Velomitor CT Piezo - Velocity Transducer: Operation Manual
Velomitor CT Piezo - Velocity Transducer: Operation Manual
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
Contact Information
The following contact information is provided for those times when you cannot contact
your local representative:
ii
CDG/125389/000/06
Additional Information
Product Disposal Statement
Customers and third parties, who are not member states of the European Union, who are
in control of the product at the end of its life or at the end of its use, are solely
responsible for the proper disposal of the product. No person, firm, corporation,
association or agency that is in control of product shall dispose of it in a manner that is
in violation of any applicable federal, state, local or international law. Bently Nevada, Inc
is not responsible for the disposal of the product at the end of its life or at the end of its
use.
iii
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
Contents
Contact Information ....................................................................................................................................ii
1. Operating Information ............................................................................... 1
1.1 Velomitor CT Application.................................................................................................................................. 1
1.2 Principle of Operation ........................................................................................................................................ 2
1.3 Compatible Monitoring Systems .................................................................................................................. 3
1.4 Customer Service ................................................................................................................................................ 3
1.5 Cable Options ........................................................................................................................................................ 4
2. Installation and Removal .......................................................................... 9
2.1 Receiving Inspection .......................................................................................................................................... 9
2.2 Positioning the Velomitor CT .......................................................................................................................... 9
2.3 Mounting Options ................................................................................................................................................ 9
2.4 Machine Case Surface Preparation ......................................................................................................... 10
2.4.1 Stud Mounting .............................................................................................................................................. 11
2.4.2 Adhesive Stud Mounting.......................................................................................................................... 11
2.5 Installation Procedures .................................................................................................................................. 11
2.5.1 Installing the Case ...................................................................................................................................... 11
2.6 Routing the Interconnect Cable ................................................................................................................ 12
2.7 Routing Flexible Metal Liquid-Tight Conduit ........................................................................................ 13
2.8 Sealing the Interconnect Cable ................................................................................................................. 16
2.9 Power and Signal Connections .................................................................................................................. 16
3. Maintenance .............................................................................................. 20
3.1 Test Setup............................................................................................................................................................. 21
3.2 Performance Test Procedure ...................................................................................................................... 22
3.3 Polarity Test Procedure.................................................................................................................................. 23
3.4 Transducer Cleaning Instructions ............................................................................................................ 23
4. Field Testing and Troubleshooting ........................................................ 24
4.1 Fault Indication #1 Cause/Solution ......................................................................................................... 24
4.2 Fault Indication #2 Cause/Solution ......................................................................................................... 25
4.3 Fault Indication #3 Cause/Solution ......................................................................................................... 25
5. Specifications............................................................................................. 26
5.1 Velomitor CT ........................................................................................................................................................ 26
5.1.1 Electrical .......................................................................................................................................................... 26
5.1.2 Environmental .............................................................................................................................................. 27
5.1.3 Mechanical ..................................................................................................................................................... 27
5.2 Mechanical Outline .......................................................................................................................................... 28
5.3 Velomitor CT Operating Range (English)................................................................................................ 29
5.4 Velomitor CT Operating Range (Metric) ................................................................................................. 30
5.5 Velomitor CT Amplitude Response ........................................................................................................... 31
5.6 Velomitor CT Phase Response.................................................................................................................... 32
5.7 Velomitor CT Noise Floor Curve ................................................................................................................. 32
6. Accessories ................................................................................................ 33
6.1 Mounting Hardware ........................................................................................................................................ 33
6.1.1 Adhesive Mount Base Kits ...................................................................................................................... 33
6.1.2 Spare Mounting Adapters ....................................................................................................................... 33
iv
CDG/125389/000/06
6.1.3 Fittings .............................................................................................................................................................. 35
6.1.4 Teflon-Coated Stainless Steel Armor ................................................................................................. 35
6.1.5 Flexible Metal Conduit ............................................................................................................................... 35
CDG/125389/000/06
CDG/125389/000/06
Section 1 - Operating Information
1. Operating Information
1.1 Velomitor* CT Application
Velocity transducers measure machinery casing vibration. They are typically used
on machines where the use of eddy current proximity transducers is not practical.
