CTC PAL Control for Waters Emp3 CDS
Installation Manual
A. Printing History
Edition Date Reason for Change
Edition 4 December 2014 Product version 2.3.3 Release
Edition 3 June 2014 Product version 2.3.1 Release
Edition 2 June 2014 Update as to Feature Release 2
improvement concerning client support
Edition 1 April 2011 Empower 3 support
2
CTC PAL Control for Waters Emp3 CDS
Installation Manual
B. Table of Content
A. Printing History 2
B. Table of Content 3
1. INTRODUCTION 5
1.1 General 5
1.2 About Screen Shots 5
1.3 Default Behavior 5
2. LICENSING, PACKAGING AND DEPLOYMENT 6
3. SUPPORTED CONFIGURATIONS 7
4. SOFTWARE INSTALLATION 8
4.1 System Requirements 8
Hardware and software requirements 8
4.2 Tested Configurations 8
4.3 Installation Instructions 8
Installing the software 8
Uninstalling the software 9
4.4 Trademarks 9
5. EMPOWER CONFIGURATION 10
5.1 Configuration Manager 10
5.2 Empower Nodes 11
5.3 Configuration Manager Dialog for use with serial port 11
5.4 Configuration Manager Dialog for use with TCP/IP 13
Obtain TCP/IP Settings from the Waters DHCP server 13
Fixed TCP/IP Settings 17
5.5 New Chromatographic System 19
Create New System 19
Connect to Existing System 19
Add instruments 20
Access Control 21
Name Selection 22
5.6 Plates 23
Define plate type VT32 23
Define plate type VT98 24
Configure Plates 25
6. FIRST USE 26
6.1 Configure for use with multiple plates 26
7. POINTS YOU NEED TO KNOW 27
7.1 Prerequisite 27
3
CTC PAL Control for Waters Emp3 CDS
Installation Manual
4
CTC PAL Control for Waters Emp3 CDS
Installation Manual
1. INTRODUCTION
1.1 General
This document describes the configuration of the CTC PAL in Empower.
5
CTC PAL Control for Waters Emp3 CDS
Installation Manual
6
CTC PAL Control for Waters Emp3 CDS
Installation Manual
3. SUPPORTED CONFIGURATIONS
The software is designed for use with Empower 3 Build 3471 Feature Release
2 and higher (for more details see 4.2 Tested Configurations).
7
CTC PAL Control for Waters Emp3 CDS
Installation Manual
4. SOFTWARE INSTALLATION
4.1 System Requirements
Before you install the CTC PAL Control for Empower, note these hardware
and software requirements and the information about the PAL environment:
8
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Run ‘Setup.exe
Setup.exe’
Setup.exe to start the installer application (InstallShield Wizard will be
displayed). Follow the instructions the installer provides.
4.4 Trademarks
PAL is a trademark of CTC Analytics AG.
Microsoft, Windows 7 and the Windows logo are trademarks of Microsoft
Corporation in the United States, or other countries, or both.
Other company, product, and service names may be trademarks or service
marks of others.
9
CTC PAL Control for Waters Emp3 CDS
Installation Manual
5. EMPOWER CONFIGURATION
This window is part of Empower and is used to access the primary
applications in Empower software. Applications started from here are assigned
to the current user.
Pressing the Configure System button will pop up the Configuration manager.
1 0
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Managing Empower Nodes involves several tasks for setting up the CTC PAL.
The menu item File > New or right-click Properties brings you to the
configuration Manager Dialog.
Use this tab to select the serial port the PAL is attached to.
1 1
CTC PAL Control for Waters Emp3 CDS
Installation Manual
On some systems the ‘OK?’-value will be ‘No’ at first. The reason may be that
the instrument is not yet initialized. The value should change to ‘Yes’ by
pressing ‘Scan Instruments’ after a few seconds.
1 2
CTC PAL Control for Waters Emp3 CDS
Installation Manual
1 3
CTC PAL Control for Waters Emp3 CDS
Installation Manual
1 4
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Click OK
Confirm that the Instrument Type is set to ctcPAL:
Click OK
1 5
CTC PAL Control for Waters Emp3 CDS
Installation Manual
On some systems the ‘OK?’-value will be ‘No’ at first. The reason may be that
the instrument is not yet initialized. The value should change to ‘Yes’ by
pressing ‘Scan Instruments’ after a few seconds.
1 6
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Click OK
PAL Instrument is added in DHCP Server Configuration
Click OK
When switching to the ‘Instruments’-tab, the following screen will appear:
1 7
CTC PAL Control for Waters Emp3 CDS
Installation Manual
On some systems the ‘OK?’-value will be ‘No’ at first. The reason may be that
the instrument is not yet initialized. The value should change to ‘Yes’ by
pressing ‘Scan Instruments’ after a few seconds.
