631D-1039281-001-MAN (1)
631D-1039281-001-MAN (1)
631D-1039281-001-MAN (1)
Protected by U.S. Patent Numbers 4785669 and 5625152. Foreign patents pending.
Some Baratron® products may not be exported to many end user countries without both US and
local government export licenses under ECCN 2B230.
2
Table of Contents
Table of Contents
iii
Table of Contents
Sensor ................................................................................................................26
Piping Considerations and Port Connections ..................................................................27
General ...............................................................................................................27
Sensor Fittings ...................................................................................................27
iv
Table of Contents
Index ............................................................................................................................................ 55
v
Table of Contents
Tables
Table 1: Definition of Symbols Found on the Unit......................................................................... 2
Tabelle 2: Bedeutung der am Gerät angebrachten Symbole ........................................................... 6
Tableau 3: Définition des symboles apparaissant sur l'unité ........................................................ 10
Tabla 4: Definición de los símbolos hallados en la unidad ............................................................ 14
Table 5: Interface Cables .............................................................................................................. 22
Table 6: Pinout of the Interface Connector (J1) ............................................................................. 34
Table 7: Highest Pressures Suggested for Proper Zero Adjustment .............................................. 38
Table 8: Over-Temperature Cutoff Value ...................................................................................... 41
Table 9: Conditions and resulting status LEDs and heater alarm relay output. ............................. 41
vi
Safety Information
Safety Information
Definitions of WARNING, CAUTION, and NOTE messages used throughout the manual.
1
Safety Information
|
Protectivee arth
On(
S u
pply
) Off (Supply) E
ar
th(g
ro
und
) (gro und)
I
E C
417
,No.
5 00
7 IEC 417, No.5008 I
E C4
17,No
.5
017 IEC4 17 ,N o.5019
Fra
meorc
hass
is Eq
uip
ote
nt
ialit
y D
ir
ectc
ur
ren
t Alternating current
I
E C4
17
,No.
5020 I
E C
417,No
.50 21 I
E C
417,No
.50
31 IEC 417, No.5032
Caution, refer to
accompanying Caution, risk of
documents electric shock Caution, hot surface
ISO 3864, No.B.3.1 ISO 3864, No.B.3.6 IEC 417, No.5041
2
Safety Information
3
Safety Information
4
Sicherheitshinweise für den Druckmeßumformer
Warnung! Das Symbol WARNUNG! weist auf eine Gefahr für das
Bedienpersonal hin. Es macht auf einen Arbeitsablauf, eine
Arbeitsweise, einen Zustand oder eine sonstige Gegebenheit
aufmerksam, deren unsachgemäße Ausführung bzw.
ungenügende Berücksichtigung zu Verletzungen führen kann.
Vorsicht! Das Symbol VORSICHT! weist auf eine Gefahr für das Gerät hin. Es
macht auf einen Bedienungsablauf, eine Arbeitsweise oder eine
sonstige Gegebenheit aufmerksam, deren unsachgemäße Ausführung
bzw. ungenügende Berücksichtigung zu einer Beschädigung oder
Zerstörung des Gerätes oder von Teilen des Gerätes führen kann.
Hinweis Das Symbol HINWEIS macht auf wichtige Informationen bezüglich eines
Arbeitsablaufs, einer Arbeitsweise, eines Zustands oder einer sonstige
Gegebenheit aufmerksam.
5
Sicherheitshinweise für den Druckmeßumformer
|
Ein (Energie) Aus (Energie) Erdanschluß Schutzleiteranschluß
IEC 417, No.5007 IEC 417, No.5008 IEC 417, No.5017 IEC 417, No.5019
Aquipotential-
Masseanschluß anschluß Gleichstrom Wechselstrom
IEC 417, No.5020 IEC 417, No.5021 IEC 417, No.5031 IEC 417, No.5032
Durchgängige Dreileiter-
Gleich- oder doppelte oder Wechselstrom
Wechselstrom verstärkte Isolierung (Drehstrom)
IEC 417, No.5033-a IEC 417, No.5172-a IEC 617-2, No.020206
6
Sicherheitshinweise für den Druckmeßumformer
Gerät nicht zusammen mit explosiven Stoffen, Gasen oder Dämpfen benutzen!