Typical applications include fans, cooling towers, and pumps. The Velomitor CT
transducer (Velomitor Cooling Tower transducer) is suited for these applications.
The Velomitor CT transducer is designed to monitor vibration frequencies ranging
from 1.5 Hz to 1 kHz. The calibrated scale factor of the transducer is typically 3.94
mV/(mm/s) (100 mV/(in/s)) and can measure velocities up to 63.5 mm/s pk (2.5 in/s
pk).
Application Alert
If you measure casing vibration to protect a machine, evaluate the
usefulness of the measurement for each application. Most common
machine malfunctions, such as unbalance or misalignment, occur on the
rotor and originate as an increase (or at least a change) in rotor vibration.
In order for any casing measurement alone to be effective for overall
machine protection, a significant amount of rotor vibration must be
faithfully transmitted to the machine casing or mounting location of the
transducer. In addition, locate the transducer on the bearing housing or
machine casing carefully. Improper installation may decrease the
transducer amplitude and frequency response or generate false signals
that do not represent vibration on that particular machine.
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
The Velomitor CT transducer is a single unit (see Figure 1-1). The transducer
consists of a piezoelectric shear-mode element and electronics. When subjected
to machinery vibration, the shear-mode assembly exerts a force on the
piezoelectric ceramic generating an acceleration signal proportional to that force.
This signal is then internally amplified and integrated to produce a low-noise
output signal proportional to velocity.
Application Alert
A piezoelectric velocity transducer subjected to a sudden mechanical
impulse may generate a low frequency signal that does not represent
actual machinery vibration. This signal could trip alert or danger relays.
CDG/125389/000/06
Section 1 - Operating Information
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
Table 1-1 describes the Bently Nevada cables to use with the Velomitor CT.
Figures 1-2, 1-3 and 1-4 show the configuration of the cable and connector. You
may use either a Bently Nevada cable or other comparable cable. If you use
another cable, be sure that it is compatible with the environment
CDG/125389/000/06
Section 1 - Operating Information
Part Number CB2W100 - AXXX (See Note)
2 3
1 4
Note
OPTIONS ARE
(1) Cable
(2) Clamping ring
(3) Adapter seal (3/4” flats)
(4) O-ring
(5) Connector body
Figure 1-2: Splash Proof Cable Assembly
The cable assembly is supplied with the clamping ring on the cable. The o-
ring (replacement p/n 04200063) and adapter seal (replacement p/n
02200677) must be installed on the transducer according to Figure 1-2.
1. Put the o-ring over the MIL-C 5015 connector.
2. Install the adapter seal over the o-ring on the MIL-C-5015 connector
and tighten approximately to 30 - 50 in-lb (3.4 - 5.6 N-m). The adapter
seal is supplied with a pre-applied thread-locker on the internal
threads.
3. Apply a thin coat of silicon grease to the external threads of the
adapter seal.
4. With the clamping ring on the cable, align the pins and keyway and
install the cable on the transducer.
5. Tighten the clamping ring onto the adapter seal.
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
6. Finish the blunt cut end as described in Section 2.
CDG/125389/000/06
Section 1 - Operating Information
Note
When removing the cable be sure to grab the connector body. DO NOT
PULL ON THE CABLE. If the adapter seal is removed, fresh thread-locker
should be applied for re-installation. Additionally, apply a small amount of
dielectric grease such as Dow Corning 4 Electrical Insulating Compound
onto the cable connector for maximum environmental protection (see
Figure 2-1)
NOTE
ORDER IN INCREMENTS OF 1.0 FOOT (0.30 m)
2 5 25 FEET (7.6 m)
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
PART NUMBER 84661 - (SEE NOTE)
NOTE
2 5 25 FEET (7.6 m)
CDG/125389/000/06
Section 2 - Installation and Removal
The standard and plate studs require you to drill and tap mounting holes using
standard drilling and tapping tools. The adhesive stud uses an epoxy adhesive to
hold the stud in place. The threaded mounting options are the preferred method
for mounting the sensor because these mountings are more reliable than the
adhesive-mount method. You may also fabricate your own mounting hardware
to meet specific needs, such as for accessing locations that have limited space.
