Strother ', M.E.: S.C. Casey' Hoffman'
Strother ', M.E.: S.C. Casey' Hoffman'
Abstract
Sensitivityparameters derived from a plot of a scanner's as it eliminates the 25%-30% change in the minor to major
true coincidence count (TCC) rates as a function of activity axes of the elliptical cross section of the average human
in a 20cm cylindrical phantom have no direct link to image head. More experimental work is needed to address these
quality. Noise equivalent count (NEC) rate curves, which issues.
incorporate the noise effects of subtracting the randoms and However, even if a set of standard phantoms is available
scatter count components provide a direct link between the issue of what performance parameters to measure
image signal-to-noise ratios and the scatter, randoms and remains. We suggest, as have others [6-91 that for the
trues coincidence count rates. We measured TCC and NEC purposes of measuring PET scanner sensitivity,image quality
curves with a standardized 20cm diameter nylon cylinder for should be defined in terms of the reconstructed image signal-
five different CI'I/Siemens PET scanners with several to-noise ratio (SNR) in a uniform object which is directly
scanner-collimator combinations: (1) 831/08-12 with lmm related to a PET scanner's noise equivalent count perfor-
collimator septa; (2) 933/08-12 and 933/08-16 with 9 1 m m mance. The basic NEC expressions are obtained below from
tapered collimator septa; (3) 931/08-12 with P l m m tapered a simple tomographicS N R derivation and are then extended
and a lmm collimator septa and the 931/08-16 with 9 1 m m to include the additional noise effects of realistic scatter and
tapered collimator septa. In addition, we compared TCC randoms corrections.
and NEC curves on the 933/08-12 with those from an The concept of noise effective (or equivalent) counts
Alderson brain phantom. In general we found that the TCC derives originally from the work of Shaw [lo]. The idea has
curves indicated peak count rates and activity levels that been used in the development of a unified theory of radiation
were as much as 50% higher than the corresponding values imaging and as a measurement of noise equivalent quanta
from NEC curves. The primary factor causing this diff- (NEQ) has been applied to the assessment of imaging
erence is the noise effect of the randoms component. In efficiency in forming a figure of merit for S N R transport
addition, we demonstrated that compared to the Alderson through an imaging system. This figure of merit, called the
brain phantom the standard 20cm cylinder is a poor pred- detective quantum efficiency (DQE) has been applied to a
ictor of count rate performance for PET brain imaging. wide range of medical imaging devices [ll].
Introduction Theory
The 20cm diameter cylindrical phantom has become Consider the S N R of a reconstructed image element
the de facto object for measuring PET scanner sensitivity. following the approach of Brownell et al. [12].
The derived performance measurements range from the low
activity sensitivity specification - count rates in kcps/pCi/cc SNR = c te(vAFt,)-'/2
typically measured at about .1 pCi/cc - to the phantom
activity levels for (1) 50% deadtime, (2) saturation and (3) where c = a constant [13].
randoms equal to trues (plus scatter). For brain scanning te = true coincidence counts in a reconstructed
the use of the 20cm diameter cylinder has been questioned image element.
by Links and Wagner [1,2] in favor of a 15cm diameter and VAR, = L(weighted variance of the samples from
several authors have reported PET scanner sensitivities and each of the m projection angles contri-
scatter fractions for a range of diameters from 14cm-20cm buting to the image element).
[3,4]. None of these test cylinders bears a particularly close
resemblance to the elliptical dimensions of the average adult The coincidence components without attenuation and dead-
human head [5] - Males:19.6cmxl5.5cm; Females:18.5cmx time correction, in each projection sample are - trues(tp),
14.5cm. A test cylinder diameter of 7 inches or 17.8cm, scatter (s ) and randoms (rp). The variance of m projection
while better than 20cm may still not be a very good standard samples [t, t sp + rp)-(spt rp)], is combined to give
1
Present Address: PET Imaging Section (llp), VA Medical Center, 1 Veterans Dr., Minneapolis, MN 55417
scanner-collimator combinations as the ratio of the counts elliptical cross-section of 19cmxl5cm, we choose a, = 0.10
under the exponential tails to the central peak counts, for a based on the 18cm cylinder value for the 831 PET scanner
scatter proftle obtained from a rod source in the center of a [l5],and f=0.36 (for an 18cm diameter).
