Spectrum Analysis Basics-Trang-5
Spectrum Analysis Basics-Trang-5
A preselector improves dynamic range if the signals in question have sufficient frequency
separation. The discussion of dynamic range in Chapter 6 assumed that both the
large and small signals were always present at the mixer and their amplitudes did not
change during the course of the measurement. But as we have seen, if signals are far
enough apart, a preselector allows one to reach the mixer while rejecting the others. For
example, if we were to test a microwave oscillator for harmonics, a preselector would
reject the fundamental when we tuned the analyzer to one of the harmonics.
Let’s look at the dynamic range of a second-harmonic test of a 3-GHz oscillator. Using
the example from Chapter 6, suppose that a –40-dBm signal at the mixer produces
a second harmonic product of –75 dBc. We also know, from our discussion, that for
every 1 dB the level of the fundamental changes at the mixer, measurement range also
changes by 1 dB. The second-harmonic distortion curve is shown in Figure 7-12. For
this example, we assume plenty of power from the oscillator and set the input attenuator
so that when we measure the oscillator fundamental, the level at the mixer is –10 dBm,
below the 1-dB compression point.
From the graph, we see that a –10-dBm signal at the mixer produces a second-
harmonic distortion component of –45 dBc. Now we tune the analyzer to the 6-GHz
second harmonic. If the preselector has 70-dB rejection, the fundamental at the mixer
has dropped to –80 dBm. Figure 7-12 indicates that for a signal of –80 dBm at the
mixer, the internally generated distortion is –115 dBc, meaning 115 dB below the
new fundamental level of –80 dBm. This puts the absolute level of the harmonic at
–195 dBm. So the difference between the fundamental we tuned to and the internally
generated second harmonic we tuned to is 185 dB!
–45
–50
–60
Internal distortion (dBc)
–70
–80
–90
–100
–110
–115
–120
–90 –80 –70 –60 –50 –40 –30 –20 –10 0
Mixed level (dBm)
and
Looking at these expressions, we see that the amplitude of the lower distortion
component (2ω 1 – ω2) varies as the square of V1 and linearly with V2. On the other side,
the amplitude of the upper distortion component (2ω 2 – ω 1) varies linearly with V1 and
as the square of V2. However, depending on the signal frequencies and separation, the
preselector may not attenuate the two fundamental tones equally.
3 dB
21 dB
27 dB
The discussion of dynamic range in Chapter 6 applies to the low-pass-filtered low band.
The only exceptions occur when a particular harmonic of a low-band signal falls within
the preselected range. For example, if we measure the second harmonic of a 2.5-GHz
fundamental, we get the benefit of the preselector when we tune to the 5-GHz harmonic.
First of all, the preselector has insertion loss, typically 6 to 8 dB. This loss comes
prior to the first stage of gain, so system sensitivity is degraded by the full loss. In
addition, when a preselector is connected directly to a mixer, the interaction of the
mismatch of the preselector with that of the input mixer can cause a degradation of
frequency response.
Some spectrum analyzer architectures eliminate the need for the matching pad or
isolator. As the electrical length between the preselector and mixer increases, the rate
of change of phase of the reflected and re-reflected signals becomes more rapid for
a given change in input frequency. The result is a more exaggerated ripple effect on
flatness. Architectures such as those used in PSA Series analyzers include the mixer
diodes as an integral part of the preselector/mixer assembly. In such an assembly, there
is minimal electrical length between the preselector and mixer. This architecture thus
removes the ripple effect on frequency response and improves sensitivity by eliminating
the matching pad or isolator.
Even aside from its interaction with the mixer, a preselector causes some degradation
of frequency response.
The preselector filter pass band is never perfectly flat, but rather exhibits a certain
amount of ripple. In most configurations, the tuning ramp for the preselector and local
oscillator come from the same source, but there is no feedback mechanism to ensure
the preselector exactly tracks the tuning of the analyzer. Another source of post-tuning
drift is the self-heating caused by current flowing in the preselector circuitry. The center
of the preselector pass band will depend on its temperature and temperature gradients,
which depend on the history of the preselector tuning. As a result, you obtain the best
flatness by centering the preselector at each signal. The centering function is typically
built into the spectrum analyzer firmware and is selected either by a front-panel key
in manual measurement applications or programmatically in automated test systems.
