S.N.

Feature Description/ Specification


1 Temperature range 300°C (minimum) to 2000°C or above

Temperature Accuracy 15° C to 35° C, ±2 °C or ±2 % of rdg, whichever is greater


Instantaneous Field of View
(IFOV) with standard lens 0.9 mRad or better
(spatial resolution)
2 Detector Type Focal Plane Array FPA, Uncooled Microthermal

3 Detector resolution: 384 x 288 pixels or higher

4 Field of view: 25° Hx 18.7°V or greater


0.5m or greater.
5 Minimum Focus Distance
The focus should be adjustable using personal computer software.
Tripod mounting with standard tripod mounting holes bottom and top
6 Mounting
mounting supported.
Main stream: 384 x 288 Pixels @ 30 Hz Frequency, 1.8 Mb Bandwidth.
7 Video
Sub stream:384 x 288 Pixels @ 30 Hz Frequency, 100 Kb Bandwidth.
Fully-Radiometric
Yes. 30Hz
Streaming
8 Wireless connectivity Required

9 Thermal sensitivity <50 mk


1 relay output: load capacity: 24 V, 1.5 A
1 optocoupled output: output capacity: 3.3 to 24 V, 35mA maximum output
10 Alarm Input/Output current.
1 channel optocoupled input: input capability: 3.3 to 24 V, 5mA-to-15mA
input current.
Compatible software for capturing images and recording Video through a
11 Analysis Software
PC installed with Windows 11 Pro.
12 Image Processing Non-uniformity correction, intelligent gain control
Should support 5 temperature measurement points, 10 temperature
Temperature measurement
13 measurement lines, 10 temperature measurement areas, support Modbus
area
output.
14 Power supply: The system must include the rated power supply.
Operating environmental
15 -20°C-125°C (sensor), 0°C-85°C (controller)
temperature:
16 Housing: Close housing made of stainless steel.
Compatible with the pyrometer must be provided.
17 Interface software:
The interface software must work on the windows operating system.
The pyrometer system should respond on ASCII commands (list of
18 Communication protocol:
commands for all the operations must be provided)
19 Mounting bracket: Suitable mounting bracket with adjustable angles
To block high-power laser radiation. The optical density of the filter at the
20 Special blocking filter: laser wavelength of 1060nm and attenuation of the signal at 2.3μm
wavelength may be mentioned.
4 Digit display indicator with controller
4 – 20 mA signal retransmission.
Accessories: Adjustable alarm switch relay – 1 Nos.
Work certificate: Should be included.
The part number and product catalog must be provided for qualification.