I 700 Specs
I 700 Specs
Product Data
No. MPDUS0212EAC
Prism Edition
INTRODUCTION
This system achieves high sensitivity and high quality by
using state-of-the-art digital technology in the T/R section,
which is the core of a diagnostic ultrasound system.
This system features advanced algorithms that run on cir-
cuits that incorporate the latest circuit technology, semicon-
ductor technology, and surface-mount technology.
This system is designed to support a full range of applica-
tions and can be used as a general purpose system or a spe- 2,1 a 2,9
cialized system, depending on the installed software.
Ergonomics Operability
The system employs a non-interlace high-definition monitor System operability is optimized for the overall clinical work-
with excellent view ability. This feature decreases operator flow in hospitals.
fatigue in long examinations.
The ergonomic design of the system ensures comfortable
and efficient examinations for operators, physicians, and
patients.
Prism Edition
23-inch wide LCD monitor, DVD/CD drive, Precision Imaging, D-THI, ApliPure+, Tissue Specific Optimization, Trapezoid Scan, Quick
TUS-AI800 Scan, Advanced Dynamic Flow (ADF), DICOM, Smart 3D, Software full keyboard, Vascularity Index, BEAM, Superb micro-vascular
Main unit
Aplio i800 Imaging (SMI), Intelligent Dynamic Micro Slice, AppLocker® Security Management, Probe connector holder, and electric lifting of
the operation panel are included.
23-inch wide LCD monitor, DVD/CD drive, Precision Imaging, D-THI, ApliPure+, Tissue Specific Optimization, Trapezoid Scan, Quick
TUS-AI700
Aplio i700
Scan, Advanced Dynamic Flow (ADF), DICOM, Smart 3D, Software full keyboard, Vascularity Index, BEAM, Superb micro-vascular
Imaging (SMI), Intelligent Dynamic Micro Slice, AppLocker® Security Management, connector holder, and electric lifting of the
8
operation panel are included.
USUR-AI900A Hardware and software kit to display reference signals (ECG waveforms etc.).
Reference Signal kit Software licence to display reference signals (ECG waveforms etc.). Op. Op. Op. Op.
USUR-AI900A/EL
*Only for limited region
Reference signal sensor unit UJUR-AI902A PCG and Pulse sensor Op. Op. Op. Op.
USSE-AI900A,
Stress Echo kit Adds the cardiac stress examination function to the system. Op. Op. Op. Op.
USSE-AI900A/EL
USWT-AI900A,
2D Wall Motion Tracking kit Adds the cardiac wall motion analysis function to the system. Op. Op. Op. Op.
USWT-AI900A/EL
USSD-AI900A,
SI-DI kit This kit adds new color mapping based on Diastolic phase Strain in 2D WMT mode. Op. Op. Op. Op.
USSD-AI900A/EL
USWT-AI901A,
3D Wall Motion Tracking kit Adds 3D cardiac wall motion analysis function to the system. Op.*3 Op.*3 Op.*3 Op.
USWT-AI901A/EL
2D Wall Motion Tracking USWT-AI904A, This kit enables cardiac wall motion analysis function for fetal heart using data from
Op. Op. Op. Op.
Fetal kit USWT-AI904A/EL a convex transducer. USWT-AI900A is required.
2D Wall Motion Tracking USWT-AI906A, Corresponds to analyze the entire four-chamber of LV, LA, RV and RA. USWT-AI900A
Op. Op. Op. Op.
Advanced kit USWT-AI906A/EL is required. (*9)
USQT-AI900A,
Quad Chamber Tracking kit Quad display for 3D Wall Motion Tracking results. Op.*3 Op.*3 Op.*3 Op.
USQT-AI900A/EL
2
MPDUS0212EAC
Main unit
This kit enables mitral valve analysis in 4D images acquired using the PEI-512VX
USMA-AI901A,
MVA 4D kit transducer. Op.*4 Op.*4 Op.*4 Op.
USMA-AI901/EL
*USMA-AI900A is required.
USVM-AI800A, This kit allows real-time 4D imaging using the PEI-512VX, PSI-50VX and
Volume Matrix kit Op. Op. Op. Std.
USVM-AI800A/EL PSI-30VX. It is included in the standard configuration of the TUS-AI900.