When activated, the centering function adjusts the preselector tuning DAC to center
the preselector pass band on the signal. The frequency response specification for most
microwave analyzers applies only after centering the preselector, and it is generally a
best practice to perform this function (to mitigate the effects of post-tuning drift) before
making amplitude measurements of microwave signals.
The external mixer’s IF output connects to the analyzer’s IF “in” port. The latest analyzers
have only one front-panel port, and this is possible because the LO frequency supplied
from the analyzer is between 3 and 14 GHz, while the IF output frequency from the
external mixer to the analyzer is 322.5 MHz. Because of the wide frequency difference
between the LO and IF signals, both signals can exist on the same coaxial interconnect
cable that attaches the analyzer and the mixer. As long as the external mixer uses the
same IF as the spectrum analyzer, the signal can be processed and displayed internally,
just like any signal that came from the internal first mixer. Figure 7-14 illustrates the
block diagram of an external mixer used in conjunction with a spectrum analyzer.
Low
band path Analog or
3.6 GHz 5.1225 GHz 322.5 MHz 22.5 MHz digital IF
Analyzer
input
LO
High 3.8 to 8.7 GHz
band
path To external
mixer
4.8 GHz 300 MHz
322.5 MHz
Preselector
Sweep generator
Display
Band Keysight 11970 Series Keysight M1970 Series Other manufacturer’s Other manufacturer’s
mixers (LO range mixers (LO range mixers (LO range mixers (LO range
3–7 GHz) 6–14 GHz) 3–7 GHz) 6–14 GHz)
A (26.5 to 40.0 GHz) 6− and 8−
Q (33.0 to 50.0 GHz) 8− and 10−
U (40.0 to 60.0 GHz) 10−
V (50.0 to 75.0 GHz) 12− and 14− 6−
E (60.0 to 90.0 GHz) N.A. 6− and 8−
W (75.0 to 110.0 GHz) 18− 8−
F (90.0 to 140.0 GHz) 16− 10−
D (110.0 to 170.0 GHz) 20− 14−
G (140.0 to 220.0 GHz) 26− 18−
Y (170.0 to 260.0 GHz ) 30− 20−
J (220.0 to 325.0 GHz) 38− 24−
(325.0 to 500.0 GHz) 58− 36−
(500.0 to 750.0 GHz) 86− 54−
(750.0 to 1,100.0 GHz) 80−
Table 7-1. Harmonic mixing modes used by X-Series analyzers with external mixers
Whether you perform harmonic mixing with an internal or an external mixer, the issues
are similar. The LO and its harmonics mix not only with the desired input signal, but also
with any other signal, including out-of-band signals, that may be present at the input.
This produces mixing products that can be processed through the IF just like any other
valid signals.
A tunable filter that performs preselection of the signals reaching the first mixer in
the internal signal path is common in most signal analyzers. External mixers that are
unpreselected will produce unwanted responses on screen that are not true signals.
A way to deal with these unwanted signals has been designed into the signal analyzer.
This function is called “signal identification.”
Signal identification
It is quite possible that the particular response we have tuned onto the display has been
generated on an LO harmonic or mixing mode other than the one for which the display
is calibrated. So our analyzer must have some way to tell us whether or not the display
is calibrated for the signal response in question. For this example, assume we are using
a Keysight M1970V 50- to 75-GHz unpreselected mixer, which uses the 6− mixing mode.
The full V-Band measurement can be seen in Figure 7-15.
Keysight X-Series signal analyzers offer two different identification methods: image shift
and image suppress. Let’s first explore the image shift method. Looking at Figure 7-15,
let’s assume we have tuned the analyzer to a frequency of 50 GHz. The 6th harmonic of
the LO produces a pair of responses, where the 6− mixing product appears on screen
at the correct frequency of 50 GHz, while the 6+ mixing product produces a response
with an indicated frequency of 49.355 GHz, which is 2 times fIF below the real response.
The X-Series analyzer has an IF frequency of 322.5 MHz, so the pair of responses is
separated by 645 MHz.
