Radar Tank Gauge
Radar Tank Gauge
Installation and
Operations Manual
www.varec.com
Varec, Inc.
5834 Peachtree Corners East, Norcross (Atlanta), GA 30092 USA
Tel: +1 (770) 447-9202 Fax: +1 (770) 662-8939
Radar Tank Gauge
000
measured value Contrast: + or +
009 008
- + E history E dist./
reset meas value
Group E
- + E
selection
00 002 003 004 005 006 007 008 051 052 053
basic setup tank medium process empty full pipe dist./ check range of start
shape property cond. calibr. calibr. diameter meas value distance mapping mapping
+
01 - flat - unknown - standard input E input F only for D and L are - ok confirm
safety settings ceiling - <1.9 - calm (see sketch) (see sketch) bypass + displayed - too small suggestion
- stilling - 1.9 … 4 surface stilling well (see sketch) - too big or specify
well - 4 … 10 - add. agitator - unknown range
… - >10 … - manual
basic setup flange:
03 030 033 034/035 reference point
mounting tank dip table dip table of measurement
calibr. gauging mode
history reset (= 555)
04 - dip table
linearisation - auto correct.
05 dip table
extended calibr. first point corrects offset
06 further points correct linearity
output D
Note!
This operating manual explains the installation and initial start-up for the level
transmitter measuring device. All functions that are required for a typical measuring
task are taken into account here.
In addition, the 7532 RTG provides many other functions that are not included in this
operating manual, such as optimizing the measuring point and converting the measured
values.
A description of the device functions for the 7532 RTG can also be found on the
enclosed CD-ROM.
Table of Contents
IOM027FVAE0804 3
Safety Instructions Radar Tank Gauge
1 Safety Instructions
FCC-approval
This device complies with part 15 of the FCC Rules. Operation is subject to the following
two conditions: (1) This device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired
operation.
Caution!
Changes or modifications not expressly approved by the part responsible for
compliance could void the user’s authority to operate the equipment.
1.4 Return
The following procedures must be carried out before a transmitter is sent to Varec for
repair:
• Always enclose a duly completed “Declaration of contamination” form. Only then can
Varec transport, examine and repair a returned device.
• Enclose special handling instructions if necessary, for example a safety data sheet as
per EN 91/155/EEC.
• Remove all residue which may be present. Pay special attention to the gasket grooves
and crevices where fluid may be present. This is especially important if the fluid is
dangerous to health, e.g. corrosive, poisonous, carcinogenic, radioactive, etc.
Caution!
• No instrument should be sent back for repair without all dangerous material being
completely removed first, e.g. in scratches or diffused through plastic.
• Incomplete cleaning of the instrument may result in waste disposal or cause harm to
personnel (burns, etc.). Any costs arising from this will be charged to the operator of
the instrument.
5
Safety Instructions Radar Tank Gauge
Safety conventions
Symbol Meaning
Explosion protection
Device certified for use in explosion hazardous area
0 If the 7532 RTG has this symbol embossed on its name plate it can be
installed in an explosion hazardous area
Electrical symbols
Direct voltage
% A terminal to which or from which a direct current or voltage may be
applied or supplied
Alternating voltage
& A terminal to which or from which an alternating (sine-wave) current
or voltage may be applied or supplied
Grounded terminal
) A grounded terminal, which as far as the operator is concerned, is
already grounded by means of an earth grounding system
2 Identification
7500RTG_nameplate_label.eps
Figure 1: Information on the nameplate of the 7500 RTG (example)
7500RTG_NMI_label.eps
Figure 2: Information on the NMi type plate for custody transfer applications of the
7500 RTG (example)
7500RTG_ATEX-PTB_label.eps
Figure 3: Information on the PTB type plate for custody transfer applications of the
7500 RTG (example)
7
Identification Radar Tank Gauge
30 Process connection
Flange Dia/Pressure Standard Material
AVJ 6"/150 lbs / RF ANSI B16.5 316/316L 11.4 kg
A3J 8"/150 lbs / RF ANSI B16.5 316/316L 19.6 kg
A5J 10"/150 lbs / RF ANSI B16.5 316/316L 28.8 kg
AWJ 6"/300 lbs / RF ANSI B16.5 316/316L 20.9 kg
AXJ 8"/300 lbs / RF ANSI B16.5 316/316L 34.3 kg
A7J 12"/150 lbs / RF ANSI B16.5 316/316L 43.2 kg
CWJ DN150 PN10 / 16 C EN 1092,1, B11) 316L 10.6 kg
CXJ DN200 PN16 C EN 1092,1, B11) 316L 16.5 kg
CZJ DN200 PN25 C EN 1092,1, B11) 316L 22.7 kg
C1J DN150 PN25 C EN 1092,1, B11) 316L 14.7 kg
C6J DN250 PN16 C EN 1092,1, B11) 316L 25.6 kg
C8J DN300 PN16 C EN 1092,1, B11) 316L 36.1 kg
XXJ with flange hub 316L
XVU Varec UNI-Flange 6"/ 304/1.4301 3.5 kg
DN150/150A,
max. 14.5LBS/PN1/1K,
compatible with:
- 6" 150LBS
- DN150 PN16
- 10K 150A
X3U Varec UNI-Flange 8"/ 304/1.4301 5.2 kg
DN200/200A,
max. 14.5LBS/PN1/1K,
compatible with:
- 8" 150LBS
- DN200 PN16
- 10K 200A
X5U Varec UNI-Flange 10"/ 304/1.4301 7.5 kg
DN250/250A,
max. 14.5LBS/PN1/1K,
compatible with:
- 10" 150LBS
- DN250 PN16
- 10K 250A
X7U Varec UNI-Flange 12"/ 304/1.4301 10.8 kg
DN300/300A,
max. 14.5LBS/PN1/1K,
compatible with:
- 12" 150LBS
- DN300 PN16
- 10K 300A
1) agreeable to DIN2527 Form C
YY9 Special version
50 Housing
C Aluminium T12-housing with separate connection compartment, coated, IP65
Y Special version
60 Gland / Entry
2 M20x1.5 cable gland
4 ½ NPT cable entry
9 Special version
80 Additional options
A Without additional options
Y Special version
9
Identification Radar Tank Gauge
Caution!
It is essential to follow the instructions concerning the unpacking, transport and storage
of measuring instruments given in the section “Incoming acceptance, transport, storage”
on page 12.
Accompanying documentation:
• Short manual (basic equalization/troubleshooting): housed in the instrument
• Operating manual (this manual)
• Operating manual: Description of the instrument functions
• Approval documentation: if this is not included in the operating manual.
TRI-CLAMP ®
Registered trademark of the company Ladish & Co., Inc., Kenosha, USA
HART ®
Registered trademark of HART Communication Foundation, Austin, USA
ToF ®
Registered trademark of the company Endress+Hauser GmbH+Co. KG, Maulburg, Ger-
many
PulseMaster ®
Registered trademark of the company Endress+Hauser GmbH+Co. KG, Maulburg, Ger-
many
PhaseMaster ®
Registered trademark of the company Endress+Hauser GmbH+Co. KG, Maulburg, Ger-
many
3 Mounting
T12 housing
2
˚
90
L00-FMR532xx-17-00-00-en-002
Note!
The 7532 RTG with planar antenna is designed for use on stilling-wells only!
11
Mounting Radar Tank Gauge
3.2.2 Transport
Caution!
Follow the safety instructions and transport conditions for instruments of more than
18 kg. Do not lift the measuring instrument by its housing in order to transport it.
3.2.3 Storage
Pack the measuring instrument so that is protected against impacts for storage and
transport. The original packing material provides the optimum protection for this.
The permissible storage temperature is -40 °C…+80 °C.
