0% found this document useful (0 votes)
162 views64 pages

Genemapper ID v3.1 e v3.2

Uploaded by

Amanda Muniz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
162 views64 pages

Genemapper ID v3.1 e v3.2

Uploaded by

Amanda Muniz
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 64

User Bulletin

GeneMapper® ID Software
November 30, 2004

SUBJECT: Installation Procedures and New Features for


GeneMapper® ID Software v3.2

In This User This user bulletin includes the following topics:


Bulletin Section 1 Installation Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Section 2 New Features: All AmpFlSTR Kits . . . . . . . . . . . . . . . . 17
Section 3 New Features and Procedures: Yfiler Kit . . . . . . . . . . . . 25
Appendix A Verification Testing . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Appendix B Troubleshooting the Installation . . . . . . . . . . . . . . . . . 57

Overview GeneMapper® ID Software v3.1 has been upgraded to v3.2. This user
bulletin:
• Provides GeneMapper ID Software v3.2 installation
requirements and procedures
• Describes the new features in GeneMapper ID Software v3.2
and, where applicable, provides procedures for using them
• Describes verification testing of GeneMapper ID Software v3.2

DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

2 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Section 1 Installation Procedures

Section 1 Installation Procedures

This section covers:


Supported Instruments and AmpFlSTR Kits . . . . . . . . . . . . . . . . . . 4
Installation Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installation Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Installation Procedures and New Features for GeneMapper® ID Software v3.2 3


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Supported Instruments and AmpFlSTR Kits


Instruments GeneMapper ID Software v3.2 is compatible with the instruments,
operating systems, and Data Collection software listed in the table
below. Compatibility is defined as the ability of the GeneMapper ID
software to perform at least one of the following functions:
A. Analyze sample files generated from the instrument
B. Coexist with the instrument’s Data Collection software
C. Run concurrently with the instrument’s Data Collection software
D. Work with the automation feature (that is, autoanalysis) in the
instrument’s Data Collection software

Data Functions (as listed above)


Computer Operating
Instrument Collection
System
Software A B C D

ABI PRISM ® 377 Macintosh® a 2.6 X — — —


DNA Sequencer
Microsoft® Windows NT® 3.0 X X X —

ABI PRISM ® 310 Macintosh® a 2.1 X — — —


Genetic Analyzer
Microsoft® Windows NT® v3.0 X X X —

Microsoft® Windows® 2000 v3.0 X X X —

ABI PRISM ® 3100 Microsoft® Windows NT® v1.1 X — — —


Genetic Analyzer
Microsoft® Windows® 2000 v2.0 X X X X

ABI PRISM ® Microsoft® Windows NT® v1.0 X — — —


3100-Avant
Genetic Analyzer Microsoft® Windows® 2000 v2.0 X X X X

a. Data generated from a 310 Genetic Analyzer or 377 DNA Sequencer using a Macintosh® platform must be
converted to a Microsoft® Windows®-based format before they can be used with the GeneMapper ID
software. For information on converting Macintosh® sample files, refer to the GeneMapper™ ID Software
Version 3.1 Human Identification Analysis User Guide (PN 4338775).

4 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Supported Instruments and AmpFlSTR Kits

Chemistry Kits GeneMapper ID Software v3.2 is specifically designed to work with


the following AmpFlSTR® kits:
• AmpFlSTR® Yfiler™ PCR Amplification Kit
• AmpFlSTR® Identifiler® PCR Amplification Kit
• AmpFlSTR® SGM Plus® PCR Amplification Kit
• AmpFlSTR® Profiler Plus® ID PCR Amplification Kit
• AmpFlSTR® Profiler Plus® PCR Amplification Kit
• AmpFlSTR® COfiler® PCR Amplification Kit
• AmpFlSTR® SEfiler™ PCR Amplification Kit
• AmpFlSTR® Profiler® PCR Amplification Kit
• AmpFlSTR Blue™ PCR Amplification Kit
• AmpFlSTR Green™ I PCR Amplification Kit

Installation Procedures and New Features for GeneMapper® ID Software v3.2 5


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Installation Requirements
Computer The computer specifications for GeneMapper ID Software v3.2 are
Specifications listed in the table below. Please note the following:
• The Minimum Requirements column lists the lowest
specifications that permit the installer to install the
GeneMapper ID software. The minimum requirements may not
provide optimal performance.
• The Recommended Requirements column lists the requirements
that are recommended by Applied Biosystems.

System
Minimum Requirements Recommended Requirements
Component

Computer • Intel Pentium® III processor, • Intel Pentium® IV processor,


540 MHz ≥2.8 GHz
• 256 MB of RAM • 512 MB of RAM
• 2-GB hard drive (free space) • 40-GB EIDE hard drive
• 20/48X IDE CD-ROM
• 10/100 NIC with RWV (internal)

Monitor 800 × 600 pixels size 1024 × 768 or higher pixels size
17-inch monitor 19-inch or larger monitor

Operating System Microsoft Windows NT® version 4.0 • Microsoft Windows® 2000
(Service Pack 5 or higher) Professional (Service Pack 4)
• Microsoft Windows® XP
Professional (Service Pack 1)

Ethernet Capability • Network card for Oracle® • Network card for Oracle®
installation installation
• TCP/IP must be installed prior to • TCP/IP must be installed prior to
Oracle installation Oracle installation

Special Considerations
• The GeneMapper ID software does not run correctly on dual-
processor Pentium® computers. A single-processor computer is
required.

6 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Installation Requirements

• GeneMapper ID Software v3.2 runs on Windows NT®,


Windows® 2000, and Windows® XP platforms only. Conversion
utilities are included for analyzing data from Macintosh®-based
sample files. For conversion procedures, refer to the
GeneMapper™ ID Software Version 3.1 Human Identification
Analysis User Guide (PN 4338775).
• The GeneMapper ID Software v3.2 automated installer operates
only on computers running Microsoft Internet Explorer. If you
are using another browser, you will need to manually launch the
installer; instructions are provided in step 2 on page 12.

Compatibility The GeneMapper ID software uses an Oracle® database.


with Oracle GeneMapper ID Software v3.2 is compatible only with the Oracle
Databases database installed with ABI PRISM ® 3100/3100-Avant Genetic
Analyzers Data Collection Software v2.0.
Be sure there is no other Oracle client or Oracle server on the
computer. GeneMapper ID Software v3.2 is not supported on any
other Oracle client or server configurations.

Special Considerations
• The version of the Oracle database in GeneMapper ID Software
v3.2 is an embedded license for use by five named users.
IMPORTANT! To accommodate more than five users, additional
GeneMapper ID software or Oracle database licenses must be
purchased.
• TCP/IP has to be installed before installing GeneMapper ID
Software v3.2.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 7


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Compatibility The table below describes compatibility of GeneMapper ID


with 3100/3100- Software v3.2 with the Data Collection software for the
Avant Genetic ABI PRISM ® 3100/3100-Avant Genetic Analyzers.
Analyzers
3100 Genetic 3100-Avant
Analyzer Genetic Analyzer Compatible with
Data Collection Data Collection Instrument Computer?
Version Version

1.1 — No. a

— 1.0

2.0 2.0 Yes

a. GeneMapper ID Software v3.2 can only exist with the Oracle database
installed by 3100/3100-Avant Genetic Analyzer Data Collection Software
v2.0.