75
70
Signal frequency (GHz)
Input Image RF 6-
RF 6+ RF 8- RF 8+ 6+
65
6-
62.500
61.855
60
55
10.36
10.47
50
7 7.5 8 8.5 9 9.5 10 10.5 11
LO frequency (GHz)
Figure 7-16 Harmonic tuning lines for the X-Series analyzers using the M1970 Series mixers
If we are tuned to a real signal, its corresponding pair will now appear at the same
position on screen that the real signal occupied in the first sweep. If we are tuned to
another multiple pair created by some other incorrect harmonic, the signal will appear to
shift in frequency on the display. The X-Series signal analyzer shifts the LO on alternate
sweeps, creating the two displays show in Figures 7-17a and 7-17b. In Figure 7-17a,
the real signal (the 6− mixing product) is tuned to the center of the screen. Figure 7-17b
shows how the image shift function moves the corresponding pair (the 6+ mixing
product) to the center of the screen.
Figures 7-17a and 7-17b display alternate sweeps taken with the image shift function.
Figure 7-18. The image suppress function displays only real signals
To make an accurate amplitude measurement, it is important that you first enter the
calibration data for your external mixer. This data is normally supplied by the mixer
manufacturer, and it is typically presented as a table of mixer conversion loss, in
dB, at a number of frequency points across the band. This data is entered into a
correction table on the signal analyzer, and the analyzer uses this data to compensate
for the mixer conversion loss. If you are using the M1970 Series harmonic mixers,
the mixer conversion loss is automatically transferred from the mixer memory to the
X-Series signal analyzer memory, which eliminates manual entry into a correction file.
The spectrum analyzer reference level is now calibrated for signals at the input to the
external mixer.
CCDF measurements provide statistical information showing the percent of time the
instantaneous power of the signal exceeds the average power by a certain number
of dB. This information is important in power amplifier design, for example, where
it is important to handle instantaneous signal peaks with minimum distortion while
minimizing cost, weight and power consumption of the device.
RF designers are often concerned with the noise figure of their devices, as noise figure
directly affects the sensitivity of receivers and other systems. Some signal analyzers,
such as the X-Series, have optional noise figure measurement capabilities available.
This option provides control for the noise source needed to drive the input of the device
under test (DUT) as well as firmware to automate the measurement process and display
the results. Figure 8-2 shows a typical measurement result, with DUT noise figure (upper
trace) and gain (lower trace) displayed as a function of frequency.
Today’s digital modulation techniques use amplitude and phase information to carry
more baseband data in limited spectrum and time. Therefore, it is essential for a signal
analyzer to handle amplitude and phase in an appropriate manner. QPSK (Quadrature
Phase-Shift Keying) is one of the simple examples of digital modulation techniques,
with which two bits of digital data are transmitted at once, or two bits per symbol.
Figure 8-4 shows an example of QPSK demodulation analysis with 89601B option AYA.
Remember you need four (2 2 ) states to transmit 2 bits at once.
As an easy and intuitive way to understand what’s going on the digital radio transmission,
use an I/Q plane, which is a two dimensional chart comprising in-phase and quadrature
components of the demodulated signal on the horizontal axis and the vertical axis,
respectively. An example of the chart is again shown on the top left window of Figure
8-4. The yellow trace called trajectory shows a vector combining phase and amplitude
moves around as time goes while red points indicates the instantaneous position of
trajectory at the time of decision when a receiver actually judges the symbol value.
Essentially, for digital radios, vectors at these decision points are most important for
modulation quality. As you can see on the bottom left window of Figure 8-4, a “scalar”
analyzer meaning traditional spectrum analyzer may be able to show the modulated
signal in frequency domain so that you can see whether the signal is properly modulated
in power wise to some extent, and you can also make sure that there is no unwanted
emission or leakage power to the adjacent channels. You need, however, some sort of
“vector” analyzer to perform meaningful analysis of modulation quality for digital data
transmission where phase information is involved.
Figure 8-5. Modulation analysis of WLAN 802.11ac signal using Keysight 89600 VSA software
A newer and much more complicated system is 802.11ac, which uses 256QAM
(quadrature-amplitude modulation). See Figure 8-5. The maximum power is limited,
so the data points are much closer in both phase and magnitude than for QPSK.
The analyzer you use to evaluate the transmitted signal must be sufficiently accurate
that it does not lead you to a false conclusion about the quality of the transmission.
Pure amplitude measurements are also required to determine signal attributes such
as flatness, adjacent-channel power levels and distortion.
These measurements are available on the Keysight X-Series signal analyzers with
appropriate options.