3.3.1 Dimensions
Ø 129 (5.08")
ENDRESS+HAUSER
Micropilot II
238 (9.37")
8 (0.31")
DIN, ANSI, E+H UNI flange
JIS (max. 1 bar)
93 (3.66")
inactive length
(2.36")
detail A:
60
flange adapter
(1.18")
(see under)
30
Ø 150 (5.91")
DN150 / 6”
L
DN200…300 / 8…12”
version with
antenna widening
detail A
flange hub for the connection
to flanges provided by the customer
mounting:
Ø 78 (3.07") 4 bolts M6 / 90°
e.g. DIN 912
Ø 62,5 (2.46")
ØD
O-Ring 85.3 x 3.53,
included
(same material as
sensor seal)
L00-FMR532xx-06-00-00-en-001
(0.35")
9
flange hub
Ø 99,5 (3.92")
Antenna version DN150 / 6" DN200 / 8" DN250 / 10" DN300 / 12"
L [mm] 93 337 490 517
Ø D [mm] no horn 190 240 290
Note!
The inactive length of 60 mm prevents condensation effects to the antenna perfor-
mance. Special versions with longer construction are available.
13
Mounting Radar Tank Gauge
100% BD
A
SD
max. level
L00-FMR53xxx-17-00-00-en-008
C
0%
Measuring range
The usable measuring range depends on the size of the antenna, the reflectivity of the
medium, the mounting location, and eventual interference reflections.
The following tables describe the groups of media as well as the achievable measuring
range as a function of application and media group. If the dielectric constant of a
medium is unknown, it is recommended to assume media group B to ensure a reliable
measurement.
1)Treat Ammonia NH3 as a medium of group A, e.g. always use a stilling well.
Measuring range
7532 RTG
> DN150
A DK(εr)=1,4…1,9 38 m/124 ft
B DK(εr)=1,9…4 38 m/124 ft
C DK(εr)=4…10 38 m/124 ft
D DK(εr)>10 38 m/124 ft
max. measuring
range with custody 20 m/65 ft
transfer approvals
15
Mounting Radar Tank Gauge
Blocking distance
The blocking distance (= BD) is the minimum distance from the reference point of the
measurement (mounting flange) to the medium surface at maximum level.
reference point of
measurement
L00-FMR53xxx-15-00-00-en-003
BD
20 mA
Max. 100%
Level
Note!
Inside the blocking distance a reliable measurement can not be guaranteed.
Standard installation
When mounting in the stilling well, please
observe engineering tips on page 14 and
the following points:
• After mounting, the housing can be
turned 350° in order to simplify access to
the display and the terminal
compartment.
• planar axis vertical to flange.
• Measurements can be performed
through an open ball valve without any
L00-FMR532xx-17-00-00-en-003
problems.
• Select antenna extension as big as possible. For intermediate sizes (e.g. 180 mm)
select next larger antenna extension and adapt it mechanically. Maximum gap allowed
between the antenna/horn of 7532 RTG and the inside of the stilling well is 5 mm (3/
16").
• The antennna extension of the 7532 RTG is mounted with defined pressure.
It is strongly recommended not to dismantle this antenna.
• Dimensions of a nozzle for manual gauging must be adapted to the dimensions of
the horn antenna used, compare to page page 23.
17
Mounting Radar Tank Gauge
7532 RTG
7532 RTG
L00-FMR532xx-17-00-00-en-004
T12 housing
2
L00-FMR532xx-17-00-00-en-005
Allen key
4 mm
tighten strongly
by hand
19
Mounting Radar Tank Gauge
L00-FMR53xxx-06-00-00-en-001
R532xx-00-00-06-en-002
21
Mounting Radar Tank Gauge
detail A:
flange adapter
(see under)
detail A
flange hub for the connection
to flanges provided by the customer
mounting:
Ø 78 4 bolts M6 / 90˚
e.g. DIN 912
Ø 62.5
O-Ring 85.3 x 3.53,
included
(same material as
sensor seal)
L00-FMR532xx-00-00-06-en-003
9
flange hub
Ø 99.5
Note!
The nozzle for manual gauging is not part of the standard offering from Varec. Please
contact Varec for further information.
L00-FMR53xxx-06-00-00-en-004
Flange DN150 DN200 DN250/300 Flange ANSI 6" ANSI 8" ANSI 10"
PN [lbs] 2)
16 16 16 PN [lbs] 2)
150 150 150
A [mm] 110 140 170 A [mm] 110 140 170
L [mm} — 300 450 L [mm] — 300 450
2)Only dimensions adapted to standard. Designed for non-pressurized operation only,
therefore thickness of flange can be reduced (e.g. 8mm).
23
Mounting Radar Tank Gauge
L00-FMR53xxx-17-00-00-en-013
4 Wiring
Wiring
25
Wiring Radar Tank Gauge
"
Caution!
Before connection please note the following:
● The power supply must be identical to the data on the
nameplate (1).
● Switch off power supply before connecting up the device. 1
● Connect Equipotential bonding to transmitter ground terminal
before connecting up the device.
● Tighten the locking screw:
It forms the connection between the antenna and the housing
ground potential. 2
When you use the measuring system in hazardous areas, make sure you comply with
national standards and the specifications in the safety instructions (XA’s).
Make sure you use the specified cable gland.
power: signal:
24 VDC 24 VDC
(16…30 V)
1 2 3 4
plant
1 2 3 4 5 ground
1 2 3 4 5
"
Caution!
Before connection please note the following:
● The power supply must be identical to the data on the
nameplate (1).
● Switch off power supply before connecting up the device. 1
When you use the measuring system in hazardous areas, make sure you comply with
national standards and the specifications in the safety instructions (XA’s).
Make sure you use the specified cable gland.
intrinsicaly safe
terminal
NRF 590 i.s. terminal board board
D+ RTD
16
S+
17
S-
18
D-
19
20 OPT1
21 OPT2 grounding single sided grounding not on
i.s. module wiring 7500 RTG
22 OPT3 on Tank Side Monitor
23 OPT4 4590 TSM
+
24 H
25 - 1 2 3 4
+ Internally
26 HART interconnected
H
27 - sensor as one HART
fieldbus loop
+ + plant
28 4
29
30
-H
+
P
- 3
+
2
1 2 3 4 5 ground
31 - - 1 1 2 3 4 5
For Micropilot
only for
S-series only!
7500 RTG
75xx_wiring_w_TSM.
The 7500 RTG is - possibly in combination with other devices - connected to a Tank
Side Monitor in a hazardous area. In this case, it is recommended that you ground
the cable screen centrally at the Tank Side Monitor and connect all devices to the
same potential matching line (PML). If, for functional reasons, a capacitive coupling is
required between local earth and screen (multiple grounding), ceramic condensers
with a dielectric strength of min. 1500 Veff must be used, whereby the total capacitance
of 10 nF must not be exceeded. Notes on grounding interconnected intrinsically safe
devices are provided by the FISCO model
27
Wiring Radar Tank Gauge
L00-FMR53xxx-04-00-00-en-001
Load HART
Minimum load for Hart communication: 250 Ω
Cable entry
Cable gland: M20x1.5 or Pg13.5
Cable entry: G ½ or ½ NPT
Supply voltage
Direct current voltage: 16…36 VDC
.
Communication Terminal voltage minimal maximal
Ex U (20 mA) = 16 V 30 V
Power consumption
Max. 330 mW at 16 V, max. 500 mW at 24 V, max. 600 mW at 30 V.
Current consumption
Max. 21 mA (50 mA inrush current).
Power supply
For stand alone operation recommended via e.g. Varec RN 221N.
mm accuracy
For measurements with mm accuracy the measured variable must be transmitted
using HART protocol to ensure the necessary resolution.
- ToF Tool
FXA 193
RN 221N
power supply
4...20 mA
signal
L00-FMR53xxx-04-00-00-en-004
dsdmdm
df das.
asdas fa
asas la.
