BTS310 Service
BTS310 Service
MS810IA110604
1
CONTENTS
1. INTRODUCTION 1
I
2.5.11. Display circuit 23
2.5.12. RS-232-C channel circuit 23
2.5.13. Fan control 23
2.5.14. Microprocessor 23
2.5.15. Power supply 23
2.5.16. Reset circuit 23
2.6. Description of the electronic circuit 23
2.6.1. Logarithmic amplifier 23
2.6.2. Analogic-digital converter 24
2.6.3. Lamp control 24
2.6.4. Filter wheel detector 24
2.6.5. Filter wheel motor control 25
2.6.6. Temperature sensor amplifier 25
2.6.7. Peltier Cell control 26
2.6.8. Pump motor control 26
2.6.9. Keyboard circuit 26
2.6.10. Printer control 27
2.6.11. Display circuit 27
2.6.12. RS-232-C channel circuit 28
2.6.13. Fan control 28
2.6.14. Microprocessor 28
2.6.15. Power supply 29
2.6.16. Reset circuit, 29
II
3.5.1. Materials needed 42
3.5.2. Access and options 42
3.5.3. Option CHECK 42
3.5.3.1. Checking method 43
3.5.4. Option MANUAL 43
3.5.5. Option AUTOMATIC 45
3.6. Check of the transformer and supply circuit 47
3.7. Check of the fan 47
3.8. Verification of the optical system 47
3.9. Check of the sensitivity with flow cuvette 48
4. CHECK TESTS 51
5. MAINTENANCE 57
III
5.12. Change of the photodiode 67
5.13. Change of the fan 68
5.14. Change of the temperature sensor 68
5.15. Change of the lamp 70
5.16. Change of the filters 70
5.17. Change of the lenses 71
5.18. Change of the program 73
5.19. General care of the photometer 73
5.20. Cleaning of the optical components 76
5.21. Cleaning of the filters 76
5.22. Cleaning of the lenses 77
5.23. Cleaning of the photodiode 77
5.24. Cleaning of the aspiration system 77
5.25. Cleaning of the flow-cuvette 78
5.26. General cleaning of the instrument 78
IV
APPENDIX V. MAINTENANCE PLAN 91
V.1. Clean 91
V.2. Change 91
V.3. Review 91
V.4. Check 91
VI.1. Accessories 93
VI.2. Authorized Spare Parts 93
SCHEMES
MODIFICATIONS
1 Modification
2 Modification
3 Modification
V
1. INTRODUCTION
Almost all the electronics of the instrument is located in a single board, allowing
the change of the printed circuit in case of failure. These circuits will be sent to
the factory where they will be repaired. A lot of check programs are
incorporated, to simplify the search and diagnosis of troubles.
This manual is conceived not only as a guide for maintenance and repair of the
instrument but also as a document for the training of the Technical Assistance
Service personnel. The functioning principles as well as the electronic circuit
are explained in order to get a global view of the photometer.
1
2. DESCRIPTION OF THE INSTRUMENT SYSTEMS
The most complete versions of the instrument are described. The components
not included in the simplified versions are indicated when absent.
Figure 2.1. shows a schema of the optical system, that is composed of the
following:
The first component is a light source, the halogen lamp (1). The diaphragm (2)
delimits a light solid angle that is then collected by the first lens (3) that sends a
beam of parallel rays to the interference filter (4). This is one of the nine
3
possible filters located in the wheel; a stepping motor actions this wheel and,
under the control of the program, positions the filter in the optical axis in front of
the light beam. Each filter monochromes one wavelength with the
characteristics described in section I.1.
A second lens (5) focuses the monochromatic beam in the center of the cuvette
(6). In the case of the flow cuvette the light path hole itself acts as a diaphragm
thus reducing the sensitivity, that will be a little lower than that of the common
cuvettes. The light passing across the cuvette is received by the detector (7),
that converts it in a electric current that will be processed and quantified by the
electronic circuits.
The optical system (figure 2.2) is composed by a holder (1) made of injected
aluminum, on which the following components are mounted:
- An aluminum block (2) containing the lamp-holder (3) with a radiator shape in
order to evacuate the heat of the lamp. Said block includes the diaphragm
(figure 2.3 (1)) that delimits the solid angle of light.
- A light protector (12), fitted together the second lens holder, avoids entrance
of parasite light.
- A cuvette-holder (13) holds both the cuvette and the photodiode detector. The
thermostatization system is fitted together (see section 2.3.).
The light received by the photodiode (1) (figure 2.4) generates a proportional
electric current (If). This current enters the LOG-100 logarithmic amplifier (2) by
the input I1. A reference current Ir (3) (nominal 100 nA) is introduced by input I2.
The output tension of LOG-100 is:
4
5
6
7
8
If
V0 = k x log -----------
Ir
2.1.5. Calibration
The global system (optical system, amplifier and converter) is not strictly linear
due to the tolerances of their components; it is thus necessary to calibrate it,
that is, to introduce correction factors to compensate the deviations inherent to
the system.
2.2.1. Components
The sample is aspirated in by the sip tubing (1). This tube, made in Teflon, has
a defined length and the instrument is calibrated in accordance with it. The
aspirated sample enters into the cuvette (2) where readings take place.
9
Aspiration is performed through a silicone tubing by means of a peristaltic pump
(3), formed by a four-roller rotor actioned by a stepping motor with a resolution
of 1/100 revolution. The sample finally passes through the pump to the waste
bottle (4).
The aspiration system (figures 2.6 and 2.7) is placed on the cuvette-holder tray.
The sipping tube (1) passes through the case by means of a steel guiding tube
(2), to which it is fixed by means of a silicone connector (3). The sipping tube is
fixed to the cuvette (4) by means of a inlet adapter (5). The peristaltic tube (6) is
connected to the cuvette by the outlet adapter (7), located in the peristaltic
pump (8) and finally connected to the waste outlet (9).
The waste bottle (1) (figure 2.8) is connected to the outlet (2) in the rear of the
instrument by means of a silicone tubing (3).
Three parameters control the sipper functioning and they must be programmed
to obtain the expected performance characteristics. Their programming is
explained in the user`s manual.
b) PUMP DELAY. It is the number of seconds the pump will wait from the
moment the sipping finishes up to the moment the pump activates again to
position the sample.
2.2.5. Programming
Programming the aspiration system means indicating the instrument the volume
in microliters to be sipped. This value may introduced in several points in the
program. For instance, one of the parameters of each stored test is the sample
volume to aspirate.
2.2.6. Calibration
The nominal flow of the pump is 110 2L/revolution; nevertheless, the accuracy
of this figure depends on the tolerances in length and diameter of the sipping
tube, that may be affected by the aging of the tube itself; this is why this value
should be calibrated from time to time. The calibration process is explained in
10
section 3.5. As it can be also done by the user, it is also explained in the user's
manual.
The general rules for maintenance of the sipping system are the same as those
indicated to the user and appear in section 5.24.
2.3.1. Components
The cuvette holder is isolated from the optical system holder (7) and contacts
one of the faces of the Peltier Cell (6). The other face of the cell is in contact
with the optical system holder.
The Peltier Cell pumps heat from one face to the other depending on the sense
of circulation of the current. The power control circuit (5) is in charge of making
that current circulate in the adequate sense inside the Cell, in order to heat or
cool depending on the instructions of the microprocessor.
When heating, heat is pumped from the environment (taken from the optical
system holder) to the cuvette holder and when cooling the opposite is done.
The optical system holder has a radiator to evacuate the heat coming from the
11
12
F ig 2 .8
13
14
cuvette holder. The temperature sensor (2) measures the temperature of the
cuvette holder that is read by the microprocessor (4) through the amplifier (3).
