0% found this document useful (0 votes)
142 views78 pages

UT285C User Manual

power quality tester

Uploaded by

z.2020
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
142 views78 pages

UT285C User Manual

power quality tester

Uploaded by

z.2020
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 78

Power Quality Analyzer

P/N:110401110254X
2023/12/05
Warning
Thank you for purchasing UT285C Power Quality Analyzer. To better use this product, please be sure to:
----Read this user manual thoroughly.
----Comply with the safe regulations and operating cautions strictly in this manual.
䕺 Failure to comply with follow precautions may cause electric shock, explosion, or fire.
䖩 Under any circumstance, please pay special attention on safety in using this device.
䖩 Pay attention to words and symbols affixed on the panel and back cover of the device.
䖩 This device must be used, disassembled and repaired by authorized professionals.
䖩 If continuous use of this device is dangerous to the user, please stop use immediately and seal the device,
and then send it to authorized body for maintenance.
䖩 The danger symbol in manual “ ” indicates that users must perform safe operation in compliance with
the manual instructions.
䖩 The extreme danger symbol in manual “ ” indicates that users must perform safe operation strictly in
compliance with the manual instructions.
䖩 Before removing and replacing the battery or the SD-Card, please power off the device and disconnect it
from tested line, current sensor, power adaptor.
䖩 Current clam p m ust be connected with the device correspondingly, otherwise the test error may be
increased.
䖩 Please do not use the device if the battery or SD card slot is missing, damaged or improperly installed.
䖩 The founder and operator of the system shall be responsible for any system security problems unrelated to
this Analyzer.
䖩 For safety, please use only the wires and accessories supplied with the Analyzer (in accordance with the
standard of IEC61010-031(2002)). When a sensor or accessories in low-voltage class are connected to the
device, which must be applicable to the system of this device.
䖩 Please always check test leads, current clamps and accessories are in optimum condition before use. Any
wire, current clamp or accessory with damaged insulation (or even partially damaged) must be repaired or
scrapped.
䖩 Comply with the environmental condition (refer to 15.3.1)
䖩 Please use personal safety equipment when necessary.
䖩 The Analyzer should be used in devices with AC/DC voltage to ground not greater than CAT IV 600V
(IEC61010-1 standard), or used in devices with voltage not greater than CAT III 1000V. It is forbidden to use
for power grids or measurement category with higher voltage.
䖩 Use only the power adapter or battery pack supplied by the m anufacturer, which is designed in accordance
with specific safety rating.
䖩 Please comply with the safety rating for the accessories or current clamp. Do not make connection with any
unused terminals.
䖩 Som e current clamps cannot be installed to or removed from exposed conductors under the dangerous
voltage.
䕺 Usage & Connection Step:
䖩 Start up the Analyzer.
䖩 Set the relevant parameters of the Analyzer according to the desired results and the network type.
䖩 Connect the voltage test leads and current sensors to the Analyzer.
䖩 Connect the ground line and/or neutral line of the voltage test leads to the ground line and/or neutral line of
the power grid.
䖩 Connect the neutral line of current sensor to the neutral line of power grid.
䖩 Connect the L1 phase lead to the L1 phase of power grid and the corresponding current sensor clamp to the
phase L1 of power grid.
䖩 Repeat the above connection steps for phases L2, L3.
Note: Complying with this operation procedure reduces connection errors to minimum and avoids wasting
time.
䕺 Disco n n ectio n Steps:
䖩 Disconnect the wiring in reverse order (disconnect the ground line or neutral line at last in general).
䖩 Remove all the leads of the Analyzer and shut it down.
䖩 Charge the battery and upload the test data when necessary.
USB is used for data transmission. Battery can be charged by the supplied special adapter.
䕺Dotted Ends Of The Current Clamp
䖩 The side with current clamp marks (L1, L2, L3, N/D or has red dot) is the current noninverting input end,
which is dotted end.
䖩 The side of 008B current clam p has red dot is the current noninverting input end, that is dotted end.
䖩 The side of 040B, 068B current clamp without screws is the current noninverting input end, which is dotted
end.
䖩 The side with the lock symbol “ ” of the 300F flexible current sensor as the current noninverting input end,
that is dotted end.
CONTENT
1. SUMMARY ....................................................................................................................................................... 1
1.1. Introduction ................................................................................................................................................ 1
1.2. Function ..................................................................................................................................................... 1
1.3. Technical Specification .............................................................................................................................. 2
2. Accessories ...................................................................................................................................................... 6
2.1. Standard Configuration ............................................................................................................................. 6
2.2. Analyzer Weight......................................................................................................................................... 6
3. Analyzer Introduce ........................................................................................................................................... 7
3.1. Overall View............................................................................................................................................... 7
3.2. Power On/Off Key ...................................................................................................................................... 7
3.3. Display Screen .......................................................................................................................................... 8
3.4. Analyzer Keys ............................................................................................................................................ 8
3.5. Wiring Connection ..................................................................................................................................... 9
3.6. Power Supply............................................................................................................................................. 9
3.7. The Bracket ............................................................................................................................................. 10
3.8. Functions Summary................................................................................................................................. 10
3.9. Abbreviations ............................................................................................................................................11
4. USE ................................................................................................................................................................ 12
4.1. Start-Up ................................................................................................................................................... 12
4.2. Setting...................................................................................................................................................... 12
4.3. Wire Connection ...................................................................................................................................... 13
4.4. Waveform Capture................................................................................................................................... 14
4.5. Display Harmonics................................................................................................................................... 14
4.6. Waveform Measurement ......................................................................................................................... 14
4.7. Alarm Detection ....................................................................................................................................... 15
4.8 Trend Chart Record Setting...................................................................................................................... 15
4.9. Energy Measurement .............................................................................................................................. 15
4.10. Data Communication With PC............................................................................................................... 15
4.11. Delete Data ............................................................................................................................................ 15
4.12. Shut Down ............................................................................................................................................. 15
4.13. Power Supply ........................................................................................................................................ 16
5. SETTING ........................................................................................................................................................ 16
5.1. Setting Select Menus............................................................................................................................... 16
5.2. Display Language .................................................................................................................................... 17
5.3. Date/Time ................................................................................................................................................ 17
5.4. Display ..................................................................................................................................................... 17
5.5. Calculation Methods ................................................................................................................................ 18
5.6. Electrical Wiring Connection Selection ................................................................................................... 18
5.7. Sensors & Ratio Selection....................................................................................................................... 18
5.8. Transient Capature Setting...................................................................................................................... 19
5.9. Trend Monitoring Parameter Setting ....................................................................................................... 20
5.10. Alarm Monitoring Parameter Setting ..................................................................................................... 21
5.11. Delete Data ............................................................................................................................................ 22
5.12. About...................................................................................................................................................... 22
6. WAVEFORM CAPTURE MODE .................................................................................................................... 23
6.1. Sub-Mode Selection ................................................................................................................................ 23
6.2. Transient Measurement Mode................................................................................................................. 23
6.3. Inrush Current Mode................................................................................................................................ 25
7. HARMONICS MODE ..................................................................................................................................... 28
7.1. Select Sub-Menu ..................................................................................................................................... 28
7.2. Phase Voltage ......................................................................................................................................... 29
7.3.Current ...................................................................................................................................................... 29
7.4. Apparent Power ....................................................................................................................................... 30
7.5. Line Voltage ............................................................................................................................................. 31
7.6. Expert Mode ............................................................................................................................................ 31
8. WAVEFORM MODE ....................................................................................................................................... 32
8.1. Select Sub-menus ................................................................................................................................... 32
8.2. True RMS................................................................................................................................................. 32
8.3. Total Harmonics Distortion....................................................................................................................... 34
8.4. PEAK Factor ............................................................................................................................................ 35
8.5. Extreme & Average Values Of Voltage & Current ................................................................................... 36
8.6. Various Test Value Displayed Simultaneously ........................................................................................ 37
8.7. Phasor Diagram Display .......................................................................................................................... 38
9. ALARM MODE ............................................................................................................................................... 39
9.1. Submenus................................................................................................................................................ 40
9.2. Alarm Mode Setting ................................................................................................................................. 40
9.3. Alarm Detection Schedual ....................................................................................................................... 40
9.4. View Alarm Log ........................................................................................................................................ 41
9.5. Delete Alarm Log ..................................................................................................................................... 42
10. TREND MODE ............................................................................................................................................. 42
10.1. Select Submenus .................................................................................................................................. 42
10.2. Presets&Starts Record Trend Charts .................................................................................................... 42
10.3. Trend Chart Monitoring Parameter Setting ........................................................................................... 44
10.4. View Record List.................................................................................................................................... 45
10.5. Delete Record ........................................................................................................................................ 46
10.6. Display Records .................................................................................................................................... 46
11. POWER AND ENERGY MODE ................................................................................................................... 49
11.1. Sub-menus............................................................................................................................................. 49
11.2. Energy Consumed ................................................................................................................................. 49
11.3. Power Factor Display Screen ................................................................................................................ 50
11.4. The Sums Of Energies Consumed Display Screen .............................................................................. 50
11.5. The Arithmetic Mean Values Of Power Factor Display Screen ............................................................. 51
11.6. Energy generate .................................................................................................................................... 51
11.7. Starting Energy Metering ....................................................................................................................... 52
11.8. Stopping Energy Metering ..................................................................................................................... 53
11.9. Reset Of Energy Metering .................................................................................................................. 53
12. SCREEN SHOT MODE ............................................................................................................................... 53
12.1. Screen Snapshots ................................................................................................................................. 53
12.2. Handling Of Screen Snapshots ............................................................................................................. 54
13. HELP ............................................................................................................................................................ 54
14. DATA UPLOADED TO THE COMPUTER.................................................................................................... 55
15. GENERAL SPECIFICATIONS ..................................................................................................................... 55
15.1. Housing.................................................................................................................................................. 55
15.2. Power Supply ........................................................................................................................................ 55
15.3. Range for use ........................................................................................................................................ 56
15.4. User safety............................................................................................................................................. 56
16. FUNCTIONAL CHARACTERISTICS ........................................................................................................... 57
16.1. Reference conditions ............................................................................................................................. 57
16.2. Electrical characteristics ........................................................................................................................ 57
17. APPENDICES .............................................................................................................................................. 59
17.1. Mathematic Formulae ............................................................................................................................ 59
17.2. Hysteresis .............................................................................................................................................. 59
17.3. Four-quadrant Diagram ......................................................................................................................... 59
17.4. Mechanism For Triggering Transient Captures ..................................................................................... 60
17.5. Capture Conditions In Inrush Current Mode ......................................................................................... 60
18. MAINTENANCE ........................................................................................................................................... 60
18.1. Important Recommendation .................................................................................................................. 60
18.2. Recharging The Battery......................................................................................................................... 60
18.3. Replacing The Battery ........................................................................................................................... 60
1. OVERVIEW
1.1. Introduction
UT285C Power Quality Analyzer is a three-phase, multifunctional and intelligent Analyzer specially designed
for on-site tests. It is characterized by large LCD, high resolution, interface in both English and Chines, shock
proof, etc. UT285C can simultaneously measure the followings: 4-channel current (ABC three-phase and
neutral wire current), 4-channel voltage (ABC three-phase voltage and neutral line voltage to ground), peak
value of current and voltage, maximum/minimum value over a period, three-phase imbalance factor, short-time
voltage flicker, transformer K factor, active power, reactive power, apparent power, power factor, displacement
power factor, active electrical energy, reactive electrical energy, apparent electrical energy, harmonic ratio,
total harmonic distortion, etc. This Analyzer is able to display real-time waveform, phasor diagram, harmonics
ratio bar chart of current and voltage. UT285C has multiple functions including capturing the instantaneous
change of voltage and current dynamically, monitoring starting current and power parameters, recording test
data for a long time, generating alarm list and trend chart, and more.
For electrical applications nowadays, more and more complex faults occur at large-scale electrical equipment
and power girds, thus it is difficult to troubleshoot such complex problems. As the requirements for the power
quality of grid become increasingly strict in various industries, we design a power quality Analyzer that can
troubleshoot electrical system rapidly and accurately, and that is capable of monitoring and analyzing the
parameters of power quality comprehensively and systematically.

UT285C Power Quality Analyzer adopts the DSP + ARM (dual processors) architecture. The DSP is used for
data collection and algorithmic processing; while the ARM is used for communication protocol and processing
of man-machine interface. Simultaneous 4-channel analog signal acquisition is achieved by 2 pieces of
AD7655 (resolution: 16 bits) of the ADI company. The highest sampling rate that the Analyzer can reach is 1
MSPS, which ensures the accuracy of the channel and the information integrity, and ensures that any transient
changes are not missing in the power grid, and makes the detection of transient waveform, surge and
undervoltage, instantaneous interruption more. The DSP working frequency is over 200 MHZ, which enables
the Analyzer to monitor the power grid in time, and to dynamically adjust the sampling rate to be synchronous
with the industrial frequency. The Analyzer is designed with a 5.6-inch color screen with resolution of 640 dots
x 480 dots, and displays the parameter, waveform, phasor diagram, and harmonic ratio for each phase by
using different colors, so that the state of parameters of power grid can be understood visibly and efficiently.
The built-in flash memory can store 60 groups of screenshots at the same time, 150 groups of captured
transient voltage/current waveforms, and 12800 groups of alarm lists. In the starting current detection model,
the Analyzer can continuously capture starting current waveform for 100 s. The built-in 2G memory is used to
store the trend curve records, and has the ability to simultaneously record 20 parameters (adjustable), with
recording performed once per minute and trend curve records stored for 300 days.

Power Quality Analyzer is also called Intelligent Three Phase Power Quality Analyzer, Multifunctional Power
Quality Analyzer, and so on. It has functions of harmonic Analyzer, phase volt-ampere meter, electric
parameter tester, and is applicable to electricity industry, petrochemical, metallurgy, railway, mining enterprises,
scientific research institution, and metrological department. It is particularly suitable for comprehensive
analysis and diagnosis on all the voltage, current, power, power, harmonic, and phase electric parameters.

1.2. Function
1.2.1. Basic function
䖩 Real-time display of waveform (4-channel voltage/4-channel current).
䖩 True R MS values of voltage and current.
䖩 The DC components of voltages.
䖩 Peak current and voltage values.
䖩 Minim um and m axim um half-cycle RMS current and voltage values.
䖩 Phasor diagram display.
䖩 Measurem ent of each harm onic up to 50th order.
䖩 Bar charts show harm onic ratios of current and voltage of each phase.
䖩 Total harmonic distortion (THD).
䖩 Active/reactive/apparent power and total value of each phase.
䖩 Active/reactive/apparent energy and total value of each phase.
䖩 Transformer K factor.
䖩㻌㻼㼛㼣㼑㼞㻌㼒㼍㼏㼠㼛㼞㼟㻌㻔㻼㻲㻕㻌㼍㼚㼐㻌㼐㼕㼟㼜㼘㼍㼏㼑㼙 㼑㼚㼠㻌㼒㼍㼏㼠㼛㼞㼟㻌㻔㻰㻼㻲㻌㼛㼞㻌㻯㻻㻿䃅㻕㻚
䖩 Short-term voltage flicker (PST).
䖩 Three phase unbalance (current and voltage).

1
1.2.2. Capture record function
䕺Transient capture function
The Analyzer can capture and detect the instantaneous changes of the voltage and current parameters of
power grid, including voltage & current fluctuation, voltage & current surge and sags, short time interruption,
transient overvoltage, impulse current, current & voltage instantaneous distortion. The Analyzer can store 150
groups of transient waveforms at the same time.

䕺Startin g cu rren t m on ito rin g


The Analyzer can monitor the surge current of the circuit and the starting current of the electrical equipment
when it is started, which facilitates the design of the installed capacity correctly. It can display the rise/fall curve
of the RMS during the starting process, the envelope curve of the starting current, the waveform of 4-channel
current and 4-channel voltage. After triggering, it can record for about 100s, and store all the instantaneous
values and waveform curve of the current & voltage in each period within 100s.

䕺 Tren d ch art reco rd an d sto re function


For all the test parameters of basic test functions (Urms, Uthd, Ucf, Uunb, Hz, Vrms, Vthd, Vcf, Vunb, PST,
Arms, Athd, Acf, Aunb, KF,W, VAR, VA, PF, COS , TAN ), 50th harmonic of voltage, 50th harmonic of current,
total 123 parameters are recorded and the trend curve can be created. Data can be recorded for a long time if
needed. (Simultaneously select 20 kinds of parameters to record data for once every one minute, can record
about 300 days).

䕺Alarm fu n ctio n
The limit value can be set for the selected parameter, to monitor if overrange occurs. Alarm log will be created
if the limit value is exceeded. For overvoltage, current overflow, unbalance degree over range, a harmonic ratio
over range, frequency over range, active power over range, total harmonic distortion over range, and others,
maximum 40 groups of alarm monitoring parameters can be set. Different monitoring parameters (including
50th harmonic, with 123 different parameters totally) and limit values can be set for each group. The minimum
time of overrange can be set. The Analyzer can store up to 12,800 groups of alarm records.

䕺Screenshot function
In any testing screen, the present screen can be captured and stored, at the same time automatically store the
record time and test mode is automatically saved at the same time. For example, voltage & current waveform,
harmonic bar chart, phasor diagram and others can be saved. The Analyzer can save a maximum of 60 groups
of screenshots.

1.2.3. Other functions


䕺Co m m u n icatio n fu n ctio n
By communicating with computer via USB cable, the monitoring software can display the waveform in the
power quality analyzing test in real time, read the detected and captured transient waveform, trend chart
records, alarm logs, screenshots and more and display them on the computer.

䕺Settin g fu n ction
The wiring method of the Analyzer and the type of power grid can be set.
Different current clamps and different voltage test ratios can be selected.
Chinese menu or English menu can be selected.

䕺Help m en u in Ch in ese/E n g lish


Relevant information can be obtained by pressing the “Help” key in each step of operation.