DELTABAR: * * * * * * * *
ONLINE
1 QUICK SETUP
2 OPERATING MENU
3 PV 352 mbar
4 SV 0 ˚C
HELP SAVE
1 2 3 4 5
9 Page
Up
6 - COMMUWIN II
- ToF Tool
Bksp Delete
Page
On
PQRS TUV W XY Z
7 8 9
,()‘ _<> +*/
. 0 -
HART
375
FIELD COMMUNICATOR
- ToF Tool
4...20 mA
signal
≥ 250 Ω
L00-FMR53xxx-04-00-00-en-005
dsdmdm
df das.
asdas fa
asas la.
DELTABAR: * * * * * * * *
ONLINE
1 QUICK SETUP
2 OPERATING MENU
3 PV 352 mbar
4 SV 0 °C
HELP SAVE
1 2 3 4 5
9 Page
Up
6 - COMMUWIN II
- ToF Tool
Bksp Delete
Page
On
PQRS TUV W XY Z
7 8 9
,()‘ _<> +*/
. 0 -
HART
375
FIELD COMMUNICATOR
Caution!
If the HART communication resistor is not built into the supply unit, it is necessary to
insert a communication resistor of 250 Ω into the 2-wire line.
29
Wiring Radar Tank Gauge
Caution!
In Ex applications, the instrument must only be grounded on the sensor side. Further
safety instructions are given in the separate documentation for applications in explosion
hazardous areas.
5 Operation
31
Operation Radar Tank Gauge
– + E
X
F
2x X
X X >3 s
F F ...
O
O O
...
S S
S ...
If, for example, the application of the instrument is to be changed, carry out the follow-
ing procedure:
L00-FMR2xxxx-07-00-00-en-005
The first two digits identify the function group:
• basic setup 00
• safety settings 01
• linearization 04
…
The third digit numbers the individual functions within the function group:
• basic setup 00 → • tank shape 002
• medium property 003
• process cond. 004
…
Here after the position is always given in brackets (e.g. "tank shape" (002)) after the
described function.
33
Operation Radar Tank Gauge
L00-FMR53xxx-07-00-00-en-003
Figure 5: Layout of the display and operating elements
5.2.1 Display
Liquid crystal display (LCD):
Four lines with 20 characters each. Display contrast adjustable through key
combination.
Selection list
L00-FMRxxxxx-07-00-00-en-003
Help text
Envelope
curve
Figure 6: Display
Symbols Meaning
ALARM_SYMBOL
This alarm symbol appears when the instrument is in an alarm state. If the symbol
flashes, this indicates a warning.
LOCK_SYMBOL
This lock symbol appears when the instrument is locked,i.e. if no input is possible.
COM_SYMBOL
This communication symbol appears when a data transmission via e.g. HART,
PFOFIBUS-PA or Foundation Fieldbus is in progress.
LED Meaning
35
Operation Radar Tank Gauge
Key(s) Meaning
O and F
or Contrast settings of the LCD
S and F
Software reliability
The software used in the 7500 RTG fulfills the requirements of
OIML R85. This particularly includes:
• cyclical test of data consistency
• non-volatile memory
• segmented data storage
The 7500 RTG continuously monitor the compliance with accuracy requirements for
custody transfer measurements according to OIML R85. If the accuracy cannot be
maintained, a specific alarm is generated on the local display and via the digital
communication
Hardware lock:
The instrument is locked by pressing the O and S and F keys at the same time.
The lock is shown on the display by the symbol and can only be unlocked again
via the display by pressing the O and S and F keys at the same time again. It is not
possible to unlock the hardware by communication.
All parameters can de displayed even if the instrument is locked.
– + E
⇒
⇓ O and S and F press simultaneous
⇓
The LOCK_SYMBOL appears on the LCD.
37
Operation Radar Tank Gauge
Hardware lock:
After pressing the O and S and F keys at the same time, the user is asked to enter the
unlock parameter
– + E
⇒ O and S and F press simultaneous
⇓
Please enter unlock code and confirm with F.
Caution!
Changing certain parameters such as all sensor characteristics, for example, influences
numerous functions of the entire measuring system, particularly measuring accuracy.
There is no need to change these parameters under normal circumstances and
consequently, they are protected by a special code known only to the Varec service
organization. Please contact Varec if you have any questions.
Caution!
A reset sets the instrument back to the factory settings. This can lead to an impairment
of the measurement. Generally, you should perform a basic setup again following a
reset.
– + E
⇒
User input ("reset" (0A3)):
• 333 = customer parameters
• 555 = History
The tank map can also be reset in the "cust. tank map" (055) function of the
"extended calibr." (05) function group.
This reset is recommended whenever an instrument with an unknown 'history' is to be
used in an application or if a faulty mapping was started:
• The tank map is deleted. The mapping must be recommenced.
39
Operation Radar Tank Gauge
– + E
⇒
Error messages
Error messages appear as four lines of plain text on the display. In addition, a unique
error code is also output. A description of the error codes is given on page 79.
• The "diagnostics (0A)" function group can display current errors as well as the last
errors that occurred.
• If several current errors occur, use O or S to page through the error messages.
• The last occurring error can be deleted in the "diagnostics (0A)" function group
with the function "clear last error" (0A2).
dsdmdm
df das.
asdas fa
asas la.
dsdmdm
df das.
asdas fa
asas la.
FMR531: LIC0001
FMR531: LIC0001 ONLINE
ONLINE
1 GROUP SELECTION
1 GROUP SELECT
2 PV 8.7 m 2 PV 8.7 m Delete
HELP SAVE
H dsdmdm
df das.
asdas fa
asas la.
FMR531: LIC0001
6
9
HELP SAVE
Page GROUP SELECTION
Up
1 CALIBRATION
Bksp Delete 2 LINEARISATION Delete
3 EXT. CALIBRATION
Page 4 SERVICE
On Bksp 5 OPERATING MODE
#%& ABC DEF
1 2 3
L00-FMR53xxx-07-00-00-xx-001
Copy
GHI
Paste
JKL
Hot Key
MNO
H dsdmdm
df das.
asdas fa
asas la.
Note!
• Further information on the HART handheld unit is given in the respective operating
manual included in the transport bag of the instrument.
41
Operation Radar Tank Gauge
Note!
Further information you may find on the CD-ROM, which is enclosed to the instrument.
Menu-guided commissioning
Connection options:
• Service-interface with adapter FXA 193 (see page 29)
• HART with Commubox FXA 191 (see page 29)
Note!
Further information on Commuwin II is given in the following Varec documentation:
• System Information: SI 018F/00/en “Commuwin II”
• Operating Manual: BA 124F/00/en “Commuwin II” operating program
Connection
The table provides an overview of the Commuwin connections.
Interface Hardware Server Device list
HART Commubox FXA 191 to HART HART Connected instrument
Computer with RS-232C interface
Interface FXN 672 ZA 673 for List of all rack bus modules:
Gateway for MODBUS, PROFIBUS, PROFIBUS the required FXN 672 must
FIP, INTERBUS, etc. ZA 672 for be selected
Computer with other
RS-232C interface or
PROFIBUS card
Note!
The 7500 RTG can also be operated locally using the keys. If operation is prevented by
the keys being locked locally, parameter entry via communication is not possible either.
43
Operation Radar Tank Gauge
6 Commissioning
6.2 Commissioning
– + E
⇒
⇓ After 5 s, the following message appears
⇓
After 5 s or after you have pressed F the follow-
ing message appears
⇓
Select the basic unit
(this message appears the first time the instrument
is switched on)
⇓
The current measured value is displayed
45
Commissioning Radar Tank Gauge
Commissioning
flange:
reference point of
measurement
BD
measuring cond.
SD
D
basic setup (standard)
measuring cond.
empty calibr. F
full calibr. L
pipe diameter
(for bypass/stilling well)
Automatic
mounting
linearisation
option
extended calibr.
...
47
Commissioning Radar Tank Gauge
Note!
• The instrument continues to measure while data entry is in progress, i.e. the current
measured values are output via the signal outputs in the normal way.
• If the envelope curve mode is active on the display, the measured values are updated
in a slower cycle time. Thus, it is advisable to leave the envelope curve mode after the
measuring point has been optimized.