The microprocessor has the thermostatization program and, depending on the
programmed temperature and the read value, it activates the power control (5)
heating or cooling as required.
The cuvette holder (13) (figure 2.2) is fixed to the optical system holder (1) by
means of four screws (14) thermally isolated. The Peltier Cell (15) is located
between the cuvette holder (13) and the optical system holder (1). The
temperature sensor, located in a plastic holder (16) is fixed to the cuvette
holder by a thread.
2.3.4. Programming
2.3.5. Calibration
The same as with other systems, the tolerances of the different components
produce a deviation between the programmed and the real temperatures, thus
making it necessary a calibration process to compensate such deviations. The
thermostatization calibration process is described in section 3.3.
The thermostatization system has two critical points: the good thermal contact
between both faces of the Peltier Cell with the optical system holder and the
cuvette holder, and the good thermal contact between the temperature sensor
and the cuvette holder.
a) A thin, uniform layer of silicone must be placed in each face of the Peltier
Cell, covering the whole surface.
b) The silicone layer should neither go beyond the surface of the Cell, nor go
inside between its two faces; it would result in a thermal crossing, thus reducing
the performance of the system.
15
c) The temperature sensor should also have a silicone layer and make a good
thermal contact with the bottom of its lodging.
The connector is located in the rear of the instrument, inside a box marked
COM 1. The control lines and the electric level of the communication channel
are compatible with the E.I.A. RS-232 standard. Figure 2.10 describes the
signals of this connector.
16
The chain includes two characters (CS1 and CS2) for checking that the
transmission has been correctly received (check sum). The checking process is
as follows: CS1 and CS2 are first transformed to hexadecimal numbers and a
control byte is formed with both. Then, the sum of the hexadecimal values of
characters 2 to 11 plus the control byte is calculated. The low-order byte of the
calculated sum must be 00 (hexadecimal).
Example:
CS1 = 32h
CS2 = 35h
2.4.4. Programming
17
18
a) When communications are activated or at the moment of connecting the
photometer if communications are already activated, the photometer transmits
a XON (11 hexa).
c) If the photometer receives a XOFF (13 hexa), it stops the transmission, but it
can transmit two characters yet.
e) When the photometer is working in a job that does not allow it to receive
information, the photometer transmits a XOFF (13 hexa) to the computer, but it
can receive two more characters yet.
b) If the computer recognizes the message, it sends the ACK control character
(06 hexa).
d) The photometer waits for the time programmed in TIME-OUT for the
reception of those messages.
e) Once the time-out has elapsed without receiving any message, the following
appears on the display:
COMMUNICATIONS
ERROR
Press ENTER
1. DISCONNECT
2. CONTINUE
19
g) By selecting DISCONNECT, the photometer ends the communications and
deactivates them. Communications can be re-started using the path:
OTHER OPTIONS/COMMUNICATIONS
i) If the photometer receives a NAK control character it will repeat the last
transmission.
j) If a NAK is received a second time, the display will show the transmission
error flag as in step e). In this case, follow the steps f) to h).
If the protocol ACK/NAK has not been programmed, the photometer transmits
without waiting for confirmation from the computer.
The block diagram (figure 2.12) is intended for a general functional overview of
the various parts of the electronic circuit.
It converts the electric current coming from the photodiode in a tension equal to
its logarithm (section 2.1.4.).
It digitalizes the tension coming from the logarithmic amplifier for its further
treatment by the microprocessor.
It supplies the lamp two different tensions. The first one (12 V) for its normal
use. The other (6 V), for the standby.
20
Although the hardware is implemented to perform this function, it is not
operative in the normal use and only in the lamp standby test (test 200, section
4.1.7.)
It is a barrier photodetector that detects the position of the filter wheel. Allows
the microprocessor to know the position of each filter thus being able to position
them in front of the light beam.
This is a circuit that, with the logic control of the microprocessor, applies the
needed energy to the stepping motor that moves the filter wheel.
This is a circuit that makes the tension level of the temperature sensor
adequate to be measured by the 10-bits converter of the microprocessor, with a
resolution enough to adjust the system.
This is a power circuit that, controlled by the microprocessor, gives the Cell the
adequate current to heat or cool the cell-holder.
This is a circuit that, based in the logic control of the microprocessor, applies
the needed energy to the stepping motor that moves the pump.
This circuit is basically formed by the keyboard itself and some protection
elements.
It is formed by the power circuits that allow the orders coming from the
microprocessor to act on the printer, either moving the motors or heating the
points of the thermal head.
21
22
2.5.11. Display circuit
This circuit is basically formed by the display itself, with the controller and the
power circuits incorporated.
It is formed by the circuits needed to transform the TTL tension levels to those
of the RS-232 rule. The ACIA is integrated in the microprocessor itself.
This is a circuit that checks the temperature of the radiator of the power supply
and changes the speed of the fan according with the heat evacuation needs. A
more silent operation of the photometer under normal conditions is thus
achieved.
2.5.14. Microprocessor
By means of its program, it is in charge of linking and controlling all the systems
of the instrument. The only autonomous circuit, not controlled by the
microprocessor, is that of the fan control.
This section describes the electronic schema of the instrument, following the
same functional structure as the block diagram (section 2.5.).
23
The logarithmic amplifier is formed by the hybrid circuit LOG-100 (U17). Its
supply tension is H 15 V. C48, C49, C50 and C51 uncouple the supply. The
photodiode is connected between ground and the input I1 (1/U17). The
reference current (section 5.1.4.) is generated in the 2.5 V reference power
supply TL431CLP (U18) and the T-circuit formed by R50, R51 and R52. Its
nominal value is 100 nA. The LOG-100 has the output OUT (7/U17) connected
to the pin K1 (3/U17), thus being the logarithmic conversion constant K=1. The
capacitor C45 stabilizes the frequency circuit.
The output tension of the logarithmic amplifier is applied to the input VIN(+)
(11/U16) of the analogic-digital converter TSC500A CPE (U16), through the RC
net formed by R49 and C44. This converter takes as a reference the U18 2.5 V
tension and is supplied at H5 V. The tensions are reached from H15 V with the
D24 and D23 zeners and are uncoupled with the capacitors C40, C41, C42
and C43. Capacitors C37, C38, C39 and the resistance R46 take part of the
converter system. Said capacitors should be polypropylene-made in order to
have very low leaks.
As the tension coming from 7/U17 can vary in the range H15 V while the input
tension of 11/U16 must not exceed the supply tension U16, the circuit formed
by D28, D29, D32, D33, R59 and R60 is in charge of keeping the input between
these values.
Power supply of the lamp is done through the circuit formed by the RG1
regulator. It receives the non-regulated tension from the 0-17 Vac coiling of the
transformer, rectified by the bridge formed by D10, D11, D12 and D13, filtered
by C19 and C20 and uncoupled in high frequency by C21. The 12 V output
tension is given by resistances R34, R35 and R36. The MOSFET P322 (T4)
transistor, controlled by the microprocessor, short-circuits the resistance R36,
thus making the output tension fall to about 6 V (standby tension).
24
to the photodetector collector as its load. From this point the logic signal is
taken and read by the microprocessor pin P71 (84/U1).
Under normal conditions, the infra-red light of the photodiode reaches the
phototransistor, that conduces at saturation and the signal is its collector is a
logic "0". When the filter wheel turns and the detection stem interrupts the light
from the photodiode, the transistor stops conducing and the signal in its
collector is a logic "1".
The control of the stepping motor actioning the filter wheel is achieved by two
integrated circuits PBL3717A (U10 and U11). These circuits are constant
current stepper motor drivers, each one controlling the current of one of the
motor coils. This current is given by the reference current entering by their
respective REF pins (11/U10 and 11/U11), that of the 12 V zener D5, the
resistances R23 (in U10) and R27 (in U11) and the internal tension dividers of
these circuits. C33 and C34 uncouple the +5 V supply of these circuits and the
other connected components complete its operation.
