Datasheet

Servo-air
System version 4.2

This document is intended to provide information to an international audience outside of the US.
Contents

Technical specifications............................................................... 3
General................................................................................................... 3
The ventilator – general................................................................... 3
Ventilation – general......................................................................... 4
User interface...................................................................................... 4
Power supply........................................................................................ 4
Gas supply............................................................................................. 4
Operating conditions........................................................................ 4
Non-operating conditions.............................................................. 4
Standards – safety and functionality.......................................... 5
Display.................................................................................................... 5
Ventilation modes – invasive ventilation................................... 5
Ventilation modes – non invasive ventilation.......................... 5
Non invasive ventilation.................................................................. 5
High Flow therapy (option)............................................................. 6
Parameter settings............................................................................ 6
Backup parameter settings............................................................ 6
Special functions................................................................................ 6
Disconnection / Suction.................................................................. 6
Monitoring and trends...................................................................... 7
Alarms..................................................................................................... 7
Autoset (alarm limits) specification............................................ 8
Aerogen nebulizers............................................................................ 8
Communication / Interface............................................................ 8
Log function......................................................................................... 8
Saving of data...................................................................................... 8
Optional equipment.......................................................................... 9
Service.................................................................................................... 9

Ordering information..................................................................... 9

Dimensional drawings................................................................... 10
Servo-air on Mobile cart.................................................................. 10
Servo-air on shelf base..................................................................... 10
Servo-air shelf base........................................................................... 10

2 GETINGE • SERVO -AIR


Servo-air
Technical specifications

General The ventilator – general

Intended use The Servo-air ventilator system is: Servo-air Servo-air on


mobile cart
• intended for respiratory support,
monitoring and treatment of pedi- Base system weight Approx. 15 kg Approx. 30 kg
atric and adult patients (33 lbs) (66 lbs)
• to be used only by healthcare pro- • Base system ap-
viders prox. 15 kg (33 lbs)
• to be used only in professional • Mobile cart ap-
healthcare facilities and for trans- prox. 15 kg (33 lbs)
port within these facilities
Dimensions of base 375 x 350 mm 647 x 547 mm
Instructions for use Please carefully read the user’s man- (W x D), see dimensional (14.8"x13.8") (25.5"x21.5") incl.
ual drawings wheels

Legal manufacturer Maquet Critical Care AB Height (incl. user inter- 489 mm (19.3") 1335 mm (52.6")
face)
Other products See separate datasheets.
Wheels N/A Four wheels with
Contact your local Getinge supplier
separate brakes
for more information.
A-weighted sound pres- <49 dB, measured at a distance of
sure level (LpA) 1 m (3.3 ft)

A-weighted sound pow- <57 dB


er level (LWA)

GETINGE • SERVO -AIR 3


Ventilation – general Gas supply

Patient range Tidal volume: Inlet gas pressure O2 200 – 600 kPa / 2.0 – 6.0 bar /
• Pediatric: 20 – 350 ml 29 – 87 PSI
• Adult: 100 – 2000 ml
Connection standards AGA, DISS, NIST, or French standard
Bias flow 2 l/min ±5 % available

Internal compressible factor Max. 0.1 ml/cmH2O Unavailable gas/loss The flow from an unavailable gas (O2)
of gas pressure is automatically compensated for so
Gas delivery system Air turbine and O2 valve
that the patient gets the preset vol-
Maximum airway pressure 125 cmH2O
ume and pressure.
Method of triggering Flow and pressure
Patient system gas Male 22 mm / female 15 mm. In ac-
Inspiratory flow range • Adult: 0 – 240 l/min connectors cordance with ISO 5356-1.
• Pediatric: 0 – 240 l/min
Gas exhaust port Male 30 mm cone
Pressure drop Max. 3 cmH2O at a flow of
60 l/s (exp. channel)

