Assignment2 solution (1)
Assignment2 solution (1)
a) Qualitative Analysis
b) Quantitative Analysis
The amount or concentration of analytes in the sample.
c) Sensitivity
The per unit analyte detector responds, i,e slope of the calibration curve.
d) Detection Limit
Smallest amount (concentration) of analytes that generate a detector response significantly different from
that of a blank.
e) Selectivity
The ability of the analysis to differentiate/distinguish the analytes from other chemical species co-exist in
the sample
f) Accuracy
A measure of how close the measured result is to the true amount/concentration
g) Precision
Reproducibility of the result from multiple/replicated measurement
h) Robust
Results unaffected by small changes in experimental conditions
i) Calibration Check
It is by analyzing a standard solution of known conc./formulation regularly to make sure our in
k) Recovery
Percentage of analyte recovered/extracted from the sample to the solution for analysis after accounting for
analyte instability/decomposition and recovery loss during the sample preparation.
l) Reagent Blank
A sample containing all components except analyte but not subject to sample preparation procedures.
m) Field Blank
A sample containing all components except analyte, including the sample preparation and exposed to the
site of sampling.
Q2.
Given Signals:
0.0054, 0.0062, 0.0060, 0.0046, 0.0056, 0.0052, 0.0044
Mean Calculation:
0.0054+0.0062+0.0060+0.0046+0.0056+0.0052+0.00447
Mean= = 0.00534
7
(0.0062−0.00534)2 = 0.0000007396
(0.0060−0.00534)2 = 0.0000004356
(0.0046−0.00534)2 = 0.0000005476
(0.0056−0.00534)2 = 0.0000000676
(0.0052−0.00534)2 = 0.0000000196
(0.0044−0.00534)2 = 0.0000008836
Standard deviation:
0.00000269646
𝑠=√ ≈ 0.000670
6
• Mean: 0.00534
To find the signal detection limit, we can use the standard deviation of the blanks.
Given Blanks:
0.0006, 0.0012, 0.0022, 0.0005, 0.0008, 0.0011, 0.0010
Mean of Blanks:
0.0006 + 0.0012 + 0.0022 + 0.0005 + 0.0008 + 0.0011 + 0.00107
𝑀𝑒𝑎𝑛𝑏𝑙𝑎𝑛𝑘𝑠 = ≈ 0.001057
7
Standard Deviation of Blanks:
𝑠=
(0.0006−0.001057)2 +(0.0012−0.001057)2 +(0.0022−0.001057)2 +(0.0005−0.001057)2 +(0.0008−0.001057)2 +(0.0011−0.001057)2 +(0.00107−0.001057)2
√ =
6
0.000565.
Detection Limit:
The detection limit is often calculated as:
a) Blank Titration
A titration performed without the analyte of interest, but with all other elements. It is used to correct the
difference between the end point and equivalent point.
b) Back Titration
A known excess amount of a standard reagent is added to react with analytes in the solution. The excess
amount of the standard reagent is determined by another(back) titration. From the difference in amount
of the reagent added and that consumed in the back titration, the amount of reagent consumed by
reactivity with the analyte can be determined.
Q4.
Weight percentages:
Given Data:
Volume of seawater = 6.23 mL
Moles of Carbon:
Since each mole of CO2 corresponds to one mole of carbon: Moles of C = 0.0000540 mol
Mass of Carbon:
Mass of C=Moles of C×Molar mass of C=0.0000540 mol×12.01 g/mol≈0.000649 g
Convert to ppm:
Since 1 mg/L = 1 ppm in water:
Convert to g/L:
104.2 𝑔/𝐿
𝐶𝑜𝑛𝑐𝑒𝑛𝑡𝑟𝑎𝑡𝑖𝑜𝑛 (𝑔/𝐿) = = 0.1042 𝑔/𝐿
1000
Volume in Liters:
a) Volumetric Flask
b) Conical Flask
Water/solvent there will not affect the number of moles of the analyte in the conical flask.
Q7:
Qualitative Analysis
Analyte standard added increases signal with very similar detection response, eg, retention time, as that
of analyte in the sample analysis (original), giving evidence that the signal was originated from the target
analyte.
Standard addition is primarily a quantitative technique. While it doesn't directly identify unknown
substances (qualitative analysis), it can be used in conjunction with qualitative techniques. For
example, if you suspect a particular analyte is present, you could use standard addition to confirm
its presence and determine its quantity. A positive response (increase in signal proportional to the
added standard) would provide further evidence for the presence of the suspected analyte.
Quantitative Analysis
Known amount of analyte added to the sample containing analyte (at the concentration to be determined),
the amount of signal increase (in the presence of very similar matrix effect) was used to estimate the
amount of analyte in the original sample by correlating the signal in the original sample analysis.
Standard addition is a powerful tool in quantitative analysis because it directly addresses matrix
effects. By adding the standard directly to the sample matrix, the calibration occurs within the
matrix itself, effectively compensating for any interferences. This leads to a more accurate
determination of the analyte concentration.
Q8
Blind sample:
A blind sample is a sample of known concentration given to the analyst who performs the tests as an
unknown value. It is a QC measure to help eliminate bias introduced by the analyst who knows
concentration or composition of the calibration check.
A blind sample is a sample of known concentration (determined independently or by another lab) that's
presented to the analyst as if it were an unknown. The analyst is unaware of the true concentration.
How it helps in quality assessment: Blind samples assess the accuracy and precision of an
analytical method and the proficiency of the analyst. By comparing the analyst's determined
concentration to the known true value, you can evaluate:
Accuracy: How close the measured value is to the true value. A large difference indicates
systematic errors in the method or analyst technique.
Precision: How reproducible the results are. If multiple blind samples are analyzed, the
consistency of the results reflects the method's precision.
Method Validation: CRMs are used to demonstrate that a method is fit for its intended purpose by
showing that it can accurately measure the analyte in a relevant matrix.
Quality Control: Regular analysis of CRMs helps monitor the ongoing performance of a method
and identify any drifts or biases that may develop over time.
Proficiency Testing: Laboratories often participate in proficiency testing programs where they
analyze CRMs and compare their results to other labs, ensuring comparability and high standards.
A fortification recovery test is adding known concentration of analyte to a sample (blank, eg). The amount
found after all experiment steps is calculated as a test of method recovery.
In a fortification recovery test, a known amount of the analyte (the "spike") is added to a sample, and the
sample is then analyzed using the chosen method. The percentage of the added analyte that is recovered
is calculated.
How it helps in quality assessment: This test assesses the method's ability to accurately measure the
analyte in the presence of the sample matrix. It helps to evaluate:
Matrix Effects: A low recovery percentage suggests that the sample matrix is interfering with the
analysis, either by suppressing or enhancing the signal. This helps identify potential issues and
optimize the method to minimize matrix effects.
Method Accuracy and Precision: Consistent and high recovery percentages (ideally close to
100%) indicate that the method is accurate and precise in the specific sample matrix.
Linearity: Performing spike recovery tests at different concentration levels helps assess the
linearity of the method over the desired analytical range.