Keysight Technologies

Parametric Instrument Accessories


Accessories Guide

Technical Overview
Table of Contents
1. Introduction............................................................................... 3
2. Connecting to Fixture............................................................... 4
2.1. Jumper cables............................................................... 4
2.2. Socket modules............................................................ 5
3. Connecting to SWM................................................................. 6
3.1. GNDU............................................................................. 6
3.2. SMU............................................................................... 7
3.3. MFCMU.......................................................................... 7
4. Connecting to Prober............................................................... 8
4.1. GNDU connections....................................................... 8
4.2. SMU/CMU connections................................................ 9
5. Accessories..............................................................................13
5.1. CMU accessories.........................................................13
5.2. B1500A peripherals.....................................................15
5.3. Connector type............................................................16
5.4. Cables...........................................................................18
5.5. Connector plates........................................................ 20
5.6. Adapters & connectors............................................... 24
1. Introduction
This guide provides information on how to connect a prober to Keysight Technologies,
Inc. semiconductor parametric measurement instruments.

This guide also provides information on accessories for a variety of Keysight


Technologies semiconductor parametric measurement instruments, such as:
Keysight B1500A—Semiconductor Device Analyzer
Keysight 4155C/4156C—Semiconductor Parameter Analyzer
Keysight E5270B—8 slot Precision Measurement Mainframe
Keysight E5260A—8 slot High Speed Measurement Mainframe
Keysight B2200A—fA Leakage Switch Mainframe
Keysight B2201A—14ch Low Leakage Switch Mainframe
Keysight E5250A—Low Leakage Switch Mainframe

Figure 1.1. B1500A with Cascade prober Figure 1.2. B1500A with Vector prober

Figure 1.3, B1500A with Suss prober Figure 1.4. ASUs on the probe station
04 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

2. Connecting to Fixture
The 16442A and 16442B test fixtures have 2.1. Jumper cables
6 triaxial input ports. You can connect
force ports of SMU to them by using Table 2.1. Jumper cables
Triaxial Cable 16494A as shown in Mini banana – Pin plug
Figure 2.1.
Approximately 200 mm
11 mm 7 mm

2 mm

1 mm
Triaxial cable
16494A-001 1.5 m
or As option of 16442A/16442B
16494A-002 3.0 m 16442A-811/16442B-811 for Black (2ea), Blue (2ea), Red (2ea)
Fixture Parts number
S 16442A
M Force 16442-61600 for Black (1ea)
SMU or
U 16442B 16442-61601 for Red (1ea)
1-6
Sense 16442-61602 for Blue (1ea)

Figure 2.1. Connecting SMUs to fixture Pin plug – Pin plug


Approximately 200 mm
7 mm 7 mm

1 mm
1 mm

As option of 16442A/16442B
16442A-812/16442B-812 for Black (2ea), Blue (2ea), Red (2ea)
Parts number
16442-61603 for Black (1ea)
16442-61604 for Red (1ea)
16442-61605 for Blue (1ea)

Mini banana – Mini clip


Approximately 200 mm
11 mm
40 mm
2 mm

As option of 16442A/16442B
16442A-813/16442B-813 for Black (2ea), Blue (2ea), Red (2ea)
Parts number
16442-61606 for Black (1ea)
16442-61607 for Red (1ea)
16442-61608 for Blue (1ea)

Mini banana – Mini banana

Approximately 200 mm
11 mm
2 mm

As option of 16442A/16442B
16442A-814/16442B-814 for Black (2ea), Blue (2ea), Red (2ea)
Parts number
16442-61609 for Black (1ea)
16442-61610 for Red (1ea)
16442-61611 for Blue (1ea)
05 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

2.2. Socket modules


Table 2.2.

Socket Modules

N1254A-204 0.1 inch

43 mm
This socket module is for measurement
16442A-801/16442B-801 0.1 inch of virtually any device—component, DIPIC,

86 mm
or small-scale circuit.

