fidas_200_series
fidas_200_series
fidas_200_series
Operating Manual
For specialized users
5200-en_V2.0_06/20
V 2.0
Copyright
For all information in this technical document as well as drawings and technical descriptions provided by us,
Palas GmbH reserves all proprietary and intellectual property rights. Duplication or dissemination to third
parties without our prior written consent is prohibited.
Technical changes reserved.
Table of contents
Table of contents
1 About this manual 5
1.1 Function of the manual 5
1.2 Target group for the manual 5
1.3 Validity of the manual 5
1.4 Related documents 5
1.5 Storage of documents 5
1.6 Presentation rules 6
2 For your safety 7
2.1 Hazards and safety instructions 7
2.2 Warnings 8
2.2.1 Warnings on the device 9
2.3 Regulations 10
2.4 Obligations of the user 10
3 Product description 11
3.1 Use 11
3.1.1 Intended use 11
3.1.2 Improper use 11
3.2 Models 12
3.2.1 Components of the model Fidas 200 12
3.2.2 Components of the model Fidas 200 E 13
3.2.3 Components of the model Fidas 200 S 14
3.3 Controller 16
3.4 Functional description 19
3.5 Environmental conditions 20
3.6 Type plate 20
3.7 Conformity 20
3.8 Dimensions 22
3.9 Minimum clearance 25
3.10 Scope of delivery 26
3.11 Optional accessories 27
4 Initial commissioning 28
4.1 Overview 28
4.2 Setting up and connecting Fidas 200 28
4.3 Setting up and connecting Fidas 200 E 30
4.4 Setting up and connecting Fidas 200 S 32
4.4.1 Installing weather station 33
4.4.2 Installing sampling tube 35
4.5 Attaching sampling head 38
5 Operation and settings 39
5.1 Turning the device on 39
5.2 Fidas Firmware 40
5.2.1 Main menu 40
5.2.2 Data screen – current measured values 42
5.2.3 Dust chart screen – course of fine dust values measured 43
5.2.4 Air sensor chart screen – course of values measured by weather
station 44
5.2.5 Accessories screen – accessories and other information 45
5.2.6 Intelligent Aerosol Drying System (IADS) screen 46
5.2.7 Weather station screen – data measured by weather station 48
5.2.8 PM data screen (alternative PM values) 49
5.2.9 Filter system screen 50
5.2.10 Particle size distribution screen 51
5.2.11 Alarms screen 52
5200-en_V2.0_06/20 3
Table of contents
4 5200-en_V2.0_06/20
1 About this manual
The manual specifies all of the tasks necessary for initial commissioning of
the device. For more information, refer to the section "Initial commissioning
[} 28].”
5200-en_V2.0_06/20 5
1 About this manual
6 5200-en_V2.0_06/20
2 For your safety
5200-en_V2.0_06/20 7
2 For your safety
2.2 Warnings
The warnings in this manual are highlighted with pictographs and signal
words. The severity of a hazard is indicated by the pictograph and signal
word.
DANGER
Type and source of the hazard
Explanation of type and source of the hazard / description of conse-
quences of failure to comply with the warning.
4 Measures to avert the hazard
DANGER
Danger to life due to electric shock
NOTICE
Damage to property
8 5200-en_V2.0_06/20
2 For your safety
5200-en_V2.0_06/20 9
2 For your safety
2.3 Regulations
Observe the following regulations and directives:
Legal requirements • Legal regulations for Health and Safety at Work
• Legal regulations for environmental protection
• Employer's Liability Insurance Association provisions
Standards and directives • The applicable safety requirements according to DIN, EN, and VDE
10 5200-en_V2.0_06/20
3 Product description
3 Product description
3.1 Use
Any use other than the intended use is prohibited. The manufacturer as-
sumes no liability for damages resulting from this. All warranty claims are
voided if changes are made to the device, including in the course of assem-
bly and installation.
All corresponding documentation must be observed for any work on the de-
vice.
