Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.

0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

Section 3 Cabinet installation

Contents

1. MOUNTING THE SYSTEM FILTER BOX.............................................................................................3

2. MOUNTING THE GRADIENT SWITCH (INTERA NOVA DUAL ONLY) ..............................................5

3. MOUNTING THE REMOTE MAGNET MONITORING UNIT (RMMU) ..................................................6

4. MOUNTING THE EMERGENCY RUN DOWN UNIT (ERDU) SWITCHES ...........................................7

5. MOUNTING THE CEILING SPEAKERS...............................................................................................8

6. MOUNTING THE MAINS DISTRIBUTION UNIT (MDU) .......................................................................9

7. MOUNTING THE MAINS DISTRIBUTION EXTENSION UNIT (MDE)................................................10

8. MOUNTING THE REFRIGERATOR SYSTEM TRANSFORMER (NOT FOR USA) ...........................11

9. MOUNTING THE PATIENT VENTILATION UNIT ..............................................................................12

10. POSITIONING CABINETS .................................................................................................................13


10.1. Gradient amplifier 271 .....................................................................................................................13
10.1.1. installation for the Intera PULSAR (Copley 271) ..........................................................................13
10.1.2. Installation of the Intera Nova, Quasar or Nova Dual (Copley 271) ..............................................13
10.2. RF-amplifier .....................................................................................................................................16
10.2.1. Installation anti-tilt brackets (S26A only) ......................................................................................17

11. CONSOLE ..........................................................................................................................................18

MR/Intera/3612-SMI-101 03.0 BCD 3 - 1


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

BCD 3 - 2 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

1. MOUNTING THE SYSTEM FILTER BOX


1. Remove the dummy plate used for testing the RF shielded room. Clean the contact areas of the System
Filter Box (SFB) and of the RF enclosure.
2. Remove the upper aluminum covers from the SFB to gain access to the handles. Also remove the
middle copper cover.
3. Clean the contact area's of the RF cage interface and of the SFB.
4. Hang the SFB on the M10 bolts (do not tighten) prepared by the RF supplier. Only lift at the handles.
5. With the SFB four (4) strips are delivered for protecting the cables against sharp edges of the hole in the
SFB, see figure 2. Place the protection strips at the positions pointed out in figure 1, mount the strips
with M5 screws with washers delivered by RF cage supplier. Figure 3 shows the result with the upper
protection strip mounted seen from the technical room.
6. Mount all remaining M5 screws with washers delivered by RF cage supplier.
7. Tighten the M10 nuts.

Figure 1 - Mounting the System filter box

M10 BOLT

CABLE RELIEF

M5 SCREWS
LIFT ONLY AT
THE HANDLES

MOUNT PROTECTION STRIPS RVV12.DRW

MR/Intera/3612-SMI-101 03.0 BCD 3 - 3


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

Figure 2 - Protection strips

2x

2x

Figure 3 - Upper protection strip seen from technical room

BCD 3 - 4 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

2. MOUNTING THE GRADIENT SWITCH (INTERA NOVA DUAL ONLY)


The end of the crate, just below the feet of the Explorer switch, can be removed before rising up the gradient
switch. To rise up the gradient switch two engineers are required.

The gradient switch must be mounted to the wall, using the four 10 cm high wall brackets. The gradient
switch can either be positioned in front of it the system filter box (See Figure 4.), partly in front of it or to the
right of it. The position in front of the system filter box or to the right of it makes routing of the gradient cables
easier. Notice that you must connect the signal cables from the examination room side of the system filter
box. Position the Explorer switch as close as possible to the system filter box cable opening, but keep 10 cm
distance between the Explorer switch and the wall.
The explorer switch can either be installed directly on the computer floor (if available) or lowered on to the
floor of the technical room.

Figure 4 – The gradient switch in front of the system filter box.

MR/Intera/3612-SMI-101 03.0 BCD 3 - 5


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

3. MOUNTING THE REMOTE MAGNET MONITORING UNIT (RMMU)


The RMMU must be mounted on the top frame of the System Filter Box (SFB).

Figure 5: Position of RMMU

RMMU

SYSTEM FILTER BOX

FHC46.WPG

BCD 3 - 6 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

4. MOUNTING THE EMERGENCY RUN DOWN UNIT (ERDU) SWITCHES


One of the ERDU switches must be mounted inside the RF shielded room and one in the control room.
Mount the ERDU switch inside the RF room to the planned position (see site planning drawings). Ask the RF
room supplier how to mount the ERDU switch in order not to damage the RF shield.

CAUTION

Be sure not to penetrate or make contact with the RF shielding when installing the ERDU switch inside the
RF room.
________________

Figure 6: Mounting the ERDU switch

100

RF SHIELDING

ERDU
SWITCH

170
STOP

SCREW 5mm OR BOLT M5


1700+/-20

OR STUD WITH NUT M5


LOCALLY SUPPLIED

DO NOT TOUCH OR PENETRATE


THE RF SHIELDING

DIMENSIONS IN mm

MR/Intera/3612-SMI-101 03.0 BCD 3 - 7


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

5. MOUNTING THE CEILING SPEAKERS


Two ceiling speakers have to be mounted. One above the patient support and the other one at the rear of
the magnet (see figures). The distance of 1200mm is an average in order to obtain a good sound quality on
both sides of the magnet. The absolute minimum distance at the rear side is 0.55m. In that case, the quality
is decreased but is still acceptable. As an alternative, it is possible to mount the speaker on the wall. Use one
of the delivered build-on boxes.