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
This table describes the required conditions for mounting the three stud types.
Table 2-1: Required Conditions for Mounting Studs
mm 0.254 TIR 0.254 TIR 0.254 TIR Minimum radius of curvature is 127
Surface mm (5 in).
Flatness in 0.010 TIR 0.010 TIR 0.010 TIR Remove paint and grease.
10
CDG/125389/000/06
Section 2 - Installation and Removal
2.4.1 Stud Mounting
Mounting studs couple the transducer to the mounting surface. Ideally, this
coupling should be as "stiff" as possible for optimum frequency response.
Standard studs are acceptable for most applications where the mounting surface
is smooth and relatively flat. A smooth, flat surface ensures a stiff transducer
coupling to the mounting surface under the torque specifications listed in Table 2-
1.
Plate studs are appropriate for applications that require an overall height
increase (approximately 0.25 in) or where mounting surfaces are slightly coarse or
slightly curved. Coarse, curved mounting surfaces decrease coupling stiffness
and could result in less than optimum frequency response. To address this
situation, the plate stud specifies a high torque to the mounting surface. The
transducer is then torqued to the plate stud as listed in Table 2-1. The higher
torque fastening the plate stud to the "rough" mounting surface improves
coupling stiffness, which in turn improves frequency response.
Adhesive studs are packaged in a kit, which contains all materials and
instructions needed for installation.
CAUTION
Contact with rotating
machinery could cause severe
injury. Be careful. Keep a safe
distance.
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
Step 1 — Verify that the ambient conditions do not exceed the maximum ratings
of the transducer listed in Table 2-1.
Step 2 — Verify that the machine's surface temperature does not exceed the
maximum temperature rating of the transducer listed in Section 5.1.2.
Step 3 — Decide which mounting stud you will be using (refer to Mounting Options
section).
Step 4 — Ensure that the mounting site is clean and dry and has no obstruction
within the diameter listed in the Table 2-1.
Step 5 — Drill and tap the mounting hole to the dimensions required by the
mounting stud. Drill the hole so that the sensitive axis of the mounted transducer
will be perpendicular to a tangential plane on the machine casing. For best
results, the hole should be within 1° of perpendicular (refer to Machine Case
Surface Preparation section on page 10).
Step 6 — Apply one drop of Loctite® 242 or an equivalent adhesive to each
threaded end of the mounting stud.
Step 7 — If using a standard stud, thread the stud into the base of the transducer
until the stud hits bottom.
Step 8 —Thread the transducer into the tapped machine casing and torque
according to Table 2-1.
CDG/125389/000/06
Section 2 - Installation and Removal
• Apply a 1/8” thick bead of Dow Corning 4 Electrical Insulating Compound
to cover the mating connector for maximum environmental protection
(see Figure 2-1). Over-application will prevent proper pin engagement.
2
(1) 3 mm (1/8”) thick DC 4
(2) Connector face
Figure 2-1: Apply Dielectric Grease to Cable Connector
Note
Do not connect to the transducer case with hard conduit. This will affect
frequency response and attenuate vibration amplitudes.
Follow these instructions when routing cable through conduit (see Figure 2-2):
Step 1 — Connect one end of the conduit to the protective enclosure or structure
in which the monitor is mounted.
Step 2 — Temporarily connect the other end of the conduit to a structure near the
transducer. This will help when pulling cable through conduit.
Step 3 — Remove sharp edges from conduit to prevent physical damage to the
integral cable.
Step 4 — Slip a 1/2" male NPT fitting (Bently Nevada p/n 03839201) over the
interconnect cable and thread the fitting into the transducer conduit adapter cap.
Tighten the fitting.
13
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
Note
The transducer conduit adapter cap has sharp threads. Exercise care to
avoid physical damage to the cable during installation.