5200-en_V2.0_06/20 11
3 Product description
3.2 Models
These operating instructions describe three models of the measuring sys-
tem:
• Fidas 200: Controller with attached aerosol sensor
• Fidas 200 E: Controller with external aerosol sensor
• Fidas 200 S: Controller with attached aerosol sensor and with weather-
proof housing
12 5200-en_V2.0_06/20
3 Product description
5200-en_V2.0_06/20 13
3 Product description
14 5200-en_V2.0_06/20
3 Product description
5200-en_V2.0_06/20 15
3 Product description
3.3 Controller
6 7
5 4 3
16 5200-en_V2.0_06/20
3 Product description
6 7
5 4 3
1 Port for LAN cable for data 2 Port for cable for power supply
transmission to external sensor unit
3 Tube to external aerosol sen- 4 Feed line to interior aerosol
sor pumps
5 Touch screen 6 USB port
7 Filter for aerosol pumps
5200-en_V2.0_06/20 17
3 Product description
12 13 14 15 16 17
11 10 9 8 7 6
18 5200-en_V2.0_06/20
3 Product description
Sampling and drying The sampling head Sigma-2 draws in ambient air.
Volume flow rate: approx. 0.3 m3/h.
The aerosol passes through the sampling tube with drying section IADS and
into the aerosol sensor. The drying section dries the air that was sucked in,
thus preventing distortion of the measurement by any remaining moisture or
mist particles.
Weather data The following weather-related factors have an impact on and control the dry-
ing effect:
• Air temperature
• Air pressure
• Humidity
The weather station measures this data. Additional data can be measured
and taken into consideration with some models:
WS300 WS500 WS600
Scattered light analysis The central element of the device is an optical aerosol sensor. Lorenz‐Mie
scattered light analysis of single particles is applied to determine the particle
size. The single particles move through an optically differentiated measure-
ment volume that is homogeneously illuminated with a polychromatic LED
light source. Each particle generates a scattered light impulse that is de-
tected at an angle of 85° to 95° degrees. The particle count is determined
based on the number of scattered light pulses. The level of the measured
scattered light pulse is a measure of the particle diameter.
5200-en_V2.0_06/20 19
3 Product description
Data processing The compiled data is processed in the controller and saved. The device can
be integrated into a higher ranking measurement system via the interfaces
(RS-232 and Ethernet).
The measured data can also be transferred via the USB port to an external
data medium. The data can then be analyzed or processed with the external
device. The software PDAnalyze should be used to process the data with an
external device. PDAnalyze is included in delivery.
Firmware The device is operated via a touch screen. The touch screen shows the user
interface of the firmware Fidas. The firmware Fidas runs on the operating
system Windows Embedded for Point of Service (WEPOS).
1 2 3
Type plate
20 5200-en_V2.0_06/20
3 Product description
3.7 Conformity
As the manufacturer, we hereby declare that this product meets the funda-
mental directives for bringing it to market in the EU.
5200-en_V2.0_06/20 21
3 Product description
3.8 Dimensions
513 159
255
1514
Main dimensions of the model Fidas 200 – all dimensions in millimeters (in this case with the weather station WS500)
22 5200-en_V2.0_06/20
3 Product description
Y
Ø 57,0
2382
2117
Y
200
Ø 48,3
952
5200-en_V2.0_06/20 23
3 Product description
1745
1384
485
661
869
600
Main dimensions of the model Fidas 200 S – all dimensions in millimeters (in this case with the weather station WS600)
Dimensions of controller
483 324
449
177
184
24 5200-en_V2.0_06/20
3 Product description
5200-en_V2.0_06/20 25
3 Product description
26 5200-en_V2.0_06/20
3 Product description
5200-en_V2.0_06/20 27
4 Initial commissioning
4 Initial commissioning
4.1 Overview
Initial commissioning of the device includes the following tasks:
• Set up and connect the device. The process is explained over the next
few pages.
• Adjust the operating system settings to suit your location. This includes:
- Time and date settings
- Network address
- Number formats
• Calibrate and verify the measuring system; refer to section "Verifying
and calibrating measuring system [} 72]”
Setup time: The tasks listed above take about two hours to complete.
28 5200-en_V2.0_06/20
4 Initial commissioning
CAUTION
There is a risk of pinching and shearing points when placing the
sampling tube on the aerosol inlet tube.
Risk of injury to unprotected fingers.
4 Wear protective gloves.
1. Hold the sampling tube up to light to verify that there are no foreign ob-
jects in the tube.
2. Carefully guide the sampling tube over the aerosol inlet tube on the con-
1 - 2 mm troller.
Important: There should be a gap of 1 to 2 mm between the sampling
tube and the contact surface. If the tube rests directly on the contact sur-
face, it can subject the aerosol sensor to mechanical strain.
3. Secure the sampling tube with a pipe clamp (provided on site). Ensure
that the tube is positioned vertically.
4. Plug the cable from the drying section IADS (on the sampling tube) into
the receptacle “IADS” on the back of the controller.