2Ocm diameter water filled phantom [SI. The closest values Axial Inter a,=
of a,, (2Ocm cylinder) available for most scanners integrated Scanner Ring Crystal Septa Septa Septa (Scatter/
the scatter tails out to l5cm from the center of the profile. Model Diam. Width Depth Width Gap' TNS)
For two data sets (see Table 1)it was necessary to estimate 931/06-l2 l a m 12Anm 18Scm 3-lmm 105mm 0.12
the scatter fraction from the values measured for other 93l/oEl2 " " lmm l25mm 0.14'
scanner-collimator combinations using the approximations 931/06-16 " 15.4mm " 3-1mm UOmm 0.15
[6,8], a, a (septa depth).' and asa (inter-septa gap).
?33/06-l2 76cm 12%m l25cm 3-1mm 105mm 0.18
Calculating NEC Rates 933/06-16 " 15Amm m r n " UOmm 025
All NEC curves were calculated using equation 7 with
aci= 0, asi=agfrom Table 1and R, = fR. The randoms rate 831/CXLl2 64cm 12Anrn 145cm lmm l25mm 0.18'
'Exparcd crystal width
(R) was reduced by the fraction of a projection subtendedby
the 2Ocm cylinder (f):(l) 831, f=0.5; (2) 933, f=0.4; (3) 931, Table 1: Relevan! physical parameters of the PET
f=0.33 . For the Alderson brain phantom with a maximum scanners tested. Calculated values, see text.
Results
Figure 1: The mean true and random coincidence count curves for cross and direct slices, and the derived noise equivalent
count (NEC) curves (see text) for the 831/08-12, 933/08-12 and 933/08-16 scanning a 2Ocm diameter phantom.
831/08-12, lmm Septa
12500 12500
loo00 10000
7500 7500
Counts/ Counts/
Second Moo
Second 5000
2500 2500
0 0
(a) 0 0.5 1 1.5 2 2.5 3 3.5 (b) 0 0.5 1 1.5 2 2.5 3 3.5
microCuries/cc microCuries/cc
933/08- 12, 3mm Tapered Septa 933/08- 12. 3mm Tapered Septa
12500 12500
10000 low0
7500 7500
Counts/ Counts/
Second 5wo Second 5ooo
2500 2500
0 0
(c) 0 0.5 1 1.5 2 2.5 3 3.5 (a 0 0.5 1 1.5 2 2.5 3 3.5
microCuries/cc microCuries/cc
10000
7500
Counts/ Counts/
Second Second 5000
2500
0
0 0.5 1 2 2.5 3 3.5 0 05 1 1.5 2 2.5 3 3.5
(e) 1.5
(0
microCuries/cc microCuries/cc
786
12500 12500
0 0
(a) 0 1 2 3 4 5 6 7 (b) 0 1 2 3 4 5 6 7
microCuries/cc rnicroCuries/cc
12500 12500
12500 12500
5ooo 5ooo
2500 ?500
0 0
0 1 2 3 4 5 6 7 0 1 2 3 4 5 6 7
(e)
rnicroCuries/cc
(0 microCuries/cc
Figure 2:The mean true and random coincidence count curves for cross and direct slices, and the derived noise equivalent
count (NEC) curves (see text) for the 931/08-12and 931/08-16scanning a 2Ocm diameter phantom.
The slopes of the early portions of the TCC curves in figures Scanner Septa scadtivity-Trues. pc9AEtivity
1 and 2 are given in Table 2. This trues sensitivity at low Model Width Cross Direct Total Trurs@pCi/cc NE@pCi/cc
activity increases by a factor of 2.0with increasing inter-septa pn/aslz 3.lmm 881r 4- 1Wr 1wr 6.0 WL 4.0
gaps and decreasing ring diameter, a trend which is slightly 9 3lm2 lmm 1 U
l 5Jk 119L 158ir 4.0 9sL 27
decreased if the trues sensitivities are converted to NEC !l3lml6 3 . l m m U S L 7% 1ssL 1% 35 105L 2.2
rates using the asfractions of Table 1. The maximum factor
?