Um der Gefahr einer Explosion vorzubeugen, darf dieses Gerät niemals zusammen mit (oder in
der Nähe von) explosiven Stoffen aller Art eingesetzt werden, sofern es nicht ausdrücklich für
diesen Zweck zugelassen ist.
7
Sicherheitshinweise für den Druckmeßumformer
8
Informations relatives à la sécurité pour le transducteur de pression
9
Informations relatives à la sécurité pour le transducteur de pression
|
Marche Terre de protection
(sous tension) Arrêt (hors tension) Terre (masse) (masse)
IEC 417, No.5007 IEC 417, No.5008 IEC 417, No.5017 IEC 417, No.5019
10
Informations relatives à la sécurité pour le transducteur de pression
PURGE DE L'APPAREIL
Après l'installation de l'unité, ou avant son enlèvement d'un système, purger l'unité complètement
avec un gaz propre et sec afin d'éliminer toute trace du produit de flux utilisé précédemment.
11
Informations relatives à la sécurité pour le transducteur de pression
12
Medidas de seguridad del transductor de presión
13
Medidas de seguridad del transductor de presión
|
Encendido Apagado
(alimentación eléctrica) (alimentación eléctrica) Puesta a tierra Protección a tierra
IEC 417, N° 5007 IEC 417, N° 5008 IEC 417, N° 5017 IEC 417, N° 5019
Precaución. Consulte
los documentos Precaución. Riesgo Precaución.
adjuntos de descarga eléctrica Superficie
ISO 3864, N° B.3.1 ISO 3864, N° B.3.6 caliente
IEC 417, N° 5041
14
Medidas de seguridad del transductor de presión
PURGUE EL INSTRUMENTO
Una vez instalada la unidad o antes de retirarla del sistema, purgue completamente la unidad con
gas limpio y seco para eliminar todo resto de la sustancia líquida empleada anteriormente.
15
Medidas de seguridad del transductor de presión
16
Chapter One: General Information
Introduction
Note Some Baratron® products may not be exported to many end user countries
without both US and local government export licenses under ECCN
2B230.
The MKS Baratron® Type 631D Absolute Pressure Transducer1 is part of the MKS family of
RoHS (Restriction of Hazardous Substances)-compliant, general purpose pressure transducers
designed to provide accurate, reliable and repeatable pressure measurements from 1000 Torr
down to 1 Torr, Full Scale (F.S.). The 631D is temperature-controlled to prevent contamination
build-up and is available in either of two operating temperatures: 150 C and 200 C. These
instruments have a small footprint and self-contained signal conditioning electronics mounted
directly above the sensor. The precision pressure transducer operates on a variable capacitance
technique2 and requires 15 VDC input to provide a 0 to 10 VDC output linear with pressure over
the sensor’s measurement range.
The 631D transducer is available with an optional trip point relay system. The relay system
provides two (2) internally-mounted solid-state trip point relays with user adjustable pressure
conditions that will turn ON or OFF each solid state relay. The trip point relay system includes a
status LED, trip point value adjustment, direction selection and signal access for each of the two
trip points.
The 631D pressure transducer is comprised of a sensor and signal conditioner. An MKS or
compatible power supply is required to complete the pressure-to-electronic output conversion. A
power supply/readout can also be used to provide a direct pressure display; the necessary Y-type
cables are available from MKS.
The 631D transducer measures absolute pressure, so only the measurement port has a connector.
During production, the reference port (PR) is pumped down to a pressure less than 10-7 Torr, and
outgassed thoroughly. A chemical getter is added to maintain the low pressure, and the port is
permanently sealed.
Temperature control of the transducer minimizes the effects of ambient or process temperature
variations typically encountered in process line environments. The sensor temperature is
continually monitored; two status LEDs provide visual indication of the unit’s temperature
control status. The LED labeled “Htr Fail Status” is normally off. If the sensor heater fails, the
LED becomes red and the associated relay becomes de-energized. The LED labeled “At Temp
Status” is yellow while the unit heats up and green when the unit is at the correct temperature.
(The unit requires a warm up time of two hours before any pressure data should be taken.)