• If the power supply fails, all preset and parameterized values remain safely stored in
the EEPROM.
Note!
All functions are described in detail, as is the overview of the operating menu itself, in
the 7500 RTG "Service manual and description of instrument functions".
– + E
⇒
This function displays the current measured value in the selected unit
(see "customer unit" (042) function). The number of digits after decimal point can be
selected in the "no.of decimals" (095) function. The length of the bargraph corresponds
to the percentile value of the present measured value with regard to the span.
– + E
⇒
Function "tank shape" (002)
– + E
⇒
This function is used to select the tank shape.
Note!
For the application of the 7532 RTG the selection "stilling well" has to be chosen in the
function "tank shape" (002).
Selection:
• dome ceiling
• horizontal cyl
• bypass
• stilling well
• flat ceiling (Typical ceiling of storage tanks: a slight slope of only a few degrees can
be neglected.)
• sphere
L00-FMR2xxxx-14-00-00-en-007
49
Commissioning Radar Tank Gauge
– + E
⇒
This function is used to select the dielectric constant.
Selection:
• unknown
• < 1.9
• 1.9 ... 4
• 4 ... 10
• > 10
3)Treat Ammonia NH3 as a medium of group A, i.e. always use a stilling well.
– + E
⇒
This function is used to select the process conditions.
Selection:
• standard
• calm surface
• turb. surface
• agitator
• fast change
• test:no filter
Note!
The phase evaluation of the 7500 RTG (see »Function "auto-correction" (031)« on
page 57) is only activated if you select the measuring conditions "standard" or
"calm surface". We strongly recommend that, in the case of rough product surfaces or
rapid filling, you activate the appropriate application parameters.
51
Commissioning Radar Tank Gauge
– + E
⇒
This function is used to enter the distance from the flange (reference point of the
measurement) to the minimum level (=zero).
L00-FMR2xxxx-14-00-06-en-008
Caution!
For dish bottoms or conical outlets, the zero point should be no lower than the point at
which the radar beam hits the bottom of the tank.
– + E
⇒
This function is used to enter the distance from the minimum level to the maximum level
(=span).
L00-FMR2xxxx-14-00-06-en-009
Note!
If bypass or stilling well was selected in the "tank shape" (002) function, the pipe
diameter is requested in the following step.
– + E
⇒
This function is used to enter the pipe diameter of the stilling well or bypass pipe.
100% 100%
L00-FMR532xx-14-00-00-xx-001
0%
0%
ø ø d2
Microwaves propagate slower in pipes than in free space. This effect depends on the
inside diameter of the pipe and is automatically taken into account by the 7500 RTG. It
is only necessary to enter the pipe diameter for applications in a bypass or stilling well.
If mounting the 7532 RTG on stilling wells with a widening of the pipe, the inner diam-
eter of the lower part of the pipe (d2 in the Fig.) must be entered. This is the part of the
stilling well, where the measurement is actually performed.
– + E
⇒
The distance measured from the reference point to the product surface and the level
calculated with the aid of the empty adjustment are displayed. Check whether the values
correspond to the actual level or the actual distance. The following cases can occur:
• Distance correct – level correct -> continue with the next function, "check
distance" (051)
• Distance correct – level incorrect -> Check "empty calibr." (005)
• Distance incorrect – level incorrect -> continue with the next function, "check
distance" (051)
53
Commissioning Radar Tank Gauge
– + E
⇒
This function triggers the mapping of interference echoes. To do so, the measured
distance must be compared with the actual distance to the product surface. The
following options are available for selection:
Selection:
• distance = ok
• dist. too small
• dist. too big
• dist. unknown
• manual
L00-FMR2xxxx-14-00-06-en-010
distance = ok
• mapping is carried out up to the currently measured echo
• The range to be suppressed is suggested in the "range of mapping (052)" function
Anyway, it is wise to carry out a mapping even in this case.
dist. unknown
If the actual distance is not known, no mapping can be carried out.
manual
A mapping is also possible by manual entry of the range to be suppressed. This entry is
made in the "range of mapping (052)" function.
Caution!
The range of mapping must end 0.5 m (20") before the echo of the actual level. For an
empty tank, do not enter E, but E – 0.5 m (20").
– + E
⇒
This function displays the suggested range of mapping. The reference point is always
the reference point of the measurement (see page 46). This value can be edited by the
operator.
For manual mapping, the default value is: 0 m.
– + E
⇒
This function is used to start the interference echo mapping up to the distance given in
"range of mapping" (052).
Selection:
• off: no mapping is carried out
• on: mapping is started
55
Commissioning Radar Tank Gauge
– + E
⇒
The distance measured from the reference point to the product surface and the level
calculated with the aid of the empty alignment are displayed again. Check whether the
values correspond to the actual level or the actual distance. The following cases can
occur:
• Distance correct – level correct -> basic setup completed
• Distance incorrect – level incorrect -> a further interference echo mapping must be
carried out "check distance" (051).
• Distance correct – level incorrect -> check "empty calibr." (005)
– + E
⇒
By this function a history reset of the device is performed, i.e. the correspondence table
between level an index values is deleted. A new correspondence table will be filled and
stored after the history reset, see page 57.
Caution!
Perform only after first installation (see »Function "auto-correction" (031)« on page 57).
In this case also effect a reset of the dip table in function "dip table mode" (033).
– + E
⇒
⇓ After 3 s, the following message appears
Note!
After basic calibration, it is wise to evaluate the measurement using the envelope curve
(function group "display" (09)).
– + E
⇒
Function "tank gauging" (030)
– +
⇒
E
Using this function, you can either enter a dip table or carry out an auto-correction.
– + E
⇒
When measuring levels with radar systems, so-called "multipath reflections" can affect
the level signal giving rise to serious measuring errors. "Multipath reflections" also
include radar beams which are received by the radar system, which have not been
reflected directly by the medium surface. They may reach the antenna via the basin wall
and the medium surface. This phenomenon is particularly noticeable with devices
mounted near to walls, as soon as the conical radar beam strikes the basin wall.
The 7500 RTG can automatically discover and correct measuring errors due to this
"multiple path" propagation. This is because it uses two independent sets of information
when evaluating reflection signals:
• Firstly, it evaluates the amplitude of the reflected energy using the so-called envelope
curve system.
• Secondly, it evaluates the phase of the reflected energy.
The decisive factor for a constant output signal is to assign the phase values to the
associated level values. This assignment is ensured using a correspondence table (index
correction table). The 7500 RTG learns this for the specific application after installation
(learning period).
Therefore, after mounting the device, and after completing the basic calibration, a
history reset must be performed (enter "yes" in the "history reset" (009) function in the
"basic setup " (00) function group).
Do not switch off the radar system during filling and emptying operations during the
teach-in phase. Switching off when there are only negligible level changes produces no
error.
" Caution!
During the learning period, fast filling/emptying or turbulent surfaces can result in
switching off and on the phase evaluation. Subsequently observed measurement errors
will disappear as soon as tank levels come back to areas measured by 7500 RTG
previously with activated phase evaluation. If the observed measurement errors are
curetted by dip table entries, the 7500 RTG will take care of these corrections and
automatically adjust the index correction table. Do NOT correct any settings in the basic
calibration or the extended calibration.
Note!
Immediately after installation, the 7500 RTG measures with the specified mm-accuracy.
Until the level range has been completely covered by the medium (setting up the
57
Commissioning Radar Tank Gauge
– + E
⇒
For level measurement in stilling wells, radar systems require highly precise pipe inner
diameter data. An mm-exact level measurement cannot be guaranteed for deviations
from the actual stilling well inner diameter of more than ± 0.1mm to the value entered
in the function group "basic setup" (00). The errors which occur as a result are linear and
can be corrected with a dip table containing at least two entries.