1.3. Technical Specification


1.3.1. Basic condition and working condition
Influence Factor Test Item Basic Condition Working Condition
Environment Temperature All parameters (23±2)°C -10°C~ 40°C
Relative Humidity All parameters 40%~ 60% <80%

Phase Voltage All parameters (100±1%)V 1.0V~ 1000V

Line Voltage Test true RMS of line voltage (200±1%)V 1.0V~ 2000V

Current Test true RMS of current (5±1%)A 10mA~ 6000A


Power grid Frequency All parameters 50Hz±0.1Hz 40Hz~ 70Hz

2
Test active power and active Cos =1 Cos : 0.2~ 1.0
Phase Difference Between energy
the current and voltage Test reactive power and Sin =1 Sin : 0.2~ 1.0
reactive energy
Harmonic All parameters <0.1% 0.0%~ 100%
Voltage Unbalance All parameters <10% 0.0%~ 100%
Working Voltage All parameters DC9.8V±0.1V DC9.5V~ 10.5V
External electric Field,
All parameters Should be avoided
Magnetic Field
Measured related parameters
The Tested Wire Position Tested wire at the center of clamp.
of current

1.3.2. General specification


Power Supply Rechargeable lithium-ion battery packs 9.6V, external charger.
Battery symbol shows power energy. When the voltage is too low, the Analyzer shuts
Battery Indicator
down automatically after 1 minute.
Working Current About 490 mA. Working for 8 hours continuously.
Display Mode LCD color screen, 640dots×480dots, 5.6 inches, display area: 116mm×88mm.
008B small sharp current clamp: 7.5mmX13mm;(optional)
040B round current clamp: 35mmX40mm;(optional)
CT Size
068B round current clamp: 68mmX68mm.
300F flexible current sensor (with Integrator) : 300mm.(optional)
Analyzer Dimensions L W H: 240mmX170mmX68 mm.
Number Of Channels 4 voltage channels & 4 current channels (4U&4I).
Line Voltage 1.0V~2000V.
Phase Voltage 1.0V~1000V.
008B current clamp: 10mA~10.0A;
040B current clamp: 0.10A~100A;
Current
068B current clamp: 1.0A~1000A;
300F flexible current sensor (with Integrator) : 10A ~ 6000A
Frequency 40Hz~70Hz.
Electricity Parameters W, VA, Var, PF, DPF, cos , tan .
Energy Parameters Wh, Varh, Vah.
Harmonic Yes, 0th~50th order
Total Harmonic Distortion Yes, 0th~50th order, each phase.
Groups Of Transient
150 groups
Records
Voltage Flicker Yes.
Starting Current Mode Yes, 100 seconds.
3 Phases Unbalance Yes.
Record 300 days (Recording 20 parameters simultaneously, with one point recorded for every 1 min).
Min/Max Recorded Value Measurement of maximum and minimum values within a period of time.
Alarm 40 different kinds of parameter can be selected. There are 12,800 groups of alarm logs.
Peak Yes.
Screenshot Store 60pcs
Menu Language English
Communication Interface USB.
In the alarm mode/trend chart record mode/transient capture mode (waiting or in progressing),
the Analyzer will not shut down automatically.
Automatic Shutdown
If there is no keystroke within 15 minutes in other test modes, the Analyzer will shut down
automatically after one minute of prompt.
3
Backlight Function Yes (Suitable for use at night and dark place)
Analyzer: 1.6kg (with battery).
008B small sharp current clamp: 170gX4 (optional)
040B current clamp: 190gX4 (optional)
Weight 068B current clamp: 510gX4
300F flexible current sensor (with Integrator) : 330gX4 (optional)
Test wires and power adapter: 900g
Total weight: about 9.2kg (with packaging).
Voltage Test Wire Length 3m.
Current Sensor Wire
2m.
Length
Working Temperature And
-10°C~40°C; below 80%Rh.
Humidity
Storage Temperature And
-10°C~60°C; below 70%Rh.
Humidity
Input Impedance Input impedance of test voltage: 1M .
Withstand 3700V/50Hz sinusoidal AC voltage for 1 minute between the wiring and housing of
Withstand Voltage
the Analyzer.
Insulation 10M between the wiring and housing of the anaylzer
Structure Double insulation, with insulated vibration-proof sheath.
Suitable Safe Standard IEC 61010 1000V Cat III / 600V CAT IV, IEC61010-031, IEC61326, Pollution Degree 2.

1.3.3. Analyzer accuracy specification (excluding the current sensor)


The following data are respectively introduced according to the basic conditions and on the base of an ideal
current sensor (completely linear and without phase shift).
The Max Error In
Measurement Range Display Resolution
Reference Range
Frequency 40Hz~ 70Hz 0.01Hz ±(0.03)Hz

Phase Voltage True RMS 1.0V~ 1000V Min resolution 0.1V ±(0.5%+5dgt)

Line Voltage True RMS 1.0V~ 2000V Min resolution 0.1V ±(0.5%+5dgt)

DC Voltage 1.0V~ 1000V Min resolution 0.1V ±(1.0%+5dgt)

Current True RMS 10mA~ 6000A Min resolution 1mA ±(0.5%+5dgt)

Peak Of Phase Voltage 1.0V~ 1414V Min resolution 0.1V ±(1.0%+5dgt)

Peak Of Line Voltage 1.0V~ 2828V Min resolution 0.1V ±(1.0%+5dgt)

Current Peak 10mA~ 6000A Min resolution 1mA ±(1.0%+5dgt)

1.00~ 3.99 0.01 ±(1%+2dgt)


Peak Factor
4.00~ 9.99 0.01 ±(5%+2dgt)
±(1%+3dgt)
Cos 0.8
Active Power 0.000W~ 9999.9kW Min resolution 0.001W
±(1.5%+10dgt)
0.2 Cos <0.8
±(1%+3dgt)
Reactive Power Inductive 0.000VAR~ Sin 0.5
Min resolution 0.001VAR
& Capacitive 9999.9kVAR ±(1.5%+10dgt)
0.2 Sin <0.5
0.000VA~
Apparent Power Min resolution 0.001VA ±(1%+3dgt)
9999.9kVA

4
±(1.5%+3dgt)
Cos 0.5
Power Factor -1.000~ 1.000 0.001
±(1.5%+10dgt)
0.2 Cos <0.5
±(1%+3dgt)
Cos 0.8
Active Energy 0.000Wh~ 9999.9MWh Min resolution 0.001Wh
±(1.5%+10dgt)
0.2 Cos <0.8
±(1%+3dgt)
Reactive Energy Inductive 0.000VARh~ Sin 0.5
Min resolution 0.001VARh
& Capacitive 9999.9MVARh ±(1.5%+10dgt)
0.2 Sin <0.5
0.000VAh~
Apparent Energy Min resolution 0.001VAh ±(1%+3dgt)
9999.9MVAh
Phase Angle -179°~ 180° 1° ±(2°)
7DQij
-32.76~ 32.76 Min resolution 0.001 :±(1°)
9$•9$
Phase Difference Of Power
-1.000~ 1.000 0.001 :±(1°)
Factor(DPF)
(1st̚20th order)±(1%+5dgt)
(21st̚30th
Voltage Harmonic Ratio
0.0%~ 99.9% 0.1% order)±(1%+10dgt)
1st-50th order (Vrms>50V)
(31st̚50th
order)±(1%+15dgt)
±(3°) harmonics for 1st – 25th
Voltage Harmonic Angle order
-179°~ 180° 1°
(Vrms>50V) ±(10°) harmonics for 26th –
50th odrer
(1st̚20th order)±(1%+5dgt)
Current Harmonic Ratio (21st̚30th
1st -50th order 0.0%̚99.9% 0.1% order)±(1%+10dgt)
(Arms>Irange/100) (31st̚50th
orde)±(1%+15dgt)
±(3°) harmonics for 1st – 25th
Current Harmonic Angle order
-179°̚180° 1°
(Arms>Irange/100) ±(10°) harmonics for 26th –
50th order
Total Harmonic Ratio
0.0%~ 99.9% 0.1% ±(1%+10dgt)
(THD or THD-) ”
Distortion Factor
0.0%~ 99.9% 0.1% ±(1%+10dgt)
(DF or THD-5 ”
Transformer K Factor 1.00~ 99.99 0.01 ± (5%)

3 Phases Unbalance 0.0%~ 100% 0.1% ± (1%)

1.3.4. Current sensor character


Max Max Error Of Max Error Of Phase
Current Sensor Type Current True RMS Range
Resolution Current True RMS Angle ij
10mA~ 99mA ±(1%+3dgt) ±(1.5°),Arms 20mA
008B current clamp 10.0A 1mA
100mA~ 10.0A ±(1%+3dgt) ±(1°)

0.10A~ 0.99A ±(1%+3dgt) ±(1.5°)


040B current clamp 100A 0.01A
1.00A~ 100A ±(1%+3dgt) ±(1°)

068B current clamp 1.0A~ 9.9A 1000A 0.1A ±(2%+3dgt) ±(3°)

5
10.0A~ 1000A ±(2%+3dgt) ±(2°)

300F flexible current 10A̚99A ±(1 % + 3dgt) ±(3°)


sensor (with 6000A 1A
Integrator) 100AA~6000A ±(1 % + 3dgt) ±(2°)
Note: Please connect the current clamp to the Analyzer with corresponding wiring. Do not make reverse
connection. The direction of same-polarity end of the current clamp:
䖩The side of current clam p m arking L1, L2, L3, N/D or with red dot is the in-phase current input end, that is,
the same-polarity end.
䖩The side of 008B current clam p with red dot is the in-phase current input end, which is the same-polarity
end.
䖩The side of 040B, 068B current clamp without screws is the in-phase current input end, that is, the
same-polarity end.
䖩The side with the lock symbol” ” of the 300F flexible current sensor is the in-phase current input end, that
is, the same-polarity end.

2. Accessories
2.1. Standard Configuration
No. Designation Quantity
1 Analyzer 1 unit
2 Carrying bag 1 pcs
3 Current sensors 008B 4 pcs (optional)
4 Current sensors 040B 4 pcs (optional)
5 Current sensors 068B 4 pcs
6 300F flexible current sensors (with Integrator) 4 pcs (optional)
7 Testing wires. 5 pcs (yellow, green, red, blue, black)
8 Crocodile clips. 5 pcs
9 Test probe. 5 pcs
10 Power adapter. 1 pcs
11 USB communication cable 1 pcs
12 Lithium battery pack. 1 (Built into the Analyzer)
13 2GB memory card 1 pcs(Plug in the Analyzer)
14 Manual 1 pcs

2.2. Analyzer Weight


No. Designation Weight
1 Analyzer 1.6Kg (with battery).
2 008B small sharp current clamp sensor. 170gX4.
3 040B current clamp sensor. 190gX4.
4 068B current clamp sensor. 510gX4.
5 300F flexible current sensor (with Integrator) 330gX4.
6 Test wires and power adapter. 900g.
7 Total weight. 9.2kg (with packaging).

6
3.3. Display Screen
3.3.1. Introduce
The backlit LCD (640×480) can display measurement value and curves diagram, device parameters, selection
curves, instantaneous values of the signals, and measurement mode.
When the device is powered up, it automatically displays the waveform mode screen. For more information
about this screen, please refer to section §8.
Present select test mode Battery remaining capacity

Present date and time

Active mode screen

Figure 3-2: Screen display figure


Automatic shutdown: When the Analyzer is not in any state of trend chart recording, transient capture,
starting current detection or alarm detection (waiting or in progressing), the Analyzer will automatically shut
down if there is no keystroke within 15 minutes.
3.3.2. Symbol
The screen displays symbols as below:
Symbol Designation Symbol Designation
V Phase voltage Display voltage and current RMS values and extrema.
Simultaneous display all voltage and current
U Line voltage
measurement value
A Current Display voltage and current vector diagram
VA Apparent power Energies consumed
Zoom in Energies generated
Zoom out The 1st page of Help
The waveform of X axis cursor indicator The 2nd page of Help
PF… Display value of PF, DPF, Tan . The 3rd page of Help
W… Display value of powers and energies The 1st group monitoring parameter of trend mode
Start Recording The 2nd group monitoring parameter of trend mode
Recording list The 3rd group monitoring parameter of trend mode
OK Confirm The 4th group monitoring parameter of trend mode
Stop Page up
Delete Page down

3.4. Analyzer Keys


3.4.1. Function keys (yellow keys)
,, , , , , 6 yellow function keys, corresponding function displayed on the LCD screen
above the keys.

3.4.2. Navigation keys


Four arrow keys, one confirm key and one return key constitute the menu of navigation module.
Key Function
Up direction or increase
Down direction or reduce
Right direction( page down)
Left direction(page up)
Confirm

8
Return

3.4.3. Test Mode Keys (6 keys)


Press the mode key to enter into the corresponding test mode:
Key Corresponding Test Mode Detail Chapter

Waveform mode: display of voltage and current waveform, max and min value, extreme
value, summary tables of each parameter measurement value, voltage and current vector §8
diagrams.
Harmonic mode: display of voltage, current, and apparent power harmonic ratios (displayed
§7
in a graph), harmonic RMS value, phase shift with respect to the fundamental.
Power and energy mode: displays the active power, the reactive power, the apparent power,
§11
power factor, phase shift of power factor, etc.
Trend mode: recording of the parameters selected in the Configuration menu. §10
Alarm mode: listing of recorded alarms exceeding the thresholds programmed in the
configuration; recording of network blackouts with half-cycle resolution (Vrms, Arms, Urms),
§9
determination of energy consumption overshoots, monitoring of compliance with a power
supply quality contract.
Capture mode: monitoring instantaneous change of the voltage and current parameters of
power grid, including the voltage current fluctuations, voltage current surge, sag and short
§6
supply interruption, temporary overvoltage, impact current and current voltage instantaneous
distortion.

3.4.4. Other keys


The other keys with following functions:
Item Function Detail Chapter

Setting key: Analyzer setting and capture mode parameter setting, can setting date and time,
display, type of connection to the network, voltage ratio, current sensor select, choice of
§5
thresholds to be detected, definition of alarms to be detected, choice of parameters to be
detected.
Snapshot mode: snapshot of current screen and retrieval of screens already stored. §12
Help key: provides information about the functions and the symbols used for the current
§13
display mode.

3.5. Wiring
3.5.1. Measurement connectors
The wire connectors are located on the top of the Analyzer. These connectors are distributed as follows:
5 voltage test ports: L1,
L2, L3, GND, N. Charge indicator
(GND is ground lead; N is
neutral lead)
Dedicated power adapter
port/ charger port

4 current sensor ports:


L1, L2, L3, N USB port

Figure 3-3: The connection port on the top

3.5.2. Charge interface & USB interface


Use only the dedicated power adapter and USB communication cable supplied with the Analyzer.

3.6. Power Supply


3.6.1. Battery level indication
The battery symbol is at the top right corner of the screen and can show the battery level. The number of bars
indicates the current battery level.
Symbol Charge State

Battery fully charged

9
Low battery

Moving bars: battery charging

The Analyzer is powered from an external power supply (the battery is fully charged)
If the battery power is too low, the Analyzer will display the following message “Low battery! The Analyzer will
turn OFF soon”. If the Analyzer is not charged by connecting the dedicated power adaptor in time, the Analyzer
will shut down after one minute.

3.6.2. Battery working time


The working time of battery is 8 hours when the battery is fully charged.

3.6.3. Recharge the battery


The battery is charged by the supplied special power adapter. The power charging connector is shown in
Figure 3-3. Please charge the battery using use the supplied special adaptor that meets the safety standard.
For battery without power, please charge it for about 5 hours. After charging is completed, the Analyzer will use
the external power supply first and not consume the rechargeable battery power. The charger indicator light is
on to indicate that the battery is charging, and the charger indicator light is off when the battery is fully charged.

3.6.4. The battery


Special lithium battery (9.6V), 4,500 mAh. The battery is built in the battery compartment.
3.6.5. Power supply by special power adapter
When the Analyzer is running from the dedicated power adapter, the battery is optional. However, there is a
risk of data loss if the main power supply is cut off during recording (when no battery is available).

3.7. The Bracket


The retractable bracket (Figure 3-4) on the back of the Analyzer keeps the Analyzer at an angle of 60°.

Retractable bracket

Battery cover

Figure 3-4: Battery cover and bracket (the Analyzer back side structure)

3.8. Functions Summary


3.8.1. Measurement functions
„ The AC voltage value between the devices (up to 1000V).
„ The AC current value (including neutral wire, up to 1000A).
„ Sustaining voltage and current value (including neutral wire).
„ Minimum and maximum half-cycle RMS value of voltage and current.
„ Peak of voltage and current (including neutral wire).
„ Power grid of 50 Hz or 60 Hz.
„ The peak factor of voltage and current (excluding neutral).
„ Calculation of the K factor (KF) (applied to calculate the harmonics current of transformers).
„ Current and voltage distortion factor (DF) (excluding neutral).
„ The total harmonic distortion of voltage and current (excluding neutral).
„ Active power, reactive power (capacitive and inductive), apparent power of each phase (excluding neutral).
„ Power factors (PF) and phase different power factors (DPF) (excluding neutral).
„ Short-term voltage flicker (PST) (excluding neutral).
„ Active, reactive (capacitive and inductive), and apparent energy (excluding neutral).
„ Current and voltage harmonics (excluding neutral) up to 50th order: harmonic ratio, RMS value, minimum
and maximum, and sequence harmonics.
„ Apparent power harmonic (50th order): harmonic ratio, RMS value, minimum and maximum.
„ The starting current and surge current of motor.

10
3.8.2. Main functions
„ Displaying waveform (voltage and current).
„ “Starting Current” function: displays the parameters used to study the motor starting.
䖩Instantaneous current value at which the cursor points.
䖩Maxim um instantaneous current value (the whole starting time).
䖩The semi-period current TRMS value of the cursor point.
䖩Maxim um semi-period current TRMS value (the whole starting time).
䖩The motor starting time.
„ Screen shot (60 groups maximum).
„ Transients function. Detect and record transient event records of power grid parameters in a period of time
(set the start and end times of the detection schedule) (up to 150 groups), store 4 complete period (Before
transient triggering: 1 period; After triggering: 3 periods).
„ Trend diagram record function (2GB memory card which with the time, date, the start and end time of setting
record--- max can record 100 groups). With histogram or curve shows the recorded parameters values and
averages in chronological order.
„ Alarm function. Alarm record log (up to 12,800 groups) (triggered according to the setting threshold value);
Set the start and end times of alarm monitoring; Displays the trigger channel of the alarm, the maximum and
minimum value after the trigger, and the duration.

3.8.3. Setting function


„ Date and time setting.
„ Screen brightness and contrast setting.
„ Choice of curve colors.
„ Choice of reactive power and reactive energy calculation mode (with or without harmonics).
„ Choice of wirings of power grid (single-phase, split-phase, 3-phase 3 wires or 3-phase 4-wires, 3-phase
5-wire).
„ Choice of voltage measurement ratio and current clamp sensors.
„ Transient monitoring of voltage and current trigger threshold setting.
„ Monitoring parameters of trend diagram setting.
„ Choice of alarm monitoring parameters, trigger threshold setting.
„ Data delete (total or partial).
„ Display the Analyzer software and hardware version numbers.
„ Choice of language (Chinese/English).