N1254A-205 0.075 inch

43 mm
16442A-802/16442B-802 0.075 inch 84 mm

N1254A-206 0.05 inch

1.8 mm
1 mm
8 mm 17.2 mm
16442A-803/16442B-803 0.05 inch
43 mm

N1254A-200 This socket module has two


transistor sockets for measurement

86 mm
5 mm 3 mm of three lead or four lead bipolar
transistors and FETs.
5 mm

3 mm
43 mm

Maximum voltage: 200 V

20 mm
16442A-821/16442B-821
84 mm

N1254A-201
43 mm

This socket module has a


28-pin dual-in-line socket.
86 mm

28 15 Maximum voltage: 230 V


16.54 mm
43 mm

16442A-823/16442B-823
1
2.54 mm
84 mm

N1254A-202
43 mm

This socket module has an


18-pin dual-in-line socket.
86 mm

1 18
Maximum voltage: 230 V
43 mm

2.54 mm

16442A-822/16442B-822 9 mm
84 mm

N1254A-207
43 mm

This insulation board has minimal leakage


current, and so is good for extremely low
86 mm

current measurements.
43 mm

16442A-800/16442B-800
84 mm
06 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

3. Connecting to SWM
3.1. GNDU
An SMU has FORCE (SENSE), GUARD, Ground unit cable
and COMMON, but a GNDU has SENSE, 16493L-001 1.5 m SWM
FORCE, and COMMON as shown in Figure or B2200A
16493L-002 3.0 m E5250A
3.1. You need some sort of GNDU to Kelvin G
N
adapter when connecting the GNDU to D
standard triaxial connections. U
Triaxial – triaxial
adapter
Common N1254A-107
Figure 3.2. Examples of a GNDU non-Kelvin
connection to SWM
Sense

Force Table 3.1. Parts information

Description Qty Part number Product


Figure 3.1. GNDU terminal
Ground unit cable – 1.5 m (Max 4.2 A) 1 — 16493L-001
Ground unit cable – 3.0 m (Max 4.2 A) 1 — 16493L-002
CAUTION Adapter for GNDU (triaxial (m) – triaxial (f)) 1 1250-2654 N1254A-107
You should use the Keysight 16493L GNDU
cable to connect the GNDU to the SWM;
do not use a standard triaxial cable. The
Kelvin triaxial cable
GNDU cable can handle the maximum 16493K-001 1.5 m
GNDU current of 4.2 A, but standard tri- or Ground unit cable
16493K-002 3.0 m 16493N-001 1.5 m
axial cables are only rated to 1 A. SWM SWM
Force Force B2200A Force B2200A
G G
(Non-Kelvin connection) N
D
N
D
Sense Sense Sense
U U
Connect the triaxial connector using an
adapter for GNDU (N1254A-107) as shown
in Figure 3.2. For this part, see Table 5.10
GNDU to Kelvin adapter
in Section 5.6 Adapters. For parts infor- N1254A-100
mation, see Table 3.1.
Figure 3.3. Examples of a GNDU Kelvin connection to SWM

Kelvin connection
Table 3.2. Parts information
Connect the dual triaxial connector using
a GNDU to Kelvin Adapter (N1254A-100) Description Qty Part number Product
as shown in Figure 3.3. For this part, see Kelvin triaxial cable – 1.5 m (Max 3.0 A) 1 — 16493K-001
Table 5.10 in Section 5.6 Adapters. For Kelvin triaxial cable – 3.0 m (Max 2.6 A) 1 — 16493K-001
parts information, see Table 3.2. Maxi-
Ground unit cable – 1.5 m 1 — 16493N-001
mum current is restricted when the Kelvin
triaxial cable is used. See Table 3.2. GNDU to Kelvin adapter 1 — N1254A-100
07 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

3.2. SMU
An SMU terminal is shown in Figure 3.4.
Kelvin triaxial cable
16494B-003 0.8 m
Kelvin triaxial cable or
Common 16493K-001 1.5 m 16494B-001 1.5 m
or or
16493K-002 3.0 m 16494B-002 3.0 m
SWM SWM
Force Force Force B2200A Force Force E5250A
(Sense) S or S
M B2201A M
Guard U Sense Sense U Sense Sense

