,
minute ventilation, and I:E ratio constant.60 They dis- Proportional-assist ventilation (PAV)63 delivers pressure-
covered that compared to a constant inspiratory flow, a control breaths with a servo targeting scheme (see Chapter
descending ramp flow (what they and many subsequent 2).49 The pressure applied is a function of patient effort: the
authors have called “decelerating flow”) resulted in a lower greater the inspiratory effort, the greater is the increase in
peak airway pressure, total respiratory resistance, work of applied pressure (Fig. 3-17). The form of PAV implemented
inspiration, dead space-to-tidal volume ratio, and alveo- on the Dräger Evita XL ventilator (called proportional pres-
lar–arterial oxygen tension gradient. They also noted an sure support) requires the operator to input desired assis-
increase in compliance and partial pressure of arterial oxy- tance values for elastance and resistance. PAV implemented
gen (PaO 2) with no changes in partial pressure of arterial on the Puritan Bennett 840 ventilator (called PAV +) uses a
carbon dioxide (PaCO 2) or any hemodynamic variables. In different algorithm. It automatically calculates the resistance
1991, Rau et al compared peak and mean airway pressure of the artificial airway, and combines resistance and elas-
for seven different inspiratory flow waveforms (including tance such that the operator enters only a single value repre-
square, ascending and descending ramps, and sinusoidal) senting the percentage work of breathing to be supported.64
under three different lung model conditions.54 For all mod- The design differences between proportional pressure sup-
els, the descending ramp flow waveform produced the low- port and PAV + lead to significant performance differences.65
est peak and the highest mean airway pressures, whereas
the ascending ramp produced the opposite: the highest
NEURALLY ADJUSTED
peak and lowest mean values. When compliance was low,
VENTILATORY SUPPORT LEVEL
mean airway pressure increased as peak airway pressure
increased. When resistance was high, peak airway pres- NAVA is a mode that applies airway pressure proportion-
sure was more affected by the peak flow setting than the ately to patient effort based on the voltage recorded from
waveform setting. diaphragmatic activity. The “NAVA level” is the constant
84 Part II Physical Basis of Mechanical Ventilation
Identify candidate
patient
Assess patient
Add PEEP
5 cm H2O or
increase by PEEPi > 5 Add PEEP
No Yes
1 cm H2O not cm H2O 75–80% of PEEPi
to exceed
8 cm H2O
No Al > 10%
Yes
Increase Pressure
No
inspiratory flow control mode
Yes
Increase
pressurization rate
Decrease
Yes Time cycled
inspiratory time
No
No
Decrease flow
cycle threshold
(% peak flow)
FIGURE 3-16 Algorithm for improving patient–ventilator synchrony. AI, asynchrony index, percent of inspiratory efforts that failed to trigger a
breath; COPD, chronic obstructive pulmonary disease; PEEPi, intrinsic PEEP (aka auto-PEEP). (Modified from, with permission, Sassoon CSH.
Triggering of the ventilator in patient-ventilator interactions. Respir Care. 2011;56(1):39–48.)
Chapter 3 Basic Principles of Ventilator Design 85
Proportional-assist ventilation
Volume Pressure
Patient
Flow, pressure, and volume delivered
Ventilator measuring effort Flow by the ventilator are adjusted
respiratory system proportionally to patient effort
characteristics
FIGURE 3-17 Pressure, volume, and flow waveforms for proportional assist ventilation.
Paw Paw
[cm H2O] [cm H2O]
PEEP PEEP
Flow Flow
[L/sec] [L/sec]
Time [s]
Time [s]
Edi Edi
[µv] [µv]
FIGURE 3-18 Airway pressure, flow, and electrical diaphragmatic activity curves in pressure support (left) and in neurally adjusted ventilatory assist
(right). Edi, electrical activity of the diaphragm; PEEP, positive end-expiratory pressure, Td, trigger delay; Tiex, inspiratory time in excess; Tin, neural
inspiratory time; Tiv, ventilator pressurization time. (Reproduced, with permission, from Piquilloud L, Vignaux L, Bialais E, et al. Neurally adjusted
ventilatory assist improves patient–ventilator interaction. Intensive Care Med. 2011;37(2):263–271.)
In pressure-control modes, the operator presets the inspi- effect on oxygenation and ventilation by increasing mixing
ratory time directly for mandatory breaths. Thus, prolong- time and decreasing dead space.69,70
ing the inspiratory time causes the ventilator to decrease
the expiratory time, possibly resulting in air trapping, larger I:E Ratio and Duty Cycle. I:E is the ratio of inspiratory time
tidal volumes, or cycle asynchrony. One must remember to expiratory time (Fig. 3-20).
that the effect on tidal volume of the inspiratory time in a
TI
pressure-control breath will depend on the respiratory sys- I : E = TI : TE = (9)
tem characteristics (i.e., the time constant). Thus, a patient TE
with a long time constant (high compliance and/or high
The I:E can also be described as the duty cycle or percent
resistance) will require a longer inspiratory time to achieve
inspiration. In engineering, the duty cycle is defined as the
full pressure equilibration, cessation of flow, and complete
time spent in active state as a fraction of the total time. In
tidal volume delivery.
mechanical ventilation, the active state is the inspiratory
Figure 3-16 illustrates an algorithm that can be used
time, and the total time is the sum of the inspiratory and
to adjust inspiratory time to improve patient–ventilator
expiratory times. It is expressed as a percentage. The larger
synchrony.62
the percentage, the longer the inspiratory time in relation to
the total cycle time.
