HV Impulse Analysis Option: Perception
HV Impulse Analysis Option: Perception
English
HBM GmbH
Im Tiefen See 45
64293 Darmstadt
Germany
Tel: +49 6151 80 30
Fax: +49 6151 8039100
Email: [email protected]
www.hbm.com/highspeed
Copyright © 2012
All rights reserved. No part of the contents of this document may be reproduced or transmitted in any form or by
any means without the written permission of the publisher.
2 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 3
4 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 5
Perception HV-IA
C Calculate Uncertainty 72
C.1 Introduction 72
D HV-IA Preferences 75
D.1 Introduction 75
D.2 Preferences dialog 76
D.3 Output formats 77
D.4 Recording location 78
D.4.1 Miscellaneous 79
D.5 Calculations 85
6 I2975-4.0 en
Perception HV-IA
1 Getting Started
1.1 Introduction
The High Voltage Impulse Analysis (HV-IA) option calculates and analyzes a
variety of parameters that are relevant in high-voltage impulse testing. The HV-
IA option is typically used as a stand-alone application for repetitive tests. This
option is focused on the results of tests and is tailored for ease of use.
The HV-IA option is a successor for the legacy HV-IA software that ran on
the Nicolet Accura and Sigma DSO's with ProView software.
Note Measurements and calculations comply with current and new standards,
including the K-factor filter for the revision of IEC60060-1 and -2.
I2975-4.0 en 7
Perception HV-IA
You can find the serial number of your key in Help About Perception
4 Click OK.
After the installation you can go to Help About Perception More... to see
all installed options.
You will need to restart the program before the changes take effect. The HV
Impulse Analysis option is now available.
8 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 9
Perception HV-IA
A1 Evaluation group
A2 Control group
A3 Digitizer setup group
A4 Toolbox group
10 I2975-4.0 en
Perception HV-IA
B Results area
The analysis settings and the results of a test will show in the results area,
enabling tolerance testing in the Analysis settings window will highlight
these results as shown below.
If tolerance testing is enabled in the Analysis settings then the results which
do not fall within the specifications of the testing parameters will show up
highlighted in yellow, see Figure 1.5, whereas a good test result will be
highlighted in green, see Figure 1.4. When tolerance testing is not enabled
the results are not highlighted.
I2975-4.0 en 11
Perception HV-IA
C Collection pane
This section explains the available controls and data show in the collection
pane.
C1 Collection name This contains the name of the collection and the
button New...
C2 Collection grid This area shows the current tests of a collection and
their status.
When a new collection is started the collection pane will appear as in Figure
1.7 "Start new collection area" on page 13.
12 I2975-4.0 en
Perception HV-IA
C4 Tests
C5 Terminals
Complete test
A Test has been done with no warnings.
B This test is on display.
Test error
A Test has been done with a warning.
B This test is on display.
C The next test will be done in the next cell.
Collection control
You can navigate in a collection by using keyboard arrows to move to
different cells. Double clicking or pressing enter will select a cell as active
or you can right click and click Select. If the black arrow , is not
pointing to start a test in the correct cell, right click and click Start next test
here.
If a cell is selected for a test and it is occupied, a warning dialog will come
up and informs you that it is not possible to start a new test because the
test storage location is not empty. Before you can continue you have to
empty the location, see Figure 1.8.
I2975-4.0 en 13
Perception HV-IA
D Display area
The main graphical area of the screen.
This area displays current waveforms and the controls and position axis to
navigate current data. Cursors can be found in their default position in the
top left of the area. Data sources can also be dragged into this area for
viewing. The display setup menu is accessed via a right click in this area.
14 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 15
Perception HV-IA
This is done before making a new collection as a collection can only use one
method of evaluation.
16 I2975-4.0 en
Perception HV-IA
Note If you are using the HV-IA sheet for the first time it is recommended to use option
C to get started.
After selecting your collection type you will be presented with a collection setup
dialog. In this dialog you will design your collection layout and properties.