• LTE/LTE-Advanced
• WLAN
• Multi-standard radio (MSR)
• GSM/EDGE
• W-CDMA
• HSDPA
• cdma2000
• 1xEV-DO
• 1xEV-DV
• cdmaOne
• NADC and PDC
• TD-SCDMA
Figure 8-7 . Even when you use fast sweeps and max hold over a period of many
seconds, the swept spectrum analyzer view of the radar signal is not very informative
Very often, we may want to save measurement data as a file, either in the spectrum
analyzer’s internal memory or on a USB mass-storage device. There are several
different kinds of data we can save this way:
An image of the display: Preferably in a popular file format, such as bitmap, GIF, PNG
or Windows metafile.
Trace data: Saved as X-Y data pairs representing frequency and amplitude points on
the screen. The number of data pairs can vary. Modern spectrum analyzers such as
the X-Series allow you to select the desired display resolution by setting a minimum of
1 up to a maximum of 40,001 frequency sweep points (buckets) on the screen. This
data format is well suited for transfer to a spreadsheet program on a computer.
Today, automated test and measurement equipment has become the norm, and nearly
all modern spectrum analyzers come with a variety of standard interfaces, including
LAN, USB 2.0 and GPIB. LAN connectivity is the most commonly used interface, as
it can provide high data-transfer rates over long distances and integrates easily into
networked environments such as a factory floor. Other standard interfaces used widely
in the computer industry are likely to become available on spectrum analyzers in the
future to simplify connectivity between instruments and computers.
Keysight’s X-Series signal analyzers literally have computer firmware running USB
ports and a Windows operating system. These features greatly simplify control and
data transfer. In addition, the X-Series analyzers can be operated remotely, and the
analyzer’s display appears on the remote computer. Details are beyond the scope of
this application note; see the operating manual for your particular analyzer.
Firmware updates
Modern spectrum analyzers have much more software inside them than do instruments
from just a few years ago. As new features are added to the software and defects repaired,
it becomes highly desirable to update the spectrum analyzer’s firmware to take advantage
of the improved performance.
The latest revisions of spectrum and signal analyzer firmware can be found on the
Keysight Technologies website. You can download this firmware to a file on your local
computer. A common method to transfer new firmware into the spectrum analyzer
is to copy the firmware onto a USB drive and then insert it into one of the spectrum
analyzer’s USB ports. Some models, such as the X-Series, allow you to transfer the new
firmware directly into the spectrum analyzer using the instrument’s Ethernet LAN port.
It is a good practice to periodically check your spectrum analyzer model’s Web page
to see if updated firmware is available.
Modern spectrum analyzers usually have a service menu available. In this area, you can
perform useful diagnostic functions, such as a test of the front-panel keys. You also can
display more details of the alignment process, as well as a list of all optional hardware
and measurement personalities installed in the instrument. When you upgrade a spectrum
analyzer with a new measurement personality, Keysight provides a unique license key
tied to the serial number of the instrument. You install this license key through the USB
port or enter it on the front-panel keypad to activate the measurement capabilities of
the personality.
Glossary of Terms
Absolute amplitude accuracy: The uncertainty of an amplitude measurement in
absolute terms, either volts or power. Includes relative uncertainties (see Relative
amplitude accuracy) plus calibrator uncertainty. For improved accuracy, some spectrum
analyzers have frequency response specified relative to the calibrator as well as relative
to the midpoint between peak-to-peak extremes.
ACPR: Adjacent channel power ratio is a measure of how much signal energy from
one communication channel spills over or leaks into an adjacent channel. This is an
important metric in digital communication components and systems, as too much
leakage will cause interference on adjacent channels. It is sometimes also described
as ACLR, or adjacent channel leakage ratio.
Amplitude reference signal: A signal of precise frequency and amplitude that the
analyzer uses for self-calibration.
Analog display: A display technology where analog signal information (from the envelope
detector) is written directly to an instrument’s display, typically implemented on a cathode
ray tube (CRT). Analog displays were once the standard method of displaying information
on spectrum analyzers. However, modern spectrum analyzers no longer use this
technology; instead, they now use digital displays.
Blocking capacitor: A filter that keeps unwanted low-frequency signals (including DC)
from damaging circuitry. A blocking capacitor limits the lowest frequency that can be
measured accurately.
Delta marker: A mode in which a fixed, reference marker has been established and a
second, active marker is available that we can place anywhere on the displayed trace.