1 8
CTC PAL Control for Waters Emp3 CDS
Installation Manual
1 9
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Add instruments
Use this page to add instruments to a new chromatographic system.
Note: You cannot drag the last instrument from an existing system. To free up
the last instrument on a system, delete the system.
Use this page to control how other users access the new system you are
creating. You can share the system with other network users and password-
protect the system.
2 0
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Access Control
Allowed Access
Owner Only
Specifies that the owner of the chromatographic system and users with
Administrator privileges can access the selected chromatographic system.
Password Required
This field specifies that users enter a password before they can access the
selected chromatographic system across databases.
Password
Specifies a password (up to 30 alphanumeric characters) is needed to access
the selected chromatographic system across databases.
Confirm Password
Requires re-entry of the password to confirm that it was spelled correctly.
2 1
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Name Selection
Use this page to name the new chromatographic system and enter any
optional comments.
System Name
Identifies the chromatographic system in the Empower database and in
Configuration Manager. Up to 30 alphanumeric characters are allowed for the
system name. Use a meaningful system name in case other users have access
to the system.
Online
If all configured system instruments are available (not configured in another
online system), this brings the new system online. Default: selected.
System Comment
Displays optional remarks (up to 250 characters). Comments that describe
the chromatographic system, its components, and its intended use can help
others if they have access to the system.
2 2
CTC PAL Control for Waters Emp3 CDS
Installation Manual
5.6 Plates
The following sections show the properties of the most commonly used CTC
tray types; the VT32 and VT98.
Defining plate type can be done from the QuickSet, QuickStart and
Configuration Manager.
2 3
CTC PAL Control for Waters Emp3 CDS
Installation Manual
2 4
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Configure Plates
Before you select the correct plates in Empower, you need to know the
names of the trays and their configured tray types in the PAL itself. This
information can be gathered through the CTC PAL Control panel.
2 5
CTC PAL Control for Waters Emp3 CDS
Installation Manual
6. FIRST USE
6.1 Configure for use with multiple plates
By default Empower is not configured to work with multiple plates/trays.
In this the case the system will default to Tray1,
Tray1 which must be present in the
PAL.
To work with the plates defined in section 5.6 open the Empower QuickStart
or QuickSet interface and select the menu item Edit > Plates.
The following screen will appear:
2. Click on the ‘Plate Type Name’ of the first row and select a VT32
plate
3. Click on the ‘Plate Layout Position’ of the first row and enter name
of the VT32 tray as configured in the PAL. By default ‘Tray1’ or
‘Tray2’.
To select sample locations, select a plate, click on the vials in the right pane
and push the Insert or Append button.
2 6
CTC PAL Control for Waters Emp3 CDS
Installation Manual
In combination with Agilent instruments (e.g. A7890, A6890, A1100 etc.) the
PAL Cycle events require the following setup:
• Menu > Setup > Objects > Out Signals >Injected > Destination >
SW-Out1
2 7
CTC PAL Control for Waters Emp3 CDS
Installation Manual
Agilent 6890:
The same applies here, the General Tab should have the Front Injector and
Back Injector disabled (see below). Otherwise the system would indicate an
Agilent autosampler is configured, while there is no autosampler present.
Empower would respond with an Instrument Failure (Empower Message
Center > the injector configured in the method is not present).
7.4 Examples
A7890 and PAL:
QuickStart > Run Samples > Control Panel
2 8
CTC PAL Control for Waters Emp3 CDS
Installation Manual
2 9
CTC PAL Control for Waters Emp3 CDS
Installation Manual
8. KNOWN LIMITATIONS
8.1 Overlapped injection
As of now, time optimization for overlapped injections is limited to two
samples. This can be achieved by building a method based on the cycle
named HS_2-
HS_2 -Vial-
Vial -Overlap.
Overlap Sample lists should only contain methods based
on overlap cycles.
The order of instruments within the system configuration should always have
the preparing instrument before the analyzing instrument (please see
example below).
This feature is not available when the CTC PAL is used in combination with
the Agilent 6890 GC or the Agilent 7890 GC with Agilent Instrument Control
Software v2.5 and up (Instrument Control Software v2.01 is supporting this
feature) controlled by Empower. The reason for this is how the Agilent 6890
and 7890 is integrated and controlled by Empower.
The Cycle Editor is located in the Tools menu of the Configuration Manager.
Cycles can only be edited on the PC that is physically connected to the PAL.
This is because Empower is unable to store and distribute Cycles.
3 0
CTC PAL Control for Waters Emp3 CDS
Installation Manual
This conversion can be achieved by starting ‘Run Samples’ for the instrument
the method was intended for. Make sure the Control Panel Status is green
and displays ‘Ready’. Now open the method, save it and close it.
When editing a method in ‘Run Samples’, always make sure it is opened for
the correct instrument, because the method will only work for this instrument
or an instrument that is configured identically, once it’s saved.