Inspiratory Pause. The inspiratory pause is the period
during which flow ceases but expiration has not begun TI
Duty Cycle = × 100 (10)
(see inspiratory time). The expiratory valves are closed TI + TE
during this period. The inspiratory pause time is part of
the inspiratory time. It is also named plateau time (PB 840, One can convert one to the other by the following formula:
Covidien, Mansfield MA), Pause time (Servo i, Maquet,) Duty Cycle
or Pause (G5, Hamilton Medical). When set directly, pause I:E = (11)
100 − Duty Cycle
time may be entered in seconds or as a percentage of the
inspiratory time. When it is activated, most ventilators Example: A duty cycle of 50% is an I:E of 1:1, a duty cycle of
will display a plateau pressure (i.e., static inspiratory hold 33% is an I:E 1:2.
pressure). Increasing the inspiratory pause time will increase The relevance of I:E is highlighted in the context of
the mean airway pressure and thus the time the lung is the time constant. The time constant is a measure of how
exposed to volume and pressure. This may have a positive quickly the respiratory system can passively fill or empty in
A
B C
IT Expiratory time
Inspiration Expiration
FIGURE 3-20 Divisions of the inspiratory and expiratory periods. A volume-controlled breath is depicted. A. End of inspiratory flow. B. Start of
expiratory flow. C. End of expiratory flow. IFT, Inspiratory flow time; IPT, inspiratory pause time; IT, inspiratory time.
88 Part II Physical Basis of Mechanical Ventilation
TABLE 3-4: EFFECT OF LUNG CONDITION ON TIME CONSTANT AND EXPIRED VOLUME
0 0 0 0 500 0 100
1 0.780 0.510 1.000 184 63%a 37%
2 1.560 1.020 2.000 68 86% 14%
3 2.340 1.530 3.000 25 95% 5%
4 3.120 2.040 4.000 9 98% 2%
5 3.900 2.550 5.000 3 99% 1%
Abbreviations: ARDS, acute respiratory distress syndrome; COPD, chronic obstructive pulmonary disease.
a
The exact value is (1 − e−1) × 100%.
response to a step change in transrespiratory pressure.23 It is safety feature (i.e., an alarm setting) with current modes of
calculated as the product of resistance and compliance. The ventilation. When a preset high-pressure alarm threshold
value obtained is the time that takes to achieve 63% of steady is crossed, the ventilator will cycle the ventilator. The goal
state. This percent change remains a constant, regardless of is to prevent the patient from exposure to hazardous pres-
the combination of resistance and compliance. It follows that sures. Pressure cycling without an alarm is the normal oper-
each time constant will lead to a 63% decrease or increase ational state for some devices (e.g., VORTRAN automatic
in volume. In Table 3-4, one can see the difference among resuscitator).
time constants for different lung conditions. In COPD, the
time constant is longer so the time required for exhalation is
VOLUME
longer than for patients with acute respiratory distress syn-
drome. This table demonstrates the effect of the time con- Volume cycling occurs when a preset volume is reached.
stant during passive exhalation using previously published71 This occurs when the operator sets a tidal volume in vol-
expiratory time constants for three conditions (normal lung ume-control modes. Volume cycling implies that inspired
was 0.78 seconds, for acute respiratory distress syndrome volume is monitored by the ventilator’s control system dur-
0.51 seconds, and for COPD 1 second). In this example, expi- ing inspiration and compared to a threshold value (the set
ration starts from a lung volume of 500 mL above functional tidal volume). But on some ventilators, despite the setting
resting capacity. When expiratory time equals one time con- of a tidal volume, the actual cycle variable is time, that is,
stant, 63% of the tidal volume will be exhaled, leaving 37% of the time it takes to deliver the set tidal volume with the set
the tidal volume yet to be exhaled. inspiratory flow. Manufacturers seldom make this distinc-
The I:E ratio can be an operator-entered value, or just dis- tion clear in the operator’s manual.
played as a calculated value based on common scenarios for Volume cycling can also be found as a default safety
mandatory breaths: feature. In PAV + (Covidien PB 840 ventilator), one of the
cycling criteria is volume. Once the operator-preset high
• Preset I:E ratio and frequency.
inspired tidal volume limit is reached, the ventilator cycles
• Preset inspiration time (TI in seconds) and frequency
the breath and alarms.
(breaths/min). The frequency sets the ventilatory period
(1/f ) and the expiratory time is the period minus TI :
FLOW
I : E = TI : [(60 ÷ rate) − TI ] (12)
• Expiratory time and inspiratory time are fixed: Flow cycling occurs when a preset flow or percentage of
the peak flow is reached for pressure-control breaths. Flow
I : E = TI : TE (13) cycling is most commonly found with the pressure-support
Note: some ventilators will synchronize inspiration and/ mode but can be added as an “advanced setting” in other
or expiration of a mandatory breath if the patient effort is pressure-control modes on at least one ventilator (Avea,
detected in a trigger/cycle window (e.g., SIMV or APRV), CareFusion). The flow-cycling threshold preset by the oper-
which may alter the I:E from the expected value based on ator has been given many names: expiratory trigger sensi-
settings. tivity (Hamilton ventilators); trigger window (Engstrom
Ohmeda); inspiratory termination peak inspiratory flow
(Dräger Evita XL); expiratory threshold (Newport); flow ter-
PRESSURE
mination (Pulmonetics LTV ventilators); PSV cycle (Avea,
Pressure cycling occurs when the ventilator reaches a pre- CareFusion); inspiratory cycle off (Servo i, Maquet); Ecycle
set peak airway pressure. Pressure cycling is most often a (V200 respironics); and E sens (PB 840, Puritan Bennett).