I2975-4.0 en 17
Perception HV-IA
Number of terminals
Choose from a number of 1 to 20 terminals using the up and down arrows. In
the terminal window, terminals will fill up with results in numerical order until all
tests are complete and then the next terminal will start to fill up.
Collection name
Each collection should have a unique name. This name is also used for the
name of the file which contains all the measured and calculated data including
the system configuration.
Comment
Use this entry field to add additional comments and remarks to the collection.
This comment can be used for reporting.
After you Click OK the collection will be ready for tests and the main screen
collection pane will look like Figure 2.4.
18 I2975-4.0 en
Perception HV-IA
Notice that the right arrow in the upper left cell of the collection grid indicates
that the next test will be saved here.
I2975-4.0 en 19
Perception HV-IA
Note The control pane will display “waiting for trigger” and the current step to be filled
in the collection matrix.
The system is ready for a triggered input or event to start recording the impulse
data. A few seconds after the trigger a physical test is done, and a pop up dialog
will ask if the test is to be accepted or rejected.
1 Click Accept to save the current results to the collection, after this a new
test can be started.
2 Click Reject if for some reason you do not want to save the results to the
collection.
20 I2975-4.0 en
Perception HV-IA
At this point you can enter a test name to replace the automatically created
name "A1". This name is created from the current terminal name and test
number. If a test name is changed the system will remember the name, this will
now become the default name for the "A1" test position. A comment can also
be added to the Comment text box. Once a comment is added and accepted
the display will look like Figure 2.7.
Note The name can be reset by clicking Reset in the Collection... settings via the
sheet menu.
The upper left hand cell in the collection table now indicates that the test
has been done; the results are within tolerance and that the next test will be
done in the cell below ("A2").
Note If you cannot see the waveform click the star key in the display control area to
fit the waveform to the screen.
The results are always displayed as described in Figure 1.3 "Lightning impulse
test results (A)" on page 11 and the following figures.
I2975-4.0 en 21
Perception HV-IA
Note If you have selected the standard IEC 60060-1 Ed. 3.0 (2010-09) and you are
doing Lightning Impulse tests and you are evaluating a tail chopped signal
then an additional not chopped signal is needed to do the calculations. The
application assumes that the first test of a terminal is a not chopped signal and
will automatically be used. If this signal is not available then you can measure
it later and do a manual recalculation on the chopped signal afterwards.
Note The calculations will automatically detect if there is a major oscillation for
Lightning or Switching Impulse tests. In those cases the standard IEC 60060-3
(2006) will automatically be used to do the calculations.
22 I2975-4.0 en
Perception HV-IA
Click the collection manager button and the dialog in Figure 2.9 should
appear.
Within this dialog you have a selection of 4 buttons, first you must select a test
or row then:
I2975-4.0 en 23
Perception HV-IA
Click or double click the test in question to view the result data of this test.
Once highlighted right click and select Re-evaluate test(s). The tests will be
recalculated one test at a time.
24 I2975-4.0 en
Perception HV-IA
A Load or save
I2975-4.0 en 25
Perception HV-IA
Collection Manager
Opens the collection manager dialog for an overview of the active
collection, see Figure 4.1 "Add Recording dialog" on page 35 for more
information.
Add Recording
Adds an already existing recording to the active collection, see Figure 4.1
"Add Recording dialog" on page 35 for more information.
C Configuration defaults
26 I2975-4.0 en
Perception HV-IA
Analysis ...
Click Analysis in the sheet menu and the dialog in Figure 3.2 will be
shown:
Select this check box if you want to automatically accept tests. The dialog
in Figure 2.6 on page 20 will not be shown.
I2975-4.0 en 27
Perception HV-IA
The Enable Tolerance Testing option is off by default. This option will
enable the program to compare the analysis results to the entered
tolerances. The default tolerance percentages are specified by the IEC
standards.
l Full wave
l Chopped wave
l Front Chopped wave
Selecting the Switching Impulse tab will only apply one set of testing
parameters in all cases.