3.9. Abbreviations
Meanings of the symbols and abbreviations used:
Symbol Designation Symbol Designation

AC and DC components MAX Maximum true RMS

AC component only MIN Minimum true RMS

DC component only ms Millisecond (unit)

ij Phase angle PEAK Voltage & current maximum (+) or minimum(-) peak

Inductive phase shift PF Power factor

Capacitive phase shift PST Short-term voltage flicker

° Degree RMS True RMS value (current or voltage)

™ Sum of each phase values t Time

L Phase (line) Tan Tangent

% Percentage THD Total harmonic distortion

A Ampere Ucf Line voltage peak factor

Acf Current peak factor Uh Line voltage harmonic

Ah Current harmonic Urms Line voltage true RMS

Akf K Factor (for transformers) Uthd Line voltage total harmonic distortion

11
Arms Current true RMS Uunb Line voltage unbalance (3 )

Athd Total harmonic distortion of current V Phase voltage

Aunb Current unbalance (3 ) VA Apparent power

AVG Average value. Vah Apparent energy

CF Peak factor (current or voltage). VAR Reactive power

DC DC component (current or voltage). VARh Reactive energy

DPF Displacement power factor Vcf Voltage peak factor

Hz Power grid frequency Vrms Phase voltage true RMS

KF Refer to Akf Vthd Voltage to neutral total harmonic distortion

W Active power Vunb Phase voltage unbalance (3 )

Wh Active energy

4. USE
The Analyzer must be set up before it can be used. Please refer to Chapter 5 of this manual for setting content.
Please comply with the following precautions when using the Analyzer:
„ Do not measure voltage to earth exceeding 1,000V RMS.
„ Before installing or removing the rechargeable battery, please ensure that the Analyzer is disconnected from
any test wires and shut down.
4.1. Start-Up
Press the key to start up the Analyzer.
After about 3 seconds, the Waveform screen is displayed.

Figure 4-1: Waveform screen


When the battery power is full, the power of the Analyzer is supplied by battery. If not, the Analyzer will display
the alarm message “Low battery! The Analyzer will turn OFF soon” (refer to § 3.6). The power of the Analyzer
can be supplied by the special external power source (figure 3-3), without the use of battery.

4.2. Setting
To set the configuration parameters and test parameters of the Analyzer, please proceed as follows:
䖩 Press key, then the Analyzer displays the setting interface.
䖩Press or key to select the parameter to be modified, then press to enter the sub-menu.

Figure 4-2: Setting screen


12
Press or and or to browse and press key to confirm in the displayed sub-menu. For specific
operations, please refer to §5.3–§ 5.10.
Note: The following points must be checked or adjusted for each measurement:
Function Reference

Define the parameters calculation methods.(reactive power/ reactive energy) §5.5

Select the type of wiring (single- phase ~ three-phase, five-wire) §5.6

Select voltage ratios and current clamp sensor type §5.7

Transient triggering threshold (transients mode) §5.8

Parameters to be recorded (trend mode) §5.9

Definition of alarm thresholds §5.10


Press to return to the configuration menu screen.
4.3. Wire Connection

Figure 4-3 test connection on the top of device


Connect the measurement lines to the device as follows:
„ Current measurement: Connect the 4 current clamps to 4-channel current interfaces of L1/A, L2/B, L3/C,
N/D correspondingly. To ensure accurate measurement, please connect the current clamps correspondingly.
Select the applied current clamp before measurement (see § 5.7).
„ Voltage measurement: Connect the 5 voltage test wires to the 5 voltage input interfaces of
L1/A, L2/B, L3/C, E/GND, N/D correspondingly. Set the voltage ratio before measurement (see § 5.7).
The measurement line should be connected to the tested circuit as shown in the following diagrams.
4.3.1. Single-phase power grid

Figure 4-4: Single-phase connection


4.3.2. Three-phase 3-wire power grid

13
Figure 4-5: 3-phase 3-wire connection
4.3.3. Three-phase 4-wire power grid

Figure 4-6: 3-phase 4-wire connection

4.3.4. Three-phase 5-wire power grid

Figure 4-7: 3-phase 5-wire connection


4.3.5. Connection steps
䖩Power on the Analyzer.
䖩Configure the voltage ratio, select current sensor and the connection type of power grid.
䖩Connect the leads and current sensors to the unit.
䖩Connect the voltage test lead of ground lead and/or neutral lead to the ground lead and/or neutral lead of
power grid, and connect the corresponding current sensor.
䖩Connect the L1 phase lead to the network L1 phase and connect the corresponding current sensor.
䖩If applicable, repeat the procedure for phases L2, L3.
Note: Please comply with the connection steps to reduce connection errors to a minimum and avoid wasting
time.

4.4. Waveform Capture


Reminder: All screen interfaces can be saved (screenshot) by pressing the key. Refer to § 12 for related
information).
When the device start up and connected to the power grid (connect with the voltage test line and current
sensor), press key to waveform capture.
4.4.1. Display of the transients mode
Refer to §6.2.
4.4.2. Display of the inrush current mode
Refer to §6.3.

4.5. Display Harmonics


Reminder: All screen interfaces can be saved (screenshot) by pressing the key (refer to § 12).
When the device start up and connected to the power grid (connect well with the voltage test line and current
sensor), press key to display harmonic.
4.5.1. Voltage harmonics display
Refer to §7.2.
4.5.2. Current harmonics display
Refer to §7.3.

4.6. Waveform Measurement


Reminder: All screen interfaces can be saved (screenshot) by press the key (refer to § 12).
When the device start up and connected to the power grid (connect well with the voltage test line and current
sensor), press key to waveform measurement.
4.6.1. Display true RMS value
Refer to §8.2.
14
4.6.2. Display measurement value of total harmonic distortion
Refer to §8.3.
4.6.3. Display PEAK factor measurement
Refer to §8.4.
4.6.4. Display Min and Max RMS value, extreme value (voltage and current)
Refer to §8.5.
4.6.5. Simultaneous display all measurement parameter
Refer to §8.6.
4.6.6. Display vector diagram
Refer to §8.7.

4.7. Alarm Detection


Reminder: All screen interfaces can be saved (screenshot) by pressing the key (refer to § 12).
When the device start up and connected to the power grid (connect well with the voltage test line and current
sensor), press key to waveform measurement.
4.7.1. Parameter setting of alarm mode detection
Refer to §9.2. to set the alarm detection parameters and threshold value.
4.7.2. Alarm schedule setting
Refer to §9.3 to set the start and stop time.
4.7.3. Automatic stop
After reaching the stop date and time set by the operator, the alarm recording will be stopped automatically.
4.7.4. Manual stop
Refer to §9.3.3. If the preset stop date or time is not reached, the operator shall stop the detection manually.
4.7.5. Viewing the alarm log
Refer to §9.4.
4.7.6. Deleting the alarm log
Refer to §9.5.

4.8 Trend Chart Record Setting


Reminder: All screen interfaces can be saved (screenshot) by pressing the key (refer to § 12).
When the device is started up and connected to the power grid (connecting well with the voltage test line and
current sensor), press key for waveform measurement.
4.8.1. Trend chart monitoring parameters setting
Refer to § 10.3.
4.8.2 Trend chart schedule setting
Refer to § 10.2.

4.9. Energy Measurement


Reminder: All screen interfaces can be saved (screenshot) by pressing the key (refer to § 12).
When the device is started up and connected to the power grid (connect well with the voltage test line and
current sensor), press key for waveform measurement.
4.9.1. Power consumption measurement
Refer to §11.2.
4.9.2. Energy generated measurement
Refer to §11.6.

4.10. Data Communication With PC


The PC software can communicate with the Analyzer through the USB interface to upload and store the
measured data, which can be easily reviewed for future reference.
Note: The record data in the Analyzer will not be deleted after they are transferred to the PC, but copied to the
PC only, and the PC cannot read the Analyzer measurement data when the Analyzer is in any of the detection
states of trend chart logging, transient capture, start current detection or alarm detection (waiting or in
progress).

4.11. Delete Data


Before a new test, the memory can be freed up by deleting the stored data. Refer to §5.11.

4.12. Shut Down


Press the key to shut down the Analyzer.
If the Analyzer is in any of the detection states of trend chart recording, transient capture, start current
detection, or alarm detection (pending or in progress), the Analyzer will not shut down without confirmation,
and the following dialog box will pop up:

15
Are you sure to turn OFF the Analyzer?
Recording in progress or in standby
YES NO

Select Yes or No by pressing the or key. Press key to confirm selecting red font.
䖩 If No is selected, recording will continue.
䖩 If Yes is selected, only the record at the time before the Analyzer is turned off will be saved, and the
Analyzer is turned off after saving the records.

4.13. Power Supply


4.13.1. Recharge battery
Refer to §3.6.3.
4.13.2. Power supply operation
Refer to §3.6.5.

5. SETTING
Press key to enter the main setting menu. Before using the Analyzer, please set up the Analyzer and
confirm the parameters. In addition, the settings can be modified if necessary. The Analyzer setting information
is saved in the Analyzer after shutdown.

5.1. Setting Select Menus


Press key to select the sub-menu, press key to confirm, and press key to return to the main screen.

Figure 5-1: Sub-menu display interface


Name Sub-menu Refer
Date/Time Date and time setting. §5.3

Contrast/ Screen contrast and brightness setting. §5.4.1


brightness Voltage curve and current curve color setting §5.4.2
Calculation
Reactive parameters calculation selection (with or without harmonics). §5.5
Method
Select the type of electric connection to the power grid (attention:
Electric Wiring §5.6
calculations depend upon the type of connection type).
Select the current sensors (008B current clamp, 040B current clamp, 068B
§5.7.1
Sensor & Ratio current clamp, flexible current clamp).
Voltage ratio setting §5.7.2
Current threshold setting §5.8.1
Transient Mode
Voltage threshold setting §5.8.2
Trend Mode Setting the request monitoring parameters of the trend record §5.9
Alarm Mode Setting the request parameters, threshold of the alarm §5.10
Delete Data Delete some or all user data. §5.11
Serial number, software and hardware version numbers, and capacity of
About §5.12
on-board memory card.
16
5.2. Display Language
To select the display language, press the yellow key under the corresponding icon on the screen (Figure 5-1).
Select zh-CN for Chinese; and select en for English. Yellow background icon indicates the current language.

5.3. Date/Time
This menu defines the system date and time. The display is as follows:

Figure 5-2: Date/Time menu


The Date/Time area is highlighted in yellow.
„ Press key to change the date/time setting. The arrows indicate the present value can be changed,
press or key to increase or reduce the value.
Press or key to select the require change value, press key to confirm.
„ Change the date setting. Press or key select the date/time item and display as yellow highlighted,
and press key to confirm. The arrows indicate the present value can be changed.
Press or key to select DD/MM/YY or MM/DD/YY or YY/MM/DD, press key to confirm.
„ Press key to return to the setting main menu.

5.4. Display
5.4.1. Contrast/Brightness setting
Set the display contrast and brightness, as shown in the Figure below:

Figure 5-3: Contrast/Brightness menu


The selected field is highlighted in yellow.
„ Press or key to modify the contrast
„ Press or key move to the next item.
„ Press or key to change the brightness.
„ Pess key return to the main menu.
5.4.2. Color setting
The menu will display the colors of the voltage and current curves. The colors available are: yellow, orange,
red, pink, brown, green, dark green, light blue, sky blue, dark blue, light grey, grey.
The display screen is shown as follows:

Figure 5-4: Colors menu


The selected field will be highlighted in yellow.
„ Press or key to select the color of the voltage and current curves
„ Press or key move to next item.
„ Press key return to the main menu.
17
5.5. Calculation Methods
Menu sets the calculation of reactive parameters (power and energy) with harmonics or not.

Figure 5-5: Calculation methods menu


„ Press or key to select the harmonic or not.
䖩With harmonics: Including harmonic when calculating reactive quantities.
䖩Without harmonics: Excluding harmonic when calculating reactive quantities.
„ Press key to return to the main menu.

5.6. Electrical Wiring Selection


The menu is used to select the correct electrical wiring mode, as shown in figure5-6:

Figure 5-6: The Connection menu


Several electrical diagrams can be selected as follow:

Figure 5-7: Four type can be selected

Set up the wiring as the following steps:


„ Pressing or key to select Single-Phase, 2-Phase, 3-phse 3-wire or 3-phase 4-wire or
3-phase 5-wire connection.
„ Press key to return to the main menu.
5.7. Sensors & Ratio Selection
5.7.1. Current sensors
In menu, press and select yellow key A icon to choose the current clamp sensors and ratios. The device
can select 4 types of current clamp sensors, and also can choose optional current clamp sensor, and set the
turns ratio according to requirements.

Figure 5-8: Current clamp and ratios screen in the Sensors and ratios menu
18
Optional current sensors:
008B current clamp: 10mA~ 10A

040B current clamp: 0.10A~ 100A

068B current clamp: 1.0A~ 1000A

300F Flexible Current Sensor (with Integrator) : 10A ~ 6000A

300F flexible current sensor, integrator mounted in handle, 2-in-1 mode:


„ 4 pcs of 300F flexible current sensors corresponds to L1,L2,L3,N:
Insert the 300F audio plug into the current clamp input port of the power quality Analyzer.
„ Toggle the switch to the right to turn on the power of the flexible current sensor, the "POWER"
indicator will be on, then toggle the switch to the left to turn off the power of the flexible current sensor.
Note: The power of the flexible current sensor is supplied by the 6F22 9V dry battery. The lit “BAT LOW”
indicator light on the right side indicates that the battery voltage is low and please replace the battery.
5.7.2. Voltage ratio
In menu, press and select the yellow V icon to set the voltage ratios.

Figure 5-9: Voltage ratios setting menu


The ratios of all channels can be set as the following:
„ (4V, 1/1) The ratios of 4 channels are 1:1.
„ (4V) The ratio of 4 channels are same.
„ (3V+VN)L1\L2\L3 is the same ratio. The ratio of N line is set independently.
„ (V1+V2+V3+VN) The ratios of 4 channels are set respectively.
䖩Press key and then the or arrow keys appear in the ratio setting. Press or key to select 4
types of ratio combinations, then press key again to confirm.
䖩Press or key to select the corresponding ratio value, which will be highlighted in yellow.
䖩 Press key to enter ratio setting mode, and then the arrow keys appear.
䖩Press or key to select the position which is to be changed, press or key to increase or decrease
the value, then press key to confirm.
䖩Press key to return to setting main menu.

5.8. Transient Capture Setting


In the menu, the captured transient current and the trigger threshold of captured voltage can be set.
5.8.1. Current thresholds
In interface, press the yellow A icon to select the current thresholds of transient capture.

Figure 5-10: Current thresholds setting menu


䖩The arrow indicates the channel of cursor.
䖩Press or key to select the channel, then press key (with arrow key displayed) to start
modification.

19
䖩Press or to select the change position and threshold current unit (mA/A), then press or key to
increase or decrease the value.
䖩Press key to confirm.
5.8.2. Voltage thresholds
In interface, press the V icon to select the voltage thresholds of transient capture.
All the voltage threshold can be set with the three combined modes below:
„ (4V)—The thresholds of 4 channels are same.
„ (3V+VN) –The threshold of L1\L2\L3 are same. The threshold of N line is set independently.
„ (V1+V2+V3+VN)—The thresholds of 4 channels are set respectively.

Figure 5-11: Voltage thresholds setting menu


䖩Press key and then the arrows key appears in the ratio setting. Press or to select the above 3
different threshold combination modes, then press to confirm.
䖩Press or key to select the corresponding require thresholds, which will be highlighted in yellow.
䖩Press key to enter into threshold setting (with the arrows displayed), then start to modify.
䖩Press or to select the required change position and threshold unit (V/kV), press or key to
increase or decrease the value.
䖩Press key to return to the main menu.

5.9. Trend Monitoring Parameter Setting


The Analyzer has a trend chart recording function (refer to § 10), which records the measured or calculated
values (e.g. Urms, Vrms, Arms, etc.). The menu is used to set the parameters to be monitored in the trend
graph recording.
Press the corresponding function icon , , , to select the required monitoring parameters
groups. The selected parameters groups are identified by the icon with a yellow background.
The configuration example is shown in the Figure below:

Figure 5-12: Trends chart record parameter selection


䖩Press the icon function key to set the monitoring parameters group 1. The selected icon is displayed
with yellow background.
䖩Press or and or key to move the cursor.
䖩Press to select/cancel. Red solid dot indicates selected, and red hollow dot indicates unselected.
Optional monitoring parameters are available:
Unit Description
Urms True RMS phase-to-phase voltage.
Uthd Harmonic distortion of the phase-to-phase voltage (2 , 3 ).
Ucf Crest (peak) factor of phase-to-phase voltage (2 , 3 ).
Uunb Phase-to- phase voltage unbalance (2 , 3 ).
Hz Network frequency.
Vrms True RMS phase-to- neutral voltage.
Vthd Total harmonic distortion of the phase-to-neutral voltage.
Vcf Crest factor of phase-to-neutral voltage.
Vunb Phase-to-neutral voltage unbalance (2 , 3 ).

20
PST Short-term flicker.
Arms True RMS current.
Athd Total harmonic distortion of the current.
Acf Crest factor of current.
Aunb Current unbalance (2 , 3 ).
KF K factor.
W Active power.
VAR Reactive power.
VA Apparent power.
PF Power factor.
DPF Displacement power factor.
Tan Tangent.
? See comment below (relate to the harmonic).
The last two lines need special settings, shown as follows:

Figure 5-13: These two lines involve harmonics


These two lines relate to the monitoring values of harmonics VAH, Ah, Vh and Uh. The user can independently
select the measured harmonic order (0 to 50th) to record the harmonic ratio of corresponding order, and can
also choose to record the odd harmonics only. Specific operations are as follows:
„ Select the monitoring harmonic parameters: the selected highlighted display in yellow, press key
and appear the arrows key , press or key to select the require parameters (VAh, Ah, Vh, and Uh),
“?” indicate is unselected. Press key to confirm, the selected values field is highlighted in yellow.
„ Press to go to the next option.
„ Select the starting harmonic order: the relate option highlighted in yellow. Press key and appear the
arrows key. Press or key to increase or decrease the starting harmonic order, then pressing
key to confirm. Press key to go to the next option.
„ Select the finishing harmonic order: (the finishing harmonic order must be higher than or equal to the
starting harmonic order) highlighted in yellow. Pres key and appear arrows key. Press or key
to increase or decrease the finishing harmonic order, then press key to confirm.
„ Press to go to the next option.
„ Record odd harmonics only: press key to select or cancel to record the odd harmonic only, the solid
red dot means selected:
Selected: Record only the odd harmonic of the two harmonic order.
Not selected: Record all harmonic order (include the odd harmonic).

Figure 5-14: Record harmonic order setting


„ Press key to return to setting main menu.
Set up another group of trend chart recording & monitoring parameter settings in the same way.