The 7500 RTG also has an automatic pipe inner diameter correction. This adjusts the
entered stilling well inner diameter (input in the function group
"basic setup" (00)) to the actual values. However, this presupposes that the value entered
in the function group "basic setup" (00) matches the actual pipe inner diameter
accurately as possible. The user-defined value entered in the function group
"basic setup" (00) can be corrected with this value. In order to do this, switch on the "pipe
diam. corr." (032) function, after a level change of at least 5 m has occurred since start-
up. The pipe diameter, which the instrument determines automatically, will then be
transmitted to the "pipe diameter" (007) function.
Note!
Only if the "pipe diameter" (007) function has changed its value, it is necessary to per-
form a "history reset" (009) and to delete the dip table after activation of the "pipe diam.
corr." (032) function. Otherwise the level change of 5 m has not yet been exceeded. The
"pipe diam. corr." (032) function must be deactivated again and the procedure should be
repeated at a later point of time.
– + E
⇒
This indicates the instrument calibration mode. The calibration mode (active) can be set
using the hardware security lock on the electronics (see Page 34).
Selection:
• inactive
• active pos.
• active neg.
active pos.
The custody mode (instrument is lead-sealed and accurate to the nearest mm) is active
and is held.
active neg.
Custody mode (instrument is lead-sealed and accurate to the nearest mm) is activated
and not held, e.g. because the signal-to-noise ratio is less than 10 dB (refer to
function "echo quality" (056) in the function group "extended calibr." (05)).
Caution!
After entering all the values and completing mounting and aligning work, enter the
Reset Code "555" in the function "history reset" (009) of the "basic setup" (00) function
group to reset the instrument history for auto-correction.
59
Commissioning Radar Tank Gauge
Dip table
The dip table is used to correct the level readings of the 7500 RTG using independently
taken hand dips. The dip table is used in particular to adapt the level gauge to the
specific application conditions as mechanical offset and tank/stilling well design.
Depending on national regulations, national inspectors will dip the tank at one to three
levels during a calibration run and check the level readings.
Only one value pair must be entered into the dip table to correct the measurement
offset.
If a second value pair is entered into the dip table, the 7500 RTG accepts the corrected
measured values identically for both value pairs. All other measured values are
determined by linear extrapolation.
If you enter more than two value pairs, the system carries out a linear interpolation
between adjacent value pairs. Outside these value pairs, extrapolation is also linear.
basic setup
history reset
dip table
alternatives
preferred choice:
hand dip and collect hand dip and
immediate measurement data over
offset correction in full measurement range
semi-automatic mode
L00-FMR53xxx-19-00-00-en-014
collect further hand dip data;
if deviation from measurements evaluate collected data;
> 4 mm and level change > 2 m: enter characteristic value
enter a new dip value pairs into the dip table
into dip table using the using the manual mode
semi-automatic mode
To collect and enter data into the dip table, two alternative procedures may be carried
out. In order not to mix up measurement values corrected by the offset or linearization
of the dip table with uncorrected measurement values, it is recommended to use the
semi-automatic mode of the dip table to enter new data pairs. In this case, the first dip
value should be entered immediately after the basic calibration. Further linearization
points should be entered only after a level change of at least 2 m (see Figure 8:,
preferred choice) and a deviation between the "uncorrected measurement value" and the
hand dip value of at least 4mm.
If this procedure can not be followed, then NO value pair should be entered into the dip
table after basic calibration. Measurement data and hand dip values should be collected
over the full measurement range and be evaluated with regard to a good linear fit. Only
then characteristic value pairs should be entered into the dip table using the "manual
mode" (see Figure 8:, right side). If further linearization is needed, further hand dip
values should be entered using only the "semi-automatic mode".
Note!
The offset should NOT be determined and entered within the close range of the antenna
(conf. definition of the safety distance) or immediately in the range of the tank bottom,
because within these ranges interferences of the radar signal may occur.
Note!
The dip table can be printed out using the ToF-Tool. Before doing this, the ToF Tool
must be reconnected to the instrument in order to update the values within the ToF Tool.
Note!
Make your inputs into the dip table in semi-automatic mode. We advise you to leave
"auto correction" (031) activated ("on") while you enter your inputs.
Caution!
After entering one or more points into the dip table, make sure that the dip table is
activated and left in the "table on" dip table mode.
– + E
⇒
This function displays the dip table status.
Display:
• table on
• table off
table on
Indicates whether the dip table is active.
table off
Indicates whether the dip table is not active.
– + E
⇒
The dip table can be switched on or off using this function.
Selection:
• manual
• semi-automatic
• table on
• table off
• clear table
• view
61
Commissioning Radar Tank Gauge
manual
The value pairs in the dip table can be read and written. You can enter the measured
value and the dip value.
–uncorrected measured value:
This is the measured value supplied by the instrument, NOT corrected by the dip
table. The choice of measured value, level or remaining fill height is dependent on
the instrument setting.
–Dip value:
This is the level or distance to flange respectively, given by the hand dip. This value
should be used to correct the measured value.
The "manual mode" of the dip table can be used to enter collected data after a series
of data pairs taken at different tank levels.
Note!
The bigger the distance between the different levels while taking hand dips, the more
accurate the linearization of the dip table will be.
semi-automatic
The value pairs in the dip table can be read. You can enter the dip value only. When there
are new value pairs, the current level or distance is accepted as the measured value.
table on
The dip table is switched on.
table off
The dip table is switched off.
clear table
The complete dip table is deleted. The table is switched off. The number of free table
entries is set to the maximum value (= 32).
View
The value pairs in the dip table can only be read. You can still select this menu option,
even if there is no dip table available. In this case, the number of free table entries is at
maximum value (= 32).
– + E
⇒
This function edits measured variable. The number behind the entry "remain" indicates
the current number of remaining free value pairs. The maximum number of value pairs
is 32; after each entry, the remaining number is decremented.
Note!
The uncorrected measured value is displayed in the "dip table" (034) function. This may
differ considerably from the measured values when a dip table is activated.
– + E
⇒
This function edits the dip value.
– + E
⇒
Use this function to enter the dip value (level or distance) which will correct the mea-
surement values.
Selection:
• new point
• edit point
• store point
• delete point
• return
• next point
• previous point
General procedure:
To enter a new point into the dip table, use
"new point" to enter the value (pairs),
"store point" to sort the new value (pairs),
"return" to go to the dip table mode and
"table on" to activate the dip table.
new point
You can enter a new point. A suggested value is displayed for the measured variable, the
dip value of the current level or the distance. With semi-automatic inputs, the current
level or the current remaining fill height is displayed as proposed dip value. The new
value pair can be altered without selecting the
"edit point" parameter.
If the table is full, you can still select this parameter. In this case, the number of free
table entries stands at minimum value (= 0).
edit point
The displayed value pair can be changed. Only the dip value can be changed with semi-
automatic input mode.
Caution!
To accept the value pair in the table, confirm it with "store point".
store point
The displayed value pair is sorted in the table.
Note!
For sorting, the following criteria must be met:
• Measured variables may not be equal but have different dip values.
• A measured variable available in the table is recognized as equal when it is closer than
1 mm to the sorting value.
• After successful sorting, the setting remains at "edit point" and the number of free
table entries is decremented.
Caution!
If the value cannot be sorted, the setting remains at the previous menu option. No
warning or error message is generated. However, the number of remaining table entries
is not decremented.
63
Commissioning Radar Tank Gauge
delete point
The currently displayed point is deleted from the table. After deletion, the previous point
is displayed. If the table only consisted of one point before deletion, then the current
measured variable is displayed as a value pair.
return
By selecting this point, you return to the function "dip table mode" (033).
next point
This scrolls down in the table. If the table is empty, you can still select this option.
However, the displayed value does not change.
previous point
This scrolls up in the table. If the table is empty, you can still select this option. However,
the displayed value does not change.
Caution!
After entering one or more points into the dip table, make sure that the dip table is
activated in the "table on" dip table mode.
