18460 Combiwash

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Specifications & more

CombiWash
>> System Specifications
>> Accessories

System Overview Cat. No. 18460


Analyzer type: 1 plate washer & shaker
2 plate incubator
Throughput (Max): 1 plate run wash & shake
Up to 2 plates incubation
Reaction / reading System 96 Well Microtiterplate
Operation mode Batch
CombiWash
Automatic microtiter plate washer and incubator

Reaction Unit / Incubation Capacity 1 chamber for up to 2 plates


Mixing capability Mixing in wash position
Incubation temperature 25° / 30° / 37.0°C fixed
Accuracy ± 5 % (Incubator plate)
Stability ± 5 % (Incubator plate)
Uniformity ± 1°C (incubator plate)
± 2°C at 25°C ambient temperature in liquid
Warm up time 15 min.
(ambient 25 °C to room incubation temp. for incubator plate)
Incubation time free chooseable (increment 1 minute)

Washing Capacity 3 x washbuffer bottles 2L


1 x waste bottle 2L
1 x distilled water system liquid bottle 2L
Wash head 8 channels (included)
Optional 12 channels (cat no 18460/5)
Plate type Flat ,round and V- bottom wells
Wash mode Plate and strip mode
Wash processing time 90 sec. (one time wash 12 channels)
Dispense volume 10-3000 µl / well in 1 µl steps
Wash accuracy ± 5 % at 300 µl
Residual volume < 5 µl
Fluid alarms Low wash buffer
Systemliquid
Waste full
Features 10 wash profiles
up to 99 wash cycles
0-24h shaking /soak time

Data Processing Interfaces RS 232


# of programable tests 100 wash protocols
User interface LCD Display, 10 soft buttons
System language English
Generals Dimensions (W x D x H) Instrument without any components: 46 x 40 x 20 cm
Space required for routine use: 56 x 60 x 40 cm
Packaging: 55 x 51 x 35 cm
43 x 26.5 x 31.5 cm
Weight Net: 10.00 kg
Gross: 12.65 kg
Electrical requirements Universal AC input: -~220 V, 50/60 Hz
±10 % (110V with transformer)
Power consumption: 80 VA without incubator
140 VA with incubator
Enviromental conditions Operating: temperature:18 – 30°C, humidity: < 85 % non condensing
Transport: temperature: -20 – 50°C, humidity: < 93 % non condensing

Accessories

Additional Accessories
Cat. No.
12-pin Manifold 18460/5
110 VA Transformer 18460/26

Standard Accessories (supplied with each instrument)


12-pin Manifold 8-pin Manifold
Quantity Cat. No.
Cat.No. 18460/5 Cat.No. 18460/4
User Manual 1 18460/1
8-pin Manifold 1 18460/4
Bottles with Cap 5 18470
Power Cable 1 18999
Fuses 2 18460/14
Set of Buttons Wash 1, Wash 2,
Wash 3, Waste, Water for Bottles 1 -------------
110 VA Transformer
Cleaning Wires 2 -------------
Cat.No. 18460/26
Dust Cover 1 -------------

Servicekits Quantity Cat. No.


Maintenance Kit (for 5 CombiWash) 1 18460/201
18460/2012-07 © 2012 Human GmbH