FOCAS-2000 Instruction-Manual Rev5
FOCAS-2000 Instruction-Manual Rev5
FOCAS-2000
䎃 䎃 Oil Concentration Detector
Instruction Manual
Read this instruction manual thoroughly before use to operate the product safely by following precautions.
Be sure to keep it in an easy-to-access place so as to refer to it whenever necessary.
Page. 1
Contents
1. General
1-1. Introduction 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 2
1-2. Product Features 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 2
1-3. Safety precautions 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 3
2. Specifications
2-1. Product Specification 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 3
2-2. External dimensions and component names 䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 4
2-3. Internal layout figure and terminal block 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 5
2-4. Relay contact output, alarm lamp movement view 䯂䯂䯂䯂䯂 page. 6
3. Connecting pipes
3-1. Mounting space 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 7
3-2. Precautions for piping work 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 7
4. Preparing for Operation
4-1. About items displayed on the screen 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 8
4-2. Outline of operating conditions 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 9
4-3. Preparing for Operation 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 9-10
5. Operation
5-1. Test-operation method by the Manual operation function 䯂䯂 page. 11-12
5-2. Test-operation method by the Automatic operation
䯂䯂䯂䯂䯂䯂 page. 13
function
5-3. Function of judging whether the detection tube is clean 䯂䯂 page. 14
6. Setting
6-1. Tabulated list of menu items 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 15
6-2. Time correction of the clock 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 16
6-3. Various settings 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 17-18
7. Operation and Maintenance 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 19
8. Maintenance and Checking
8-1. Testing function 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 19-24
8-2. Maintenance 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 25-27
8-3. Viewing the operational log 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 28-30
8-4. Cleaning the sensor (brush washing) 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 31
8-5. Cleaning the filter 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 31
8-6. Checkup and replacement timing of
䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 32
desiccant (silica gel)
8-7. Replacing the clock backup battery 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 32
8-8. Replacing the operational log memory card 䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 33
8-9. Replacing the fuses 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 33
9. Troubleshooting 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 34-37
10. Service parts 䯂 Option parts 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 38
11. Request 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 38
12. Display of groups and description of statuses 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 39-40
13. Additional Information
13-1. About the Operational log of normal operation cycle 䯂䯂䯂䯂䯂 page. 41
13-2. Technical Instruction ”Calibration Overdue” 䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂䯂 page. 42
Page. 2
1
䢢 .䢢 䢢
伪䢢 1-1.䢢 䢢
䢢
䢢
䢢
䢢
1.
●
●
●
2.
3.
伪䢢 1-2.䢢 䢢
䢢
●
●
●
●
●
●
・
Page. 3
1
䢢 .䢢 䢢
伪䢢 1-3.䢢 䢢
●
DANGER
●
WARNING ●
CAUTION ●
䢰
●
2.䢢 䢢
伪䢢 2-1.䢢 䢢
䢢
䢢
到 到 䢢
凚 凛
凬 到 䢢 凬 到 䢢
凬 到
到
凚 凛 䢢 䢢 䢢凚 )
兟 到 䢢 到 凘 䢢
凬
凬 凬
傎 傎 傎
凚到 凛 凚 凛䢢
凚到 凛 凚 凛䢢
凚 凛 凚 凛
φ φ
凚φ 到 φ 凛
凬 凚 凛
Page. 4
䎃 .䎃 Specifications䎃
2
䦟䎃 2-2.䎃 External dimensions and component names䎃
䎃
Component names sample water outlet 251
Screw-in type joint 165 4.0
122 (R1/4 䕮8) 116 1.0 4-㱂10
ᵈᗧ
Upper lid
CAUTION
260 1.2
䬺ᵈᗧ䬦䬵ਅ䬤䬓䫻
Case cover
283.5
䬢䭛䭘䬽ᵈᗧ䭡ᕃ䭚䬷䫺ᯏེ䬽៊்䭓ṳ㔚╬䭡↢
䬧䭚ᕟ䭛䬛䬑䭙䫺䭍䬮ᯏ⢻ᡰ㓚䬷䬹䭙䭍䬨䬽䬶චಽ
䬺ᵈᗧ䬦䬵ਅ䬤䬓䫻
Relay pipe
Main body
Solenoid
MODEL OF TYPE APPROVAL
Snap latch
Esc Enter
ဳ
CANADA NO,
㪬㪪㪚㪞㩷㪥㫆㪅
EC
ᑼ
MO D E L
ㅧ ⇟ ภ
SERIAL NO.
ㅧ ᐕ
15ppm Bilge Alarm
FOCAS-2000
valve unit Operation switch
DATE 20䇭䇭䇭䋭
Filter
Maximum 265
Earth screw
(M4)
Panel configuration Maximum 354
A Display
15ppm Bilge Alarm
LCD display
MODEL FOCAS-2000
A Display
POWER AL. 1 AL. 2
POWER Green lamp lights when applying current.
AL. 1 Red lamp blinks when oil alarm is activated or
Power lamp Alarm2 lamp (red) AL. 2 the unit is defective.
(green) Alarm1 lamp (red) B Operation switch
B Operating switch Esc Used to set items and select a function excluding the Auto Run function.
Esc Enter
Used to change display items and select or change set values.
Enter Used to determine and execute a setting.
Page. 5
䢢 .䢢 Specifications䢢
2
伪䢢 2-3.䢢 Internal layout figure and terminal block䢢
2-3-1 Inside component layout of FOCAS-2000 Memory card for FOCAS-2000 white or black with marking (circle painted all red)
Desiccant (silica gel)
余 For replacement of the desiccant, see
Operation record memory "8-6 Checkup and replacement timing
of desiccant (silica gel)."
余 For replacement of the memory card, see
"8-8 Replacing the operational log memory."
余 Use (or press) this when the machine is frozen and cannot be recovered. 3 External output alarm1 signal (Alarm1)
COM
COM
NO
NC
NO
NC
FG
L
N
䢢
2
䎃 .䎃 Connecting pipes䎃 䎃
3
䦟䎃 3-1.䎃 Mounting space
䚚䎔
2m or less overall length Screwed oint
Within 1.5m head
the top of Bilge Separator.
lower than 2m down from
Do not install Bilge Alarm
(recommended)
(R1/4 䕮8 )
Drain outlet
Bilge Alarm
opper
Drainage
eight of
Sample water inlet
Bilge 3m or less overall length
Separator Screwed oint
Sample water inlet (R1/8 䕮8 ) 䚚䎔
䯴0.03䱊0.3MPa䯵
Be sure to set Bilge Alarm above Fresh water inlet
the outlet level of Bilge Separator. (Equivalent of Mineral water)
䯴0.03䱊0.6MPa䯵
opper Place the valve near the
concentration detector.
Allow the pipe upstand around Bilge Separator outlet to
provide Sample water for testing.
(In accordance with MEPC.107(49) 6.1.1.䯵
䎃
ƕ Use fresh water equivalent of mineral water, which is indispensable for sensor correction.䎃
ƕ In order to prevent dust contamination from fresh water line, please be sure to carry out cleaning of the line.
● Tubes to be connected to the sample water inlet, fresh water inlet, and outlet should be copper tubes of 8 mm in
Sample water and outer diameter. 䚚䎔
fresh water pipes ● Normal operating pressure 0.03 to 0.3 MPa for sample water and 0.03 to 0.6 MPa for fresh water
䚚䯽 When the oil concentration detector is shipped from the factory, its connection ports are for tubes of 8 mm in
diameter (as standard). Use an accessory reducing oint to connect a 10mm-dia. copper tube.
Apply seal tape to the threaded part of the copper tube before coupling the tube to the port.
ƕ䎃 The length from the sample water outlet to the end of piping shall be 2 m or less and the fall shall be within 1.5m.
Be sure to provide hoppers.
ƕ䎃 Piping length between Bilge Separator and Bilge Alarm shall be 3m or less. Bilge Alarm shall be set above the
Restrictions on piping
Outlet height of Bilge Separator. Do not install Bilge Alarm lower than 2m below the top of Bilge Separator.
Operation pressure of Bilge Alarm is 0.03Mpa 0.3Mpa. In consideration of recommended pressure for Bilge
Separator, ad ust the operation pressure.䎃
Caution Be sure to follow the above instructions to run the oil concentration detector normally.
Page. 8
4.䎃 Preparing for Operation䎃
䎃
䎃 䦟 4-1. About items displayed on the screen
z Standby state
State Display item䎃 Description
z Automatic operation
State Display item䎃 Description
Example of display during "operation" 1RGTCVKPIAAAAAAZZZZZZ The "AL1" alarm lamp is off when the oil
Oil concentration < Alarm1 oil alarm 8CNXG㧧10A㨉AAAAAAAZZ=RRO? concentration is less than Alarm1 oil
point alarm point.
State of signal output from Oil concentration
the 3-way valve
ON = The bilge drains. Less than oil alarm point
Example of display during "operation" 1RGTCVKPIAAAAAAZZZZZZ The "AL1" alarm lamp blinks at the
Oil concentration䟮Alarm1 oil alarm 8CNXG㧧1((㨉AAAAAAAZZ=RRO? same time.
point State of signal output from the Oil concentration The bilge does not drain.