– + E
⇒
Here you can select which information is shown on the display:
• envelope curve
• env. curve+FAC (for FAC see 7500 RTG "Service manual and description of
instrument functions")
• env. curve+suppress. (i.e. the tank map is also displayed)
L00-FMRxxxxx-07-00-00-en-004
Function "recording curve" (09B)
This function determines whether the envelope curve is read as:
• single curve
or
• cyclic
– + E
⇒
Note!
If the envelope curve mode is active on the display, the measured values are updated in
a slower cycle time. Thus, it is advisable to leave the envelope curve mode after the
measuring point has been optimized.
65
Commissioning Radar Tank Gauge
Note!
• Each parameter that is changed must be confirmed with the RETURN key!
• The "Next" button moves you to the next screen display:
67
Commissioning Radar Tank Gauge
Note!
You can also specify the pipe diameter in this display.
measured value
distance
Step 5/5:
After the first installation of the device, initialise the index correction table (compare
page 57) by activating the history reset 555.
69
Commissioning Radar Tank Gauge
Note!
• Each parameter that is changed must be confirmed with the RETURN key!
• the "Next" button moves you to the next screen display:
71
Commissioning Radar Tank Gauge
7 Maintenance
The 7500 RTG measuring instrument requires no special maintenance.
Exterior cleaning
When cleaning the exterior of measuring devices, always use cleaning agents that do not
attack the surface of the housing and the seals.
Replacement
After a complete 7500 RTG or electronic module has been replaced, the parameters can
be downloaded into the instrument again via the communication interface. Prerequisite
to this is that the data were uploaded to the PC beforehand using the ToF Tool /
Commuwin II.
Measurement can continue without having to carry out a new setup.
After an antenna component has been replaced, a new calibration must be carried out.
73
Maintenance Radar Tank Gauge
8 Accessories
Various accessories, which can be ordered separately from Varec, are available for the
7500 RTG.
L00-FMR53xxx-00-00-06-en-001
Commubox FXA 191 HART
For intrinsically safe communication with ToF Tool or Commuwin II via the RS 232C-
interface.
Commuwin II
Operating software for intelligent instruments.
75
Accessories Radar Tank Gauge
9 Troubleshooting
77
Troubleshooting Radar Tank Gauge
L00-FMR53xxx-19-00-00-en-010
W103 initialising - please E²PROM storage not yet wait some seconds; if
wait finished warning prevails, exchange
electronics
A116 download error checksum of stored data not restart download of data
repeat download correct
79
Troubleshooting Radar Tank Gauge
A270 custody switch switch for custody transfer check position of custody
undef may be defective switch
check position exchange electronics
W621 simulation ch. 1 on simulation mode is active switch off simulation mode
A671 linearization ch1 not linearization table is in edit activate linearization table
complete, not mode
usable
W681 current ch1 out of current out of range (3,8 mA check calibration and
range ... 21,5 mA) linearization
A warning or Depending on the configuration See table of 1. See table of error messages
alarm has error messages (see page 79)
occurred. (see page 79)
→ level/ullage (040)
→ max. scale (046)
expected
→ diameter vessel (047)
→ Check table
actual no ↓
E m/ft
4 mA/0% t→
Measurement in yes → 1. Is bypass or stilling well
bypass or stilling selected
well? in tank shape (002)?
no ↓
no ↓
expected
4 mA/0% t→
81
Troubleshooting Radar Tank Gauge
20 mA/100%
actual
expected
4 mA/0% t→
During filling/ Multiple echoes yes → 1. Check the tank shape (002),
emptying e.g. "dome ceiling" or
the measured "horizontal cyl"
20 mA/100%
value jumps 2. In the range of the blocking dist.
downwards actual
(059) there is no echo evalua-
tion
→ Adapt the value
expected
3. If possible, do not select
central installation position
4 mA/0% t→
4. Perhaps use a stilling well
4 mA/0% t→
E 641 (loss of If the instrument is configured to Hold noise level during Repeat once more empty calibr.
echo) after by loss of echo the output is set to any the initialisation (005).
turn on the value/current. phase to high. Caution!
power supply Before conformation change with
O or S to the edit mode.
Note!
You can order spare parts directly from your Varec service organization by giving the
order code and the serial number which is printed on the measuring transducer
nameplate (see page 7). The corresponding spare part number also appears on each
spare part. Installation instructions are given on the instruction card that is also
delivered.
Note!
If the calibration seal is broken, the national calibration authority should normally be
informed within 24 hours.
83
Troubleshooting Radar Tank Gauge
7532 RTG
L00-FMR532xx-00-00-06-en-001
O-ring
85,32 x 3,53 FKM (green)
85,32 x 3,53 HNBR12-70
(black)
O-ring
85,32 x 3,53 FKM (green)
85,32 x 3,53 HNBR12-70
(black)
L00-FMR532xx-00-00-06-en-004
Antenna size Seal
Modification nameplate
When ordering parts that are listed in the product structure (see page 8) , a check must
be made as to whether the instrument description on the nameplate is still valid, e.g. for:
• an antenna component,
• an electronics module,
• an RF module,
• a VU 331 operating and display module,
• a housing cover with window.
85
Troubleshooting Radar Tank Gauge
Housing T12
The complete order code must be specified when ordering the replacement housing so
that the correct nameplate can be delivered, e.g.
• 7532 RTG-AACWJAC1AA
Caution!
• It is not possible to convert a standard instrument into an Ex instrument by simply
exchanging parts.
• When repairing certified instruments, the relevant regulations must be followed.
• For FM approved instruments, it is forbidden to make any changes to the instrument
that are not expressly authorized in the operating manual. Contravening this
prohibition can invalidate the approval for operation of the instrument.
9.5 Return
If you need to send a 7500 RTG back to Varec for repair, please send a completed copy
of the form printed on the last page.
Please carry out the measures described in section 1.4, “Return” on page 5 before
sending a 7500 RTG back for repair.
9.6 Disposal
In case of disposal please separate the different components according to their material
consistence.
10 Technical data
Application
Application The 7500 RTG is used for highly accurate level measurement in
storage tanks and can be applied in custody transfer applications. It
meets the relevant requirements according to OIML R85 and API 3.1B.
• The 7532 RTG with planar antenna is specifically and only suited for
stilling well applications with ranges up to 38 m (124 ft).
Equipment architecture The 7500 RTG can be used for measurement in a stilling well as well
as in free space. The different instrument versions are applied as
follows:
• The 7532 RTG with planar antenna is the preferred device in stilling
wells ≥ 150 mm.
• The instruments are equipped with a passive 4…20 mA output with
HART protocol.
Input
Measured variable The measured variable is the distance between a reference point
(mounting flange) and a reflective surface (e.g. medium surface). The
measured value and all parameters are displayed using either metrical
SI-units or US/UK-units (inch, ft, …).
The level is calculated based on the tank height entered.
The level can be converted into other units (volume, mass) by means
of a linearization.
In order to compensate for non-linear effects like movement of the
tank roof, an additional correction table (diptable) can be entered.
Output
Signal on alarm Error information can be accessed via the following interfaces:
• Local display:
– Error symbol (see page 35)
– Plain text display
– LED’s: red LED continuously on = alarm, red LED flashes = warn-
ing
• Current output
• Digital interface
87
Technical data Radar Tank Gauge
Auxiliary energy
Power supply For stand alone operation recommended via e.g. Varec RN 221N/Z.
Performance characteristics
Proof of accuracy for The accuracy of each 7500 RTG is established through a calibration
custody transfer ver- certificate that records the absolute and relative error at 10
sions equidistant points during the final test. For stilling well measurements
with 7532 RTG, a NMI / PTB calibrated tape with an absolute accuracy
of 0.25 mm is used.
Each 7500 RTG is delivered with the PTB and NMi type approval.
Additional initial factory verifications for custody applications are
available on demand for all 7500 RTG.
Maximum fill speed By the first pass trough of measuring range: 100 mm/min., after it
unlimited.
