4900 Multigas Installation and Operator Manual Rev B04
4900 Multigas Installation and Operator Manual Rev B04
4900 Multigas Installation and Operator Manual Rev B04
Continued safe and reliable operation of this equipment is conditional on all installation, operation and
maintenance procedures being carried out in accordance with the appropriate manuals, by personnel having
appropriate qualifications, experience and training. Failure to observe the requirements of the manual may result
in the user being held responsible for the consequences and may invalidate any warranty. Servomex accepts no
liability for unauthorized modifications to Servomex supplied equipment.
Servomex has paid particular attention to Health and Safety throughout this manual. Where special precautions
need to be taken due to the nature of the equipment or product, an appropriate safety icon and warning
message is shown. Special attention should be made to section 2 – Safety, where all such messages are
summarized.
This document, and all specifications and drawings within the document, are the property of Servomex, unless
agreed otherwise by contract. They must not be reproduced, copied or transmitted in any form, or by any means,
or used as the basis for the manufacture or sale of apparatus, programs or services without permission.
In line with our continuous policy of research and development, we reserve the right to amend models and
specifications without prior notice therefore the information in this document is subject to change without notice
and does not represent a commitment on the part of Servomex Group Limited. This handbook is accurate at the
date of printing but will be superseded and should be disregarded if specifications or appearance are changed.
Check the internet for updates to the manual. The latest revision of this manual is available in Adobe Acrobat
format at www.servomex.com.
The customer agrees that in accepting and using this instrument Servomex’s liability arising from or in any way
connected with this instrument shall be limited exclusively to performing a new calibration or replacement or
repair of the instrument or transducer, at Servomex’s sole option, as covered by Servomex’s warranty. In no
event shall Servomex be liable for any incidental, consequential or special damages of any kind or nature
whatsoever, including but not limited to lost profits arising from or in any way connected with this instrument or
items hereunder, whether alleged to arise from breach of contract, express or implied warranty, or in tort,
including without limitation, negligence, failure to warn or strict liability.
Servomex is a registered trademark of Servomex Group Inc. The use of all trademarks in this document is
acknowledged.
Revision: 890000M/001/B04
Installation: To take commissioning to the point where the analyzer is powered and
operational. The installer is advised to read this manual completely
before commencing installation.
Configuration: How to set up the clock, passwords, alarm levels, analogue outputs,
relays and other parameters.
Review: How to display analogue output / input settings, relay allocation, alarms,
faults and analyzer identity without changing the analyzer settings.
This symbol warns of specific hazards due to high voltages which, if not
taken into account, may result in personal injury or death.
This symbol highlights where you must take special care to ensure the
analyzer or other equipment or property is not damaged.
Hint: Hints give helpful tips and highlight information which is useful for you to be
aware of, for example, specific operating conditions.
Product overview
The 4900 Multigas Analzyer is designed to meet the needs of regulatory emissions
monitoring and providing feedback / feedforward control purposes in industrial and
combustion related processes.
The analyzer is a highly customizable, very low maintenance instrument that can monitor
up to four gases at one time. It uses ultra-stable paramagnetic and non-dispersive infrared
(NDIR) with gas filter correlation (GFX) technologies plus single beam single wavelength
infrared (SBSW IR).
General description
The 4900 Multigas analyzer is simple to operate, with an intuitive user interface that will
display data from external sources. The chassis accepts up to four gas modules. It
provides power, gas connections and other support functions to the gas sensor modules
and processes their outputs to provide the sample gas concentrations. Gas measurements
are shown on the analyzer display and at the same time are sent out of the analyzer to
other devices using serial, milliamp (mA), voltage or digital communications protocols.
The analyzer supports up to four external analog input signals that can then be displayed
on the screen as measurement signals, and output through the analog and/or the serial
outputs or accessed using Modbus or PROFIBUS protocols. These external input signals
can be recorded, used to activate relays, or trigger auto-calibration / validation routines or
low / high alarms.
Included with each analyzer ordered:
• 4 Relays contacts provided as standard (up to 8 relays per option board, 32 relays max
with 4 option boards)
Included with each transducer ordered:
Note: When the optional auto-calibration option is configured, the manual calibration
process will use the auto-calibration valves to select the calibration gases as
required.
Product identification
Figure 1-1: Standard 4900 Multigas Gas Analyzer configured for rack-mount
Note: The AquaExact 1688 port is not currently available for use in the 4900 Multigas.
Figure 1-3: The Extended Chassis version of the 4900 Multigas Gas Analyzer
Table 1-1 provides a general overview of the connectors on the back of the analyzer.
Connections for optional features will be inactive unless the optional feature has been
purchased.
ID Description ID Description
J19 Option board 1&2 4-20 mA J36 Option board 3&4 4-20mA
inputs inputs
J20 4 – 20 mA output / voltage J37 4 – 20 mA output / analog
output (I1± / V1±) voltage output (I3± / V3±)
J21 4 – 20 mA output / voltage J38 4 – 20 mA output voltage
output (I2± / V2±) output (I4± / V4±)
PROFIBUS Profibus connector RS-232 RS232
Sample requirements
For best performance the flow supplied to the analyzer should be kept at a constant value
and the analyzer must be freely vented to atmosphere, for both process sampling and for
calibration gas input.
Flow Rate: Nominal 1500 mL/min (Min 500 mL/min, Max 2000 mL/min)
Temperature: 5 to 45°C / 41 to 113°F
Dew point: 5°C / 9°F below minimum ambient
Condition: Oil free, non - condensing, filtered to 1m
Vent: Connect the outlet of the analyzer to a separate atmospheric
vent, free from any back-pressure
Warm up Time: Typically, 24 hours from cold start to 20°C / 68°F
High sensitivity measurements may take longer to warm
up.
Make sure that if pressurized gases are used to keep the pressure below
8psig (55kPa g).
General warnings
Before you attempt to install, commission or use the 4900 Multigas
analyzer, read this manual carefully.
Do not connect the 4900 Multigas analyzer to a power source until all
relays, input/ output signals and plumbing connections are made.
The 4900 Multigas analyzer is only suitable for installation in safe areas.
Make sure that all floors or platforms where you install the 4900 Multigas
analyzer are large enough for you to move freely and to change position.
The 4900 Multigas is not suitable for use with corrosive samples.
Electrical warnings
Always observe the appropriate electrical safety codes and regulations.
Make sure that the electrical installation of the analyzer conforms with all
applicable local and national electrical safety requirements.
Make sure the analyzer is provided with a sound earth connection via the
electrical supply plug.
Make sure the electrical supply coupler or plug is easily accessible for
disconnection from the electrical supply.
All signal and electrical supply cables must be rated for temperatures of
70°C or higher.
Make sure that the cables that you connect to the analyzer are routed so
they do not present a trip hazard.
Route AC power wiring as far from the analyzer and its wiring as
possible.
The 4900 Multigas analyzer includes the following external markings on the rear panel and
correspond to:
Do not connect any cables carrying mains voltage or cables that have
inadequate insulation between line and mains to any of the I/O
connectors.
Follow the appropriate safety instructions and be aware of any warnings about potential
hazards.
The 4900 Multigas analyzer is only suitable for installation in safe areas.
Follow the instructions in this section to safely install the 4900 Multigas
analyzer.
Make sure that all floors or platforms where you install the 4900 Multigas
analyzer are large enough for you to move freely and to change position.
Unpacking
Read this manual carefully BEFORE you remove the 4900 Multigas Gas
Analyzer from its shipping container, or you attempt to install,
commission or use the equipment.
1. Remove the analyzer and any other equipment from its packaging.
2. Remove the protective plastic covers from the sample gas inlets and outlets on the
rear of the analyzer (Figure 3-1).
Gas
outlets
Hint: Remove the RED and BLACK protective covers before connecting to the
sample gas pipework. If you do not intend to use the analyzer immediately,
wait to remove the plastic covers until just before connecting to the sample
gas pipework.
3. Inspect the analyzer and the other items supplied and check that they are not
damaged. If any item is damaged, contact Servomex or your local Servomex agent
immediately.
4. Check the packing list to ensure you have received all the items ordered. If any item is
missing, contact Servomex or your local Servomex agent immediately.
5. If you do not intend to use the analyzer immediately:
a. Refit any protective plastic covers that you may have removed.
b. Place the analyzer and any other equipment supplied back in its protective
packaging.
c. Store the analyzer as described in section 15.1.
6. Read Section 2 – Safety before proceeding.
Hint: Keep all shipping packaging and documentation for future use when
moving, storing or returning the analyzer for service or repair.
Figure 3-3: Sliding rack installation for Extended Chassis version (M indicates
metric value in mm)
Make sure the analyzer is provided with a sound earth connection via the
electrical supply plug.
All signal and electrical supply cables must be rated for temperatures of
70°C or higher.
Make sure that the cables that you connect to the analyzer are routed so
they do not present a trip hazard.
Follow the instructions given below when you install the analyzer. If you
do not, the analyzer warranty may be invalidated, the analyzer may not
operate correctly, or it may be damaged.
Make sure your electrical supply can provide the necessary maximum
power consumption.
Refer to Table 3-1 to identify which screw terminal (J20, J21, J37, and J38) is connected
to which gas transducer position.
Connect the cable shielding to the ground point on the rear of the analyzer. The ground
points are marked with the symbol, or if more convenient the screw terminals at pins
3 and 6 may be used.
The analyzer must supply power for any mA input devices. Do not
allow devices on the inputs to supply power or the input readings may
not be valid.
Analog input number “X” shares the same isolated reference as the
analog output with the same number. This reference is isolated from
the rest of the chassis as well as from the other analog inputs and
outputs. Do not allow equipment wired to an input number X to be
grounded to the same frame as equipment reading from output number
X, or both input and output X readings may not be valid.
Refer to Table 3-1 to identify which screw terminal (J20, J21, J37, and J38) is connected
to which gas transducer position.
Connect the input cable wires to the pins on J19 (inputs #1 and #2), J36 (inputs #3 and
#4), as shown in Table 3-4. The (X) in the table indicates the input location in position 1,
2, 3, or 4 that matches with the label on the back of the analyzer:
Connect the cable shielding to the ground point on the rear of the analyzer. The ground
points are marked with the symbol. Connect an external mA loop device “+” to the
COMPL(X) pin and connect the external device “-“ to the IN(X)+ pin. The RET(X) pins
cannot currently be used.
The relay connections are separated from the analyzer mains circuits by
reinforced insulation. The terminals must only be connected to circuits
that are themselves separated from mains voltages by at least reinforced
insulation.
The analyzer relays are accessible via the connectors J9-J16 and J26-J33. Connect one
end of your cable wire to the screw terminal for the relevant relay connector as shown in
Table 3-3. Each connector has two relays assigned to it where the “X” in XNO, XC, XNC
represents the relay number, NO is Normally Open, C is Closed and NC is Normally
Closed.