3-way valve Equal to or more than oil alarm point
OFF = The bilge does not drain.
z Washing
State Display item䎃 Description
Current state: Operating
Shows these items when the washing is Feed fresh water into the detection tube
in progress. 1RGTCVKPI to wash the inside of the tube when the
Shows the remaining washing time 9CUJKPIZZZ sample water pressure goes under 0.03
(in seconds). MPa after the bilge pump stops.
Details: Washing in progress Remaining time
Page. 9
4.䎃 Preparing for Operation䎃
䎃
䦟 4-2. Outline of operating conditions
Ɣ The FOCAS-2000 oil concentration detector automatically starts to operate when detecting a water pressure of 0.03 MPa or more at the
sample water inlet.
Ɣ After starting to run, the oil concentration detector waits for a preset time (warm-up time) until the sample water state becomes stable so as not
to detect a residue in the area between the sample water pipeline and the detection tube. In this state, the three-way valve output signal is off
and the bilge does not drain.
Ɣ The oil concentration detector has two oil alarm points (Alarm1 and Alarm2). You can set an alarm point (in ppm) in the range of 5 to 15 ppm.
Ɣ The Alarm1 oil alarm point is in synchronism with the three-way valve output signal. When the oil concentration is less than the specified oil
alarm point, the bilge drains. By the way, when the oil concentration exceeds the oil alarm point, the three-way valve output signal is output to
immediately stop drainage of the bilge independently of the setting of Alarm1 oil alarm delay.
Ɣ When the water pressure at the sample water inlet falls under 0.03 MPa, the oil concentration detector stops operation, feed fresh water into
the detection tube for a preset time period to keep it clean and leave no residue in the tube, and enters the standby state.
4. Turn on power to the oil concentration detector. Make sure the "POWER" lamp (green) lights.)
5. When the oil concentration detector is turned on, the alarm lamps blink and the LCD screen shows the following.
ᵏᵓᶎᶎᶋᴾᵠᶇᶊᶅᶃᴾᵟᶊᵿᶐᶋ
9CTPKPIA=%?AAAAAAAAAAA
46%A.QUVAXQNVCIGAAAAAAAA
MODEL FOCAS-2000
Esc Enter POWER AL. 1 AL. 2
Ŷ Use the Manual Operation function to test-operation the oil concentration detector after installing and constructing it or in the
following cases:
Ɣ when the bilge separator is disabled but you want to execute an alarm operation test including three-way valve output signals.
Ɣ when a pressure of 0.03 MPa or higher cannot be applied to the sample water inlet
Ŷ The Manual Operation enables you the following:
Ɣ The manual operation is implemented for 5 minutes independently of the setting state of the pressure switch. When the manual operation
ends, ”Main Menu” returns. And then the oil concentration detector enters the standby state.
Ɣ The alarm operation test can be executed when the concentration is less than the specified oil alarm point during manual operation.
Ŷ Manual test operation procedure
Make sure the oil concentration detector is in the standby state. ("Standby" and the current time
5VCPFD[
are displayed on the screen portion of the oil concentration detector.)
ZZZZZZZZZZZZ
Close the valve provided before the sample water inlet.
Open the upper lid of the sensor unit and the drain valve to empty the sensor unit. Close the
drain valve and feed mineral water into the detection tube (to about 80% of the detection tube)
from the top of the sensor unit (from which the upper lid is removed), and close the upper lid.䎃
Keep on pressing “Esc” for two seconds longer in the standby state. Enter “Main Menu.”
㧷CKP㨉/GPW Press "Ÿ" twice. "Test Mode" is displayed under "Main Menu."
6'56㨉/QFG Press "Enter." A screen appears to select respective TEST function items.
&GVGEV㨉2TGUUWTG㨉㨉AA㨉㨉㨉AA
This screen is displayed for about 5 seconds and the manual operation starts.
AAAZZZZZZAZZZZZZ
/CPWCN㨉TWPPKPI㨉AZZZZZZ Shows an oil concentration during operation. This screen shows the state of the three-way valve
8CNXG10ZZRRO output signal and the measured oil concentration.
When the oil concentration is less than the oil alarm point (when "Alarm" is not output), you can
execute an alarm test including the three-way valve output signal.
Press "Enter" for 2 seconds or longer. The alarm test starts. (Continued to the next page.)
Page. 12
5.䎃 Operation䎃
䎃
Continued from the preceding page. (Manual test-operation procedure)
When the alarm test starts, the oil concentration detector artificially increases the oil concentration
to enable the alarm test. "sim" appears in the center of the top line of the screen.
The ppm value "x" is increased by one each time you press "Ÿ.” (You can reduce the value
similarly by pressing "ź" but cannot reduce the value under the current measured concentration
/CPWCN㨉TWP㨉UKOAZZZZZZ value.) When the measured oil concentration exceeds the oil alarm point (Alarm1 and Alarm2),
8CNXG10ZZRRO the oil concentration detector enters the alarm state and turns off the three-way valve output 䎃
signal.You can confirm the alarm operation. Both AL1 and AL2 lamps blink. The alarm test is
disabled when the oil concentration of the sample water goes over the oil alarm point in the test.)
/CPWCN㨉TWP㨉UKOAZZZZZZ Press "Esc" after confirming the alarm test.
8CNXG1((ZZARRO About 5 minuites later with “Esc” unpressed, “sim” disappears from the screen. 䎃
The operation ends. “Main Menu” returns. And then the oil concentration detector enters the
standby state.
"sim" disappears and "Manual running" returns.
/CPWCN㨉TWPPKPIZZZZZZ This screen shows the ON/OFF state of the three-way valve output signal and the measured oil
8CNXG10ZZARRO concentration. Press "Esc" for 2 seconds and unpress it.
The operation ends and the oil concentration detector enters the washing state.
/CPWCNTWPPKPI
The remaining time (in seconds) is displayed to the right of "Washing."
9CUJKPIZZZ
When the washing operation ends, the oil concentration detector waits for judgment of whether
the inside of the detection tube is clean.
/CPWCNTWPPKPI "Wait Stable" appears and the remaining time (in seconds) is displayed to the right of "Wait
9CKV5VCDNGZZ Stable."
6'56㨉/QFG After the washing operation ends, "TEST Mode" appears to select “Manual running” of the TEST
/CPWCNTWPPKPI function. Press "Esc"
㧷CKP㨉/GPW
"Main Menu" returns. Press "Esc."
6'56㨉/QFG
5VCPFD[
"Standby" returns.
ZZZZZZAAZZZZZZ
Ŷ Use the Automatic Operation to test-operation the oil concentration detector after installing and constructing it or in the following cases:
Ɣ when you want to operation the bilge pump and execute an alarm test including the three-way valve output signal.
Ɣ when a water pressure of 0.03 MPa or higher can be applied to the sample water inlet.
Ŷ The Automatic Operation enables you the following:
Ɣ When a water pressure of 0.03 MPa or higher is applied to the sample water inlet, the oil concentration detector always starts operation. When
detecting a low water pressure (lower than 0.03 MPa) for a preset time period, the oil concentration detector ends the washing operation and
returns to the standby state.
Ɣ The alarm operation test can be executed when the concentration is less than the specified oil alarm point during auto operation.
Ŷ Test operation procedure by the Automatic Operation
5VCPFD[AAAAAAAAAAAAAAAA Check and make sure the bilge separator is ready to run or when both sample water and fresh water
AAAZZZZZZAAAZZZZZZ can be supplied to the oil concentration detector. In the standby state while "Standby" and the
current time are displayed on the screen, the oil concentration detector automatically starts operation
when the bilge pump and the bilge separator runs or sample water of 0.03 MPa or higher is supplied
to the sample water inlet.
&GVGEV㨉2TGUUWTG㨉㨉AA㨉㨉㨉AA
This screen is displayed for about 5 seconds and the auto operation starts.
AAA㨉ZZZZZZAZZZZZZ
This function judges whether the detection tube is clean when the washing process ends after manual/auto operation.
9CTPKPI㨍GNNVQQ&KTV[
ZZZZZZZZZZZZ
9CTPK㨘I=?
#FJGUKQPQHFKTV
Ɣ When warning message still remains on screen after the sensor cleaning (with brush), the fresh water may be
contaminated. Do not use industrial water or circulating reclaimed water in the fresh water line since such water may
be cloudy or contaminated.
%#76+10 Ɣ This oil concentration detector is equipped with a “Zero-point compensation” function. However never use this function
when “Warning: cell too dirty” or ”Warning [112]” (to alarm dirtiness of sample water detection tube) is displayed on
screen. If used, a detector trouble may occur.
During standby, a clock time is
Standby Esc Return to standby displayed as shown below.