Note The HV-IA application supports three different types of available IEC working
standards. For Lightning and Switching Impulse you either can select IEC
60060-1 Ed. 2.0 (1989-11) or IEC 60060-1 Ed. 3.0 (2010-09), the (2001-09)
standard uses the k-factor filtering method. For Current Impulse the standard
IEC 62475 Ed. 1.0 (2010-09) will always be used.
28 I2975-4.0 en
Perception HV-IA
Collection...
Click Collection in the sheet menu and the dialog in Figure 3.3 will be
shown:
Also in this tab is Logical name. Each test is given a unique name. Select
the Terminal name and Test number using the related drop down boxes.
Now you can change the default logical name by typing in the new name
in the Logical name box.
Click Reset to set all logical names back to their defaults (A combination
of the Terminal name and Test number).
Note For more information on collections please see Figure 1.6 "Collection pane" on
page 12.
I2975-4.0 en 29
Perception HV-IA
Digitizer...
Click Digitizer in the sheet menu and the dialog in Figure 3.4 will be shown:
Current Channel
The current channel is optional and can be used to measure a current
signal. Make this field empty if you do not use a current signal.
30 I2975-4.0 en
Perception HV-IA
Both the voltage channel and the current channel have high voltage
attenuator and low voltage attenuator settings.
The High Voltage attenuator and Low voltage attenuator settings are
combined to reduce the signal at the digitizer input to less than 100 V. If
the HV attenuator ratio is 2000:1 and the LV attenuator ratio is 50:1 then
the total attenuation will be 100 000:1.
Note The Perception setting Technical Units Multiplier of the selected hardware
channel is automatically set to 100 kV/V.
When connected to Active hardware, the following will become active for
data input:
l Span (Voltage span/range)
This is the Peak-to-peak scale that the digitizer can measure at the
input. It defines the physical measurement range.
l Technical unit
The technical units (y scale) as displayed in displays, tables, meters
or reports.
l Trigger Level (The level used to trigger the acquisition)
If the trigger level is set to e.g. 100 kV then the acquisition is triggered
if the signal goes above 100 kV or below -100 kV.
Note The Perception Trigger mode is set to Dual and the primary level is set to the
entered trigger level while the secondary level is set to the negative value of
this entered trigger level.
I2975-4.0 en 31
Perception HV-IA
E Digitizer Extra
The Digitizer Extra tab is only available if the HVIA miscellaneous
preference Allow working with 2nd channel has been enabled. For
more information see the appendix “HV-IA Preferences - Miscellaneous”
on page 79.
Voltage Channel
This channel measures the voltage or current signal of the extra channel.
This channel defaults to the second channel of the first recorder.
32 I2975-4.0 en
Perception HV-IA
Both the voltage channel and the current channel have high voltage
attenuator and low voltage attenuator settings.
When connected to active hardware, the following will become active for
data input:
F Miscellaneous
If you have multiple monitors you can open the HV Impulse Analysis sheet
into a new window. This is only possible if the option Workbooks in your
key have been enabled. See “Perception Data Acquisiton” manual.
I2975-4.0 en 33
Perception HV-IA
4 Commands – How to
4.1 Introduction
This section describes, in short summaries, how to carry out some of the most
useful functions in the HV-IA sheet.
Click the Add recordings button the Add Recording dialog will be
shown.
34 I2975-4.0 en
Perception HV-IA
Select multiple files from the recordings folder using the Ctrl key and then click
Open. The file will then be loaded into the Collection Recording Loader where
you may select a Channel for it to be placed.
The application will look for empty test locations to place each of the new files.
If you would like to place the files in different positions do the following:
I2975-4.0 en 35
Perception HV-IA
Once you are satisfied click Load recordings. While loading, the status will
change to one of the following fields.
l Waiting The Load recordings button has been pressed, the recording is
waiting to be read.
l Reading The program is busy reading and doing calculations on the current
recording.
l Done Recording has been read, calculations have been done, the data is
now part of the collection.
Note Some recordings can contain multiple channels, if this is the case, select the
correct channel in the "U" and "I" columns.