Using the latest single sided, dual-electrode Inconel® sensor design, coupled with low impedance,
fixed-frequency bridge signal conditioner, these instruments are capable of withstanding high
17
Chapter One: General Information
overpressure conditions (45 psia). The high overpressure limit, coupled with the high temperature
operation, ensures maximum zero stability. The 631D transducer diaphragm design results in
extremely low hysteresis and the ability to specify a 0.50% accuracy as a percentage of Reading
instead of as a percentage of Full Scale. Since the accuracy is based on a percentage of Reading,
the transducer can be used over a wide dynamic measuring range without the need for re-ranging,
thus it provides accurate process monitoring over a variety of operating conditions.
This product can be used for pressure and vacuum measurements with a wide variety of process
gases, including (but not limited to) toxic, pyrophoric, and oxidizing materials such as silane
(SiH4), dichlorosilane (Cl2H2Si), trichlorosilane (Cl3HSi), anhydrous ammonia (NH3), ozone (O3),
and de-ionized water vapor (DIW). However, it is not recommended for use with corrosive or
explosive gases, so please contact the MKS Applications group or your local MKS Account
Manager if you plan to use the product with such gases.
This product contains electronic components made with non-toxic metals, and the packaging
materials are recyclable plastics. However, at the end of life of the product, decommissioning
must be performed in accordance with your local environmental regulations in consideration of
the chemicals and gases that this product has been exposed to.
Chapter One, General Information, (this chapter) introduces the product and describes the
organization of the manual.
Chapter Two, Installation, explains the environmental requirements and describes how to mount
the instrument in your system.
Chapter Three, Overview, gives a brief description of the instrument and its functionality.
Chapter Four, Operation, describes how to use the 631D unit and explains all the functions and
features.
Chapter Five, Optional Trip Point Relay System, discusses the optional relay system and its
parameters, including measuring and adjusting the trip point values and specifying the trip point
direction.
Chapter Six, Maintenance and Troubleshooting, provides a few maintenance recommendations
and a brief troubleshooting section. The Type 631D transducer is highly reliable and requires
little maintenance.
Appendix A, Product Specifications, lists the specifications of the instrument.
Appendix B, Model Code Explanation, describes the model code.
18
Chapter One: General Information
Customer Support
Standard maintenance and repair services are available at all of our regional MKS Calibration and
Service Centers, listed on the back cover. In addition, MKS accepts the instruments of other
manufacturers for recalibration using the Primary and Transfer Standard calibration equipment
located at all of our regional service centers. Should any difficulties arise in the use of your Type
631D instrument, or to obtain information about companion products MKS offers, contact any
authorized MKS Calibration and Service Center. If it is necessary to return the instrument to
MKS, please obtain an RMA Number (Return Material Authorization Number) from the MKS
Calibration and Service Center before shipping. The RMA Number expedites handling and
ensures proper servicing of your instrument.
Please refer to the inside of the back cover of this manual for a list of MKS Calibration and
Service Centers.
19
Chapter One: General Information
20
Chapter Two: Installation
Note Do not discard any packing materials until you have completed your
inspection and are sure the unit arrived safely.
If you find any damage, notify your carrier and MKS immediately. If it is necessary to return the
unit to MKS, obtain an RMA Number (Return Material Authorization Number) from the MKS
Service Center before shipping. Please refer to the inside of the back cover of this manual for a
list of MKS Calibration and Service Centers.
Caution Only qualified individuals should perform the installation and any
user adjustments. They must comply with all the necessary ESD and
handling precautions while installing and adjusting the instrument.
Proper handling is essential when working with all highly sensitive
precision electronic instruments.
Unpacking Checklist
Standard Equipment:
• Type 631D Unit
• Type 631D Instruction Manual (this book)
21
Chapter Two: Installation
Optional Equipment:
• Interface cables, refer to Interface Cables, page 22, for details
• Electrical Connector Accessories Kit: 631D-K1 (contains a mating connector)
• MKS Type PR4000B Electronic Display Unit
• All MKS automatic pressure controllers (Type 651, 1651, and others)
Note The Type 651 pressure controller must include the high power supply
option (2.0 Amperes) to work with a 631D transducer. A Type PR4000B
must also be equipped with the 1.5 Ampere output to be used with this
Baratron.