Connect the wires in your cable to the screw terminals on the relevant connectors as
shown in Table 3-4 showing Relay X (J odd numbers) and Relay Y (J even numbers)
positions on the jumpers listed in Table 3-3.
For setting up the Zero Gas relay when Auto-Cal is purchased, connect the wires in your
cable to the screw terminals of the relevant connectors J11, J15, J28, and J32 (based
upon the number and position of the transducers in the analyzer) shown in Table 3-5. “Y”
Table 3-5: Zero Gas Relay Screw Terminal Pin Connections for Auto-Cal
For setting up the Span Gas and Sample Gas relays when Auto-Cal is purchased, connect
the wires in your cable to the screw terminals to the relevant connectors of J12, J16, J29,
and J33 as shown below to control the Span Gas Relays and / or Sample Gas Relays
shown in Table 3-6. “X” represents the connector J12 for gas #1, J16 for gas #2, J29 for
gas #3 and J33 for gas #4 and Pins 1 – 3 control the Span Gas Relays while Pins 4 – 6
control the Sample Gas Relays.
Table 3-6: Span and Sample Gas Relay Screw Terminal Pin Connections for Auto-
Cal
Only use the power supply cord provided with the unit.
Make sure the analyzer is suitable for use with your electrical supply
voltage and frequency (Section 12). If the analyzer is not suitable, it may
not operate correctly, or it may be damaged if you operate it.
The analyzer is supplied with an electrical supply cable and plug, configured for your
electrical supply. Connect the electrical supply to the analyzer as follows:
1. Turn the Power Switch on the back of the unit to OFF: press the “O“ on the On/Off
switch shown in Figure 3-5 A.
2. Fit the IEC plug on the end of the electrical supply cable provided to the electrical
supply socket on the rear of the analyzer (Figure 3-5 B).
A
B
Figure 3-5: Power switch (A) and electrical supply socket (B) on rear of analyzer
3. Plug the other end of the electrical supply cable into your electrical supply outlet.
4. Check the earth (ground) continuity between your electrical supply outlet earth
(ground) and the functional earth (ground) terminal on the rear of the analyzer.
5. If a local earth bonding is required, the functional earth stud can be used. The earth
ground cable must be kept to less than 3 meters to comply with EMC standards.
This does not replace the earth conductor on the electrical supply
socket which must always be connected. Therefore never cut or
remove any of the metal pieces from the supplied plug.
It is essential that the analyzer is isolated from the sample system until
any cleaning solvents are fully purged from the pipelines. Failure to take
this precaution may lead to contamination of the transducer, which will
be observed as an offset and drift in output.
Hint: This section gives simple instructions about connecting the sample and
calibration gas pipelines to the analyzer.
Sample and calibration gases pass into the analyzer via Sample Inlet 1 (Figure 3-6 A) and
out via Sample Outlet 1 used (Figure 3-6 B).
Sample Inlet 2 and Sample Outlet 2 are installed if required for a Nitric Oxide (NO)
transducer (for use with an external Nitrogen Dioxide (NO2) converter) or a Carbon
Monoxide (CO) transducer. (see Figure 3-6).
Figure 3-6: Gas inlets and outlets for two sample streams.
Note: It is recommended an external filter is fitted (see accessories in Section 16) at the
analyzer inlet or, if preferred, at a convenient point in the sample line prior to the
inlet.
Figure 3-7: Gas Flow Meter location on front panel of the standard analyzer
A
1
To switch off the analyser, press the “O” on the On/Off switch on the rear of the analyser
(Figure 4-1 A).
If you intend to leave the analyzer off for an extended period of time, for example, when
carrying out plant maintenance and will not use the analyzer for several days:
• Turn off the analyzer and disconnect the electrical supply cable from the analyzer.
• Purge the transducers with Zero Air or Nitrogen gas to remove any sample gas.
• Close off the sample gas inlet and outlets using a shut off valve or the protective caps
supplied with the analyzer.
It is essential that the analyzer is isolated from the sample system until
any cleaning solvents are fully purged from the pipelines. Failure to take
this precaution may lead to contamination of the transducer, which will
be observed as an offset and drift in output.
Hint: When the electrical supply to the analyzer is switched on, a series of beeps
will be heard, the readings are displayed on screen and the clock in the upper
right hand corner of the screen starts running.
Figure 4-2: On/Off switch on the rear of the analyzer, (A) points to the Power Switch
“I” for ON position.
Figure 4-3: Home screen for a 3-transducer analyzer showing the O2 warming up
Hint: Figure 4-3 shows the unit in Warm Up mode, indicated by the three wavy lines
shown in the upper left corner encircled by the red box.
Purchased Options
Unpurchased Option
Figure 5-1: Setup Screen Icons showing 3 mA Inputs activated (purchased) and the
bottom mA Input is grayed out (not purchased)
Introduction
The user interface is a touchscreen display with icon-driven menus to allow for easy and
intuitive operation of the 4900 Multigas analyzer. Figure 5-2 shows the main display for a
four-transducer analyzer with the Main Menu option icon in the lower right-hand corner.
4900 Multigas analyzers share several standard features with optional features dependent
on the configuration of transducers, options purchased and setup preferences.
B
A
Figure 5-2: Home screen for a 4-transducer analyzer (A) is the Main Selection Touch
Screen and (B) indicates the Main Menu Icon
A Touchscreen display Screens, horizontal bars and icons are displayed on the
touchscreen depending on the information and operation
being engaged.
B Touchscreen icons The icons displayed depend on the current screen
capabilities / function. When touched the icons will
produce a new screen or icon list.
Note: If there are no menu interactions for 60 seconds, the display reverts back to this
Home screen. This timer can be adjusted in the Settings section.
To interact with a specific gas transducer, touch on the horizontal bar on the home screen
display for that gas transducer when more than one gas stream is displayed or touch the
single gas screen. A single screen display will show the gas transducer specific user
interactions that are available for the selected item.
Note: The remainder of the software descriptions will show a single transducer window
when appropriate.
General techniques
The general navigation route through the user interface screens is described by a
sequence of icons that you must touch to get to the desired screen. A shortened visual
description of the sequence of icons to be touched is used in this manual to help you
navigate easily to the various screens.
For example, to reach the Alarms screen (a sub-screen of the Measurement branch) you
must press the following sequence of icons:
> >
Familiarity of the icons below will allow easier navigation of the menus.
Displays the Main Menu screen that contains the four main
Main Menu branch icons: Measurement, Diagnostics, Maintenance, and
Settings.
Displays the next set of functions onto the screen. The new
Next list will always appear in a new column to the right of the
arrow.
>
The Measurement choices available are shown in Figure 5-3 and show up as icons in the
column to the right of the Main Menu list. The icon background now turns blue
indicating it is the active Main Menu choice as you navigate forward through the various
choices.
Note: The Main Menu branch stays visible all the time unless you are in a special screen
or the Home page. This allows you to access the other Main Menu choices easily.
Main Menu
branch icons
Next list of
functions
Figure 5-3: The user interface of the Main Menu screen with the Measurement
branch active.
When a Main Menu icon is selected further icons associated with that function are
displayed as seen in Figure 5-3. New icons associated with that function will appear to the
right of the icon just touched or it may transfer you to a new screen.
In the case of the Main Menu Branch icons, if there are more functions associated with the
main function activated and they do not all fit onto one screen then the Next List key
will be present in the lower right corner or on the bottom of the list. When touched, more
functions will appear in a new list to the right of the old list for selection.
The Main Menu Branch will remain visible as the farthest column on the left. Details are
shown later in this section.
Figure 5-4 shows a Step Series of Screens that are launched when the Settings Main
Menu icon is touched:
>
• The first series of functional icons that belong to the Settings section are displayed in
the column of icons to the right of the Main Menu icons (Figure 5-4 A).
> >
• To display the next set of functions touch which brings up the second set of
Setting functions (Figure 5-4 C) to the right of the first set (Figure 5-4 A). Note that
is activated as the background color is now blue (Figure 5-4 B).
A B C D
Figure 5-4: The User Interface Menu screen with the Settings branch activated
Note: The Main Menu branch stays visible all the time even while navigating through the
three sets of functional icons of the Settings branch. In this case, the first set of
icons also remain visible and only the third colum of icons is replaced when the
Next List icon is touched.
A
F
B
G
C, D
E H
Hint: If no icon is pressed for 1 minute in any other menu branch, Home screen is
automatically displayed. You will also then have to re-enter your password to
access any password-protected screens. The “Home screen return” value can
be increased in the Settings Menu from 1 to 3 or 5 minutes.
Note: The first column of icons on each menu screen is the same for all analyzers. Once
one of the four menu branches are selected that relevant icon background changes
blue to show which menu screen is active (see Figure 6-1 below).
Note: The four main menu branches are shown in bold in Table 5-1.
Note: If you touch the green OK icon it will display the date and time when the analyzer
was last started.
If a problem occurs with the system, the Status icon changes to one of the symbols shown
in the table below.
Alarm Indicates that there is an alarm on the system. Touch the icon to
display the Alarm Selection screen. An example is shown in
Figure 5-7.
Faults Indicates a fault with the transducer or analyzer: a
communication failure with the transducer, an over-temperature
condition, or out of specification where the measured value is out
of the maximum range for the transducer (see Figure 5-8). Touch
the icon to display a message in the text bar describing the fault.
Figure 5-8: Home screen (3 transducer analyzer), showing warning screen with fault
description
Note: In Figure 5-8 the Calibration icon was touched and “Reading: 20.02” was
displayed at the bottom of the screen indicating the O2 value is out of specification.
Diagnostics
> Section 6.1.3
Maintenance
> Section 6.1.4
Settings
> Section 6.1.5
6.1.2 Measurement
Note: Each Alarm and Relay have their own settings. Only one set is shown as example.