Hold pressed for 2 seconds or more 5VCPFD[
AAAAZZZZZZAAZZZZZZ
Page. 15
Esc
(Wrong operation prevention)
Main Menu
6.䎃 Setting
Clock setting Various settings Testing function Maintenance Operational log display Firmware version
Clock Adjustment Setting Test Mode Maintenance Mode Operational Log FOCAS-2000
See Section 6-2. See Section 6-3. See Section 8-1. See Section 8-2. See Section 8-3. Firmware X.XX
Ɣ On the Clock Adjustment screen, you can set the clock of the oil concentration detector with the radio time signal.
5VCPFD[
AAAZZZZZZAAZZZZZZ
Keep on pressing "Esc" for two seconds or longer in the standby state.
/CKPA/GPWAAAAAAAAAAAAAAA Main menu for clock adjustment is displayed.
%NQEM㨉#㨎㨔WUVOGPVA㨉AAAAAA Press "Enter."
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉 A clock setting item is displayed.
'06AZZZZZZAAZZZZZZ Press "Enter."
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉 A clock setting starts.
'06AZZZZZZAAZZZZZZ Press "Enter."
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉 Press "Enter" to set a year. The cursor blinks at a digit you can enter a value.
'06AZZZZZZAAZZZZZZ Press "Ÿ" to select a value. Press "Enter" after selecting a value.
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉
'06AZZZZZZAAZZZZZZ
Press "Ÿ." The cursor moves to a month entry digit.
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉 Press "Enter" to set a month. The cursor blinks at a digit you can enter a value.
'06AZZZZZZAAZZZZZZ Press "Ÿ" to select a value. Press "Enter" after selecting a value.
Repeat the above to set days, hours, minutes, and seconds.
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉 Set seconds after setting minutes.
'06AZZZZZZAAZZZZZZ Press "Ÿ." The cursor moves to a second entry digit.
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉 Press "Enter" to set seconds. The cursor blinks at a digit you can enter a value.
'06AZZZZZZAAZZZZZZ Press "Ÿ" to select a value. Press "Enter" after selecting a value.
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉 Press "Ÿ." The cursor moves to the "ENT" position.
'06ZZZZZZAAZZZZZZ Press "Enter." to end the setting.
%NQEM㨉#㨎㨔WUVOGPV㨉㨉㨉㨉 A clock setting item is displayed.
'06AZZZZZZAAZZZZZZ Press "Esc.”
/CKP/GPW Main menu for clock adjustment is displayed back.
%NQEM#㨎㨔WUVOGPV Press "Esc"
5VCPFD[
Return to standby
ZZZZZZAZZZZZZ
Setting data
Value set
Value set at
at
Setting items Variable range
Variable range Step
Step factory
factory Units
Units
shipment
shipment
Warming-up time Warming up
Warming up 30䢢 to
30䎃 to 300
300 11 30
30 sec.
sec.
Oil alarming point 1 (alarm signal 1/AL1 lamp) Alarm 11
Oil Alarm 5䢢 to
5䎃 to 15
15 11 15
15 ppm
ppm
Oil alarming delay 1 (alarm signal 1/AL1 lamp) Alarm 11 Delay
Delay 0䢢 to
0䎃 to 55 11 00 sec.
sec.
Oil alarming point 2 (alarm signal 2/AL2 lamp) Alarm 22
Oil Alarm 5䢢 to
5䎃 to 15
15 11 15
15 ppm
ppm
Oil alarming delay 2䎃
2䢢 alarm
(alarmsignal
signal2/AL2
2/AL2lamp)
lamp) Alarm 22 Delay
Delay 0䢢 to
0䎃 to 600
600 10
10 00 sec.
sec.
Alarm suppression Alarm output
Alarm output::OFF
OFF// ON
ON
(during warm-up, washing operation, and setting)
setting) Suppress
Alarm Suppress ---
--- (Alarmsuppression)
(Alarm suppression) ---
---
Alarm suppression:
Alarm suppression:ON
ON
Washing timer (automatic washing time) Washing Time
Auto Washing Time 30䢢 to
30䎃 to 600
600 11 60
60 sec.
sec.
Washing timer (manual washing time) Washing Time
Man. Washing Time 10䢢 to
10䎃 to 600
600 11 50
50 sec.
sec.
During standby Stand by Current
Standby Current 0䢢 to
0䎃 to 44 11 00 mA
mA
During washing Washing Current
Current 0䎃
0䢢 to
to 44 or
or 20
20 to
to 24
24 11 44 mA
mA
Exceeding the oil alarming point Over Oil
Oil Alarm
Alarm Density
Density or
or 20
20 to
to 24
24 11 Densit
Densit mA
mA
RECout Exceeding the oil concentration of 30
30
ppm (measuring limit) Over range
range Current
Current 20䎃
20䢢 to
to 24
24 11 20
20 mA
mA
During setting Maint. Current
Current 00 or
or 20
20 to
to 24
24 11 00 mA
mA
With equipment fault Defect Current
Current 00 or
or 20
20 to
to 24
24 11 00 mA
mA
䢢䎃
Setting method
5VCPFD[
Standby䎃 Keep on pressing "Esc" for two seconds or longer in the standby state.
ZZZZZZAAZZZZZZ
/CKPA/GPWAAAAAAAAAAAAAAAA A clock adjustment item is displayed. Press "Ÿ."
Clock adjustment menu
%NQEM㨉#㨎㨔WUVOGPVAAAAAA㨉㨉A 䯴Return to standby press "Esc"䯵
/CKPA/GPWAAAAAAAAAAAAAAAA A setting menu item is displayed. Press "Enter."
Setting menu
5GVVKPIAAAAAAAAAAA㨉AAAAAA 䯴Return to standby press "Esc"䯵
Warm-up time setting 5GVVKP㨑㨉㨉㨉㨉AAAAAAAAAAAAAA Press "Enter."
screen 9CTOKPIAWRAAAAAAAAAZZZUGE 䚚䎃 This example sets a warm-up time.䎃
The cursor blinks at a digit you can enter a value.
5GVVKP㨑㨉㨉㨉㨉AAAAAAAAAAAAAA The cursor blinks at a value entry digit ("xx" in the example).
Changing a set value Select a value by using "ź" or "Ÿ" and press "Enter." The value is
9CTOKPIAWRAAAAAAAAAZZZUGE
set and changing ends.䎃
Returning to the setting 5GVVKP㨑㨉㨉㨉㨉AAAAAAAAAAAAAA
Press "Ÿ." The next setting item is displayed on screen.䎃
item screen 1KN#NCTOAAAAAAAZZRRO
Ɣ The above setting method in a black box is also applicable to the other setting items. (see the next page for the other setting items.)
Ɣ Press "Esc" to stop the setting operation halfway. The setting item screen is displayed back.
The new setting becomes valid after it is saved. Select the screen below and press "Enter" to save it.
%#76+10 5GVVKP㨑㨉㨉㨉㨉AAAAAAAAAAAAAA
5CXGARCTCORTGUUA'PVGT
…(see "Caution" on the next page.)
Page. 19
7. Operation and Maintenance
● The FOCAS-2000 oil concentration detector is a high precision instrument.
● It is recommended for you to periodically check and maintain FOCAS-2000 as shown below to assure its performance for a long time.
伪䢢 Requirements
Item Maintenance interval Remarks
Every 2 weeks or every 10 measurements For cleaning method, see "8-4. Cleaning the
Sensor cleaning (with brush)
whichever comes first. sensor (brush washing)."䢢
For checking method, see ”8-1-2. Zero-point
Zero-point check Every month
check.”
Filter cleaning Every 3 months For cleaning method, see "8-5. Cleaning the filter."
Replace desiccant when the indicator change in
color as below two types,
Became pink(normally from blue to light purple)
Desiccant (silica gel) check Every month Became red dot on white base(normally black
dot on white base)
For desiccant (silica gel) replacement,
see directions attached to desiccant (silica gel)
which is an optional parts
Replacement of the clock backup For battery replacement, see "8-7 Replacing the
Every 2 years
battery clock backup battery."
Call your local Fellow Kogyo distributor or Sales
Accuracy check Every 2 years
Division of Fellow Kogyo.
Every 1 year or when the sensor unit is Call your local Fellow Kogyo distributor or Sales
Performance check
replaced Division of Fellow Kogyo.
Call your local Fellow Kogyo distributor or Sales
Update of precision certificate Every 5 years
Division of Fellow Kogyo.
The left screen is displayed for about 5 seconds and the manual operation process starts.
The oil concentration in the operation process is displayed together with the ON/OFF state of the
three-way valve output signal.
Press "Esc" for 2 seconds. The manual operation process ends and the oil concentration
detector goes to washing process.
When the washing process ends, the oil concentration detector waits for judgment of cleanliness
of the detection tube. "Wait Stable" is displayed together with the remaining time (in seconds)
to the right of "Wait Stable."
A. Normal with the clean glass
→Manual running
The screen to select a manual
operation item of the testing
function is displayed back
B. Glass is stained
→Warning: cell too Dirty
After the judgment ends, the screen to select a manual operation item of the testing function is
displayed back.
Press "Esc."
Ɣ This may be caused by the dirty detection tube due to aging or contaminated water supplied from the fresh water inlet.