IMPORTANT! To install GeneMapper ID Software v3.2 on a


computer connected to a 3100/3100-Avant Genetic Analyzer that
uses Data Collection Software v2.0, the Data Collection software
must be running. If the Data Collection software is not running, the
GeneMapper ID software does not register with the Data Service
software.

Login When installing GeneMapper ID Software v3.2, you must:


Requirements • Have Administrator privileges on the local computer (that is,
have complete and unrestricted access to the local computer)
• Log in to the local computer (not a network domain)
• Be sure there is no other Oracle client or Oracle server on the
computer. GeneMapper ID Software v3.2 is not supported on
these configurations.

8 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Installation Requirements

PDF File Reader To open the user documentation included on the GeneMapper® ID
Software v3.2 installation CD, use a program that reads PDF files,
such as the Adobe Acrobat® Reader software.
The GeneMapper ID Software v3.2 installer does not install Acrobat
Reader. If you do not have a program that reads PDF files, you can
download Acrobat Reader from either the Adobe or Applied
Biosystems websites:
• www.adobe.com
• www.appliedbiosystems.com/support/software.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 9


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Installation Procedures
What Gets The installer on the GeneMapper® ID Software v3.2 CD installs:
Installed • GeneMapper® ID Software v3.2
• Oracle® database
• Panel folder that contains AmpFlSTR® kit panels and bins
• Microsatellite and SNP genotyping example data
• Electronic (PDF) versions of the user guide and tutorial
• Size standard folder that contains various size standards
• Default plot and table settings

Before You Begin Before you begin the installation, be sure all requirements are met.
the Installation See the references listed below.
• “Computer Specifications” on page 6
• “Compatibility with Oracle Databases” on page 7
• “Compatibility with 3100/3100-Avant Genetic Analyzers” on
page 8
• “Login Requirements” on page 8
• “PDF File Reader” on page 9

Installation Time The total time required to install the GeneMapper ID software is
approximately 60 minutes. This time may vary, depending on your
system configuration.

Procedure If necessary, remove previous versions of the


Flowchart GeneMapper® Software (page 11).

Install or upgrade the GeneMapper ID software


(page 11).
Installation
Procedures

Register the GeneMapper ID software (page 14).

Launch the GeneMapper ID software (page 14).

10 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Installation Procedures

Remove If you have either GeneMapper® Software v3.0 or v3.5 on your


Non-Validated computer, you must remove it before installing the GeneMapper ID
GeneMapper software.
Software To remove GeneMapper software v3.0 or v3.5:
(If Necessary)
1. Log in to the local computer as an Administrator.

2. Select Start > Control Panel to open the Control Panel


window.

3. Double-click Add or Remove Programs to open the Add or


Remove Programs dialog box.

4. Select the GeneMapper software program, then follow the


prompts to uninstall/remove the GeneMapper software.

5. If prompted to do so, restart the computer.

6. Continue with “Install the Software” below.

Install the To install GeneMapper ID Software v3.2:


Software
1. Log in to the local computer as an Administrator.
IMPORTANT! If you do not have Administrator privileges,
log off of the computer and log back in as a user with
Administrator privileges.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 11


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

To install GeneMapper ID Software v3.2: (continued)

2. Insert the GeneMapper® ID Software v3.2 CD into the CD-


ROM drive.
The installer launches automatically, displaying the Main
Menu.

Note: If the installer does not launch automatically:


a. On the desktop, right-click the My Computer icon and
select Explore to open the My Computer window.
b. Double-click the CD drive to open the
GeneMapper ID v3.2 folder.
c. Double-click the Setup.exe file to launch the installer.

3. Click Installation Readme to view the


InstallationReadme.txt file on the GeneMapper® ID
Software v3.2 CD.
Note: The InstallationReadme.txt file may contain
installation information that is more current than the
information in this user bulletin.

12 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Installation Procedures

To install GeneMapper ID Software v3.2: (continued)

4. Click Install GeneMapper® ID Software v3.2. to start the


InstallShield® Wizard.

5. Follow the onscreen instructions to install the software.


IMPORTANT! During installation, DOS commands are
executed and the DOS window is opened. Do not delete,
close, or click in the DOS window. If you click in it, press the
Esc key.

6. When the installation is complete, click Finish.

7. If prompted to do so, restart the computer.

8. Continue with “Register the Software” on page 14.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 13


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Register the To register GeneMapper ID Software v3.2:


Software
1. Locate the registration code on the Getting Started Card in
the software package.

2. IMPORTANT! Keep the registration code in a place where


you can easily retrieve it. You will need the registration code
the first time you start the GeneMapper ID software (below).
In addition, if you need to reinstall the software at any time,
you are prompted for the registration code again.

3. Complete the registration card and return it to Applied


Biosystems.

4. Continue with “Start the Software” below.

Start the To start GeneMapper ID Software v3.2 for the first time:
Software
1. Log in to the local computer.

2. On the desktop, click Start > Programs > Applied


Biosystems > GeneMapper > GeneMapper ID v3.2.
The first time you start the GeneMapper ID software, the
Product Registration dialog box opens.

Note: If the GeneMapper ID software fails to start, see


“Troubleshooting the Installation” on page 57.

14 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Installation Procedures

To start GeneMapper ID Software v3.2 for the first time:

3. Register the software:


a. In the Product Registration dialog box, enter your name,
organization, and registration code.
b. Click OK.
The software verifies the registration code, then displays the
Login dialog box.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 15


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

To start GeneMapper ID Software v3.2 for the first time:

4. The first time you launch the GeneMapper ID software,


complete the Login dialog box as follows:
a. Leave the User Name as gmid and leave the Password
field blank.
b. From the Database Host drop-down list, select the local
computer (that is, select the name of the computer you
are currently working on).
c. Click OK.
The Enter Password dialog box opens.

5. Create your password:


a. Leave the Old Password field blank.
b. In the Password field, enter a new, user-defined
password.
c. Confirm the password.
d. Click OK.
The GeneMapper ID software starts.

16 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Section 2 New Features: All AmpFlSTR Kits

Section 2 New Features: All AmpFlSTR Kits

This section covers:


New Features Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Export Combined Table Format . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Changes to the Electropherogram Displays . . . . . . . . . . . . . . . . . . 22

New Features Summary


Table 1 below provides a brief summary of the new features included
in GeneMapper ID Software v3.2. For more detailed information, see
the pages referenced in the table.

Table 1 New features summary

See
Feature Description
Page

Export Combined Table • When exporting from the Samples view, you can now export 18
format samples that do not pass sizing along with samples that pass
sizing. This feature combines columns from the sample table
and the genotype table and exports them as a single table.

• You now have two display options when exporting samples:


– One line per marker
– One line per sample (This is similar to the Make Allele table
in the Genotyper® Software.)

Changes to the • The software now provides the option to display labeled peak 37
electropherogram assignments for all size standards. This allows you to quickly
displays identify peaks visually and perform a size precision test.

The labeled peak assignments are printable.

