FS 205 Introduction
FS 205 Introduction
Presenter: Time:
01 Instrument overview
CONTENTS
02 Operation introduction
01 Instrument overview
Detection System of Finecare
Finecare FIA Meter III Plus (FS-205) Finecare FIA Meter Plus SE (FS-114)
Appearance
Channels 1 1 20
Items 32 53 53
Temperature Temperature can be Temperature can be controlled
without
Control controlled around 25 ℃ around 25 ℃
Heat Dissipation without fan heatsink Fan cooling
Displayer small touch screen 8-inch color touchscreen 10-inch color touchscreen
Printer thermal printer thermal printer thermal printer & external printers
Compare
Touch Screen
Thermal Printer
Configuration
Vent
Instrument Port Area
Test Name
Fluorescence
Immunoassay
Principle of Finecare System
Photoelectric Converter
Antigen
Fluorescent Labeled
Antibody
Capture Antibody
Emergency
Room Hospital lab Clinics
02 Operation Introduction
Test Procedure
1 Preparation
2 Software operation
1. Put the instrument on a level, dry surface, away from direct sunlight. Connect the instrument with
provided power adapter, switch on the instrument.
2. The instrument will boot up, load on the system and enter the Login interface.
3. After 1 minute self-check, input Username (default: Adm) and Password (default: 123456) to login.
Test Procedure
➢ Software Operation
① ② ③ ④ ⑤ ⑥
A: Functions
C: Result Display
F: Carousel Info.
D: Instrument Info.
Test Procedure
➢ Software Operation
4. Click“Reagent”button.
5. Insert the ID chip into the instrument.
6. Click“Read ID”button to memorize the information in the ID chip.
Test Procedure
➢ Software Operation
7. Click“Test”, a“Test”window will pop out. Edit the basic information and choose the sample
type and sample size.
8. Click“Add”button to add the Test request to“Wait Measuring List Area”.
9. Click“Edit”to enter the patient details.
Test Procedure
➢ Software Operation
10. Click“Start”to start the test, the access will automatically open.
11. The information“Please insert test device”will shown in“Instrument Info. Area”
Test Procedure
➢ Sample Adding
12. Open the package and take out the test device, draw the accurate sample volume which is in
accordance with the instructions of test kits to the buffer to mix thoroughly.
13. Add indicated volume of sample-buffer mixture into the sample well of Test Device, immediately
insert the test device into the holder.
Test Procedure
➢ Testing
14. Once the test device is inserted, the test device access closes. The carousel rotates to transfer
the test device into the test area. The bar code on the test device is then scanned. A time
countdown will appear on the screen.
15. When the test is finished, the results will show on the screen.
16. The test results will be automatic kept in the system,and can be query in “Test record” page.
Test Procedure
➢ Testing
17. Click “Print”to print the result. Also can export the result to the LIS system.
Test Procedure
➢ Stop Testing
17. After test, the instrument will automatically discard the test card, but if you want to manually kick
out the reagent card, here are two ways.
1) When the reagent card is shown in Carousel Info. area and start to countdown, click the regular
circle, then click“Discard card”button, the instrument will kick out that reagent card;
2) If want to kick out all the reagent card in the instrument, click“Test”, then click“Stop”button,
the instrument will immediately stop testing and kick out all the reagent cards in the instrument;
Test Procedure
Test Procedure
➢ Notices — Sample
➢ Notices — Operation
➢ Notices — Instrument
The reagent card should be insert into position, after insert, hand
should be taken out immediately from the instrument.
When the access is running, please don’t touch the access or insert
the reagent.