36 I2975-4.0 en
Perception HV-IA
The system will perform the same calculations for the second channel as for
the first channel. The results of the second channel will be saved in a separate
test. This test is located on the right-hand side of the first test in the collection
grid. The test is also a different color (blue). Before you start any test, the light
blue cell background indicates that you are working with an extra channel.
After you have done the first measurement, the first cell and the one directly
next to it are both filled.
You can see the second channel results and other data by selecting its
corresponding cell in the collection grid, just as you do in normal operation.
I2975-4.0 en 37
Perception HV-IA
You cannot switch back to normal operation while a collection is open. First,
you need to close the collection and then you can enable or disable the Extra
Channel (see Figure 5.3).
Figure 5.3: Impulse Analysis Test Setup dialog with the Extra Channel disabled
A This selection can be modified after you close the current collection
You can only change the Enabled Extra Channel option if you close the
collection first (see Figure 5.4)
Figure 5.4: Impulse Analysis Test Setup dialog with the Extra Channel enabled
A Enabled Extra Channel
38 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 39
Perception HV-IA
All of the data sources; consisting of numbers, strings and waveforms are
stored here, the main sections can be expanded by clicking on the respective
plus sign.
The HV-IA data sources are structured like a tree as shown in Figure A.1. The
main branch for this sheet is called HVIA.
The sub nodes of this HVIA entry are A, B, C, … , Collection, LI. The nodes
A,B,C correspond to the number of used terminals and contain the calculation
results per test.
40 I2975-4.0 en
Perception HV-IA
The sub nodes per test contain all the calculation results for a specific test:
The variables marked by the icon are waveforms. These waveforms can be
measured or calculated data. The waveforms can be displayed by any
Perception display. If the name of the waveform ends with “_cmp” then we are
dealing with a normalized waveform, the maximum of a normalized waveform
is set to 1 and the minimum is set to 0
I2975-4.0 en 41
Perception HV-IA
For detail on the meaning of each variable please refer to the tables in the
Lightning impulse, Switching impulse and Current impulse parameters chapters
starting on page 46.
If a data source has the same name as another data source but ends with a k
or µ then this data sources refers to the same parameter; the only difference is
that the parameter value is expressed differently:
l Name ends with k: Parameter value is divided by 1000. For example 1000
-> 1 or 1000 V becomes 1 kV.
l Name ends with µ: Parameter value is multiplied by 1e06. For example
0.000001 becomes 1 or 0.000001 s becomes 1 µs.
The last node will be LI, SI or CI depending on the selected test/evaluation type.
This node represents the current selected test in the control area, see Figure
A.5. The LI, SI or CI nodes show the current or active test results, Figure A.5
shows test A2 is highlighted, this means that LI, SI or CI will reflect the results
obtained in node A2.
42 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 43
Perception HV-IA
44 I2975-4.0 en
Perception HV-IA
In the example above, the LI node will have the same “sub-node source
structure” as represented by the data source HVIA, see Figure A.6.
Depending on the standard and test/evaluation type used, the variables can
differ and will have another meaning. Therefore we will explain the different
cases.
I2975-4.0 en 45
Perception HV-IA
Before explaining each individual variable we will show a lightning impulse with
some characteristic measured information.
46 I2975-4.0 en
Perception HV-IA
The above display can be constructed with your current HV-IA application, see
following chapter for more details. In Figure A.8 we see the main parameters
of the LI calculations. The table below describes these and all other parameters.
Description of the data sources when dealing with lightning impulse
measurement, calculated with the standard IEC 60060-1 Ed. 2.0 (1989-11).
I2975-4.0 en 47
Perception HV-IA
48 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 49
Perception HV-IA
The following paragraph gives a basic overview of the used calculation method,
for exact details we refer to the above mentioned standard.
See Figure A.9, start with U(t) this is the recorded curve (red) and the base
curve Ub(t) (green).