Zero / Low
Span Set-up Set & Run Zero / Low
Calibration
Audible Alarm
On / Off
Alarm Following
On / Off
Alarm Latching
On / Off
Alarm Off
Audible Alarm
Flow Alarm
On / Off
Alarm Following
On / Off
Alarm Off
Transducer Temp
Value Low
Transducer Temp
Alarm Value
Threshold
Transducer Temp
Alarm Value
Hysteresis
Audible Alarm
On / Off
Alarm Following
On / Off
Alarm Latching
On / Off
Alarm Off
Analog Output
Settings Follow Meas Value
During Cal On / Off
Jam Condition
Zero
Jam Condition
2 mA
Jam Condition
21.5 mA
Show
Transducer
Temp
Filter Type 2
Filter Type 3
Filter Type 4
Filter Type 5
Filter Type 6
Activate / Deactivate
Transducer
Firmware Rev
Transducer
Serial Number
Transducer Auto-Validation
Auto-Cal Run Auto-Validation
Sequence Selection
Settings Setup Now
(#1, #2, #3)
Abort Auto-
Validation
Set Auto-Validation
Date / Time / Repeats
Abort Auto-
Calibration
Transducer
Diagnostics
Decimal Place
Adjustment
Select 2 or 4 decimal GFX or Paramagnetic
places sensors only
Analyzer
Electronic Temp
Analyzer SN
Display System
Log
Display
Calibration Log
6.1.4 Maintenance
Update Firmware
Delete Calibration
Log
Delete Data
Recording Log
Upload System
Settings Files
Clear Errors
6.1.5 Settings
RS485
Full Duplex
Half Duplex
RS485 only
2 Stop Bits
RS485 Only
Delta F Legacy 485
Functions
Modbus
Off
Streaming
Modbus Word
Swapping: On / OFF
Relay
Alarm Assignment Select Transducer
Assignment
Measurement #1, #2,
etc.
Transducer Alarm
“Service in
Progress”
Transducer Fault
Alarm
Range Change
Operator
Keystroke Recording
On / Off
Normal Operation –
Relay ON
Manual Relay
Override
Normal Operation –
Relay OFF
Forced ON Operation
– Relay ON
Analyzer Date
DD, MM, YYYY
Analyzer Time
HH:MM:SS
High Value
Text input
Units
Text input
Display Text /
Analyte / Device
Text input
Record Analog Input
Choose mA Input
Contact Closure
(left side)
Stop Auto-Cal
Concentration ppm
Units
mg/m3
IP Static
DHCP Active
Screen View
Single / Split Screens
Decimal Place
Adjustment
Select 2 or 4 decimal GFX or Paramagnetic
places sensors only
Increase / Decrease
Screen Brightness
mA Output adjust
Jam Condition
Zero
Jam Condition
21.5 mA
Block Time
Global Block
Constants
Averaging
Off, 1 sec, 2 sec, 5
sec, 10 sec, 20 sec, 30
sec, 60 sec, 120 sec
The Measurement branch has four pages of icons (shown in Figure 7-1 to Figure 7-4).
Touch the Measurement icon to see the first page of icons appear on the right of the
Measurement icon (Figure 7-1). Touch the Next icon and a new column of active
icons will appear over the last column on the right.
Note: The measurement selected determines the active settings and diagnostics functions.
Not all of the features are available to every transducer type, so some screens may
look different depending upon the allowed features.
Note: Page 1 displays first.
To go to the subsequent pages, press the icon.
To go back to the previous page, press the icon.
Note: The transducer firmware revision, temperature, and serial number display in a text
box at the bottom of the screen when the relevant button is pressed. The
information is valid for the specific gas transducer that was selected.
Note: The Manual calibration and validation descriptions and settings are described later
in the manual in Section 11 along with the Auto-Cal optional features.
> >
The Alarms screen shows how each of the eight alarms is currently configured. The
standard analyzer is supplied with 2 alarms for each transducer, and a further 2 or 6
alarms per transducer can be optionally purchased. Figure 7-6 shows a CO transducer
where the option for 8 total alarms has been purchased. Note that none of the alarm bells
are grayed out. Alarm #1 is set to High Alarm, Alarm #3 is set to Low Alarm, Alarm #2 is
set to the Low Flow Alarm. Alarms 4, 5, 6, 7 and 8 are not yet configured.
Figure 7-6: Alarms screen showing CO with 8 alarms available: #1, #2, #3 have
already been configured.
Note: During a calibration, an alarm will only be activated if the alarm 'Follow' option
is selected (see Section 7.6).
2. The Alarm setup screen for the selected alarm is displayed. Figure 7-8 shows Alarm
Figure 7-8: Measurement Alarm setup screen for Transducer #1 and Alarm #4
In Figure 7-8 the following alarms have been set or deactivated: Audible alarm is on
, Do Not Follow is on, Do Not Latch is on and no Measurement value
alarm (Hi or Lo) has been selected.
Note: Most of the icons are paired with the Deactivate and Activate icon next to the
Alarm icon on the same row. The exceptions are the High and Low
Measurement Alarm icons that are located at the bottom right side with the
deactivate icon between them in a column.
Figure 7-9: Measurement Alarm setup screen for Transducer #1 and Alarm #1
In Figure 7-9 the following alarms have been set or deactivated: Audible alarm is off
, Do Not Follow is on, Do Not Latch is on and the Hi Measurement value
alarm has been selected.
Note: Figure 7-8 and Figure 7-9 show examples of the different options available for
specific transducers or system.
3. Press the required icon on the row horizontally across the top of the screen to either
configure or activate the alarm as the main alarm function.
4. Table 7-3 shows the available alarm functions and descriptions, some of which can be
used together on the same alarm.
Alarm Mode available but Alarm available but is not configured to activate
deactivated under any condition. Can use this to quickly
remove an alarm configuration that had been
previously assigned.
Alarm Mode Low Alarm will be activated when a sample
measurement is lower than the pre-set alarm
level.
Alarm Mode High Alarm will be activated when a sample
measurement is higher than the pre-set alarm
level.
Alarm Value Threshold Set the value at which the High or Low alarm
will be activated.
Alarm Value Hysteresis These values determine when an activated
measurement alarm condition will be
deactivated.
Audible Alarm Mode is On Activates the audible alarm so the alarm
sounds when the alarm limit is triggered.
Audible Alarm is Deactivates the audible alarm. Also used to
deactivated silence the alarm when the Audible Alarm Mode
is On and has been triggered.
Alarm Following Activates the alarm during a calibration. If the
alarm threshold is reached the alarm will sound.
Alarm Following Deactivates the alarms during a calibration.
deactivated This is the default mode.
Alarm Latching The alarm condition remains activated (both
visible and audible) until the alarm is manually
deactivated (see below).
Alarm Latching The alarm condition remains activated until the
deactivated sample measurement value falls within the
normal limits, at which point the alarm is
automatically deactivated.
This icon is also used to deactivate all Alarm
Latching conditions.
The Threshold Level is the value at which the measurement will trigger the alarm. To
set the Threshold value for the high or low alarms for the measurement range of the
transducer use the following sequence (specific example below is for Alarm #1 setting the
High Alarm):
B
Figure 7-10: Alarm Threshold values set by pressing (A) and number key pad (B).
1. Touch the number to the right of the icon as shown in Figure 7-10 A.
2. Use the number keypad shown in Figure 7-10 B to type in the threshold value then
press to accept the value or to cancel the entry.
3. To save the alarm settings to the current alarm, press the icon shown in the left
screen in Figure 7-10 A.
4. To abort the changes and return to the original settings prior to entering the Alarm
Setting screen, press the icon. This icon appears in place of the Return icon
in Figure 7-10 A if changes have been made to the settings.
Icon Function
Figure 7-11: Home screen showing alarm settings as red lines and text on the bar
graph
Low alarm Once the low alarm condition has been activated, the alarm condition
will not be deactivated until the value of a sample measurement is
above the low alarm level + hysteresis level.
High alarm Once the high alarm condition has been activated, the alarm
condition will not be deactivated until the value of a sample
measurement is below the high alarm level – hysteresis level.
Examples:
a. If a 'Low' alarm has an alarm level of 15 %-vol and a hysteresis level of 1
%-vol, the alarm is activated when a sample measurement is < 15 %-vol,
and the alarm is not deactivated until a sample measurement is > 16%-vol.
b. If a 'High' alarm has an alarm level of 3 ppm and a hysteresis level of 1
ppm, the alarm is activated when a sample measurement is > 3 ppm, and
the alarm is not deactivated until a sample measurement is < 2 ppm.
To set the Hysteresis level associated with a measurement alarm use the following
sequence (specific example below is for Transducer #1 and Alarm #1):
1. Touch the number to the right of the icon as shown in Figure 7-12 A.
2. Use the number keypad shown in Figure 7-12 B to type in the threshold value.
A
B
Figure 7-12: Alarm Hysteresis values set by pressing (A) and number key pad (B).
Hint: Ensure that the measurement alarm and hysteresis levels are not too close to
the expected sample measurements. Otherwise minor and potentially
acceptable variations in sample gas concentrations will result in spurious
alarms.
Hint: If you configure one measurement alarm as 'low' and the other measurement
alarm as 'high', ensure that the 'high' alarm with its hysteresis levels is higher
than the 'low' alarm with its hysteresis levels. Otherwise, the analyzer can be
permanently in an alarm condition until the hysteresis levels are adjusted.
Press the following sequence of icons to view the details of the activated alarm (Figure
7-14 A).
> >
Figure 7-14: Alarms screen showing Alarm #1 (A) High Measurement Range
triggered
Figure 7-15: Measurement Screen - Alarm Limits (in red) on Range Limit bar
To make any changes press the following sequence from the Home page to get to the
Analog Output Range Setting configuration panel in Figure 7-17:
Touch to reach the screen shown in Figure 7-16. Select the measurement to be
configured by touching the appropriate “Tx #”. For example, in Figure 7-16 the
measurement “Tx #2” has been selected and is highlighted. Touch to enter the
“Analog Output Range Setting” configuration screen (see Figure 7-17). This screen allows
configuration of Auto-ranging (activated icon highlighted in the red box A) and User
expanded Range values (shown in red box B).
Note: Several of the icons on this screen look the same ( ) but have very different
functions. Please use the guidance in the document and Figure 7-17 as to which
icon represents the various functions.
Note: More than one mA output may be assigned to a measurement (Tx #) provided an
additional option board is ordered. Each mA output will require its range setting to
be configured.
Table 7-7: Standard Navigation tools in the User Range Value screen
Icon Function
The User Expanded Range (or over-range) setting is used to define a new over-ranging
maximum value that the serial output and analyzer display will reset to when the sample
gas measurement value exceeds the User High Range value. This range will always be
less than the intrinsic measurement range of the transducer and allows more precision
across a smaller measurement range when over-ranging takes place.
To set a new User Expanded Range for the transducer, touch the icon shown inside
the red box labelled B in Figure 7-17 and use the numeric key pad in Figure 7-18 to enter
the new maximum range value.
Once a new User Expanded Range value is set, and the sample measurement
value is above its maximum value, depending upon the analyzer settings the display and
Initial Settings:
• User High Range set to 10 ppm, User Low Range set to 0 ppm, User Expanded
Range set to 100 ppm.
• The serial output values will be scaled to the User High Range of 10 ppm.
• The display bar and output values will be scaled to the User High Range of 10
ppm.
Any time the gas measurement value is at or above the User High Range of 10 ppm:
• The serial output value will remain fixed at 10 ppm.
• The analyzer display will continue to show the measurement values up to the User
Expanded Range maximum of 100 ppm.