The U8 amplifier and its associated circuit make these tensions adequate to be
read with the maximum resolution by the 10-bit converter of the Hitachi H8/510
microprocessor. The adequate tension exits by 6/U8 and is applied to the input
AN0 (83/U1) of the microprocessor through the resistance R59.
As the tension coming from 6/U8 may vary in the range H15 V and the input
tension of 83/U1 must not exceed the reference tension of the converter
applied to AVCC (87/U1) (2.5 V), the circuit formed by D34, D35, D27, D30 and
R64 is in charge of keeping this input between this value and ground.
25
(See schema E81001, sheets nº 1 and 2)
The full-wave rectification in both senses is achieved using two triacs and a
center-tapped coiling. Each triac may let a positive or negative half-wave pass,
when convenient. The circuit formed by the amplifier TL072CP (U15), that
works as a comparator, and the transistor BC548 (T2), that makes the output
1/U15 (H15 V) adequate to logic level (0 at +5 V), form a polarity detector of the
alternating tension supplying the power control circuit.
This signal is applied to the microprocessor by the pin P73 (86/U1) in order for
it to know which triac and when should be activated. The microprocessor
activates the triacs by means of two signals leaving by pins P84 (93/U1) and
P61 (74/U1) and that, together with the polarity detection signal (collector T2)
are applied to the programmable logic circuit GAL16V8 (U21) to generate the
two control signals of the triacs (12/U21 and 17/U21) and avoid a simultaneous
activation caused by a program error. Circuits 8/U6, 12/U6 and transistors T8
and T3 form the activation step of the triacs. Should it occur, two fuses F in the
coil common protect the transformer.
The result of this procedure is that each triac supplies the Peltier Cell a half-
wave, in the same sense, depending on the heating or cooling requirements.
This circuit, formed by the integrated circuits U12 and U13, is similar to that
used for the filter wheel control, with the only difference of the values of
resistances R30 and R33, as the current of this motor is different. Thus refer to
section 2.6.5. for its description.
The membrane-type keyboard forms a contact matrix and detects the pressed
key by scan (schema E81005). The keyboard is connected to the printed circuit
I81004 by J17. The lines of this connector pass directly to connector J11
(schema E81003) that carries them to the printed circuit I81002.
The scan is generated by the microprocessor lines P80 to P83 (89/U1 to 92/U1,
respectively). The standstill of these lines is a logic "1". One of them and only
one is active with a logic "0". Lines P40 to P42 (56/U1 to 58/U1) and P45 to
P47 (61/U1 to 63/U1) read the status of the keyboard. If there is no key
26
pressed, the status is a logic "1" due to the polarization resistances R13 to R18.
Instead, if a key is pressed, it will contact any of the scan lines and when it is at
zero the corresponding reading line will be the same. This is how the
microprocessor detects the pressed keys.
Diodes D1 and D4 prevent the short circuit between a scan line at "0" and
another at "1" that could happen when pressing two keys of the same reading
line simultaneously.
The schema of the printer appears in E81003 (box marked "PRINTER"). Four
sets of uni-polar coilings belong to the two stepping motors that control it. One
moves the head and the other the paper feed.
Eight thermal points form the printer head and a microswitch detects the initial
position (HOME) of the head. This microswitch is polarized by resistance R54
and the signal is read by the microprocessor pin P72 (85/U1).
Both the motors and the thermal points are controlled by parallel lines exiting
the microprocessor (schema E81001). The motors are controlled by lines P30
to P37 (47/U1 to 54/U1 respectively) and the points by P10 to P17 (4/U1 to
11/U1 respectively). These logic signals are converted into power signals by
means of two circuits 2803A (U19 and U20). Capacitors C57, C62, C58 and
C59 uncouple the power supply of the printer.
The LCD display is connected to the microprocessor bus. To the display (see
box marked "DISPLAY" in the schema) arrive the data bus, the LCD control line
that is selected by the display in the microprocessor access cycles, as well as
/HWR to write data and A0 to select internal registers.
A tension given by resistances R55 and R56 is given by the line VEE, given the
adequate vision angle (that cannot be varied).
27
lines, TxD2 (transmitter data, 96/U1) and RxD2 (receiver data 95/U1). Flux
control lines are not implemented and so it can only be achieved by means of
the XON-XOFF handshake (section 2.4.5.).
The logic level of these lines is TTL. To adapt them to the RS-232-C E.I.A. rule,
the integrated circuit MAX232CPE (U14) is used; with its associated capacitors
it generates the needed positive and negative tensions. The communication
lines enter and exit by the J6 connector (Figure 2.7).
The supply of the fan is done by an independent power supply, formed by the 0
to 24 V transformer coiling, the rectifier bridge formed by D6, D7, D8 and D9
and capacitors C18 and C56 that generate the non-regulated tension. The
voltage to apply to the fan is controlled by RG7 regulator. The output tension is
given by the divider formed by R37 and NTC1, that contacts the power supply
radiator and makes the tension increase when it becomes hot. In this way, the
higher is the load or the external temperature, the faster is the operation of the
fan, thus achieving a more silent function under normal conditions.
2.6.14. Microprocessor
This microprocessor does not incorporate internal memory, but has got lines to
connect it externally. This memory if formed by a EPROM (U4) that contains the
program. The one installed is 2 Mbytes (27C2001) although a 4 Mbytes one
can be installed in a future for further enlargements of the program. The printed
circuit is also prepared to accept FLASH memories (able to be recorded under
program control). The permanent data such as test parameters or calibration
coefficients are located in the E2PROM 2816 (U2). A 43256, 32 Kbytes
memory is installed as working RAM.
Attention: The maximum access time of the memories should be 200 nsec.
28
The power supply is in charge of supplying the various tensions to the circuit.
The lamp supply is described in section 2.6.3. and the fan supply in section
2.6.13.
Besides this power supplies, there is that of H15 V, formed by the regulators
RG2 and RG3 with their associated circuits and that of +5 V. This power supply
departs from a 0 to 12 V transformer coiling. The alternative tension is rectified
by the bridge D19, D20, D21 and D22, filtered by C28 and C29 and uncoupled
in high-frequency by C30. The regulated tension obtained in this way is applied
to three regulators in order to better uncouple the impedances. RG4 generates
the +5 V to supply the thermal head of the printer, RG5 supplies its motors and
RG6 generates the +5 V for the logic circuits.
The reset circuit that initializes the microprocessor is formed by the integrated
circuit TL7705ACP (U7). This circuit monitors the supply voltage in such a way
that when it is below a certain threshold the signals RESET and /RESET are
activated. The signal /RESET re-initializes the microprocessor while the signal
RESET blocks the line /OE of the E2PROM, by means of the transistor T1, thus
being a "hardware" protection against sporadic writings during tension
increases or decreases.
29
30
3. CALIBRATIONS AND VERIFICATIONS
With the only exception of the peristaltic pump calibration, that can be also
performed by the user, access to the tests and calibrations procedures is done
through the menu OTHER OPTIONS/SERVICE.
PASSWORD........ =
Introduce the access code (appendix VII). While being input, asterisks appear
in the display. In case of mistake, type "C" to repeat.
1 PHOTOMETER
2 PELTIER
3 FILTERS WHEEL
31
Test 302: Reading Stability
Test 303: Precision
Test 304: Carry-over
Test 305: Accuracy
Test 900: QC Procedure Activation
Test 901: Restore Default Personalization
OPEN: The user can modify the contents of all the locations.
In order to let the user modify only some of the locations, while the others
remain closed, the option OPEN/CLOSED can be selected in the option
PROGRAMMING TESTS of the menu SERVICE, thus allowing to close
individual locations. This option does not appear in the test programming by the
user.
