Anaesthetic Monitoring Devices
Anaesthetic Monitoring Devices
Anaesthetic Monitoring Devices
Anesthetic
Monitoring
Devices to Use &
What the Results Mean
Jeff Ko, DVM, MS, Diplomate ACVA, and
Rebecca Krimins, DVM
T
he main fundamental aspects of anesthetic monitor-
ing are:
series discussing the goals of 1. Oxygenation (circulatory and respiratory func-
tion)
anesthetic monitoring as well 2. Ventilation (respiratory function)
as associated procedures 3. Circulation (circulatory function with an empha-
sis on cardiac output).
and equipment. In the first These three elements work simultaneously in order to maintain
adequate tissue and organ perfusion with oxygenated blood.1-3
article, the authors answered Oxygenation and ventilation are essential for maintaining a
high oxygen level in the blood, while cardiac output plays a
questions about anesthetic pivotal role in maintaining tissue and organ perfusion with highly
monitoring, including why oxygenated blood. In this way, oxygenation, ventilation, and cir-
culation each play a critical role in providing oxygen to tissues.
itÕs performed, information
PHYSIOLOGY FUNDAMENTALS
obtained during monitoring, Oxygen Delivery
Oxygen delivery is a product of blood oxygenation and cardiac
and important components output.1-3
of the anesthetic process. • Blood oxygenation is represented by blood oxygen con-
tent, which is the total amount of oxygen carried by the
blood, including oxygen dissolved in plasma and oxygen
bound to hemoglobin.1-3
• Cardiac output is maintained by a complex interaction
involving heart rate, stroke volume, peripheral vascular
resistance (afterload), blood volume returning to the heart
(preload), and blood viscosity.1-3
Table 2. cardiorespiratory & Physiologic Parameters in the Anesthetized Dog & cat
Variable reference interval (Dogs) reference interval (cats)
Circulation heart rate (beats per min) 60–120 120–160
systolic blood pressure (mm hg) 90–140 90–140
Diastolic blood pressure (mm hg) 60–90 60–90
Mean arterial blood pressure (mm hg) 70–90 70–90
Ventilation respiratory rate (breaths per min) 8–16 12–24
Tidal volume (ml/breath) 10–15 10–15
Arterial blood ph 7.35–7.45 7.35–7.45
Paco2 (mm hg) 35–45 35–45
Bicarbonate (mmol/l) 22–26 22–26
end-tidal co2 (mm hg) 35–45 35–45
Oxygenation spo2 (%) ≥ 95 ≥ 95
Pao2 (mm hg) ≥ 100 ≥ 100
Other Body temperature (°F) 98–101 98–101
hematocrit (%) 34–59 28–47
Total protein (mg/dl) 5–8.3 5.9–8.4
Blood glucose (mg/dl) 90–150 90–150
Blood lactate (mmol/l) <2 <2
Urine output (ml/kg/h) 1–2 1–2
CO2 = carbon dioxide; PaCO2 = partial pressure of carbon dioxide in the arterial blood; PaO2 = partial pressure
of oxygen in the arterial blood; SpO2 = saturation level of oxygen in hemoglobin as measured by pulse oximetry
Figure 1. A schematic representation of clinical anesthesia monitoring equipment to ensure proper tissue perfusion
with well-oxygenated blood. Continous systemic surveillance can provide an early warning system, prompting
immediate intervention.
Disadvantages:
• Technical skill required to place the catheter
• Expense of equipment required
• Frequent maintenance (of pressure transducer,
pressure tubing, saline flushing) and calibration
• Potential hematoma formation and thrombosis
(rare)
• Infection at the catheter insertion site and acute
blood loss if an inadvertent disconnection occurs
between the catheter and artery (rare).
Placement
The BP cuff is inexpensive and
placed on a limb extremity or the
base of the tail and leveled with
the heart during measurement.
The width of the blood pressure
cuff should be 40% of the limb
circumference.
VENTILATION
Various aspects of respiratory function can be mea-
Figure 5. An oscillometric monitor with BP cuff
sured using a respirometer, capnometry, and/or blood
used on a front limb of a dog: Notice the oscillo-
gas analysis.
metric BP provides systolic (96 mm Hg), diastolic
(63 mm Hg) and mean arterial blood pressures (74
mm Hg) with heart rate (109 beats/min). Respirometer
Respirometry assesses tidal volume and minute vol-
ume in the anesthetized patient. Minute volume (mL/
• BP has to be measured each time by the anes- min) is the product of tidal volume (mL) and respira-
thetist because the equipment cannot be set for tory rate of the patient:
automatic measurements.