5200-en_V2.0_06/20 29
4 Initial commissioning
1 2 3 3 2 1
Hose connections and cable connections between controller and sensor unit
1. Set up the controller and the sensor unit in a suitable place within the
measuring station.
2. Plug the weather station cable into the receptacle “weather station” on
the back of the controller.
3. Connect the LAN cable to the external sensor unit and the controller.
4. Connect the power supply cable to the external sensor unit and the con-
troller.
5. Connect the aerosol hose to the external sensor unit and the controller.
6. To connect to a network: Connect a network cable.
7. Connect the network cable but do not switch it on yet.
30 5200-en_V2.0_06/20
4 Initial commissioning
CAUTION
There is a risk of pinching and shearing points when placing the
sampling tube on the aerosol inlet tube.
Risk of injury to unprotected fingers.
4 Wear protective gloves.
1. Hold the sampling tube up to light to verify that there are no foreign ob-
jects in the tube.
2. Carefully guide the sampling tube over the aerosol inlet tube on the sen-
1 - 2 mm sor unit.
Important: There should be a gap of 1 to 2 mm between the sampling
tube and the contact surface. If the tube rests directly on the contact sur-
face, it can subject the aerosol sensor to mechanical strain.
3. Secure the sampling tube with a pipe clamp (provided on site). Ensure
that the tube is positioned vertically.
4. Plug the cable from the drying section IADS (on the sampling tube) into
the receptacle “IADS” on the back of the controller.
5200-en_V2.0_06/20 31
4 Initial commissioning
1 2 3 4 5 6 7
32 5200-en_V2.0_06/20
4 Initial commissioning
5200-en_V2.0_06/20 33
4 Initial commissioning
34 5200-en_V2.0_06/20
4 Initial commissioning
5200-en_V2.0_06/20 35
4 Initial commissioning
1. Release the union nut and remove it along with the clamping ring and
the sealing ring on the clamping ring.
2. Slide the clamping ring with the sealing ring over the sampling tube.
3. Guide the cable from the drying section IADS into the opening in the
weatherproof housing.
4. Carefully insert the sampling tube into the opening. Work carefully to
avoid damaging the cable.
36 5200-en_V2.0_06/20
4 Initial commissioning
CAUTION
There is a risk of pinching and shearing points when placing the
sampling tube on the aerosol inlet tube.
Risk of injury to unprotected fingers.
4 Wear protective gloves.
1. Hold the sampling tube up to light to verify that there are no foreign ob-
jects in the tube.
2. Carefully guide the sampling tube over the aerosol inlet tube on the
1 - 2 mm controller.
Important: There should be a gap of 1 to 2 mm between the sampling
tube and the contact surface. If the tube rests directly on the contact
surface, it can subject the aerosol sensor to mechanical strain.
3. Secure the sampling tube with the pipe clamp. Ensure that the tube is
positioned vertically.
4. Screw in the union nut and tighten by hand.
5. Plug the cable from the drying section IADS (on the sampling tube) into
the receptacle “IADS” on the back of the controller.
5200-en_V2.0_06/20 37
4 Initial commissioning
Prerequisites
Attaching the sampling head is the last step required to set up the device.
Once the sampling tube is fixed in place with a pipe clamp, attach the sam-
pling head.
Assembly
1. Place the sampling head on the sampling tube.
2. Secure the sampling head with two screws.
38 5200-en_V2.0_06/20
5 Operation and settings
The warm-up time, from switching on the device until valid measured data is
available, is about 10 to 15 minutes.
5200-en_V2.0_06/20 39
5 Operation and settings
40 5200-en_V2.0_06/20
5 Operation and settings
data Shows the course of the measured PM fractions and Data screen – current mea-
the concentrations. sured values [} 42]
dust chart Shows the course of the fine dust values measured. Dust chart screen – course of
fine dust values measured
[} 43]
air sensor Shows the course of the values measured by the Air sensor chart screen –
chart weather station. course of values measured by
weather station [} 44]
accessories Shows the menu for accessories as well as additional Accessories screen – acces-
information (IADS, weather station, filter system, parti- sories and other information
cle size distribution, PM values, set alarm, calibrate
[} 45]
weather station).
datalogger Data transfer from the internal memory to an external Datalogger screen – measured
storage medium (e.g. flash drive). data [} 54]
settings/cali- • Continuous monitoring of calibration. Settings/calibration screen
bration • Switch on and off calibration features. [} 56]
device status Provides an overview of critical system parameters Device status screen [} 58]
(volume flow, coincidence, pump output, weather sta-
tion, IADS, calibration, LED temperature and operat-
ing mode.
expert user Changes to expert mode. Expert mode [} 61]
menu
This mode is password-protected.
shut down Shuts down the device and the operating system. Completely switching off de-
vice [} 63]
menu Returns to the main menu from the various menus. -
device ready Device status. Device status screen [} 58]
or device ready The device is ready for operation.
