OXILYSER 3: DIGITAL PASSIVITY METER

• Easy measurement
DISCOVER THE DIFFERENCE on any surface,
horizontal or vertical

• LED color and scale


Easy and accurate passivity measurement of (0 - 100) indication
austenitic and duplex stainless steel surfaces
• Pre programmed for
most common
The corrosion resistance of stainless steels depends on the quality of a very thin
stainless steel types
oxide layer, the so called ‘passive layer’. Maintaining this passive layer is one of
the most important aspects in maintaining the quality of stainless steel. Damage
• Robust design
of the passive layer, in time, will cause severe corrosion. Therefor preserving the
passive layer is a must. With the Oxilyser 3 you measure passivity fast and
accurately. Besides a 0-100 passivity scale, a LED (green/red) will indicate
passivity or not. The Oxilyser 3 helps you to avoid corrosion problems.

PRINCIPLE
The principle of the Oxilyser 3 is based upon measuring the rest potential
of the stainless steel surface combined with a sophisticated algorithm for
evaluation. The electrolyte is a non toxic organic acid and filter paper strips
function as electrolyte bearer.
This method is the most direct way for measuring passivity. The scale of the
LCD display is divided into 100 units. Passivity of three stainless steel groups
can be measured:

1. 0% Molybdenum (AISI 304, 321, 304L, Wst. Nr. 1.4301, 1.4541, etc.).
2. 2% Molybdenum (AISI 316, 316L, 316Ti, Wst. Nr. 1.4404, 1.4571, etc.).
3. Duplex stainless and 6% Mo steel such as Wst. Nr. 1.4462, duplex 2205,
duplex 2507, 254 SMO and 1943 hMo.

APPLICATIONS
• Checking effectivity of pickling and passivating treatment
• Checking passivity after unloading tanks with aggressive fluids like acids
• Checking passivity after cleaning/rinsing, e.g. stainless steel storage
tanks, transport tanks (shipping, industry, trucks, etc)

ADVANTAGE
The Oxilyser replaces the palladium chloride test as it is a more reliable and
quick standard for passivity testing. Direct measurement, built in algorithm for
passivity testing. The unit operates on a standard industrial quality 9V battery.
SPECIFICATIONS
• Objective result on LCD display. Scale 0 (active) to 100 (fully passive)
• Passive or not LED passivity indication for three stainless steel groups
• Measuring procedure fits into quality control and assurance procedures
and corrosion monitoring systems.

OXYLISER KIT CONTENT:


• Oxilyser 3 base unit with 9V battery OXILYSER 3
• Shockproof Epoxy Double Bridge Ag/AgCl2 reference electrode. Gel filled, PASSIVITY METER
this makes measuring up side down and on vertical surfaces possible
• Red connector clamp
• Box with filter paper (for ± 100 measurements) • Easy measurement
• Potassium Chloride 60ml bottle (for conditioning of electrode) on any surface,
• Electrolyte refill 60 ml bottle (for ± 100 measurements) horizontal or vertical
• D.I. water (rinsing) liquid 60 ml bottle
• LED color and scale
INSTRUCTIONS FOR USE: (0 - 100) indication
1. Ensure the surface is clean (and degreased) before testing
• Pre programmed for
2. Connect electrode and connector clamp to most common
Oxilyser 3 unit (respect colour coding of connectors) stainless steel types

3. Select corresponding grade of stainless steel to be • Robust design


tested (304, 316 or duplex); in case unknown:
select 304 or apply molybdenum drop test

4. Determine test area on stainless steel surface

5. Take out a piece of filter paper and wet with a few


drops of electrolyte, apply strip on surface to be tested

6. Remove transparent cap from electrode, clean


electrode tip with some drops of demineralised water

7. Place electrode on wetted filter paper and connector


clamp on surface within 10 cm of the filter paper,
ensure the filter paper stays very wet

8. Wait for 10 seconds until the value on the LCD display


stabilizes and the LED light gives a green (pass) or
red (fault) value

9. After use: Fill transparent cap with potassium chloride and


place cap on electrode (to prevent it from drying out)

10. Disconnect cables and put components back in case

Passivity Value on LCD Display Interpretation


After ~3-10 seconds
0% Mo 2% Mo 6% Mo/
(304 a.o.) (316 a.o.) duplex
< 30 < 35 <40 AND ▼ Active and contaminated, surface will not re-passivate.
30-55 35-60 40-65 AND ▲ Wait until stable value has been reached.
30-55 35-60 40-65 AND ▼ OR ═ Bad passivity, surface will not re-passivate.
>55 > 60 >65 AND ▼ Wait until stable value has been reached.
>55 >60 >65 AND ▲ OR ═ Good passivity.

This information is based on data that we believe to be accurate and reliable and is intended to be used by persons having the knowledge and skill to
apply this product properly. We assume no responsibility for results or damages incurred from their use by the Buyer. No li ability (including liability for
negligence) will attach to Marine Care or its servants or agents arising out of or in connection with or in relation to this product information.