• When switching from the “align by base pair” to the “align by


data point” views for the X-axis, the labels associated with
the peaks are now retained in both views.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 17


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Export Combined Table Format


GeneMapper ID Software v3.2 can export data in the Export
Combined Table format. This new format:
• Allows you to track all samples in an electrophoresis run. That
is, when exported from the Samples view, samples that do not
pass sizing are exported along with samples that pass sizing.
• Provides two display options: one line per marker and one line
per sample

Size Quality As shown in Figure 1 below, the size quality status is displayed in the
Status in the SQ column of the Samples view:
Samples View • Samples that do not pass sizing are flagged with a red circle.
• Samples that pass sizing are flagged with a green square.
When you create an Export Combined Table, both the passing and
low-quality samples are exported together, as shown in Figure 2 on
page 19 and Figure 3 on page 20. The procedure for creating an
Export Combined Table is on page 20.

These samples do
not pass sizing.

These samples
pass sizing.

Figure 1 Size quality status

18 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Export Combined Table Format

Two Export Two display options are available to export samples that do not pass
Display Options sizing with samples that pass sizing:
• One line per marker
This option displays information for each marker on a separate
line for a given sample. In Figure 2 below for example, lines 2
through 15 display the information for each marker in sample
file A1012531.fsa.

Figure 2 One line per marker

Installation Procedures and New Features for GeneMapper® ID Software v3.2 19


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

• One line per sample


This option displays the information for each sample on just one
line. In Figure 3 below for example, line 2 displays information
for Sample Name 12531. The subsequent lines display
information for different samples.

Figure 3 One line per sample

Creating an To create an Export Combined Table:


Export Combined
Table 1. After analyzing your samples, go to the toolbar and select
File > Export Combined Table to open the Export
Combined Table window.

20 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Export Combined Table Format

To create an Export Combined Table:

2. From the Look in drop-down list, select an appropriate


destination folder for the Export Combined Table.

3. From the Export File As drop-down list, select a file type.

4. From the Merge panel, select the format style:


• One line per marker, or
• One line per sample
Note: See page 19 and page 20 for an example of each
format.

5. If you selected One line per marker in step 4,


select/deselect the Include all marker information
checkbox, as necessary.
As shown in the pop-up text below, if you select this option,
marker information for samples with no sizing data will be
included in the Export Combined Table.

6. Type in a File name.

7. From the Files of type drop-down list, select the appropriate


file type/extension.

8. Click Export Combined Table.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 21


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Changes to the Electropherogram Displays


In GeneMapper ID Software v3.2, two changes have been made to
the electropherogram displays:
• Labeled peak assignments can be displayed for all size standards
• Labels are retained in both the “align by base pair” and “align by
data point” views for the X-axis

Labeled Peak GeneMapper ID Software v3.2 now provides the option to display
Assignments for labeled peak assignments for all size standards in the
All Size electropherograms. (That is, all defined size standard peaks are
Standards labeled in the electropherogram per the size standard definitions).
This allows you to quickly identify peaks visually and perform a size
precision test.

To display peak assignments for all size standards:

1. In the Samples view, highlight all sample files in the project.

2. Click the Display Plots icon.

3. From the Plot Setting drop-down list, select AmpFLSTR


Genotyping.

4. Hide all dyes except the size standard dye (that is, red for
ROX™ dye or orange for LIZ® dye).

5. Select View > Show Size Standard Label.

6. Verify that all values for the 250-bp fragment peaks are
within ± 0.5 bp.

22 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Changes to the Electropherogram Displays

To display peak assignments for all size standards: (continued)

7. If desired, select File > Print to print the electropherogram


plots. The examples below show both labeled and unlabeled
peak assignments.

Labeled peak assignments

Unlabeled peak assignments

Installation Procedures and New Features for GeneMapper® ID Software v3.2 23


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Labels Retained The X-axis can be displayed in either base pairs or data points. In
in Base Pair and GeneMapper ID Software v3.1, when switching from the “align by
Data Point Views base pair” to the “align by data point” views for the X-axis, the labels
associated with the peaks were not retained in both views. In
GeneMapper ID Software v3.2, the labels are automatically retained
in both views, as shown below.

Figure 4 Align by base pair

Figure 5 Align by data point

24 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Section 3 New Features and Procedures: Yfiler Kit

Section 3 New Features and Procedures:


Yfiler Kit

This section covers:


New Features and Procedures Summary. . . . . . . . . . . . . . . . . . . . . 25
Allele Calling Parameters for New Marker Repeat Types . . . . . . . 27
Plus Stutter Filtering and −2-bp Filtering Workaround. . . . . . . . . . 27
Creating HID Analysis Methods for the Yfiler Kit. . . . . . . . . . . . . 30
Creating a Table Setting and Uploading Exported Haplotype(s) . . 37

New Features and Procedures Summary


Two features have been added in GeneMapper ID Software v3.2 to
facilitate analysis of the AmpFlSTR® Yfiler™ PCR Amplification
Kit:
• Allele calling parameters for new marker repeat types
• Plus stutter filtering
Additionally, Applied Biosystems has developed several procedures
for using GeneMapper ID Software v3.2 with the Yfiler kit:
• Workaround for the DYS19 locus (−2-bp filtering)
• Creating HID analysis methods for the Yfiler kit
• Exporting haplotypes from GeneMapper ID Software v3.2
Table 2 on page 26 provides a brief summary of these new features
and procedures. For more detailed information, see the pages
referenced in the table.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 25


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Table 2 New features and procedures summary

See
Feature/Procedure Description
Page

New feature: Allele calling In GeneMapper ID Software v3.1, allele calling parameters 27
parameters for new were only available for markers with tetranucleotide repeat
marker repeat types motifs. In GeneMapper ID Software v3.2, however, allele
calling parameters are available for four marker repeat types:
tri-, tetra-, penta-, and hexanucleotide.
The four allele calling parameters appear in the Allele tab of the
Analysis Method Editor (as shown in Figure 6 on page 29). You
may enter all related analysis values directly into the Tri, Tetra,
Penta, or Hexa column fields, which allows you to see all
values at once.

New feature: Plus stutter To aid in interpreting genotype profiles, two new fields have 27
filtering (for the DYS392 been added to the Allele tab of the Analysis Method Editor
locus) specifically to filter out the DYS392 plus stutter:
• Plus Stutter Ratio
• Plus Stutter Distance

Procedure: Workaround Laboratories may choose to implement a workaround for the


for the DYS19 locus DYS19 locus by using the Minus A Ratio and Minus A Distance
(−2 bp filtering) fields in the Allele tab of the Analysis Method Editor. These
fields can be used to filter out the −2-bp stutter that is
observed at the DYS19 locus.

Procedure: Creating HID Use the Analysis Method Editor in GeneMapper ID Software 30
analysis methods for the v3.2 to set analysis parameter values for analyzing the Yfiler kit
Yfiler kit data.

Procedure: Creating a Using the Table Setting Editor, you can create a table setting in 37
table setting and GeneMapper ID Software v3.2 to export haplotypes
uploading exported specifically for searching the Yfiler Haplotype Database for
haplotype(s) for searching profile match estimation.
profiles with the Yfiler™
Haplotype Database

26 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Allele Calling Parameters for New Marker Repeat Types

Allele Calling Parameters for New Marker Repeat


Types
Peak Detection The peak detection analysis parameters allow you to specify the
Analysis minimum peak height to be detected for analysis. This, in turn,
Parameters controls the number of peaks analyzed. Peaks falling below the
parameters specified are displayed in the electropherogram, but are
not labeled and will not be genotyped.
In the Allele tab of the Analysis Method Editor, a number of
parameters can be set to control allele calling, including:
• Bin set
• Marker-specific stutter ratio
• Amelogenin cutoff
• Marker repeat type

New Marker In GeneMapper ID Software v3.1, allele calling parameters were only
Repeat Types available for markers with tetranucleotide repeat motifs. In
GeneMapper ID Software v3.2, however, allele calling parameters
are available for four marker repeat types: tri-, tetra-, penta-, and
hexanucleotide.
The four allele calling parameters appear in the Allele tab of the
Analysis Method Editor (as shown in Figure 6 on page 29). You may
enter all related analysis values directly into the Tri, Tetra, Penta, or
Hexa column fields, which allows you to see all values at once.

