H8 Hemoglobin Analyzer (HPLC)

Technical Specs
Methodology High - Performance Liquid Chromatography (HPLC)
Test Modes Standard Mode, Variant Mode, Thalassaemia Mode
Test Range 3% - 18%
Precision CV ≤ 1.5%
Test Speed 130 Secs / T for Variant Mode, 380 Secs / T for thalassaemia Mode
Venous Blood, Finger Peripheral Blood, Lyophilized Whole Blood
Sample Type
Peripheral Blood, 500ul ( 150 Dilution Ratio )
Auto Sample Station 10 Positions
Photometer 415nm + 500nm Detector
Chromatography Column Available Tests ≥ 800T
Filter ≥ 400T
Display 10.1 " TFT True Color LCD Touch Screen
Software Linux Software with Self - Diagnosis to Monitor and Detect System Errors
Reagent Kit Eluent A, Eluent B, Eluent C, Hemolysin, Calibrator, QC Material ( Weight Sensor ± 1% )
Information Input Scanner or Touch Keypad
Storage 4000 Sample Results
Connection USB, LAN, LIS Compatible
Printer Thermal Printer and External Laser Printer
Operation Temperature 10 ~ 30 °C ( 50 ~ 86 °F )

Humidity ≤ 80%

Power AC 100-240V 50/60HZ 120VA

Dimensions 600mm x 360mm x 540mm ( 23.6’’H x14.2 ''W x 21.3'' D )

Weight 49KG ( 108 Ibs )

Barcode Scanner QC Curve

NGSP International Federation


of Clinical Chemistry
and Laboratory Medicine

Lifotronic Technology Co., Ltd.


H8 Hemoglobin Analyzer (HPLC)
Address: 4th Floor, Building 15, 1008 Songbai Road, Nanshan District, Shenzhen, 518055, China

Tel: 86-755-29060026 Fax: 86-755-29060036 Email: [email protected]

Web: en.lifotronic.com en.lifotronic.com


H8 Hemoglobin Analyzer (HPLC)
The fully automated Lifotronic H8 Hemoglobin
Analyzer offers the fast throughput of HbA1c results in
130 seconds, with Hb variant detection, providing the
outstanding solution for quick and reliable diabetic
monitoring. No sample preparation and very little
hands-on time by the operator is required for the
H8 Analyzer.

The Elements of Hemoglobin

The content of HbF is very low in the blood of adults


Hb

Adult hemoglobin HbA HbA2 HbF Hb(Variant)

97% 2.5% 0.5% HbS,HbD,HbC,HbE and etc.


IFCC NGSP Fetal hemoglobin
Non-glycated Hb HbA0 HbA1
Glycated Hb
94% 6%
Gold Standard of Diabetes Diagnose
Form from the combination between HbA and glucose, and is 60~70% of total HbA1
Glycosylated hemoglobin (HbA1c) is widely recognized as a Gold Standard to monitor diabetes, which can indicate the HbA1a HbA1b HbA1c

average plasma glucose concentration over 8 ~ 12 weeks. Product of the combination between HbA and fructose / lactose

HPLC Technology – Gold Standard Methodology


HPLC Methodology
! NGSP and IFCC Certified
High-Pressure Liquid Chromatography (HPLC), to separate HbA1c, HbF, HbA2 directly with measuring the absorbance ! HbA1c Results in 130 Seconds
points continually to form chromatogram. Using normal distribution curve fitting auto-iterative algorithm to get Fully Automated - To Minimize Operation Hassles HbA0
precise HbA1c testing result, excluding interference of variant and unstable hemoglobin. Standard Analysis Mode will ! Primary Tube Sampling with Cap Piercing
report HbA1a, HbA1b, HbF, La1c, HbA1c, HbA0 peak areas and ratio. And the result also includes IFCC, NGSP and ! Fully Automated Start - up, Maintenance and Shutdown
ADAG value for diverse client needs. ! Barcode Scanner for Sample Identification
Precise and Reliable – To Serve You Consistently
! HbA1c Inter Measuring CV≤1.5%& Intra Measuring CV’s ≤3% to
y = 0.988x + 0.0873
R² = 0.9944 12
20 Enable Exceptional Result Management
18 10
! Superior Quality Chromatographic Resolution to Eliminate Interferences
16
14 8 Dual Wavelength Detection – To Avoid Interference
12
系列1 ! To Avoid the Reagent Peak Interference
H8

10 6 系列2

8 ! More Anti-interference Abilities, the Mutation Factor Interference to the Peak Can be Easily Counteracted
6 4
4 ! To Eliminate the Nonspecific Absorption of Hemoglobin
2 2

0
Degasser – For Better Result Accuracy
0 2 4 6 8 10 12 14 16 18 20 0
1 2 3 4 5 6 7 8 9 10
! More Stable Pressure, More Accurate Flow Rate
D-10
! To Reduce Background Absorption and Improve Detection Sensitivity
Correlation between H8 and D-10 Precision Study ! To Improve the Separation Effect of Column and Prolong Its Lifetime
Compact Size – To Minimize Space Requirements
! Small Footprint Reduces Bench Space Needed en.lifotronic.com