5200-en_V2.0_06/20 41
5 Operation and settings
Data screen
42 5200-en_V2.0_06/20
5 Operation and settings
Touch clear charts to clear the display and start depiction of a new
course of the fine dust measured values.
5200-en_V2.0_06/20 43
5 Operation and settings
Note If the weather station provides other values, e.g. precipitation, wind direction
or wind speed, this data is also saved. Graphic display of the values can be
found under:
Main menu > accessories > weather station
Also refer to: Section "Weather station screen – data measured by weather
station [} 48].”
44 5200-en_V2.0_06/20
5 Operation and settings
Accessories screen
5200-en_V2.0_06/20 45
5 Operation and settings
Setting Effect
off (for testing) The IADS is switched off. But the internal tube in the IADS is
heated to +2K, based on the ambient temperature, to prevent
condensation inside of the IADS and the optical sensor.
remove volatile / moisture The IADS dissolves volatile components of the aerosol and com-
compensation (default setting for op- pensates the water condensation as well as the consequent parti-
eration as certified measuring equip- cle growth when the relative humidity is greater than 60%.
ment) The setpoint temperature is set dynamically as a factor of the am-
bient temperature and humidity.
remove volatile and semi- The IADS dissolves volatile and semi-volatile components of the
volatile (for testing) aerosol and compensates the effect of humidity on particle size.
The internal IADS heater is set to a constant temperature of
75 °C.
46 5200-en_V2.0_06/20
5 Operation and settings
Note The last setting is not saved. When the device is started up the next time, it
is reset to the default remove volatile / moisture compensation.
To change the setting permanently, change the file promo.ini:
To do this, set the numerical value for off to 0 in the line IADS-Modus=1,
and set remove volatile / moisture compensation to 1 an-
dremove volatile and semi-volatile to 2.
5200-en_V2.0_06/20 47
5 Operation and settings
Note If the model of the weather station used cannot supply certain values, the
default settings appear.
48 5200-en_V2.0_06/20
5 Operation and settings
These are followed by the PM values based on the density or the form factor
specified in the screen expert user menu > sensor/calibration.
These values are designated as “classic.”
The lower PM values are calculated pursuant to EN 481 and are applied par-
ticularly when measuring the quality of indoor air.
5200-en_V2.0_06/20 49
5 Operation and settings
Features:
Name of button Feature
50 5200-en_V2.0_06/20
5 Operation and settings
Menu path Main menu > accessories > particle size distribution
Note The end values on the Y-scales showing discrete distribution (red curves)
can be adjusted as needed. To do this, select the respective scale and over-
write the value (external keyboard or soft keyboard required).
5200-en_V2.0_06/20 51
5 Operation and settings
Alarms screen
Notification by e-mail
If this feature is activated (X in the box), the device sends a message to the
stated address if one of the status parameters exceeds the limits. Refer to
the section “Device status screen [} 58].”
52 5200-en_V2.0_06/20
5 Operation and settings
Menu path Main menu > accessories > calibrate weather station
Note Changes are not applied until after the system is restarted.
5200-en_V2.0_06/20 53
5 Operation and settings
Datalogger screen
Button Feature
enter comment Open a field in which a comment on the measured data can be en-
tered. The comment is saved every time the the measured data is
saved. This continues until the comment is deleted or changed.
copy datafiles to D:\ Copies measured data from the internal memory onto a flash drive.
54 5200-en_V2.0_06/20
5 Operation and settings
Button Feature
copy textfiles to D:\ Copies all text files created onto a flash drive.
delete all textfiles Deletes all text files created.
There is no repeated query of whether the text files really should be
deleted.
5200-en_V2.0_06/20 55
5 Operation and settings
Error message If the average deviation over the course of 40 hours is greater than 3.5 raw
data channels, an error is triggered. When this happens, the device has to
be verified with the MonoDust 1500; refer to the section "Verifying and cali-
brating measuring system [} 72].”