Figure 3.4. SMU terminal

Figure 3.5. Examples of a SMU Kelvin connection to SWM


Kelvin connection
When Kelvin connection is used, connect
a Keysight 16493K Kelvin triaxial cable for
B2200A/B2201A as shown in Figure 3.5. Triaxial cable
16494A-001 1.5 m
For E5250A, connect a Keysight 16494B or
Kelvin Triaxial Cable as shown in Figure 16494A-002 3.0 m
SWM
3.5. You can also use two Keysight 16494A E5250A
triaxial cables in place of one Kelvin Tri- S
M Force
axial Cable. U

Connect the FORCE and SENSE lines to Sense

ports separately.
Figure 3.6. Examples of a SMU non-Kelvin connection
to SWM

Non-Kelvin connection
When non-Kelvin connection is used,
connect the Keysight 16494A triaxial cable SWM
Hcur CMU input cable
between the SMU and SWM as shown in 16494F-001 2.0 m B2200A
or
Figure 3.6. B2201A
M Hpot CMH
F
C
3.3. MFCMU M
U Lpot CML
MFCMU of B1500A is connected to CMH
and CML port of B2200A/ B2201A by Lcur
using CMU input cable 16494F-001 as
shown in Figure 3.7.
Figure 3.7. Examples of a MFCMU connection to SWM
08 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

4. Connecting to Prober
4.1. GNDU connections
An SMU has FORCE (SENSE), GUARD, Shielding box
and COMMON, but a GNDU has SENSE,
FORCE, and COMMON as shown in Figure Triaxial cable
Adapter for GNDU
16493L-001 1.5 m
4.1. You need some sort of GNDU to Kelvin or
N1254A-107
adapter when connecting the GNDU to 16493L-002 3.0 m
G Probe cable (triaxial) Prober
standard triaxial connections.
N
D
U
Common
Connector plate
Sense
DUT
Force

Figure 4.1. GNDU terminal Figure 4.2. Examples of a GNDU non-Kelvin connection

Table 4.1. Parts information


CAUTION
Description Qty Part number Product
You should use the Keysight 16493L
Ground unit cable – 1.5 m (Max 4.2A) 1 — 16493L-001
GNDU cable to connect the GNDU to the
Ground unit cable – 3.0 m (Max 4.2A) 1 — 16493L-002
connector plate; do not use a standard
triaxial cable. The GNDU cable can handle Adapter for GNDU (triaxial (m) – triaxial (f)) 1 1250-2654 N1254A-107
the maximum GNDU current of 4.2 A, but
standard triaxial cables are only rated to
1 A. Shielding box
Kelvin triaxial cable
16493K-001 1.5 m
Non-Kelvin connection or
16493K-002 3.0 m
Connect the triaxial connector using an Probe cable (triaxial)
Force Force
adapter for GNDU (N1254A-107) as shown G Prober
in Figure 4.2. For this part, see Table 5.10 N
D Probe cable (triaxial)
in Section 5.6 Adapters. For parts infor- U Sense Sense

mation, see Table 4.1.

GNDU to Connector plate DUT


Kelvin adapter
Kelvin connection N1254A-100

Connect the dual triaxial connector using Figure 4.3. Examples of a GNDU Kelvin connection

a GNDU to Kelvin Adapter (N1254A-100)


as shown in Figure 4.3. For this part, see Table 4.2. Parts information
Table 5.10 in Section 5.6 Adapters. For
Description Qty Part number Product
parts information, see Table 4.2. Maxi-
mum current is restricted when the Kelvin Kelvin triaxial cable – 1.5 m (Max 3.0 A) 1 — 16493K-001
triaxial cable is used. See Table 4.2. Kelvin triaxial cable – 3.0 m (Max 2.6 A) 1 — 16493K-002
GNDU to Kelvin adapter 1 — N1254A-100
09 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

4.2. SMU/CMU connections


An SMU terminal is shown in Figure 4.4. Shielding box
Kelvin triaxial cable
16493K-001 1.5 m
Common or
16493K-002 3.0 m
Force Force
Probe cable (triaxial)
Force Prober
(Sense) S
M Probe cable (triaxial)
Guard U Sense Sense