If the instrument has been closed to the user not allowing him to modify all of
any of the locations, the Technical Assistance Service can do it using this
option, that is the same as that of the user's procedure, including the option to
open or close an individual location.
When the instrument is wholly closed, the user cannot modify the units (the
option UNITS does not appear in the PROGRAMMING MENU). In only some of
the locations are closed, although he can access to the unit table, he cannot
modify those assigned to a closed location. The Technical Assistance Service
can do it by this option.
This option allows loading of parameters into the memory from a personal
computer. A simple application program to perform this option is not available
for the first series of the instrument and thus this option is reserved for internal
use in the factory.
3.1.7. Option SAVING TESTS
32
This option allows saving the parameters of the tests in a personal computer. A
simple application program to perform this option is not available for the first
series of the instrument and thus this option is reserved for internal use in the
factory.
- Calibrated neutral filters kit (supplied with the service tools kit).
3.2.2. Procedure
a) Switch the instrument on and wait for at least 20 minutes before beginning
calibration readings.
b) Select the following path in the menu tree: OTHER OPTIONS / SERVICE
/password/CALIBRATION/PHOTOMETER/CALIBRATE.
STABILIZATION TIME = 0
Press 3.
CALIBRATION
33
Reading Filter ... 1 = 340
CALIBRATION
Reading Filter ... 2 = 405
CALIBRATION
Reading Filter ... 3 = 505
Input the indicated wavelengths or press ENTER if they are already introduced.
d) Then the instrument asks for the data of the reference standards. First the
serial number of the kit (printed in the box) and then the values assigned for
each filter and standard. Thus, the display shows:
KIT NUMBER
Actual KIT ............. = 0
New KIT .................. = _
CALIBRATOR 1 340
0.000 : 4.000
Actual value = 0.249
New value =_
Input all the values of the calibrators for each filter. (In case of mistake, go to
the erroneous value by means of the cursor and correct).
The last value corresponds to calibrator number 4 and 505 nm. The display
shows:
CALIBRATOR 4 505
0.000 : 4.000
Actual value = 1.988
New value =_
Once all the values are input, the calibration process begins.
e) Insert the required calibrator in the cuvette holder, to perform the calibration
reading.
34
g) Finally, the instrument asks through the display if the performed calibration is
to be stored.
STORE CALIBRATION ?
Store only if the obtained results are inside the tolerance margins.
This procedure describes the way to calibrate the thermostating system. Both
the cuvette-holder temperature sensor and its associated electronic circuit have
some tolerances that must be compensated in order to have the cuvette
temperature adjusted with the required accuracy. Calibration allows, using a
thermometer, the calculation of the correction coefficients needed to perform
this compensation.
- Thermometer Fluke 51
- Calibrated thermal sensor (ref.: FLUKE 80pk-1)
OTHER OPTIONS/SERVICE/password/CALIBRATE/PELTIER
1 LIST
2 CHECK
3 CALIBRATE
A list is obtained with the parameters of the last calibration (see example at the
end of the procedure). If no calibration has been performed before, the
following appears:
PELTIER CALIBRATION
------------------------------------
NOT CALIBRATED
The calibration data appear in the display in the case of the versions of the
instrument without printer. First at 25qC (for instance):
TEMP. READ = 25
35
REAL TEMP. = 23.40
E. ABSOLUTE = - 1.600
E. RELATIVE = - 6.400
TEMP. READ = 37
REAL TEMP. = 36.60
E. ABSOLUTE = - 0.400
E. RELATIVE = - 1.081
b) Press "ESC/NO". The factors calculated in the calibration will appear. For
instance:
OFFSET = 1033
TANGENT. = -21.27
TEMPERATURE .............. _
PUT A TEMPERATURE
SENSOR AND PRESS
ENTER
THERMOSTATING
Wait 5 minutes
The displayed waiting time decreases. After the five minutes, the photometer
asks a new temperature. Check temperature in the external thermometer.
36
OTHER OPTIONS/SERVICE/password/CALIBRATE/PELTIER
INSERT TEMPERATURE
SENSOR AND PRESS
ENTER
STEP 1
Thermostating at 25qC
Wait 5 minutes
The displayed waiting time decreases. After the five minutes, the display
shows:
STEP 1
Thermostating at 25qC
Cuvette temp..... = _
STEP 2
Thermostating at 37qC
Wait 5 minutes
The displayed waiting time decreases. After the five minutes, the display
shows:
STEP 2
Thermostating at 37qC
Cuvette temp.....= _
PELTIER CALIBRATION
37
CORRECTION FACTORS
OFFSET TANGENT
446.000 0.526
ABSOLUTE ERROR (Ea): Difference in qC between the read and the real
temperatures. Ea = Tr - T1
Ea
Er = 100 x
T1
SAVE CALIBRATION ?
f.1) Compare the obtained results with the tolerances indicated in section II.2.
f.2) If the values are outside the tolerance ranges, or to abandon the process
without memorizing the new correction factors, press ESC/NO to return to the
preceding menu or END to go to the main menu.
f.3) If the values are inside the tolerance ranges, press ENTER/YES to store
the new correction factors. The display shows:
Please WAIT
38
Once storage is completed the menu indicated in section 3.2.2. appears again.
g) To check that the calibration at 25qC and 37qC has been performed, go to
option CHECK (section 3.2.4.). Check should be done with the same
temperature sensor used in calibration. Tolerance: see section II.10.
h) In the versions of the instrument without printer the calibration data appear in
the display. First, those of 25qC. For instance,
TEMP. READ = 25
REAL TEMP. = 23.40
E. ABSOLUTE = - 1.600
E. RELATIVE = - 6.400
h.1) Press "CURSOR DOWN". The calibration data for 37qC will appear. For
instance,
TEMP. READ = 37
REAL TEMP. = 36.60
E. ABSOLUTE = - 0.400
E. RELATIVE = - 1.081
h.2) Press "ESC/NO". The calibration factors will appear. For instance,
OFFSET = 1033
TANGENT. = -21.27
h.3) Compare the obtained values with the tolerance ranges (see f.1)
SAVE CALIBRATION ?
If the values are inside the tolerance ranges, press "ENTER/YES". If storage of
calibration is not desired, press "ESC/NO".
This procedure describes the way to adjust the optimal position of the filters in
the optical path of the reading group of the photometer. The theoretical position
is established by the wheel reference photodetector but, because of the
mechanical tolerances, a further correction is necessary to optimize centering.
Calibration is performed with the filter in the position 1 of the wheel (340 nm).
39
3.4.1. Access and options
1 MANUAL
2 AUTOMATIC
b) Insert the flow cuvette in its holder making sure that the position is correct
and fill it with distilled water by means of the WASH key.
WHEEL CALIBRATION
SENSITIVITY = 20.1 nA
STEP 52 (+0)
Explanation:
STEP: Indicates the number of absolute steps done by the wheel from the
reference photodetector until filter 1 is in the current reading position. The
number between brackets indicates the increase or decrease in relation with
the number without correction (see section II.3.).
d) Using the horizontal cursors, increase (cursor right) or decrease (cursor left)
the number of steps until the position with maximum sensitivity is found. If
several consecutive steps show the same sensitivity, select the one in the
center of the range.
e) Once the calibration is completed, press ENTER/YES to store the new value
and leave the process.
40
To leave the process without storing the new values, thus keeping the former
ones, press ESC/NO (return to the preceding menu) or END (return to the main
menu).
b) Insert the flow cuvette in its holder making sure that the position is correct
and fill it with distilled water by means of the WASH key.