Figure 7 - Mounting positions of the ceiling speakers

ALTERNATIVE ON WALL

F=25N
F=25N

MR-1
MR-1

MINIMAL 550mm IN CASE OF MINIMAL ROOM LAYOUT

1000 1200 1200 1000

3000
3000

CEILING SPEAKER AREA CEILING SPEAKER AREA


88

9-25

182

Dimensions in mm

BCD 3 - 8 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

6. MOUNTING THE MAINS DISTRIBUTION UNIT (MDU)


The MDU must be mounted to four threaded rods mounted in the technical room. Arrange a local electrician to
connect the mains to the MDU via the MDU extension unit.

Figure 8 - Mounting the MDU

600

A
DETAIL A

THREADED
RODS
M8
780

MDU
15
1800

CABLE OUTLETS

MAINS INLET

DIMENSIONS IN mm

MR/Intera/3612-SMI-101 03.0 BCD 3 - 9


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

7. MOUNTING THE MAINS DISTRIBUTION EXTENSION UNIT (MDE)


Also the MDU extension must be mounted to four studs mounted in the technical room. Arrange a local
electrician to connect the hospital mains to the MDU.

Figure 9- MDU extension

BCD 3 - 10 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

8. MOUNTING THE REFRIGERATOR SYSTEM TRANSFORMER (NOT FOR


USA)
The refrigerator system transformer must be mounted to four threaded rods in the technical room.

Figure 10 - Mounting the refrigerator system transformer

DETAIL A

THREADED
RODS
M6

500 15
A

SHEARING FORCE: 250N


REFRIGERATOR SYSTEM
TRANSFORMER
500
400

550

MAINS INLET
SITE DEPENDENT

MAXIMUM CABLE
DISTANCE
TRANSFORMER TO
COMPRESSOR 4 mtr.
DIMENSIONS IN mm
FH127.DRW

MR/Intera/3612-SMI-101 03.0 BCD 3 - 11


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

9. MOUNTING THE PATIENT VENTILATION UNIT


Mount the patient ventilation unit against the wall inside the technical room.
Two mounting brackets are supplied with the unit. Mount these brackets to the wall.
Mount the ventilation unit on top of these brackets using four dampers.
An air filter must be installed at the air outlet of the patient ventilation unit.

Figure 11 - Mount patient ventilation unit

FRONT VIEW LEFT SIDE VIEW


? DETAIL A
?

THREADED M6 RODS

? 15

?
SHEARING FORCE: 250 N

?
370 RF-SHIELDING

DO NOT MAKE CONTACT OR


30 PENETRATE THE RF-SHIELDING
? 0

TO MAGNET

DETAIL A AIR IN

TOP VIEW RF FEEDTHROUGH


RF-SHIELDING
EXAMINATION ROOM

TECHNICAL ROOM WALL


MIN. 120

MIN. BENDING
60 RADIUS 300 mm
0

800

Dimensions in mm :
Service and operational area

BCD 3 - 12 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

10. POSITIONING CABINETS


Position all cabinets into the technical room according to the Site-Planning drawings. For dimensions of
cabinets and service area consult the Intera Planning Reference Data.
Level all cabinets using a spirit level and tighten all feet.

10.1. GRADIENT AMPLIFIER 271


Unpacking:

WARNING

Under no circumstances move the packaging if it is partly unpacked. When the front riser assembly is
removed, the front wheels will rest on metal plates. The rear wheels are elevated and the cabinet will rest on
the riser blocs. The cabinet then will be slightly tilted forward. To move the cabinet with its packing in this
configuration is very dangerous.

The riser is marked with:

DO NOT DISCARD
DO NOT TRANSPORT UNLESS THIS ASSEMBLY IS INSTALLED

_______________

1. Remove all screws circled in red. Remove front panel and the right side panel of the crate.
2. Remove the front riser assembly. This supports the weight of the cabinet and protects the casters
during shipment.
3. Remove the ramp assembly and its support, attached at the inner right side panel. Place the heavy
wooden bar halfway under the ramp for additional support. Attach the ramp assembly into the slots at
the front of the pallet.
4. Use the nylon straps to pull the amplifier down the ramp to floor level.

10.1.1. INSTALLATION FOR THE INTERA PULSAR (COPLEY 271)


1. Move the cabinet to where it will be installed. The cabinet including the swivel wheels is 160,6 cm
high.
2. Remove the straps and plastic covering from the cabinet.
3. Adjust the cabinet levelers for stability.