Figure 4.4. SMU terminal


Connector plate
DUT

Kelvin connection
Figure 4.5. Examples of a SMU Kelvin connection
These instructions apply when all
connections are Kelvin. Two probes must
contact the wafer in this connection. Table 4.3. Parts information

Description Qty Part number Product


Connect a Keysight 16493K Kelvin
Kelvin triaxial cable – 1.5 m (Max 3.0 A) 1 — 16493K-001
triaxial cable (or two Keysight 16494A
triaxial cables) between the SMU and the Kelvin triaxial cable – 3.0 m (Max 2.6 A) 1 — 16493K-002
connector plate. Connect the FORCE and
SENSE lines to probes separately. Connect
the triaxial connector on the probe Shielding box
cable as shown in Figure 4.5. To prevent Triaxial cable
16494A-001 1.5 m
oscillations, use short cables. For parts or
information, see Table 4.3. 16494A-002 3.0 m
Probe cable (triaxial) Prober
S
Non-Kelvin connection M
U Force
These instructions apply when all
connections are non-Kelvin. Connect the Sense
Keysight 16494A triaxial cables between Connector plate DUT
the SMU and the Connector plate.

Connect the triaxial connector on the Figure 4.6. Examples of a SMU non-Kelvin connection
probe cable as shown in Figure 4.6. For
parts information, see Table 4.4. Table 4.4. Parts information

Description Qty Part number Product


Triaxial cable – 1.5 m 1 — 16494A-001
Triaxial cable – 3.0 m 1 — 16494A-002
10 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

Kelvin to non-Kelvin connection


These instructions apply when the Shielding box
connections up to the Connector Plate Kelvin triaxial cable
are Kelvin, but the probe is a non-Kelvin 16493K-001 1.5 m Probe cable (triaxial) Prober
or
connection. 16493K-002 3.0 m
Force Force
Connect a Keysight 16493K Kelvin Tee triaxial BNC
S
triaxial cable (or two Keysight 16494A M 1250-1551
triaxial cables) between the SMU and the U Sense Sense DUT
Connector plate.

Connector plate Triaxial cable


Connect the triaxial connector on the
16494A-004 0.4 m
probe cable as shown in Figure 4.7.

Figure 4.7. Examples of a SMU Kelvin to non-Kelvin connection


Connect the FORCE and SENSE lines on
this side of the probe input terminal. A
Tee Triaxial BNC adapter can be used. For Table 4.5. Parts information
parts information, see Table 4.5.
Description Qty Part number Product

Note: When connecting the cable and Kelvin triaxial cable – 1.5 m 1 — 16493K-001
tee connector as shown in Figure 4.7, Kelvin triaxial cable – 3.0 m 1 — 16493K-002
space restrictions make an adjacent Triaxial cable – 0.4 m 1 — 16494A-004
connector unusable. Since three Triaxial tee-connector 1 1250-1551 —
connectors are necessary for a Kelvin
connection a connector plate must be
used. Select the correct connector plate
for your application. To make two Kelvin
connections, use a Keysight 16495H. To
make three Kelvin connections, use a
Keysight 16495J.
11 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

ASU connection
Each ASU has one Dsub control cable, one
Triaxial cable and
triaxial cable, and (optionally) two coaxial CMU cable Dsub cable
cables connected to it. The cables can be N1300A-001 1.5 m 16493M-001 1.5 m Shielding box
or or
connected to the ASU inside the shielding N1300A-002 3.0 m 16493M-002 3.0 m
box through the 16495K plate.
H
The ASU outputs should be connected to R Probe cable (triaxial)
S Dsub Dsub
the DUT by using triaxial cables as shown M
U A
in Figure 4.8. S
Force Force
U
Probe cable (triaxial)
When making IV measurements with the Hcur

