HME_DR_GM85_Datasheet_v1.3
HME_DR_GM85_Datasheet_v1.3
HME_DR_GM85_Datasheet_v1.3
GM85
PIONEERING MOBILE EXPERIENCE
GM85 GM85
Mechanic Collimator
Column Type Collapsible Type 4-axis auto blade
SIZE (W (cm) x L (cm) x H (cm)) 55.5 x 125.8 x 139.6 Light LED
Weight (kg) 349 Auto Filter 0.1, 0.2, 0.3 mmCu
Tube Reach (cm) 79.3 ~ 135.5 (Center of column to Focal spot) Illuminance > 160 lux
Focal Spot Range (cm) 55 ~ 203 Inherent Filteration Over 1.4 mmAl @75kVp
Column Rotation ±315° Rotation Angle -90 ~ 180°
Tube Rotation (arm axis) ±180° ETC Measuring Tape
DAP Integration*
Tube Rotation (tube axis) +90° ~ -30°
* Option
HVG Detector
Power (kW) 32 40 Model S4335-W S4343-W S3025-W
kVp 40 ~ 150 Active Matrix Size (pixel) 2,466 x 3,040 3,036 x 3,040 1,750 x 2,108
mA 10 ~ 500 Effective Area (cm) 34.5 x 42.6 42.5 x 42.6 24.5 x 29.5
Exposure time range From 1ms to 10 seconds Weight (kg) [Typical] 3.1 3.7 1.63
Exposure Parameter Setting 49kV 400mA 49kV 400mA A/D Conversion (bits) 14 / 16 16 16
@Max. Power Rating 64kV 500mA 80kV 500mA DQE (0.1 lp/mm) [Typical] 73% 80% 78%
80kV 400mA 100kV 400mA
100kV 320mA 125kV 320mA Battery Life 160 images for 4 hours
128kV 250mA 150kV 250mA Battery charging time 6 hrs. (tethered), 3 hrs. (battery charger)
150kV 200mA
Tube Driving
Nominal Tube Voltage (kVp) 150 Speed Max. 5.6 km/h
Heat Capacity (HU) 300 kHU Soft Driving O
* Soft Driving : The adaptive driving control with sensing the force of user.
CT-GM85 1.3-161012-EN