Note: If the user does not set an Expanded Range and sets a User High Range of
10 ppm, then for a gas measurement value above 10 ppm the Serial output will
remain at 10 ppm and the analyzer display will continue to show the measurement
values up to the maximum of the intrinsic range of the transducer ie 1000 ppm.
Example 2 Auto-Range ON
Initial Settings:
• User High Range set to 10 ppm, User Low Range to 0 ppm, User Expanded Range
set to 100 ppm.
• The serial output values will be scaled to the User High Range of 10 ppm.
• The display bar and output values will be scaled to the User High Range of 10
ppm.
Measured Value User Low User High Expanded
≤ 10 ppm (0 ppm) (10 ppm) (100 ppm)
Serial Output X X
Any time the gas measurement value is at or above the User High Range of 10 ppm:
• The serial output value will automatically rescale to the User Expanded Range
maximum of 100 ppm.
• The display bar graph value will automatically rescale to the User Expanded Range
maximum of 100 ppm.
Measured Value User Low User High Expanded
≥ 10 ppm (0 ppm) (10 ppm) (100 ppm)
Serial Output X X
Display Output X X
Note: If the user does not set an Expanded Range and sets a User High Range of
10 ppm, then for a gas measurement value above 10 ppm the Serial output and
the analyzer display will automatically rescale the measurement values up to the
maximum of the intrinsic range of the transducer ie 1000 ppm.
Each transducer is supplied with an associated 4-20mA output. On the mA output screen
select either to “Follow” or “Not Follow” the concentration changes during
calibration. If the “Not Follow” is selected then the mA output signal will freeze at the
last measured value until the calibration has finished.
You can make small adjustments to the mA output using the icon (see 7.4).
A jam condition occurs when the incoming data has a Fatal Flag or when the Transducer
has a Communications Fault that stops the data. By assigning a jam condition the user is
able to distinguish between valid and non-valid readings.
Once the adjustments and flags are set, touch the icon to save the mA output
information or the icon to quit the screen without saving.
Note: You must touch the GREEN ENTER key for this change to occur. If you touch
the key the entered values will be displayed but will default to 6 decimal places.
Note: To change the number of decimal places displayed, you have to enter a value with
the required number of decimal places and save it by touching .
The measurement record option allows the operator to save time-stamped concentration
readings in seconds, minutes and hours as a text file to the analyzer which can then be
exported to a USB stick. It is possible to configure multiple record files if there are two or
more transducers.
Figure 7-21: Measurement Record (left) and Record Timing Setup (right) screens
To store a result:
1. Touch the clock icon to toggle through the time options until the required time
page displays.
2. Type 1 on the numeric pad to select 1 second, 1 minute or 1 hour, depending on the
time interval selected.
3. Touch the icon to start the logging.
4. To stop the logging, set the time interval to zero by typing 0 on the numeric pad.
To view a result:
1. The log must be saved onto a memory stick before it can be viewed.
2. Insert a memory stick into the USB socket located at the bottom of the right side of the
front panel.
3. Use the following sequence to navigate to the Diagnostics branch page to access the
USB icon and touch to save the files to the USB stick:
>
4. Several system files including the measurement logs will be written onto the memory
stick. Column 1 is the ppm measured value; column 2 is the date; column 3 is the
time.
50 27/7/2016 09:34:09
50 27/7/2016 09:44:09
50 27/7/2016 09:54:09
50 27/7/2016 10:04:10
Note: To stop the logging you must set the time interval to zero. Type 0 on the
numeric pad and press key to accept the value.
Icon Function
If you use your finger or Stylus and click on the center of the screen, a Data Recording
Plot will show up, presenting the last 10 hours of data that was recorded. See Figure 7-22.
Figure 7-23: Data Filter screen showing two filter types with short, medium, long
time averages (no filter is selected at this time).
To activate the Derived NOx measurement, touch the NO measurement on main screen
(see Figure 7-24) then touch the sequence of icons shown above to reach page 3 of the
Measurement branch (see Figure 7-24). Touch the NOx Derived icon to access the
screen shown in Figure 7-26.
7.8.3 How the Derived NOx value is implemented in the 4900 Multigas analyzer
An analyzer fitted with the NO transducer (GFX 1210) will automatically have this derived
value available as a choice. When activated the units displayed will be determined by the
analyzer global setting for concentration unit (ppm or mg/Nm3). If the global setting is
changed from one to the other then the NOx Derived value will also change. In general if
you are using this term the output concentration units should be set to Mass (mg/Nm3).
The analyzer first takes the measured NO concentration value and multiplies it by the User
Selected value of either 2.5% or 5.0% (indicative of the maximum amount of NO2 in the
flue gas). Table 7.10 shows the factors applied by the analyzer to calculate the derived
NOx.
Note: In all cases the NOx is calculated after NO value is reported
If the Global Setting is in PPM-V (not mass based but volume based) then the Derived NOx
will be in PPM. If NO concentration is measured at 10 ppm then NOx will be calculated as
follows:
User Selected 2.5%: Derive the NOx using 2.5% NO2 on top of the NO 1.0256
(100/97.5) then multiply this by the 10 ppm-v NO value measured by the NO
transducer.
User Selected 5.0%: Derive the NOx using 5.0% NO2 on top of the NO 1.053 (100/95)
then multiply this by the 10 ppm-v NO value measured by the NO transducer.
Example 2
If the Global Setting is in mg/m3 then the Derived NOx will be in mg/m3 (note mg/m3 is
based upon mg/Nm3) and a further conversion is needed to convert the NO mass into a
reported mass of NOx as NO2. So if NO is at 10 mg/m3 then NOx will be:
User Selected 2.5%: Derive the NOx using 2.5% NO2 on top of the 1.0256 NO
(100/97.5) then multiply this by the mass ratio of NO2 to NO 1.5298 (2.05 NO2 mg/m3
÷ 1.34 NO mg/m3) then multiply this by 10 mg/m3 NO value measured by the NO
transducer.
User Selected 5.0%: Derive the NOx using 5.0% NO2 on top of the NO 1.053 (100/95)
then multiply this by the mass ratio of NO2 to NO 1.5298 (2.05 NO2 mg/m3 ÷ 1.34 NO
mg/m3) then multiply this by 10 mg/m3 NO value measured by the NO transducer.
Transducer Diagnostics
This section provides access to the diagnostics for installed transducers that support
diagnostics at this time. We are working on allowing user access to diagnostics for all our
transducers and these new features will roll out in successive firmware updates. Figure
7-28 shows and example of the diagnostics for a GFX transducer.
The diagnostic values are meant for interpretation by service personnel and be used to
report the values back to a Servomex service person for troubleshooting the problem.
>
Diagnostics
Menu Branch
Save system files to Save the system files to a USB memory device.
USB Unit will beep if you do not have a USB device
installed.
Relay test menu To view and test the relays that were purchased.
When background is blue then the relay is closed.
System Log Display up to 100 pages of logged system
activities with most current date first. Use the
arrow keys to navigate through the pages.
Calibration Log Display up to 100 pages of logged calibration
activities with the most current date first. Use the
arrow keys to navigate through the pages.
A USB thumb drive must be installed into the slot located on the bottom right side of the
front panel prior to accessing the Log Save icon . The text files will be written to the
USB drive when this icon is touched. The analyser will beep if there is no USB in the port
and the system will stop.
For example, if relay #8 alarm state was triggered then relay #8 would show closed
prior to entering the Diagnostics menu. When exiting the Relay Diagnostics page, relay #8
will be set back to the closed position even if the relay had been opened on the
diagnostics page.
If Auto-Cal function is purchased the last three relays on each transducer option card are
pre-assigned and cannot be changed by the user (see Figure 8-5). The relay number icons
are now replaced by new icons that represent each of the Auto-Cal Zero, Span and
Measure functions.
Relay #6 is pre-set as the Zero or Low Span function , Relay #7 is pre-set as the High
Span function , and Relay #8 is pre-set as the Sample / Measurement function
for Transducer #1. Relays #14, #15, #16 are pre-set in the same manner for Transducer
#2.
Touch the Maintenance icon to see the available icon choices to the right. The
Maintenance branch menu has one page, shown in Figure 9-1.
Maintenance
Menu Branch
> >
This screen displays the amount of memory used, how much space is still available in
kilobytes and the percent of total disk space that is used. The displayed icons allow
selective freeing up of disk space and to copy the data from the disk to a USB memory
device. (see Figure 9-2)
Delete Keystroke Logs To delete all of the Keystroke Logs from the
Analyzer Memory.
Delete Calibration Logs To delete all of the Calibration Logs from the
Analyzer Memory.
Save Data to USB To copy the recorded data from the Analyzer
Memory to a USB stick.
Delete Data Recording To delete all of the Data Recording Logs from
Logs the Analyzer Memory.
>
Settings
Menu
Branch
active
> >
Settings
Menu: Next
List active
Digital and Analog Set up for any of the analog and/or digital inputs if
Inputs purchased.
Pump Off Option is not available for 4900 Multigas therefore
the icons are greyed out.
Pump On Option is not available for 4900 Multigas therefore
the icons are greyed out.
Data Reporting Change the units from ppm to mg/Nm3 or the
Units Selection reverse.
Modbus TCP/IP Used to activate and select mode of the Modbus
TCP/IP communications (over ethernet port)
Screen Functions Assign ICONS to Favorites shortcut page, switch
from Single to Multiple gases displayed, adjust
brightness, adjust screen “home” timer.
Analog Output Used to reset the display range bar and the mA
Range Settings output to a user defined maximum limit (User
Extended Range) and to define a lower
measurement range (User High Range and User
Low Range limits) on each of the transducers see
Section 7.3.5 for details.
Used to configure the 4-20mA or the 0 – 10V
Analog Output
output (see Section Configure the measurement
Settings
mA Output7.4 for details).
Note: To delete the last digit typed in, press the icon.
RS485 function: output To set the intervals for the output frequency (in
frequency setup seconds) via a numerical entry screen.
RS485 – Modbus:
To switch on Modbus word swapping.
Word swapping on
The function of the relay can be assigned using the relay configuration settings menu
and selecting the appropriate relay. Each relay is set up to respond to any function or
combination of functions including Faults, Service in Progress, Any type of Alarm. Once
an activity has been assigned to the relay, the icon will take on a symbol reflecting the
new assigned function (see Figure 10-11).
Note: If Auto-Cal is purchased then Relays at positions #6, #7 and #8 on each of the
transducers will be pre-assigned to output as Zero, Span and Sample /
Measurement modes. They cannot be reassigned (see Figure 10-11).
Figure 10-11: Available relays on the system with Auto-Cal pre-assigned relays.
If there are more relays available, then use the Next icon to access the next page
(see Figure 10-11).