Chapter 3 Basic Principles of Ventilator Design 89
During a breath in the pressure-support mode, the flow. Flow and volume are not directly controlled during this
ventilator provides enough initial flow to achieve the set period on any current ventilator. The most common value
inspiratory pressure. The initial flow is high and then decays controlled is pressure relative to atmospheric pressure (zero-
exponentially. Some ventilators have a preset default value gauge pressure).
for flow cycling (range: 5% to 30% of peak inspiratory flow);
others allow the operator to adjust it (range: 1% to 80% of
peak inspiratory flow). Only one device (e500, Newport Positive End-Expiratory Pressure
Medical, Costa Mesa, CA) has automatic adjustment of the
flow-cycling criteria. This device has a proprietary algorithm The PEEP is established by the ventilator exhalation valve. A
called FlexCycle. It will change the cycle criterion from common source of confusion is the term continuous positive
10% to 50% of peak flow based on measurements of airway airway pressure versus PEEP. Continuous positive airway
pressure, the expiratory time constant, and expert-based pressure is generally considered to be a mode on mechanical
rules applied through a closed-loop system.72 ventilators (or a mode of treatment for sleep apnea), whereas
A default cycle criterion of 25% to 30% of the peak PEEP is the elevation of the baseline pressure during any
flow seems inappropriate as a “fit all” measure. The goal mode of ventilation and is generally a setting for a mode. Until
of adjusting the flow-cycling criterion is to avoid expira- recently, the selection of PEEP has been a relatively arbitrary
tory asynchrony.59 In expiratory asynchrony, the ventila- process and the meaning of “optimum PEEP” is debatable.77
tor ends inspiration before or after the patient inspiratory Now, Hamilton Medical has developed the INTELLiVENT
effort. We must remember that flow is a manifestation of the system for the G5 ventilator that uses an algorithm for auto-
respiratory system characteristics, respiratory muscle effort matic targeting of PEEP and FIo2. A closed-loop algorithm
(inspiratory and expiratory) and the integrity of the lung- based on expert rules defines the response of the ventilator
ventilator circuit. If the respiratory system has a prolonged to measured ventilation variables, end-tidal carbon dioxide
time constant, a standard flow-termination criterion may be and pulse oximetry.
inappropriate as it will prolong inspiration. That may be the
case for patients with COPD, where the standard criterion of P-Low. P-low is one of the settings entered for so-called
25% may be too low, and lead to expiratory asynchrony and “bilevel” modes like APRV (Fig. 3-21). P-low is just another
increased work of breathing.73,74 Finally, a leak in the ventila- name for PEEP. Similar to PEEP, the settings are dependent
tor circuit (mask) or in the patient (endotracheal cuff or a on the user. There is, however, a large discrepancy with
bronchopleural fistula) may lead to lack of decay in the flow the objective of PEEP. In APRV, P-low is set to zero.78 The
curve and thus asynchrony.72 goal is to maintain lung recruitment with the use of auto-
Figure 3-16 illustrates an algorithm that can be used to PEEP induced by short T-low settings. P-low can also be set
adjust the flow-cycle threshold to improve patient–ventilator based on the biphasic model,79 where complete exhalation is
synchrony.62 allowed and P-low is then set with the same goals as PEEP.
Pressure
thereby generating air trapping.29 Adjusting T-low on the Level 4 events are based entirely on patient condition.
ventilator to manually maintain %PEF at 50% to 75% may They may include events such as changes in gas exchange,
be a tedious process, which may seem simple on paper, but dead space, oxygenation, and cardiovascular functions.
in a spontaneously breathing patient can become a true Ventilators generally monitor these events and exter-
challenge. Newer ventilators, like the Dräger Evita Infinity nal monitors are required for alarms (the exception being
V500, have attempted to make the process easier by allowing exhaled carbon dioxide-level alarms built into the ventilator
the operator to set a trigger threshold based on a percentage display).
of peak expiratory flow. Currently, ventilators do not display alarm settings
as levels of priority. Instead, they tend to lump them all
together on one screen that shows alarm limits and controls
Alarms for changing them (Fig. 3-22). How to set alarm thresholds
is a complicated topic that has been studied but for which
Ventilator alarms bring unsafe events to the attention of the little information is available regarding mechanical venti-
clinician. Events are conditions that require clinician aware- lation. The goal is to minimize false alarms and maximize
ness or intervention. Events can be classified according to true alarms. A high false alarm rate leads to clinician habit-
their level of priority.81 uation and can also lead to inappropriate responses. In a
Immediately life-threatening events are classified as Level recent study of an intensive care unit, 1214 alarms occurred
1. They include conditions like insufficient or excessive gas and 2344 tasks were performed. On average, alarms
delivery to the patient, exhalation valve failure, control cir- occurred six times per hour; 23% were effective, 36% were
cuit failure, or loss of power. Level 1 alarm indicators should ineffective, and 41% were ignored.82 In another intensive
be mandatory (cannot be turned off by the operator), redun- care unit study, alarms occurred at a rate of six per hour.
dant, and noncanceling. Approximately 40% of the alarms did not correctly describe
Level 2 events range from mild irregularities in machine the patient condition and were classified as technically false;
function to dangerous situations that could threaten patient 68% of those were caused by manipulation. Shockingly, only
safety if left unattended. Some examples are failure of the 885 (15%) of all alarms were considered clinically relevant.83
air–oxygen blending system, inadequate or excessive PEEP, Although these studies did not address mechanical venti-
autotriggering, circuit leak, circuit occlusion, inappropriate lator alarms specifically, it is not hard to imagine similar
I:E ratio, and failure of the humidification system. Alarms in results for such a study.
this category may be self-canceling (i.e., automatically turned Ventilator alarms are usually set by the operator as either
off if the event ceases) and are not necessarily redundant. an arbitrary absolute value or a percentage of the current
Level 3 events indicate changes in the amount of ventila- value. Examples would be airway-pressure alarms (high
tor support provided to the patient consequent to changes in and low) set at the current value plus or minus 5 cm H2O
the patient’s ventilatory drive or respiratory system mechan- or low and tidal volume/minute ventilation set at plus or
ics and the presence of auto-PEEP. These events often trigger minus 25% of the current value.81 The problem is that the
the same alarms as Levels 1 and 2. parameters for which alarms are important, and these three
Chapter 3 Basic Principles of Ventilator Design 91
FIGURE 3-22 Alarm screen from the G5 ventilator. (Reproduced, with permission, from Hamilton Medical.)
in particular, are highly variable, with significant portions so that, for example, significant events or gradual changes in
at extreme values.84 Thus, limits set as absolute values or patient condition can be identified (Fig. 3-24). In addition,
percentages may reduce safety for some extreme values ventilators often provide an alarm log, documenting such
while increasing nuisance events for other values. An alter- things as the date, time, alarm type, urgency level, and events
native approach might be a type of “smart alarm,” whereby associated with alarms, for example, when activated and
the alarm limits are automatically referenced to the current when canceled. Such a log could be invaluable in the event of
value of the parameter such that extreme values have tighter a ventilator failure leading to a legal investigation.
limits. Further research is needed to identify optimiza-
tion algorithms (i.e., minimize both harmful and nuisance
WAVEFORMS AND LOOPS
events).