50 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 51
Perception HV-IA
52 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 53
Perception HV-IA
54 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 55
Perception HV-IA
56 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 57
Perception HV-IA
58 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 59
Perception HV-IA
60 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 61
Perception HV-IA
Figure A.13: Review Sweep area with the calculated results (zoomed)
62 I2975-4.0 en
Perception HV-IA
The markers will use these formulas and they are defined as follows:
I2975-4.0 en 63
Perception HV-IA
B.1.2 Challenge
The main challenge for any measurement equipment is to operate as safely
and accurately as possible in severe electromagnetic environments. High
voltages in combination with fast rise/fall times create additional problems.
Each piece of wire acts as an antenna. Each series resistor in a connection
between two points creates a potential difference and hence a voltagedrop,
which in turn results in current flow. The higher the voltages and the faster the
rise/fall times are, the more serious the effects.
64 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 65
Perception HV-IA
66 I2975-4.0 en
Perception HV-IA
All of the above-mentioned voltages at the generator output, the test object, the
voltage divider input and output are the potential difference between the
terminal and an earth. To have the impulse attenuator and the digitizer working
on the same reference level, they also have to be connected to the same
protective earth. For this purpose, the impulse attenuator and the Sigma
100 HV are equipped with protective earth screw terminals, so a proper
protective earth connection with a copper strip can be made. The protective
earth connection serves additional purposes. First, the metal housing is acting
as a Faraday cage and the solid connection to protective earth allows offloading
of the energy picked up from electromagnetic fields and keeps the inside free
of disturbances.
With HBM’s battery operated and fiber optic isolated digitizers ISOBE5600t,
two new test set-up’s can be realized:
l Measurement point in control room
l Measurement point in test cell
For both of these set-up’s a few things need to be taken into account.
I2975-4.0 en 67
Perception HV-IA
68 I2975-4.0 en
Perception HV-IA
In this set-up, all the measurement equipment is still located in the well shielded
control room. The PC and the ISOBE5600m (receiver with memory) are mains
powered and earthed via this mains connection. The ISOBE5600t (battery
operated digitizer) is connected to the ISOBE5600m only by means of fiber
optic cables, so there is NO electrical connection between them. The digitizer
does get the input signal from the impulse attenuator, which is properly
grounded to the test cells earth connection.
I2975-4.0 en 69
Perception HV-IA
The ISOBE5600 set up with measurements done from inside the test cell
70 I2975-4.0 en
Perception HV-IA
N ISOBE5600m (Receiver)
O Shielded compartment in test cell
In this set-up, half of the measurement equipment is still located in the well
shielded control room, the PC and the ISOBE5600m (receiver with memory).
They are mains powered and earthed via the mains connection.
Ideally, the ISOBE5600t, the impulse attenuator and the connecting short cable
are all placed together in the same shielded and earthed box or compartment.
The size and performance of such a box is up to the test cells requirement and
part of the users installation. HBM does not offer such boxes.
I2975-4.0 en 71
Perception HV-IA
C Calculate Uncertainty
C.1 Introduction
The method to calculate the SW uncertainty per parameter in this appendix is
based on the method described in Appendix B of IEC 61083-2 Ed. 2.0 (CDV).
The values uB71 listed in the table below show the standard uncertainty of the
algorithms used in the HV Impulse analysis option of the Perception software
for each kind of waveshape and each parameter. The values uB72 show the
standard uncertainty of the reference values according to tables B.1 to B.6 in
the IEC 61083-2 Ed. 2.0 (CDV).
The last column with the values uB7 shows the resulting standard uncertainty
as the result of uB71 and uB72.
The symbol uB7 follows the numbering used in IEC 60060-2:2010 (clauses
5.2.1.3 to 5.9).
72 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 73
Perception HV-IA
NOTES
1 Each calculated parameter for every waveform is within the acceptance
limits of the standard.
3 The settings used (sampling rate, resolution and noise) of the TDG is listed
for each group of waveforms. The choices were made to be in line with the
hardware specifications (ISOBE5600t and 6600HV digitizer) as well as to
take worst case scenarios of user’s choice into account (sampling speed
for slow waveforms like SI and OSI). The algorithms were verified with
higher sampling speeds for SI and OSI as well.