Interface Cables
As of July 20, 2009, all products shipped to the European Community must comply with the EMC
Directive 2004/108/EC, which covers radio frequency emissions and immunity tests. MKS
products that meet these requirements are identified by application of the CE Mark.
MKS offers the following interface cables to connect the 631D unit to your system.
Interface Cables
Cable Number Number of Power Supplies
CB631-1- xx 1
CB631S-1- xx 1
xx designates the length of the cable, in feet
22
Chapter Two: Installation
• Mount the Type 631D transducer components correctly: refer to Figure 1: Recommended
Orientation of the Type 631D Transducer, page 26, for the recommended orientation of the
sensor
• Maintain a solid system ground for proper operation and safety to personnel
23
Chapter Two: Installation
Environmental Requirements
Caution DO NOT use a thermal blanket on the 631D unit. The 631D
transducer is designed to maintain its specified operating
temperature without external assistance.
24
Chapter Two: Installation
Setup
General
To obtain maximum accuracy from your Type 631D transducer, be sure the interconnecting
piping and the transducer components themselves are properly installed. Some general guidelines
are:
• Mount the transducer such that vibration, shock, and temperature fluctuations are
minimized, and that the product’s inlet always points downward or horizontally.
Environmental Considerations
The 631D high temperature transducer is designed to operate at its specified operating
temperature. To meet this high operating temperature, you must follow these guidelines:
25
Chapter Two: Installation
Sensor
The 631D transducer can be installed in any orientation as long as the inlet tube is not facing
upwards. However, for best results, it is recommended that the Px port of the unit face downward
(as shown in Figure 1) to reduce the chance of contaminants entering the device and degrading its
performance. The sensor is zeroed at the factory in the vertical orientation (shown in Figure 1).
You should zero the unit again after installation, following the steps in How To Zero the Type
631D Transducer, page 37.
If you anticipate vibration, orient the plane of the diaphragm parallel with the major oscillating
axis. The edge of the diaphragm is indicated by the dashed line in Figure 1.
Note If vibration causes the unit to be oriented such that the Px port is not
facing downward, you must decide the best orientation for optimum
performance.
Edge of the
Sensor
Diaphragm
Px Port
26
Chapter Two: Installation
General
• Position a line tap above or to the side of the process line, and position the sensor above
the line tap
This allows any liquids which may be present to drain into the process line.
• Install the interconnecting piping so that it slopes away from the sensor to the process line
A slope of 1 inch per foot (8 centimeters per meter) is adequate.
• Purge the interconnecting piping (close to the process tap) when needed to prevent
sediment buildup
When purging, be sure to isolate the sensor from the interconnecting piping.
Sensor Fittings
The following additional fittings are available for the Type 631D transducer:
2. Insert the tubing through the compression nut and O-ring all the way to the shoulder.
27
Chapter Two: Installation
28
Chapter Three: Overview
Sensor
The Type 631D transducer contains an absolute sensor that is referenced to high vacuum (less
than 10-7 Torr). This pressure sensor utilizes a single-sided, dual-electrode/AC bridge circuit
design. In this design, two capacitance electrodes are deposited upon a ceramic disk in a
concentric “bull’s-eye” arrangement. The disk is positioned close to an Inconel® tensioned
diaphragm to form two capacitors in an AC bridge circuit. The other side of the diaphragm is
exposed to the process gas.
The diaphragm deflects with changing pressure—force per unit area—independent of the
composition of the measured gas. This deflection causes a capacitance change between the
diaphragm and the adjacent electrode assembly. When pressure is equal on both sides of the
diaphragm, the bridge is balanced. As pressure deflects the diaphragm toward the electrode, the
center capacitance changes more than the outer capacitance, causing the bridge to become
unbalanced and an AC voltage to be generated.
Signal Conditioner
The signal conditioner contains the state-of-the-art, low impedance balanced bridge circuitry,
self-compensated for thermal stability with ambient temperature changes. The output is linear
with pressure and provides a 0 VDC output at Zero pressure and a 10 VDC output at sensor Full
Scale.