Note: There can be up to eight relays per transducer giving a maximum of 32 if four
transducers are purchased or extra option boards added.
Figure 10-12: Activity Assignment Menu for Relay #12 in an analyzer with two
transducers (CO and O2).
Multiple activities can be assigned to one relay. In Figure 10-12 the Calibration, Fault and
Out of Range functions for the CO transducer are all active (so the icons show a dark blue
background) and are assigned to Relay #12.
Those activities can also be linked to different transducers present in the analyzer. In
Figure 10-12 Relay #12 can be assigned alarms for Transducer #1 (CO) and / or
Transducer #2 (O2). To assign an alarm to Transducer #2 (in this case it is O2):
• touch the alarm icon located below the O2 symbol as shown in Figure 10-12
• touch the icon for Alarm #1 as shown in Figure 10-13 to set the alarm
• Touch the icon to set the Low Limit alarm (Figure 10-14).
• Touch the icon to set the Hi Limit alarm (Figure 10-14).
• Touch the icon to return to the relay assignment page (Figure 10-14).
• Touch the icon (see Figure 10-12) to save the relay information.
Touch the Password Assignment icon to see the Operator , Master and
Keystroke Record icons in the column furthest right (see Figure 10-15).
> >
Once a master and/or operator password are set all protected icons and operations will
prompt for a password. If no passwords are set, then the protected icons and operations
will not prompt to have a password entered.
To reset or deactivate the OPERATOR password enter the OPERATOR password as a
blank / empty field. You will need the Master password to re-enter either the MASTER or
OPERATOR password. If you have forgotten the MASTER password a remote recovery
service password can be provided with a call to your local customer service center.
The function will cycle through three states each time the relay icon is touched:
Note: There are three general functions to keep in mind that are available for relay
settings and diagnostic testing that are accessed by the icon sequences below:
> > This sequence is used to set and assign tasks / states to the
relays described above in Section 10.2.
> > This sequence will over-ride the original state by forcing the
relay to be On or Off. The condition is permanent until the
relay is switched back to Normal Operation above in Section
10.4. When you exit the Manual Over-ride menu, any changes
will be made permanent.
1. Touch the first set of up or down arrows on the left-hand side to select how the date
will be displayed (DD/MM/YY or MM/DD/YY or YY/DD/MM).
2. To set the date (Figure 10-18 shows the DD/MM/YY format):
a. Touch the second set of up or down arrows from the left-hand side to select
the day (from 1 to 31).
b. Touch the third set of up or down arrows from the left-hand side to select
the month (where 1 is January, and 12 is December).
c. Touch last set of up or down arrows from the right-hand side to select the
year.
3. Touch the icon to accept the value or the icon to leave the screen without
updating the value.
5. Touch the icon to accept the value or the icon to leave the screen without
updating the value.
A
B
Figure 10-20: Settings sub-screen for mA Analog (A) and Digital (B) Inputs
Figure 10-21: Active mA Analog Inputs that are purchased appear as blue pictures /
text
The mA Analog Input is configured by touching areas on a graphic display. Figure 10-22
shows an example of the mA Analog Input #2 set up for an external Temperature Sensor
A
B
C H
D G
E F
Figure 10-22: Screen for entering the mA analog input #2 example.
D Set and alarm for the analog input H Record analog signal
Alarms can be assigned to the mA Input, in the same way as the standard transducer gas
measurements. Figure 10-23 shows the mA Analog Input #2 with a High alarm threshold
value set to “21.5” (Figure 10-23 B) however the function is currently disabled (Figure
10-23 D).
C D
E
F
Analog Input #2 set up Set up a specific analog input, in this case #2.
Threshold value Input the threshold value that will trigger the
Alarm.
Hysteresis value Input the hysteresis value at which the Alarm will
reset.
There are two Digital Inputs for each option board for a total of up to 8 for a fully-
configured analyzer. The Digital Input can be used to control the calibration of a transducer
or declare the state of an Analog Input.
Digital Input #2
Figure 10-24: Digital Inputs screen for setting up the digital inputs - total of 8 shown
in this example)
Figure 10-25 A shows the location of the Calibration operational icons and the Analog
inputs shown in Figure 10-25 B.
A
Figure 10-25: Digital Inputs screen for Digital Input #2 showing the assignments for
(A) Calibration control operational icons and (B) Analog input that can be controlled.
A Digital Input operation is essentially a binary ON / OFF state that detects the contact
closure between the labeled pins on the back-panel connectors. The Digital Input can only
be assigned one action.
To configure a Digital Input operation the user must select both the operation and the
target device.
Each Digital Input can be assigned as OFF or one of six possible operations. The available
operations are shown in the box labelled A in Figure 10-25. From top to bottom, left to right
for the transducer:
As another example, if the user were to select Ext Cal and within the box labelled A
in Figure 10-25 then mA Input #3 within the box labelled B in Figure 10-25 then it
would mean a declaration that the device attached to Analog In #3 is currently in
calibration.
Note: It is important to verify that both device and operation are highlighted before leaving
this page because some selections may cause the other buttons to be reset.
Digital Input #8 Set Up Set up a specific digital input, in this case #8.
Stop / Disable Auto- Digital input is selected to stop and disable the
Cal* Auto-Cal routine. *
Zero or Low Span Digital input is selected to trigger the Zero or Low
Calibration Span Calibration. *
* Next action - a gas number or a mA Input must then be selected to assign the measurement gas or
input to the operation.
Note: This is a global setting that will change all GFX Infrared measurements from ppm to
mg/m3 units or mg/m3 to ppm units.
Note: The mg/m3 units are calculated as mg/Nm3).
Screen Settings
> > > >
Touch the to activate the “Add Assignment” mode (top right-hand corner in Figure
10-29).
a) The is now activated so touch on any of the open slots which will bring you to
the Main Menu. See Figure 10-30.
b) Since no Icon is selected at this point the grayed out “unavailable” icon is
showing. See Figure 10-30. You need to select an icon and if the “available” is
present at the bottom of the page then your selection is allowed.
c) To assign the Manual Calibration section for the O2% Transducer (shown in Figure
10-30) to the Favorites touch the following sequence to get to the Manual Calibration
Section.
>
The screen shown in Figure 10-31 will appear.
Touch the Home Icon to return to the HOME page. The Favorites Icon Page can
be accessed by touching (see Figure 10-33). The is located above the Main
Menu icon for a Multiple Gas screen or just to the left of this icon for a Single
Gas screen.
Figure 10-33: Home Page after an Icon has been added to Favorites
NOTE: Some Back buttons operations will take you back to the Favorites Page and
some will not once Favorites have been set and used.
NOTE: If Password Protection is activated, you will need the operator password to access
the Favorites page each time. You will need the master password to Hide or Not Hide the
Favorites as well as to add or remove assignments.
The Analog Output Range Settings activate auto-ranging and set values for the Expanded,
High and Low User ranges for each measurment (Tx #). To make any changes press the
following sequence from the Home page.
Detailed information on these adjustments is given in Section 7.3 “Configuring the User
Ranges for mA Output and Screen Display.”
The Analog Output Settings are used to adjust the mA output, set the jam mode under
fault conditions and behavior during a calibration. To make any changes press the
following sequence from the Home page.
The Home Page in Figure 10-34 shows each transducer is assigned to a single analog
output. For our example, we touch measurement #3 to display the measurement menu.
Touch the Analog Output Settings icon as shown in Figure 10-35 to reach the
Analog Output assignments page.
See Figure 10-36. The Analog Output Assignments page shows the chosen measurement
#3/output #3 is highlighted. Outputs #1 and #2 are shown in gray which means they are
already assigned. The icon for output #4 is displayed in blue but is not highlighted, which
means it is unassigned and available to be assigned.
To adjust the Range Settings for output #4, touch the Range Settings icon for output #4,
(ringed in Figure 10-37). See Section 7.3.1 for a description of how to adjust the Range
Settings.
The measurement page for measurement #3 will now show two scales, one for each
analog output (see Figure 10-38). Compare the measurement page in Figure 10-38 with
the one in Figure 10-35.
Note: You will only need to calibrate the higher of the two ranges.
> >
The block averaging filter icon is reached through the Settings branch (see Figure
10-40). Touch the to select the sampling interval (see Figure 10-41). The filter is
inactive when is selected.
When the block averaging filter is active the analyzer gathers up all data during a
designated length time interval for each sensor. Available intervals are 1 sec, 2 secs, 5
secs, 10 secs, 20 secs, 30 secs, 60 secs, 90 secs, and 120 secs. When the time interval is
up it reports a single average for that time and begins gathering the next interval’s
data. For example, if the interval is set for 5 sec, new measurements are reported only
every 5 seconds and the value is the average from the preceding 5 seconds up to that
report.
The block averaging is a global setting that applies to all the transducers. The block
averaging takes place after any sensor-specific filtering has been applied.
Introduction to Calibration
The calibration of the transducers in the analyzer may be checked or adjusted either
manually or automatically (requires the Auto-Cal option) and the activity will be logged as
an entry in the calibration history log. The Auto-Cal feature allows the user to set up to
three automatic timing sequences on each of the transducers. These sequences can be
any combination of calibration or validation processes (described in Section 11.6).
External valves or a single multiport switching valve will need to be installed by the
customer to enable the use of the various Span and Zero gases as well as the Sample gas
stream. These valves can then be assigned to the relays associated with each transducer
for controlling Zero, Span and Sample gas. For the Auto-Cal option three relays per
transducer will be pre-assigned as Zero, Span and Sample mode automatically.
If the analyzer is still warming up a ( ) symbol will appear in the corner for each of the
gas icons. If a calibration is attempted, a warning message will appear with an option to
proceed. As each gas transducer warms up this symbol will disappear and at this time a
calibration or validation can be performed on that transducer.
Note: For optimal performance, allow the analyzer to run and stabilize for at least 24
hours from a cold start at 20oC (68 oF). For the higher sensitivity measurements, this
time may be longer.
The operator is guided through the setup sequence choices by a series of menus, icons,
and editable text. Each gas sensor / transducer parameter including output concentration
units, triggers and ranges, must be setup individually.
Hint: The required calibration interval depends on the reliance that you place upon
the accuracy and consistency of the measurements made by the analyzer.
Adjust the calibration interval according to your requirements and the drift
characteristics of your analyzer.
Manual calibration
> >
Zero or Low Span To set the Low Span calibration point, generally
set to “0” using a zeroing gas like Nitrogen.
Analog Output Touch this icon to set the parameters for the
Adjustment analog output
1. Run the High Span calibration gas through the analyzer and wait 15 minutes.
a. Allow the span value to stabilize for one minute.
b. Touch the high span measurement icon shown on the screen (see
Figure 11-1).