Individual points can lie outside of the limits. This can happen and is not a
cause for concern. Calibration is fine in this case, too.
Settings/calibration screen
56 5200-en_V2.0_06/20
5 Operation and settings
Button Meaning
sensor calibration Opens a window with the aerosol sensor settings; refer to the
section "Setting aerosol sensor offset [} 81].”
activate calib mode Activates calibration mode.
All subsequent data is marked as if it were measured during
the calibration process. The data will not be taken into consid-
eration for an evaluation, unless this is selected in PDAnalyze.
clean optical sensor Routine that heats the IADS to 75°C and at the same time al-
ternately brings the pumps from 0 l/min to the maximum flow
rate.
The process takes 15 minutes and cannot be interrupted.
activate calib mode Ends calibration mode and activates automatic mode.
5200-en_V2.0_06/20 57
5 Operation and settings
sensor flow The measured flow in between indicates the current flow rate. The set-
data is invalid. point is 4.8 l/min. The flow rate is standardized to standard
atmospheric temperature and pressure (SATP), i.e. based on
25°C and 1013 hPa.
If the actual value deviates more than 5% from the setpoint,
the measured data is invalid. An error indication appears.
flow velocity in between shows the detection volume,
which is the amount of air that is tested.
An error is indicated if the particle velocity deviates more than
15% from the setpoint.
The setpoint is the value velocity_calibrated entered in
the promo.ini file.
coincidence The measured Detection of more than one particle in the optical detection
data is invalid. volume.
58 5200-en_V2.0_06/20
5 Operation and settings
5200-en_V2.0_06/20 59
5 Operation and settings
Status log The Status log on the right side of the screen shows activities occurring
on the device, along with the date and time.
Display Meaning
60 5200-en_V2.0_06/20
5 Operation and settings
5200-en_V2.0_06/20 61
5 Operation and settings
Interface screen
62 5200-en_V2.0_06/20
5 Operation and settings
2. Touch system
5200-en_V2.0_06/20 63
5 Operation and settings
64 5200-en_V2.0_06/20
5 Operation and settings
5200-en_V2.0_06/20 65
5 Operation and settings
_palassupport.exe Starts the Teamviewer software. The Open the software for remote
software assigns a new password maintenance and support.
every time it is started.
_palassupport_default.exe Starts the Teamviewer software with Open the software for remote
a standardized password. maintenance and support.
Counter-win32.100###.exe Executable file. Starts the firmware. Start the Fidas firmware.
The file with the highest number
starts the newest version.
Data_auto_NNNNN_YYYY_MM_DD. File containing measured values. Save files regularly.
data.promo The file name contains the date on
Delete old files regularly.
which the file was created.
Promo.ini Control file with device settings. Save the file regularly and
when settings have been
changed.
Default.set.pdcontrol Control file with device settings. Save the file regularly and
when settings have been
changed.
66 5200-en_V2.0_06/20
5 Operation and settings
The file can be found on the desktop under startup > Fidas.
XXXX
XXXX
5200-en_V2.0_06/20 67
5 Operation and settings
Type= Fidas 200 Type designation of device group. Includes the models
Fidas 200 S and Fidas 200 E.
Ser# XXXX Serial number of device. Please do not change.
Password= XXXX Password for internal use. Please do not change.
User_device= XXX User-defined ID (3 characters)
Fidas 100_enabled= yes Plug-in for the top level user interface. Please do not
change.
Fidas 200/210_enabled= yes Plug-in for the top level user interface. Please do not
change.
Fidas 300/310_enabled= yes Plug-in for the top level user interface. Please do not
change.
My-palas.com_au- no Initializes automatic connection of the device to the
tostart= Palas website every time the device is started. Facili-
tates data upload and remote maintenance. Prerequi-
site: The device is registered on the website.
Velocity_calibrated= 9.3 m/s Particle velocity setpoint from factory calibration
PM10_slope= 1.000 Slope of fraction PM10. Default = 1. The value can be
changed on the basis of the TÜV (German technical in-
spection association) report 936/2122795.
PM10_intercept= 0 Axis intercept of fraction PM10. Default = 0. The value
can be changed on the basis of the TÜV (German tech-
nical inspection association) report 936/2122795.
PM4_slope= 1 Slope of fraction PM4. Default = 1.
PM4_intercept= 0 Axis intercept of fraction PM4. Default = 0.
PM2.5_slope= 1 Slope of fraction PM2.5. Default = 1. The value can be
changed on the basis of the TÜV (German technical in-
spection association) report 936/2122795.