5.10. Alarm Monitoring Parameter Setting


The menu is used to set the parameters and thresholds to be monitored in alarm mode (refer to § 9). User
can set 40 groups of alarm monitoring parameters.

Figure 5-15: Alarm mode menu

21
„ Use the or key to select different setting alarm parameter groups.
„ Press key to select present position parameter and then arrows key appear.
„ Press or key to select the setting parameters (Vah, Ah, Uh, etc., refer table in § 5.9), then press
key to confirm. The selected parameter is highlighted in yellow.
„ Press or keys to horizontally select each parameter, then press key to confirm (with the arrows
key displayed. Press or key to select the required setting value, then press key to confirm.
Set the other values of each parameter in the same way.
Define each alarms (optional):
䖩The alarm types (Vah, Ah, Uh, Vh, Tan, PF, DPF, VA, VAR, W, Athd, Uthd, Vthd, KF, Hz, Aunb, Vunb, Vrms,
Acf, Ucf, Vcf, PST, Arms, Urms and Vrms) (refer the abbreviations in § 3.9).
䖩The harm onic order range (between 0 and 50, for Vah, Ah, Uh and Vh value).
䖩The alarm trigger channel (3L: 3-phases, L1, L2, L3 can be triggered individually; N: neutral can be
triggered).
䖩The alarm judgement direction (> or < can be selected for Arms, Urms, Vrms, Hz only; otherwise only one
direction).
䖩The alarm triggering threshold (The threshold unit of W, VAR, VA, Arms, Urms, and Vrms can be set).
䖩E xceed threshold trigger m in. time (can set 0s - 99 minutes).
䖩The alarm hystersis value (The percentage increases or decreases from the corresponding alarm threshold,
optional values can be selected with 1%, 2%, 5% or 10%. If the selected percentage is exceeded, the
alarm will be stopped. Please refer §17.2).
„ Press key to select or cancel the alarm parameter groups. Red solid dot indicates selected, and red
hollow dot indicates unselected.
„ Press or icons of the yellow background function key to display the alarm monitoring parameter
groups of next page or previous page.
„ Press key to return to setting main menu.

5.11. Delete Data


The menu is used to delete partial or all of the data stored in the Analyzer (trend chart record, transient
record, starting current, alarm, screenshot, Analyzer setting, monitoring parameter setting, etc.).

Figure 5-16: Delete data menu


„ Partial deletion:
䖩Press or key to select the option to be deleted. The selected option will be highlighted in yellow.
䖩Press to select/unselect. Red solid dot indicates selected, and red hollow dot indicate unselected.
Note: If the setting parameter is selected for deletion, the Analyzer screen will display message “After the
configuration is deleted, the device will be turned off”.
䖩Press icon function key, then the icon will be highlighted in yellow background and to indicate
the “ready-to-delete” state. Press the icon again to cancel the delete state.
䖩In the state of ready to delete, press key to confirm deleting the selected parameters.
䖩Press key to return to setting main menu.

„ Delete all:
䖩Press icon function key to select all the stored data (Red solid dot indicates selected), then the screen will
display the message “After the configuration is deleted, the device will be turned off”.
Note: Once the set parameter is selected for deletion, the message “After the configuration is deleted, the
device will be turned off” will be displayed on the screen.
䖩 Press icon function key to cancel all the selection, red hollow dot indicate is unselected.
䖩Press key return to setting main menu.

5.12. About
The screen will display the serial number of the device, firmware version, DSP software version, icon version,
and SD card capacity.

22
Figure 5-17: About menu

Press key to return to the main setting menu.

6. WAVEFORM CAPTURE MODE


6.1. Sub-Mode Selection
The waveform capture sub-modes are listed in the screen below, and the following section will be introduced
individually.

Transient capture (refer § 6.2).

Inrush current mode refer §


6.3).

Figure 6-1: Waveform capture mode


Enter the sub-modes as the following steps:
䖩 Press or key to select, the selected mode will be highlighted in yellow.
䖩 Press key to confirm.
䖩Press key to return to waveform capture main menu.

6.2. Transient Measurement Mode


Transient measurement mode can be used to detect transient event records, browse transient records, and
select the deletion of transient records. 150 groups of transient measurement records can be saved.

Battery remaining capacity


Present mode
Present date & time
Present memory used, black
indicates used memory, white
indicates remaining memory Open the detection record list

Set detect parameters (refer § 6.2.1)

Set the current & voltage Confirm and start detection


threshold
Figure 6-2: Detection schedule in transient detection mode

6.2.1. Setting & starting detection


Press icon function key, and then the detection schedule is displayed.

23
Present mode The start and stop times of transient
detection.

Number of transient records


Indicates that detection starting
File name

The icon is stop key displayed


when the program starts.

Figure 6-3: The analyzer are performing transient schedule

6.2.1.1. Step 1: Parameters setting


䖩Press or key to select the start time (the selected field is highlighted in yellow). Press key to enter
the value setting mode, and the arrows key appears at the start date & time optional item value.
䖩Press or to increase or decrease the value, and press or key move to next item.
Note: The start time must be later than the present time.
䖩Press key to confirm the start date and time of detection.
䖩Press or key to select the stop time(the selected field is highlighted in yellow), press key to enter
into setting values, then the arrows key appears at the stop date & time optional item value.
䖩Press or to increase or decrease the value, press or key move to next item.
Note: The stop date and time must be later than the start date and time.
䖩Press key to confirm the stop date and time.
䖩Setting the number of records in the same way. The total number of transient records that saved by the
Analyzer is 150 groups maximum.
䖩Set the file name in the same way. The file name can be composed of numbers 0-9 and letters A-Z, the
length of the file name is 8 characters at most.
䖩Press icon function key to enter into the interface of setting voltage and current threshold. After the
setting is completed, press key to return to the waveform capture mode directly.
6.2.1.2. Step2: Start detection
Press OK icon function key to start detection at the preset start and stop times.
䖩The OK icon disappears and the icon appears instead.
䖩Before the real time reaches the preset start time, “Detection on standby” will be displayed, and the icon
will flashes on the top display bar.
䖩W hen the real time reaches the preset start time, “Detection in progress” will be displayed.
䖩After the real time reaches the preset stop time and the detection is completed, the detection schedule and
the OK icon (bottom right-hand corner of the screen) will be displayed again. At this time, the next detection
schedule can be set.
Note: The transient capture is related to voltage and/or current and is consistent with the setting trigger
threshold. If triggered by an event, the Analyzer will save the current waveform and voltage waveform.
Press key to return to the main menu of waveform capture.

6.2.1.3ˊStop detection
Before the real time reaches the preset stop time, the user can press the corresponding yellow function key
to stop the transient detection, and the OK icon will appear again at this time.

6.2.2. Display transient measurement record


The saved transient detection records are displayed by following the steps below:
䖩Press icon function key to display the transient detection record list, shown as follow:

24
Present mode Select transients curve require
display:
-ALL: display all transient records
-4V: display 4-wire voltage
Present memory used, black
transients record.
indicates used memory, white
-4A: display 3-wire current and
indicates remaining memory
neutral current transients record
-L1, L2, or L3: Corresponding the
1,2,3 phase to display each phase
Present page and total page
voltage and current record.
-N: display neutral current transient
record.
Press , icon function key
to browse the record list of up The icon function key select whether to filter
and down page the list of transient records.

Figure 6-4: Transient list screen

䖩Press or key to select the desired transient detection record. The selected line will be displayed in
bold font. Then press key to confirm. The related transient detection record will be displayed as curves.

Curve number

Present display
Press or key to
curve in record
select the displayed curve
position

: Zoom out

: Zoom in

Figure 6-5: Transient waveform curves


䖩Press or key to select the desired curve, press or key to move the cursor leftwards and
rightwards, or long press to move the cursor quickly.
䖩Press key to return to the transient detection record list.

6.2.3. Delete transient measurement record


The delete icon will be displayed only when displaying transient record list. The deletion steps are as
follows:

Figure 6-6: Delete transient measurement record


䖩Press or key to select the transient record to be deleted. The selected record will be displayed in
black bold font.
䖩Press icon function key to prepare deletion. The icon displayed in yell background indicates selected
and ready to delete. Press icon function key or key to cancel the delete state.
䖩In the state of ready to delete, press key to confirm deleting the record.

6.3. Inrush Current Mode


In inrush current mode, the Analyzer can monitor the current waveform and voltage waveform when the
electrical equipment starts up, and there are two submenus of RMS and PEAK in the capture record (refer to
6.3.2).
25
The Analyzer can store only one group of inrush current waveform capture chart.
6.3.1. Capture Schedule
Inrush current capture schedule is as follows:
Press key to enter into the corresponding menu and display the capture schedule interface.

Present mode

Setting capture start threshold

Select triggering channel


Select alarm hysteresis
: Open the
completed capture Setting capture start time
inrush current records.

OK:Confirm and start to capture


Figure 6-7: Setting the capture schedule of inrush current

6.3.1.1. Step 1: Parameter setting


Specific operations are as follows:
䖩Press or key to select the start threshold setting (the selected item will be highlighted in yellow),
Press key to enter and set relate values, then the arrows key appears on various start threshold
value.
䖩Press or to increase or decrease the value and press or to move to the next item and setting.
䖩Press key to confirm.
Operate in the same way to set the triggering channel, Hysteresis and Start time.
Note: For more information about the hysteresis, please refer to § 17.2.
6.3.1.2. Step 2: Start capture
Press OK icon function key to start capture with the pre-set date and time.
䖩The OK icon disappears and the icon appears instead.
䖩Before reaching the preset start time, the screen will display “capture pending” and the icon flashes in the
screen’s upper display bar.
䖩After reaching the preset start time and meeting the activation condition, the screen will display “Capture in
progress”, and the memory occupation indicator bar will be displayed at the top of the screen. (The black
bar is the memory used; the white bar is the memory available), the memory card occupancy indicator bar is
only displayed during the capture process and disappears after the capture is completed.
䖩If the stop conditions are m et (refer to 17.5) or if the memory of the Analyzer is full, the capture will stop
automatically and the "Capture Schedule" and OK icons will appear again.
Note: The Analyzer can only store one group of inrush current capture waveform, the user needs to delete the
previous capture before the next capture can be made.
Press key to return to the waveform capture menu.
6.3.1.3. User stops waveform capture
User presses icon (bottom right corner of the screen) function key to stop waveform capturing, and the OK
icon will appear in the same place.

6.3.2. Inrush Current Capture Parameter Display


Display the record parameters of inrush current capture by following the steps below:
„ Press icon function key to open the completed inrush current record, shown as follow:

Figure 6-8: Capture parameters interface

26
No. Function Refer
(1) RMS mode § 6.3.3
(2) PEAK mode § 6.3.4
„ Press the RMS or PEAK icon function key to select the capture parameter to be displayed in RMS mode or
PEAK mode. The Analyzer displays the corresponding current waveform and voltage, and the user can
move the cursor along the waveform curve to zoom in or zoom out to observe the waveform. Related
Information:
䖩 The cursor pointing on the curve are instantaneous current or voltage values.
䖩 The m axim um instantaneous current (whole inrush cycle).
䖩 Half-cycle current RMS value of the cursor mark point.
䖩 The m axim um half-cycle RMS current (whole inrush cycle).
䖩 Maximum instantaneous value in inrush cycle (PEAK).
䖩 The starting tim e and the motor starting period.
Caution: Voltage must be available before the motor can be started at a stable and correctly servo-controlled
frequency.

6.3.3. True RMS Current


In RMS mode, the Analyzer displays the half-cycle true RMS trend curve chart of the start-up process.
6.3.3.1. Displaying 3A in RMS mode
The related information are as follow:

Memory state AMAX: maximum half-cycle


RMS value
Current scale value Select the desired display curve chart:
3 V: displays the 3-phase voltage in
Cursor, press or inrush current mode
key to move 3 A: displays the 3-phase current in
inrush current mode.
L1, L2, L3: display the current &
voltage of phases 1, 2, and 3,
Instantaneous position respectively.
relative to the start point Hz: displays the frequency of power
of the inrush current (t=0) grid.
A1, A2, A3: the instantaneous
current of each phases at the
: Zoom out, : Zoom in
position of the cursor.

Figure 6-9: Displaying 3A in RMS mode

6.3.3.2. Displaying L1 in RMS mode


The related information are as follows:
Memory state MAX: maximum half-cycle RMS
value of the inrush current capture
Current & voltage scale value
V: measured voltage.
A: measured current.
Cursor, press or key to V1: the instantaneous voltage of
move
phase 1 at the position of the
cursor.
Instantaneous position A1: the instantaneous current of
relative to the start point of phase 1 at the position of the
the inrush current (t=0)
cursor.
: Zoom out, : Zoom in.

Figure 6-10: Displaying L1 in RMS mode


Note: L2 and L3 can display the true RMS trend curves of phase 2 and phase 3 half-cycle current and voltage
starting process, same as L1 display screen.

6.3.4. Instantaneous Value of Inrush Current in PEAK in mode


Displaying the envelope curve and the captured waveform of inrush current in PEAK mode.
6.3.4.1. Displaying 4A in PEAK mode
The related information are as follows:

27
Memory state AMAX: maximum half-cycle RMS
value of the inrush current
Peak absolute value
capture
Current scale value Select desired display curve
chart:
Cursor, press or key to 4 V: display 3-phase and neutral
move voltages
Instantaneous position 4 A: display 3-phase and neutral
relative to the start point current.
of the inrush current (t=0) L1, L2, L3: display the current &
voltage of phases 1, 2, and 3,
A1, A2, A3: the instantaneous respectively.
current of each phase at the N: display the neutral current and
position of the cursor voltage.
: Zoom out, : Zoom in.

Figure 6-11: Displaying 4A in PEAK mode

6.3.4.2. Displaying A1 PEAK


The related information are as follows:

Memory state MAX: maximum half-cycle RMS


value.

Current scale value


Cursor, press or key to A1: the instantaneous current of
move phase 1 at the position of the
cursor

Instantaneous position
relative to the start point of
the inrush current (t=0) : Zoom out, : Zoom in.

Figure 6-12: Displaying A1 PEAK

Note: A2 and A3 can display the current envelope curve record of phase 2 and phase 3, the above is the
display screen of A1. V1, V2, V3 display the voltage envelope curve chart of phase 1, phase 2, and phase 3,
similar to the display screen of A1.

7. HARMONICS MODE
In harmonic mode, the Analyzer can display each harmonic ratio of voltage, current and apparent power,
measure harmonic currents generated by nonlinear loads, and analyze the problems caused by harmonics
(heating case of neutral wires, conductors and motors, etc.).

7.1. Select Sub-Menu


The harmonic mode submenus are listed in the following screens and will be introduced separately in the
following sections.
Select the measurement type with the following function keys at the bottom of the screen.

Harmonic analysis of line voltage Press or key to select


each phase and expert mode

Harmonic analysis of apparent


power
:Zoom out
Current harmonic analysis

Harmonic analysis of phase


voltage
:Zoom in
Figure 7-1: Harmonic mode screen

28
7.2. Phase Voltage
This sub-menu displays the phase voltage harmonic.
Note: The display curve is determined by the wiring type (refer to §5.6):
Single-phase: no choice (L1 only).
2-phase: 2L, L1, L2.
3-phase 4wire or 5-wire: 3L, L1, L2, L3, -,+ (expert mode).
The 3-phase wiring is shown in the Figure, which is also valid for the other submenus.
7.2.1. Displaying 3L phase-voltage harmonic
The information are shown as below:
Present mode
Harmonic order selection. Press
or keys to move the cursor
Instantaneous frequency
Press or key to display
The cursor corresponding each phase and expert mode
harmonic order parameters.
Horizontal lines indicate the
Vh05: harmonic order
harmonics order (odd order)
%: Harmonic ratio relative to
Harmonic ratio relative to the
fundamental wave.
fundamental is shown as a
V: The harmonic RMS
percentage (fundamental: 1st
voltage
harmonic)
+000°: Phase shift relative to
DC: DC component
the fundamental wave
1~25: harmonic order.
Automatically displays 26th to 50th
order when the cursor exceeds
25th order
Figure 7-2: Displaying 3L phasevoltage harmonic
Expert mode displayed in 3 x 3L or L1, L2 and L3 (*) (3-phase wiring only - refer to §7.6)
7.2.2. Displaying L1 phase-voltage harmonic
The information are shown as below:
The cursor corresponding Harmonic order selection, Press
harmonic order parameters. or keys to move the cursor
Vh05: harmonic order
%: Harmonic ratio relative to Press or key to display
fundamental wave. each phase and expert mode
V: The harmonic RMS voltage
+000°: Phase shift relative to Horizontal lines indicate the
the fundamental wave harmonics order (odd order)
Max, min: harmonic ratio of Harmonic ratio relative to the
maximum and minimum (press fundamental is shown as a
key to reset). percentage (fundamental: 1st
THD: total harmonic distortion. harmonic)
DC: DC component
1~25: harmonics order.
Automatically displays 26th to
50th order when the cursor
exceeds 25th order
Figure 7-3: Displaying L1 phase-voltage harmonic
Note: L2 and L3 display the phase voltage harmonics of phase 2 and phase 3 respectively, same as the L1
screen display.
7.3. Current
The sub-menu displays the current harmonics.
7.3.1. Displaying 3L current harmonic
The information are shown as below:
Harmonic order selection, Press
Present mode or keys to move the cursor
Instantaneous frequency
Press or key to display
The cursor corresponding each phase and expert mode
harmonic order parameters Horizontal lines indicate the
Ah03: harmonic order harmonics order (odd order)
%:Harmonic ratio relative to Harmonic ratio relative to the
fundamental wave fundamental is shown as a
A: Harmonic RMS current percentage (fundamental: 1st
+000°: Phase shift relative to the harmonic)
fundamental wave 1~25: harmonics order.
Automatically displays 26th to 50th
order when the cursor exceeds 25th
Figure 7-4: Displaying 3L current harmonic order
29
7.3.2. Displaying L1 current harmonic
The information are shown as below:
Present mode Harmonic order selection, Press
or keys to move the cursor
Instantaneous frequency Press or key to display
each phase and expert mode
The cursor corresponding
harmonic order parameters .
Horizontal lines indicate the
Ah03: harmonic order harmonics order (odd order)
%:Harmonic ratio relative to Harmonic ratio relative to the
fundamental wave fundamental is shown as a
A: Harmonic RMS current percentage (fundamental: 1st
+000°: Phase shift relative to harmonic)
the fundamental wave 1~25: harmonics order.
Automatically displays 26th to 50th
order when the cursor exceeds
25th order
Figure 7-5: Displaying L1 current harmonic
Note: L2 and L3 display the current harmonics of phase 2 and phase 3 respectively, same as the L1 screen
display.