• When vasoconstriction or hypotension occurs, the minute volume = tidal volume × respiratory rate
signal can be relatively weak and difficult to obtain.
A respirometer measures expiratory volume and can
Oscillometric Method be placed between the expiratory limb of an anes-
Oscillometric BP monitoring (Figure 5) is used to thetic machine and the anesthetic breathing hose.
measure systolic, diastolic, and mean arterial BP. Alternatively, it can be connected to a tightly-fitting face
Capnometry
Figure 6. A respirometer attached to the face
End-tidal carbon dioxide (ETCO2) monitoring allows
mask of a dog to measure tidal volume and
exhaled CO2 to be measured noninvasively and
respiratory rate
reflects PaCO2. Monitoring ETCO2 is useful for:
• Determining optimal ventilation efficiency
mask and used to assess ventilation in a nonintubated, • Diagnosing respiratory, airway, or device prob-
sedated patient (Figure 6). lems, such as apnea, hypoventilation, airway dis-
The true usefulness of respirometry is to pinpoint connection, airway obstruction, leak in the endo-
whether respiratory rate and/or tidal volume are inad- tracheal tube cuff, exhaustion of CO2 absorbent,
equate (based on the equation of minute volume) when and incompetent one-way valve of anesthetic
end-tidal CO2 or PaCO2 is elevated. rebreathing circuit
• Reflection of adequate cardiac output production
Blood Gas Analysis (eg, sudden, acute drops in cardiac output are
Arterial blood gas analysis and resultant partial pres- associated with decreases in ETCO2 measure-
sure of CO2 in arterial blood (PaCO2) can be used ments due to poor pulmonary circulation).
to assess ventilation in the anesthetized patient. Capnometry is the measurement and numerical
The arterial blood sample is collected directly from display of ETCO2 during the respiratory cycle (inspi-
an artery or through a preplaced arterial catheter ration and expiration). A capnometer measures and
(Figure 3). displays the readings without a graphic presentation;
Although more accurate than a respirometer, PaCO2 capnography refers to the comprehensive measure-
measurement requires a blood gas analyzer and arte- ment of CO2 using a graphic recorder that displays CO2
rial blood samples. If arterial blood samples are not concentration in real-time, throughout the respiratory
available, venous blood samples may be used instead. cycle (Figures 7 and 8).
Figure 7. Side-stream capnography: Notice the Figure 8. A main-stream anesthetic agent analyzer, with
adaptor (white color) is connected between a an adaptor connected between a breathing circuit and
breathing circuit and the endotracheal tube to the endotracheal tube, analyzes isoflurane concentration
sample the inspired and expired gas (via the (inspired concentration of 1.6% and expired concentration
transparent tubing); the sample is analyzed in of 1.3% ) and end-tidal CO2 (ETCO2 of 36 mm Hg, respira-
the monitor. Also visible is an esophageal stetho- tory rate of 17 breaths/min) of the anesthetized dog. Notice
scope (black tubing) inserted into the esophagus the main-stream adaptor is bulkier than the side-stream
of the dog for auscultation of heart and lung adaptor (Figure 7) and the exhaled gas is analyzed (blue
sounds, and a pulse oximeter probe (white tubing) light) within the adaptor using infrared technology.
on the patient’s tongue.
This technique:
• Is relatively inexpensive.
• Allows for continuous monitoring of ventilation.
• Limits the need for invasive procedures, such as
arterial blood gas sampling.
• Provides valuable information on the respira-
tory status of the patient, including whether the
patient has had an inadvertent esophageal intu-
bation.
• Identifies if the patient is rebreathing carbon
dioxide (eg, too much dead space within breath-
ing circuit, inadequate oxygen flow rate used for
non-rebreathing circuit, CO2 absorbent exhaust-
ed, or a mechanical problem with one-way valve Figure 9. Pulse oximetry with a lingual probe
of anesthetic rebreathing circuit). placed on the tongue of a dog. Notice the place-
ment of wet gauze between the probe and the
What the Numbers Mean tongue to minimize air pockets and improve the
ETCO2 measurements usually underestimate actual contact between the tongue and the probe.
PaCO2 measurements by 5 to 10 mm Hg. The discrep-
ancy between ETCO2 and PaCO2 measurements is
due to alveolar dead space. In the conscious patient,
usually attached to the patient’s tongue, lip, ear,
ETCO2 and PaCO2 measurements are very similar.
or interdigital space)
However, the anesthetized patient has increased
• Provides the pulse rate of the patient.
alveolar dead space, which results in lower ETCO2
Pulse oximeter function may be affected by many
measurements.
factors, including:
• Normal PaCO2 measurements in the anesthetized
• Motion artifact (eg, shivering or body movement)
dog and cat are equal to 35 to 45 mm Hg.