29
Chapter Three: Overview
Zero Pot
Coarse Zero
Switch
Htr. Fail
Status LED
At Temp
Status LED Interface Connector
Figure 2: Top View of the Type 631D Transducer Without Optional Relay System
The transducer with the optional trip point relay system has the same functionality as the
transducer without the relay system described above. The relay system brings an additional two
LEDs, two adjustment POTs and three test plugs to the top of the transducer. (Refer to Figure 3:
Top View of the Type 631D Transducer With Optional Relay System, below, showing the
30
Chapter Three: Overview
location of the trip point relay system components.) The trip point status LEDs glow green when
the corresponding solid-state relay is turned ON and are off when the corresponding solid-state
relay is turned OFF, depending on the relationship between the pressure output and the trip point
value. (Refer to Chapter Five: Optional Solid-State Relay System, page 43 for a complete
description of the relay system.)
Relay system
described in
Zero Pot Chapter Five
Coarse
Zero
Switch
Htr. Fail
Status LED
At Temp
Interface Connector
Status LED
Figure 3: Top View of the Type 631D Transducer With Optional Relay System
31
Chapter Three: Overview
Coarse Zero
Switch
Note All dimensions are nominal and are listed in inches with millimeters
referenced in parentheses.
32
Chapter Three: Overview
Interface Connector
The Interface connector (J1) is a 15-pin male type D-subminiature connector. Refer to Figure 2:
Top View of the Type 631D Transducer, page 30, for the location of the Interface connector for
the 631D without the optional relay system. Refer to Figure 3: Top View of the Type 631D
Transducer With Optional Relay System, page 31, for the location of the Interface connector for
the 631D without the optional relay system. It contains the pins to connect the 631D unit to a
power supply or power supply/readout. Table 6 specifies the pinout of the Interface connector.
Note The state of the relay contacts at pins 13, 14 are for an energized heater
alarm relay (normal operation) – refer to Heater Status LEDs and Heater
Alarm Relay, page 40 for a description of relay functionality. To use a
631D transducer with any non-MKS power supplies or power
supply/readouts, consult the manufacturers’ specifications for electrical
and power considerations, and Appendix A: Product Specifications, page
49, for electrical requirements of the 631D transducer.
33
Chapter Three: Overview
Pressure Output
The 631D transducer provides a 0 to 10 VDC output signal at pressures of 100% of its full-scale
range and below.
Overpressure Considerations
As with any transducer, the 631D transducer performs valid measurements up to the unit’s Full
Scale. At pressures of roughly 10% over the Full Scale range, a high (≥ 11 VDC) output signal is
produced to warn of invalid readings. This high output signal varies from device to device, and is
not linear with further changes in pressure. At pressures somewhat beyond 110% of the Full
Scale range, the diaphragm touches the fixed capacitor plate. The capacitor plate serves as a
protective overpressure stop to prevent damage to the diaphragm up to 45 psia for an absolute
transducer.
Note The output signal of the 631D above its full-scale measurement range is
not linear or repeatable, and should not be used for quantitative
measurements or process control. MKS does not guarantee that the
product will meet any of its published performance specifications above
110% of its full-scale range.
Labels
Three labels appear on the 631D unit: a wiring diagram label, a “Danger Hot” label, and a serial
number label.
The wiring diagram label shows how to connect a single or dual power supplies to the 631D unit.
Refer to Figure 5: Wiring Diagram Label, page 34, for the location of the label on the 631D unit.
34
Chapter Three: Overview
35
Chapter Three: Overview
36
Chapter Four: Operation
Note If available pressures are not sufficiently low to set the transducer zero,
you may use a vacuum leak detector with sufficient vacuum pumping (to
achieve proper zeroing pressures). In this case, mount the transducer on
the leak detector in the same plane of orientation as it will be during
actual use.
37
Chapter Four: Operation
Note Allow 2 hours for units to warm up. Ensure that the transducer is fully
stabilized before you adjust the zero.
4. Adjust the ZERO potentiometer with a small screwdriver until the readout displays zero
(0.000). Typically, the ZERO potentiometer provides ample control under normal conditions.
However, if the ZERO potentiometer fails to provide sufficient adjustment, additional zero
range capability is available with the COARSE ZERO switch, located on the side of the unit
(refer to Figure 2: Top View of the Type 631D Transducer, page 30, or Figure 3: Top View
of the Type 631D Transducer With Optional Relay System, page 31, for location of the
ZERO potentiometer and coarse zero switch.) Refer to How To Adjust the Coarse Zero
Switch, page 39, for more information.