Hysteresis 0.3 mm
Software reliability The software used in the 7500 RTG fulfills the requirements of OIML
R85. This particularly includes:
• cyclical test of data consistency
• non-volatile memory
• segmented data storage
The 7500 RTG continuously monitor the compliance with accuracy
requirements for custody transfer measurements according to OIML
R85. If the accuracy cannot be maintained, a specific alarm is
generated on the local display and via the digital communication.
Inventory Control Ver- All device types can be delivered as "Inventory Control Versions" with a
sions reduced accuracy of ± 3mm (under reference conditions).
To these versions, the calibration certificate or custody transfer type
approval is NOT attached.
The "Inventory Control Versions" can be selected by choosing the
option »R« in feature »70« in the order code section »Custody transfer
approvals« on page 8.
Operating conditions
Operating conditions
Environment
Process conditions
89
Technical data Radar Tank Gauge
Mechanical construction
Human interface
CE approval The measuring system meets the legal requirements of the EC-
guidelines. Varec confirms the instrument passing the required tests
by attaching the CE-mark.
Ex approval XA 081F-A
Installation 7500 RTG (T12 / EEx ia IIC T6…T1)
PTB 00 ATEX 2067 X, Equipment marking: (II 1/2 G)
Ordering Information
Accessories
see page 75
Supplementary Documentation
11 Appendix
91
Appendix Radar Tank Gauge
basic setup 00 tank shape 002 medium property 003 process cond. 004 empty calibr. 005 full calibr. 006
dome ceiling unknown standard enter value enter value
horizontal cyl. DK: < 1.9 calm surface
bypass DK: 1.9 … 4 turb. surface
stilling well DK: 4 … 10 add. agitator
flat ceiling DK: > 10 fast change pipe diameter 007
sphere test: no filter enter value
safety settings 01 output on alarm 010 outp. echo loss 012 delay time 014
MIN -10% 3.6mA alarm in case of echo loss
MAX 110% 22mA hold max. 4000 sec.
hold default: 30 s
user specific output on alarm 011 ramp %MB/min ramp %/min 013
enter value enter value
extended calibr. 05 selection 050 dist./meas value 008 check distance 051 range of mapping 052 start mapping 053
mapping D and L distance = ok input of off
are displayed dist. too small mapping range on
manual
dist. unknown
dist. too big
common
extended map..
pres. Map dist 054 cust. Tank map 055
is displayed inactive
active
reset
output 06 commun. address 060 no. of preambels 061 low output limit 062 fixed current 063
off off
on on fixed cur. value 064
display 09 language 092 back to home 093 format display 094 no. of decimals 095
enter time decimal x
default: 100 s 1/16” x.x
x.xx
x.xxx
diagnostics 0A present error 0A0 previous error 0A1 clear last error 0A2 reset 0A3 unlock parameter 0A4
for reset code for reset code
see manual see manual
system parameter 0C tag no. 0C0 protocol+sw-no. 0C2 software no. 0C3 serial no. 0C4
custody transfer
dist./meas.value 008 check distance 051 range of mapping 052 start mapping 053 history reset 009
distance = ok input of off no
dist. too small mapping range on yes
D and L
dist. too big
are displayed
manual
dist. unknown
safety distance 015 in safety dist. 016 ackn. alarm 017 overspill protection 018
from blocking alarm no standard
distance warning yes german WHG
default: 0.1m self holding
dist./meas.value 008
D and L
are displayed
echo quality 056 offset 057 output damping 058 blocking dist. 059
is displayed will be added to the enter value is displayed
measured level default: 5 s
sim. off
sim. level simulation value 066
sim. volume
sim. current
sep. character 096 display test 097 plot settings 09A recording curve 09B
. point off envelope curve single curve
, comma on env.curve+FAC cyclic
env.curve+cust.map
measured dist. 0A5 measured level 0A6 application par. 0A8 custody mode 0A9
not modified inaktiv
modified aktiv positiv
aktiv negativ
93
Appendix Radar Tank Gauge
00 basic setup V0 measured value tank shape medium property process cond. empty calibr. full calibr. pipe diameter
01 safety settings V1 output on alarm output on alarm outp. echo loss ramp %span/min delay time safety distance in safety dist. ackn. alarm overspill protection
03 mounting calibr. V2 tank gauging auto correction pipe diam. corr. dip table mode dip table dip table dip table handl. dip table state
04 linearisation V3 level/ullage linearisation customer unit table no. input level input volume max. scale diameter vessel
05 extended calibr. V4 check distance range of mapping start mapping pres. Map dist cust. Tank map echo quality offset output damping blocking dist.
06 output V5 commun. Address no. Of preambels low output limit fixed current fixed current simulation simulation value output current
09 display V6 language back to home format display no. of decimals sep. character
0D service V7
0A diagnostics V9 present error previous error clear last error reset unlock parameter measured dist. measured level application par. custody mode
0C system parameter VA tag no. protocol+sw-no. software no. serial no. distance unit download mode
Appendix
95
Appendix Radar Tank Gauge
Note!
A detailed description of the function groups, functions and parameters is given in the
the 7500 RTG "Service manual and description of instrument functions".
L00-FMR53xxx-15-00-00-en-001
Input
The reflected radar impulses are received by the antenna and transmitted into the
electronics. A microprocessor evaluates the signal and identifies the level echo caused
by the reflection of the radar impulse at the product surface. The unambiguous signal
identification is accomplished by the PulseMaster® software based on many years of
experience with time-of-flight technology. The mm-accuracy of the 7500 RTG is
achieved with the patented algorithms of the PhaseMaster® software.
The distance D to the product surface is proportional to the travel time t of the impulse:
D = c · t/2,
with c being the speed of light.
L=E–D
Reference point for "E" is the lower surface of the process connection.
The 7500 RTG is equipped with functions to suppress interference echoes. The user can
activate these functions. They ensure that interference echoes (e.g. from edges and weld
seams) are not interpreted as level echo.
97
Appendix Radar Tank Gauge
Output
The 7500 RTG is commissioned by entering an empty distance E (=zero), a full distance
F (=span), and an application parameter. The application parameter automatically
adapts the instrument to the measuring conditions. The data points “E” and “F”
correspond with 4mA and 20mA for instruments with current output. They correspond
with 0 % and 100% for digital outputs and the display module.
A linearization with max. 32 points, based on a table entered either manually or semi-
automatically, can be activated locally or remotely. This function provides a
measurement in engineering units and a linear output signal for spheres, horizontal
cylindrical tanks, and vessels with conical outlet.
SPS
Service-Adapter
Commubox
FXA 193
FXA 191
HART
HART
2x Messumformer- L00-FMR53xxx-14-00-06-en-002
Anzeige- und ENDRESS + HAUSER
speisegerät Bedienmodul %
RN 221 N VU 331
– + E
dsdmdm
df das.
asdas fa
asas la.
DELTABAR: * * * * * * * *
ONLINE
1 QUICK SETUP
2 OPERATING MENU
3 PV 352 mbar
4 SV 0 °C
HELP SAVE
6
9 Page
Up
Bksp Delete
Page
On
PQRS TUV W XY Z
7 8 9
,()‘
.
_<>
0
+*/
-
HART-Handbediengerät
FXA 191 375
FIELD COMMUNICATOR
DXR 375
oder
DXR 375
On-site operation:
• with display and operating module VU 331,
• with a Personal Computer, FXA 193 and the operating software ToF Tool.
The ToF Tool is a graphical operating software for instruments from Varec that operate
based on the time-of-flight principle (radar, ultrasonic, guided micro-impulse). It
assists with commissioning, securing data, troubleshooting, and documentation of the
measuring point.
Remote operation:
• with HART handheld DXR 35,
• with a Personal Computer, Commubox FXA 191 and the operating software
COMMUWIN II respectively ToF Tool.
• With a Personal Computer, TSM (Tank Side Monitor) and the operating software
FuelsManager.