A read-out indicates the relative frequency separation and amplitude difference between
the reference marker and the active marker.
Digital display: A display technology where digitized trace information, stored in memory,
is displayed on an instrument’s screen. The displayed trace is a series of points designed
to present a continuous-looking trace. While the default number of display points
varies between different models, most modern spectrum analyzers allow the user
to choose the desired resolution by controlling the number of points displayed. The
display is refreshed (rewritten from data in memory) at a flicker-free rate; the data in
memory is updated at the sweep rate. Nearly all modern spectrum analyzers have
digital flat-panel LCD displays, rather than CRT-based analog displays that were used
in earlier analyzers.
Display detector mode: The manner in which the signal information is processed prior
to being displayed on screen. See Neg peak, Pos peak, Normal, Average and Sample.
Digital IF: An architecture found in modern spectrum analyzers in which the signal is
digitized soon after it has been downconverted from an RF frequency to an intermediate
frequency (IF). At that point, all further signal processing is done using digital signal
processing (DSP) techniques.
Display range: The calibrated range of the display for the particular display mode and
scale factor. See Linear and Log display and Scale factor.
Displayed average noise level: The noise level as seen on the analyzer’s display
after setting the video bandwidth narrow enough to reduce the peak-to-peak noise
fluctuations such that the displayed noise is essentially seen as a straight line. Usually
refers to the analyzer’s own internally generated noise as a measure of sensitivity and
is typically specified in dBm under conditions of minimum resolution bandwidth and
minimum input attenuation.
Drift: The very slow (relative to sweep time) change of signal position on the display
as a result of a change in LO frequency versus sweep voltage. The primary sources
of drift are the temperature stability and aging rate of the frequency reference in the
spectrum analyzer.
Dynamic range: The ratio, in dB, between the largest and smallest signals simultaneously
present at the spectrum analyzer input that can be measured to a given degree of accuracy.
Dynamic range generally refers to measurement of distortion or intermodulation products.
Envelope detector: A circuit element whose output follows the envelope, but not the
instantaneous variation, of its input signal. In a superheterodyne spectrum analyzer, the
input to the envelope detector comes from the final IF, and the output is a video signal.
When we put our analyzer in zero span, the envelope detector demodulates the input
signal, and we can observe the modulating signal as a function of time on the display.
Frequency accuracy: The degree of uncertainty with which the frequency of a signal
or spectral component is indicated, either in an absolute sense or relative to some
other signal or spectral component. Absolute and relative frequency accuracies are
specified independently.
Frequency span: The frequency range represented by the horizontal axis of the display.
Generally, frequency span is given as the total span across the full display. Some earlier
analyzers indicate frequency span (scan width) on a per-division basis.
Frequency stability: A general phrase that covers both short- and long-term LO
instability. The sweep ramp that tunes the LO also determines where a signal should
appear on the display. Any long term variation in LO frequency (drift) with respect to
the sweep ramp causes a signal to slowly shift its horizontal position on the display.
Shorter-term LO instability can appear as random FM or phase noise on an otherwise
stable signal.
NOTE: On some earlier spectrum analyzers, full span referred to a sub-range. For
example, with the Keysight 8566B, a microwave spectrum analyzer that used a
mechanical switch to switch between the low and preselected ranges, full span referred
to either the low, non-preselected range or the high, preselected range.
Gain compression: That signal level at the input mixer of a spectrum analyzer at which
the displayed amplitude of the signal is a specified number of dB too low due just to
mixer saturation. The signal level is generally specified for 1-dB compression and is
usually between +3 and –10 dBm, depending on the model of spectrum analyzer.
GSM: The global system for mobile communication is a widely used digital standard for
mobile communication. It is a TDMA-based system in which multiple communication
streams are interleaved in time, enabling them to share a common frequency channel.
IF gain/IF attenuation: Adjusts the vertical position of signals on the display without
affecting the signal level at the input mixer. When changed, the value of the reference
level is changed accordingly.
IF feedthrough: A raising of the baseline trace on the display due to an input signal at
the intermediate frequency passing through the input mixer. Generally, this is a potential
problem only on non-preselected spectrum analyzers. The entire trace is raised because
the signal is always at the IF; mixing with the LO is not required.