33m
l23.lmm
of 1.8 change in the NEC sensitivity over all scanners at low
2oereglindtr UBr 75k 1ssL m 2 5 75kl.6
activity, corresponds to a 41.8 or 36% SNR advantage based
Ht.apbanhm loslr 6.Ok WL 164k 4.1 118k 3.1
on equation 5. Comparing cross versus direct slices
between scanners, the maximum sensitivity change betweem mm16 s l m m 182k loak 2U7k l2tk 1.7 67k 1.1
the direct slices of the 931/08-12and the cross slices of the 8
31m
l2 lmm l4& 95k 178k 107k l.7 62k 1.1
933/08-16is 3.7 for trues, and 3.3 for NEC’s, a 82% S N R ‘cps/&i/cc @ O.lpCj/cc
advantage. However, if sensitivities are expressed per mm
of axial FWHM resolution [6]the advantage of the 933/08- Table 2 For all PET scanners tested the measured values
16 is reduced by approximately 7/9=0.8 [16]. Expressing for:(l) low activity sensitivity (trues for cross, direct and
sensitivity per mm of axial FWHM resolution makes the total);(2) saturation counting levels of true and NEC count
831/08-12the most sensitive scanner at low activities. rates @ phantom activity levels.
787
The low activity sensitivity for the head phantom compared direct slices is 4.9-7.lkcps and 3.3-4.lkcps respectively,
to the 20cm cylinder in Table 2 is decreased by 20% because corresponding to SWR advantages of 20% and 11%.
of the smaller volume of activity and increased attenuation The efficiency with which increased activity is used at mid
due to the skull. to high activity levels may be estimated by comparing the
Figures 1 and 2 demonstrate the much lower peak average NEC rates for a given scanner’s cross or direct slices
counting (saturation) rates and phantom activity levels at at the peak counting rate activity and at half that activity.
which these rates occur for the NEC curves compared to the The ratio for any pair of cross slice or direct slice rates at
TCC curves. These results are summarized for total trues these two activities, ranges from 1.19-1.23 giving a
and NEC counting rates in Table 2. In contrast to the low consistently small S N R advantage of about 10% for a
activity sensitivitiesthe peak rates and corresponding activity doubling of the activity concentration. Finally consider the
levels tend to decrease with increasing inter-septa gaps and relative cross/direct slice efficiency advantage for each
decreasing ring diameter. individual scanner tested. For the 4 PET scanners with 3mm
In particular consider the change of collimator on the tapered septa the cross/direct slice ratio is between 1.7 and
931/08-12. When moving from the 3mm tapered septa to 1.8 for the activity range up to the peak counting rate for
the lmm septa there is an average 18% increase in low both TCC and NEC curves. For the 931/08-12 and 831/08-
activity, trues sensitivitywhich correspondsto an average 8% 12 with lmm septa the cross/direct ratio is 1.65 and 1.5
SNR advantage based on NEC rates. The randoms rate respectively. These NEC ratios imply S N R advantages for
rises much more sharply in Figure 2(c) for the lmm septa the cross slice over the direct slice of 34% to 22%.
than in Figure 2(a) for the 3mm tapered septa. This causes Figure 3 demonstrates the effect of reducing the
the NEC curves of Figure 2(d) to turn over at a lower count randoms rate when testing with a real, brain sized object as
rate and activity level than in Figure 2(b) so that by opposed to the 20cm diameter cylinder. Compare Figure 3,
approximatelylpCi/cc the lmm septa collimator has lost its (a) and (b) to Figure 1, (c) and (d). The small SNR
8% S N R advantage and the 3mm tapered septa starts to advantage of about 8% for the 20cm cylinder, based on low
perform at a higher NEC rate. activity NEC rates is lost by OSpCi/cc in favor of the brain
Similarly for the 933/08-12 in Figure l(c) and (d) phantom. At almost twice the activity concentration of the
compared to the 933/08-16 in Figure l(e) and (f) the 20cm cylinder the brain phantom has a peak NEC counting
increased inter-septa gap results in a large increase in the advantage of about 1.53 giving a SNR advantage of 24%.
randoms rate at a given activity level. Based on the NEC
counting rates in Figure l(d) and l(f) the average 12% S N R 933/08-12, 3mm Tapered Septa
17500
advantage of the 933/08-16 at low activities has been lost by 15000
lpCi/cc. In addition, if the rates are expressed per mm of 12500
FWHM axial resolution the 933/08-16 has no image S N R
Counts/ 10~00
advantage over the 933/08-12 at any activity level. This
Second 7500
result is in contrast to the 34% sensitivity advantage of the
933/08-16 based on the standard, low activity sensitivityrates 5000