WHEEL CALIBRATION
SENSITIVITY = 20.1 nA
STEP 52 (+0)
Explanation:
STEP: Indicates the number of absolute steps done by the wheel from the
reference photodetector until filter 1 is in the current reading position. The
number between brackets indicates the increase or decrease in relation with
the number without correction (see section II.3.).
WHEEL CALIBRATION
SENSITIVITY : 23.1 nA
STEP 52 (+ 2)
SAVE CALIBRATION ?
e) Check that the number of adjustment steps is inside the tolerance (section
II.3.).
41
This procedure describes the way to calibrate the flux of the peristaltic pump
and the positioning of the sample in the cuvette. The pump flux depends on the
number of steps done by the motor and on the internal diameter of the
peristaltic tubing. Slight differences in the internal diameter in the different
tubings cause variations in the flow, that can also be affected by the
deformations caused by the repeated use. Also slight differences in length or
diameter affect the positioning of the sample. This calibration is intended to
compensate these problems of tolerances and aging of the tubing.
1 CHECK
2 MANUAL
3 AUTOMATIC
This option allows to enter a sample volume between 100 and 4000 PL and to
perform pump cycles to check the programmed parameters. On selecting this
option the following appears on the display:
CHECK PUMP
Insert VOLUME = _
Input the volume in PL to be checked and press ENTER. The display shows:
CHECK PUMP
Insert TUBE
From this moment on, as many aspiration cycles as desired can be performed.
42
Check that calibration has been performed. To do this:
a) Take the Volume Calibration Tube and fill it with water till the upper mark
(3.0 mL).
d) Insert the Volume Calibration Tube, with 3.0 mL of water, and perform one
aspiration cycle. Then check that the length of sample in the tube before the
cuvette is that indicated in section II.11. and that the level of water remaining in
the tube is between the two lower marks.
b) Insert the flow cuvette in its holder, making sure that it is in the correct
position.
1 SAMPLE VOLUME
2 POSITIONING
3 PUMP DELAY
PERISTALTIC PUMP
MANUAL CALIBRATION
__________________________
d.1) SAMPLE VOLUME (Calibration of the aspirated volume). Input the number
of steps the pump must do to aspirate 1 mL. The theoretical value is 917,
equivalent to 1.09PL/step. The number to be input is approximate and should
be determined by "error and trial", using the option "CHECK" in the calibration
menu (section 3.5.3.). When selecting this option the menu shows:
SAMPLE VOLUME
Actual value = 917
New value =_
43
SAMPLE VOLUME CALIBRATION
----------------------------------------------------
Actual value = 917
New value = 950
POSITIONING
Actual value = 150
New value... = _
Input the new value (i.e. 155). The following will be printed:
POSITIONING CALIBRATION
------------------------------------------
d.3) PUMP DELAY: Calibration of the time (in seconds) the pump will wait from
the moment the sipping finishes up to the moment the pump activates again to
position the sample. When selecting this option the display shows:
PUMP DELAY
Actual value = 2
New value... = _
Input the new value (i.e. 4). The following will be printed:
PUMP DELAY ADJUST
-----------------------------------
Actual value = 2
New value = 4
f) Check that the calibration has been properly performed (section 3.5.3.1.).
44
a) Switch the photometer on and wait for at least 20 minutes.
b) Insert the flow cuvette in its holder, making sure that it is in the correct
position.
1 SAMPLE VOLUME
2 POSITIONING
PERISTALTIC PUMP
AUTOMATIC CALIBRATION
----------------------------------------------
Press ENTER
d.1) Pipette precisely 5 mL of distilled water into a test tube and place it in
making sure that the Teflon tube end is at the bottom of the tube.
d.2) Press ENTER. The photometer aspirates a certain volume at normal speed
and the remaining at a reduced speed. In this moment, the display shows:
d.3) Observe carefully the bottom of the tube and press ENTER exactly when
the last drop enters into the aspiration tube. The display shows:
SAMPLE VOLUME
Actual value = 917
New value = 950
45
e) Select POSITIONING. The display shows:
e.1) Insert the tube with distilled water and press ENTER. The pump is
activated and, after a few seconds, the display shows:
REMOVE TUBE
Press ENTER
Please WAIT
POSITIONING
Actual value = 150
New value... = 155
POSITIONING CALIBRATION
----------------------------------------------
Actual value = 150
New value = 155
The display shows the preceding value (actual value) and the new one
calculated in this calibration process (new value).
g) Check that the calibration has been properly performed (section 3.5.3.1.)
a) Check that the tension selector position is that of the supply voltage (rear
part of the instrument). If not, select the correct one by slipping the selector
(with the help of a screw-driver).
46
b) Check the ground connection. With the mains cable unplugged, connect the
ohmmeter between the ground connection terminal in the power socket and the
checkpoint TP14 in the main board and check that the resistance is lower than
0.5 ohm.
c) Disconnect the connector J7 of the main board (transformer) and plug the
instrument. Using a voltmeter, read the alternating tensions in the transformer
windings and check that they are inside the tolerance limits in the table of
section II.4.
d) Unplug the instrument and connect J7. Plug the instrument again.
If the optical system has been dismounted in order to change any of its
components (lens, Peltier cell, etc.) or by any other reason, it is necessary to
verify its alignment, once re-assembled.
f) Check that the light spot is inside the outer circle (the inner circle simulates
the 1.5mm light pass in the flow cuvette.
g) If point f) is correct, the reading group is centered. Continue with point k).
h.1) Extract the lamp-holder set, giving attention to its current position.
47
h.2) Check if the lamp is properly positioned, from the point of view of deepness
and inclination.
If the lamp is properly positioned, place it again but rotating it 180º. That is, in
the other only possible position. Check it again as described in point f).
h.3) If problems are found in the lamp-holder set, correct them, position it again
and check as described in point f).
h.4) If after these operations the problems still exist, change the lamp and go
back to point f).
i) Place the photodiode again in its lodgment, taking care of its cleanness.
k) The final, functional check, of the centering of the light beam on the optical
axis of the instrument is carried out by a sensitivity test (section 8.8) with flow
cuvette and distilled water (or a macro-cuvette and distilled water in the
versions without flow cuvette).
b) Clean the external faces of the cuvette with a mixture of ethylic alcohol and
ether, then dry it with a soft paper (section 10.20.).
d) Wash the cuvette with 1 mL of washing solution and then with abundant
distilled water, by means of the "WASH" key.
e) Make sure that the instrument is on for at least 20 minutes.
f) Fill the cuvette with distilled water, by means of the "WASH" key. Check the
absence of air bubbles inside the cuvette.
g) With the cuvette filled with water measure the sensitivity of each filter. To do
it, activate TEST 300 (section 4.1.11.). Make sure that, during the test, the
cover is closed.
h) The reading process for each filter begins. A list of sensitivities (expressed in
nA) for each filter is printed.
For the versions without printer the values are shown in the display. Pressing
"ENTER" the sensitivity for each filter appears.
i) If the photometer finds a value outside the tolerance range (section II.9), it will
indicate it with a horizontal line.
48
j) Check that all the values are equal or higher than those listed in the Table in
section II.7.
If any of the data is below the corresponding expected value, check the
following:
49
4. CHECK TESTS
The test described in this chapter may be activated when needed and in any
order. To retrieve a test, follow the path:
OTHER OPTIONS/SERVICE/password/TEST
Then input the test number and press ENTER. The test number is printed in the
versions with printer.
Press each key; the beeper must sound. When pressing ESC or END the test
is left. To check these particular keys it is necessary to carry this test twice,
leaving it once with each key.
4he beeper sounds 10 times. Once completed, the test is automatically left.
A set of characters is displayed. Press ENTER and all the matrix points appear
activated thus allowing check for defective ones. Leave by pressing ESC.