Plus Stutter Filtering and −2-bp Filtering


Workaround
The following have been reported:
• Plus stutter at the DYS392 locus
• −2-bp stutter at the DYS19 locus
Further descriptions of these loci are provided on page 28. Allele
calling parameters that can be used as workarounds are provided on
page 28.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 27


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Plus Stutter The DYS392 trinucleotide repeat locus displays the typical −3-bp
(DYS392) stutter but also a larger +3-bp stutter. Sequencing analysis of this
+3-bp stutter revealed that the product contains an additional repeat
unit relative to the true allele peak.
Based on developmental validation studies conducted by
Applied Biosystems, the highest percent stutter observed for any
DYS392 allele was 7.9%.

−2-bp DYS19 It has been reported that the DYS19 tetranucleotide repeat locus
Stutter displays the typical −4-bp stutter but also a smaller −2-bp stutter.
Based on developmental validation studies conducted by
Applied Biosystems, the highest percent stutter observed for any
DYS19 locus was 10.21%.

Allele Calling The Allele tab of the Analysis Method Editor is used to set allele
Parameters calling parameters for data processing.
Select the Use marker-specific stutter ratio if available checkbox.
When this checkbox is selected, the software uses the stutter ratios
defined in the panel. These values are provided in the panels supplied
by Applied Biosystems.
To filter out the reproducible DYS19 tetranucleotide stutter, the
Minus A Ratio and Minus A Distance fields can be used.
Additionally, two new fields specifically developed to filter out the
DYS392 trinucleotide repeat plus stutter have been added to the
Analysis Method Editor: Plus Stutter Ratio and Plus Stutter Distance.
Figure 6 on page 29 shows allele calling parameters that can be used
with GeneMapper ID Software v3.2.

28 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Plus Stutter Filtering and −2-bp Filtering Workaround

Figure 6 Allele calling parameters

Note: The circled values shown in Figure 6 above are based on


developmental validation studies conducted at Applied Biosystems
and should only be used as reference values. Each laboratory should
perform their own internal validation studies to determine the
appropriate values to enter. The DYS19 workaround is not locus-
specific. All peaks for the tetranucleotide loci falling within the range
specified in the Minus A Ratio and Minus A Distance fields will be
filtered.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 29


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Creating HID Analysis Methods for the Yfiler Kit


Use the Analysis Method Editor to set analysis parameter values for
analyzing the Yfiler kit data.
Note: The size distribution for the Yfiler kit is ~100 to 330 bp. As a
result, the 450-bp size standard peak does not need to be included in
the size standard definition for analysis.

Workflow 1. Import the Yfiler kit panels and bin sets.


See the GeneMapper™ ID Software Version 3.1 Human
Identification Analysis Tutorial (PN 4335523) for procedures on
importing panels and bin sets.
2. Create a new HID_Advanced analysis method for use with the
AmpFlSTR Yfiler kit (below).
3. Create a new HID_Classic analysis method for use with the
AmpFlSTR Yfiler kit (page 36).

Create the To create the HID_Advanced analysis method:


HID_Advanced
Analysis Method 1. Select Tools > GeneMapper Manager to open the
GeneMapper Manager.

2. Select the Analysis Methods tab, then click New to open the
New Analysis Method dialog box.

3. Select HID, then click OK to open the Analysis Method


Editor with the General tab selected.

4. Select the analysis method settings shown in Table 3a on


page 31 through Table 3e on page 35.
IMPORTANT! You must select your settings on all the tabs
before you click OK to save the analysis method and return
to GeneMapper Manager!

30 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Creating HID Analysis Methods for the Yfiler Kit

Analysis Method IMPORTANT! The values shown in Table 3b on page 32 through


Settings Table 3e on page 35 are based on developmental validation studies
conducted at Applied Biosystems and should only be used as
reference values. Each laboratory should perform their own internal
validation studies to determine the appropriate values to enter.
Table 3a Analysis method settings: General tab

Tab Settings

General

Enter a name of
your choosing
here. See the
Notes below.

Notes for the General tab:


1. Enter a name of your choosing. For example, enter:
– AmpFlSTR_Yfiler_AdvancedMode, if you are creating an HID_Advanced analysis method, or
– AmpFlSTR_Yfiler_ClassicMode, if you are creating an HID_Classic analysis method
2. Select your instrument from the drop-down list.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 31


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Table 3b Analysis method settings: Allele tab

Tab Settings

Allele

Notes for the Allele tab:


1. Be sure to select the Use marker-specific stutter ratio if available checkbox. When this
checkbox is selected:
– The software uses the stutter ratios defined in the panel.
– The stutter ratios include all loci except for the DYS392 +3 stutter and the DYS19 −2 stutter. The
stutter ratios are defined in the Analysis Method Editor under the Allele tab, if applicable.
2. See “Plus Stutter Filtering and −2-bp Filtering Workaround” on page 27 for a discussion of the
Tetra column values in the Minus A Ratio/Distance fields and Tri column values in the Plus Stutter
Ratio/Distance fields.

32 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Creating HID Analysis Methods for the Yfiler Kit

Table 3c Analysis method settings: Peak Detector tab

Tab Settings

Peak Detector

Select
Advanced or
Classic here,
depending on
the analysis
method you are
creating. See the
Notes below.

Notes for the Peak Detector tab:


From the Peak Detection Algorithm drop-down list, select:
• Advanced, if you are creating an HID_Advanced analysis method, or
• Classic, if you are creating an HID_Classic analysis method

Installation Procedures and New Features for GeneMapper® ID Software v3.2 33


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Table 3d Analysis method settings: Peak Quality tab

Tab Settings

Peak Quality

Notes for the Peak Quality tab:


The values shown in the Peak Quality tab have been chosen based on empirical data generated at
Applied Biosystems and should only be used as reference values. Each laboratory should perform
their own internal validation studies to determine the appropriate values to enter.

34 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Creating HID Analysis Methods for the Yfiler Kit

Table 3e Analysis method settings: Quality Flags tab

Tab Settings

Quality Flags

Notes for the Quality Flags tab:


The weighted values in the Quality Flag Settings section were ranked on a scale from most important
to least important. This ranking was performed by the Applied Biosystems HID Group.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 35


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Create the To create the HID_Classic analysis method:


HID_Classic
Analysis Method 1. Select Tools > GeneMapper Manager to open the
GeneMapper Manager.

2. Select the Analysis Methods tab, then click New to open the
New Analysis Method dialog box.

3. Select HID, then click OK to open the Analysis Method


Editor.

4. Select the analysis method settings shown in Table 3a on


page 31 through Table 3e on page 35.
IMPORTANT! You must select your settings on all the tabs
before you click OK to save the analysis method and return
to GeneMapper Manager!