7.4. Apparent Power


The sub-menu displays apparent power harmonic.
7.4.1. Displaying 3L apparent power harmonic
The information are shown as below:

Present mode
Harmonic order selection, Press
Instantaneous frequency or keys to move the cursor

The cursor corresponding Press or key to display


harmonic order parameters each phase and expert mode
VAh03: harmonic order Horizontal lines indicate the
%:Harmonic ratio relative to harmonics order (odd order)
fundamental wave Harmonic ratio relative to the
+000°: Phase shift relative to the fundamental is shown as a
fundamental wave percentage (fundamental: 1st
: indicates that the harmonic harmonic)
generates electrical energy 1~25: harmonics order.
:Indicates that the harmonic Automatically displays 26th to 50th
consumes electrical energy order when the cursor exceeds
25th order
Figure 7-6: Displaying 3L apparent power harmonic

7.4.2. Displaying L1 apparent power harmonic


The information are shown as below:
Present mode
Harmonic order selection, Press
The cursor corresponding or keys to move the cursor
harmonic order parameters cursor.
VAh03: harmonic order Press or key to display
%:Harmonic ratio relative to each phase and expert mode
fundamental wave
+000°: Phase shift relative to Horizontal lines indicate the
the fundamental wave harmonics order (odd order)
: indicates that the Harmonic ratio relative to the
harmonic generates electrical fundamental is shown as a
energy percentage (fundamental: 1st
:Indicates that the harmonic)
harmonic consumes electrical 1~25: harmonics order.
energy. Automatically displays 26th to
Max,min:harmonic ratio of 50th order when the cursor
maximum and minimum, press exceeds 25th order
key to reset
Figure 7-7: Displaying L1 Apparent power harmonic L1

Note: L2 and L3 display the apparent power harmonics of phase 2 and phase 3 respectively, same as the L1
screen display.

30
7.5. Line Voltage
The corresponding submenu displays the line voltage harmonic and there are 2-phase and 3-phase wirings
only.
7.5.1. Displaying 3L line voltage harmonic
The information are shown as below:
Present mode Harmonic order selection, Press
or keys to move the cursor
Instantaneous frequency Press or key to display
each phase and expert mode
The cursor corresponding Horizontal lines indicate the
harmonic order parameters harmonics order (odd order)
Uh03: harmonic order Harmonic ratio relative to the
%:Harmonic ratio relative to fundamental is shown as a
fundamental wave percentage (fundamental: 1st
V: Harmonic RMS voltage. harmonic)
+000°: Phase shift relative to DC: DC component
the fundamental wave 1~25: harmonics order.
Automatically displays 26th to 50th
order when the cursor exceeds
25th order
Figure 7-8: Displaying 3L line voltage harmonic

7.5.2. Line voltage harmonic L1 display


The information are shown as below:
The cursor corresponding
harmonic order parameters Harmonic order selection, Press
Uh03: harmonic order or keys to move the cursor
%:Harmonic ratio relative to
Press or key to display
fundamental wave
each phase and expert mode
V: Harmonic RMS voltage.
+000°: Phase shift relative to the Horizontal lines indicate the
fundamental wave harmonics order (odd order)
Max,min: harmonic ratio of Harmonic ratio relative to the
maximum and minimum, press fundamental is shown as a
key to reset percentage (fundamental: 1st
THD: total harmonic distortion. harmonic)
DC: DC component
1~25: harmonics order.
Automatically displays 26th to 50th
order when the cursor exceeds
25th order
Figure 7-9: Line voltage harmonic L1 display screen

7.6. Expert Mode


The expert mode is only available for 3-phase connections. This mode is used to show the harmonic effects of
the neutral line heating or the rotating motors. Press the or keys to select the expert mode. The
selected item will be highlighted in yellow and displays the expert mode on the screen.
In the expert mode screen, two sub-menus of and can be selected (refer to next page).
7.6.1. Expert mode display phase voltage
Submenu shows the effect of phase voltage harmonics in neutral heating or in rotating motors.
The information are shown as below:

Include a positive sequence


Include a zero sequence
harmonic
harmonic
Include a negative
sequence harmonic

%: Harmonic ratio relative to


fundamental wave

Figure 7-10: Phase voltage expert mode display

31
7.6.2. Expert mode display current
Submenu shows the effect of current harmonics in neutral heating or in rotating motors.
The information are shown as below.
Include a zero sequence
harmonic
Include a positive
sequence harmonic
Include a negative
sequence harmonic

%: Harmonic ratio relative to


fundamental wave

Figure 7-11: Current expert mode display

8. WAVEFORM MODE
In waveform mode, the Analyzer displays voltage and current curve chart, measured values and, calculated
values (excluding power, energy and harmonics).

8.1. Select Sub-menus


The sub-menus are listed on the screen below and described individually in the paragraphs that follow.
Users can select the type of measurement by pressing the yellow function key at the bottom of the screen.

Present mode Press or key to select


the requested display curve
Measure voltage, current
Display (the RMS value, peak
peak factor
value, maximum value and
minimum value of voltage or
Measure total harmonic current)
distortion Simultaneous display all the
measured values of voltage or
Measure the true RMS value current (RMS, DC, THD, DF,
of voltage and current CF, PST)
Display the phase chart

Figure 8-1: Waveform mode screen

8.2. True RMS


The submenu can display the waveform of the measured signal and the true RMS value of voltage and current
within a certain period of time.
Note: Which curve is displayed depends on the type of external wiring (refer to § 5.6):
„ Single-phase: no choice (L1)
„ 2-phase: 2V, 2A, L1, L2
„ 3-phase, 3- or 4-wire: 3U, 3V, 3A, L1, L2, L3
„ 3-phase 5-wire:
䖩For , and : 3U, 3V, 3A, L1, L2 and L3
䖩For , and : 3U, 4V, 4A, L1, L2, L3 and N
The wiring type of the screen shown is 3-phase 5-wire.

Including a zero
sequence harmonics
Including a positive sequence
harmonics
Including a negative
Sequence harmonics

%˖Harmonic distortion relative to the


fundamental.

32
8.2.1. 3U RMS display screen
The screen showed displays the 3-phase voltage of 3-phase system.
The information are shown as below:

Instantaneous frequency Line voltage waveform

Phase voltage true RMS Instantaneous value of the


waveform to which the cursor
Cursor. Press or key to is pointing
move t: The time corresponding the
Long press to move quickly start stage (milliseconds)
U1: Line voltage
instantaneous value between
Voltage value axis line phase 1 and phase 2 (U12)
(Automatic scale adjustment) U2: Line voltage
instantaneous value between
phase 2 and phase 3 (U23)
Figure 8-2: 3U RMS display screen U3:Line voltage instantaneous
value between phase 3 and
phase 1 (U31)

8.2.2. 4V RMS display screen


The screen showed displays the 3-phase phase voltage and neutral line to earth voltage of 3-phase system.
The information are shown as below:

Instantaneous frequency Phase voltage waveform curve

Phase voltage true RMS


Waveform instantaneous value
pointed by the cursor.
Cursor. Press or key to t: Relative start time (ms).
move. V1: Phase voltage instantaneous
value of L1.
V2: Phase voltage instantaneous
Voltage value axis line value of L2.
(Automatic scale adjustment) V3: Phase voltage instantaneous
value of L3.
VN: Neutral line earth voltage
instantaneous value.
Figure 8-3: 4V RMS display screen

8.2.3. 4A RMS display Screen


This screen displays the three phase currents and the neutral current of three-phase system.
The information are shown as below:
Instantaneous frequency of Current waveform curve
network.

Waveform instantaneous value


RMS currents
pointed by the cursor.
t: Relative start time (ms).
Cursor. Press or key to A1: Current instantaneous value of
move
L1.
Long press to move quickly.
A2: Current instantaneous value of
L2.
Current value axis line A3: Current instantaneous value of
(Automatic scale adjustment) L3.
AN: Neutral line current
Figure 8-4: 4A RMS display screen instantaneous value.

8.2.4 Neutral line RMS display screen


This screen displays current and earth voltage of neutral line.
The information are shown as below:

33
Instantaneous frequency RMS currents
of network.

RMS voltage Waveform instantaneous value


pointed by the cursor.
t: Relative start time (ms).
Cursor. Press or key to move VN: Earth voltage instantaneous
value of neutral line.
AN: Current instantaneous value of
Current and voltage value axis neutral line
line
(Automatic scale adjustment)

Figure 8-5: Neutral line RMS display screen

Note: L1, L2, and L3 display the current and voltage in phases 1, 2, and 3, respectively. The screen is identical
to the one displayed for the neutral.

8.3. Total Harmonics Distortion


The submenu can display the measured signal waveform within a certain period of time and the total
harmonic distortion rate of voltage and current.
8.3.1. 3U THD display screen
This screen displays the phase voltage waveforms and harmonic distortion rate within a certain period.
The information are shown as below:
Instantaneous frequency of network Phase voltage waveform curve

Waveform instantaneous value


Harmonic distortion rate of each
pointed by the cursor.
curve t: Relative start time (ms).
U1: Phase voltage instantaneous
Cursor. Press or key to move value between phases 1 and phase 2.
U2: Phase voltage instantaneous
value between phases 2 and phase 3.
Voltage value axis line U3: Phase voltage instantaneous
(Automatic scale adjustment) value between phases 3 and phase 1.

Figure 8-6: 3U THD display screen


8.3.2. 3V THD display screen
This screen displays the phase voltage waveforms and harmonic distortion rate within a certain period of time.
The information are shown as below:

Instantaneous frequency of Phase voltage waveform curve


network

Harmonic distortion rate of Waveform instantaneous value


each curve pointed by the cursor.
t: Relative start time (ms).
Cursor. Press or key to V1: Phase voltage instantaneous value
move of L1.
V2: Phase voltage instantaneous value
of L2.
V3: Phase voltage instantaneous value
Voltage value axis line of L3.
(Automatic scale adjustment)

Figure 8-7: 3V THD display screen

8.3.3. 3A THD display screen


This screen displays the current waveforms and harmonic distortion rate within a certain period.
The information are shown as below:

34
Instantaneous frequency of Current waveform curve
network.

Harmonic distortion for each Waveform instantaneous value


curve. pointed by the cursor.
t: Relative start time (ms).
Cursor. Press or key to A1: Current instantaneous value
move of L1.
A2: Current instantaneous value
of L2.
Current value axis line A3: Current instantaneous value
(Automatic scale adjustment) of L3.

Figure 8-8: 3A THD display scree


Note: L1, L2, and L3 display the total harmonic distortion rate of voltage and current for phases 1, 2and 3
respectively.

8.4. PEAK Factor


The submenu can display the measured signal waveform within a certain period of time and the peak
factor of voltage and current.

8.4.1. 3U CF display screen


This screen displays the phase voltage waveforms and peak factor within a certain period.
The information are shown as below:

Instantaneous frequency of Phase voltage waveform curve


network.

Waveform instantaneous value


Peak factor of each curve.
pointed by the cursor.
t: Relative start time (ms).
Cursor. Press or key to U1: Phase voltage instantaneous
move value between phases 1 and phase 2.
U2: Phase voltage instantaneous
value between phases 2 and phase 3.
U3: Phase voltage instantaneous
Voltage value axis line value between phases 3 and phase 1.
(Automatic scale adjustment)

Figure 8-9: 3U CF display screen

8.4.2. 3V CF display screen


This screen displays the phase voltage waveforms and peak factor within a certain period.
The information are shown as below:
Instantaneous frequency of Phase voltage waveform curve
network.

Waveform instantaneous value


Peak factor of each curve
pointed by the cursor.
t: Relative start time (ms).
Cursor. Press or key to
V1: Phase voltage instantaneous
move
value of L1.
V2: Phase voltage instantaneous
value of L2.
V3: Phase voltage instantaneous
Voltage value axis line value of L3.
(Automatic scale adjustment)

Figure 8-10:3V CF display screen

8.4.3. 3A CF display screen


This screen displays the current waveforms and peak factor within a certain period.
The information are shown as below:

35
Instantaneous frequency of Current waveform curve
network.

Peak factor of each curve Waveform instantaneous value


pointed by the cursor.
Press or key to move t: Relative start time (ms).
cursor A1: Current instantaneous
value of L1.
A2:Current instantaneous value
of L2.
Current value axis line
(Automatic scale adjustment) A3:Current instantaneous value
of L3.
Figure 8-11: 3A CF display screen
Note: L1, L2, and L3 display the peak factor of voltage and current for phases 1, 2and 3 respectively.

8.5. Extreme & Average Values Of Voltage & Current


The Submenu can display the RMS value, maximum value, minimum value, positive peak maximum
value, negative peak minimum value of voltage and current.
8.5.1. 3U Max & Min value display screen
This screen displays the RMS value, maximum value, minimum value, positive peak maximum value, negative
peak minimum value of line voltage.
The information are shown as below:
The various values of the column
correspond to phases 1, 2 and 3 Instantaneous frequency of
respectively. network.
RMS: Line voltage true RMS Value
MAX: Line voltage RMS maximum
value (The maximum value
accumulated since power on or since
the last press or key)
MIN: Line voltage RMS maximum
value (same as above).
PK+: Line voltage positive peak
maximum value (same as above).
PK-: Line voltage negative peak
maximum value (same as above).
Figure 8-12: 3U Max & Min value display screen
Note: The Analyzer calculates the maximum and minimum values every half cycle (e.g. 10ms at 50Hz) and
refreshes the measurement results every 300ms.
8.5.2. 4V Max & Min value display screen
This screen displays the RMS value, maximum value, minimum value, positive peak maximum value, negative
peak minimum value of phase voltage and neutral line voltage.
The information are shown as below:

The various values of the column Instantaneous frequency of


correspond to phases 1, 2 and 3 network.
respectively.
RMS: Phase voltage true RMS Value
MAX: Phase voltage RMS maximum The Column various values of
value (The maximum value neutral line: RMS, PK+, PK-.
accumulated since power on or since
the last press or key)
MIN: Phase voltage RMS maximum
value (same as above).
PK+: Phase voltage positive peak
maximum value (same as above).
PK-: Phase voltage negative peak
maximum value (same as above).
Figure 8-13: 4V Max&Min. value display screen
Note: The Analyzer calculates the maximum and minimum values every half cycle (e.g. 10ms at 50Hz) and
refreshes the measurement results every 300ms.

36
8.5.3. 4A Max & Min value display screen
This screen displays the RMS value, maximum value, minimum value, positive peak maximum value, negative
peak minimum value of current and neutral line current.
The information are shown as below:
The various values of the column Instantaneous frequency of
correspond to phases 1, 2 and 3 network
respectively.
RMS: Current true RMS Value
MAX: Current RMS maximum The Column various values of
value (The maximum value neutral line: RMS, PK+, PK-.
accumulated since power on or
since the last press or key)
MIN: Current RMS maximum value
(same as above).
PK+: Current positive peak
maximum value (same as above).
PK-: Current negative peak
maximum value (same as above).
Figure 8-14: 4A Max & Min value display screen

Note: The Analyzer calculates the maximum and minimum values every half cycle (e.g. 10ms at 50Hz) and
refreshes the measurement results every 300ms.

8.5.4. L1 Max & Min value display screen


This screen displays the RMS value, maximum value, minimum value, positive peak maximum value, negative
peak minimum value of phase voltage and current.
The information are shown as below:

Voltage values column. Current value column: Same as


RMS: Phase voltage true RMS Value voltage value for each parameter
MAX: Phase voltage RMS maximum information
value (The maximum value
accumulated since power on or since
the last press or key)
MIN: Phase voltage RMS maximum
value (same as above).
PK+: Phase voltage positive peak
maximum value (same as above).
PK-: Phase voltage negative peak
maximum value (same as above).

Figure 8-15: L1 Max & Min. value display screen


Note: The Analyzer calculates the maximum and minimum values every half cycle (e.g. 10ms at 50Hz) and
refreshes the measurement results every 300ms.
L2, L3, and N respectively display the earth voltage and current RMS value, maximum value, minimum value,
positive peak value, and negative peak value of phase 2, phase 3, neutral line, and the relevant information
display is same as L1.

8.6. Various Test Value Displayed Simultaneously


The submenu displays all of the voltage and current measurement values (RMS, DC, THD, DF, CF, PST, KF).
8.6.1. Simultaneous display of various values in 3U
This screen displays the RMS, DC, THD, DF, and CF values of the line voltages.
The information are shown as below:
The various values of the column Instantaneous frequency of
correspond to phases 1, 2 and 3 network
respectively.
RMS: True RMS value of 1s
DC: DC component.
THD: Total harmonic distortion.
DF: Distortion factor.
CF: peak factor of 1s

Figure 8-16: Simultaneous display of various values in 3U

37
8.6.2. Simultaneous display of various values in 4V
This screen displays the RMS, DC, THD, DF, CF and PST values of the phase voltages and neutral line
voltage.
The information are shown as below:
The various values of the column Instantaneous frequency of
correspond to phases 1, 2 and 3 network.
respectively.
RMS: True RMS value of 1s
The Column various values
DC: DC component.
THD: Total harmonic distortion. of neutral line: RMS, DC
DF: Distortion factor.
CF: peak factor of 1s
PST: short-time voltage flicker (10
minutes).

Figure 8-17: Simultaneous display of various values in 4V

8.6.3. Simultaneous display of various values in 4A


This screen displays the RMS, DC, THD, DF, CF, and KF values of the phase current and neutral line currents.
The information are shown as below:

The various values of the column Instantaneous frequency of


correspond to phases 1, 2 and 3 network.
respectively.
RMS: True RMS value of 1s Neutral values: RMS.
DC: DC component.
THD: Total harmonic distortion.
DF: Distortion factor.
CF: peak factor of 1s
KF: transformer K factor

Figure 8-18: Simultaneous display of various values in 4A

8.6.4. Simultaneous display of various values in L1


This screen displays the RMS, DC, THD, DF, CF of phase 1 voltage, and DC, PST parameters of voltage, KF
value of current.
The information are shown as below:
The various values of the Instantaneous frequency of
column correspond to phases network.
1, 2 and 3 respectively.
RMS: True RMS value of 1s Current values.
DC: DC component. RMS: true RMS value for 1s
THD: Total harmonic THD: total harmonic distortion.
distortion. DF: distortion factor.
DF: Distortion factor. CF: peak factor of 1s
CF: peak factor of 1s KF: transformer K factor.
PST: short-time voltage flicker
(10 minutes).

Figure 8-19: L1 Simultaneous display of various values in L1


Note: L2 and L3 display the voltage and current various values of phase 2 and phase 3 respectively, and the
relevant information display is same as L1.

8.6.5. Simultaneous display of various values of neutral line


This screen displays the voltage RMS, DC value and the current RMS value of neutral line.