This function is used to check such a stain of the detection tube.
ƔPlease check once a month.
䦟䎃 Zero-point check by feeding mineral water from the upper lid portion (recommended)
5VCPFD[
Keep on pressing "Esc" for two seconds or longer.
AAZZZZZZZZZZZZ
㧷CKP/GPW
Press "Ÿ" twice.
%NQEM#FLWUVOGPV
㧷CKP/GPW Press "Enter.” The TEST mode is set..
Press "Ÿ.”
6'56/QFG
6'56/QFG
Press "Enter.” The zero-point check starts.
'ZCO<GTQRRORTGUU'PVGT
9CUJVJGKPPGTEGNN Open the upper lid and the drain valve and empty the sensor unit.
Close the drain valve, feed mineral water into the sensor unit, and open the drain valve once
CPFRTGUUGPVGT
more to empty the sensor unit. Press "Enter.”
䎃
(KNNOKPGTCNYCVGTKPVQ Repeat a set of feeding mineral water into the sensor unit and draining water from the sensor
unit twice or more. Close the drain valve, feed mineral water into the sensor unit, and close
VJGEGNNCPFRTGUUGPVGT
the upper lid. Press "Enter.”
'ZCO<GTQRRO'UESWKV Shows a time required until the supplied mineral water becomes stable in the detection tube
(until the oil concentration detector becomes ready to measure the oil concentration).
AAAAAAAAAAAAAAAAAAAAZZ
'ZCO<GTQRRO'UESWKV Measures the concentration of oil in the mineral water.
(You can terminate the measurement anywhere. However, if you do not press the "Esc"
/GCUWTGZZRRO
switch for 10 minutes, the oil concentration detector stops measurement.)
Keep on pressing "Esc" for two seconds or longer to stop measurement.
6'56/QFG
Press "Enter.”
'ZCO<GTQRRORTGUU'UE
㧷CKP㨉/GPW
"Main Menu" is displayed back.
6'56㨉/QFG
With this, the zero-point check by feeding mineral water from the upper lid portion is completed.
Page. 22
8.䎃 Maintenance and Checking䎃
䦟 8-1-3. Zero-point check 䯴MeasureWater press Enter䯵
Ɣ This function simply checks the oil concentration of water supplied from the fresh water inlet.
Ɣ Execute manual washing to cleaning the sensor before starting this check function (Zero-point check by feeding water
from the fresh water inlet).
Ɣ Do not feed industrial water or circulating reclaimed water to the fresh water inlet. When the fresh water is cloudy or
%#76+10 contaminated, the oil concentration may not be measured correctly when the sensor is cleaned with the contaminated
water.
Ɣ To check whether the measurement is influenced by the water supplied from the fresh water inlet, the zero-point check
function (using mineral water) is used after sensor cleaning.
Ɣ If “Warning: cell too dirty” is displayed, this zero-point check function (by using water fed through the fresh water inlet) is used to check whether
the fresh water is contaminated.
䦟䎃 Zero-point check procedure by feeding water through a fresh water line on a ship
5VCPFD[
Keep on pressing "Esc" for two seconds or longer.
ZZZZZZAAZZZZZZ
㧷CKP㨉/GPW
Press "Ÿ" twice.
%NQEM#FLWUVOGPV
㧷CKP㨉/GPW
Press "Enter" and then press "Ÿ" twice.
6'56㨉/QFG
6'56/QFG
Press "Enter.” The zero-point check starts.
/GCUWTG9CVGTRTGUU'PVGT
/GCUWTG9CVGT'UESWKV Open the fresh water valve to feed water to the fresh water line.
ZZ A time to flow water is displayed on the lower right part of the screen.
/GCUWTG9CVGTAAAA'UESWKV Shows a waiting time before the water fed into the fresh water line becomes stable.
ZZ A remaining time is displayed on the lower right part of the screen.
/GCUWTG9CVGT'UESWKV The oil concentration detector measures the concentration of oil in water fed from the fresh water
/GCUWTGZZARRO inlet. (You can terminate the measurement anywhere. However, if you do not press the “Esc”
switch for about 10 minutes, the oil concentration detector stops measurement.)
Keep on pressing "Esc" for two seconds or longer to end measurement.
6'56/QFG
Press "Esc.”
/GCUWTG9CVGTRTGUU'UE
㧷CKP㨉/GPW
"Main Menu" is displayed back.
6'56㨉/QFG
With this, the zero-point check by feeding water from the fresh water line on the ship is completed.
Page. 23
8.䎃 Maintenance and Checking䎃
䦟䎃 8-1-4.䎃 RECout output test 䯴Current Output xxmA䯵
Ɣ This test function is used to check the operation of a connected instrument and cable connections.
Ɣ This function can output a current of 0 to 24 mA at intervals of 1 mA.
5VCPFD[
Keep on pressing "Esc" for two seconds or longer.
ZZZZZZAAZZZZZZ
㧷CKP/GPW
Press "Ÿ" twice.
%NQEM#FLWUVOGPV
㧷CKP/GPW
"TEST Mode" is displayed. Press "Ÿ" three times.
6'56/QFGA
6'56/QFG
Press "Enter.” "Current output" menu is displayed.
%WTTGPV1WVRWVO#
6'56/QFG The cursor blinks at a value entry digit. Select a value by pressing "ź" or "Ÿ."
Press "Enter." The specified current is output.
%WTTGPV1WVRWVZZO#
Press "Esc.”
6'56/QFG
"Main Menu" is displayed back.
/CPWCNTWPPKPI
Upper row: Status of the sample water valve; XXX OFF: Closed; XXX ON: Open
Page 2 Lower row: Status of the freshwater valve; XXX OFF: Closed; XXX ON: Open
Page 3 1, 2, 3 and 4 Level values for each sensor; Temp Sensor temperature
䯲each sensor level and temperature value are checked while FOCAS-2000 is
measuring oil content in sample water or fresh water
<Reference>
If reset or power-off occurred after density measurement, following
figures are displayed. They are not abnormal.
1=0000, 2=0000, 3=0000
4=0000, Temp=00.0
䯲 In the Automatic operation state, the Alarm1 and Alarm2 operation tests are enabled when the oil concentrations are less than
the oil alarm points.
䯲 Information on Pages 1, 3, 5, and 6 will be required by your local Fellow Kogyo distributor or Sales Division of Fellow Kogyo when oil
concentration detector errors and problems are repaired. Please let us know the information when requested.
Page. 25
8.䎃 Maintenance and Checking䎃
䦟䎃 8-2. Maintenance䎃 (Maintenance Mode)
Ɣ The oil concentration detector has Manual Washing, Zero-Point Compensation, Operation Check, and Reset After Zero-Point Compensation
functions. (See "6. Setting (Outline) - Ŷ Tabulated list of menu items.")
Ɣ To use a testing function, keep on pressing "Esc" for 2 seconds or longer in the standby state.
(The testing function is not available while the oil concentration detector is running.)
Ɣ Select a maintenance function by operating "ź" or "Ÿ" in the "Maintenance Mode."
䦟䎃 8-2-4. Function for recovery after zero-point compensation (Recover press Enter)
Ɣ This function is used to reset the compensated value (compensated by "8-2-2 Zero-point compensation") to the initial value (not compensated).
5VCPFD[
Keep on pressing "Esc" for two seconds or longer
ZZZZZZZZZZZZ
㧷CKP/GPW
Press "Ÿ" three times.
%NQEM#FLWUVOGPV
㧷CKP/GPW "Maintenance Mode" is set.
/CKPVGPCPEG/QFG Press "Enter.”
/CKPVGPCPEG/QFG
Press "Ÿ" four times.
9CUJKPIRTGUU'PVGT
/CKPVGPCPEG/QFG Press "Enter.”
4GEQXGTRTGUUA'PVGT Resets the zero-point-compensated value to the previous value (before compensation).
/CKPVGPCPEG/QFG
Press "Esc.”
4GEQXGTRTGUU'UE
㧷CKP/GPW
"Main Menu" is displayed back.
/CKPVGPCPEG/QFG
With this, resetting to the uncompensated value is completed.
Page. 28
8.䎃 Maintenance and Checking䎃
䦟䎃 8-3.䎃 Viewing the operational log 䎃
Ɣ This function is used to read the operational log (that has recorded bilge alarm operations, warning and error operations) on the screen.
Ɣ This function can display the contents of the operation log wholly or in groups and roughly or in detail.
Ɣ You can view the content of the operation log only while standby. (You cannot view while operate.)
Ɣ "Dump Log" as a viewing item is an optional function. (For more information, call your local Fellow Kogyo distributor or Sales Division of Fellow Kogyo.)
Ɣ This function can also view the content of operational log memory card of FOCAS-1800 (with some limitations). (see 䚚1 of “section 6-1” )
䦟䎃8-3-1. Data groups of operational log
Ɣ To facilitate viewing of the content of the operational log, the recorded data can be classified into some functional groups. The groups are
explained below.
5VCPFD[
Keep on pressing "Esc" for two seconds or longer.