What Next? After creating analysis methods for the Yfiler kit, continue with data
analysis as follows:
1. Import sample files
2. Select analysis parameters
3. Analyze the project

For More For more information, refer to the following documents:


Information • GeneMapper™ ID Software Version 3.1 Human Identification
Analysis User Guide (PN 4338775)
• GeneMapper™ ID Software Version 3.1 Human Identification
Analysis Tutorial (PN 4335523)
• AmpFlSTR® Yfiler™ PCR Amplification Kit User’s Manual
(PN 4358101)

36 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Creating a Table Setting and Uploading Exported Haplotype(s)

Creating a Table Setting and Uploading Exported


Haplotype(s)
Using the Table Setting Editor, you can create a table setting in
GeneMapper ID Software v3.2 to export haplotypes specifically for
searching the Yfiler™ Haplotype Database for profile match
estimation.

Database Online The Yfiler Haplotype Database online search tool searches for
Search Tool haplotypes generated with the Yfiler kit. The online search tool
allows you to:
• Estimate the frequency of a given Y-chromosome haplotype in
specified populations. (Frequency calculations are based on the
haplotype data generated from the 17 loci included in the Yfiler
kit. This data was compiled from more than 3000 samples from
a range of populations.)
• Estimate the frequency of haplotypes in a number of reference
populations.
• Search complete or partial profiles generated with the Yfiler kit.
• Compare the discrimination capacity of the Yfiler kit relative to
the recombination of Ystr loci.

Workflow 1. Use the Table Setting Editor to create a table setting in


GeneMapper ID Software v3.2 (below).
2. Upload the exported haplotype(s) into the Yfiler Haplotype
Database for profile match estimation (page 40).

Create a Table Define custom Samples and Genotypes views for displaying the
Setting Yfiler kit data.
To create a table setting:

1. From the GeneMapper Manager, select the Table Settings


tab.

2. Click New to open the Table Setting Editor.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 37


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

To create a table setting: (continued)

3. Select the General tab and complete it as follows:


a. Enter a Name of your choosing.
b. If desired, enter a Description.

4. Select the Samples tab and complete it as follows:


a. In the Column Settings section, deselect all options.
b. In the Font Settings section, leave the default values.

38 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Creating a Table Setting and Uploading Exported Haplotype(s)

To create a table setting: (continued)

5. Select the Genotypes tab and complete it as follows:


a. In the Column Settings section, select:
– Sample Name
– Marker
– Allele
Deselect all of the other options.
b. In the Font Settings section, leave the default values.
c. In the Allele Settings section:
– Enter 2 in the Number of Alleles field.
– Deselect the Keep Allele, Size... check box.

6. Click OK to save the table settings, close the Table Setting


Editor, and return to the GeneMapper Manager.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 39


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Upload the Upload the exported haplotype(s) into the Yfiler Haplotype
Exported Database.
Haplotype(s)
To upload the exported haplotype(s):

1. If you are not already there, open the project in GeneMapper


ID Software v3.2.

2. From the Table Settings drop-down list, select the table


setting you created specifically to export haplotype(s) for
searching the Yfiler Haplotype Database.

3. For the remaining upload procedures, see:


• The Yfiler Haplotype Database:
http://www.appliedbiosystems.com/yfilerdatabase/
or
• The AmpFlSTR® Yfiler™ PCR Amplification Kit User’s
Manual (PN 4358101)

40 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Creating a Table Setting and Uploading Exported Haplotype(s)

Appendix A Verification Testing

This appendix covers:


Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Materials and Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Category 1: Peak Detection and Genotyping (Reproducibility) . . . 45
Category 2: Algorithm Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Category 3: Data Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Category 4: Workflow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Results for Category 1: Peak Detection and Genotyping
(Reproducibility) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Results for Category 2: Algorithm Testing . . . . . . . . . . . . . . . . . . . 53
Results for Category 3: Data Handling . . . . . . . . . . . . . . . . . . . . . . 54
Results for Category 4: Workflow. . . . . . . . . . . . . . . . . . . . . . . . . . 55
Results Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

Installation Procedures and New Features for GeneMapper® ID Software v3.2 41


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Introduction
Test Plan The test plan, which Applied Biosystems has defined as verification
of the software, was designed to evaluate the performance of
GeneMapper ID Software v3.2 for the human identification
communities.

Test Categories Applied Biosystems performed verification of GeneMapper ID


Software v3.2 to assess the performance, robustness, and feature
design in four specific categories:
• Peak Detection and Genotyping (Reproducibility) (page 45)
• Algorithm Testing (page 46)
• Data Handling (page 48)
• Workflow (page 48)

Evaluation GeneMapper ID Software v3.2 verification was performed to:


• Confirm and document the functionality of the new software
features
• Assess the functionality of the minor modifications made to the
analysis method
Our findings demonstrate that GeneMapper ID Software v3.2 (with
its default settings and panels and bin sets) is valid for forensic,
paternity, and databasing analyses. GeneMapper ID Software v.3.2
accurately detects, genotypes, and performs quality checks when
performing STR analysis.

42 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Materials and Methods

Materials and Methods


This section describes the AmpFlSTR® kit, sample types,
instruments, and software used for the GeneMapper® ID Software
v3.2 verification testing.

Kit The AmpFlSTR® Yfiler™ PCR Amplification Kit was used to


amplify each sample type for the verification testing.

Sample Types • Single-source DNA Samples – Eighty (80) DNA samples from
four (4) population groups
• Mixture Samples – Mixtures of male and male DNA were
amplified. These samples were mixed in defined ratios of 1:1,
3:1, 10:1, and 15:1, respectively. Additionally, mixtures of male
and female DNA were amplified. These samples were mixed in
defined ratios of 1:1000, 1:2000, 1:4000, and 1:8000,
respectively.
Note: The human genomic DNA samples were quantified using
Quantifiler™ Human DNA Quantification Kit and
Quantifiler™ Y Human Male DNA Quantification Kit. The
samples were amplified using a GeneAmp® PCR System 9700
with a silver 96-well block. The recommended cycling
conditions were used, as outlined in the AmpFlSTR® Yfiler™
PCR Amplification Kit User’s Manual (PN 4358101). The
amplified samples were injected on three or more
ABI PRISM ® instrument platforms.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 43


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Instruments and Table 4 below lists the instruments, Data Collection software
Software versions, computer operating systems, and run modules that were
used to process the samples for verification testing.

Table 4 Forensic validated instrument platforms

Data
Computer Operating
Instrument Collection Run Module
System
Software

ABI PRISM ® 310 v3.0 Microsoft® Windows NT® • GS STR Pop4 (1 ml) F
Genetic Analyzer • GS STR Pop4 (1 ml) G5v2

ABI PRISM ® 3100 v1.1 Microsoft® Windows NT® • GS 36_Pop4_F


Genetic Analyzer • GS 36vb_Pop4_G5
v2.0 Microsoft® Windows 2000

44 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Category 1: Peak Detection and Genotyping (Reproducibility)

Category 1: Peak Detection and Genotyping


(Reproducibility)
The first test category evaluated genotype concordance using single-
source population DNA samples, male:male mixture samples, and
male:female mixture samples. All samples were amplified with the
Yfiler kit.
Two comparison studies were conducted to verify genotype
concordance using DNA samples previously amplified with the
Yfiler kit. These data were used to compare the genotype results from
data analysis with GeneScan® Software v3.7.1 and Genotyper®
Software v3.7 (both running on the Windows NT operating system)
to GeneMapper ID Software v3.2 (running on the Windows 2000
operating system) using the Advanced mode.