Many ventilators display waveforms (sometimes called “sca-
lars”) of airway pressure, volume, and flow as functions of
VENTILATOR OUTPUTS (DISPLAYS) time. Such displays are useful for identifying the effects of
changes in settings or mechanics on the level of ventilation.85
Display Types They are also very useful for identifying sources of patient–
ventilator asynchrony, such as missed triggers, flow asyn-
Ventilator output displays represent the values of monitored chrony, and delayed or premature cycling.86 They can also
parameters that result from the operator settings. There are display one variable against another as an x-y or “loop” dis-
four basic ways to present the monitored data: as numbers, as play. The most common loop displays show pressure on the
waveforms, as trend lines, and in the form of abstract graphic horizontal axis and volume on the vertical axis, or volume
symbols. on the horizontal axis and flow on the vertical axis. Pressure-
volume loop displays are useful for identifying optimum
PEEP levels (quasistatic loops only) and over distension.
NUMERIC VALUES
Flow-volume loops are useful for identifying the response to
Data are most commonly represented as numeric values such bronchodilators. Figure 3-25 is an example of a composite
as FIo2, peak, plateau, mean and baseline airway pressures, display showing numeric values, waveforms, and loops.
inhaled/exhaled tidal volume, minute ventilation, and fre-
quency. Depending on the ventilator, a wide range of calcu-
ADVANCED GRAPHICS
lated parameters may also be displayed including resistance,
compliance, time constant, airway occlusion pressure at 0.1 As ventilators have become more complex, their displays have
second (P0.1), percent leak, I:E ratio, and peak inspiratory/ become more confusing and difficult to use. A recent trend is
expiratory flow (Fig. 3-23). to move away from the traditional display screens in favor of
a more integrative approach using creative graphic elements.
For example, one study showed that observers detected and
TRENDS
treated obstructed endotracheal tubes and auto-PEEP prob-
Many ventilators provide trend graphs of just about any lems faster with graphical rather than conventional displays.
parameter they measure or calculate. These graphs show how They also reported significantly lower subjective workloads
the monitored parameters change over long periods of time, using the graphical display.87
92 Part II Physical Basis of Mechanical Ventilation
FIGURE 3-23 Digital display of monitored and calculated parameters from the Newport e360 ventilator. (Reproduced, with permission, from
Newport Medical.)
Hamilton Medical was the first to make use of innova- pre-spontaneous breathing trial (SBT), SBT in progress,
tive picture graphics on their G5 ventilator. They created and post-SBT phase.
a graphic representation of the lungs, called a “dynamic Dräger Medical recently introduced a similar graphic
lung panel,” that visually displays information about resis- display called “Smart Pulmonary View.” The shapes of the
tance and compliance by the shape and color of the lungs graphic elements quickly indicate relative values of respira-
and airways (Fig. 3-26). This panel is supplemented by a tory system resistance and compliance as well as the balance
unique graphic, called the “vent status panel,” which dis- between mandatory and spontaneous breaths (Fig. 3-27).
plays key parameters (e.g., oxygenation, ventilation, and Digital values are also displayed.
spontaneous breathing activity). Furthermore, the display
shows when each item is in or out of an acceptable zone
and for how long. This makes weaning status easy to iden- THE FUTURE
tify. Preliminary data88 suggest that this display reduces the
time required for clinicians to identify common problems,
for example, normal, restrictive, and obstructive lungs;
Better Operator Interfaces
occluded endotracheal tube, right main-stem intubation, As modes have become more complex, the operator
interfaces on ventilators with computerlike displays has
become cumbersome. Multiple options for control settings
tend to get lost in layers of different screen views. Worse,
screen views are often customizable such that if strict con-
trol is not exerted by an individual hospital department,
each ventilator will be “stylized” by individual operators
and chaos will ensue. Clearly, flexibility is a double-edged
sword.
Very few studies have been published on ease of use or
the problems with current displays. We need to identify
optimal ways for ventilator displays to provide three basic
functions: to allow input of control and alarm parameters,
to monitor the ventilator’s status, and to monitor the ven-
tilator–patient interaction status. There is a long way to
go before the user interface provides an ideal experience
FIGURE 3-24 Trend display from the Avea ventilator. (Reproduced, with these functions. This may be a fruitful area of future
with permission, from CareFusion.) research.6,8
Chapter 3 Basic Principles of Ventilator Design 93
SIMV Adult
AutoFlow
.500 25 10 5
0 Main
.300 Single Vol.% FIo2
blech
-25
Data...
.100
0
–10 0 20 40 cm H2O
-50
0 .200 .400 .600 L
Ref.
35 Special
Flow Procedure...
L/min L/min MV
50
25
4.5
0
L/min MVspn
–25
–50 0
.700 L bpm Ftotal
V
.500
10
.300
L VTi
.100
0
0 2 4 6 8 10 12 14 16 Sec
.450
Apnea-vent. On 1.0: 2.5
35 .450 1.7 10 5 0
Ext.
I:E 02 VT Tinsp f PEEP Psupp Int.
FIGURE 3-25 An example of both scalar and loop displays. (Reproduced with permission from Draeger Medical GmbH, Luebeck, Germany.)