ZZZZZZZZZZZZ
㧷CKP/GPW
Press "Ÿ" four times.
%NQEM#FLWUVOGPV
㧷CKP/GPW The Operational Log function is set.. Press "Enter."
1RGTCVKQPCN.QI Press "Ÿ.”
The total number of log items is displayed on the lower right part of the screen. (This value is displayed on every screen below.)
1RGTCVKQPCN㨉.QI
Enables viewing of all contents of the operational log.
#NN5VCVWUZZZZ
Press "Ÿ.”
1RGTCVKQPCN.QI
Power ON/OFF record group. Ɣ Viewing method
2QYGTZZZ
Press "Ÿ.” 1. Select an item you
want to view and
1RGTCVKQPCN.QI Operation-related group. press "Enter."
1RGTCVKQPZZZ Groups related to Automatic operation, Manual opration, Washing, etc.
Press "Ÿ.” 2. You can get the
1RGTCVKQPCN.QI Alarm1- and Alarm2-related group. details from Pages
1KN#NCTOZZZ Group related to oil alarm points (preset oil concentration values, etc.) (Outline) below.
Press "Ÿ.” (See the next page.)
1RGTCVKQPCN.QI Error/alarm group.
3. By pressing "Esc,"
&GHGEVZZZ Group related to Error / Warning occurrence, etc.
Press "Ÿ.” "All Status" (to
1RGTCVKQPCN.QI Equipment error group. enable viewing all
5[UVGOZZZ Group related to sensor unit troubles, etc.
contents of the
operational log) is
Press "Ÿ.” displayed back on
1RGTCVKQPCN.QI Compensation, check, and calibration record group. screen.
%CNKDTCVKQPZZZ Group related to performance checks, zero-point compensation, etc.
Press "Ÿ.”
1RGTCVKQPCN㨉.QI TEST function group.
6'56ZZZ Group related to relays, alarms, current output check, etc.䎃
Press "Ÿ.”
1RGTCVKQPCN.QI Optional function.
&WOR.QIRTGUU'PVGT Outputs a simple operational log to the outside.
Press "Ÿ.”
1RGTCVKQPCN.QI Free memory space available.
#XCKNCDNG/COQT[ZZ The remaining free memory space is displayed on the lower right part of the screen.
When you press "Esc" here, "Main Menu" screen (see below) is displayed back.
When you press "Ÿ" here, "All Status" is displayed back to enable you to view all contents of the operational log.
㧷CKP/GPW
Press "Esc."
1RGTCVKQPCN.QI
5VCPFD[
Press "Esc.”
ZZZZZZZZZZZZ
With this, the explanation of each functional group item of the content of the operational log is completed.
Outlined and detailed displays of each group are explained below.
Page. 29
.䢢 Maintenance and Checking
8.䎃
伪䎃䢢 8-3-2. Outlined and detailed displays of the operational log䎃䢢
䦟
Ɣ
● Here we'll explain how to view details of the operational log from the outlined display and also explain their screens.
䯲The oil alarm point (Alarm Setting) record will be explained on the next page"8-3-3 Oil alarm
䚚 alarm point
point (Alarm
(Alarm Setting)
Setting) record
recorddisplay
display screen)"
since their screens are different. Regarding details of the log please see “12.Display of groups and description of statuses”.
Outlined display
1RGTCVKQPCNA.QI
This example shows how to view an Alarm1
#NNA5VCVWUAAAAAAAAAZZZZZ ON operational log group in the All Status
mode.
Item & sorting Total number of data elements written
The left screen is displayed after "Enter" is
pressed.
Press "Enter." The selected group item of "All Status" is displayed.
䎃
Display of a group item
Code No. Year / Month / Day Hours : Minutes : Seconds
Content of oil alarm point record You can also view old records by pressing
"ź" on this screen.
䦟䎃 Cleaning procedure
1. Prepare the dedicated washing liquid and washing brush.
2. Be sure to immerse the washing brush in water to soften it and remove excessive water from the brush.
3. Turn off power to the oil concentration detector, close the sample water valve and the fresh water inlet
valve, and remove the upper lid.
(Use an adjustable wrench or the like to open the upper lid if it is too hard to open.)
4. Open the drain port (in the sensor unit), drain the sample water detection tube, and close the drain port.
5. Feed the washing liquid to 80% of the detection tube.
6. Wash the inside of the detection tube by moving the washing brush up and down and turning the brush.
Wash the inside wall of the detection tube by turning the brush along the wall surface a shown right.
7. Open the drain port, drain water from the detection tube, and close the drain port.
Feed mineral water to 80% of the detection tube, open the drain port, drain water from the sample water
detection tube, check and make sure the drain seal washer is fitted to the drain, and close the drain port.
8. Turn on power to the oil concentration detector, manually operation the oil concentration detector by using
mineral water according to the manual running procedure (see "Ŷ 8-1-1 Manual running"), and make sure
"Warning: cell too dirty" is not displayed. When the warning is displayed, repeat manual washing, sensor Washing brush
cleaning, and manual running (see 8-1-1), and check whether the detection tube is clean.
9. Open the upper lid, check the state of the upper part of the detection tube and the O ring of the upper lid.
Ɣ Wash the inside of the detection tube by moving the
Wipe off stain or dust (if any) on the O ring with clean waste cloth. Firmly fasten the upper lid by your washing brush up and down and turning the brush.
hands. (Firmly fasten it by your hands. You need not use an adjustable wrench or the like.) 䎃
10. Open the sample water valve and the fresh water inlet valve (to the original positions).
11. Feed fresh water according to "Ŷ8-2-1 Manual washing (Washing press Enter)" and end the manual
washing.
䦟䎃 8-5.䎃 Cleaning the filter
Clean the filter when "Warning: cell too dirty" is frequently displayed when the oil concentration detector is running or at the end of running.
The filter should be cleaned about every 3 months. 䎃
Ɣ When the water supplied from the sample water inlet contains high concentration of oil or contaminants or when
%#76+10 industrial water or circulating reclaimed water is supplied from the fresh water inlet, the filter may be clogged with oil
film, contaminant, and/or dust and will lose its performance. If the filter is clogged too much, oil concentration detector
operation will be disabled.
Sample water filter body Fresh water filter body 䦟䎃 Filter cleaning procedure
1. Close thesample water inlet valve and the fresh water inlet
Sample water seal washer Fresh water seal washer
valve respectively to the OFF position to stop water supply.
Sample water filter 2. Rotate and remove the filters by using an adjustable wrench.
Fresh water filter
3. Clean each filter with a soft plastic brush and neutral
detergent solution.
4. Rinse each filter fully with water.
5. Remount the filters.
Page. 32
8.䢢 Maintenance and Checking䢢
伪䢢 8-6.䢢 Checkup and replacement timing of desiccant (silica gel)䢢
CAUTION ● Be sure to close the case cover and lock the snap latch before operation the oil concentration detector.
Inspect the color of desiccant silica gel in the indicator every month. When desiccant (silica gel) is
change in color, replace it by new one. There are two type depends on the delivered date.
● Turn off power supply and make sure all lamps are off.
CAUTION ● Do not reverse the polarities of the battery.
● Do not use any metallic screw driver to demount the battery. If done, the battery may be damaged.
Open ■Preparation
Top Prepare replacement fuses.
Fuse for oil concentration detector: φ5 - 20 mm, ceramic tube type
Close ䷵Fuse for detector 3.15A, 250V (1 pc.)
䢢 䢢 䢢 3.15A, 250V Fuse for three-way valve φ5 - 20 mm, glass tube type 2A, 250V (2 pcs.)
■ Demounting
1. Turn the fuse holder counterclockwise while pressing it with your finger.
䷶Fuse for 2. Take up the fuse holder.
Three-way valve 3. Pull out the fuse from the cap of the fuse holder.
䢢 䢢 䢢 2A, 250V
■ Remounting
1. Insert the fuse into the cap of the fuse holder.
2. Turn the fuse holder clockwise with the fuse on the cap.
3. Check and make sure the fuse holder cap is tight.
Page. 34
9.䎃 Troubleshooting
Ŷ Principal measures against failure
Symptom Possible cause Action / measures
No power supply Ɣ The oil concentration detector fuse was Ɣ Check the supply voltage (90 to 240 V).
blown out by an overvoltage on the oil Turn off power to the oil concentration detector and
concentration detector. replace the fuse by a new one.
Inactive three-way valve Ɣ The three-way valve fuse was blown out Ɣ Check the cable between the oil concentration detector
by an overload on the three-way valve and the three-way valve for short-circuit or damage.
terminals. Check the three-way valve for troubles. Replace the
fuse by a new one.
Asynchronous three-way valve and Ɣ An Alarm1 delay time has been set. Ɣ Check the setting data and ranges (see 6-2-1) and set
Alarm1 operations the Oil Alarm Delay 1 to "0 sec" (see "6-2-2. Setting
method").
䚚䎃 The Alarm1 output should work in synchronism with the
AL1 lamp.
䚚 You cannot set a delay for the three-way valve output
signal (VALVE).