Single-Source The first comparison study consisted of:


Population DNA • Eighty (80) samples from the Yfiler kit population study
Samples
• Four (4) positive controls
• One (1) negative control
• Six (6) ladders
These samples were electrophoresed on the ABI PRISM ® 3100
Genetic Analyzer and consisted of 1222 alleles from the samples, 60
alleles from the controls, and 137 alleles from the ladders. The
samples were examined for reproducibility.

Mixture Samples The second comparison study consisted of mixture data produced
during the Yfiler kit developmental validation. This study was
conducted to verify that the same mixture samples (that is, PCR
product) gave concordant genotype calls between the
GeneScan/Genotyper software and the GeneMapper ID software
applications.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 45


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

The mixture samples were electrophoresed on three different


instruments and consisted of five (5) male:male and five (5)
male:female mixture samples. Table 5 below lists the mixture ratios
that were used in the mixture experiment.

Table 5 Mixture ratios

Male:Male Male:Female

1:1 1:1000

3:1 1:2000

10:1 1:4000

15:1 1:8000

Category 2: Algorithm Testing


The second test category evaluated the functionality of GeneMapper
ID Software v3.2 to accurately filter out stutter based on the Marker
Specific Stutter ratio specified in the panel and in the Allele tab of
the Analysis Method Editor.

Stutter Evaluation In the Allele tab of the Analysis Method Editor, one of the parameters
that can be set to filter out stutter is the Use marker-specific stutter
ratio if available check box.
Each allele within a locus displays a percent stutter that is
reproducible. The Use marker-specific stutter ratio if available
check box was selected in order to use the stutter ratio information
from the panel. Using GeneMapper ID Software v3.2, 27 population
samples were examined and the percent stutter recorded. The 27
population samples consisted of:
• 432 (n−4), (n−5), and (n−6) stutter peaks
• 27 DYS19 (n−2) stutter peaks
• 27 DYS392 (n+3) stutter peaks

46 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Category 2: Algorithm Testing

Spectral Pull-Up Laboratories that have implemented the ABI PRISM 3100 Genetic
PQV Analyzers into their workflow to electrophorese DNA samples may
observe pull-up peaks within ±2 data points (in a different color)
from the main allele peak, which is causing the pull-up peak. To this
end, Applied Biosystems has made modifications to the analysis
method to flag spectral pull-up peaks in the range of ±2 data points.
The spectral pull-up PQV flag was evaluated to ensure that a peak
meeting specified criteria for spectral pull-up is appropriately
activated. The two criteria required to flag a peak as spectral pull-up
are as follow:
• The peak is ≤ 5% (or as specified by the value in the Pull-up
ratio field in the Peak Quality tab of the Analysis Method
Editor) of the main allele peak causing the pull-up
• The peak is within the ±2 data points range from the main allele
peak in a different color
Note: After installing GeneMapper ID Software v3.2, laboratories
should create new analysis methods specifically for use with
GeneMapper ID Software v3.2. Creating new analysis methods will
ensure that spectral pull-up peaks within ±2 data points from the
main allele peak are flagged appropriately.

Spectral Pull-UP PQV Study


To test the functionality of the spectral pull-up PQV, twenty-eight
(28) samples from the single-source population study were
electrophoresed on both the ABI PRISM ® 310 and 3100 Genetic
Analyzers (that is, the same PCR product was electrophoresed on two
different instruments) and were evaluated using the following
criteria:
• Each analyzed sample was evaluated for the presence of pull-up
peaks by noting whether or not the spectral pull-up PQV was
flagged.
• The data point (that is, ±0 data point, ±1 data point, or ±2 data
points) of the pull-up peak relative to the main allele peak was
recorded.
• The percent height of the pull-up peak relative to the main allele
peak was recorded (that is ≤ 5% of the main allele peak causing
the pull-up)

Installation Procedures and New Features for GeneMapper® ID Software v3.2 47


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Category 3: Data Handling


The third test category evaluated the Export Combined Table format
and the size standard printing functionality.

Export Combined The purpose for evaluating the Export Combined Table format was to
Table Format verify that samples that do not pass sizing (that is, low-quality
samples) are exported along with samples that pass sizing.

Printing Labeled In GeneMapper ID Software v3.2, the ability to display and print
Size Standard labels on the size standards in different plot configurations was
Peaks tested.

Category 4: Workflow
The fourth test category evaluated the ability of GeneMapper ID
Software v3.2 to:
• Display labeled peak assignments for all size standards
• Retain labels in both the “align by base pair” and “align by data
point” views for the X-axis

Displaying All samples within the Samples view were highlighted and only the
Labeled Peak orange dye color was selected, as shown below.
Assignments for
Size Standards

The size standards were checked to ensure that all defined size
standard peaks were labeled appropriately.

Retaining Labels The electropherogram plots were displayed to verify that labels
in Base Pair and associated with the peaks were retained when switching back and
Data Point Views forth between the “align by base pair” and “align by data point”
views for the X-axis.

48 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Results for Category 1: Peak Detection and Genotyping (Reproducibility)

Results for Category 1: Peak Detection and


Genotyping (Reproducibility)
Single-Source The first parameter evaluated the Peak Detection and Genotyping
Population DNA category to assess genotyping concordance. Applied Biosystems
Samples compared the genotype results from data analysis with GeneScan
Software v3.7.1 and Genotyper Software v3.7 (both running on the
Windows NT operating system) to GeneMapper ID Software v3.2
using the Advanced algorithm.
The analysis parameter settings for the software packages were
defined according to Applied Biosystems default analysis parameters
with a defined peak amplitude threshold (PAT) of 50 RFUs.
As shown in Figure 1 below, a side-by-side comparison between
Genotyper Software v3.7 and GeneMapper ID Software v3.2 was
evaluated to ensure the consistency of genotype results, where the
evaluation criteria was that all samples give concordant genotypes.
After the trained scientist completed data analysis, each population
sample tested gave 100% concordant genotype calls between the two
software programs (see Table 6 on page 50).

Genotyper Software v3.7 GeneMapper ID Software v3.2

Figure 1 Example of the Yfiler kit allelic ladder analyzed with


Genotyper Software v3.7 (left) and GeneMapper ID Software v3.2
displaying concordant results (right)

Installation Procedures and New Features for GeneMapper® ID Software v3.2 49


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

As shown in Table 6 below, each sample type tested gave concordant


genotype calls between the two software programs.