Better Patient Interfaces Certainly, humidification systems using heated wires and
automatic-temperature control have evolved, but we still
The interface between a modern ventilator and the patient are not capable of measuring and directly controlling a pri-
is a piece of plastic tubing, that is, the “patient circuit,” mary variable of gas conditioning: humidity. Indeed, after
whose design has not changed much in several decades. all this effort at evolving humidification systems, there are
data to show that simple, unheated circuits provide better
humidification of inspired gas.89 In addition, the compli-
ance of the patient circuit degrades the accuracy of flow
delivery and must be “compensated” for by complex math-
ematical algorithms. It seems to us that a major revolu-
tion in patient-interface design would be to simply make
the patient circuit a permanent part of the ventilator and
treat water molecules the way we treat molecules of oxygen,
nitrogen, helium, and nitric oxide. But to do this, ventilator
manufacturers would have to merge with humidifier manu-
facturers and collaborate in systems design rather than see-
ing the patient circuit and humidifier as devices separate
from the ventilator (see Chapter 2).
12:58:43
VC-SIMV
PS
AutoFlow
45.7
Cdyn 45.7 Take
reference
11-Nov-2010
120 12:34:25 Rpat mbar/L/s
mL/mbar R 7.6
VT
Spon
Mand
Cdyn 79
9.3
VTe mL
D
760
MVe L/min
RR
C
8.88 12.00
2.00
A FIo2 Vol%
21
FIGURE 3-27 Example of picture graphic display from the Dräger Evita Infinity V500 ventilator showing the Smart Pulmonary View. A. The move-
ment of the diaphragm indicates synchronized mandatory breaths or supported (triggered) breaths. B. The blue line around the trachea indicates the
resistance R. The higher the resistance, the thicker the line. The numeric value is also displayed. C. The blue line around the lungs indicates the com-
pliance Cdyn. The higher the compliance, the thinner the line. The numeric value is also displayed. D. Diagram displaying the relationship between
spontaneous breathing and mandatory ventilation. The following parameters are displayed in different colors: spontaneous tidal volume (VT spon),
spontaneous respiratory rate (RR spon), mandatory tidal volume (VT mand), and mandatory respiratory rate (RR mand). (Reproduced, with permis-
sion, from Draeger Medical GmbH, Luebeck, Germany.)
commercial applications, is to develop “closed-loop” target- CO2 monitoring and pulse oximetry. These extra data,
ing systems based on mathematical models of physiologic along with advanced targeting software algorithms, allow
processes, or artificial intelligence, or combinations thereof, the ventilator to automatically select and adjust minute
with the goal of automating the moment-to-moment adjust- ventilation, PEEP, and FIo2. This makes INTELLiVENT
ment of ventilator output to patient needs. The best example “… the world’s first complete closed-loop ventilation solu-
so far is a mode called INTELLiVENT-ASV (G5 ventilator, tion that offers automated adjustment of oxygenation and
Hamilton Medical) and is currently available only in Europe. ventilation.”90
This mode is an improvement on the optimal target- Along with the new targeting systems, this mode also
ing scheme that is the basis of the mode called ASV (see provides a unique operator interface that Hamilton refers
Chapter 2). Like ASV, INTELLiVENT-ASV is a form of to as the “Ventilation Cockpit,” an apparent reference to the
pressure control intermittent mandatory ventilation using “autopilot” feature in airplanes. The interface is designed
adaptive-pressure targeting to automatically adjust inspi- to facilitate understanding complex information in a visu-
ratory pressure to maintain a target tidal volume, which, ally intuitive way. In addition to displaying the usual digi-
in turn, is selected by an optimization model. An “optimal” tal parameters and waveforms, the new mode offers several
targeting scheme attempts to either maximize or minimize other screens. The “Dynamic Lung” screen integrates data
some performance metric.49 In the case of ASV, the venti- on lung mechanics, end-tidal carbon dioxide (PETCO2),
lator attempts to select a tidal volume and frequency (for and pulse oximetry (SpO2), and offers a metric called the
passive ventilation) that minimizes the work rate of ven- “heart–lung interaction” index (Fig. 3-28). A graphic ele-
tilation for the patient’s particular state of lung mechan- ment called the “Ventilation Map” plots PETCO2 against peak
ics. As the lung mechanics change, the ventilatory pattern airway pressure as shown in Figure 3-29. Another display,
changes. ASV requires that the operator input the patient’s the “Oxygenation Map,” is very similar to the Ventilation
weight, however, so that the ventilator can calculate an esti- Map: it provides detailed information about the oxygen-
mated minute ventilation requirement. The operator must ation status based on the major physiologic input, as mea-
also manage PEEP and FIo2. INTELLiVENT-ASV takes sured by pulse oximetry (SpO2), and the resulting treatment
ASV a step further by adding input data from end-tidal (PEEP/FIo2).
Chapter 3 Basic Principles of Ventilator Design 95
REFERENCES
1. McPherson SP, Spearman CB. Respiratory Therapy Equipment. St.
Louis, MO: CV Mosby; 1977.
2. Sanborn WG. Microprocessor-based mechanical ventilation. Respir
Care. 1993;38(1):72–109.
3. Morch ET. History of mechanical ventilation. In: Kerby RR, Smith RA,
Desautels DA, eds. Mechanical Ventilation. New York, NY: Churchill
FIGURE 3-28 The operator interface of the Hamilton G5 ventilator Livingstone; 1985:1–58.
with INTELLiVENT-ASV option, called the “Ventilation Cockpit.” This 4. Russell IF, Ross DG, Manson HJ. Fluidic cycling devices for inspira-
screenshot shows the “Dynamic Lung” display including the “Heart– tory and expiratory timing in automatic ventilators. J Biomed Eng.
Lung Interaction” index. (Reproduced, with permission, from Chatburn 1983;5(3):227–234.