38
Chapter Four: Operation
Note Use the COARSE ZERO switch only if the ZERO potentiometer fails to
provide sufficient adjustment.
Note Allow 2 hours for units to warm up. Ensure that the transducer is fully
stabilized and the “At Temp Status” LED is glowing green, before you
adjust the zero.
4. Center the ZERO potentiometer located at the top of the transducer by adjusting the screw
to leave an equal amount of adjustment both clockwise and counterclockwise.
5. Remove the plug that covers the COARSE ZERO switch. Refer Figure 4: Dimensions of the
Type 631D Transducer, page 32.
6. Turn the COARSE ZERO switch to a position that produces the output signal closest to
0.000 Volts.
8. Cover the COARSE ZERO switch with the plug removed in step 5.
39
Chapter Four: Operation
Over-Temperature Protection
The heater implements two temperature sensors, one dedicated to the heater control circuit and
one dedicated to the over-temperature protection. The over-temperature protection limit is used
as one layer of protection by the heater alarm circuit, refer to Table 8: Over-Temperature Cutoff
Value for the over-temperature value for each of the operating temperatures. If the transducer
temperature reaches the over temperature cutoff, the “Htr Fail Status” LED will glow red, the
heater alarm relay becomes de-energized and the “At Temp Status” LED will glow yellow. Note
that once the temperature drops below the overtemperature cutoff, the heater power will be
activated, the heater alarm relay energized, the “At Temp Status” LED will continue to glow
yellow and the “Htr Fail Status” LED will turn off.
40
Chapter Four: Operation
150 185
200 230
Table 9: Conditions and resulting status LEDs and heater alarm relay output.
41
Chapter Four: Operation
42
Chapter Five: Optional Solid-State Relay System
General Information
The 631D transducer provides the option for two (2) internally mounted trip point relays with
user adjustable pressure conditions that will turn ON/OFF each solid state relay. The trip point
direction – whether the relay turns ON above or below the selected pressure condition – is chosen
using the appropriate model code when purchasing the 631D transducer (refer to Appendix B:
Model Code Explanation, page 51). A visual indication of the status of each trip point relay is
provided by a dedicated LED that illuminates when the trip point relay is turned ON. The relay
gets turned OFF and the LED turns off when the pressure returns to the selected trip point value
plus hysteresis (refer to Trip Point Direction, page 44 for a description of how hysteresis affects
the relay switching).
The interface connector houses the contact pins of both trip point relays to equip the user with
remote indication of the relay status. The three pins of each relay produce an open/closed output;
for example, if Trip Point A is turned ON, pins 3 & 4 will be an open circuit. Refer to Table 6:
Pinout of the Interface Connector (J1), page 34, for the pinout of the Interface (J1) Connector.
43
Chapter Five: Optional Solid-State Relay System
Access to
Trip Point
Voltage
Signals
Figure 8: Location of the Trip Point Adjustment Pots and Trip Point Signal Output
44
Chapter Five: Optional Solid-State Relay System
LED OFF
Relay Turned OFF
TRIP POINT
VALUE
0.030 Volts
TRIP POINT
HYSTERESIS
Figure 9: LED and Relay State When Trip Point Direction Set Above Trip Point Value
If the preset direction for Trip Point B is “below”; the trip point is turned ON as the pressure falls
below the specified trip point level. The trip point is not turned OFF until the pressure rises
above the value defined by the hysteresis, as shown in Figure 10. Figure 10 also depicts the relay
being turned ON when the initial pressure condition applied to the unit is below the trip point
value.
TRIP POINT
HYSTERESIS
0.030 Volts
TRIP POINT
VALUE
LED ON
Relay turned ON
LED OFF
IF Initial Pressure < Trip Point Value Relay turned OFF
Figure 10: LED and Relay State When Trip Point Direction Set Below Trip Point Value
Note that regardless of trip point direction the time response of the 631D transducer will have a
minor affect of the time between (1) a change in the pressure condition necessary to cause a
change in the relay state and (2) the relay actually changing state. The time response is specified
in Appendix A: Product Specifications, page 49.