PEEP regulation Microprocessor controlled


valve

Rise time, expiratory flow mea- <12 ms for 10 – 90 % response


Operating conditions
surement at flow of 3 – 192 l/min
Operating temperature +10 to +40ºC (+50 to +104ºF)
Expiratory flow range 0 – 192 l/min
Relative humidity 15 to 95% non-condensing

Atmospheric pressure 660 to 1060 hPa

Lowest pressure in pa- -400 cmH2O


tient circuit
User interface

Type TFT-LCD touchscreen

Size 300 x 248 mm (11.8" x 9.8")


Non-operating conditions
Viewing area 12" XGA, 1024x768 pixels with a 24-
bit color palette
Storage temperature -25 to +60ºC (-13 to +140ºF)

Storage relative humidity 10 to 95% non-condensing

Storage atmospheric pres- 470 to 1060 hPa


sure
Power supply

Power supply, auto- 100 –240 V AC ±10%, 50 – 60 Hz


matic range selection

Plug-in battery module:


• Battery backup (Li-ion) • Two battery module slots. One bat-
tery is delivered with the ventilator.
• Battery capacity • Rechargeable, 14.4 V, 6.6 Ah each
• Battery backup time • Approximately 2 h (factory new
battery)
• Recharge time • Approximately 3 h/battery

External 12 V battery 12.0 V – 15.0 V DC, 15 A

4 GETINGE • SERVO -AIR


Standards – safety and functionality Ventilation modes – invasive
ventilation
The device complies with re-
0123
quirements and classification IIb
Controlled ventilation • PC (Pressure Control)
of Medical Devices Regulation
• VC (Volume Control)
(EU) 2017/745.
• PRVC (Pressure Regulated Volume
CE Mark Notified Body number: Control), option
0123.
Supported ventilation: • PS/CPAP (Pressure Support / Con-
Classification IEC 60601-1: 2005 + A1:2012, tinuous Positive Airway Pressure)
Class I, continuous operation. • VS (Volume Support), option

Applied parts: AUTOMODE (option) • Control mode: VC <–>


- Equipment making physical Support mode: VS
contact with the patient and • Control mode: PC <–>
the gas path ways. Type B Support mode: PS
- Nebulizer patient unit and • Control mode: PRVC <–>
­cable. Type BF Support mode: VS

Standards • Applied parts, i.e. equipment Combined ventilation • SIMV (VC) + PS (Synchronized In-
making physical contact with termittent Mandatory Ventilation)
the patient, are described in • SIMV (PC) + PS
Ventilation patient connec- • SIMV (PRVC) + PS (option)
tion – System flowchart. • Bi-Vent/APRV (Airway Pressure
• ISO 80601-2-12:2011 Release Ventilation), option
• ISO 80601-2-55:2018
• EN 13544-1:2007 + A1:2009

Ingress protection IP 21

Electromagnetic compati- According to limits specified in


bility (EMC) IEC 60601-1-2:2014
Ventilation modes – non invasive
ventilation

Controlled ventilation NIV PC, option

Supported ventilation: NIV PS, option


Display
Views • Basic view
• Advanced view
• Loops view
• Distance view
Non invasive ventilation
• Family view
• Servo Compass view (option) Max. leakage compen- • Adult:
• Each of the screen layout views offers a sation level - Inspiratory: up to 240 l/min*
specific combination of displayed wave- - Expiratory: up to 65 l/min
forms, loops and presented values. • Pediatric:
- Inspiratory: up to 240 l/min*
Real time wave- • Pressure
- Expiratory: up to 25 l/min
forms • Flow
• Volume
* up to 180 l/min with 100% O2 concentra-
Loops • Pressure – Volume
tion
• Volume – Flow
Disconnection flow • Low: 7.5 l/min
A reference loop and two overlaying loops can be displayed.
(configurable) • High: 40 l/min
Servo Compass Visualizes volume (VT/PBW) and pressure • Disabled: Deactivates disconnec-
(total or driving) in relation to set targets tion detection
in invasive modes.
Connection detection Manual or automatic via bias flow