RL. Computer control of mechanical ventilation. Respir Care. 2004; 5. Chatburn RL. Classification of mechanical ventilators. In: Branson
49:507–515.) RD, Hess DR, Chatburn RL, eds. Respiratory Care Equipment.
Philadelphia, PA: Lippincott Williams & Wilkins; 1999:359–393.
6. Uzawa Y, Yamada Y, Suzukawa M. Evaluation of the user interface
simplicity in the modern generation of mechanical ventilators. Respir
Early studies of INTELLiVENT show that compared to Care. 2008;53(3):329–337.
ASV, patients ventilated with INTELLiVENT spent more 7. Lasocki S, Labat F, Plantefeve G, et al. A long-term clinical evaluation
of autoflow during assist-controlled ventilation: a randomized con-
time with optimal ventilation and less time with nonsecure trolled trial. Anesth Analg. 2010;111(4):915–921.
ventilation. In addition, INTELLiVENT delivered lower vol- 8. Vignaux L, Tassaux D, Jolliet P. Evaluation of the user-friendliness of
umes and pressures for equivalent gas exchange.91 In another seven new generation intensive care ventilators. Intensive Care Med.
2009;35(10):1687–1691.
preliminary study, patients managed with INTELLiVENT 9. Otis AB, Fenn WO, Rahn H. Mechanics of breathing in man. J Appl
spent less time with nonsecure and nonoptimal ventilation Physiol. 1950;2(11):592–607.
(3%) compared to conventional ventilation (47%, P = 0.03) 10. Loring SH. Mechanics of the lungs and chest wall. In: Marini JJ,
after cardiac surgery.92 Slutsky AS, eds. Physiological Basis of Ventilatory Support. New York,
NY: Marcel Deckker; 1998:177–208.
We speculate that modes of the future will continue this 11. Rodriquez D Jr, Blakeman TC, Dorlac W, et al. Maximizing oxygen
trend toward automation and include protocols for auto- delivery during mechanical ventilation with a portable oxygen con-
matic weaning for various populations of patients. They will centrator. J Trauma. 2010;69 Suppl 1:S87–S93.
12. Barash PG, Cullen BF, Stoelting RK, et al, eds. Clinical Anesthesia. 6th
provide means for communication with electronic medical ed. Philadelphia, PA: Lippincott Williams & Wilkins; 2009.
13. Hess DR, Fink JB, Venkataraman ST, et al. The history and physics of
heliox. Respir Care. 2006;51(6):608–612.
14. Tassaux D, Jolliet P, Thouret JM, et al. Calibration of seven ICU venti-
lators for mechanical ventilation with helium-oxygen mixtures. Am J
Respir Crit Care Med. 1999;160(1):22–32.
15. Berkenbosch JW, Grueber RE, Dabbagh O, McKibben AW. Effect of
helium-oxygen (heliox) gas mixtures on the function of four pediatric
ventilators. Crit Care Med. 2003;31(7):2052–2058.
16. Brown MK, Willms DC. A laboratory evaluation of 2 mechanical
ventilators in the presence of helium-oxygen mixtures. Respir Care.
2005;50(3):354–360.
17. Venkataraman ST. Heliox during mechanical ventilation. Respir Care.
2006;51(6):632–639.
18. Imanaka H, Hess D, Kirmse M, et al. Inaccuracies of nitric oxide
delivery systems during adult mechanical ventilation. Anesthesiology.
1997;86(3):676–688.
19. Montgomery FJ, Berssenbrugge AD. Inhaled nitric oxide delivery and
monitoring. J Clin Monit Comput. 1999;15(5):325–335.
20. Sydow M, Bristow F, Zinserling J, Allen SJ. Variation of nitric oxide
concentration during inspiration. Crit Care Med. 1997;25(2):
365–371.
21. Sinderby C, Navalesi P, Beck J, et al. Neural control of mechanical ven-
tilation in respiratory failure. Nat Med. 1999;5(12):1433–1436.
FIGURE 3-29 The operator interface of the Hamilton G5 ventilator 22. Nikischin W, Gerhardt T, Everett R, et al. Patient-triggered ventilation:
with INTELLiVENT-ASV showing details of ventilation and oxygen- a comparison of tidal volume and chest wall and abdominal motion as
ation management. trigger signals. Pediatr Pulmonol. 1996;22(1):28–34.
96 Part II Physical Basis of Mechanical Ventilation
23. Chatburn RL. Fundamentals of Mechanical Ventilation. Cleveland 48. MacIntyre NR. Patient-ventilator interactions: optimizing conven-
Heights, OH: Mandu Press; 2003. tional ventilation modes. Respir Care. 2011;56(1):73–84.
24. John J, Bjorklund LJ, Svenningsen NW, Jonson B. Airway and body 49. Chatburn RL, Mireles-Cabodevila E. Closed-loop control of mechani-
surface sensors for triggering in neonatal ventilation. Acta Paediatr. cal ventilation: description and classification of targeting schemes.
1994;83(9):903–909. Respir Care. 2011;56(1):85–102.
25. Bernstein G. Patient-triggered ventilation using cutaneous sensors. 50. Mireles-Cabodevila E, Chatburn RL. Work of breathing in adap-
Semin Neonatol. 1997;2(2):89. tive pressure control continuous mandatory ventilation. Respir Care.
26. Chatmongkolchart S, Williams P, Hess DR, Kacmarek RM. Evaluation 2009;54(11):1467–1472.
of inspiratory rise time and inspiration termination criteria in new- 51. Jaber S, Delay JM, Matecki S, et al. Volume-guaranteed pressure-sup-
generation mechanical ventilators: a lung model study. Respir Care. port ventilation facing acute changes in ventilatory demand. Intensive
2001;46(7):666–677. Care Med. 2005;31(9):1181–1188.