PM2.5_intercept= 0 Axis intercept of fraction PM2.5. Default = 0. The value
can be changed on the basis of the TÜV (German tech-
nical inspection association) report 936/2122795.
PM1_slope= 1 Slope of fraction PM1. Default = 1.
PM1_intercept= 0 Axis intercept of fraction PM1. Default = 0.
PMtotal_slope= 1 Slope of fraction PMtotal. Default = 1.
PMtotal_intercept= 0 Axis intercept of fraction PMtotal. Default = 0.
PM_alternative yes / no Yes = The mesage ”alternative PM” value is active un-
der “accessories.”
textfile= yes / no Yes = saving data in a text file is activated.
textfile_interval= 60s Interval for saving data in a text file.
PM_autoadjust= no Default. Please do not change.
IADS_modus= 1 Mode of drying section IADS
68 5200-en_V2.0_06/20
5 Operation and settings
Weathersta- 1
tion_scale_p=
Weathersta- 1
tion_scale_h=
Weatherstation_off- 1 Entry of possible offset factors for adjusting the weather
set_T= station; obtained from comparison to transfer standard.
Weatherstation_off- 1
set_p=
Weatherstation_off- 1
set_h=
Weatherstation_equa- X_corr = scale * Function applied to calibrate the weather station.
tion= x + offset
5200-en_V2.0_06/20 69
5 Operation and settings
70 5200-en_V2.0_06/20
5 Operation and settings
The reports indicate that the measurement results related to gravimetry can
be optimized by modifying the slope and the intercept of data conversion.
Optimum correlation according to TÜV Rheinland:
PM10 slope = 1.058
PM10 intercept = -1.505
PM2.5 slope = 1.076
PM2,5 intercept = -0.339
We recommend reading the data with the default and offsetting it in your
own database and calculating the respective values.
5200-en_V2.0_06/20 71
6 Verifying and calibrating measuring system
Resources needed
The following resources are needed:
• Calibration dust MonoDust 1500
• Fidas calibration certificate
• Plug DN 16 mm to seal the aerosol inlet
• Plug DN 12 mm to seal the IADS inlet
• Plastic hose, approx. 30 cm long
• External flow meter, set to SATP conditions (25°C, 1013 hPa)
• HEPA* filter that fits IADS tube inlet
72 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
Prerequisites
• Permitted temperature range during calibration: +5°C to +40°C.
• The device has to have been running for at least one hour before it can
be calibrated. It does not reach a thermally stable state until then.
• The device has to be in calibration mode:
Main menu > settings/calibration >
activate calib mode
Screen settings/calibration
5200-en_V2.0_06/20 73
6 Verifying and calibrating measuring system
Check that aerosol tube and sensor unit - Checking that aerosol
are unobstructed sensor is unobstructed
[} 75]
Check that controller shows no leakage Net leakage of controller Checking that controller
Maximum leakage permitted: 0.07 l/min shows no leakage
[} 76]
Check device for leakage Net leakage of total system Checking device for
Maximum leakage permitted: 0.49 l/min leakage [} 78]
Check aerosol sensor zero point Cn Checking aerosol sensor
Correct value: 0.0 zero point [} 80]
74 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
5200-en_V2.0_06/20 75
6 Verifying and calibrating measuring system
When checking, the aerosol tube has to briefly be sealed with a plug (or with
your thumb).
76 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
5200-en_V2.0_06/20 77
6 Verifying and calibrating measuring system
When checking, the IADS tube has to briefly be sealed with a plug.
78 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
5200-en_V2.0_06/20 79
6 Verifying and calibrating measuring system
Cn: 0.00
Performing test
1. Detach the sampling head.
2. Attach a HEPA filter to the IADS tube inlet.
3. Select Main menu > data.
4. Observe the value of the parameter Cn: The value has to return to zero
within one minute.
5. Remove the HEPA filter.
80 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
Prerequisites
• Calibration mode is activated.
• The device has been tested for leakage and verified.
• The device’s zero point has been checked.
Performance
1. Select Main menu > expert user menu (password 1-).
2. Select settings/calibration.
ð The value changes as the controller varies the offset adjustment voltage.
When the displayed curve passes the minimum, the displayed value has
to be below 0.2 mV.
5200-en_V2.0_06/20 81
6 Verifying and calibrating measuring system
Prerequisites
• Calibration mode is activated.
• The device has been tested for leakage and verified.