SMUs, all measurements are Kelvin. The CMU-cur/AUX in


Wire furnished with ASU
SENSE line information is fed back to the M Hpot
(both pin terminals Prober
F
SMU via the Dsub cable. C CMU-pot
M
Lpot
U Probe cable (triaxial)
When making CV measurements, the CMU-pot

four-terminal pair (4TP) connections are Lcur


correctly terminated inside of the ASUs. CMU-cur/AUX in
A
S DUT
The ASUs also allow for a current return H U Probe cable (triaxial)
path between the outer conductors R
S Force Force
to stabilize the series inductance and M
U
improve accuracy. You must connect the Dsub Dsub
furnished wire between the two ASUs to
enable this feature. Triaxial cable and Connector plate
Dsub cable 16495K
16493M-001 1.5 m
or
16493M-002 3.0 m

Figure 4.8. Examples of an ASU connection

Table 4.5. Parts information

Description Qty Part number Product


Triaxial cable & Dsub cable for ASU – 1.5 m 2 — 16493M-001
Triaxial cable & Dsub cable for ASU – 3.0 m 2 — 16493M-002
CMU cable for B1500A – 1.5 m 1 — N1300A-001
CMU cable for B1500A – 3.0 m 1 — N1300A-002
Connector plate 1 — 16495K
12 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

SCUU positioner-based
probing mode
The SCUU should be connected to the
SCUU cable Shielding box
DUT by using triaxial cables as shown in N1301A-002 3.0 m
Figure 4.9.

When making IV measurements using the Sense


SMUs, the outputs of the SCUU function S
M SCUU
exactly the same as the Kelvin outputs of U N1301A-100
Force
an SMU (both FORCE and SENSE lines are
available). Probe cable (triaxial)
Hcur

Note: In Figure 4.9 the SENSE lines are not Force


Prober

used. M H pot
F
C Sense Probe cable (triaxial)
When making CV measurements using the M
Lpot
U
MFCMU, only the FORCE outputs of the Force
SCUU are used. SCUU FORCE output1 DUT
Lcur
corresponds to CMH, and SCUU FORCE
Sense
output2 corresponds to CML.
Force
S
The SCUU can be fixed on the prober by M
U
using SCUU Magnet Stand (N1301A-110). Sense

Connector plate
16495K

Figure 4.9. Examples of a SCUU remote docking mode

Table 4.9. Parts information

Description Qty Part number Product


SMU CMU unify unit (SCUU) 1 — N1301A-100
SMU CMU unify unit cable (3.0 m) 1 — N1301A-102
Connector plate 1 — 16495K
13 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

5. Accessories
5.1. CMU accessories
B1500A has the multi frequency
capacitance measurement unit (MFCMU).
There are two types of CMU connection as
follows.
–– Standard four terminal pair (4TP)
connection
–– SCUU positioner-based probing
mode

According to your situation, you can select Figure 5.1. Standard four terminal pair (4TP)
the type of connection.

Figure 5.2. SCUU positioner-based probing


14 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

Table 5.1. CMU accessories overview

N1300A-001/N1300A-002 16494F-001

N1301A-100 N1301A-100 N1301A-100

N1301A-102 SCUU w/SCUU Cable, GSWU

N1301A-110 SCUU w/SCUU magnet N1301A-200

N1301A-201 Pin cable for GSWU GSWU w/GSWU cable


15 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

Table 5.2. CMU accessories item list

Item Description OP instruction


N1300A CMU cable for B1500A
N1300A-001 CMU cable for B1500A (1.5 m)
N1300A-002 CMU cable for B1500A (3.0 m)
N1301A CMU accessories for B1500
N1301A-100 SMU CMU unify unit (SCUU)
N1301A-102 SMU CMU unify unit cable (3 m)
N1301A-110 SCUU magnet stand To fix SCUU on prober top etc.
N1301A-200 Guard switch unit (GSWU) Need to be specified with N1301A-201 or N1301A-202.
N1301A-201 Guard switch unit cable (1 m) Please specify N1301A-201 or N1302A-202 when N1301A-200 is ordered.
N1301A-202 Guard switch unit cable (3 m)
16494F CMU input cable
16494F-001 CMU input cable (2 m)

5.2. B1500A peripherals


Table 5.3. B1500A peripherals overview

16444A-001 16444A-002 16444A-001 16444A-003

Table 5.4. B1500A peripherals item list

Item Description OP instruction


16444A B1500A accessories
16444A-001 USB keyboard Keyboard includes mouse pad and mouse pointer
16444A-002 USB mouse USB Mouse
16444A-003 Stylus pen For use with the B1500A touch screen
16 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

5.3. Connector type


There are two main styles of Kelvin Table 5.5. Overview of instrument connector types
SMU connections. When connecting
B1500A Type A
instruments together, you need to make
sure that you have the correct type of
Kelvin triaxial cables.