USVM-AI700A,
Volume Matrix 2 kit This kit enables real-time 3D imaging using PSI-40VX. Op. Op. Std. Std.
USVM-AI700A/EL
USAV-AI900A, This kit enables analysis of the Aortic Valve in 3D images acquired using the
AVA kit Op.*6 *10 NA NA Op.
USAV-AI900A/EL PEI-512VX.
Fetal Heart MPI USFH-AI600A, This kit enables Myocardial Performance Index (MPI). The MPI value can be
Op. Op. Op. Op.
Measurement kit USFH-AI600A /EL calculated from the time change curve in TDI.
USFP-AI900A, This kit enables the automatic generation of standard fetal heart views from a
Smart Fetal Heart kit Op.*7 Op.*7 Op.*7 Op.*7
USFP-AI900A/EL 4 chamber volume data set.
Pencil Connector unit UIPC-AI900A This unit is used to add connectors for pencil transducers. Op. Op. Op. Op.
TEE hanger kit UAEH-AI901A TEE hanger for the PET-508MA. Op. Op. Op. Op.
This kit consists of preset data suitable for CV examinations, a CV sticker, and a
CV kit UACV-AI900A Op. Op. Op. Op.
startup screen.
This kit is used to add the STC knob on the operating panel.
UIST-AI900A Op. Op. Op. Op.
Use in combination with UIUB-AI900 is not possible. For V5.1 or before.
STC kit
This kit is used to add the STC knob on the operating panel.
UIST-AI901A Op. Op. Op. Op.
Use in combination with UIUB-AI900 is not possible. For V6.0 or later.
The ultrasound scanning condition export function exports the scan conditions
Ultrasound scanning
USVL-AI900A, (currently selected image mode etc.) of the diagnostic ultrasound system to an
condition export function Op.*4 Op.*4 Op.*4 Op.
USVL-AI900A/EL external device.
Kit Applicable transducer: PET-512VX. (*9)
USHI-AI900A,
CHI kit Adds the Contrast Imaging function to the system. Op. Op. Op. Op.
USHI-AI900A/EL
USCQ-AI900A, Adds the Time Curve Analysis (TCA) function to the system. The USHI-AI900A is
CHI-Q kit Op. Op. Op. Op.
USCQ-AI900A/EL required.
USCQ-AI901A, Function for calculating characteristic value parameters by Fitting carve. The USCQ-
Fitting Curve kit Op. Op. Op. Op.
USCQ-AI901A/EL AI900A is required.
3
Prism Edition
Main unit
USHI-AI901A,
High frame rate CHI kit This kit enables Contrast Harmonic Imaging with high frame rate. Op. Op. Op. Op.
USHI-AI901A/EL
This kit enables the visualization and analysis of the Direction/Velocity of bubbles
USVF-AI900A,
Contrast Vector Imaging kit by tracking individual contrast bubbles and displays such parameters with different Op. Op. Op. Op.
USVF-AI900A/EL
colors.
USMV-AI900A, This unit is required for using the 4D transducer or the motor-driven TEE transducer. Pre-
4D kit Op. Op. Op.
USMV-AI900A/EL Pre-installed in WHC model. installed
Image processing technology that makes 3D/4D images of fetuses and anatomical
USLM-AI900A, Pre-
Luminance kit structures appear more realistic. The USMV-AI900A or USVM-AI800A or Smart Sensor Op. Op. Op.
USLM-AI900A/EL installed
3D kit is needed. Pre-installed in WHC model.
This kit allows 3D display for displaying the internal walls of hollow organs and
USFT-AI900A,
Fly Thru. kit structures from endoluminal viewpoints as viewed using an endoscope. Op. Op. Op. Op.
USFT-AI900A/EL
USMV-AI900A or USVM-AI800A or Smart Sensor 3D kit is needed.
Used for calculation of the volume by extracting the contours for regions with
Auto Volume Measurement USOB-AI900A, lower brightness in the 3D Volume image acquired in 4D mode. Enables volume
Op. Op. Op. Op.
kit USOB-AI900A/EL measurement, eg, antral follicle count. USMV-AI900A is required.
USMV-AI900A or USVM-AI800A or Smart Sensor 3D kit is needed.