Step 5 — Use a "fish" line to carefully pull the interconnect cable through the
conduit.
Step 6 — Tighten the conduit compression fitting to the 1/2" NPT fitting on the
transducer cap.
Step 7 — Secure the conduit at 0.6 m (2 ft) intervals to prevent excessive vibration.
Step 8 — Trim and secure excess, unfinished interconnect cable. If trimming,
maintain a service loop in the cable.
14
CDG/125389/000/06
Section 2 - Installation and Removal
10
6 8
5
4
3
2
7
11
(1) Machine
(2) Transducer
(3) Transducer cap
(4) 1/2” male NPT fitting
(5) 1/2” compression fitting
(6) 1/2” flexible conduit
(7) Moisture trap
(8) Junction box
(9) Monitor
(10) Rigid conduit
(11) Drain
15
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
Routing Armored Cable
If flexible, metal conduit cannot be used, the cable can be routed through
Teflon®-coated stainless steel armor. Use the steps in “Routing Flexible Metal
Liquid-Tight Conduit” for routing armor. Route the cable through protected areas
to reduce the chance of damage.
Note
In Section 6 - Installation Accessories, the 3/8" armor compression fitting,
Bently Nevada part number 03850000, will fit most Teflon®-coated 3/8"
stainless steel armor that has a maximum outer diameter, including the
teflon® thickness, of 0.543 inches.
Figure 2-3 shows a simple block diagram of the transducer. The internal circuitry
automatically sets the DC output voltage bias from the constant current. The DC
bias and AC signal appear between "A" and "B". Compatible Bently Nevada
monitoring systems power the Velomitor CT transducer without additional
external circuitry.
16
CDG/125389/000/06
Section 2 - Installation and Removal
1 5
2 3 6
7 9
4
8
(1) Velomitor CT
(2) Crystal
(3) Electronics
(4) Shielded cable
(5) Shield, braided wire
(6) “A”, white wire
(7) “B”, black wire
(8) 3mA current diode 1N5309 (BN P/N 00643485)
(9) External DC voltage supply (current limited < 10 mA)
Figure 2-3: Block Diagram of Velomitor CT Transducer
17
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
The monitor end of some interconnect cables will be prepared by the factory. The
cable consists of a white lead, black lead, and braided shield. The white lead will
be referred to as connection "A" and the black lead will be connection "B". Steps 1
through 7 show how to finish the cable, if necessary. Follow these instructions for
connecting the transducer to the monitor:
1 2
Trim or secure the excess cable after Remove 100 mm (4 in) of the outer jacket.
routing. Maintain a service loop in the Do not cut inner conductors or braided
cable before trimming. shield.
100
1 3 (4.0)
(1) Conduit
3 4
Pull the braided shield apart. Twist the braided shield and connect an
insulated 18 AWG to 22 AWG drain wire
as shown.
(1) Wrap
18
CDG/125389/000/06
Section 2 - Installation and Removal
5 6
Solder drain wire to twisted, braided Add shrink tubing to cable end.
shield.
7 8
Add terminal lugs if required. Connect "A", white lead, to terminal "A"
on monitor.
1
Connect "B", black lead, to terminal "B" on
monitor.
2 3
19
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
3. Maintenance
This section shows how to check the performance of the Velomitor CT transducer.
The transducer system does not require verification at regular intervals. You
should, however, verify operation by using the following test setup and
performance procedures if any of the following conditions occur:
• The performance of the system changes or becomes erratic
• You suspect that the transducer is not calibrated properly
Table 3-1 lists the recommended maintenance equipment. If the equipment is
not available, contact the nearest Bently Nevada, Inc. field office, or return the
transducer to the factory or a testing laboratory for verification.
