Assignment PH Meter
Assignment PH Meter
Muhammad Safeer
Chemist
E-1205
pH Meter
pH Measurement:
The pH rate of a material is directly linked to the degree of the hydrogen ion
[H+] and the hydroxyl ion [OH-] concentrations.
The quantitative data rendered via the pH meter shows the ratio of the
movement of an acid or base in terms of hydrogen ion activity.
If the H+ density is higher than OH-, the substance is acidic; i.e., the pH amount
is less than 7.
If the OH- intensity is higher than H+, the substance is basic, including a pH
value higher than 7.
If identical quantities of H+ and OH- ions are present, the substance is neutral,
with a pH of 7.
Calibration of pH Meter:
To achieve highly accurate measures the pH meter must be calibrated before each
measurement. Also, calibration is executed once per day of work. Calibration is
required because the glass electrode does not provide reproducible electrostatic
potentials across longer periods of time.
i. First, make sure the correct measurement mode is selected in pH meter.
ii. Then wash the electrode with distilled water and don’t wipe the electrodes,
otherwise it will generate an electrostatic charge on the electrodes.
iii. All standards and samples are brought to 25-degree centigrade before use.
iv. Now immerse the electrodes within the standard buffer solution or calibration
solution (pH 7/pH 10). Make sure the endpoint of the electrodes is fully
immersed within the buffer solution. To create the homogenous sample, stir
the electrodes within the buffer.
v. Now enter in pH calibration mode by pressing the CAL/MEAS key. The CAL sign
will be displayed.
vi. The measured reading will be shown by the primary display whereas the pH
standard buffer solution reading is displayed by the smaller secondary screen.
vii. Wait until the pH values get stabilized.
viii. After stabilization confirm the calibration by pressing the HOLD/ENTER key.
ix. Now the pH meter is calibrated to the current buffer solution.
x. Finally, wash the electrodes with distilled water and placed it into the buffer
solution.
Advantages of pH Meter:
pH Calibration is low-priced and robust.
Pocket size pH Meters are user friendly.
Accounts are reliable and specific.
Disadvantages of pH Meter:
Heat affects the output readings.
pH Calibration utilizing glass electrodes need to be clean as deposition on the
electrodes influences the readings.
Precautions:
Touch the electrode with responsibility- it is breakable!
Store the electrode always in immersed condition within the solution approved
by the company or neutral solution of KCl (3M-4M).
Always maintain the inner level of filling liquid beyond the level of measured
solution.
Load electrode (the flowing type) by exact filling solution (as suggested by
manufacturer – normally KCl solution, 3M to saturated) to not let it dry inside.
You may store the electrodes as dry if they are not used for a long period of
time to prevent aging. The Aging only happens when the electrode is moist.
Don’t do it with gel electrodes – certain must be put in a strong solution of KCl
only.
If the electrodes are dried soak them at least 24 hours before using.