2. The new screen (see Figure 11-2) will display the target value along the top bar.
a. The target value will be the last span value that was used to calibrate that
transducer.
b. If the target value is not correct for the calibration gas you are using,
change the target value to the certified gas bottle concentration value using
the numeric keypad (see Figure 11-2).
c. accepts the new entry, deletes the last typed value and
cancels the entry.
3. Press to accept the new entry and launch the manual calibration run screen
shown in Figure 11-3.
b. The Run icon will stay red (see Figure 11-3) while the gas measured
value is:
• For High Span: less than 80% of the High Span target value (in the
example above this would be ≥ 40 ppm).
• For Zero / Low Span: greater than 20% of the High Span target value
(in the example above this would be ≤ 10 ppm).
c. When the gas measured value is near to the target value the Run
icon will turn from RED to GREEN (Figure 11-4).
• For High Span: >80% of the High Span target value (in the example
above this would be ≥ 40 ppm).
• For Zero / Low Span: < 20% of the High Span target value (in the
example above this would be ≤ 10 ppm).
5. When the calibration procedure is finished touch the Back icon to terminate
“Service In Progress” and bring up the measurement screen (Figure 11-1).
Note: The User can over-ride the warning given by the Run icon by touching the
icon. The Calibration sequence will continue even though the measured gas value
is not near the target value. This is not recommended as the red icon may be
flagging there is something wrong with the sampling system, the calibration gas or
the transducer.
Run Calibration with Start the Manual Calibration run. The icon turns
measured value <20% green when the measured value is within 20% of
away from target. the Target value.
The optional Auto-Cal feature provides validation and calibration sequences that can be
triggered automatically. The validation sequences allow the user to validate the reading
against preset Zero and Span gas value and tolerance. The calibration sequences allow
validation followed by calibration to update the transducer calibration equation.
Any of the sequences can be set to run at user chosen times and dates automatically. To
validate or calibrate multiple transducers at the same time using a gas blend then each
transducer date, time and repeats must be set up to be the same. The Span value for the
Calibration or Validation must then be added for each gas transducer based upon the
certified value of that gas constituent in the blended cylinder.
External valves or a single multiport switching valve will need to be installed and
connected to the transducer relays by the customer to enable the “Service in Progress”
signal and the Auto-Cal routine to trigger the Zero, Span and Sample modes on each
transducer.
If more than one transducer is present in the analyzer then all of the transducers will turn
Yellow indicating they are no longer producing “good” sample gas readings and are in the
“Service in Progress” mode (see Figure 11-6). Transducers that are in the Auto-Cal or
Auto-Val sequence will show up as black text on the screen while those that are not in
progress are in grey text. In Figure 11-6 the NO and SO2 gas transducer sequence timings
were configured to be the same and they are undergoing Auto-Cal simultaneously as
indicated by the BLACK text.
When Auto-Cal is purchased a manual validation check or calibration adjustment can still
be made, using the same transducer relays and valves established in the Auto-Cal
configuration.
The auto-validation and auto-calibration processes can be initiated by four methods:
1. by an internal timer;
2. by an external contact closure;
3. by operator manual request using the analyzer front panel user interface;
4. by an external Modbus or PROFIBUS command
The instrument will only respond to a request for Auto-Cal from the internal timer or
external input if there are no faults indicated. Auto-Cal can be initiated from the user
interface when there is a fault condition.
If any one of these events occurs while the Auto-Cal cycle is in progress the request will
be ignored. Touching the keypad during Auto-Cal will initiate the abort sequence.
The Auto-Cal feature offers custom configurations for automatic Validation and Calibration
operations. The general sequence uses three gas streams (Zero, Span and Sample) and
the Zero gas or the Span gas can be skipped by setting the timing (t1, t2..) to zero,
skipping that step.
Up to three sequences can be assigned to each transducer (see Figure 11-7). They can
be of any form of auto-calibration or auto-validation as needed.
For example, if you need to perform a daily Zero and Span check / validation then you
could assign this to sequence #1 using the auto-validation function. In sequence #2 you
might assign a weekly auto-calibration sequence.
An overview of how the Auto-Cal validation sequences are handled is given in Section
11.7 and an overview of how the Auto-Cal calibration sequences are handled is given in
Section 11.8.
NO
Start Auto
Run ZERO / LOW Record ZERO/ LOW
Validation / Time >t1? YES
Gas As Read
Check
NO
Compare ZERO As
Value within
Read to ZERO
Time > t2? Run HIGH SPAN GAS YES Tolerance?
Tolerance
Resume SAMPLE
Time > t3? YES
GAS Reporting
Point Function
1 Zero gas routed to transducer and set Service In Progress flag and
turn on any Service In Progress relays.
t1 Time required for Zero gas readings to stabilize.
2 Zero gas “As Found” value recorded and compared to “0” ± tolerance.
If Zero gas “As Found” value Passes - Skip to Point 3.
If Zero gas “As Found” value Fails – Log Failure and turn Gas icon
and Auto-Cal icon RED.
3 Span gas routed to transducer.
t2 Time required for Span gas readings to stabilize.
4 Span gas “As Found” value recorded and compared to High Span
value ± tolerance.
If Span gas “As Found” value Passes - Skip to Point 5.
If Span gas “As Found” value Fails – Log Failure and turn Gas icon
and Auto-Cal icon RED.
5 Sample gas routed to analyzer.
t3 Time required for Sample gas readings to stabilize and begin reporting
gas measurments. This cannot be set to “0” and is defaulted to 1 sec
before restoring the measurment values.
6 Clear Service In Progress and turn off any Service In Progress
relays and begin reporting Sample gas measurements.
Final output will be Pass or Fail flag
Point Function
1 Zero gas switch to analyzer skipped as t1 set to “0”
t1 Zero gas Time set to “0”
2 Zero gas “As Found” read and comparison skipped as t1 set to “0”
3 Span gas routed to transducer and set Service In Progress flag and
turn on any Service In Progress relays.
t2 Time required for Span gas readings to stabilize.
4 Span gas “As Found” value recorded and compared to High Span
value ± tolerance.
If Span gas “As Found” value Passes - Skip to Point 5.
If Span gas “As Found” value Fails – Log Failure and turn Gas icon
and Auto-Cal icon RED.
5 Sample gas routed to analyzer.
t3 Time required for Sample gas readings to stabilize and begin reporting
gas measurments. This cannot be set to “0” and is defaulted to 1 sec
before restoring the measurment values.
6 Clear Service In Progress and turn off any Service In Progress
relays and begin reporting Sample gas measurements.
Final output will be Pass or Fail flag
A
C
B
D
E F
G H I
Figure 11-11: Auto-validation setup screen is active (t2 is skipped)
Table 11-5: Auto-Validation setup screen input function labels for Figure 11-11
Symbol Function
A Touch this text box to enter High Span concentration bottle tag / certified
value (shown as 200.000 ppm tag value).
B Touch this text box to enter High Span concentration tolerance value as a
percentage of the High Span value or a fixed value (shown as ±0.500
ppm fixed concentration value).
C Touch this text box to enter High Span concentration tolerance units as
concentration (ppm, mg/m3 or %-v) or a percent of the span value
(shown as ppm).
D Touch this text box to enter Zero or Low Span concentration bottle tag /
certified value. Default is 0.000 (shown as 0.000 ppm value).
E Touch this text box to enter Zero or Low Span concentration tolerance
value as a percentage of the High Span value or a fixed value (shown as
±5.000 ppm fixed value).
F Touch this text box to enter Zero or Low Span concentration tolerance
units as concentration (ppm, mg/m3 or %-v) or a percent of the span
value (shown as units as ppm).
G Touch this text box to enter Time required for Zero gas readings to
stabilize, then take an “As Found” reading (see Section 11.7.2 for
location of #1,t1,#2). Shows t1 as 30 seconds.
H Touch this text box to enter Time required for High Span gas readings to
stabilize, then take an “As Found” reading (see Section 11.7.2 for
location of #3,t2,#4). Shows t2 as Off indicating the High Span Validation
check section will be skipped.
I Touch this text box to enter Time required for Sample gas readings to
stabilize before measurement reporting begins reading (see Section
11.7.2 for location of #5, t3, #6). Shows t3 as 30 seconds.
Symbol showing where Measurement of Sample begins. Has no function.
Compare ZERO As
Value within
Read to ZERO
Time > t2? Run HIGH SPAN GAS YES Tolerance?
Tolerance
NO
YES
Compare ZERO As
Record ZERO/ LOW Value within Update HIGH SPAN
Read to ZERO YES RUN SAMPLE Gas Time > t4?
As Read Tolerance? Calibration Curve
Tolerance
NO YES
(1) Abort Calibration sequence
(2) Send As Read values to Audit Log
Resume SAMPLE
(3) Store As Read Value on Analyzer Validation FAILS
GAS Reporting
(4) GAS Species RED
(5) CAL Icon RED
Symbol Function
1 Zero gas routed to transducer and set Service In Progress flag and turn
on any Service In Progress relays.
2 Zero gas “As Found” value recorded and compared to “0” ± tolerance.
4 Span gas “As Found” value recorded and compared to High Span value
± tolerance.
If Span gas “As Found” value Passes - Calibrate to High Span value.
Skip to Point 5.
If Span gas “As Found” value Fails – then Auto-Calibration sequence is
terminated, Validation Failure flag is issued, and the Gas Component and
Auto-Cal icon turn RED. Skip to Point 7 and switch gas to sample stream.
If Zero gas “As Found” value Passes Calibrate to Low Span value which
is generally “0”. Skip to Point 7.
If Zero gas “As Found” value Fails – then Auto-Calibration sequence is
terminated, Validation Failure flag is issued, and the Gas Component and
Auto-Cal icon turn RED. Skip to Point 7 and switch gas to sample stream.
t4 Time required for Sample gas readings to stabilize and begin reporting
gas measurments.
8 Clear Service In Progress and turn off any Service In Progress relays
and begin reporting Sample gas measurements.
Final output will be Pass (Recalibration) or Fail (No Calibration,
Failure Flag set, Switch to Sample gas)
Symbol Function
2 Zero gas “As Found” read and comparison skipped as t1 set to “0”
3 Span gas routed to transducer and set Service In Progress flag and turn
on any Service In Progress relays.
4 Span gas “As Found” value recorded and compared to High Span value
± tolerance.
If Span gas “As Found” value Passes - Calibrate to High Span value.
Skip to Point 5.
If Span gas “As Found” value Fails – then auto-calibration sequence is
terminated, Validation Failure flag is issued, and the Gas Component and
Auto-Cal icon turn RED. Skip to Point 7 and switch gas to sample stream.
6 Zero gas “As Found” value recorded and compared to Low Span “0”
value ± tolerance.