Table 3-1: Recommended Maintenance Equipment
Recommended Equipment Specification
MB Dynamics
Model PM50 Exciter Shake Table
Bently Nevada
Model 330180-50-00 Proximitor* Sensor
Model 330101-00-08-10-02-00 Probe
Model 330130-040-00-00 Extension Cable
Power Supply, Bently Nevada TK15 -24.0 Vdc with minimum output current of 20
mA and less than 5 mV pp noise
AISI 4140 Steel Target Material 0.762 mm (0.030 in) thick, 21.6 mm (0.85 in)
diameter, 0.41 m rms (16 µin rms) roughness
20
CDG/125389/000/06
Section 3 - Maintenance
1 2 3
6
5 7
13
11
10 14
12
15
(1) Oscilloscope
(2) To Pin A, white wire
(3) To Pin B, black wire
(4) 3mA current diode, BN P/N 00643485
(5) Probe
(6) Target
(7) Cable
(8) Velomitor CT
(9) Armature
(10) Proximitor Sensor
(11) TK15
(12) Extension cable
(13) Shake table
(14) Power amplifier
(15) Signal generator
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
Step 2 — Mount the 4140 steel target to the shake table so that it is rigidly
attached to the moving armature.
Note
The probe and target assembly need to be rigid for the testing to be
accurate. Any mechanical "buzzing" sounds caused by a loose test setup
will affect results.
Step 3 — Mount a 3300 XL/ 8 mm, 1-metre probe (part number 330101) so that it
is isolated from the motion of the shake table.
Step 4 — Connect the probe to a 4-metre extension cable (part number 330130-
040-00-00) and Proximitor Sensor (part number 330180-50-00).
Step 5 — Apply -24 Vdc power to the Proximitor Sensor and monitor the output
with a voltmeter or oscilloscope.
Step 6 — Mount the transducer on the armature of the shake table using an
appropriate 3/8-24 UNF stud and torque to 5.5 N m (40 in lbs).
Step 7 — Support the interconnect cable of the transducer to prevent the
armature from rubbing against the stationary part of the shake table. If the
armature rubs against the stationary part of the shake table, the output signal
from the Velomitor CT transducer and probe will be distorted.
Step 8 — Adjust the probe-to-target gap so that the Proximitor Sensor output is at
mid range, -10.0 ± 0.5 Vdc.
Step 9 — Apply a constant current of 3 mA to the Velomitor CT transducer and
monitor the output with a voltmeter or oscilloscope.
22
CDG/125389/000/06
Section 3 - Maintenance
3.3 Polarity Test Procedure
Use this test to verify the proper phase response. Improper phase will adversely
affect the use of the transducer for machinery diagnostics.
23
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
24
CDG/125389/000/06
Section 4 - Field Testing and Troubleshooting
4.2 Fault Indication #2 Cause/Solution
Unusually low vibration with non-machine-related low-level, broadband noise.
Transducer signal is not isolated, is shorted to the case, or has noise coupled
to the signal.
Measure the resistance between the "A" terminal of the transducer and the
transducer case and the resistance between the "B" terminal of the transducer
and the transducer case. If either resistance measurement is not 1 GΩ or greater,
replace the transducer. Dispose of the transducer according to federal, state, and
local requirements. If the transducer is still covered by its Service Plan, return it to
your local Bently Nevada, Inc. office.
25
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
5. Specifications
5.1 Velomitor CT
Parameters are specified at 25 °C (77 °F) unless otherwise indicated.
Note
Operation outside the specified limits will result in false readings or loss of
machine monitoring.