Image frequencies: Two or more real signals present at the spectrum analyzer input
that produce an IF response at the same LO frequency. Because the mixing products all
occur at the same LO and IF frequencies, it is impossible to distinguish between them.
Image response: A displayed signal that is actually twice the IF away from the frequency
indicated by the spectrum analyzer. For each harmonic of the LO, there is an image pair,
one below and one above the LO frequency by the IF. Images usually appear only on
non-preselected spectrum analyzers.
Input attenuator: A step attenuator between the input connector and first mixer of
a spectrum analyzer. Also called the RF attenuator. The input attenuator is used to
adjust level of the signal incident upon the first mixer. The attenuator is used to prevent
gain compression due to high-level or broadband signals and to set dynamic range by
controlling the degree of internally generated distortion. In some analyzers, the vertical
position of displayed signals is changed when the input attenuator setting is changed,
so the reference level is also changed accordingly. In modern Keysight analyzers, the
IF gain is changed to compensate for input attenuator changes, so signals remain
stationary on the display, and the reference level is not changed.
Input impedance: The terminating impedance that the analyzer presents to the signal
source. The nominal impedance for RF and microwave analyzers is usually 50 ohms. For
some systems, e.g., cable TV, 75 ohms is standard. The degree of mismatch between
the nominal and actual input impedance is given in terms of VSWR (voltage standing
wave ratio).
Linear display: The display mode in which vertical deflection on the display is
directly proportional to the voltage of the input signal. The bottom line of the graticule
represents 0 V, and the top line, the reference level, some nonzero value that depends
upon the particular spectrum analyzer. On most modern analyzers, we select the
reference level, and the scale factor becomes the reference level value divided by the
number of graticule divisions. Although the display is linear, modern analyzers allow
reference level and marker values to be indicated in dBm, dBmV, dBuV, and in some
cases, watts as well as volts.
LO emission or feedout: The emergence of the LO signal from the input of a spectrum
analyzer. The level can be greater than 0 dBm on non-preselected spectrum analyzers
but is usually less than –70 dBm on preselected analyzers.
Marker: A visible indicator we can place anywhere along the displayed signal trace. A
read out indicates the absolute value of both the frequency and amplitude of the trace
at the marked point. The amplitude value is given in the currently selected units. Also
see Delta marker and Noise marker.
Measurement range: The ratio, expressed in dB, of the maximum signal level that
can be measured (usually the maximum safe input level) to the displayed average noise
level (DANL). This ratio is almost always much greater than can be realized in a single
measurement. See Dynamic range.
Negative peak: The display detection mode in which each displayed point indicates
the minimum value of the video signal for that part of the frequency span or time interval
represented by the point.
Noise figure: The ratio, usually expressed in dB, of the signal-to-noise ratio at the input
of a device (mixer, amplifier) to the signal-to-noise ratio at the output of the device.
Noise power bandwidth: A fictitious filter that would pass the same noise power as
the analyzer’s actual filter, making comparisons of noise measurements among different
analyzers possible.
Noise sidebands: Modulation sidebands that indicate the short-term instability of the
LO (primarily the first LO) system of a spectrum analyzer. The modulating signal is noise,
in the LO circuit itself or in the LO stabilizing circuit, and the sidebands comprise a noise
spectrum. The mixing process transfers any LO instability to the mixing products, so
the noise sidebands appear on any spectral component displayed on the analyzer far
enough above the broadband noise floor. Because the sidebands are noise, their level
relative to a spectral component is a function of resolution bandwidth. Noise sidebands
are typically specified in terms of dBc/Hz (amplitude in a 1-Hz bandwidth relative to the
carrier) at a given offset from the carrier, the carrier being a spectral component viewed
on the display.
Positive peak: The display detection mode in which each displayed point indicates the
maximum value of the video signal for that part of the frequency span or time interval
represented by the point.
Preselector: A tunable bandpass filter that precedes the input mixer of a spectrum
analyzer and tracks the appropriate mixing mode. Preselectors are typically used only
above 2 GHz. They essentially eliminate multiple and image responses and, for certain
signal conditions, improve dynamic range.
Real-time spectrum analyzer: A method of signal analysis in which all signal samples
are processed for some sort of measurement result or triggering operation. There are no
gaps between time acquisitions while nonreal-time operations leave gaps.