Different Alarm1 and Alarm2 Ɣ Oil alarm points and delay time set for Ɣ Correct the set values according to "6-2-2. Setting
operations Alarm1 and Alarm are invalid. method." To synchronize Alarm1 and Alarm2 operations,
the set values for Oil Alarm Points 1 and 2 of Oil Alarm
Delays 1 and 2 should be identical.
No alarm operation Ɣ The connection with the alarm panel and Ɣ Check the cable connection. When the cable
cable connections may be wrong. connection is right, check cables for disconnection or
damage. Replace the cable if defective.
Ɣ The connected alarm panel may be Ɣ Inspect the alarm panel.
defective. Ɣ Check whether the terminal operation (output) of the oil
Ɣ The Alarm1/Alarm2 relay malfunctions. concentration detector is normal in the standby state
according to "7-1-6 Alarm Operation check."
Inactive measuring instrument Ɣ The connection of the measuring Ɣ Check the cable connection. When the cable
connected to the RECout terminal instrument may be wrong or the connection is right, check cables for disconnection or
connection cable may be defective. damage. Replace the cable if defective.
Ɣ The connected measuring instruments may Ɣ Inspect the connected measuring instruments.
bedefective. Ɣ Check whether the terminal operation (output) of the oil
Ɣ RECout output failure concentration detector is normal according to "7-1-4
RECout output check."
Inoperable switch Ɣ The operation switch malfunctions.䎃 Ɣ Check whether the switch button can be pressed.
The button may be locked by paint when the system is
installed and built. Remove paint if the button is locked
by paint and check whether the button is operable.䎃
䎃 Ɣ The oil concentration detector is frozen. Ɣ Press the RESET switch and make sure the clock ounts
up on the "Standby" screen. Then check whether the
button is operable.
No runnning Ɣ The oil concentration detector is frozen. Ɣ Press the RESET switch and make sure the clock
䎃 counts up on the "Standby" screen. When the
pressure of water supplied from the bilge pump to the
sample water inlet becomes 0.03 MPa or higher, check
whether the oil concentration detector runs.
䎃 Ɣ The pressure may be low or the pressure Ɣ When the pressure of water supplied from the bilge
switch is disabled. pump to the sample water inlet becomes 0.03 MPa or
higher, check whether the oil concentration detector runs.
No clock count (in seconds) on the Ɣ The oil concentration detector is frozen. Ɣ Press the RESET switch and make sure the clock
LCD display screen䎃 counts up on the "Standby" screen.
Ɣ When the oil concentration detector cannot be recovered by the above operations, call your local Fellow Kogyo distributor or Sales Division of
Fellow Kogyo.
In the maintenance works below, power to the oil concentration detector must be shut off to prevent electric shocks and
device failures.
Ɣ Demounting the operational log memory card, remounting the memory card, and replacing the clock backup battery.
(See "2-3-2 Terminal arrangement of terminal base and inside connection of FOCAS-2000.")
%#76+10 Ɣ Cleaning the sensor. (See "8-4 Cleaning the sensor (brush washing).")
Ɣ Cleaning the filter. (See "8-5 Cleaning the filter.")
Ɣ Replacing silica gel. (See directions attached to desiccant (silica gel) which is an optional parts.)
Ɣ Other works. (See the instruction sheets of the replacement parts if available.)
Page. 35
9.䢢 Troubleshooting
In the case "Warning" or "Error" is displayed on the LCD display screen
● To assure the performance, the oil concentration detector (FOCAS-2000) is so designed to stop running when detecting a warning or error
state. In such a state, the AL1 and AL2 alarm lamp blink with a warning or error code on the LCD display screen.
● Press the RESET switch or turn off and on power to the oil concentration detector when the warning or error state cannot be canceled, and
when an unknown symptom (excluding the LCD displays listed below) is displayed, and when the "Enter" switch is disabled to temporarily
cancel the warning state.
Symptom (LCD display) Possible cause Action / measures
● The insulating sheet is still on the ● Remove the insulating sheet from the battery holder
battery. and set the clock right.
● Power has not been supplied to the oil ● Replace the clock backup battery by a new one and set
concentration detector for a long time the clock right.
and the clock backup battery has been ● Set the clock right.
exhausted. See "4-3 Preparing for Operation --- 6.2 Time correction
● The clock is not right. of the clock."
● No operational log memory card ● Turn off power to the oil concentration detector, mount
the operational log memory card, and turn on power to
the oil concentration detector again.
● If the memory card which was used ● All you can do is to confirm log (contents of memory
in a certain FOCAS-2000 is inserted card).
in another FOCAS-2000 unit, density
measurement cannot be performed.
● Outside the measurement range of the ● Perform the zero-point check (see 8-1-2) and check
sensor unit whether the oil concentration detector can run normally.
● Inspect the bilge separator when the oil concentration
detector can run normally.
䯲䢢 When the error still occurs, turn off the oil concentration
䢢 detector and replace the sensor unit.
● Failure of the light source system in the ● Turn off power and clean the sensor (with brush).䢢
sensor of the sensor unit Turn on power again, and perform the zero-point check.
● Inspect thebilge separator when the oil concentration
● Light in the sensor is blocked due to detector can run normally.
stains in the supplied sample water. 䯲䢢 When the error still occurs, turn off the oil concentration
䢢 detector and replace the sensor unit.
● Condensation or water leak in the ● Check the color of the desiccant (silica gel).
sensor unit. When the desiccant (silica gel) indicator is pink, turn off
the detector and replace the sensor unit.
● The ambient temperature is outside the ● Use the oil concentration detector in the service
operating temperature of the sensor temperature range specified by "2-1. Product
unit. specification."
● The detection tube in the sensor unit is ● When the valve is in the pipeline before the sample
empty. water inlet or fresh water inlet, check whether the valve
is in the "Open" (to supply) position. When the error
still occurs although the valve is in the "Open" position,
clean the fresh water line filter and check once more.
● Contaminants ● The detection of the oil concentration detector is greatly
affected by the contaminants.
● Perform the zero-point check (see 8-1-2) and check
whether the oil concentration detector runs normally.
● Inspect the bilge pump when the oil concentration
detector runs normally.
● The detection tube in the sensor unit is ● Clean the filter (see "8-5 Cleaning the filter") and clean
contaminated. the detection tube (see “”8-4 Cleaning the sensor (brush
washing)).
● When the oil concentration detector cannot be recovered by the above operations, call your local Fellow Kogyo distributor or Sales Division of
Fellow Kogyo.
Page. 36
9.䎃 Troubleshooting
In the case "Warning" or "Error" is displayed on the LCD display screen
Symptom (LCD display) Possible cause Action / measures
Ɣ The operational log memory card is Ɣ Turn off power to the oil concentration detector, mount
9CTPKPI=? demounted while power is supplied to the memory card correctly, and turn on the oil
/GOQT[ECTFGLGEVGF the oil concentration detector. concentration detector again.
Ɣ The operational log memory card is not Ɣ Turn off power to the oil concentration detector, mount
mounted correctly. the memory card correctly, and turn on the oil
concentration detector again.
Ɣ Sensor unit transmission error. Ɣ Turn off power to the oil concentration detector and
'TTQT=? check the connection of Connector 1 (see "2-3 Internal
5GPUQTYCUPQTGURQPUG layout figure and terminal base"). If the error still
occurs, turn off power to the oil concentration detector
and replace the sensor unit.
Ɣ Failure in the display PC board or the Ɣ Turn off power to the oil concentration detector, mount
'TTQT=? operational log memory card. the operational log memory card correctly, and turn on
/GOQT[%QOO'TTQT power to the oil concentration detector again. If the
error still occurs, replace the operational log memory
card by a new one.
Ɣ The operational log memory card has no Ɣ Replace the operational log memory card by a new
9CTPKPI=? free space. card. When the operational log memory cards must
/GOQT[ECTFHWNN be replaced so often since measurement is repeated
so frequently or power is turned on and off so
frequently, check the bilge pump pressure source and
the power supply source.
Ɣ The connectors of the sensor unit are Ɣ Confirm the connection of Connector 2 according to
'TTQT=? disconnected. "2-3 Internal layout figure and terminal base."
%JGEMUGPUQTRQYGTECDNG
Ɣ The connectors of the sensor unit are Ɣ Confirm the connection of Connector 1 according to
'TTQT=? disconnected. "2-3 Internal layout figure and terminal base."
%JGEMUGPUQTUKIECDNG
Ɣ When the oil concentration detector cannot be recovered by the above operations, call your local Fellow Kogyo distributor or Sales Division of
Fellow Kogyo.
In the case the other data is displayed on the LCD display screen
Symptom (LCD display) Possible cause Action / measures
Ɣ Execute manual washing (see”8-2-1”) and clean the
Ɣ The detection tube is not clean.
9CTPKPI㨍GNNVQQ&KTV[ sensor (with brush (see “8-4”)).
A㨉AZZZZZZZZZZZZ Ɣ The water supplied at fresh water inlet is not Ɣ Perform the zero-point check (see”8-1-2”) and check
clean. whether the oil concentration detector can run
normally. Inspect the fresh water supplied at fresh
water inlet when the oil concentration detector can
run normally.