GETINGE • SERVO -AIR 5


High Flow therapy (option) Backup parameter settings

Flow setting range • Adult: 5 – 60 l/min Parameter Adult Pediatric


range range
• Pediatric: 2 – 30 l/min
Inspiratory tidal volume (ml) 100 – 2000 20 – 350

Pressure level above PEEP in 5 – 100 5 – 80


backup (cmH2O)

Pressure level above PEEP in NIV 5 – 60 5 – 60


Parameter settings backup (cmH2O)

Respiratory rate in backup 4 – 100 4 – 150


Parameter Adult Pediatric (breaths/min)
range range
I:E ratio 1:10 – 4:1 1:10 – 4:1
Tidal volume (ml) 100 – 2000 20 – 350
Ti (s) 0.1 – 5 0.1 – 5
Minute volume (l/min) 0.5 – 60 0.3 – 20

Apnea, time to alarm (s) 15 – 45 2 – 45

Max. apnea time in Automode (s) 7 – 12 3 – 15

Pressure level above PEEP (cmH2O) 0 – 100 0 – 80


Special functions
Pressure level above PEEP in NIV 0 – 60 0 – 60
(cmH2O)
Special function Setting range
PEEP (cmH2O) 0 – 50 0 – 50
Manual breath Initiation of 1 breath (In SIMV mode
PEEP in NIV (cmH2O) 2 – 20 2 – 20
initiation of 1 mandatory breath)
Respiratory rate (breaths/min) 4 – 100 4 – 150
Static measurements Insp. or exp. hold (0 – 30 seconds)
SIMV rate (breaths/min) 1 – 60 1 – 60
Nebulization 5 – 30 min/Continuous/Off
Breath cycle time, SIMV (s) 1 – 15 0.5 – 15
O2 boost level Off, 1 – 79 %
PHigh (cmH2O) 2 – 50 2 – 50
O2 boost function Activate O2 boost up to 1 minute
THigh (s) 0.2 – 30 0.2 – 30
Leakage compensation Automatic in all non invasive
TPEEP (s) 0.1 – 10 0.1 – 10 modes

PS above Phigh in Bi-Vent/APRV 0 – 98 0 – 78 Circuit compensation On/Off


(cmH2O) (not available in NIV)

O2 concentration (%) 21 – 100 21 – 100 Previous mode Activates previously used mode

I:E ratio 1:10 – 4:1 1:10 – 4:1 Backup ventilation Backup On/Off

Ti (s) 0.1 – 5 0.1 – 5 Apnea management Several parameters

TPause (s) 0 – 1.5 0 – 1.5

TPause (% of breath cycle time) 0 – 30 0 – 30

Flow trigger (l/min) 0–2 0–- 2

Pressure trigger (cmH2O) -1 to -20 -1 to -20 Disconnection / Suction


Insp. rise time (% of breath cycle time) 0 – 20 0 – 20
Pre-oxygenation time Max. 2 min
Insp. rise time (s) 0 – 0.4 0 – 0.2
Post-oxygenation time Max. 1 min
End inspiration (% of peak flow) 1 – 70 1 – 70
Patient disconnected High priority alarm activated after
End inspiration (% of peak flow) in NIV 10 – 70 10 – 70
1 min
Decelerating flow pattern in VC (%) 0–100 0–100
Adjustable oxygen level 21 – 100 %
Flow adaptation in VC on/off on/off

6 GETINGE • SERVO -AIR


Monitoring and trends Alarms

Peak airway pressure Ppeak Alarm Adult range Pediatric range

Pause airway pressure Pplat Airway pressure (upper 16 – 100 16 – 90 cmH2O


alarm limit) cmH2O
Mean airway pressure Pmean
Airway pressure NIV (upper 16 – 70 cmH2O 16 – 70 cmH2O
Driving pressure Pdrive
alarm limit)
Positive end expiratory pressure PEEP
Respiratory rate (upper 2 – 160 2 – 160
Spontaneous breaths per minute RR sp
alarm limit) breaths/min breaths/min
Respiratory rate RR
Respiratory rate (lower 1 – 159 1 – 159
Spontaneous expiratory minute volume MVe sp alarm limit) breaths/min breaths/min