Table 6 Genotype concordance between Genotyper Software


v3.7 and GeneMapper ID Software v3.2

Number of Number of Concordant


Allele Source
Samples Alleles Alleles

Population samples 80 1222 100%

Positive and negative 5 60 100%


controls

Allelic ladders 6 137 100%

Mixture Samples After the trained scientist completed data analysis, Applied
Biosystems documented three observations when comparing the
results produced from GeneScan Software v3.7.1 and Genotyper
Software v3.7 (both running on the Windows NT operating system)
to the results produced from GeneMapper ID Software v3.2. Of the
272 loci examined, 267 loci produced concordant results between the
software packages and 5 loci/alleles produced non-concordant
results. These observations are further described on page 51.
Figure 2 below shows the genotyping concordance results when
comparing the analyzed data from GeneScan Software v3.7.1 and
Genotyper Software v3.7 (both running on the Windows NT
operating system) to the results from GeneMapper ID Software v3.2

Figure 2 Genotyping concordance results

50 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Results for Category 1: Peak Detection and Genotyping (Reproducibility)

• Observation 1: 267 (98.20%) of the loci/alleles compared


produced concordant genotype calls.
• Observation 2: Three (1.08%) of the loci/alleles were
non-concordant because these loci/alleles fell just below the
50-RFU cutoff threshold value in GeneMapper ID Software v3.2
and were not genotyped.
• Observation 3: Two (0.72%) of the loci/alleles were
non-concordant because the alleles analyzed with Genotyper
Software v3.7 did not filter the stutter or the Minus A. The label
is removed from Peak A (the stutter peak) if Peak B (the true
allele) meets two criteria:
– Peak B is higher than Peak A by the specified percentage, and
– Peak B is within the specified proximity size (in base pairs)
range relative to Peak A
Table 7 below shows the results for observation 3.

Table 7 Observation 3: non-concordant loci

GeneMapper ID
Genotyper Software v3.7
Software v3.2
Marker/Locus
Allele 1 Allele 2 Allele 3 Allele 4 Allele 1 Allele 2

DYS439 11 OL 12 13 12 13

DYS456 15 16 OL 17 15 17

As shown in Figure 3 and Figure 4 on page 52, these two loci/alleles


did not meet these criteria due to −A product, resulting in the
detection of a third peak (Peak C). The filtering in Genotyper
Software v3.7 compared Peak A to Peak C. GeneMapper ID Software
v3.2 appropriately filtered the stutter and the shoulder peak using
defined bin sets

Installation Procedures and New Features for GeneMapper® ID Software v3.2 51


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Genotyper Software v3.7 GeneMapper ID Software v3.2

Figure 3 DYS439 Marker

Genotyper Software v3.7 GeneMapper ID Software v3.2

Figure 4 DYS456 Marker

52 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Results for Category 2: Algorithm Testing

Results for Category 2: Algorithm Testing


Stutter Evaluation Twenty-seven (27) population DNA samples were examined,
consisting of:
• 432 (n−4), (n−5), and (n−6) stutter peaks
• 27 DYS19 (n−2) stutter peaks
• 27 DYS392 (n+3) stutter peaks
The results were as follows:
• All (n−4, n−5, and n−6) stutter peaks were below the stutter
percent as defined in the panel and were properly filtered.
• The DYS19 locus (n−2) stutter peak was below the stutter
percent as defined in the Allele tab of the Analysis Method
Editor and were properly filtered.
• The DYS392 locus (n+3) stutter peak was below the stutter
percent as defined in the Allele tab of the Analysis Method
Editor and were properly filtered.

Spectral Pull-Up Twenty-eight (28) single-source population samples were evaluated


PQV for spectral pull-up. For each pull-up peak, the following was
recorded:
• Size
• Peak height
• Data point
• Spectral pull-up PQV activated
Seventy-eight (78) spectral pull-up peaks were identified. As shown
in Figure 5 on page 54, 97.44% (76) of these pull-up peaks were
correctly flagged by GeneMapper ID Software v3.2. The remaining
2.56% (2) of pull-up peaks that were not flagged were produced by
off-scale data. If a spectral pull-up peak is caused by an off-scale
allele peak, the spectral pull-up flag is not triggered for that peak.
This is because the ratio of pull-up to off-scale peaks cannot be
calculated (since the peak height for off-scale data cannot be
determined).

Installation Procedures and New Features for GeneMapper® ID Software v3.2 53


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Applied Biosystems recommends that you carefully review all off-


scale yellow PQV flags. Applied Biosystems recommends that you
dilute or re-run samples to obtain on-scale data to ensure the PQVs
reported by data analysis are accurate and to reduce spectral
separation (that is, pull-up).

Figure 5 Results for the spectral pull-up PQVs flagged

Results for Category 3: Data Handling


Export Combined Projects with their associated settings (that is, analysis methods, table
Table Format settings, plot settings, matrices, if applicable, and size standards)
consisting of samples that passed sizing as well as samples that did
not pass sizing were exported. The tables were exported from the
Samples view using the Export Combined Table format. Eighty (80)
samples from the Yfiler kit population study were used.
The results obtained verified that GeneMapper ID Software v3.2 is
100% functional in its ability to:
• Export projects using the Export Combined Table format
• Track all samples in an electrophoresis run (samples that do not
pass sizing as well as samples that pass sizing)

54 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Results for Category 4: Workflow

Printing Labeled The purpose of this test was to verify that the size standard plots
Size Standard could be printed in different display configurations.
Peaks The results obtained verified that GeneMapper ID Software v3.2 is
100% functional in its ability to display and print labels on size
standards in different plot configurations. For example, displaying
five labeled size standards separated into five panes, using three
levels of magnification.

Results for Category 4: Workflow


Displaying All size standards from eighty (80) DNA samples from the Yfiler kit
Labeled Peak single-source population study were examined. All peaks were
Assignments correctly labeled.

Retaining Labels The results obtained verified that GeneMapper ID Software v3.2 is
100% functional in its ability to switch back and forth between the
“align by base pair” and “align by data point” views for the X-axis.

Results Summary
Applied Biosystems verification testing demonstrates that human
identification laboratories can successfully adopt GeneMapper ID
Software v3.2.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 55


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

56 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Results Summary

Appendix B Troubleshooting the Installation

This appendix covers:


Troubleshoot the Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Uninstall the Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Troubleshooting Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
How to Obtain Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

Installation Procedures and New Features for GeneMapper® ID Software v3.2 57


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Troubleshoot the Installation


Workflow If the installation appears to be successful, but the GeneMapper ID
software fails to start and reports the error “Failed to connect to
database,” perform the troubleshooting procedures in this appendix in
the order shown below.
1. Check the Oracle® database installation (page 58).
2. Configure the Oracle database manually (page 58).

Check the Oracle Check to see if the Oracle database instance was installed correctly.
Database
Installation To check the Oracle database installation:
1. Select Start > Programs > Accessories > Command Prompt.
2. Type sqlplus, then press Enter. You should see:
SQL*Plus: Release 8.1.7.0.0...
3. Type the user name system, then press Enter.
4. Type the password manager, then press Enter. If:
• A “Connected to:” message is displayed, the Oracle
database was installed correctly. Proceed with “Configure
the Database Manually” on page 58.
• If the Oracle database was not installed correctly, an error
message appears. Complete Table 8, “Troubleshooting
Checklist,” on page 61, then contact Technical Support.

Configure the If the Oracle database was installed correctly, perform the following
Database steps in the Command Prompt window to configure the database
Manually manually.

To configure the database:


1. Return to the open Command Prompt window (it should still be
open from step 4 above).
2. At the SQL prompt, type exit to exit the SQL Plus application.
3. At the drive prompt, type cd /d x:\GeneMapper\database,
where x is the drive where the GeneMapper ID software is
installed, then press Enter.

58 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Uninstall the Software

4. At the drive prompt, type ifadb x:\oraclegm\oradata\ifa,


where x is the drive where the GeneMapper ID software is
installed, then press Enter.
• If “Succeeded” messages are displayed, the database is
configured correctly. Proceed to “Start the Software” on
page 14.
• If “Error” messages are displayed, the database is not
configured correctly. Proceed to “Uninstall the Software”
on page 59.