A set of characters is sent through the line TxD (pin 5 of the connector COM1)
(figure 5.7) that must be received through the line RxD (pin 3 of the connector
COM1) checking that the received characters coincide with those emitted. To
carry it out it is necessary to make a jumper between said lines. If the test is
correct the display indicates "PASSED"; otherwise, it indicates "NO PASSED".
Leave by pressing ESC.
51
This test has been conceived to keep the instrument operating for hours in such
a way that its power elements are periodically activated. It is used in the factory,
before the functional QC. It may be helpful for the Technical Assistance Service
to keep the instrument under observation or to check it after repair.
During this test, the system thermostatizes at 37qC permanently. The peristaltic
pump and the filters wheel work for 1 minute each, in 5-minutes cycles. Finally
the printer (if present) prints the operation number and an asterisks line.
This test is not deleted when switching the instrument off. So, when switched
on again it begins again once the autotest is completed (although the operation
numbers begin in 1 again).
To leave the test switch the instrument off and switch it on again while holding
the key "ENTER".
Each time the ENTER key is pressed the intensity of the lamp commutes from
normal to standby.
In releases 1.1. and 1.2 of the program, although the test is implemented, the
standby is not operative in the normal function. Leave by pressing ESC.
a) Using the horizontal cursor keys, turn the pump rotor until the arrow is in
front of that of its support.
d) When stopped again, the two arrows should be in the same original position.
52
When doing this test the photometer asks:
TEMPERATURE ............... = _
PUT A TEMPERATURE
SENSOR AND PRESS
ENTER
THERMOSTATING
Wait for 5 minutes
The instrument thermostatizes until the desired temperature is reached and the
5-minutes value decreases. Once finished check that the external thermometer
indicates the programmed temperature, with the tolerance indicated in section
II.10.
This test allows to know the current generated in the photodiode for a given
amount of light, as a measurement of the sensitivity of the instrument for each
filter.
The photometer reads the sensitivity for all the filters programmed in the filters
table and prints a table (in versions without printer results appear in the
display).
This test can be performed with or without cuvette. When using the flow-cuvette
it is necessary to fill it with water by means of a WASH cycle.
The sensitivities read must be equal or higher than those indicated in the tables
of Appendix II, section 5 (sensitivity without cuvette) or Appendix II, section 7
(sensitivity with flow cuvette filled with water).
If the photometer finds a value outside the tolerance range (section II.9.), it will
indicate it with a horizontal line.
53
This test allows to know the noise detected by the converter in the signal
coming from the logarithmic amplifier.
Check that case and cover are closed to avoid entrance of light.
The instrument selects the 340 nm filter and zeroes. After approximately 1
minute, the results are displayed. They must be inside the tolerance ranges of
the Table in section II.6.
This test allows analysis of a set of repeated readings of the same sample.
Program all these parameters, depending on the sample. The duration of the
test will be the result of the interval time and the number of intervals.
When required, insert the baseline and then the sample. If the sample is the
same as the baseline, the zero stability will be tested.
For a routine test, make the BASELINE with calibrator nº 0 of the calibration kit
(SERVICE TOOLS, code 005) and use calibrator nº 4 as sample.
The number of the reading, the time and the absorbance are printed (only
versions with printer). At the end, statistical data are printed: average, CV,
maximum and minimum absorbances.
This test allows to perform successive readings changing the sample every
time. If the sample is always the same, this test checks the precision of the
instrument (repetitivity between identical samples).
54
When required, insert the baseline and then as many samples as programmed
intervals. At the end, some statistical data are printed: average, CV, maximum
and minimum absorbances.
- Reading Filter
- Stabilization Time
- Sample Volume
Insert the baseline when required, then the first sample for 5 consecutive times
(R1 to R5) and then the second sample for 5 consecutive times again (R6 to
R10). The photometer calculates the contamination using the formula:
R6
Contamination = ( - 1 ) x 100
Xav
This test allows to check the accuracy when reading a sample of know
absorbance.
- Reading Filter
- Stabilization Time
- Sample Volume
INSERT
REFERENCE = _
Input the theoretical value of the sample by means of the numeric keyboard.
Then the instrument asks:
INSERT SAMPLE
55
Introduce the sample. The following data are printed:
- Reading Number.
- Absorbance applying the calibration.
- Absorbance without applying the calibration.
- Reference value (Theoretical value of the sample).
- Absolute and relative errors applying the calibration.
- Absolute and relative errors without applying the calibration.
This test is for internal use in the factory and has no direct application for the
Technical Assistance Service.
This test resets the personalization to the default parameters listed in Appendix
IV.
It includes 4 tests:
- Program Checksum
- Tests Checksum
- Calibrations Checksum
- Personalization Checksum
These tests control the integrity of the data and the program used by the
microprocessor. They are run automatically in the AUTOTEST and cannot be
activated by the user. Further information appears in the user`s manual.
56
5. MAINTENANCE
b) Remove the printer paper roll and take the aspiration tube out its guiding
tube.
c) Remove the two screws (1) located in the lower part of the instrument.
d) Remove the two screws (2) located in the rear part of the instrument.
e) Carefully lift the case, pull it up and incline it to the rear, taking care that the
connecting strip that joins it to the main board is not stretched.
f) If the case if to be fully removed, disconnect the flat band (J1) and the supply
strip (J2) from the main board.
g) Connect the flat band (J1) again, as well as the supply strip (J2) (take into
account the polarization of the connectors).
i) Place the printer paper again and the aspiration tube (figures 2.6 and 2.7).
d) Remove the 5 Allen screws (1) fixing the radiator to the base.
57
58
e) Remove the 3 screws (2) fixing the main board to the respective separators.
f) Put a thin and uniform silicone layer in the lower part of the radiator (3).
g) Place the board again and fix it with the 5 Allen screws (1) and the 3 screws
(2).
h) Solder the photodiode cable thoroughly cleaning the solder resin with
alcohol. Resin remaining in this point may cause instability of the reading
system.
a) Disconnect, in the board itself, the keyboard band (J17) and the printer
connectors (J13, J14 and J15).
b) Disconnect the flat strip and the supply connector in the main board (J1 and
J2).
d) Place the board taking into account that the LED's should be properly
placed.
f) Connect the keyboard strip (J17) and the printer connectors (J13, J14 and
J15).
g) Connect the main board, the flat band and the supply connector (J1 and J2).
The printer is fixed to the case by means of a metallic support. To change the
printer proceed as follows (see figure 5.3):
a) Disconnect the connectors J13, J14 and J15 from the display board.
59
60
c) Remove the 4 screws (3).
61
To place the printer again, proceed as follows:
d) Fix the printer to the metallic support with the 4 screws (3).
f) Connect the connectors J13, J14 and J15 of the display board.
The keyboard is glued to the case and, when removed, it cannot be used again.
Thus make certain that it is defective before removal. To change it, proceed as
follows (see figure 5.4):
d) Place the new transformer fixing it to the base with the 4 screws (4).
62
63
5.7. Removal of the cuvette holder tray
b) Remove the waste tube from the outlet (8) in the lower part of the tray.
c) Disconnect the peristaltic pump motor strip (J6) from the main board.
d) Remove the 2 screws (6) and the screw (7) that fix it to the optical support.
f) Place the tray and fix it to the optical support with the 2 screws (6) and the
screw (7).
g) Connect the peristaltic pump motor strip (J6) to the main board.
h) Connect the waste tube to the outlet (8) in the lower part of the tray.
e) Put the wheel (5) in its place, taking into account to position the belt (8) and
the two washers (19) as indicated in the figure.
If the filter holders were removed, place them again in the same order as they
were. If order is changed, the filters table must be re-programmed.