27. Ferreira JC, Chipman DW, Kacmarek RM. Trigger performance of 52. Davis K Jr, Branson RD, Campbell RS, Porembka DT. Comparison of
mid-level ICU mechanical ventilators during assisted ventilation: a volume control and pressure control ventilation: is flow waveform the
bench study. Intensive Care Med. 2008;34(9):1669–1675. difference? J Trauma. 1996;41(5):808–814.
28. Marchese A, Sulemanji D, Chipman D, et al. Performance of current 53. Yang SC, Yang SP. Effects of inspiratory flow waveforms on lung
ICU ventilators during pressure and volume ventilation. Respir Care. mechanics, gas exchange, and respiratory metabolism in COPD
2011;56(7):928–940. patients during mechanical ventilation. Chest. 2002;122(6):2096–2104.
29. Neumann P, Golisch W, Strohmeyer A, et al. Influence of different 54. Rau JL, Shelledy DC. The effect of varying inspiratory flow waveforms
release times on spontaneous breathing pattern during airway pres- on peak and mean airway presures with a time-cycled volume ventila-
sure release ventilation. Intensive Care Med. 2002;28(12):1742–1749. tor: a bench study. Respir Care. 1991;36:347–356.
30. Sassoon CS. Mechanical ventilator design and function: the trigger 55. Shelledy DC, Rau JL, Thomas-Goodfellow L. A comparison of the
variable. Respir Care. 1992;37(9):1056–1069. effects of assist-control, SIMV, and SIMV with pressure support on
31. Messinger G, Banner MJ. Tracheal pressure triggering a demand-flow ventilation, oxygen consumption, and ventilatory equivalent. Heart
continuous positive airway pressure system decreases patient work of Lung. 1995;24(1):67–75.
breathing. Crit Care Med. 1996;24(11):1829–1834. 56. Hewlett AM, Platt AS, Terry VG. Mandatory minute volume. A
32. Banner MJ, Blanch PB. Tracheal pressure ventilator control. Semin new concept in weaning from mechanical ventilation. Anaesthesia.
Respir Crit Care Med. 2000;21(3):233–243. 1977;32(2):163–169.
33. Messinger G, Banner MJ, Blanch PB, Layon AJ. Using tracheal pres- 57. Tehrani FT. Automatic control of mechanical ventilation. Part
sure to trigger the ventilator and control airway pressure during con- 1: theory and history of the technology. J Clin Monit Comput.
tinuous positive airway pressure decreases work of breathing. Chest. 2008;22(6):409–415.
1995;108(2):509–514. 58. Medical D. EvitaXL, Intensive Care Ventilator Software 6.1n Instructions
34. Branson RD, Campbell RS, Davis K Jr, Johnson DJ 2nd. Comparison for use. Lübek, Germany: Drager Medical; 2007.
of pressure and flow triggering systems during continuous positive 59. Kondili E, Akoumianaki E, Alexopoulou C, Georgopoulos D.
airway pressure. Chest. 1994;106(2):540–544. Identifying and relieving asynchrony during mechanical ventilation.
35. Goulet R, Hess D, Kacmarek RM. Pressure vs flow triggering during Expert Rev Respir Med. 2009;3(3):231–243.
pressure support ventilation. Chest. 1997;111(6):1649–1653. 60. Al-Saady N, Bennett ED. Decelerating inspiratory flow waveform
36. Imanaka H, Nishimura M, Takeuchi M, et al. Autotriggering caused by improves lung mechanics and gas exchange in patients on intermittent
cardiogenic oscillation during flow-triggered mechanical ventilation. positive-pressure ventilation. Intensive Care Med. 1985;11(2):68–75.
Crit Care Med. 2000;28(2):402–407. 61. Polese G, Lubli P, Poggi R, Luzzani A, Milic-Emili J, Rossi A. Effects
37. Sager JS, Eiger G, Fuchs BD. Ventilator auto-triggering in a patient of inspiratory flow waveforms on arterial blood gases and respiratory
with tuberculous bronchopleural fistula. Respir Care. 2003;48(5): mechanics after open heart surgery. Eur Respir J. 1997;10:2820–2824.
519–521. 62. Sassoon C. Triggering of the ventilator in patient-ventilator interac-
38. Remmers JE, Gautier H. Servo respirator constructed from a positive- tions. Respir Care. 2011;56(1):39–51.
pressure ventilator. J Appl Physiol. 1976;41(2):252–255. 63. Younes M. Proportional assist ventilation, a new approach to ventila-
39. Schertel ER, Schneider DA, Howard DL, Green JF. A phrenic nerve- tory support. Theory. Am Rev Respir Dis. 1992;145(1):114–120.
actuated electronically controlled positive-pressure ventilator. J Appl 64. Kondili E, Prinianakis G, Alexopoulou C, et al. Respiratory load
Physiol. 1987;62(5):2121–2125. compensation during mechanical ventilation—proportional assist
40. Sinderby C, Beck J, Spahija J, et al. Inspiratory muscle unloading by ventilation with load-adjustable gain factors versus pressure support.
neurally adjusted ventilatory assist during maximal inspiratory efforts Intensive Care Med. 2006;32(5):692–699.
in healthy subjects. Chest. 2007;131(3):711–717. 65. Chatburn RL, Wear JL. Comparison of two different proportional
41. Prinianakis G, Kondili E, Georgopoulos D. Effects of the flow wave- assist ventilation algorithms. Respir Care. 2009;54:1583.
form method of triggering and cycling on patient-ventilator inter- 66. Roze H, Lafrikh A, Perrier V, et al. Daily titration of neurally adjusted
action during pressure support. Intensive Care Med. 2003;29(11): ventilatory assist using the diaphragm electrical activity. Intensive Care
1950–1959. Med. 2011;37(7):1087–1094.
42. Chatburn RL. Understanding mechanical ventilators. Expert Rev 67. Fabry B, Haberthur C, Zappe D, et al. Breathing pattern and additional
Respir Med. 2011;4(6):809–819. work of breathing in spontaneously breathing patients with different
43. Chatburn RL, Volsko TA. Documentation issues for mechanical venti- ventilatory demands during inspiratory pressure support and auto-
lation in pressure-control modes. Respir Care. 55(12):1705–1716. matic tube compensation. Intensive Care Med. 1997;23(5):545–552.