Type A connectors have two screws, one


guide pin, and two triaxial terminals. Type
B connectors have one screw and two
triaxial terminals. These connectors are
also different in their width and thickness.

4156C+41501B Type A

4155C Single triaxial

E5260A/E5270B Type A

B2200A/B2201A inputs Type A


17 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

Table 5.5. Overview of instrument connector types, (continued)

B2200A/B2201A outputs Type B

E5250A inputs Type B

E5250A outputs (E5252A cards) Type B

16442A/B Type A

16493K/16494B/16493N cables Type A

16494B cable Type B


18 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

5.4. Cables
Several possibilities exist for connecting Table 5.6. SMU cables overview
the Source Monitor Units (SMUs) to the
16493K 16494B
DUT. The great variety of available cables
can be categorized into mainly three
types:

Triaxial cables have three layers or


leads: a center conductor for the signal,
a middle conductor that is actively
Kelvin triaxial cable Kelvin triaxial cable
driven to be at the same potential as the
center conductor to eliminate leakage
Connector type Type A Type A Type B
currents, and a grounded (Common) outer
conductor. SMU B1500A OK OK NG
E526X OK OK NG
Kelvin triaxial cables are similar to triaxial E527X OK OK NG
cables, except that both the Force and 415X OK OK NG
Sense lines from the SMU run through 41501 OK OK NG
the same cable. The Force and Sense SWM input B220XA OK OK NG
lines share a common driven guard and E5250A NG NG OK
a common outer shield. Kelvin triaxial SWM output B220XA NG NG OK
cables typically give better low-current E5250A NG NG OK
measurement performance than you Length option 0.4 m N/A N/A
would get from using two separate triaxial 0.8 m N/A 16494B-003
cables for the Force and Sense lines. 1.5 m 16493K-001 16494B-001
3.0 m 16493K-002 16494B-002
The GNDU cable is a special triaxial cable 4.0 m N/A N/A
designed to handle the large currents (up
to 4.2 A) that the ground unit can sink. In
the GNDU cable the center conductor is
SENSE, the middle conductor is FORCE,
and the outer conductor is Common.
19 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

Table 5.6. SMU cables overview, (continued)

16493M 16494A 16493J

Triaxial & Dsub cable For ASU Triaxial cable Interlock cable

Connector type — Triax —


SMU B1500A OK OK OK
E526X N/A OK OK
E527X OK OK OK
415X N/A OK OK
41501 N/A OK N/A
SWM input B220XA N/A OK N/A
E5250A N/A OK N/A
SWM output B220XA N/A OK N/A
E5250A N/A OK N/A
Length option 0.4 m N/A 16494A-004 N/A
0.8 m N/A 16494A-003 N/A
1.5 m 16493M-001 16494A-001 16493J-001
3.0 m 16493M-002 16494A-002 16493J-002
4.0 m N/A 16494A-005 N/A

Table 5.7. GNDU cables overview

16493N 16493L

Ground unit cable Ground unit cable

Connector type Type A Triaxial Triaxial


GNDU B1500A N/A OK OK
E526X N/A OK OK
E527X N/A OK OK
415X N/A N/A N/A
41501 N/A OK OK
SWM input B220XA OK OK OK
E5250A NG OK OK
SWM output B220XA NG OK OK
E5250A NG OK OK
Length option 1.5 m 16493N-001 16493L-001
3.0 m N/A 16493L-002
20 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