L00-FMR53xxx-14-00-06-en-003
99
Appendix Radar Tank Gauge
The selected variable is the basis for the subsequent calculation of the current amount
of product in a tank, along with other measured variables such as (average) temperature
and pressure.
In addition to this, the accuracy of every single device is documented using a calibration
certificate, which is issued in the factory after the device has been tested on a reference
test rig.
On request, a separate initial verification of the devices can be carried out with a
National Standards Authorities inspector present, who issues a preliminary test certifi-
cate for every device. In the initial verification, the device is tested to ensure that it com-
plies with the limit of error in legal metrology, which lies at +-2mm for radar measuring
devices in Germany.
Essentially, this proves that the devices are weight and measure approved. The devices
must not, however, be used in custody transfer mode straight away.
The measuring device is not approved until after the approval after installation by the
Standards Authorities. For this, the device's level measurement is compared with the
tank gauging by a National Standards Authorities inspector using manual dips (also "Ini-
tial verification"). As a rule, a quiescent tank gauging is dipped by hand three times in a
row and then compared with the value displayed by the level radar. Depending on
national regulations, the transfer error limit, calculated as the arithmetic mean of the
absolute deviations of all three measurements, must not exceed double the limit of error
in legal metrology (compare, for example, the German "Eichordnung" or the American
"API 3.1B", in which the necessary procedures are also defined).
Depending on national regulations, this test is repeated with various tank gaugings.
Using linearization tables to compensate any non-linearities that occur in measurement
is permitted. For this, the 7500 RTG level radar offers a special dip table, compare
“Mounting calibration with VU 331” on page 57.
After the measurement has been approved by an inspector, he seals the level radar at
the stamp position and thereby also secures the programming status of the device.
7500 RTG radar devices continuously monitor the compliance with accuracy
requirements for custody transfer measurements according to OIML R85. If, for
example, the accuracy cannot be maintained due to quick surface movements, this is
reported via a separate alarm in the local display (displays "#" symbol) and via digital
communication.
101
Appendix Radar Tank Gauge
Fuels Manager
Software 7500 RTG
HART
4 wire
RTU 8130
(remote
terminal
unit) Tank Side Monitor
Supply voltage:
16…253 VAC
L00-FMR53xxx-14-00-06-en-004
Pressure
Index
A Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Alarm. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 G
Antenna extension . . . . . . . . . . . . . . . . . . . . . . . 75 Green LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Antenna size . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Group selection . . . . . . . . . . . . . . . . . . . . . . . . . 31
Approval after installation . . . . . . . . . . . . . . . . . 100
Auto-correction . . . . . . . . . . . . . . . . . . . . . . . . . . 57
H
Handheld unit DXR 275 . . . . . . . . . . . . . . . . . . . 41
B HART . . . . . . . . . . . . . . . . . . . . . . . . . . . 28–29, 41
Basic Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Housing T12. . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Basic Setup VU 331 . . . . . . . . . . . . . . . . . . . . . . 49
Blocking distance . . . . . . . . . . . . . . . . . . . . . . . . 16
I
Bypass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Initial verification. . . . . . . . . . . . . . . . . . . . . . . . 100
Installation in stilling well . . . . . . . . . . . . . . . . . . 11
C Installation in tank (free space) . . . . . . . . . . . . . 11
Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Interference echo . . . . . . . . . . . . . . . . . . . . . . . . 54
Calibration to regulatory standards disturbed . . . 35 Interference echo mapping . . . . . . . . . . 54–55, 68
CE approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
K
Check distance . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Key assignment . . . . . . . . . . . . . . . . . . . . . . . . . 36
Cleaning agents . . . . . . . . . . . . . . . . . . . . . . . . . 73 L
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46, 53
Communication symbol. . . . . . . . . . . . . . . . . . . . 35
Liquid crystal display (LCD) . . . . . . . . . . . . . . . . 34
Commuwin II . . . . . . . . . . . . . . . . . . . . . . . . . 43, 75
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Commuwin II-Operating Programm . . . . . . . . . . 43
Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Contrast settings . . . . . . . . . . . . . . . . . . . . . . . . . 36
Current consumption. . . . . . . . . . . . . . . . . . . . . . 28 M
Custody locking switch . . . . . . . . . . . . . . . . . 34, 36 Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54–55
Custody mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 Marker at instrument flange . . . . . . . . . . . . . . . . 17
Customer unit . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Maximum measured error . . . . . . . . . . . . . . . . . 88
Measure value actualisation. . . . . . . . . . . . . . . . 48
D
Measured value . . . . . . . . . . . . . . . . . . . . . . . . . 49
Declaration of conformity . . . . . . . . . . . . . . . . . . 10
Measured value display . . . . . . . . . . . . . . . . . . . 31
Degree of protection . . . . . . . . . . . . . . . . . . . . . . 30
Measuring principle . . . . . . . . . . . . . . . . . . . . . . 87
Diameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . 14
Dielectric constant. . . . . . . . . . . . . . . . . . . . . . . . 50
Medium conditions . . . . . . . . . . . . . . . . . . . . . . . 50
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Minimum level . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Dip Table Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Modification nameplate . . . . . . . . . . . . . . . . . . . 85
Dip table status . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Dip value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Mounting calibration . . . . . . . . . . . . . . . . . . . 57, 70
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Mounting kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Distance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46, 53
N
E
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Electrical symbols . . . . . . . . . . . . . . . . . . . . . . . . . 6
Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Empty calibration. . . . . . . . . . . . . . . . . . . . . . 46, 52
No.of decimals . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Envelope curve . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Notes on safety conventions and symbols. . . . . . 6
Error messages. . . . . . . . . . . . . . . . . . . . . . . . . . 40
O
F
Operating menu . . . . . . . . . . . . . . . . . . . . . . . . . 33
Foam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 31, 37
Foam surface . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Ordering structure . . . . . . . . . . . . . . . . . . . . . . . . 8
Full calibration . . . . . . . . . . . . . . . . . . . . . . . . 46, 52
Function groups . . . . . . . . . . . . . . . . . . . . . . . . . 33
IOM027FVAE0804 103
Radar Tank Gauge
Q
Quick wiring guide . . . . . . . . . . . . . . . . . . . . . . . 25
R
Range of mapping . . . . . . . . . . . . . . . . . . . . . . . 55
Reapproval. . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Recommendations for the stilling well . . . . . . . . 17
Red LED. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Reference point of measuring . . . . . . . . . . . . . . 46
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Reset customer parameters . . . . . . . . . . . . . . . . 39
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
RF approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
S
Safety conventions. . . . . . . . . . . . . . . . . . . . . . . . 6
Sample hatch . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . 10
Sealing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Select the language . . . . . . . . . . . . . . . . . . . . . . 45
Selection menu . . . . . . . . . . . . . . . . . . . . . . . . . 31
Service adapter FXA 193 . . . . . . . . . . . . . . . . . . 75
Slot width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Slots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Smallest possible measuring range . . . . . . . . . . 14
Software history . . . . . . . . . . . . . . . . . . . . . . . . . 86
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Stilling well . . . . . . . . . . . . . . . . . . . . . . . . . . 15, 53
Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Switching on the measuring device . . . . . . . . . . 45
T
Tank Gauging . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Tank shape. . . . . . . . . . . . . . . . . . . . . . . . . . 49, 67
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Terminal compartment . . . . . . . . . . . . . . . . . . . . 28
ToF Tool . . . . . . . . . . . . . . . . . . . 41, 66, 69–70, 91
Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Trouble-shooting . . . . . . . . . . . . . . . . . . . . . . . . 77
Trouble-shooting instructions . . . . . . . . . . . . . . . 77
Turn housing . . . . . . . . . . . . . . . . . . . . . . . . 11, 19
U
Unlock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Notes
IOM027FVAE0804 105
Radar Tank Gauge
www.varec.com
Varec, Inc.
5834 Peachtree Corners East, Norcross (Atlanta), GA 30092 USA
Tel: +1 (770) 447-9202 Fax: +1 (770) 662-8939