44. Mireles-Cabodevila E, Chatburn RL. Mid-frequency ventilation: the- 68. Elsasser S, Guttmann J, Stocker R, et al. Accuracy of automatic tube
oretical basis and clinical predictions. Paper presented at: American compensation in new-generation mechanical ventilators. Crit Care
Thoracic Society International Conference; Monday, May 19, 2008; Med. 2003;31(11):2619–2626.
Toronto, Canada. 69. Aboab J, Niklason L, Uttman L, et al. CO2 elimination at varying
45. Chatburn RL, Mireles-Cabodevila E. Closed-loop control of mechani- inspiratory pause in acute lung injury. Clin Physiol Funct Imaging.
cal ventilation: description and classification of targeting schemes. 2007;27(1):2–6.
Respir Care. 2011;56(1):85–102. 70. Devaquet J, Jonson B, Niklason L, et al. Effects of inspiratory pause on
46. Chiumello D, Pelosi P, Croci M, et al. The effects of pressurization CO2 elimination and arterial PCO2 in acute lung injury. J Appl Physiol.
rate on breathing pattern, work of breathing, gas exchange and patient 2008;105(6):1944–1949.
comfort in pressure support ventilation. Eur Respir J. 2001;18(1): 71. Arnal JM, Wysocki M, Nafati C, et al. Automatic selection of breath-
107–114. ing pattern using adaptive support ventilation. Intensive Care Med.
47. Murata S, Yokoyama K, Sakamoto Y, et al. Effects of inspiratory rise 2008;34(1):75–81.
time on triggering work load during pressure-support ventilation: a 72. Du HL, Yamada Y. Expiratory asynchrony. Respir Care Clin N Am.
lung model study. Respir Care. 2010;55(7):878–884. 2005;11(2):265–280.
Chapter 3 Basic Principles of Ventilator Design 97
73. Tassaux D, Gainnier M, Battisti A, Jolliet P. Impact of expiratory trig- 83. Siebig S, Kuhls S, Imhoff M, et al. Intensive care unit alarms—how
ger setting on delayed cycling and inspiratory muscle workload. Am J many do we need? Crit Care Med. 2010;38(2):451–456.
Respir Crit Care Med. 2005;172(10):1283–1289. 84. Mullins R, Chatburn RL. Reference data for determining ventilator
74. Tassaux D, Michotte JB, Gainnier M, Gratadour P, Fonseca S, alarm limits. Respir Care. 2010;55(11):1520.
Jolliet P. Expiratory trigger setting in pressure support ventilation: 85. Waugh JB, Deshpande VM, Harwood RJ. Rapid Interpretation of
from mathematical model to bedside. Crit Care Med. 2004;32(9): Ventilator Waveforms. Englewood Cliffs, NJ: Prentice Hall; 1999.
1844–1850. 86. de Wit M. Monitoring of patient-ventilator interaction at the bedside.
75. Beck J, Gottfried SB, Navalesi P, et al. Electrical activity of the dia- Respir Care. 2011;56(1):61–72.
phragm during pressure support ventilation in acute respiratory fail- 87. Wachter SB, Johnson K, Albert R, et al. The evaluation of a pul-
ure. Am J Respir Crit Care Med. 2001;164(3):419–424. monary display to detect adverse respiratory events using high
76. Piquilloud L, Vignaux L, Bialais E, et al. Neurally adjusted ventilatory resolution human simulator. J Am Med Inform Assoc. 2006;13(6):
assist improves patient-ventilator interaction. Intensive Care Med. 635–642.
2011;37(2):263–271. 88. Mullins R, Cook SE, Chatburn RL. Evaluation of a new graphical
77. Hess DR. How much PEEP? Do we need another meta-analysis? interface for mechanical ventilation. Respir Care. 2009;54(11):1581.
Respir Care. 2011;56(5):710–713. 89. Solomita M, Palmer LB, Daroowalla F, et al. Humidification and
78. Habashi NM. Other approaches to open-lung ventilation: airway secretion volume in mechanically ventilated patients. Respir Care.
pressure release ventilation. Crit Care Med. 2005;33(3 Suppl): 2009;54(10):1329–1335.
S228-S240. 90. Intellivent product documentation. Hamilton Medical. 17/03/2011.
79. Baum M, Benzer H, Putensen C, et al. [Biphasic positive airway pres- Available at: http://www.hamilton-medical.com/Product-documenta
sure (BIPAP)--a new form of augmented ventilation]. Anaesthesist. tion.1074.0.html. Accessed July 5, 2011.
1989;38(9):452–458. 91. Arnal J, Wysocki M, Demory D, et al. Prospective randomized cross-
80. Putensen C, Zech S, Wrigge H, et al. Long-term effects of spontane- over controlled study comparing adaptive support ventilation (ASV)
ous breathing during ventilatory support in patients with acute lung and a fully close loop control solution (Intellivent®) in adult ICU
injury. Am J Respir Crit Care Med. 2001;164(1):43–49. patients with acute respiratory failure. Am J Respir Crit Care Med.
81. MacIntyre NR, Branson RD. Mechanical Ventilation. 2nd ed. St. Louis: 2010;181:A3004.
Saunders; 2009:153–156. 92. Lelluche F, Bouchard PA, Wysocki M, et al. Prospective randomized
82. Gorges M, Markewitz BA, Westenskow DR. Improving alarm perfor- controlled study comparing conventional ventilation versus a fully
mance in the medical intensive care unit using delays and clinical con- closed-loop ventilation (IntelliVent®) in post cardiac surgery ICU
text. Anesth Analg. 2009;108(5):1546–1552. patients. Am J Respir Crit Care Med. 2010;181:A6035.
This page intentionally left blank