If Zero gas “As Found” value Passes Calibrate to Low Span value which
is generally “0”. Skip to Point 7.
If Zero gas “As Found” value Fails – then auto-calibration sequence is
terminated, Validation Failure flag is issued, and the Gas Component and
Auto-Cal icon turn RED. Skip to Point 7 and switch gas to sample stream.
t4 Time required for Sample gas readings to stabilize and begin reporting
gas measurments. This cannot be set to “0” and is defaulted to 1 sec
before restoring the measurment values.
8 Clear Service In Progress and turn off any Service In Progress relays
and begin reporting Sample gas measurements.
Final output will be Pass (Recalibration) or Fail (No Calibration,
Failure Flag set, Switch to Sample gas)
Hint: The key difference between the sequences shown in Figure 11-13 and Figure
11-14 is that there is no initial Zero “As Found” value recorded in the sequence
shown in Figure 11-14. The sequence in Figure 11-13 will support regulatory
compliance where both the Zero gas and the Span gas “As Found” readings
need to be recorded prior to any calibration curve changes from the High Span
or the Low / Zero Span calibrations.
A
B
C
D
E F
G J
H I
Figure 11-15: Auto-calibration setup screen is active
Symbol Function
A Touch this text box to enter High Span concentration bottle tag / certified
value. Figure 11-15 shows 20.900% tag value.
B Touch this text box to enter High Span concentration tolerance value as a
percentage of the High Span value or a fixed value. Figure 11-15 shows
±1.000 %-vol fixed value.
C Touch this text box to enter High Span concentration tolerance units as
concentration (ppm, mg/m3 or %-vol) or a percent of the span value (%).
Figure 11-15 shows units as %-vol.
D Touch this text box to enter Zero or Low Span concentration bottle tag /
certified value. Default is 0.000. Figure 11-15 shows 0.000% value.
E Touch this text box to enter Zero or Low Span concentration tolerance
value as a percentage of the High Span value or a fixed value. Figure
11-15 shows ±1.000 %-vol fixed value.
F Touch this text box to enter Zero or Low Span concentration tolerance
units as concentration (ppm, mg/m3 or %-v) or a percent of the span
value. Figure 11-15 shows units as %-vol.
G Touch this text box to enter Time required for Zero gas readings to
stabilize, then take an “As Found” reading (see Section 11.8.2 for
location of #1, t1, #2). Figure 11-15 shows t1 as 30 seconds.
H Touch this text box to enter Time required for High Span gas readings to
stabilize, then take an “As Found” reading, and Calibrate if reading
outside of tolerance (see Section 11.8.2 for location of #3, t2, #4). Figure
11-15 shows t2 as 30 seconds.
I Touch this text box to enter Time required for Zero gas readings to
stabilize, then take an “As Found” reading, and Calibrate if reading
outside of tolerance (see Section 11.8.2 for location of #5, t3, #6). Figure
11-15 shows t3 as 30 seconds.
J Touch this text box to enter Time required for Sample gas readings to
stabilize before measurement reporting begins (see Section 11.8.2 for
location of #7, t4, #8). Figure 11-15 shows t4 as 30 seconds.
Symbol Description
A, D High Span and Zero Span Bottle Concentration value. The units here can
be set at any time from the Measurement Branch but are based upon the
intrinsic transducer value.
If the transducer intrinsic value is (ppm) concentration, then the units will
be displayed as ppm and can be changed to mg/m3.
B, E High Span and Zero Span Tolerance Value. The units are set in this
screen.
If the Span or Zero gas value “As Read” is outside of this tolerance, then
the auto-validation or auto-calibration will fail and no changes will be
made to the transducer calibration.
If the Span or Zero gas value “As Read” is inside of this tolerance, then
the auto-validation passes and auto-calibration recalibrate the transducer
equation using the new value.
C, F High Span and Zero Span Tolerance Value Units. The units are set in
this screen.
If the Threshold Value Unit selected is (%-vol) then the value represents
a concentration value.
If the Threshold Value Unit selected is (%) then the value represents a
percentage calculation of the concentration value entered regardless of
the unit type (%-vol, ppm, mg/m3).
If the Threshold Value Units are (mg/m3) then the value represents a
fixed concentration value above and below the Span or Zero
concentration in mg/m3.
If the Threshold Value Units are (ppm) then the value represents a fixed
concentration value above and below the Span or Zero concentration in
ppm.
This section applies to both the auto-calibration and the auto-validation sequences and is
used to set up the date, time and repeats for the automatic runs.
Touch the icon in Figure 11-17 to get to the Date screen to set the Day/Month/Year
for the first auto-calibration / auto-validation event. The first set of arrows allows you to
choose to display the values in MM/DD/YY, DD/MM/YY, or YY/MM/DD.
• Touch to cancel the changes and go back to Sequence Setup initial screen
Touch the icon in Figure 11-18 to accept the Date and take you to the Time screen
to set the Hour/Minute/Second for the auto-calibration / auto-validation event.
• Touch to cancel the changes and go back to Sequence Setup initial screen
Touch the icon in Figure 11-19 to accept the Time and take you to the Repeat
Timing screen. This is used to repeat the auto-validation / calibration sequence on a
regular basis, every XX Days or every YY Hours.
• Touch to cancel the changes and go back to Sequence Setup initial screen
• Touch to cancel the changes and go back to Sequence Setup initial screen
The voltage input is activated by applying (+) 5 – 24 VDC to DIN 1A and (–) 5 – 24 VDC to
DIN 1B. The contact closure input is activated by a contact closure created between DIN
1B and DIN 1C.
Details on setting up the various input signals are found in Section 10.7.
Figure 11-23 shows an example of a Calibration Log File as viewed on the analzyer
screen. Use the and arrow to scroll up or down for more text information on the
Auto-Cal operations and results. The information presented here is the same as in the
output text log from above.
Mechanical specification
Dimensions: (Width x Height x Length)
Bench top: 430 x 140.5 x 544.2 mm
17 x 5.5 x 21.4 inches
Bench top with extension chassis: 430 x 265.5 x 544.2 mm
17 x 10.5 x 21.4 inches
Rack mount: 481.5 x 132.5 x 544.2 mm
19 x 5.2 x 21.4 inches
Rack mount with extension chassis: 481.5 x 265.5 x 544.2 mm
19 x 10.5 x 21.4 inches
Weight: Main unit: 14kg (30.9 lb.) approx.
Expansion chassis: 13.7 kg (30.2 lb.) approx.
Electrical specification
Electrical supply:
Voltage: 100 to 240 Vac, 50 to 60 Hz
(± 10% maximum fluctuation)
Environmental limits
The equipment is suitable for indoor use only.
Ambient temperature range:
Operation: 5 to 45 °C
Storage: 0 to 50 °C
Operating ambient pressure range: 101.3 kPa ± 10% (1.013 bar ± 10%)
Operating ambient humidity range: 10 to 90% RH, non-condensing
Operating altitude range: -500 metres (below sea level) to 2000 metres
(above sea level)
Ingress protection: IP20
Do not open or attempt to remove the analyzer cover yourself. If you do,
you will invalidate any warranty on the analyzer, and the analyzer may
not operate safely or provide accurate measurements.
Routine checks
You only need to carry out simple maintenance procedures annually. Carry out the
following regular checks to ensure continuous and safe operation of the monitor.
Fire Hazard: Only use the same type and rated fuse as recommended.
If you think that an electrical supply fuse has failed, use the following procedure to inspect
the fuses and replace them if necessary:
1. Open the fuse panel on the rear of the analyzer (Figure 13-1). To do this, carefully
insert a small screwdriver into the gap on the right of the panel and press the clip to
open the panel.
2. Pull the red fuse holder out of the panel (Figure 13-2).
Both live and neutral lines have fuse protection. The neutral fuse is shown (1 in Figure
13-3); the live fuse is located in the underside of the red fuse holder (2 in Figure 13-3).
3. Remove the top (neutral) fuse from the holder and check the continuity across the
fuse.
If there is continuity, the fuse has not failed, so refit it into the fuse holder.
If there is no continuity, fit a new fuse into the fuse holder.
Make sure the fuses are the correct type and rating. The fuse type and
rating is shown on the rear panel to the left of the mains connector.
Make sure you fit the fuse in the correct position in the fuse holder as
shown in (Figure 13-3).
4. Repeat step 3 for the bottom (live) fuse which is located on the underside of the red
fuse holder.
5. Push the fuse holder back into the fuse panel and close the panel door. It will click into
place.
Preventative maintenance
To minimize unscheduled analyzer downtime, ensure the proper operation of the analyzer
and to comply with the guidelines of applicable regulatory bodies, we recommend that you
utilize an annual preventative maintenance program for your analyzer.
The preventative maintenance program consists of an annual inspection of the analyzer,
and repair of any faults, to ensure that the analyzer meets its original factory specification.
Contact Servomex or your local Servomex agent to arrange for a preventative
maintenance contract.
Error codes
The following codes may be displayed in the event of a problem or error. The table gives a
suitable remedy.
Jam conditions
See section 10.1 “Configure and use the mA outputs."
Storage
Refit any protective plastic covers and place the analyzer and any associated equipment in
its original packaging before storage. Alternatively, seal it inside a waterproof plastic bag,
sack, or storage box.
Store the analyzer and any associated equipment in a clean, dry area. Do not subject it to
excessively hot, cold, or humid conditions.
Disposal
Dispose of the analyzer and any associated equipment safely, and in accordance with all
of your local and national safety and environmental requirements.
Hint: If you send the analyzer to Servomex or your local Servomex agent for
disposal, it must be accompanied by a correctly completed decontamination
certificate and a Return Authorization Number (RAN) (17.1Appendix E).
15.2.1 Disposal in accordance with the Waste Electrical and Electronic Equipment
(WEEE) Directive
The label shown in Figure 15-1 is fitted to the analyzer.
Fuse Kit
Main Fuse
Fuse Kit
Main Fuse
O2 0 - 25% 05200941
0 - 50% 0S1520702
0 - 25% 01520703
0 - 10% 01520704
0 - 5% 01520705
0 - 2.5% 01520706
0 - 1.0% 01520707
CO 0 - 10% 01522704
0 - 2.5% 01520706
0 - 1.0% 01522707
B.2 Connections
Earth / ground
RS232
RS485
Figure B-1: Rear panel of the analyzer showing RS232 and RS485 connectors
> >
RS485 function: output To set the intervals for the output frequency (in
frequency setup seconds) via a numerical entry screen.
RS485 – Modbus:
To switch on Modbus word swapping.
Word swapping on
RS485 – Modbus:
To switch off Modbus word swapping.
Word swapping off
Table B-1: Serial output data frame, start, measurement and end sequences
C.1 Introduction
This appendix details the implementation and use of the Modbus protocol in the 4900
Multigas analyzer.