• The device’s zero point has been checked.
• The aerosol sensor offset has been set.
5. Compare the value sensor flow rate to the value measured with
the flow meter.
The flow rate under normal operating conditions is about 4.8 l/min. The
value determined using the flow meter should be 4.8 ± 0.15 l/min.
If the value determined using the flow meter deviates more than 0.15 l/min.:
4 Adjust the flow rate.
2. Select sensor/calibration.
4. Enter the flow rate displayed by the flow meter by calibrate flow
sensor offset.
Be aware of the decimal point.
5. Confirm with accept.
82 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
Prerequisites
• Calibration mode is activated.
• The device has been tested for leakage and verified.
• The device’s zero point has been checked.
• The aerosol sensor offset has been set.
• The flow rate has been calibrated.
Performance
1. Select Main menu > expert user menu (password 1-).
2. Select settings/calibration.
5200-en_V2.0_06/20 83
6 Verifying and calibrating measuring system
Prerequisites
• Calibration mode is activated.
• The device has been tested for leakage and verified.
• The device’s zero point has been checked.
• The aerosol sensor offset has been set.
• The flow rate has been calibrated.
• Once / upon initial startup:
The average particle velocity has been set.
84 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
5200-en_V2.0_06/20 85
6 Verifying and calibrating measuring system
2. Select sensor/calibration.
86 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
ð A keypad opens.
3. If the measured value is lower than the reference value, increase the
value.
If the measured value is higher than the reference value, decrease the
value.
If the values are the same, calibration of the particle size is completed.
5200-en_V2.0_06/20 87
6 Verifying and calibrating measuring system
If the values are the same, calibration of the particle velocity is completed.
If the values are not the same, the value preset velocity has to be ad-
justed:
1. Touch calibrate velocity.
ð A keypad opens.
2. Enter the value of the parameter velocity at cursor (dust).
3. Touch accept.
88 5200-en_V2.0_06/20
6 Verifying and calibrating measuring system
2. Touch system
5200-en_V2.0_06/20 89
7 Maintenance
7 Maintenance
Regular maintenance of the device extends the service life and improves op-
erational reliability.
The maintenance tasks described here can be performed by the customer or
by the operator of the device. More extensive work, which is not described in
this manual, may be performed only by Palas specialists or by persons or or-
ganizations authorized by Palas. Unauthorized changes or modifications will
lead to loss of warranty. Palas will not be liable for damage caused by unau-
thorized changes or modifications.
Please contact Palas or your sales partner for maintenance and repairs not
described in this manual.
Palas GmbH
Greschbachstrasse 3b
D-76229 Karlsruhe
Germany
www.palas.de
Updating firmware
Upon customer request, an update to the latest firmware version can be per-
formed in the course of maintenance performed at the Palas plant. Ask us
about this.
90 5200-en_V2.0_06/20
7 Maintenance
5200-en_V2.0_06/20 91
7 Maintenance
WARNING
Harmful substances in the filter pose a health risk.
If the cover on the filter is open, harmful substances from the filter can es-
cape into the ambient air and endanger persons.
4 Do not use compressed air when the filter cover is open.
4 Close the removed filter airtight and dispose of it properly.
4 Do not attempt to clean the filter. The filter cannot be regenerated.
Proceed as follows:
1. Remove the clear protective cap from the filter by turning it to the left
and pulling it off.
If the cap cannot be turned by hand, use a pliers with soft contact sur-
faces. A cloth or a rubber mat can be used as contact surfaces.
2. Check the state of the interior sealing ring (Ø 30 mm). Replace the
sealing ring if necessary.
3. Extract the filter by turning it to the left and pulling it off.
4. Insert a new or cleaned filter by turning it to the right.
5. Turn the protective cap to the right by hand to put it back into place.
92 5200-en_V2.0_06/20
7 Maintenance
Sampling tube, aerosol inlet tube and gravimetric filter on aerosol sensor
5200-en_V2.0_06/20 93
7 Maintenance
NOTICE
Damage caused by unsuitable cleaning agents
Destruction of the optical glass.
4 Optical glass may be cleaned only with a special cloth for optics.
4 Do not touch optical glass with fingers.
4 Do not use cleaning agents such as acetone to clean the optical
glass.