10.1.2. INSTALLATION OF THE INTERA NOVA, QUASAR OR NOVA DUAL (COPLEY 271)
With the Intera Nova, Quasar or Nova Dual system, a Copley 271 master cabinet and a Copley 271 slave
cabinet is part of the configuration. The Copley 271 master cabinet has the gradient chain interface rack
(GCI) at the upper position in the cabinet. The Copley 271 slave cabinet does not have such a gradient chain
interface rack (GCI).

Mechanical interconnection:

Procedure:
1. Move the cabinets to where it will be installed and remove the straps and plastic covering from the
cabinets.
2. Master cabinet must be installed left and the slave cabinet right, faced from the front side. The
cabinets including the swivel wheels are 160,6 cm high.
3. Remove the right side panel from the master gradient amplifier rack (seen from the front).
4. Remove the left side panel from the slave gradient amplifier rack (seen from the front).

MR/Intera/3612-SMI-101 03.0 BCD 3 - 13


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

5. Move the cabinets as close as possible together.


6. Adjust the cabinet levelers to level both racks and for stability.
7. Bolt the cabinets together (See Figure 12). (Bolts are shipped with the ‘cable kit dual rack 271’ which is
used to interconnect both racks electrically.)
The master cabinet and slave cabinet must be bolted together for stability and guarantee
EMC.

Figure 12 - Copley 271 master and slave rack

GCI

Remove the side panels and use the


holes to bold the racks together, at the
front and rear side.

Slave rack

Master rack

BCD 3 - 14 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

Electrical interconnection:

The ‘cable kit dual rack 271’ is used to interconnect the Copley 271 master rack and the Copley 271 slave rack.
For installation of these cables refer to the following drawings in the drawings manual:
• Z3-6.4.1 Layout gradient amplifier Copley 271;
• Z3-6.4.2 Axis connectors and 40kW DC power supply connectors Copley 271;
• Z3-6.4.4 Front view GCI;
• Z5-6.4.1 Internal signal cable plan Copley 271 dual

Remarks to the internal cabling:


• The ADCI unit in top rear side of the Copley 271 slave cabinet is not used. The cable coming form J7 of
the X axis of the Copley 271 slave cabinet must be connected to the ADCI unit slave connector of the
Copley 271 master cabinet.
• Ensure that the ‘master’ labeled connectors are connected to the master rack and that the ‘slave’ labeled
connectors are connected to the slave rack for the cables which interconnects J8 of the X,Y and Z
master axis to the corresponding slave axis and the cable which interconnects J3 of the 40 kW DC
power supplies.
• The high voltage cables (+HV and –HV), which interconnect the high voltage outputs of the 40kW DC
power supply in the Copley 271 master rack and the Copley 271 slave rack are connected to the output
terminals of each of these power supplies. (Only for the early racks, the high voltage cables must be
connected to the bus bar directly behind the 300A fuse in the back of the rack.) See Figure 13.

Figure 13 – high voltage interconnection

The high voltage interconnection cables


are directly connected to the output
terminals of both DC power supplies.

MR/Intera/3612-SMI-101 03.0 BCD 3 - 15


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

10.2. RF-AMPLIFIER
For Intera 1.0T, 1.5T and 3.0T, the RF amplifier must be positioned next to the NTDAC.

Figure 14- Location RF Amplifier

1.0T, 1.5T and 3.0T SYSTEMS

NTDAC

S23/S24/S26
RF POWER
AMPLIFIER

EMPTY

GR91.WPG

This is the S23 (Intera 1.5T), S24 (Intera 1.0T) and S26 (Intera 3.0T).

NOTE

The Mains Power for the S23, S24 or S26A (single phase) and S26B (three phase) RF amplifiers comes
directly from the MDU.

________________

NOTE

The RF-Output on the S23, S24 and S26 RF amplifiers is located on the rear of the amplifier.
________________

BCD 3 - 16 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Intera Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Section 3
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

10.2.1. INSTALLATION ANTI-TILT BRACKETS (S26A ONLY)


To prevent that the S26A RF Amplifier will tip over when leaning against the unit, two anti-tilt brackets must
be installed.

Screws

MR/Intera/3612-SMI-101 03.0 BCD 3 - 17


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED
Section 3 Intera 0.5 T Intera 1.0 T Intera 1.5 T Intera 3.0 T Intera
Omni/ Power/ Omni/ Power/ Master/ Explorer/ Master/
Standard
Stellar Pulsar Stellar Pulsar Nova Nova Dual Quasar

11. CONSOLE
Position the console table against the wall in the operator’s room. Position the “Operator equipment cabinet”
underneath the table.

Figure 15 - Console

Figure 16 – Console with units

OPERATOR CONSOLE
DISPLAY

AUDIO
MODULE
KEYBOARD
MOUSE

OPERATOR EQUIPMENT CABINET

BE

HOST
COMPUTER
BH

SERVICE STORAGE CABINET

FRONT VIEW
AS28.DRW

The service storage cabinet can also be mounted at the right-hand side under the table.
The height of the table can be adjusted.

BCD 3 - 18 03.0 MR/Intera/3612-SMI-101


Copyright © 2003 Philips Medical Systems Nederland B.V.
ALL RIGHTS RESERVED