Inspired minute volume MVi Expired minute volume 1 – 60 l/min 0.02 – 30 l/min
(upper alarm limit)
Expired minute volume MVe
Expired minute volume 0.5 – 40 l/min 0.01 – 20 l/min
Leakage fraction (%) Leakage
(lower alarm limit)
Inspired tidal volume VTi
End expiratory pressure 1 – 55 cmH2O 1 – 55 cmH2O
Expired tidal volume VTe
(upper alarm limit)
End expiratory flow Flowee
End expiratory pressure 0 – 47 cmH2O 0 – 47 cmH2O
Measured oxygen concentration O2 conc. (lower alarm limit)

Dynamic compliance Cdyn No patient effort (Apnea) 15 – 45 s 2 – 45 s


alarm
Static compliance Cstatic
Automatic return to support
Inspiratory resistance Ri
mode on patient triggering
Expiratory resistance Re
No consistent patient effort Yes, described in User’s manual
Work of breathing, ventilator WOBvent
High continuous pressure Yes, described in User’s manual
Work of breathing, patient WOBpat
O2 concentration Set value ±5 vol% or ≤18 vol%
Elastance E
Gas supply Below 200 kPa (2.0 bar/29 PSI),
P 0.1 P 0.1
above 600 kPa (6.0 bar/87 PSI)
Shallow Breathing Index SBI
Battery • Limited battery capacity:
Ratio of expired tidal volume to predicted VT/PBW 10 min
body weight • No battery capacity: less than
3 min
Ratio of expired tidal volume to body VT/BW
• Low battery voltage
weight
Leakage too high Yes, described in User’s manual
Switch to backup (/minute) Trended value
only Technical Yes, described in User’s manual

Backup (%/min) Trended value


only

GETINGE • SERVO -AIR 7


Autoset (alarm limits) specification Communication / Interface

Autoset (alarm limits) Invasive ventilation, controlled Serial ports • Two RS-232C ports. For data com-
specification modes only
munication via the Servo Communi-
High airway pressure: Mean peak pressure +10 cmH2O or cation Interface (SCI).
at least 35 cmH2O
Servo Communica- A protocol for data communication
Expiratory minute vol- Mean expiratory minute volume tion Interface (SCI) with external devices
ume (upper alarm limit) +50 %
Alarm output con- • 4-pin modular connector for com-
Expiratory minute vol- Mean expiratory minute volume nection (option) munication of all active alarms
ume (lower alarm limit) -50 % • Switching capability: Max. 40 V DC,
max. 500 mA, max. 20 W
Respiratory rate (upper Mean respiratory rate +40 %
alarm limit)
Data transfer via USB • For transfer of trends, logs, screen-
Respiratory rate (lower Mean respiratory rate -40 %
port shots and recordings to a USB
alarm limit)
memory stick
End expiratory pressure Mean end expiratory pressure
Ethernet port • The network connection (LAN) port
(upper alarm limit) +5 cmH2O
is for service use
End expiratory pressure Mean end expiratory pressure
(lower alarm limit) -3 cmH2O

Log function
Aerogen nebulizers
Event log • Alarms
• Ventilator settings
Aerogen nebu- Pro Solo
• Apnea periods
lizers
• Immediate functions
Size W 50 x L 50 x H 45 mm W 48 x L 25 x H 67 mm
Service log • Technical alarms
(W 2.0" x L 2.0" x H 1.8") (W 1.9" x L 1.0" x H 2.6")
• Test results
Weight Approx. 25 g (0.88 oz) Approx. 14 g (0.49 oz)
• Service records
Particle size 1 – 5 µm mass median aerodynamic diame- • Software installation
ter (MMAD) • Configuration information

Flow rate >0.2 (average: ~0.4) ml/min

Max. volume 10 ml 6 ml

Residual volume <0.1 ml for 3 ml dose

Control cable 1.8 m (5.9 ft) Saving of data

Recording of current 30 seconds of data will be recorded


waveform and param- (15 seconds before and 15 seconds
eter values after activation). Up to 40 recordings
can be stored.