45
Chapter Five: Optional Solid-State Relay System
46
Chapter Six: Maintenance and Troubleshooting
General Maintenance
For best accuracy and repeatability, MKS recommends that you:
47
Chapter Six: Maintenance and Troubleshooting
Troubleshooting
Caution Damage may occur to the sensor if the sediment is removed with a
harsh solvent, or by scraping. Therefore:
• Do not scrape the inside of the process connection(s).
• Only clean the process connection(s) with solvent appropriate
for your process.
• Ensure the voltage to the transducer is ±15 VDC (±5%) @ 1.0 Amps
• Verify that the total current being drawn by the manometer is within the specifications of
the power supply
If the solutions presented in this chapter fail to resolve the problem, return the transducer to MKS
for servicing. When returning the instrument to MKS, please obtain an RMA Number (Return
Material Authorization Number) from the MKS Calibration and Service Center before shipping.
The RMA Number expedites handling and ensures proper servicing of your instrument.
Refer to the inside of the back cover of this instruction manual for a complete list of MKS
Calibration and Service centers.
48
Appendix A: Product Specifications
Physical Specifications
Dimensions3 3.15” diameter, 4.55” length, maximum
(8.00 cm diameter, 11.55 cm length)
Performance Specifications
Accuracy4
150 C; 200 C ± 0.5% of Reading
49
Appendix A: Product Specifications
Environmental Specifications
Ambient Operating Temperature Range
150 C Units 10° to 50° C* (50° to 122° F)
* Minimum 50 ft/min air flow is required if the ambient
temperature is between 45 to 50 C (113 to 122 F)
200 C Units 10° to 50° C** (50° to 122° F)
** Minimum 150 ft/min air flow is required if the
ambient temperature is between 40 to 50 C (104 to
122 F)
50
Appendix B: Model Code Explanation
Model Code
The options of your transducer are identified in the model code when you order the unit.
The model code is identified as follows:
631DWWXYZTUV
where:
631D WW X Y Z T UV
Type Number
Full Scale Range
Engineering Units
Fitting
Accuracy
Temperature
Trip Points
51
Appendix B: Model Code Explanation
Pascal L
Mbar M
Fittings (Y)
Your fitting choice is designated by a single letter code.
Ordering Code
½” diameter tube A
Swagelok 8-VCR, female B
NW-16-KF D
1½” Tri-Clover M
2” Tri-Clover N
Accuracy (Z)
The standard accuracy (±0.5% of Reading) is specified by a letter F in this field.
Temperature (T)
TheThe operating temperature of the transducer is specified by a single letter.
Ordering Code
150 C H
200 C P
52
Appendix B: Model Code Explanation
53
Appendix B: Model Code Explanation
54
Index
Index
A L
Accuracy, 17, 50, 52 Labels, 34, 35
Air flow, 24, 25, 50 hot, 35
Alarm, heater, 40 serial number, 35
wiring diamgram, 34
C
Cables, 17, 22, 23 M
CE compliance, 50 Maintenance, 47
Connectors, 30, 31, 33 Manual organization, 18
Customer support, 18 Model code, 51
Mounting, 25, 26
D
Dimensions, 32, 49 O
Output signal, 17, 29, 34, 49
E
Overpressure, 17, 23, 34, 49
Environmental, 24, 25, 50
Overtemperature, 40, 41
F
P
Fittings, 27, 49, 52
Pinout, connector, 33, 43
Full scale range, 17, 49, 51
Piping considerations, 27
55
Index
S
Sensor, 17, 26, 29
Sensor fittings, 27
Setup, 25, 26
Signal conditioner, 17, 29
Standard sensor fittings, 27
T
Temperature
coefficients, 50
operating, ambient, 24, 25, 50
operating, sensor, 17, 40, 49, 52
storage, 24, 50
Time response, 50
Trip point relay system, 17, 29, 43
direction, 43, 44, 53
hysteresis, 45
value, 43, 53
Troubleshooting, 48
V
Ventilation, 24, 25, 50
Vibration, 23, 26
W
Warm up, 17, 23
Z
Zero adjustment
coarse zero switch, 37, 39
zero pot, 37, 38
56