8.7. Phasor Diagram Display


The submenu shows the absolute value, phase and unbalance of the voltage or current at the basic
frequency.

38
8.7.1. Display of phasor diagram in 3V
This screen displays the absolute value of the phase voltage, the phase angle of the phase voltage relative to
the current and the unbalance of the phase voltage at the basic frequency.
The information are shown as below:
Phasor of phase 1,2,3. Instantaneous frequency of
V1,V2,V3: Phase voltage network
absolute value at fundamental
frequency.
ĭ: Phase angle of phase 1 to
phase 2. Channel number and color
ĭ: Phase angle of phase 2 to
phase 3.
ĭ: Phase angle of phase 3 to
phase 1.
Phasor diagram.
Vunb: Phase voltage
unbalance.

Figure 8-20: Display of phasor diagram in 3V

8.7.2. Display of phasor diagram in 3U


This screen displays the absolute value, the phase angle between the line voltages, and the unbalance of the
line voltage at the base frequency.
The information displayed on this screen is same as the information described in Section 8.7.1, and the only
difference is that the displayed values are related to the line voltage of 3U.
8.7.3. Display of phasor diagram in 3A
This screen displays the absolute value, the phase angle between each phase current and the unbalance of
the current at the base frequency.
The information displayed on this screen is same as the information described in Section 8.7.1, the only
difference is that the displayed values are related to the line voltage of current.
8.7.4. Display of phasor diagram in L1
This screen displays the absolute value of the voltage and current of phase 1, and the phase angle of the
phase 1 voltage relative to the current at the fundamental frequency.
The information are shown as below:
V1: Absolute value of phase Instantaneous frequency of
voltage at fundamental network
frequency.
A1: Absolute value of current at Channel number and color
fundamental frequency

ij9$: Phase angle of phase


voltage to current

Figure 8-21: phasor diagram display screen of L1

Note: L2 and L3 respectively display the absolute value of the voltage and current of phase 2 and phase 3 and
the phase angle of the corresponding phase voltage to current at the fundamental frequency, and the display of
relevant information is same as L1.

9. ALARM MODE
In the alarm mode, the Analyzer can monitor whether each measured value exceeds the threshold and
generate alarm log. The values that can be monitored are: Vrms, Urms, Arms, PST, Vcf, Ucf, Acf, Vunb, Aunb,
Hz, KF, Vthd, Uthd, Vdf, W, VAR, VA, DPF, PF, Vh, Uh, Ah, Vah, etc.
The user should set an alarm threshold (hysteresis) first and then can start the alarm monitoring .The
monitored values:
䖩Defined in the Set/Alarm Mode screen (Section 5.10).
䖩Select the setting parameter group (solid red dot means selected; hollow red circle means unselected).
The Analyzer can record up to 12800 alarm logs: the user can upload the stored alarm records to the PC and
save them for later retrieval and viewing.

39
9.1. Submenus
All submenus of alarm mode are listed on the screen as shown in the figure below, and the following chapters
will introduce respectively.
The user can press each yellow function key on the keyboard below the screen to select the corresponding
submenu.

Alarm log list (refer § 9.4).

Enter alarm mode setting (refer


§ 9.2).
Set the alarm detection schedule
(refer § 9.3).

Figure 9-1: Alarm mode screen

OK and symbols have the following functions:


䖩OK: Confirm alarm detection settings and start alarm detection (refer § 9.3.2).
䖩 : Actively stop alarm detection (refer § 9.3.3).

9.2. Alarm Mode Setting


The submenu displays the list of alarm settings (refer 5.10). Press this key to define or modify the alarm
settings.
The information are shown as below:
Present mode: setting mode Alarm hysteresis (select
values: 1%,2%,5% or 10%).
Selected (red solid dot) or cancel
Minimum duration: refers to the
the alarm
time when the monitored
Alarm parameters: (VAh, Ah, Uh,
parameter value exceeds the
Vh, Tan, PF, Cos, VA, VAR, W,
threshold to trigger the alarm (set
Athd, Uthd, Vthd, KF, Hz, Aunb,
from 0s to 99 minutes)
Vunb, Acf, Ucf, Vcf, PST, Arms,
Urms, Vrms). Alarm trigger threshold.
Harmonic order (0th~50th, for Vah, Alarm direction (>or< the
Ah, Uh and Vh. threshold to trigger)
Alarm channel (3L: 3 phases are
monitored separately, any of the 3 Page icon, press the
phases may be triggered; N: corresponding yellow function
monitor the neutral line; : key to turn page
monitor the total power of VA,
VAR, W; : monitor the average
value of Tan, PF, Cos)
Figure 9-2: Alarm mode setting

Reminder: Use the keys to view various values.


Set the alarm step as follows:
䖩 Press to select field value. And arrow appears.
䖩 Press or to input value (Vah, Ah, Uh, etc.), and press key to confirm.
Set other values as same operation.
䖩 Move yellow cursor to the first column, then press key to select the set alarm. Red solid dot indicates
selected. When the condition is met, the setting alarm can be triggered, and the alarm log is generated.
Note: Repeating step 3 can cancel the selected alarm.
䖩Press key to return to the alarm detection schedule screen.

9.3. Alarm Detection Schedule


The submenu sets the start time and end time of the alarm schedule.

40
Figure 9-3: Alarm detection schedule screen
9.3.1. Step1: Set the alarm schedule
Detailed operations as follows:
„ Press or key to select the start time (highlighted in yellow), and press to start setting time. The
corresponding time and date area will display the arrows .
Press or to increase or decrease each value, then press or key to select the request modify
value.
Note: The alarm start time must be later than present time.
„ Press to confirm the alarm start time setting.
„ Press or key to select the stop time setting (highlighted display). Press to set the stop time. The
corresponding time and date area will display arrows .
Press or to increase or decrease each value, then press or key to select the request modify value.
Note: The alarm start time must be after present time.
„ Press to confirm the alarm stop time setting.
9.3.2. Step2: Start alarm detection
Press the OK icon function key and the Analyzer starts alarm detection with the set start and stop time.
„ The OK icon disappears and the icon appears at same place.
„ When waiting to start detection, the Analyzer screen displays "detection standby..." and the icon
flashes on top of screen.
„ When the alarm detection starts, “detecting....” will be displayed.
„ When alarm detection is completed, “detection schedule” will be displayed again, and the OK icon is
displayed. User can set the new alarm detection.
9.3.3. Voluntary stoppage of alarm campaign
Before the stop time of alarm is reached, the user can press the icon yellow function key (lower right of the
screen) to actively stop the alarm detection, and the OK icon will be displayed again at the same position after
stopping.

9.4. View Alarm Log


The submenu displays the alarm log. The Analyzer can store up to 12,800 alarm logs. Press the
icon yellow function key to view the alarm log.
Note: In mode, the selected wiring type will not influence each phase selection and parameters
monitoring of alarm, users decide the selection.
Each data information shown as follow:
Alarm log memory usage. The Trigger threshold and trigger
black bar indicate memory direction > or <.
used.
Trigger amplitude value (minimum
or maximum)
Alarm date and time
Press or key select to
The phase of alarm trigger display the alarm log of trigger
phase
The monitored parameters
Alarm duration
Figure 9-4: Alarm log screen

Reminder: The default units of the monitored parameters and corresponding trigger thresholds and trigger
amplitudes in the alarm log are as follows:
Table 9-1
Monitored parameters Default units Units with the suffix Example
m: mean mA 2.5: mean 2.5A
Arms A
k: mean kA 423m: mean 423mA
41
326: mean 326V
Vrms/Urms V k: mean kV
1.2k: mean 1.2kV
W parameter: 315 mean
K: mean kW/kVar/kVA 315W
W/VAR/VA W/Var/VA
M: mean MW/MVar/MVA W parameter: 5.8k mean
5.8kW
Hz Hz 50.00: mean 50.00Hz

9.5. Delete Alarm Log


The submenu can delete all the alarm log. Detailed operation steps as follows:
Press yellow function key to select the submenu. The icon displayed in yellow indicates ready to
delete, then press the yellow function key to cancel the delete state. The icon displayed in gray indicates
not delete.
In the state of ready to delete, press key to confirm delete and all the alarm logs will be deleted and cleared.

Figure 9-5: Alarm log screen in delete mode

10. TREND MODE


In trend mode, the Analyzer can record the measured parameters selected in Set/Trend mode. (Refer § 5.9).

10.1. Select Submenus


The submenus are shown in the below screen, and described in the follow sections.
Users can use the yellow function keys at the bottom of the screen to select the corresponding submenu.

Trend record list (refer § 10.4).

Enter trend mode and select Trend monitoring schedule


record parameters (refer § setting (refer § 10.2).
10.3).

Figure 10-1 Trend mode screen

Press OK icon yellow function key to start trend record monitoring (refer § 10.2).

10.2. Presets & Starts Record Trend Charts


The submenu specifies the parameters of the new record detection schedule.

42
SD memory card usage. The black Select the record parameters
bar indicate memory used. configuration group (refer § 10.3).

Start time and stop time of record

Record period, interval time

Enter trend mode and select


record parameters (refer § 10.3) File name
Press OK yellow function key to
start record.

Figure 10-2: Preset record screen


10.2.1. Stage 1: Set the measurement parameters of requesting record
Operate as follows:
„ Press or keys to select the requested setting parameters area. The selected area is
highlighted in yellow. Press to enter the setting, and arrows appears.
„ Press or keys to select the parameters setting group form to , then press to confirm.
Reminder: Parameter setting groups from to are all set in the setting/trend mode (refer §5.9).
Users can also refer to §10.3 for the parameters selection and setting steps instruction of trend chart.
„ Press or keys to select the “Start Time”, and the selected field will be highlighted in yellow. Press
to enter and display arrows when setting the start date and time. Press or to increase or decrease
the value, press or to switch year, month, day, hour, minute etc.
Note: The start date and time must be later than the present date and time.
„ Press to confirm the “Start Time” of record.
„ Press or keys to select “Stop Time”, and the selected field will be highlighted in yellow. Press to
enter and display arrows when setting the stop date and time. Press or to increase or decrease the
value, press or to switch year, month, day, hour, minute etc.
Note: The stop date and time must be later than the start date and time. For the longest recording time, please
refer to § 10.6.4.
„ Press to confirm the “Stop Time” of record.
„ Press or keys to select “Record Period”, press key to enter and display arrows .
„ Press or to increase or decrease the values (optional value:1 s, 5 s, 20 s, 1 min, 2 min, 5 min, 10 min,
or 15 min).
„ Press to confirm the “Record Period” setting.
Note: The record interval period is the average interval cumulative time for measuring each recorded value
(that is, the interval of recording data once).
„ Press or to select the “File Name” and the file name are will be highlighted in yellow, press to enter
the edit mode to enter the file name (The length should not exceed 8 characters. Different names should be
used for the records, and there should be no spaces in the middle of the file name).
The available characters are the letters from A to Z and the digits from 0 to 9. Use the or keys to
display a character and or key move to the next character.
„ Press key to confirm the “File Name”.
10.2.2. Stage 2: Start recording
„ Press OK icon yellow function key (the bottom right corner of the screen), then the Analyzer starts
recording with the start and stop times set by the user.
The OK icon disappears and the icon appears instead in this time.
„ The Analyzer displays “Recording stand by...” before reaching the start time and the icon continues
flashing in the screen’s top display bar.
The Analyzer will display “Recording...” after reaching the start time.
Recording

Press icon yellow function


key to actively stop.

Figure 10-3: A trend chart monitoring display screen


43
„ The “Recording Schedule” screen and OK icon will appear again (bottom right corner of the screen)
after reaching the stop time, and the recording will be stopped automatically. The OK icon reappears at
corresponding position.
10.2.3. Active stop present record
Before reaching the set stop time, press icon yellow function key (the bottom right corner of the screen), then
the recording will be stopped automatically, and the OK icon will reappear.

10.3. Trend Chart Monitoring Parameter Setting


This submenu displays the selection list of trend chart monitoring parameters (refer to §5.9), and the user can
set and change the monitoring parameter selection by using this shortcut key.
The information are shown as below:

Monitoring parameters list, press 4 different parameter setting


key move to pages
request parameter, press key
to select or cancel. Red solid dot
indicates selected, red hollow dot
indicates unselected. Select all the parameters
VAh, Ah, Vh and Uh etc.
harmonics record parameters
selection. “?” indicate not select
any harmonic Cancel select all the parameters.

Figure 10-4: Trend chart monitoring parameter setting screen


Set a record operation as follows:
Example for parameter setting group 1:
„ Press icon yellow function key, with the display highlighted in yellow.
„ Press or and or keys to select parameters, then press to confirm, and display with red solid dot
Reminder: User can select the following parameters:
Unit Designation

Urms RMS phase-to-phase voltage (2 , 3 ).

Uthd Total harmonic distortion of the phase-to-phase voltage (2 , 3 ).

Ucf Crest (peak) factor of phase-to-phase voltage (2 , 3 ).

Uunb Phase-to-Phase voltage unbalance (2 , 3 ).

Hz Frequency of power grid.

Vrms RMS phase-to-neutral voltage.

Vthd Total harmonic distortion of the phase-to-neutral voltage.

Vcf Crest factor of phase-to-neutral voltage.

Vunb Phase-to-neutral voltage unbalance (2 , 3 ).

PST Short-term flicker.

Arms RMS current.

Athd Total harmonic distortion of the current.

Acf Crest factor of current.

Aunb Current unbalance (2 , 3 ).

KF K factor.

W Active power.

VAR Reactive power.

VA Apparent power.

44
PF Power factor.

&RVij Phase shift of power factor.

7DQij Tangent.

? See comment below.


Specific function of the last two lines.
The position as follow:

Figure 10-5: These two lines involve harmonics

The selected starting harmonic


order The selected last harmonic order

VAh, Ah, Vh and Uh etc.


harmonics record parameters Select recording odd harmonics
only
selection. “?” indicate not select
any harmonic

Figure 10-6: Recording the setting of harmonic parameters in trend mode

These two lines involve the recording of 4 different harmonics, as Vah, Ah, Vh, and Uh. For each harmonic, the
user can select the desired recording harmonic order (0th-50th), or only record odd harmonics. The operation is
as follows:
„ Enter the desired record parameters: The cursor will be highlighted in yellow. Press and then the arrows
appear. Select the value (Vah, Ah, Vh, and Uh) for which harmonics are to be recorded by pressing
or . The red solid dot identifies your selection.
Confirm by pressing . The values field is highlighted in yellow.
„ To select the starting harmonic order: With the field highlighted in yellow, press and then the arrows
appear, select the order from which the harmonics are to be recorded by pressing or , then
validate by pressing .
Press or to go to the next field.
„ To select the last harmonic: With the second field (greater than or equal to the starting harmonic order)
highlighted in yellow, press . Select the highest harmonic order to be recorded by pressing or , then
validate by pressing .
Press or to go to the next field.
„ For the odd harmonic only:
To select or deselect this function, press . The red solid dot identifies your selection:
䖩 Selected: Only odd harmonics between the two orders of harmonics specified in the previous points are
recorded.
䖩 Not selected: All harmonics (even and odd) between the two orders of harmonics specified in the previous
points are recorded.

10.4. View Record List


The submenu displays recordings already made. Press the yellow key corresponding to the icon to
view the list.
The data shown as follow:

45
Record the proportion of storage
Record stop time
space, the black bar is the used
space

Record start
time
Record file name

Figure 10-7: Record list display screen

10.5. Delete Record


The submenu is used to delete the records. Operate as follows:
„ Press or key to select the record to be deleted. The selected record is bolded.
„ Press key to select the submenu. The icon displayed yellow indicates ready to delete, and
press the function key to cancel the delete state. The icon displayed gray indicates not delete.
„ In the state of ready to delete, press key to delete the selected record.

Figure 10-8: Record list screen in delete mode.

10.6. Display Records


10.6.1. Record information

Record parameters: name, start The monitoring parameters


time, stop time, record period record page and total stored
page

Press key to switch to the


Monitoring parameters of next monitoring parameter
selected records. Press the pages. Press and keys to
yellow key corresponding to select up or down page
the parameter icon to open
the corresponding curve.
Figure 10-9: Trend mode record list sub menu screen
10.6.2. Trend curves record
Date &time Position of the viewing window in
the record, black bar represents the
present screen.
Record group
Parameter name of trend curves.
Prompt left side have trend curve
Prompt left side have trend curve

Press or key to select


Cursor. Press or key to request curve
move
Zoom out, switch the time scale

Figure 10-10: Trend curve record screen

46
Remark: The cursor position “----” indicates errors or missing values in the record.
When the curve display period is 20 seconds. Since the period of the record is one second, each point of this
curve corresponds to a value recorded in a one-second window once 20 seconds. There is therefore a
substantial loss of information (19 values out of 20), but the display is rapid.
With the increase of the display period, the loss values will be more. This case the user can select to activate
the MIN-AVG-MAX mode. After the MIN-AVG-MAX mode is activated, each point of the curve represents the
mean of the total sampling points every period (such as the display period is 20 seconds, sampling period is 1
second, each display point of the curve represents the mean of 20 values recorded every second.).

Date & time Parameter name of trend


curves.

Arithmetic mean value at period Press or key to select


of time request curve

The MIN-AVG-MAX mode icon Zoom in, switch the time scale.

Figure 10-11: Vrms with MIN-AVG-MAX


Since the MIN-AVG-MAX mode is enabled, each point on the curve represents the arithmetic mean value
within the display period, so the display is more accurate and there is no data loss, but the display speed is
relatively slow (For display time, refer §10.6.3).
Date & time
Press or key to select
The recorded values of the cursor request curve
locate

The function icon display with grey


indicate the MIN-AVG-MAX mode is
not activated

Figure 10-12: Ucf (L2) without MIN-AVG-MAX


When the MIN-AVG-MAX mode is not activated, display the curve of 60 values recorded of the cursor, the
display is rapid. Date of cursor. Recorded values of the
cursor (minimum, mean, and
Curve of the maxima, gray maximum).

Curve of the mean, green

Curve of the minima, gray To select the display curve,


press the or key.
Yellow indicates MIN-AVG-MAX
mode is activated. Press the
corresponding yellow function
keys to deactivate.
Figure 10-13: Ucf (L2) with MIN-AVG-MAX

Date of cursor The sum of the active powers


of the three phases is
Power values of the cursor presented in bar graph form.

Gray column indicates the


position of the cursor. To select the display bar
graph, press the or
key.
The MIN-AVG-MAX mode has
not been activated.

Figure 10-14: total active power without MIN-AVG-MAX

47
Date of cursor The sum of the active powers
of the three phases is
Arithmetic mean of display presented in bar graph form.
period.