Uninstall the Software


If the software installation is not successful and the “Troubleshoot the
Installation” procedures on page 58 have not resolved the problem,
you may need to uninstall the GeneMapper ID software, then contact
Technical Support.

Workflow 1. Log onto the local domain of your computer as a user with
Administrator privileges (page 59).
2. Uninstall the software (page 59).
3. Complete the Troubleshooting Checklist, then contact Technical
Support (page 61).

Log Onto the To uninstall the GeneMapper ID software, you must:


Computer • Log onto the local computer (not a network domain)
• Have Administrator privileges on the local computer (that is,
have complete and unrestricted access to the local computer)

Uninstall the This procedure is for a standalone configuration only. This procedure
Software removes the GeneMapper ID Software v3.2 and the Oracle Database
Standard Edition v8.1.7.

To uninstall the software:


1. Save and close all applications (including the GeneMapper ID
software) and windows before proceeding with the uninstall.
2. Select Start > Settings > Control Panel.
3. Double-click Add or Remove Programs.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 59


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

4. Select GeneMapper ID v3.2, then click Change/Remove.


5. In the Welcome window, select Remove, then click Next.
6. At the prompt, click OK to confirm the uninstall. The
uninstaller removes all GeneMapper ID Software v3.2 and
Oracle Database Standard Edition v8.1.7 files and settings from
the computer.
7. When the uninstall is complete, the Maintenance Complete page
opens. Select Yes, I want to restart my computer now, then
click Finish.
8. Look for the x:\GeneMapper folder and delete it if it still exists
(sometimes the folder get left behind after an uninstall).
9. Complete the “Troubleshooting Checklist” on page 61.

60 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
Troubleshooting Checklist

Troubleshooting Checklist
Before you contact Applied Biosystems Technical Support for
troubleshooting assistance, complete the Troubleshooting Checklist
(pages 61 to 62).

Checklist Table 8 Troubleshooting Checklist

Check Information for Technical Support

Summarize the problem:

Have you been able to repeat the problem?


If yes, list the steps that you perform:
1.
2.
3.
4.
5.
6.
7.

Applied Biosystems personnel that you have contacted:


❏ Field Applications Specialist
❏ Field Service Engineer
❏ Technical Support
❏ Sales Representative
❏ Order Administration
❏ Other

Computer specifications
• Operating system:
• Version:
• Processor:
• Memory:
• Hard disk space:
• Hard disk configuration:

Installation Procedures and New Features for GeneMapper® ID Software v3.2 61


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
GeneMapper® ID Software

Table 8 Troubleshooting Checklist (continued)

Check Information for Technical Support

Software installed
❏ Data Collection software version:
❏ Status of Data Collection Services:
❏ GeneMapper ID software version:
❏ Other Applied Biosystems software:

Computer login information


• User privileges:
• Local or networked domain:

Software configuration installed


❏ Instrument
❏ Standalone

Instrument and instrument computer information


• Model:
• Data Collection software version:
• Status of Data Collection Services:
• Other Applied Biosystems software:

• Capillary length:
• Capillary lot number:
• Run module:
• Dye set:

Chemistry kit or reagent, with version number:

Be prepared to send to Technical Support:


• Exported panels
• Exported bins
• Exported size standard definition
• Exported analysis method
• Sample (.fsa) files
• GeneMapper_log.txt
• PanelImportLog.txt
• Printed results

62 User Bulletin
DRAFT
November 30, 2004 11:05 am, 4352543A.fm
How to Obtain Support

How to Obtain Support


For the latest services and support information for all locations, go to
Be sure to complete http://www.appliedbiosystems.com, then click the link for
the Troubleshooting
Checklist (page 61)
Support.
before contacting At the Support page, you can:
Technical Support.
• Search through frequently asked questions (FAQs)
• Submit a question directly to Technical Support
• Order Applied Biosystems user documents, MSDSs, certificates
of analysis, and other related documents
• Download PDF documents
• Obtain information about customer training
• Download software updates and patches
In addition, the Support page provides access to worldwide telephone
and fax numbers to contact Applied Biosystems Technical Support
and Sales facilities.

Installation Procedures and New Features for GeneMapper® ID Software v3.2 63


DRAFT
November 30, 2004 11:05 am, 4352543A.fm
© Copyright 2004, Applied Biosystems. All rights reserved.
For Research Use Only. Not for use in diagnostic procedures.
Notice to Purchaser: License Disclaimer.
Purchase of this software product alone does not imply any license
under any process, instrument or other apparatus, system, composition,
reagent or kit rights under patent claims owned or otherwise controlled
by Applera Corporation, either expressly, or by estoppel.
The GeneMapper ID Software has not undergone specific
developmental validation for human identification applications. Human
identification laboratories analyzing single-source or parentage samples
that choose to use the GeneMapper ID Software for data analysis
should perform their own developmental validation studies.
Information in this document is subject to change without notice.
Applied Biosystems assumes no responsibility for any errors that may
appear in this document. This document is believed to be complete and
accurate at the time of publication. In no event shall Applied Biosystems
be liable for incidental, special, multiple, or consequential damages in
connection with or arising from the use of this document.
Purchase or receipt of this software does not include or guarantee any
training by persons belonging to Applied Biosystems or Applera
Corporation.
TRADEMARKS:
ABI PRISM, AmpFlSTR, Applied Biosystems, COfiler, GeneMapper,
GeneScan, Genotyper, Identifiler, LIZ, Profiler, Profiler Plus, and SGM Worldwide Sales and Support
Plus are registered trademarks and AB (Design), AmpFlSTR Blue,
AmpFlSTR Green, Applera, iScience (Design), Quantifiler, ROX, SEfiler, Applied Biosystems vast distribution and
and Yfiler are trademarks of Applera Corporation or its subsidiaries in the service network, composed of highly trained
U.S. and/or certain other countries. support and applications personnel,
Acrobat is a registered trademark of Adobe Systems Incorporated. reaches 150 countries on six continents.
GeneAmp is a registered trademark of Roche Molecular Systems, Inc. For sales office locations and technical support,
please call our local office or refer to our
Macintosh is a registered trademark of Apple Computer, Inc.
Web site at www.appliedbiosystems.com.
Oracle is a registered trademark of Oracle Corporation.
Applera is committed to providing the
Pentium is a registered trademark of Intel Corporation.
world’s leading technology and information
Windows and Windows NT are registered trademarks of Microsoft for life scientists. Applera Corporation
Corporation.
consists of the Applied Biosystems and
All other trademarks are the sole property of their respective owners. Celera Genomics businesses.
This product includes software developed by the Apache Software
Foundation (http://www.apache.org/). Copyright © 1999–2000 The
Apache Software Foundation. All rights reserved. Headquarters
This product includes software developed by the ExoLab Project 850 Lincoln Centre Drive
(http://www.exolab.org/). Copyright © 2000 Intalio Inc. All rights Foster City, CA 94404 USA
reserved. Phone: +1 650.638.5800
Toll Free (In North America): +1 800.345.5224
Fax: +1 650.638.5884

11/2004

www.appliedbiosystems.com Part Number 4352543, Rev. A


Stock Number 112UB06-01

You might also like