Ɣ When the zero-point compensation is Ɣ Execute manual washing (see”8-2-1”) and clean the
<GTQEQORA AAAA omitted. sensor(with brush (see “8-4”)).
(CKN\GTQEQORGPUCVKQP Ɣ When an oil concentration of 5 ppm or more
is detected in the zero-point compensation
process (due to the influence by oil or
contaminant left in the detection tube)
Ɣ When the zero-point compensation is Ɣ Execute manual washing (see”8-2-1”) and clean the
<GTQEQORA 㧛 omitted. sensor(with brush (see “8-4”)).
9CTPKPIEGNNVQQ&KTV[ Ɣ When an oil concentration over 5 ppm is
detected in the zero-point compensation
process (due to the influence by oil or
contaminant left in the detection tube)
Page. 37
䢢 7URXEOHVKRRWLQJ
Ŷ䢢 :KHQ\RXJHWWKHPHVVDJH³:DLWLQJIRU3UHVVXUH6:´VT
䢢 Ɣ䢢 ³:DLWLQJIRU3UHVVXUH6:´FDQDSSHDUZKHQ)2&$6ILQLVKHVRSHUDWLRQ±WKLVLV 刍7KHPHVVDJHVD\LQJWKDW³:DLWLQJIRU
127DPDOIXQFWLRQ 3UHVVXUH6:´刏
䢢 Ɣ䢢 7KHPHVVDJHDSSHDUVZKHQWKHFKDWWHUSUHYHQWLRQIXQFWLRQRSHUDWHVWRSUHYHQW
UHSHWLWLRQRIWKHVDPHEHKDYLRUDQGRUUDQGRPEHKDYLRU,WGRHVQRWPHDQWKDW
)2&$6LVZDLWLQJIRUWKH3UHVVXUH6ZLWFKWRWXUQRQDIWHUVDPSOHSUHVVXUHGURSV
䢢 Ɣ䢢 7KLVPHVVDJHDFWXDOO\PHDQV³7KH3UHVVXUH6ZLWFKUHPDLQV21´7KHPHVVDJHZLOO
EHGLVSOD\HGZKHQ :DLWLQJIRU3UHVVXUH6:
7KH)2&$6LVVKXWGRZQE\KROGLQJGRZQWKH(VFNH\GXULQJDXWRRSHUDWLRQ [[[[[[[[[[[[
7KHRLOGHIHFWDODUPKDVEHHQWULJJHUHGGXULQJDZDVKLQJRSHUDWLRQDIWHUDQDXWR
RSHUDWLRQWKHQWKHZDVKLQJLVFDQFHOHGZLWKWKHVDPSOHSUHVVXUHDW03DDQG
DERYH
䢢 Ɣ䢢 7KHPHVVDJHGLVDSSHDUVZKHQWKHVDPSOHSUHVVXUHGURSVEHORZ03D
Ŷ䢢 +RZWRGHDOZLWK³:DLWLQJIRU3UHVVXUH6:´
Ɣ䢢 ´:DLWLQJIRU3UHVVXUH6:´FDQRFFXULQVWDQGE\ZKHQDQDXWRRSHUDWLRQLVFDQFHOHGE\KROGLQJGRZQWKH(VFNH\
Ɣ6WRS2:6WRORZHUVDPSOHSUHVVXUHLQFDVH³:DLWLQJIRU3UHVVXUH6:´DSSHDUV7KHPHVVDJHGLVDSSHDUVZKHQWKHVDPSOH
SUHVVXUHGURSVEHORZ0SD䢢 7KHUHDUHWZRZD\VWRFOHDUWKHPHVVDJHDVIROORZV
2:6VWDUWVUXQQLQJ
ź
6DPSOHSUHVVXUHULVHVZLWKVDPSOHZDWHUVXSSO\ :KLOH2:6UXQQLQJ
ź ELOJHSXPSLVZRUNLQJ
)2&$6VWDUWVDXWRRSHUDWLRQ 6KXWGRZQ)2&$6
ź +ROGGRZQWKH(VFNH\
GXULQJZDVKLQJDIWHUDXWR
7KH(VFNH\LVKHOGGRZQGXULQJDXWRRSHUDWLRQ
RSHUDWLRQ
ź
7KHHQGRIDXWRRSHUDWLRQOHDGVWRZDVKLQJRSHUDWLRQ 7KHUHDUHWZRZD\VWRIL[
³:DLWLQJIRU3UHVVXUH
ź 6:´5HIHUWKHIORZ
)2&$6JRHVEDFNWRVWDQGE\DIWHUZDVKLQJRSHUDWLRQ FKDUWVEHORZ
ź
³:DLWLQJIRU3UHVVXUH6:´DSSHDUV
Fuse for the valve 2000-020 1 set 2A, 250V Upper lid O-ring 2000-060 1 pc 4DP-31
Electromagnetic valve (2 pcs)
unit
Washing brush 2000-030 1 pc Drain seal W 2000-070 1 pc M5
䦟䎃 Option parts
Rating/ Rating/
Part name Part No. ty. specification Part name Part No. ty. specification
Sensor unit 2000-210 1 pc Calibrated Upper lid 2000-260 1 pc
Electromagnetic valve unit 2000-220 1 pc Lithium battery 2000-270 1 pc CR2450
Behavior verification kit 2000-230 1 set Fresh water filter set 2000-280 1 set
Desiccant (silica gel) 2000-240 1 set Sample water filter set 2000-290 1 set
Operational log memory 2000-250 1 pc
card
● When effecting every-5-years renewal of precision certificate, you can replace calibrated sensor unit (sold separately䯂with Precision Cert.)
on board and use it..
● You can check the operation (e.g., PSC inspection) of the oil concentration detector by the Operation Test Kit (separately sold).
(The accuracy will not be assured.)
11.䎃 Request
The FOCAS-2000 oil concentration detector is a high precision instrument capable of detecting oil in parts per million. To prevent the product
against failures and maintain its accuracy, the product must be calibrated within 5 years by the manufacturer or the agent authorized by the
manufacturer and have the IOPP certificate renewal examination.
(It is permitted to replace your product by a calibrated and verified product.)
● For ensuring precision, it is recommended that the user have its equipment calibrated or verified for precision by the agent or
manufacturer every one to two years.
● Checkup / repair or precision verification after shipment must be performed in accordance with IMO regulations by a person who has
skills certified by Fellow Kogyo.
● Breaking the seal inside the equipment is deemed as a fraudulent act. Never break the seal.
● Do not access any portion except for portions specified in this operation manual. Access to unspecified portions is prohibited by law.
Contamination is minimized through keeping the detection tube clean by automatically performing washing and filling the detection portion
with fresh water after using the equipment.
● Ensure to install a fresh water line. Fresh water for use is recommended to be equivalent to mineral water.
● Care must be taken if the equipment is used in a place with the possibility of freezing. Drain water completely from the detective tube
by using the manual drain function if the equipment is stored in the environment of low temperatures or left with the detection tube
water-filled for a prolonged time.䎃
Page. 39 Page. 38
䢢
12. Display of groups and description of statuses.
FOCAS-2000䢢 Display of groups and description of statuses.
FOCAS-2000 䢢 儔兏兠儻⾲♧僎儝優兠儣儝僔ෆᐜ
Group Code No, Contents (at the time of Code No, Contents (at the time of
indication 儗兠儭 occurrence) 儗兠儭 cancel release)
儔兏兠儻⾲♧ ␒ྕ ෆᐜ凚Ⓨ⏕凛 ␒ྕ ෆᐜ凚ゎ㝖凛
All Status ALL record data is indicated.