C.2 References
Document "MODBUS over Serial Line Specification & Implementation guide V1.0 Nov 02”
located on Modbus web site modbus.org.
03 Illegal data value A value contained in the query data field is not an
allowable value. This indicates a fault in the
structure of the remainder of a complex request.
This does NOT mean that a value to be stored in a
register is incorrect as Modbus has no means of
determining what is legal for any particular
register.
C.6 Addressing
Addresses in Modbus ADU (application data unit), run from 1 – N, whereas addresses in
the Modbus PDU (protocol data unit) run from 0 – N. This appendix gives addresses in the
ADU model. Depending on the particular Modbus master, addresses may have to be
entered as they are given or have 1 subtracted from them. For example, to read register
101 an address of 100 may be needed.
14 date: Day ✓ ✓ ✓
1 0 0 Service in Progress ✓ ✓
0=Not In Service Mode, 1=Service Mode.
Instrument MUST be set to Service in Progress before any calibration or override actions are performed.
C.11 Measurements
Supports Function Code
2 AVFinishing ✓
3 AVFailState ✓
4 Number of Cal / Val
Points
✓
5 Select Cal/val point ✓ ✓ ✓
6 LastCal/val Point n
Reading
✓
8 LastCalPoint n Target ✓
10 LastCalPoint n Delta ✓
12 Last Cal point n Time ✓
13 Last Cal point n Date ✓
15 Cal point passed/failed ✓
AVFinishing 0=Not Finishing, 1=Finishing
AVFailState 0=Not in Fail State, 1=In Fail State
Select Cal/val point 0=zero 1=span; This must be written to read corresponding values for last reading,
target, delta. Delta is expressed as floating-point proportion ((target-last reading)/target).
13 Pressure
Compensated ✓
Measurement
15 Filtered Measurement ✓
17 Transducer
temperature
✓
4181 1 Tx 0 Alarm Active
(4n)
✓
1 Fault ✓
2 Service in progress ✓
3 Out of Specification ✓
4 Maintenance required ✓
5 Transducer
maintenance fault
✓
6 Transducer error ✓
7 Transducer fatal fault ✓
8 WarmingOn ✓
9 Reserved ✓
10 Reserved ✓
11 Calibration fault ✓
12 Communication fail ✓
13 Transducer not
detected
✓
14 Autoval / cal failed ✓
15 Remote calibration/val
denied
✓
4201 2 Tx 0 Heartbeat toggling at 1
(4n+1) Hz
✓
1 Remote service in
progress
✓
2 Transducer calibration
mode
✓
3 Auto validation/
calibration
✓
4 Incorrect transducer
type
✓
8765 Minimum
measurement
✓ ✓
D.1 Safety
Read the rest of this manual carefully before you use this appendix.
D.2 Description
The PROFIBUS option adds DPv0 synchronous and DPv1 asynchronous communication
capabilities to the 4900 Multigas Analyzer. The PROFIBUS option board provides the
following features:
• Supports DPv0/DPv1 PROFIBUS communication with RS485 transmission
• DPv1 communications with class I and class II Master devices
• Separate DPv0 modules to optimize bus load
• Profibus Address can be set from the User Interface
After changing the address, turn the unit off and on again for it to respond to the new
address.
Wait for parameterization: There are no communications between the analyser and the
PROFIBUS master device.
Data exchange: The analyser and the PROFIBUS master device are
communicating, using DPv0 cyclic messages.
Slot Module Input byte data offset Output byte data offset
1 Measurement One 0 0
2 Measurement One Status 6 6
3 Measurement Two 12 12
Example Configuration 2:
Slot Module Input byte data offset Output byte data offset
1 Measurement Three 0 0
2 Measurement One 6 6
3 Measurement One Status 12 -
4 Measurement Two 18 12
5 Measurement Two Status 24 -
Data format
Endianness: All data transfers are word aligned, and the analyzer uses
Big-Endian order, as specified in PROFIBUS standards. In
Big-Endian order, the Most Significant Byte (HI) comes
before the Least Significant Byte (LOW). Bit order follows
the same rule and bit 0, is the least significant bit.
Floating Point Numbers: All floating point numbers (e.g. 12.34, -1012.32, etc.), are
digitally represented using the IEEE-754 format. Single
precision floating point numbers are used throughout and
since they require 32 bits of data, they occupy 2 words. The
most significant word is transferred first.
Low/Hi Byte: The LOW byte of a word is the least significant byte. The
Most significant Byte is called the HIGH Byte.
Word
# Words Read/Write Name
offset
000=Idle
001=Pre-warning
0-2 Auto Validation State 010=Inerting
011=Flushing
100=Validating
000=zero gas
LOW
3-5 Auto Validation Gas 001=span gas
011=sample gas
0=not finishing
6 Auto Validation finishing
1=finishing
0=normal
7 Auto Validation fail state
1=failed
0 0= off (normal);
Calibration mode on/off 1= on (alarms masked, jamming etc).
Write 1 to turn calibration mode on.
1 0= failed
Cal point passed/failed
HI 1= passed
2 Reserved Reserved
3 0=disabled
Enable Baseline Subtraction
1=enabled
4-7 Reserved Reserved
Read /
Slot Index # Words Data
Write
The field definitions for these indexed data spaces follow sequentially (definitions within a
table also follow as commented).
0 Warming On
1 Heater fault
2 Sample heater fault
3 Calibration fault
Legacy flow n
Relay info n
0 State
1 Card not detected
2 Service in progress
0 iAOUT 1 status[8]
1 iAOUT 2 status[8]
2 iAOUT 3 status[8]
3 iAOUT 4 status[8]
4 iAIN 1 status[8]
5 iAIN 2 status[8]
6 iAIN 3 status[8]
7 iAIN 4 status[8]
8 iDin Card Not Detected[8]
9 iLegacy Pressure 1 status [8]
10 iLegacy Pressure 2 status [8]
11 iLegacy Pressure 3 status [8]
12 iLegacy Pressure 4 status [8]
13 iLegacy Flow Alarm 1 status[16]
14 iLegacy Flow Alarm 2 status[16]
15 iLegacy Flow Alarm 3 status[16]
16 iLegacy Flow Alarm 4 status[16]
17-20 Chassis Temperature
D.6.2 The communication starts, but there are frequent line drops
Make sure that the ‘PROFIBUS Cabling Guidelines’ have been followed. Refer to
www.profibus.com for the latest guidelines.
If possible decrease the network baud rate.
Date:
Address: Address:
Phone: Phone:
Email: Email:
Name: Name:
Address: Address:
Phone: Phone:
Email: Email:
Shipping
Instructions:
(UPS/FedEx
Acct, P&A)
Details of problem or
failure:
Shipping information:
Please contact the relevant Servomex Support Office (below) for your location to obtain the shipping
address.
We look forward to receiving your request. If you have any questions regarding this form, please
contact your local Servomex Support office.
Thank you for choosing Servomex.
Application / process:
Company name:
Name: Title:
Phone: Email:
Signature:
Date:
IMPORTANT NOTICE:
Servomex ensures that all products dispatched to customers have been suitably purged and cleaned
prior to packaging so that no hazards from the use of factory calibration gases or liquids are present.
1520 CO2 ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓ ✓
1522 CO ✓ ✓ ✓
Transducer FSD
Gases measured % O2
Range 0 – 100%
Minimum recommended output range 0 – 5%
Intrinsic error ± 0.1% O2
Linearity error < 0.05% O2
Inherently linear, dependent on calibration gases
Repeatability < 0.1% O2
Lower Detection Limit (LDL) 0.02% O2
(95% confidence interval)
Response (T90) at 1500ml/min < 15 s
Zero drift / week < 0.05% O2
Span drift / week < 0.1% O2
Output fluctuation (peak to peak) ± 0.05% O2
Ambient pressure coefficient Directly proportional to analyzer
vent pressure
Zero temperature coefficient / 10 °C change ± 0.1% O2
Span temperature coefficient / 10 °C change 1% of reading or ± 0.1% O2,
whichever is the larger.
Sample flow effect range over full flow range < 2% of reading or 0.1% O2,
whichever is the larger.
Argon -0.22%
Carbon dioxide -0.26%
Halothane -1.93%
Helium -0.29%
Note that the error is directly proportional to the concentration of the background gas in the
sample being measured and, in most cases, can be ignored. A detailed listing of these
measurement errors for a wide variety of background gases are listed in Section 17.1H.2
below.
XINT is a reported concentration adjustment that can be set in the Paramagnetic
Measurement menus. It is a scaling factor that is used to correct the current transducer
reading. The default value is 1.0. If required the operator can change this value to
compensate for the background gas. For example: If you are measuring oxygen in a
background of carbon dioxide -0.26 should be entered as the XINT value.
If XINIT compensation is to be used, care must be taken to insure that the value used for
the background gases is correct. During a calibration, no XINT compensation is applied
and it is assumed that the calibration gas sample has negligible cross-interference.
20 °C 50 °C
20 °C 50 °C
20 °C 50 °C
20 °C 50 °C
20 °C 50 °C
20 °C 50 °C
Transducer FSD
Gases measured GFX SO2 GFX SO2 GFX SO2 GFX NO GFX NO
high range standard high
high standard
sensitivity range
sensitivity range
Intrinsic error 1% of 1% of 1% of 1% of 1% of
reading, or reading, or reading, or reading, or reading, or
Linearity error
20 ppm* 5 ppm* 2 ppm* 3 ppm* 2 ppm*
Repeatability
Lower Detection Limit 0.41% of 0.41% of 0.41% of 0.41% of 0.41% of
(LDL) (95% reading or reading or reading or reading or reading or
confidence interval) 8.2ppm* 2.1 ppm* 0.82 ppm* 1.23 ppm* 0.82 ppm*
Intrinsic error 1% of 1% of 1% of 1% of 1% of
reading, or reading, or reading, or reading, or reading, or
Linearity error
5 ppm* 2 ppm* 1 ppm* 0.5 ppm* 0.5 ppm*
Repeatability
Lower Detection Limit 0.41% of 0.41% of 0.41% of 0.41% of 0.41% of
(LDL) (95% confidence reading or reading or reading or reading or reading or
interval) 2.1 ppm* 0.82 ppm* 0.41 ppm* 0.21 ppm* 0.21 ppm*
Ambient pressure 1% 1%
coefficient
Of reading per 1% change in
analyser vent pressure
Ambient temperature 3% of 3% of
coefficient / 10 °C change reading, or reading, or
2 ppm CH4* 1.5 ppm CH4
*
Viton ✓ ✓ ✓
Borosilicate glass ✓
Platinum ✓
Electroless Nickel ✓
Gold ✓
Calcium Fluoride ✓
Nickel ✓
Sapphire ✓
Epoxy resin ✓