Gray column indicates the


position of the cursor.
To select the display curve,
press the or key.
The MIN-AVG-MAX mode
has been activated.

Figure 10-15: Total active power with MIN-AVG-MAX


This curve differs slightly from the previous one because, with the MIN-AVG-MAX mode, there is no loss of
information.
Energy calculation mode. Date of the cursor (ending
Starting date of the selection. date of the selection).
Gray column indicates the
time domain has been
calculated. Calculation results of total
active power.
Press yellow key
corresponding to the
icon to active the energy
calculation mode. To select the display bar bar
graph, press the or
The MIN-AVG-MAX mode has key.
not been activated.
Figure 10-16: Total active energy without MIN-AVG-MAX
Energy calculation steps:
„ Press yellow key corresponding to the icon to active the energy summation mode. The current cursor
time is the starting time of the energy calculation.
„ Press the or to move the cursor. Calculate the time of the cursor to which the stop time points. The
device calculates the energy between the start time and stop time automatically.
Note: The location of the starting time cannot be exceeded when the cursor moves leftwards.

10.6.3. The time of displaying the curve in the different scale


The following table indicates the time needed to display the curve on screen as a function of the width of the
display window for a recording period of one second:
Time Span Unit Time Waiting time
5 days 2 hours 30 seconds
2.5 days 1 hour 15 seconds
15 hours 15 minutes 4 seconds
10 hours 10 minutes 2 seconds
5 hours 5 minutes 1 second
1 hour 1 minutes 1 second
20 minutes 10 seconds 1 second
5 minutes 5 seconds 1 second
1 minute 1 second 1 second
„ Press the or key to change the scale of the display
„ Press the or key to move the cursor
„ Press the or key to change the display phase curve.
„ Above operation may restart the loading/calculation of the values from the beginning.

10.6.4. The longest recording time can be programmed


The longest time is based on the number of the recording parameters selection and the sampling period, the
typical condition as follows:
Typical longest time can be
Parameter of selected Sampling period
programmed

48
All parameters (total 123) 1 second 10 days

1~20 parameters 1 second 62 days

All parameters (total 123) 5 seconds 50 days

1~20 parameters 5 seconds 300 days

All parameters (total 123) 1 minutes 600 days

1~20 parameters 1 minutes 3600 days


The above table indicates that, the less the selected parameters, the greater the sampling period, and the
longer the longest recording time.

11. POWER AND ENERGY MODE


The key displays power and energy measurements.

11.1. Sub-menus
The sub-menus are listed in the screen below and described individually in the paragraphs that follow.
The sub-menus are selected using the yellow keys on the keypad below the screen.
The date and time at which The date and time at which
energy metering starts energy metering stops.
Select displayed corresponding
Display of energy generated power and energy, use or
.
™: sum of 3 phases.
: Arithmetic mean of 3 phases
Display of energy consumed (such as PF)
Start of energy metering.

Display of PF, DPF, and tan Reset of energy meter.


Figure 11-1: The Power and energy mode screen

11.2. Energy Consumed


The sub-menu displays active power, reactive powers (capacitive and inductive), and apparent power.
11.2.1. The energies consumed screen for the 3 phases (3L)
This screen displays the following information:

Figure 11-2: The energies consumed screen for the 3 phases (3L)

Unit Designation
W Active power.
Wh Active energy consumed.
VAR Reactive power (inductive or capacitive ).
VARh Reactive energies consumed (inductive or capacitive ).
VA Apparent power.
VAh Apparent energy consumed.

49
11.2.2. The energies consumed screen for phase L1
This screen displays the following information:

Figure 11-3: The energies consumed screen for phase L1

Unit Designation
W Active power.
Wh Active energy consumed.
VAR Reactive power (inductive or capacitive ).
VARh Reactive energies consumed (inductive or capacitive ).
VA Apparent power. ( : sum of 3 phases)
VAh Apparent energy consumed.
PF Power factor.
&RVij Phase shift of power factor
7DQij Tangent factor
ij9$ Phase shift of phase-to-neutral voltage with respect to current.
Note: Filters L2 and L3 display the same information for phases 2 and 3. ™ screen displays total power and
energy consumed values for the 3 phases.

11.3. Power Factor Display Screen


This screen page is available only with the 3L filter. To display the information, press the yellow key on the
keypad corresponding to the icon.
The following data is displayed:

Power factor

Phase shift of power factor

Tangent factor

Phase angle between voltage


and current

Figure 11-4: Power factor screen for the 3 phases (3L)

11.4. The Sums Of Energies Consumed Display Screen


To display the information, select the ™ icon of the right-hand filter. This screen displays the following
information:

50
Figure 11-5: The sums of energies consumed display screen

Unit Designation
W Total active power.
Wh Total active energy consumed.
VAR Total reactive power, inductive or capacitive .
VARh Total reactive energies consumed, inductive or capacitive .
VA Total apparent power.
VAh Total apparent energy consumed.

11.5. The Arithmetic Mean Values Of Power Factor Display Screen


To display the arithmetic mean values for the 3 phases (for power factor, phase shift of power factor and
tangent), select the icon of the right-hand button.
This screen displays the following information:
Mean values of the power factor on
each phase

Mean values of the phase shift of


the power factor on each phase

Mean values of the tangent


factor on each phase.

Figure 11-6: Arithmetic mean values screen for the 3 phases

11.6. Energy generate


The sub-menu displays the active power, the reactive powers (capacitive and inductive), the apparent
power, and all associated energies generated.

11.6.1. The energies generated screen for the 3 phases (3L)


This screen displays the following information:

Figure 11-7: Energies generated screen for the 3 phases (3L)

51
Unit Designation
W Active power.
Wh Active energies generated.
VAR Total reactive power, inductive or capacitive .
VARh Total reactive energies generated, inductive or capacitive .
VA Total apparent power.
VAh Total apparent energy generated.

11.6.2. The energies generated display for phase L1


This screen displays the following information:

Figure 11-8: The energies generated display for phase L1

Unit Designation
W Active power
Wh Active energies generated
VAR Total reactive power, inductive or capacitive
VARh Total reactive energies generated, inductive or capacitive
VA Total apparent power.
VAh Total apparent energy generated.
PF Power factor
&RVij Phase shift of power factor.
7DQij Tangent factor
ij9$ Phase angle between voltage and current
Note: Filters L2 and L3 display the same information for phases 2 & 3. The screen displays total power and
energy generated values for the 3 phases.

11.6.3. The sums of energies generated display screen


To display the information, select the icon.
This page displays:
䖩The total active power,
䖩The total active energy generated,
䖩The total reactive power, inductive or capacitive ,
䖩The total reactive energies generated (inductive and capacitive ),
䖩The total apparent power,
䖩The total apparent energy generated.

11.7. Starting Energy Metering


To start metering, press the yellow key on the keypad corresponding to the icon:

52
The date and time at which
energy metering starts
The icon appears after metering
starts. To stop energy metering,
press the yellow key on the keypad
corresponding to the icon.

Figure 11-9: Power and energies mode screen when energy metering is started

11.8. Stopping Energy Metering


To stop energy metering, press the yellow key on the keypad corresponding to the icon.
The date and time at which Display the date and time at
energy metering starts which metering stops after press
the icon.

After stop metering, if not


reset, press the yellow key The icon appear after stop
corresponding to the icon metering, press the yellow key
to continue metering. corresponding to the icon to
reset metering.

Figure 11-10: Power and energies mode screen when energy metering is stop
11.9. Reset Of Energy Metering
To reset metering, press the yellow key on the keypad corresponding to the icon, The icon
displayed yellow indicate ready to delete, then press the yellow key corresponding to the icon to cancel
the delete state.
In the state of ready to delete, press the key to confirm. All energy values (consumed and generated) are
reset. Note: refer to the 4-quadrant power diagram in § 17.3.

12. SCREEN SHOT MODE


The key can be used to:
䖩Capture a m aximum of 60 screens for future reference (see §12.1).
䖩Display previously saved screen snapshots (see § 12.2).
Saved screens may then be transferred to a PC using the USB.
12.1. Screen Snapshots
Press for approx. 3 seconds to take any screenshot (including the , , , , , and
mode screens).
During the capture, the icon appears in the top left corner of the screen instead of the icon for the active
mode ( , , , , , , ). The active mode icon reappears when you release the key:
the device has saved the image.
Reminder: The device can save a maximum of 60 screen snapshots. If the user attempts to take a 61st screen
snapshot, first, upload the pictures you need to the computer through USB, then delete the device’s snapshots
before capturing new pictures.

Figure 12-1: The snapshot list display screen


53
12.2. Handling Of Screen Snapshots
This handling concerns stored screen snapshots, i.e.:
ƾDisplay of the list of screen snapshots (see § 12.2.2).
ƾViewing of one of the screen snapshots (see § 12.2.3).
ƾDeletion of one or more of the screen snapshots (see § 12.2.4).
12.2.1. Available functions
To enter screen snapshot mode, briefly press the key.
Reminder: Holding the key down for approximately 3 seconds triggers the screen snapshot function (See
§ 12.1).
Indicator of available image Screen page, the current
memory. The black bar page / total number of pages.
represents memory used.

List of saved snapshots:


The icons identify the type
of snapshot saved (such
as :harmonic mode).
The date and time of each Delete the snapshots of cursor
screen snapshot.

Figure 12-2: Example of the snapshot list display screen

12.2.2. Viewing the list of snapshots


Press briefly to display this list. The screen presents the list of snapshots (see figure 12-2).
12.2.3. Viewing a snapshot from the list
To view a snapshot, proceed as follows:
䖩 Press . The icon is active and the snapshot list screen is displayed (see figure 12-2).
䖩 Select the snapshot to be viewed using the or and or keys. The date and time of the
selected snapshot are bolded.
䖩 Press to display the selected snapshot. The top left corner of the screen displays the icon,
alternating with the icon corresponding to the active mode (such as :harmonic mode).
䖩 Press to return to the list of screen snapshots.
12.2.4. Deleting a snapshot from the list
To delete a snapshot, proceed as follows:
From the list of snapshots (see Figure 12-2 for example).
䖩 Select the snapshot to be deleted using the or and or keys. The date and time of the
selected snapshot are bolded.
䖩 Press the yellow key on the keypad corresponding to the icon, the icon displayed yellow
indicates ready to delete, then press the yellow key corresponding to the icon to cancel the delete state.
䖩 In the state of ready to delete, press the key to confirm.
Press to delete the selected snapshot. The snapshot is deleted from the list.

13. HELP
The key provides information about the functions and symbols used in the current display mode.
The following information is displayed:

The current mode and help List of help information


mode icon

Help page 2

Help page 1

Figure 13-1: Example of the help page for the powers and energies mode, page 1

54
14. DATA UPLOADED TO THE COMPUTER
Download the data uploading software of corresponding model from official website. (https://www.uni-trend.com)
Then connect the device to the PC using the USB cord supplied with the device, start the device, then open the
data software to click on the computer’s Power Quality Analyzer.exe. Wait for software to automatically search
and connect the device. For directions for using the data export software, refer to its user manual.
Note: The transfer does not delete the data, just copy to the PC. In alarm/trend chart record/transient capture
mode (pending or ongoing), PC cannot read the data.

15. GENERAL SPECIFICATIONS


15.1. Housing
Housing Rigid shell overmoulded with a red thermo-adhesive elastomer.
5 voltage measurement sockets.
4 special current connectors.

Connectors One connector for the specific mains power unit.


One connector for the USB link.
One connector for the SD memory card. This connector is located in the battery compartment on the back
of the device.
Keys Function, navigation, and mode. Can be used with gloves on.
Hand strap Located on the side of the device, use to operation more convenient.
Stand To hold the device in an inclined position.
Battery cover To access the battery on the back of the Analyzer.
Overall: 240×170×68mm
Screen: 640×480 pixels
Dimensions
W×H: 118mm×90mm;
Diagonal: 148mm
Weight Host: 1600 g (with battery).

15.2. Power Supply


15.2.1. External mains power supply
Range for use Input AC100V-240V, 50Hz/60Hz, Output DC12V, Maximum output current 3A.
Maximum input power 36VA
15.2.2. Battery supply
The device can be used without a connection to mains power. The battery also makes it possible to use the
device during power outages.

Battery Rechargeable lithium-ion battery pack 9.6V


Capacity 4500 mAh
Life At least 500 charge-discharge cycles.
Charging current approx. 0.6A
Charging time approx. 8 hours
Service T° [0 °C; 50 °C].
Charging T° [10 °C; 35 °C].
Storage 30 days:[-20 °C ; 50 °C]
Storage T° Storage for 30 to 90 days: [-20 °C; 40 °C].
Storage for 90 days to 1 year: [-20 °C; 30 °C].

15.2.3. Consumption
At 10% luminosity 410 mA

55
At 50% luminosity 490 mA
At 100% luminosity 590 mA

15.3. Range for use


15.3.1. Environmental conditions
15.3.1.1. Climatic conditions
The following table shows the conditions regarding ambient temperature and humidity:
Temperature Range Humidity Range
Reference environmental conditions 20đ~26đ 45%RH~75%RH
Application condition 0đ~50đ 10%RH~85%RH
Store condition˄With battery˅ -20đ~50đ 0%RH~90%RH
Store condition˄Without battery˅ -20đ~75đ 0%RH~90%RH
Caution: At temperatures above 40°C, the device must be powered by the battery alone OR by the mains
power unit alone; use of the device with both the battery AND the mains power unit is prohibited.

15.3.1.2. Altitude
Use: [0 m; 2 000 m]
Storage: [0 m; 10 000 m]

15.3.2. Mechanical conditions


Under IEC 61010-1, the device is regarded as a PORTABLE DEVICE (HAND-CARRIED).
„ Operating position: any position.
„ Reference position in operation: on a horizontal plane, resting on its stand or lying flat.
„ Rigidity (IEC 61010-1): force of 30 N applied to any part of the housing, the device being supported (at
40°C).
„ Fall (IEC 61010-1): 1 m in presumed worst-case position; the requirement is no permanent mechanical
damage and no functional degradation.
„ Tightness: IP 50 as per NF EN 60529 A1 (IP2X electrical protection for the terminals).

15.3.3. EMC electromagnetic compatibility


15.3.3.1. Immunity as per IEC 61326:1-2006
„ Immunity to electrostatic discharges (as per IEC 61000-4-2)
Level 1: Severity: 4 kV in contact
standard: Standard A
Level 2: Severity: 8 kV in air
standard: Standard A
„ Immunity to radiated fields (as per IEC 61000-4-3 and IEC 61000-4-8)
Severity: 10V.m-1
Standard: standard B
„ Immunity to rapid transients (IEC 61000-4-4)
Severity: 2 kV on voltage inputs and power supply
1 kV on current input
Standard: standard A
„ Immunity to electric shocks (as per IEC 61000-4-5)
Severity: 2 kV on voltage inputs in differential mode
1 kV on voltage inputs in common mode
Standard: standard A
„ Conducted RF interference (as per IEC 61000-4-6)
Severity: 3 V on voltage inputs and power supply
Standard: standard A
„ Voltage interruption (as per IEC 61000-4-11)
Severity: 100% loss over one period of the power supply
Standard: standard A
15.3.3.2. Emissions as per IEC 61326:1-2006
„ Class A equipment.

15.4. User safety


„ Application of safety rules as per IEC standard 61010-1 (protective impedances on voltage inputs).
„ Pollution type 2.

56
„ Double insulation on I/O with respect to earth ( symbol).
„ Double insulation between the voltage inputs and power supply and the other I/O ( symbol).
„ Indoor use.

16. FUNCTIONAL CHARACTERISTICS


16.1. Reference conditions
This table indicates the reference conditions of the quantities to be used by default in the characteristics.
Ambient temperature (23±2)°C
Humidity (relative humidity 40%~ 60%
Atmospheric pressure [860hPa~ 1060hPa]
Phase-to-neutral voltage [(50±1%) Vrms; (500±1%) Vrms] without DC (< 0.5 %)
Frequency of electrical network 50Hz±0.1Hz, 60Hz±0.1Hz
Phase shift 0° (active power ), 90° (reactive power)
Harmonics <0.1%
Voltage unbalance <10%
Voltage ratio 1
Current ratio 1
Power supply Battery only
Electric field <1V/m
Magnetic field <40A/m

16.2. Electrical characteristics


16.2.1. Voltage input characteristics
0 Vrms to 1000 Vrms AC+DC phase-to-neutral and neutral-to-earth.
0 Vrms to 2000 Vrms AC+DC phase-to-phase. (On condition of compliance with 1000 Vrms with respect to
earth in CAT III).

16.2.2. Current input range


008B current clamp: 10mA~ 10A.
040B current clamp: 0.10A~ 100A.
068B current clamp: 1.0A~ 1000A.
Optional current transformer: device input current 1mA~ 500mA.