74 I2975-4.0 en
Perception HV-IA
D HV-IA Preferences
D.1 Introduction
The HV-IA application can be modified by using special HV-IA preferences. You
can modify not only the user interface, but some behavior as well.
I2975-4.0 en 75
Perception HV-IA
76 I2975-4.0 en
Perception HV-IA
● Nr. digits time results: Set the number of digits for all time results in the
Lightning impulse test results area.
● Nr. decimals time Set the number of decimals for all time results in
results: the Lightning impulse test results area.
● Nr. digits U/I results: Set the number of digits for all Voltage and Current
results in the Lightning impulse test results area.
● Nr. decimals U/I Set the number of decimals for all Voltage and
results: Current results in the Lightning impulse test results
area.
I2975-4.0 en 77
Perception HV-IA
Enabling this preference creates a new folder to store the individual recordings
(*.pNRF files) for each test when a new collection is started. The new folder will
be a sub-folder of the base folder as defined in the preferences recordings
location dialog. The collection name is used for defining a sub-folder name.
HINT/TIP
78 I2975-4.0 en
Perception HV-IA
D.4.1 Miscellaneous
This section contains various preferences, most of which are related to the
HV-IA application user interface.
I2975-4.0 en 79
Perception HV-IA
80 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 81
Perception HV-IA
82 I2975-4.0 en
Perception HV-IA
I2975-4.0 en 83
Perception HV-IA
84 I2975-4.0 en
Perception HV-IA
D.5 Calculations
The standard IEC 60060-1 Ed 3.0 (2010-09) tries to generate a test signal using
a double exponential curve fitting algorithm. If this curve fitting fails, the
calculations cannot be performed in accordance with this standard.
Calculations are performed in accordance with the new version of the IEC
60060-1 Ed 3.0 standard by default, although calculations can be performed in
accordance with the older version of this standard. By enabling the following
setting, the calculations will be automatically performed using the older
standard IEC 60060-1 Ed. 2.0 (1989-11) when the new standard is selected
and the curve fitting has failed.
I2975-4.0 en 85
Perception HV-IA
Index
C I
N
D
New Test Set-up
Data sources
Measurement point in control room ................. 68
Introduction ...................................................... 40
Measurement point in test cell ......................... 70
H
O
HV Impulse Analysis Menu .................................. 25
Open a collection .................................................. 17
Accept test without confirmation ...................... 27
Collection name ............................................... 18
Collection menu items ..................................... 26
Comment ......................................................... 18
Collection Setup .............................................. 29
Number of terminals ........................................ 18
Configuration defaults ..................................... 26
Tests per terminal ............................................ 18
Digitizer Extra .................................................. 32
Output formats ...................................................... 77
Digitizer Setup ................................................. 30
Load or save .................................................... 25
Miscellaneous .................................................. 33 P
Test Analysis Setup ......................................... 27
HV Impulse Test Set-up Preferences dialog
Measurement Equipment Set-up ..................... 65 HV-IA settings ................................................. 76
HV-IA uncertainty summary ................................. 72
HVI-IA Preferences
Introduction ...................................................... 75
86 I2975-4.0 en
Perception HV-IA
Warranty ................................................................. 3
I2975-4.0 en 87
88 I2975-4.0 en
Head Office
HBM
Im Tiefen See 45
64293 Darmstadt
Germany
Tel: +49 6151 8030
Email: [email protected]
France
HBM France SAS
46 rue du Champoreux
BP76
91542 Mennecy Cedex
Tél:+33 (0)1 69 90 63 70
Fax: +33 (0) 1 69 90 63 80
Email: [email protected]
UK
HBM United Kingdom
1 Churchill Court, 58 Station Road
North Harrow, Middlesex, HA2 7SA
Tel: +44 (0) 208 515 6100
Email: [email protected]
USA
HBM, Inc.
19 Bartlett Street
Marlborough, MA 01752, USA
Tel : +1 (800) 578-4260
Email: [email protected]
PR China
HBM Sales Office
Room 2912, Jing Guang Centre
Beijing, China 100020
Tel: +86 10 6597 4006
Email: [email protected]