94 5200-en_V2.0_06/20
7 Maintenance
5200-en_V2.0_06/20 95
7 Maintenance
1 2
1 2 3 4
96 5200-en_V2.0_06/20
7 Maintenance
5200-en_V2.0_06/20 97
7 Maintenance
Two fuses (fine wire fuse M2A) are built into the device to protect it against
overvoltage. The fuses can be accessed on the rear of the device. The fuses
are located in the fuse holder in the socket of the power plug.
To replace the fuses, proceed as follows:
1. Shut down the firmware and switch off the device.
2. Disconnect the power plug.
3. Pull out the fuse holder using a small screwdriver.
4. Replace the fuse.
5. Reinsert the fuse holder.
6. Reinsert the power plug.
Note Check whether the device starts up as usual and operates normally. Pay at-
tention to the startup of the device in order to rule out any damage to the
components – especially if the fuse was blown. If any irregularities are de-
tected, contact the manufacturer.
98 5200-en_V2.0_06/20
8 Errors
8 Errors
Self-diagnostics
The device has a self-diagnostics system. Several parameters are monitored
and an error is indicated if a limit is exceeded. check device status
then appears in the screen.
The screen device status shows the values and the status of the individ-
ual parameters. Detailed information can be found in the section "Device sta-
tus screen [} 58].”
Even if the device indicates device ready, the monitored values can be
helpful in detecting tendencies. Since some values change slowly as op-
posed to suddenly, changes can be indications of potential problems. This is
why we recommend observing the device status to be able to optimally plan
maintenance.
Depending on the communication profile used, status data may be output via
the interface. This is the case e.g. with the communication profile “Bayern-
Hessen.”
Most of the problems related to the monitored parameters can be solved by
cleaning or calibration processes.
5200-en_V2.0_06/20 99
9 Decommissioning
9 Decommissioning
To decommission the device when it is no longer needed, proceed as fol-
lows:
1. Switch off the device.
2. Unplug it from the power supply.
3. Disconnect all cables and lines.
4. Disassemble the device back to the components included in the deliv-
ery; refer to the section “Scope of delivery.”
100 5200-en_V2.0_06/20
10 Packaging and transportation
5200-en_V2.0_06/20 101
11 Disposal
11 Disposal
The device, accessories, and transport packaging are largely made of recy-
clable raw materials.
Transport packaging You can dispose of the transport packaging at collection points.
4 Observe the applicable national regulations.
Device and accessories The device and accessories must not be disposed of in household waste.
1. Ensure that the used device and any accessories are properly dis-
posed of.
2. If a battery is installed in the device, make sure that it is properly dis-
posed of.
3. Observe the applicable national regulations.
102 5200-en_V2.0_06/20
12 Spare parts
12 Spare parts
The following spare parts are available:
Art. no. Designation
5200-en_V2.0_06/20 103
13 Declaration of conformity
13 Declaration of conformity
The Manufacturer
Palas GmbH
Greschbachstraße 3b
D-76229 Karlsruhe
Germany
Karlsruhe, 01.03.2016
www.palas.de
104 5200-en_V2.0_06/20
14 Technical data
14 Technical data
Interfaces USB, LAN (Ethernet), RS-232, WLAN
Measurable particle sizes 0.18 – 18 µm
3 measuring ranges
Measurable particle mass 0 – 10 mg/m3
Measured values PM1, PM2,5, PM4, PM10, PMTotal, Cn
Air pressure, air temperature, relative hu-
midity
Size channels 64 (32 per decade)
Measuring principle Optical light scattering by single particles
Quantity measuring range Cn 0 - 20,000 particles/cm3
Time resolution 1 s - 24 h
In type-approved operation: 15 min
Volume flow rate 4.8 l/min ≘ 0.3 m3/h
Light source LED
Power consumption Normal operation: 60 W
Maximum: 200 W
Data storage 5 GB
Operating system Windows Embedded for Point of Service
(WEPOS)
Electrical connection 115/230 V, 50/60 Hz
Fuse T2A
Weight Controller: 9.30 kg
Sampling head: 2.25 kg
Sampling tube: 4.50 kg
Ambient temperature during Fidas 200 and Fidas 200 E: +5°C to +40 °C
operation Fidas 200 S: -20°C to +50 °C
Noise emissions < 70 dB(A)
Sampling system Aerosol drying with IADS
(Intelligent Aerosol Drying System)
Sampling head Passive sampler Sigma-2
5200-en_V2.0_06/20 105
Palas GmbH
Greschbachstraße 3 b
76229 Karlsruhe
Germany
Tel.: +49 721 96213-0
Fax: +49 721 96213-33
www.palas.de
mail(at)palas.de