5.5. Connector plates


The following connector plates work Table 5.8. Parts information to fix connector plates
with a variety of Keysight semiconductor Description Qty Part number Product
parametric measurement instruments,
Screw 1 0515-0986 —
such as the B1500, B2200/B2201, E5250,
Spring washer 1 2190-0584 —
E5260/E5270, 4155/4156 and 41501.
Terminal washer 1 3050-0891 —
Prober vendor also provides connector Nut 1 0535-0031 —
plates. If you need the information of those
connector plates, please kindly contact
each prober vendor. Table 5.9. Connector plates

If you need parts to fix a connector plate, 16495F—Half size connector plate for matrix
see Table 5.8. Order the quantity you need. 16495F-001 has 12 triaxial through connectors (female to female), an Intlk
(The number of holes in a connector plate connector, and a GNDU connector (triaxial through, female to female). The
is four.) back of the Intlk connector is designed for soldering.

16495F-002 has 12 triaxial connectors, an Intlk connector, and a GNDU


connector. The back of each connector is designed for soldering.

16495G—Full size connector plate for matrix


16495G-001 has 24 triaxial through connectors (female to female), an Intlk
connector, and a GNDU connector (triaxial through, female to female). The
back of the Intlk connector is designed for soldering.

16495G-002 has 24 triaxial connectors, an Intlk connector, and a GNDU


connector. The back of each connector is designed for soldering.

16495H—Half size connector plate for analyzer


16495H-001 has 6 triaxial through connectors (female to female), 6 BNC
through connectors (female to female), an Intlk connector, and a GNDU
connector (triaxial through, female to female). The back of the Intlk connector
is designed for soldering.

16495H-002 has 6 triaxial connectors, 6 BNC connectors, an Intlk connector,


and a GNDU connector. The back of each connector is designed for soldering.

16495J—Half size connector plate for analyzer


16495J-001 has 8 triaxial through connectors (female to female), 4 BNC
through connectors (female to female), an Intlk connector, and a GNDU
connector (triaxial through, female to female). The back of the Intlk connector
is designed for soldering.

16495J-002 has 8 triaxial connectors, 4 BNC connectors, an Intlk connector,


and a GNDU connector. The back of each connector is designed for soldering.

16495K—Half size connector plate with universal cable holder


16495K-001 does not have any connectors. This plate provides the cable hole and the cover with
rubber used to block the light from the cable hole. This plate will be used with the Keysight E5288A
Atto Sense/Switch Unit (ASU) or the Keysight N1301A-200 Guard Switch Unit (GSWU) that will be
installed in the shielding box. The cable hole is to pass the cables from the ASU or GSWU to the
instrument such as the Keysight B1500A, E5270B, and so on. And the cables will be fixed to the
shielding box by using the cover with rubber that will close the opening of the cable hole.
21 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

A. Front view

INPUT

Common 1 1

Force 3 4

(sense) Triaxial connector


(for SMU)
Guard 5 6

7 8

Signal line
9 10

Common Coaxial connector


(for MFCMU) 11 12

Common

GN DU Intlk !

Sense Triaxial connector


(for GNDU)
Force

INTLK terminal

B. Rear view

Ground terminal
Ground terminal
Triaxial Triaxial
Common terminal
connector connector Common terminal

The difference between


the Keysight 16495F, H and J
is the number of the triaxial
and coaxial connectors.

Coaxial Coaxial
connector connector

INTLK terminal INTLK terminal


Triaxial Triaxial
connector connector
(for GNDU) (for GNDU)
Keysight 16495H-001 Keysight 16495H-002

Figure 5.3. Front view and rear view of connector plate


22 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

115
105 5
50
5

12
4 - Ø 3.2
INPUT

1 1

3 4

5 6

210 220
196 Metal cover 7 8

9 10

11 12

G ND U Intlk !

Unit: mm
Metal cover

Keysight 16495H-002
(The difference between the Keysight 16495 F, H and J
is the number of the triaxial and coaxial connectors.)

Note: Metal cover is for preventing electrical shock.