5.1.1 Electrical
Sensitivity
3.94 mV/(mm/s) (100 mV/(in/s)) ± 5% at 100 Hz
Frequency response
±1.0 dB, 3.0 Hz to 900 Hz
±3.0 dB, 1.5 Hz to 1.0 kHz
Velocity range
63.5 mm/s pk (2.5 in/s pk) (see Figure 5-2 and Figure 5-3)
Transverse sensitivity
Less than 5% of the axial sensitivity
Amplitude linearity
± 2% to 63.5 mm/s pk (2.5 in/s pk)
Mounted resonant
frequency
9 kHz minimum (stud mounted, except quick disconnect)
Power requirement
(Refer to Figure 3-1)
DC voltage
22 to 30 Vdc
Bias current
2.5 to 5 mA
Output bias voltage
10.1±1.0 Vdc, Pin A referenced to Pin B
Dynamic output
impedance
<400 Ω, typical
Noise Floor
Broadband (1.5 Hz to
1 kHz)
0.109 mm/s rms (4300 μin/s rms)
Spectral
1 Hz 0.062 mm/s/√Hz (2440 μin/s/√Hz)
2 Hz 0.0203 mm/s/√Hz (800 μin/s/√Hz)
5 Hz 0.0051 mm/s/√Hz (200 μin/s/√Hz)
10 Hz 0.0020 mm/s/√Hz (80 μin/s/√Hz)
20 Hz 0.00061 mm/s/√Hz (24 μin/s/√Hz)
26
CDG/125389/000/06
Section 5 - Specifications
Grounding
Internal electronics are isolated from case
Note
Operation outside the specified limits will result in false readings or loss of
machine monitoring.
5.1.2 Environmental
Temperature range
Operating
-40 °C to 85 °C (-40 °F to 185 °F)
Storage
-40 °C to 100 °C (-40 °F to 212 °F)
Shock limit
5000 g pk
Humidity limit
100% condensing, non-submerged
5.1.3 Mechanical
Dimensions
See Figure 5-1
Mounting surface
33 mm (1.3 in) diameter
Mounting torque
4.5 ± 0.6 N-m (40 ± 5 in-lbf)
Case material
316L Stainless Steel
Connector
316L Stainless Steel MIL-C-5015
Polarity
Pin A goes positive with respect to pin B when velocity is
from base to top of the transducer
Weight (no cable)
366 gm (12.9 oz), typical
Mounting angle:
Any Orientation
27
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
2 3
6 7
28
CDG/125389/000/06
Section 5 - Specifications
29
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
30
CDG/125389/000/06
Section 5 - Specifications
105
Scale Factor (mV/(in/s) pk)
100
95
90
85
80
75
1 10 100 1000 10000
Frequency(Hz)
4.1
3.9
3.7
3.5
3.3
3.1
2.9
1 10 100 1000 10000
Frequency(Hz)
31
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
80
70
60
Phase Lag (degrees)
50
40
30
20
10
0
-10
1 10 100 1000
Frequency(Hz)
Figure 5-6: Typical Phase Lag between Output and Case Velocity
0.040
Amplitude (in/s pk)
0.030
0.020
0.010
0.000
0 2 4 6 8 10
Frequency (Hz)
32
CDG/125389/000/06
Section 6 - Accessories
6. Accessories
6.1 Mounting Hardware
6.1.1 Adhesive Mount Base Kits
These kits are designed for machines with thin casings, which do not permit
drilling and tapping a mounting hole.
04284020-AXX
A: Mounting Stud Option
01 Contains material (adhesive and bases) for 2 each
3/8-24 UNF adhesive-mount base. One kit per 2
Velomitor CT's.
02 Contains materials (adhesive and bases) for 2
each M6x1 adhesive-mount base. One kit per 2
Velomitor CT's. (Requires part number 80755-01
3/8-24 to M6x1 adapter)
33
CDG/125389/000/06
Velomitor CT Piezo-Velocity Transducer Operation Manual
A B C D E
A B C D E F
3
A B
A B C
34
CDG/125389/000/06
Section 6 - Accessories
6.1.3 Fittings
Conduit fittings let you connect flexible, metal, liquid-tight conduit or armor to the
Velomitor CT transducer.
03839201 1/2" NPT straight, male conduit fitting for
connecting flexible, liquid-tight conduit to the
transducer or a weatherproof enclosure
106924-AXX
A: Armor Length Option in Feet
Order in increments of 10 feet (3.0 metres)
Minimum Length: 10 feet (3.0 metres)
Maximum Length: 60 feet (18.2 metres)
14847-AXX
A: Flexible Conduit Length Option in Feet
Order in increments of 1.0 feet (0.3 metres)
Minimum Length: 01 foot (0.3 metres)
Maximum Length: 99 feet (30.2 metres)
35
CDG/125389/000/06