Saving screenshots Up to 40 screenshots can be stored.

Export files Recordings, screenshots and the


event log can be saved together in
an export file and transferred to a
USB memory stick.

8 GETINGE • SERVO -AIR


Optional equipment

Optional equipment Weight Dimensions Maximum load

Mobile cart 15.0 kg (33.0 lbs) W 647 x L 547 x H 860 mm -


(W 25.5" x L 21.5" x H 33.9")

Shelf base 3.0 kg (6.6 lbs) W 340 x L 270 x H 43 mm -


(W 13.4"x L 10.6" x H 1.7")

Humidifier holder 0.6 kg (1.3 lbs) W 76 x L 125 x H 140 mm 12 kg (26.5 lbs)


(W 3.0" x L 4.9" x H 5.5")

Support arm 179 1.9 kg (4.2 lbs) Length 900 mm (35.4") • 1 kg (2.2 lbs) at 180°
• 1.5 kg (3.3 lbs) at 90°
• 3 kg (6.6 lbs) at 45°

Cable holder for handle 0.1 kg (0.2 lbs) W 138 x L 92 x H 155 mm 5 kg (11.0 lbs)
(W 5.4" x L 3.6" x H 6.1")

Waterbag/IV pole 0.4 kg (0.9 lbs) W 148 x L 26 x H 1007 mm 1.5 kg (3.3 lbs)
(W 5.8" x L 1.0" x H 39.6")

Gas cylinder restrainer kit 1.0 kg (2.2 lbs) Upper: W 104 x L 65 x H 48 mm Two 4.5-liter bottles
(W 4.1" x L 2.5" x H 1.9")

Lower: W 106 x L 162 x H 76 mm


(W 4.1" x L 6.4" x H 3.0")

Y piece holder - W 26 x L 52 x H 46 mm -
(W 1.0" x L 2.0" x H 1.8")

Service

Regular maintenance Once every 12 months or at least after 5000 operating hours

Ordering information
Servo-air, ventilator system and accessories: See separate information:
“System Flow Chart Servo-air“ (Order no: 68 91 363).

GETINGE • SERVO -AIR 9


Dimensional drawings
Servo-air on Mobile cart Servo-air on shelf base

475 mm
18.70”
334 mm
300 mm 13.15”
11.81” 34 mm
1.33”
1335 mm

1115 mm
52.56”

248 mm
43.90”

9.76”
860 mm
33.86”

489 mm
19.25”

275 mm
311 mm

10.83”
12.24”
547mm 647 mm 375 mm 350 mm
21.54” 25.47” 14.76” 13.78”

Servo-air shelf base


43 mm
1.69”

340 mm 270 mm
13.39” 10.63”

*Country specific part and directly removable parts excluded

10 GETINGE • SERVO -AIR


Notes

GETINGE • SERVO -AIR 11


© Maquet Critical Care AB 2021 · Subject to modifications · Servo-air, Servo Compass are trademarks by Maquet Critical Care AB ·
Subject to modifications · Order No. 68 91 357 Rev 02 · Servo-air, System version 4.2 – Datasheet· English · 2021-07

Getinge is a global provider of innovative solutions for operating rooms, intensive care units, sterilization departments and
for life science companies and institutions. Based on our firsthand experience and close partnerships with clinical experts,
healthcare professionals and medtech specialists, we are improving the everyday life for people, today and tomorrow.

Servo-air may be pending regulatory approvals to be marketed in your country. Contact your Getinge representative for
more information. This document is intended to provide information to an international audience outside of the US.

Manufacturer · Maquet Critical Care AB · Röntgenvägen · 2 SE-171 54 Solna · Sweden · +46 (0)10 335 73 00