• Ambient light (eg, fluorescent light affecting
• ETCO2 (and PaCO2) values higher than 45 mm
proper reading of pulse oximeter)
Hg are consistent with hypoventilation. They
• Poor peripheral blood flow due to hypotension or
alert the anesthetist that the patient may need to
vasoconstriction
breathe more frequently and/or with a larger tidal
• Electrical noise from surgical instruments (eg,
volume. Hypoventilation is usually associated
electrocautery)
with anesthetic-induced respiratory depression.
• Increased blood carboxyhemoglobin and methe-
• ETCO2 (and PaCO2) values lower than 35
moglobin levels
mm Hg may be evidence of hyperventilation.
• Dark pigmentation of the skin or tongue.
Hyperventilation is often associated with pain,
light plane of anesthesia, and/or increased body
Relationship Between SpO2 and PaO2
temperature.
• Normal pulse oximeter readings (SpO2) in anes-
For a complete list of normal cardiorespiratory and
thetized dogs and cats breathing 100% oxygen
physiologic parameters in the anesthetized patient, see
should be 98% to 100%.
Table 2, page 25.
• Normal PaO2 in the anesthetized dog and cat
breathing 100% oxygen should be greater than
OXYGENATION
200 mm Hg and can be as high as 650 mm Hg.
A patient’s oxygenation status can be measured by
• A SpO2 of 90% corresponds to a PaO2 of 60 mm
blood gas analysis for PaO2 (see Blood Gas Analysis
Hg, which indicates hypoxemia. Hypoxemia is
under VENTILATION), hemoximetry, and pulse
insufficient oxygenation of arterial blood and is
oximetry. Hemoximetry and pulse oximetry both
considered to be present if:
analyze hemoglobin saturation in oxygen, with
» PaO2 is less than or equal to 60 mm Hg
hemoximetry providing a more precise measure-
» SpO2 is less than or equal to 90%.
ment. However, it is unusual to use hemoximetry in
In the clinical setting, PaO2 can be estimated using
veterinary practice.
pulse oximetry:
PaO2 = SpO2 – 30
Pulse Oximetry
(for pulse oximeter readings between 75% and 90%)
Pulse oximetry (Figure 9):
This formula only applies to a certain range of pulse
• Calculates the percentage of oxyhemoglobin and
oximeter readings because of the linear relationship
reduced hemoglobin present in arterial blood
between the PaO2 and SpO2 values on the mid portion
• Provides noninvasive, continuous detection of
of the hemoglobin disassociation curve. Outside of
pulsatile arterial blood in the tissue bed (probe is
these values, this rule cannot be applied.
Urine Output
Under general anesthesia, normal urine output (1–2 mL/
kg/H) represents adequate kidney perfusion and assumes
proper perfusion of other organs as well. For patients
with renal failure, severe dehydration, or acute hemor-
rhage, it is important to monitor and check urine output
during surgery and recovery. If a urinary catheter is not
preplaced, palpation of the animal’s bladder or visual
estimation of the volume of urine voided can provide a
crude estimate of urine production.
Lack of urine production represents hypoperfusion of
the kidneys and other tissues, as well as possible kidney
failure. Until adequate urine production is achieved,
rehydration and an inotropic agent may be used to treat
hypotension or tissue hypoperfusion during anesthesia
and the recovery period. Patients that have low urine
output with normal tissue perfusion and blood pressure
may be in renal failure and might require treatment with
diuretics. ■
References
1. Cilley re, Scharenberg AM, Bongiorno PF, et al. Low oxygen delivery
produced by anemia, hypoxia, and low cardiac output. J Surg Res 1991;
51(5):425-433.
2. Marx G, reinhart K. venous oximetry. Curr Opin Crit Care 2006; 12(3):263-
268.
3. Mellema M. Cardiac output, wedge pressure, and oxygen delivery. Vet Clin
North Am Small Anim Pract 2001; 31(6):1175-1205.
4. Allen Se, Holm JL. Lactate: Physiology and clinical utility. J Vet Emerg Crit
Care 2008; 18(2):123-132.
5. Tas O, de rooster H, Baert e, et al. The accuracy of the Lactate Pro hand-
held analyser to determine blood lactate in healthy dogs. J Small Anim
Pract 2008; 49(10):504-508.