グ㘓⾲♧ 傿僟僌僔グ㘓儫兠儣僸⾲♧
40 AC Power ON 41 AC Power OFF 㟁※僔 ON 僎 OFF
Power 42 External reset ---- ------------------- እ㒊兎償儧儬
㟁※㛵㐃 43 Stand by ---- ------------------- ᚅᶵ≧ែ
44 Power down detect ---- ------------------- ౪⤥㟁※పୗ᳨ฟ
50 Warming up ---- ------------------- 儊儍兠元兗儔儆儧儻
53 Enter Manual Operation 54 Exit Manual Operation ᡭື㐠㌿僔㛤ጞ僎⤊
Operation 55 Start Washing 56 Stop Washing Ὑί兆兠儭僔㛤ጞ僎⤊
㐠㌿≧ἣ 59 Manual stop ---- ------------------- 㐠㌿ືసṆ
5A Pressure S/W ON 5B Pressure S/W OFF ᅽຊ儝儈儧儥僔 ON凡OFF
5E Renew Parameter ---- ------------------- タᐃ⠊ᅖ僔ኚ᭦
66 伫伫/伫伫Alarm Setting ---- ------------------- AL.1凞2 ㆙ሗⅬタᐃ
Oil Alarm 60 Alarm 1 ON 61 Alarm 1 OFF AL. 1 ㆙ሗⅬタᐃ㉸傮
Ἔศ㆙ሗ 62 Alarm 2 ON 63 Alarm 2 OFF AL .2 ㆙ሗⅬタᐃ㉸傮
64 Oil over 30ppm 65 Oil under 30ppm Ἔศ 30ppm ㉸傮
0A Maintenance mode. 2A ret fm Maintenance mode 充兗優儮兗儝兆兠儭⛣⾜僎㏥ฟ
01 Over Range 21 ret fm Over Range ᐃ㝈⏺㉸傮僔᳨ฟ僎ゎ㝖
02 Sensor Down 22 ret fm Sensor Down 償兗儙ග※儤儊兗
03 Sensor Error 23 ret fm Sensor Error 償兗儙㒊僔⤖㟢兟Ỉἐ
04 Temperature Error 24 ret fm Temperature Error 償兗儙㒊僔 ᗘ儌免兠
07 No Water found 27 ret fm No Water found ᳨ฟ⟶ෆ傲✵
08 No Sample flow 28 ret fm No Sample flow ᳨ฟ⟶ෆヨᩱỈ僐傽
09 No Water flow 29 ret fm No Water flow ᳨ฟ⟶ෆΎỈ僐傽
11 Foreign material 31 ret fm Foreign Material ᳨ฟ⟶ෆ僔␗≀ΰධ
Defect 12 Adhesion of dirt ---- ------------------ ᳨ฟ⟶ෆ僔ở僲
ᶵჾ␗ᖖ 13 PCB Open ---- ------------------ ᇶᯈ㛤ᨺ儌免兠
(㐠㌿ 15 RTC adjusted ---- ------------------ ィㄪᩚ兟ㄪᩚ
儌免兠➼)
17 Sensor was no response ---- ------------------ 償兗儙兠兊儯儧儬ᛂ⟅僐傽
14 memory Card ejected 1A memory card inserted 充兆兎儏兠儭⬺╔
18 Memory Comm. Error ---- ------------------ 充兆兎儏兠儭㏻ಙ儌免兠
19 Memory card full ---- ------------------ 充兆兎儏兠儭✵傳ᐜ㔞僐傽
1B no format memory ---- ------------------ ᮍึᮇ充兆兎儏兠儭僔᳨ฟ
1C format memory card ---- ------------------ 充兆兎儏兠儭ึᮇฎ⨨
1D no memory card ---- ------------------ 充兆兎儏兠儭ᮍᤄධ㉳ື
1E Memory ID different ---- ------------------ 㐪催ᶵჾ␒ྕ僔充兆兎儏兠儭ᤄධ
1F FOCAS-1800 card found ---- ------------------ FOCAS1800 僔充兆兎儏兠儭ᤄධ
Page. 40
䢢
12. Display of groups and description of statuses.
Group Code No, Contents (at the time Code No, Contents (at the time of
indication 儗兠儭 of occurrence) 儗兠儭 cancel release)
儔兏兠儻⾲♧ ␒ྕ ෆᐜ凚Ⓨ⏕凛 ␒ྕ ෆᐜ凚ゎ㝖凛
70 Difference over 5ppm ---- ------------------ Zero Ⅼ⿵ṇ 5ppm ㉸傮
71 Zero adjust fail ---- ------------------ Zero Ⅼ⿵ṇᮍ
償兗儙兊儯儧儬㟁※免儈兗儕兠儺兏
74 Check sensor power cable ---- ------------------
␗ᖖ
償兗儙兊儯儧儬ಙྕ免儈兗儕兠儺兏
76 Check sensor sig. cable ---- ------------------
␗ᖖ
System
78 Sensor was Changed ---- ------------------ 償兗儙兊儯儧儬
ᵓᡂ
79 Ext. osc. Doesn’t work ---- ------------------ 償兗儙兊儯儧儬እ㒊Ⓨჾ僔Ṇ
᧯సഛ
7A Sensor has no calibrated ---- ------------------ 償兗儙兊儯儧儬ᮍᰯṇရ僔᥋⥆
7B Sensor has no serial ---- ------------------ 償兗儙兊儯儧儬儛兎儆兏↓傽
7C RTC lost voltage ---- ------------------ ィᮍタᐃ(㟁ụ僐傽)
7D RTC no ack. ---- ------------------ ィᶵ⬟㏻ಙ儌免兠
7E Software update ---- ------------------ 儡儹儬᭩傮᭦᪂
7F System Error ---- ------------------ 儛儝優兄充兆兎僔ㄞ㎸⬟
84 Zero adjust 85 Zero adjust done Zero Ⅼ⿵ṇ僔ᐇ⾜僎⤊
Calibration 88 Performance check 89 Performance check done ືస☜ㄆ僔ᐇ⾜僎⤊
⿵ṇ傎☜ㄆ 8A Accuracy check 8B Accuracy check done ⢭ᗘ☜ㄆ僔ᐇ⾜僎⤊
♫ෆᰯṇグ㘓 Zero ☜ㄆ僔ᐇ⾜僎⤊
8C Exam mineral water zero 8D Exam mineral water done
凚㟁☢ᘚ㛢凛
Zero ☜ㄆ僔ᐇ⾜僎⤊
8E Exam clean water zero 8F Exam clean water done
凚㟁☢ᘚ㛤凛
95 Recover (of zero offset) ---- ------------------ Zero ⿵ṇ್๓僑ᡠ傿
A0 Operate Alarm Relay ---- ------------------ AL.1凞2 兎児兠ືస
TEST A4 Output Manual Current ---- ------------------ 㟁ὶฟຊ☜ㄆ
ヨ㦂ືస
A5 Start simulation A6 Stop simulation ㆙ሗືసヨ㦂僔ᐇ⾜僎⤊
䯲Clarification of abbreviation
␎⛠ㄝ᫂
2A到29,31䢢 ret fm=return from
15,7C,7D RTC.= Real Time Clock, 䢢 ack .=acknowledge
8C到8F Exam=Examination
66 伫伫=5到15ppm
76 sig.=signal
79 Ext. osc. Doesn’t =External oscillator does not
Page. 41
13. Additional Information䢢
Sample water flows, the measurement starts. All record data (All Status) and VALVE terminal output are as follows.
凚factory default settings凛
䢢
VALVE
Terminal 䢢
Code No. Contents Description
output
(AC OUT凛
Automatic operation by detecting the
Start 䢢 OFF 䢢 䢢 5A Pressure S/W ON pressure of the sample water line over
0.03MPa or more.
䢢 OFF Stabilization wait time of measurement
䢢 䢢 50 Warming Up 30 sec.
is about 30 seconds.
䢢
The above is the record of the normal operation cycle.
兟If there is an equipment abnormality or oil alarm comes out, it will be added to the operation log.
兟The pressure switch ON is recorded only when the sample water pressure rises to 0.03 MPa or more
䢢 from the standby state.
兟During operation, the sample water pressure drops to less than 0.03 MPa, the pressure switch turns OFF,
䢢 and when the sample water pressure becomes 0.03 MPa or more within 90 seconds, the pressure switch
ON is not recorded and normal operation is continued. After that, when the sample water pressure drops
again to less than 0.03 MPa for about 90 seconds or more, cleaning is started without recording the
pressure switch OFF, but it is not abnormal.
兟While the warming-up or alarm operation starts from normal operation, if the sample water pressure fluctuates
䢢 and the pressure switch repeatedly turns OFF and ON, the pressure switch OFF may be recorded several
䢢 times, but it is not abnormal.
兟The measurement operation continues until the start of washing mode is recorded after pressure switch off.
兟When you look at the detailed display of the alarm setting of the operational log, VALVE OFF (V=OF is
䢢 displayed) is displayed, but it becomes VALVE ON immediately after recording the alarm setting.
䢢
䢢
Page. 42
13. Additional Information䢢
䢢 兟 “Calibration Overdue” is displayed in standby at the end of FOCAS-2000 operation when the expiration date of the
last calibration approaches.
䢢 兟 This message “Calibration Overdue” is to inform you that the next calibration is required.
䢢 䢢 It doesn’t stop FOCAS-2000 operation.
䢢 兟 When “Calibration Overdue” is displayed in standby mode and / or at the end of operation, press
䢢 䢢 the [Enter] key to turn it off.
䢢 兟 The expiration date of FOCAS-2000 calibration is the date written on its latest precision certificate.
䢢 䢢 Once the “Calibration Overdue” message appears, please check the calibration due date and contact
䢢 䢢 us on how to proceed.
䢢䢢䢢
䢢
䎹䏈䏕䏖䏌䏒䏑䎃䎘䎃䎝
䣘䣧䣴䣵䣫䣱䣰䢵䢼䣑䣥䣶䣱䣤䣧䣴䢢䢴䢲䢳䢷
䎃䎶䏈䏓䏗䏈䏐䏅䏈䏕䎃䎕䎓䎕䎔
䣊䢴䢹䢰䢳䢲䢰䢴䢷䢲䢲
䎕䎓䎕䎔䎑䎜䎑䎕䎘䎓䎓 䎋䢪 䎬䣋 䎌䢫