Metal cover is furnished with 16495 F/G/H/J-002 only. 91 12

Figure 5.4. Dimensions of the connector plate (Keysight 16495F/H/J)

Triaxial Coaxial Triaxial Coaxial

A. Connector type (16495F/G/H/J-001) B. Solder type (16495F/G/H/J-002)

Figure 5.5. Connection type of connector plate (rear)


23 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

230
220 5
50
5

12
4 - Ø 3.2

196 210 220

Unit: mm

206 12

Figure 5.6. Dimensions of the connector plate (Keysight 16495G)


24 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

5.6. Adapters and connectors


If you need to make a connection between Table 5.10. Parts information
a triaxial connector and a coaxial connec- Item Description Qty Part number Illustration
tor, use a triaxial-coaxial adapter.

Table 5.10 shows the adapter informa- Force


tion provided by Keysight. If you need an GNDU to f
N1254A-100 1 —
adapter which is not provided, contact a Kelvin adapter Sense
m
local parts vendor. Common
f
WARNING Triaxial (m) –
N1254A-101 1 1250-2648
To avoid injury, always be sure to use a Coaxial (f)
shielding box and Interlock when you make Triaxial (f) –
a measurement. N1254A-102 1 1250-2649
Coaxial (m)

Do not use the adapters in Table 5.10 Triaxial (m) –


N1254A-103 1 1250-2650
when the GUARD line connects to the Coaxial (f)
outer conductor of the connector or Triaxial (f) –
N1254A-104 1 1250-2651
adapter. Because they connect GUARD Coaxial (m)
line and outer conductor of the connector Triaxial (f) – 1250-2652
or adapter, the potential of the GUARD N1254A-105 1
Coaxial (m)
terminal is equal to the output voltage
Triaxial (m) –
and there is a potential for high voltage N1254A-106 1 1250-2653
Coaxial (f)
electrical shock at the adapter and
connector.
Triaxial (m) –
N1254A-107 1 1250-2654
Triaxial (f)
25 | Keysight | Parametric Instrument Accessories Guide – Technical Overview

Keysight B2900A Series Precision For more information on Keysight


Source/Measure Unit Technologies’ products, applications or
www.keysight.com/find/b2900a services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus

Americas
Keysight B1500A Semiconductor Canada (877) 894 4414
Brazil 55 11 3351 7010
Device Analyzer
Mexico 001 800 254 2440
www.keysight.com/find/b1500a United States (800) 829 4444

Asia Pacific
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Keysight B1505A Power Device Japan 0120 (421) 345
Analyzer/Curve Tracer Korea 080 769 0800
www.keysight.com/find/b1505a Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 6375 8100
myKeysight Europe & Middle East
www.keysight.com/find/mykeysight Austria 0800 001122
A personalized view into the information most relevant to you. Belgium 0800 58580
Finland 0800 523252
Three-Year Warranty France 0805 980333
www.keysight.com/find/ThreeYearWarranty Germany 0800 6270999
Keysight’s commitment to superior product quality and lower total cost Ireland 1800 832700
of ownership. The only test and measurement company with three-year Israel 1 809 343051
warranty standard on all instruments, worldwide. Italy 800 599100
Luxembourg +32 800 58580
Keysight Assurance Plans Netherlands 0800 0233200
www.keysight.com/find/AssurancePlans Russia 8800 5009286
Up to five years of protection and no budgetary surprises to ensure your Spain 800 000154
Sweden 0200 882255
instruments are operating to specification so you can rely on accurate
Switzerland 0800 805353
measurements.
Opt. 1 (DE)
www.keysight.com/go/quality Opt. 2 (FR)
Opt. 3 (IT)
Keysight Technologies, Inc.
United Kingdom 0800 0260637
DEKRA Certified ISO 9001:2008
Quality Management System
For other unlisted countries:
Keysight Channel Partners www.keysight.com/find/contactus
www.keysight.com/find/channelpartners (BP-09-23-14)
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.

www.keysight.com/find/b1500a
www.keysight.com/find/b1505a
www.keysight.com/find/b2900a

This information is subject to change without notice.


© Keysight Technologies, 2013 - 2014
Published in USA, August 3, 2014
5989-5695EN
www.keysight.com