Reference level: The calibrated vertical position on the display used as a reference
for amplitude measurements. The reference level position is normally the top line of
the graticule.
Resolution bandwidth: The width of the resolution bandwidth (IF) filter of a spectrum
analyzer at some level below the minimum insertion loss point (maximum deflection
point on the display). For Keysight analyzers, the 3-dB bandwidth is specified; for some
others, it is the 6-dB bandwidth.
Rosenfell: The display detection mode in which the value displayed at each point is
based upon whether or not the video signal both rose and fell during the frequency
or time interval represented by the point. If the video signal only rose or only fell, the
maximum value is displayed. If the video signal did both rise and fall, then the maximum
value during the interval is displayed by odd-numbered points, the minimum value,
by even-numbered points. To prevent the loss of a signal that occurs only in an even-
numbered interval, the maximum value during this interval is preserved, and in the next
(odd-numbered) interval, the displayed value is the greater of either the value carried
over or the maximum that occurs in the current interval.
Scale factor: The per-division calibration of the vertical axis of the display.
Sensitivity: The level of the smallest sinusoid that can be observed on a spectrum
analyzer, usually under optimized conditions of minimum resolution bandwidth, 0-dB
RF input attenuation and minimum video bandwidth. Keysight defines sensitivity as
the displayed average noise level. A sinusoid at that level will appear to be about 2 dB
above the noise.
Signal analyzer: A spectrum analyzer that also uses digital signal processing to
perform other more complex measurements such as vector signal analysis.
Span accuracy: The uncertainty of the indicated frequency separation of any two
signals on the display.
Spectrum analyzer: A device that effectively performs a Fourier transform and displays
the individual spectral components (sine waves) that constitute a time-domain signal.
Phase may or may not be preserved, depending upon the analyzer type and design.
Time gating: A method of controlling the frequency sweep of the spectrum analyzer
based on the characteristics of the signal being measured. It is often useful when
analyzing pulsed or burst modulated signals’ time-multiplexed signals and intermittent
signals.
Units: Dimensions of the measured quantities. Units usually refer to amplitude quantities
because they can be changed. In modern spectrum analyzers, available units are dBm
(dB relative to 1 milliwatt dissipated in the nominal input impedance of the analyzer),
dBmV (dB relative to 1 millivolt), dBuV (dB relative to 1 microvolt), volts, and in some
analyzers, watts. In Keysight analyzers, we can specify any units in both log and
linear displays.
Vector diagram: A display type commonly used when analyzing digitally modulated
signals. It is similar to a constellation display, except that in addition to the detected
symbol points, the instantaneous power levels during state transitions are also plotted
on an IQ graph.
Vector display: A display type used in earlier spectrum analyzer designs, in which the
electron beam was directed so that the image (trace, graticule, annotation) was written
directly on the CRT face, not created from a series of dots as in the raster displays
commonly used today.
Video: In a spectrum analyzer, a term describing the output of the envelope detector.
The frequency range extends from 0 Hz to a frequency typically well beyond the widest
resolution bandwidth available in the analyzer. However, the ultimate bandwidth of the
video chain is determined by the setting of the video filter.
Video amplifier: A post-detection, DC-coupled amplifier that drives the vertical deflection
plates of the CRT. See Video bandwidth and Video filter.
Video bandwidth: The cutoff frequency (3-dB point) of an adjustable low-pass filter
in the video circuit. When the video bandwidth is equal to or less than the resolution
bandwidth, the video circuit cannot fully respond to the more rapid fluctuations of the
output of the envelope detector. The result is a smoothing of the trace, i.e., a reduction
in the peak-to-peak excursion of broadband signals such as noise and pulsed RF when
viewed in the broadband mode. The degree of averaging or smoothing is a function of
the ratio of the video bandwidth to the resolution bandwidth.
Video filter: A post-detection, low-pass filter that determines the bandwidth of the
video amplifier. Used to average or smooth a trace. See Video bandwidth.
Zero span: That case in which a spectrum analyzer’s LO remains fixed at a given
frequency so the analyzer becomes a fixed-tuned receiver. The bandwidth of the
receiver is that of the resolution (IF) bandwidth. Signal amplitude variations are displayed
as a function of time. To avoid any loss of signal information, the resolution bandwidth
must be as wide as the signal bandwidth. To avoid any smoothing, the video bandwidth
must be set wider than the resolution bandwidth.