64
5.9. Change of the filter wheel motor
b) Disconnect the strip of the motor from the main board (J4).
d) Change the motor (2) placing the belt (3) as indicated in the figure.
f) The screw lodgings have a little slackness to allow stretching the belt. Avoid
excess stretching.
a) Disconnect the peristaltic pump motor strip (J6) from the main board.
c) Remove the pump and substitute it fixing the new one with the 4 screws (5).
d) Connect the peristaltic pump motor strip (J6) to the main board again.
e) Desolder the Peltier Cell cables from those going to the connector (J10).
65
66
f) Remove the 4 screws (4).
i) Take a new cell and place a thin and uniform silicone layer on both faces.
j) Place the cell in its position taking into account that its black cable should be
in the front part of the optical system.
k) Place the cuvette holder and fix it with the 4 screws (14). During this
operation, take special care to maintain the cuvette holder parallel to the cell,
and the cell parallel to the optical support, and progressively screw the four
screws, in order to avoid breakage of the ceramic faces of the cell.
l) Once the cell is positioned, check that no silicone overflows, that could
termically short-circuit the two faces of the cell. Thus the case, clean it carefully.
m) Solder the cell cables again to those of the connector J10, according to the
color codes.
b) Remove the screw fixing the ground terminal (21) to the cuvette holder.
d) Take the support (3) and the separator (2) away, backwards.
67
i) Move the support (3) forward until the separator (2) contacts the photodiode.
f) Place a new fan, taking into account that the air flux is outwards.
g) Place the grid again (10) and fix the set with the 4 screws and nuts (9).
c) Take the support (3) and the separator (2) away, backwards.
68
69
d) Desolder the cable from the sensor pins (4).
f) Move the support (3) forward until the separator (2) contacts the photodiode.
b) Wait until the lamp radiator is cold, before removing the lamp holder.
e) Remove the lamp holder (4), loosen the Allen screw (3) and remove the
lamp.
f) Insert the new lamp (5), fully introducing the pins. Tighten the Allen screw (3)
until the lamp is tightly secured. Do not touch the glass bulb with the fingers; to
manipulate the lamp, the cover itself may be used, cutting it by the terminals
side and pressing until they come out.
g) Re-insert the lamp holder in its place. Position the flange (2) and tighten the
Allen Screw (3).
i) The lamp does not require any adjustment at all. Nevertheless, it is necessary
to check the correct centering. To carry it out, use the option LAMP TEST in the
menu OTHER OPTIONS .
If a filter with the wavelength different from that of the filters installed in the
instrument is to be installed, it should be inserted in one of the two free
positions of the filters wheel. The filter will come fitted in a filter holder, marked
70
with the corresponding wavelength and thus it is only necessary to manipulate
the filter holder. Proceed as follows (see figure 5.8):
b) Insert the filter holder (1) in one of the free positions, by simply pressing it.
Do not touch the filter faces with the fingers.
d) Now it is necessary to incorporate the new filter into the FILTERS TABLE. To
do it, switch the instrument on and select the path: PROGRAMMING/FILTERS
TABLE and add the wavelength value in the position where it has been
inserted. The filter is thus ready to be used.
If by any reason it is necessary to take the filter away from its holder, proceed
as follows (see figure 5.9):
e) Using the corresponding tool (4) from the SERVICE TOOLS KIT (code 005),
unscrew the filter subjection ring (3).
f) Take the filter (2) away from its holder taking care not to touch the faces with
the fingers.
g) Insert the new filter, taking into account the direction of the arrow (5), and fix
it to the holder by means of the screwed ring, using the tool indicated in point
e).
e) Using the corresponding tool from the SERVICE TOOLS KIT (code 005),
unscrew the lens holder set to be removed (4 or 11).
71
72
If it is necessary to take the lens away from the inside of the filter holder set,
proceed as follows (see figure 5.10):
f) Using the corresponding tool (4) from the SERVICE TOOLS KIT (code 005),
unscrew its fixation ring (3).
g) Remove the lens (2) from its holder (1) taking care not to touch its faces with
the fingers.
h) Insert the new lens and fix it in the holder with the screwed ring using the tool
indicated in point f).
a) Never use detergents or abrasive products for cleaning the outside of the
instrument. Use only a cloth with water and neutral soap.
c) The cuvette holder tray is equipped with watertight joint in order to prevent
penetration of liquid into the inner part of the instrument. If liquid is spilt into the
tray, clean it with damp paper or cloth. A drainage hole connected to the
outside by a silicone tubing is in the front left part of the tray to facilitate
drainage of poured liquid.
d) If a cuvette breaks into the cuvette holder or for any reason liquid pours into
it, the holder has a drainage hole connected to the outside. It will, however, be
necessary to rinse and dry the inside of the cuvette holder.
e) Cover the photometer with its dust cover when not in use.
73
74
75
5.20. Cleaning of the optical components
The following indications should be taken into account for the cleaning of the
optical components:
b) Recommended material:
- The area for manipulating the instrument should be clean and in order.
- As the components are fragile, they should be treated carefully; a fall could
result in breakage.
- Avoid touching the active areas with the fingers. Lenses, filters and
photodiode should be held by their sides; lamps by the connecting terminals.
- To clean the components, first take dust away using the rubber bulb; thus
scratches caused by small particles on the surface can be avoided, when
rubbing with paper.
- In the case of persistent or greasy dirt, clean with a paper soaked with the
alcohol/ether solution and then with a dry paper. Sometimes, for instance when
cleaning the filters or the photodiode window, the cotton ear picks may be
helpful together with the paper in the most delicate parts.
- After any cleaning it is convenient to repeat blowing with the rubber-bulb thus
eliminating any residual paper or cotton nap.
a) Dismount the filter holders from the wheel and take the filters away as
indicated in section 5.16.
b) Clean them as indicated in section 5.20.
c) Mount them again in their holders and the holders in the wheel as indicated
in section 5.16.
76
5.22. Cleaning of the lenses
a) Dismount the lens holder from the optical support and take the lenses away
as indicated in section 5.17.
c) Mount them again in their holder and the holders in the optical support, as
indicated in section 5.17.
At the end of the working day, wash thoroughly with a detergent solution such
as the one provided with the instrument. Lastly, rinse with distilled water and
empty the circuit by performing wash cycles with air.
Finally, to maximize the life of the peristaltic tubing, it is advisable to take it out
from its mounting, so that it remains loose and without tension. On starting a
new work session, re-insert it in its place.
b) If the outside end of the sip tubing has deteriorated, a few millimeters may be
cut making a perpendicular and clean cut. In this case, the POSITIONING
parameter must be re-calibrated.
77
Cleanness of both the outside and the inside of the flow-cuvette is very
important. Proceed as follows:
b) To clean the outside, use alcohol and then dry with a soft paper (section
5.20).
It is important to avoid dust in the instrument, that could affect the optical
system. Carefully remove dust from the inside of the instrument, specially from
the fan vanes.
78
APPENDIX I. TECHNICAL SPECIFICATIONS
- Repeatability: cv<= 1.2 % at 505 nm between 0.05 and 0.1 A (20 readings)
cv<= 0.1 % at 505 nm between 1.5 and 2.0 A (20 readings)
- Peltier system
- Temperature range: 23 to 40qC
- Programming steps: 1qC
- Stability: r 0.2qC
Temperature sensor
- Resolution: 10 mV/qC
- Output tension: 10 (mV) x T (qKelvin)
79
I.3. Printer (only in versions with printer)
I.4. Display
I.5. Electronics
Amplifier
A/D Converter
Calibrations by software
I.6. Communications
80
I.7. Power supply
Peristaltic pump
SAMPLE CARRY-OVER
VOLUME (%)
Table Y.1
81
82
APPENDIX II. CALIBRATION TOLERANCES TABLES
Adjustment steps:
Maximum: + 10
Minimum : - 5
83
II.4. AC Voltage of the Transformer Secondaries
340 50
405 200
420 400
450 1000
492 1500
505 1500
530 2000
546 3000
578 4000
600 5000
630 6000
670 7000
If a line appears instead of the value, it means that the sensitivity of this filter is
outside the reading range of the zero signal of the photometer (section II.9.)