16.2.3. Characteristics of the device alone (excluding the current sensor)


Respectively introduce the following data (on the basic of base conditions and the ideal current sensors,
perfectly linear, no phase shift).
The maximum error in the range of
Measurement Range Display resolution
the reference
Frequency 40Hz~ 70Hz 0.01Hz ±(0.03)Hz
True RMS phase-to-neutral
1.0V~ 1000V Min resolution 0.1V ±(0.5%+5dgt)
voltage
True RMS phase-to phase
1.0V~ 2000V Min resolution 0.1V ±(0.5%+5dgt)
voltage
DC voltage 1.0V~ 1000V Min resolution 0.1V ±(1.0%+5dgt)

True RMS current 10mA~ 1000A Min resolution 1mA ±(0.5%+5dgt)


Peak of phase-to-neutral
1.0V~ 1414V Min resolution 0.1V ±(1.0%+5dgt)
voltage
Peak of phase-to-phase
1.0V~ 2828V Min resolution 0.1V ±(1.0%+5dgt)
voltage
Current peak 10mA~ 1414A Min resolution 1mA ±(1.0%+5dgt)

1.00~ 3.99 0.01 ±(1%+2dgt)


Peak factor
4.00~ 9.99 0.01 ±(5%+2dgt)

57
±(1%+3dgt)
Cos 0.8
Active power 0.000W~ 9999.9kW Min resolution 0.001W
±(1.5%+10dgt)
0.2 Cos <0.8
±(1%+3dgt)
Reactive power, inductive or 0.000VAR~ Sin 0.5
Min resolution 0.001VAR
capacitive 9999.9kVAR ±(1.5%+10dgt)
0.2 Sin <0.5
0.000VA~
Apparent power Min resolution 0.001VA ±(1%+3dgt)
9999.9kVA
±(1.5%+3dgt)
Cos 0.5
Power factor -1.000~ 1.000 0.001
±(1.5%+10dgt)
0.2 Cos <0.5
±(1%+3dgt)
0.000Wh~ Cos 0.8
Active energy Min resolution 0.001Wh
9999.9MWh ±(1.5%+10dgt)
0.2 Cos <0.8
±(1%+3dgt)
Reactive energy, inductive or 0.000VARh~ Min resolution Sin 0.5
capacitive 9999.9MVARh 0.001VARh ±(1.5%+10dgt)
0.2 Sin <0.5
0.000VAh~
Apparent energy Min resolution 0.001VAh ±(1%+3dgt)
9999.9MVAh
Phase angle -179°~ 180° 1° ±(2°)
7DQij
-32.76~ 32.76 Min resolution 0.001 :±(1°)
9$•9$
Phase shift of power factor
-1.000~ 1.000 0.001 :±(1°)
(DPF)
Harmonic ratio
0.0%~ 99.9% 0.1% ±(1%+5dgt)
(order 1 to 50) (Vrms>50V)
±(3°) harmonics of order 1 to 25
Harmonic angle
-179°~ 180° 1°
(Vrms>50V)
±(10°) harmonics of order 26 to 50
Total harmonic ratio
0.0%~ 99.9% 0.1% ±(1%+5dgt)
(THD or THD-) ”
Distortion factor
0.0%~ 99.9% 0.1% ±(1%+10dgt)
(DF or THD-5 ”
Transformer K factor 1.00~ 99.99 0.01 ± (5%)

3 phases unbalance 0.0%~ 100% 0.1% ± (1%)

16.2.4. Current sensor characteristics (after linearization)


Sensor errors are offset by a typical correction inside the device. This typical correction, applied to the phase
and amplitude, depends on the type of sensor connected (detected automatically) and the gain in the current
acquisition channel used.
Type of current sensor True RMS current Max error of true RMS current Max error of phase angle ij

10mA~ 99mA ±(1%+3dgt) ±(1.5°),Arms 20mA


008B current clamp
100mA~ 10.0A ±(1%+3dgt) ±(1°)

0.10A~ 0.99A ±(1%+3dgt) ±(1.5°)


040B current clamp
1.00A~ 100A ±(1%+3dgt) ±(1°)

1.0A~ 9.9A ±(2%+3dgt) ±(3°)


068B current clamp
10.0A~ 1000A ±(2%+3dgt) ±(2°)

10A̚99A ±(1%+3dgt) ±(3°)


300F flexible current sensor
100A̚6000A ±(1%+3dgt) ±(2°)

58
17. APPENDICES
17.1. Mathematic Formulae
17.1.1. Network frequency and sampling
Sampling is controlled by (locked to) the network frequency so as to deliver 256 samples per cycle from 40 Hz
to 70 Hz. This locking is essential for the calculations of reactive power, unbalance, and harmonic ratio and
angles.
The frequency is measured by analyzing ten consecutive positive-going zero crossings in the first voltage
channel (V1) or first current channel (I1) after digital low-pass filtering and digital suppression of the DC
component.
The time of the zero crossing is determined precisely by linear interpolation between two samples to achieve
resolution better than 0.002%.
The signals are acquired using a 16-bit converter and (for current acquisition) dynamic gain switches.

17.2. Hysteresis
Hysteresis is a screening principle that is often used after detection of a threshold stage in Alarm mode (See §
5.10) and in Inrush current mode (see § 6.3). A correct hysteresis setting avoids repeated changes of state
when the measurement oscillates about the threshold.

17.2.1. Surge detection


With a hysteresis of 2%, for example, the return level for surge detection is equal to (100% - 2%) or 98% of the
reference voltage threshold.

Figure 17-1: An example of return level for surge detection

17.2.2. Undervoltage or blackout detection


With a hysteresis of 2%, for example, the return level for undervoltage detection is equal to (100% + 2%) or
102% of the reference voltage threshold.

Figure 17-2: an example of return level for undervoltage detection

17.3. Four-quadrant Diagram


This diagram is used for power and energy measurements .

Figure 17-3: Four-quadrant diagram for power and energy


59
17.4. Mechanism For Triggering Transient Captures
The sampling rate is a constant 256 samples per cycle. When a transient capture is started, each sample is
compared to the sample from the preceding cycle. The preceding cycle defines the mid-point of the trigger
envelope and is used as reference. As soon as a sample is outside the envelope, the triggering event occurs;
the representation of the transient is then captured by the device. The cycle preceding the event and the three
following cycles are saved to memory.

Figure 17-4: A graphic representation of the transient capture triggering mechanism

17.5. Capture Conditions In Inrush Current Mode


Reminder: The capture depends on a triggering (start) event and a stop event. If a capture ends with a stop
event or if the recording memory of the device is full, the capture stops automatically.
The capture stop threshold is calculated as follows:
[Stop threshold [A]] = [Start threshold [A]] × (100 – [stop hysteresis [%]]) ÷ 100
Here are the conditions for triggering and stopping captures:
Triggering channel Triggering and stop conditions
Triggering condition 䋽 [A1 half-cycle RMS value] > [Triggering threshold]
A1
Stop condition 䋽 [A1 half-cycle RMS value] < [Stop threshold]
Triggering condition 䋽 [A2 half-cycle RMS value] > [Triggering threshold]
A2
Stop condition 䋽 [A2 half-cycle RMS value] < [Stop threshold]
Triggering condition 䋽 [A3 half-cycle RMS value] > [Triggering threshold]
A3
Stop condition 䋽 [A3 half-cycle RMS value] < [Stop threshold]
Triggering condition 䋽 [[the half-cycle RMS value of one current channel]> [Triggering threshold]
3A
Stop condition 䋽 [the half-cycle RMS values of all current channels] < [Stop threshold]

18. MAINTENANCE
18.1. Important Recommendation
For maintenance, use only the spare parts specified. The manufacturer cannot be held liable for any accident
that occurs following a repair not performed by its customer service department or by an approved repairer.

18.2. Recharging The Battery


The battery charge is managed by the device when connected to the AC network via the mains power unit
supplied.
䖩 For safety reasons and to ensure the correct operation of the charger, the storage battery must be replaced
with the power off.
䖩 Do not throw the battery into a fire.
䖩 Do not expose the battery to a temperature in excess of 75°C.
䖩 Do not short-circuit the terminals of the battery.
䖩 W hen the battery is fully recharged, please remove the power adapter (if external power supply is not used).

18.3. Replacing The Battery


For safety reasons, advice that replace the battery only with the original model
To replace the battery, proceed as follows:
䖩 To eliminate all risk of electric shock, disconnect the power supply cord and connected devices.
䖩 Turn the device over.
䖩 Use a cross screwdriver to unscrew the two quarter-turn screws on the back of the housing. Then open the
battery cover.
䖩 Gently rem ove the old battery, replace a new original battery (To avoid damaging the battery connection, do
60
not pull on the wires).
䖩 Put the battery compartm ent cover back in place and screw the 2 quarter-turn screws back in.
䖩 R eboot the device to confirm .
Note: If the battery is disconnected, it must then be fully recharged, even if it is not replaced, so that the device
will know the battery charge condition (this information is lost when the battery is disconnected).

61
Data Analysis Software
User Manual
CONTENTS

1. SUMMARY ....................................................................................................................................................... 1
2. INSTALLATION ................................................................................................................................................ 1
2.1.Software Installation ................................................................................................................................... 1
2.1.1.Directly open ............................................................................................................................................ 1
2.1.2.Software installation ................................................................................................................................ 1
2.2.Drive installation ......................................................................................................................................... 1
2.2.1.Directly installation................................................................................................................................... 1
2.2.2.inf file installation ..................................................................................................................................... 1
3. SOFTWARE START-UP................................................................................................................................... 1
3.1.Start-up steps ............................................................................................................................................. 1
4. REAL-TIME DATA ............................................................................................................................................ 1
4.1. Waveform mode ........................................................................................................................................ 2
4.2ˊHarmonic Mode........................................................................................................................................ 2
4.3ˊPower Energy Mode ................................................................................................................................ 2
5. MONITER RECORD ........................................................................................................................................ 3
5.1. Trend Recording ........................................................................................................................................ 3
5.2. Alarm Recordings ...................................................................................................................................... 4
5.3. Transient capture recording....................................................................................................................... 5
5.4. Inrush Current Capture Recording ............................................................................................................ 5
5.5. Snapshots .................................................................................................................................................. 7
6. IMPORT DATA FILE ......................................................................................................................................... 8
7. ANALYZER CALIBRATION TIME .................................................................................................................... 8
1. SUMMARY
The software, Power Analyzer Transfer Data View, is based on Windows system and matches with the device.
This Data analysis software is used for viewing waveforms, harmonics, power energies, and can be exported
to the type of real-time test data file. Real-time test data file is saved to your computer, so when you need to
re-import into the analysis software to view; the device can be used to save the trend recording, alarm
recording, transient recording, the inrush current recording, snapshots uploaded to the computer and display,
and can be stored in the form of long-term data files to retain power for when you need to re-import into the
analysis software to view.
Note: When an alarm campaign is initiated or a search for transients, an inrush current capture, or a trend
recording is pending or in progress, the device will not respond to the data analysis software. You can upload
the data after the capture completed. When the software connects with the device successfully, the device’s
mode should be through the data software to switch instead of pressing the keys of device directly.
The functions as follows:
„ Monitoring test data and waveforms, monitor waveforms, harmonics and power energies in real time.
„ Upload and view historical monitoring data.
„ To zoom in, zoom out, move left and right, sorting data and waveforms.
„ Export data files (*.pqaf).
„ Export excel data report.
„ Save pictures on the current screen.
„ Import the previously saved data files (*.pqaf).

2. INSTALLATION
2.1. Software Installation
Download the “Power Quality Analyzer.exe” from the official website of Uni-Trend.
Then install it according to the software instructions.
If you fail to install the software, please contact the local seller or distributor.

3. SOFTWARE START-UP
3.1.Start-up steps
Connect the device to the computer using the USB cord, then switch the device on, click the desktop shortcut:
Power Quality Analyser.exe. to start-up the software. After the software start-up, it will detect automatically the
computer and connect with the computer. In general, it needs about 5 seconds to upload all the test data, and
then enters the Waveform mode by default.
The following information is displayed:

Test mode

Analyzer status Display waveform,


recording. (According
to different mode

Time of computer
Connect status
Figure 3-1: the software normal start-up and connect display screen
As shown in Figure 3-1, the test mode have two parts: real-time test and monitor record. The real-time mode
can be selected waveform mode, harmonic mode and power energy mode. The monitor record can be
selected trend recordings, alarms, transient recordings, inrush currents, snapshots. Yellow highlight indicates
the current mode.

4. REAL-TIME DATA
Real-time data is used to display the device's test data and waveforms in real time, including waveform,
harmonic, and power energy mode. The software reads the real data from the device every 1 second, and

1
refreshes the display. The user can click on the icon to suspend update to observe the waveform, click on
the icon to continue refreshing the waveforms.

4.1. Waveform mode


Click the Real-time Test/Wave to enter into the Waveform mode, and to display the test waveforms and data.
The following information is displayed:
Export and save buttons.

The display value of each phase.

Move the cursor.

Display the sub-menu.

Cursor (the black vertical line).

Select the curve to be displayed.

Real-time waveform.

Instantaneous value of the cursor.

Figure 4-1: Display the waveform in Real-time Test Mode


„ Export the data files: export the waveform data, the current curve waveform and the current status of the
device information, and saved as the extension .pqaf. You can re-import when needed, and the display as the
real-time monitoring, cannot be switched only in the import mode. (See § 6).
„ Save to excel: export the waveform data, the current curve waveform and the current status of the
device information, and saved in the form of excel list.
„ Save to bmp: saved the current display screen in the form of bmp.
„ Display sub-menu: select to display RMS, THD, CF, maximum, minimum, parameter list and phasor
diagram.
„ Select the curve to be displayed. Click the buttons (3U, 4V, 4A, L1, L2, L3, N) to select the displayed curve.
Note: in these three modes, the test mode of device will vary with the test mode of software to switch, that is,
the software and the device are in the same mode.

4.2ˊHarmonic Mode
Click the Real-time Test/Harmonic to enter into the Harmonic mode, display of the harmonic bar charts and the
relevant data.
The following information is displayed:

Export and save


buttons
Display by
Display the sub-menu

Harmonic bar chart

Select the bar chart to be


displayed
Harmonic order

Figure 4-2: display the harmonic in Real-time Test


„ Display list: check: display of the list of harmonic data each phase in the form of table;
Uncheck: display of the list of harmonic data each phase in the form of bar chart.
„ Display sum-menu: to select display of the harmonic bar chart of V, A, VA, U.
„ Select the bar chart to be displayed: click the buttons (3L, L1, L2, L3, -,+ expert mode) to switch the
diagrams to be displayed.

4.3ˊPower Energy Mode


Click the Real-time Test/Power Energy to enter into the Power Energy mode, display of the power and the

2
metering energy.
The following information is displayed:
„ Start calculate: press the Start calculate button, the energy starts to calculate, and displays the start time in
the lower left corner of the window.
„ Stop calculate: after starting calculation, press the Stop calculate button to stop calculate the energy, and
displays the stop time in the lower right of the window.
„ Reset of energy metering: after stop calculation, press the Clear button to reset the energy values in the
tables.
Export and save buttons

Energy calculate buttons

Display the sub-menu

Figure 4-3: Displaying power energy in real-time test

5. MONITER RECORD
Monitor record including trend recordings, alarms, transient recordings, inrush currents, snapshots. It is used
to upload the data which is saved in the device to the computer for view and export.

5.1. Trend Recording


Click the Monitor Record/Trend to upload the recording and the number of the records, and displays in tables.
The following information is displayed:
Export and save buttons.

The recording list.

Figure 5-1: list of trend recording


„ Read and save all the trend charts: readout all the trend recording of the device and save to the default
address: C:\PowerQuality\trend.

In the list of trend recording, the field where the mouse is bolded, click the left mouse button, upload and
display the selected trend recordings.
The following information is displayed:

3
Export and save buttons.

Parameter list of the trend


recording.

Figure 5-2: Parameter list of the trend recording


The parameter list if the parameters that are selected and recorded in the trend mode. All parameters
displayed in the figure. Black buttons indicate the selected parameters; gray buttons indicate the unselected
parameters. Click the corresponding button to enter the trend curve waveform.
The following information is displayed:
Return to the trend recording list.

Export and save buttons.

Location of the zone displayed in


the record.

Value of the cursor.

Current parameter name.

Date and time of the cursor.

Select the curve to be displayed.

Trend curve chart.

The activated MAX-AVG-MIN mode.

The button of start metering energy


(when the selected parameter is W,
VAR, VA).
Figure 5-3: Curve of the trend recording
5.2. Alarm Recordings
Click the Monitor Record/Alarm to upload the alarm log and the number of the records, and displays in table.
The following information is displayed:
Export and save buttons.

The list of alarm log.

Press the trigger alarm channel to


display the alarm log by
classification.

Turn the page

Figure 5-4: Alarm list screen

4
5.3. Transient capture recording
Click the Monitor Record/Transient to upload the recorded transients and the number of the records, and
displays in table.
The following information is displayed:

Read all the records and save


to the default address.

The recording list.

Figure 5-5: Instantaneous recordings

„ Read and save all the transient recordings: readout all the transient recordings of the device and save to
the default address: C:\PowerQuality\ tran.
In the list of transient recording, the field where the mouse is bolded, click the left mouse button, upload and
display the selected transient recordings. The following information is displayed:

Return to the transient recording


list.

Location of the zone displayed in


the record.

Date and time of the cursor.

Select the curve to be displayed.

Transient waveform.

Instantaneous value of the cursor.

Figure 5-6 transient waveform display

5.4. Inrush Current Capture Recording


Click the Monitor Record/Inrush to upload the inrush current recording, and displays in table.
The following information is displayed:

5
The inrush current recording list.

Figure 5-7: Inrush current recordings

In the inrush current recording list, click the Inrush Waveform button to upload the inrush current recordings,
when the upload is complete, the inrush current waveform will be displayed.

The following information is displayed:


Return to the inrush current
recording list.

The maximum RMS value (over


the entire capture).

Switch RMS and PEAK.

Location of the zone displayed in


the record.

Select the curve to be displayed

RMS curve in the half-cycle.

Figure 5-8: RMS of inrush current

Return to the inrush current


recording list.

The maximum PEAK value (over


the entire capture).

Switch RMS and PEAK.

Location of the zone displayed in


the record.

Select the curve to be


displayed.

PEAK envelope.

Figure 5-9: PEAK of inrush current

When the PEAK envelope of the inrush current is displayed, click the In button to display the recorded
waveform curves of the inrush current.

6
The following information is displayed:
Return to the inrush current
recording list.

The maximum RMS value (over


the entire capture).

Switch RMS and PEAK.

Location of the zone displayed in


the record.

Select the curve to be displayed

Waveform.

Figure 5-10: Waveform of inrush current

5.5. Snapshots
Click the Monitor Record/Snapshot to upload the snapshots and the number of the snapshots, and displays in
table.
The following information is displayed:

Read all the records and save to


the default address.

The recording list.

Figure 5-11: Snapshots table

„ Read and save all the snapshots: readout all the snapshots of the device and save to the default address:
C:\PowerQuality\ photo.
In the list of snapshot, the field where the mouse is bolded, click the left mouse button, upload and display the
selected snapshots.
The following information is displayed:

7
Figure 5-12: uploaded snapshots displayed screen

6. IMPORT DATA FILE


In the screen of Real-time Test and Monitor Record, to export the file which extension is .pqaf, press the
Export button. The file saved the current real-time data or monitor records. The data can be re-imported to the
software to display in need. Proceed as follows:
Disconnect the USB cord or turn the device off, then close the data analysis software, then click the Power
Quality Analyser.exe to restart the software.
The following information is displayed:

Figure 6-1: open the software without connecting to the USB

Select Import saved data file, press Confirm to enter the Import data file mode. In this status, it displays neither
real-time data waveform nor device's status.
Import method: click the menu bar: File\Import, select the corresponding file which extension is .pqaf, then the
real-time data and monitor records are displayed as §4 and §5. In this status, the display screen accords to the
import data mode, and the mode cannot be switched.

7. Instrument CALIBRATION TIME


In the normal, the device and computer connect successfully with the USB, if there is deviation between the
device time and computer time, you can calibrate the device time to be consistent with the computer time. Click
the menu: File\ calibrate Instrument time. Displayed as the following figure (the red block):

8
Figure 7-1: Analyzer calibration time

The company is not responsible for other losses caused by use.


The contents of this user manual cannot be used as a reason to use the product for special purposes.
The company reserves the right to modify the contents of the user manual.
The contents in the User Manual are subject to change without further notice.

You might also like