84
II.7. Sensitivity of the Filters with Flow Cuvette
340 20
405 60
420 100
450 200
492 300
505 500
530 750
546 800
578 900
600 1000
630 1750
670 2000
If a line appears instead of the value, it means that the sensitivity of this filter is
outside the reading range of the zero signal of the photometer (section II.9.).
Maximum: 24 v
Minimum: 15 v
Maximum: 30000 nA
Minimum : 10 nA
The trail of the sample that remains in the aspiration tubing before entering into
the cuvette should be between 0 and 10 mm.
85
NOTE: All the values are statistical and can suffer variations when new
production series will be released. Thus the case, this appendix will be
updated.
86
APPENDIX III. MENUS TREES
2 PELTIER 1 LIST
2 CHECK
3 CALIBRATE
4 TEST PROG.
1 CREATE NEW TEST
2 REVIEW
5 UNITS PROG.
6 LOAD MEMORY 3 DELETE
7 STORE TEST 4 COPY
5 MOVE
3 CALIBRATE PUMP
4 COMMUNICATIONS
1 CHECK
2 MANUAL
1 SAMPLE VOLUME
2 POSITIONING
3 PUMP DELAY
3 AUTOMATIC
1 SAMPLE VOLUME
2 POSITIONING
87
LEVEL I LEVEL II LEVEL III LEVEL IV LEVEL V
2 PELTIER 1 LIST
2 CHECK
3 CALIBRATE
4 TEST PROG.
1 CREATE NEW TEST
2 REVIEW
5 UNITS PROG.
6 LOAD MEMORY 3 DELETE
7 STORE TEST 4 COPY
5 MOVE
88
APPENDIX IV. DEFAULT PARAMETERS
PERSONALIZATION
FILTERS TABLE
1................................... 340
2................................... 405
3................................... 420
4................................... 505
5................................... 546
6................................... 578
7................................... 670
8................................... ---
9................................... ---
UNITS TABLE
1................................... Pg/dL
2................................... mg/dL
3................................... g/L
4................................... U/L
5................................... Pmol/L
6................................... mmol/L
7................................... nkat/L
8................................... Pkat/L
COMMUNICATIONS
CONFIG. RS-232:
Bauds........................... 9600
Bits............................... 8
Parity ........................... None
Stop bits....................... 1
PROTOCOLS
XON-XOFF .................. NO
ACK-NAK..................... NO
89
90
APPENDIX V. MAINTENANCE PLAN
V.1. Cleaning
V.2. Change
- Aspiration tubing
- Peristaltic tubing
- Inner waste outlet tubing
- Outer waste outlet tubing
- Lamp (if blackened)
V.3. Review
V.4. Check
(1) Remove filters, lenses and photodiode and proceed with the general
cleaning.
91
92
APPENDIX VI. SPARE PARTS AND ACCESSORIES
VI.1. Accessories
Code Description
--------------------------------------------------------------------------------------------------------
User's Manual
AC4460 Paper Lodgment Cover
AC964 Paper Roll TB 50 KS 4AO mtp 401 ( )
AC3778 Flow Cuvette Hellma 178, 712-OS 8.5 mm (#)
AC944 Cuvette, Hellma 6030
TU2791 Teflon Tubing, Hellma 040.203 (#)
AC3594 Cuvette Outlet Connector (#)
TU1643 Cylindrical Silicone Connector (#)
AC3486 Peristaltic Tubing (#)
AC3114 Silicone Tubing 3x6 (200 cm) (#)
AC3495 Adapter for test tubes
AC3703 Waste Bottle (#)
*** Bottle 100 mL washing solution (#)
FU656 Fuse, 1 A
AC3097 Fuse, 2 A
*** Supply Cable
AC3979 Dust Cover
LA2151 Halogen Lamp, 12 v, 20 w
FI4649 Filter Set, 340 nm
FI4650 Filter Set, 405 nm
FI4651 Filter Set, 420 nm
FI4658 Filter Set, 492 nm
FI4652 Filter Set, 505 nm
FI4653 Filter Set, 546 nm
FI4654 Filter Set, 600 nm
FI4659 Filter Set, 620 nm
FI4655 Filter Set, 670 nm
FI5557 Filter Set, 578 nm
*** Cover
** P.C. Board Micro I81002
ME4430 Reading Group, complete
DI2964 Photodetector S-1336-8BQ
IN4342 Switch EATON E30M11J01
ZO648 Fuse Holder D (06.26)
VA4343 Supply Filter FD-1Z
DI844 Diode LM-335
MO3560 Approved fan
ME3215 Fan Grid
93
TR4344 Transformer E81007
IN4345 Tension Selector 18-000-0016
VA1467 Terminal MOLEX 4809
ZO476 Connector, 2.54 H 4
ZO474 Connector, 2.54 H 3
CI4533 PAL20V8, programmed (Release G81001A)
CI4534 PAL16V8, programmed (Release G81002A)
(**) EPROM 27C2001
(**) E2PROM 2864
ZO4648 Mechanized Lamp Holder
DI4438 Wheel Photodetector, mounted
PC4449 Display Circuit I81004, checked
IM3559 Printer SEIKO STP312-F256 B ( )
VA4413 Display Protector
*** Keyboard
AC4290 Calibration Box
94
APPENDIX VII. PASSWORD
251293
95
96
1 MODIFICATION
LOCALIZATION: Monocard
* The modification was implemented from the main board number 4088 on.
1-1
2- MODIFICATION
Modify the codes of accessory and spare parts list. And improve their
description.
2-1
2-2
APPENDIX VI. SPARE PARTS AND ACCESSORIES
VI.1. Accessories
Code Description
--------------------------------------------------------------------------------------------------------
User's Manual
AC10450 Paper Lodgment Cover
AC10410 Paper Roll (*)
AC10451 Flow Cuvette Hellma (#)
AC10452 Macro Cuvette Hellma
TU10249 Teflon Tubing Hellma (#)
AC10411 Cuvette Outlet Connector (#)
TU10412 Cylindrical Silicone Connector (#)
AC10413 Peristaltic Tubing (#)
AC10414 Silicone Tubing 3x6 (200 cm) (#)
AC10453 Adapter for test tubes
AC10454 Waste Bottle (#)
*** Bottle 100 mL washing solution (#)
FU10416 Fuse, 1 A
FU10417 Fuse, 2 A
CA10455 European mains lead
CA10456 American mains lead
AC10457 Dust Cover
LA10418 Halogen Lamp, 12 v, 20 w
FI10458 Filter Set, 340 nm
FI10459 Filter Set, 405 nm
FI10460 Filter Set, 420 nm
FI10461 Filter Set, 492 nm
FI10462 Filter Set, 505 nm
FI10463 Filter Set, 546 nm
FI10464 Filter Set, 578 nm
FI10465 Filter Set, 600 nm
FI10430 Filter Set, 620 nm
FI10467 Filter Set, 670 nm
*** Cover
** P.C. Board Micro
ME10545 Reading Group, complete
DI10242 Photodetector
IN10488 Mains switch
2-3
ZO10407 Fuse Holder
VA10489 Mains Filter
DI10246 Temperature sensor set
MO10408 Fan
ME10490 Fan Grid
TR10558 Transformer
IN10492 Voltage Selector
ZO10493 Lamp Holder
DI10409 Wheel Photodetector
PC10587 Display Circuit
IM10494 Printer (*)
VA10546 Display Protector
*** Keyboard
AC10495 Calibration Box
2-4
3. MODIFICATION
3-1
3-2