USVS-AI900A,
4D ADF/SMI kit This kit enables imaging in ADF and SMI modes using abdominal VX transducer. Op. Op. Op. Op.
USVS-AI900A/EL
USVC-AI900A, This kit enables imaging in CHI mode using abdominal VX transducers.
4D CHI kit Op. Op. Op. Op.
USVC-AI900A/EL USHI-AI900A is required.
USPF-AI900A, This kit enables output of volume data for Smart 3D, Mecha4D, Smart Sensor 3D,
3D printer format export kit Op. Op. Op. Op.
USPF-AI900A/EL and Volume Matrix 3D to 3D printers.
USEL-AI901A, This kit enables Elastography (with FLR measurement) with linear and convex
14 Elastography-FLR kit
USEL-AI901A/EL transducers. (*1)
Op. Op. Op. Op.
USEL-AI900A,
Elastography kit The kit enables Elastography without FLR measurement. (*2) Op. Op. Op. Op.
USEL-AI900A/EL
USSW-AI900A, This kit allows tissue stiffness to be visualized by generating images that show shear
14 Shear Wave kit
USSW-AI900A/EL wave propagation.
Op. Op. Op. Op.
Upper limit and range are expanded from 200 kPa to 700 kPa to provide measure-
USSW-AI901A,
Shear Wave Hard kit ments of stiffer targets with the PLI-1205BX. Op. Op. Op. Op.
USSW-AI901A/EL
USSW-900A is required.
USQS-AI800A,
Quad view for SWE kit This kit allows display of shear wave images in four-frame display. Op. Op. Std. Std.
USQS-AI800A/EL
4
MPDUS0212EAC
Main unit
USDM-AI900A, This kit enables visualization of dispersion between frequencies for propagation
Dispersion Map for SWE kit NA NA Op. Op.
USDM-AI900A/EL speed for Shear Wave.
USFN-AI900A, CT/MRI/US volume data is loaded, and a CT/MRI/US planar image and an ultrasound
Smart Fusion kit Op. Op. Op. Op.
USFN-AI900A/EL image at the same position are displayed together. The UIFR-AI900A is required.
This kit allows display of a guideline in the image by simulating the pathway of the
USSN-AI900A,
Smart Navigation kit needle and the position of the needle tip based on positional information acquired Op. Op. Op. Op.
USSN-AI900A/EL
using the magnetic sensor.
This kit is used to generate the magnetic field for acquiring positional information
for transducers and needles in Smart Fusion, Smart Navigation, Smart Sensor 3D
Magnetic Generator kit UIFR-AI900A Op. Op. Op. Op.
and Smart Body Mark modes. Sensor securing adopters and magnetic sensors are
provided (one of each for the PVT-375BT, PVT-375SC, PVI-475BT and the PVI-475BX).
This magnetic sensor is added for performing Smart Fusion, Smart Navigation, and
Sensor kit for Fusion unit UIFR-A501A Op. Op. Op. Op.
Smart Sensor 3D using multiple transducers. The UIFR-AI900A is required.
Used for the Smart Fusion function. Performs automatic position matching between
USFN-AI901A,
Auto Registration kit the ultrasound volume data acquired in a previous exam and the real-time 2D Op. Op. Op. Op.
USFN-AI901A/EL
ultrasound image.
This pole cart allows the magnetic field generator included in the system main unit
Fusion Pole Cart UZWT-A500A Op. Op. Op. Op.
to be positioned independently.
UAFS-001A For PVT-382BT / PVT-482BT / PVI-482BX. (*5) Op. Op. Op. Op.
UAFS-008A For PLI-1205BX / PLI-2004BX / PLI-705BX / PLI-3003BX / PVI-574BX / PVT-574BT. (*5) Op. Op. Op. Op.
UAFS-009A For PSI-30BX / PSI-30VX / PSI-40VX / PSI-50VX / PST-28BT / PST-65BT. (*5) Op. Op. Op. Op.
5
Prism Edition
Main unit
This kit enables that the information in the DICOM data from the digital
USMB-AI900A,
Breast Scan Guide kit Mammography MLO/CC images to be used to create an overlay position on the Op. Op. Op. Op.
USMB-AI900A/EL
Ultrasound body mark in reference mode.
USSB-AI900A, It automatically traces and displays the anatomical position of transducer mark
Smart Body Mark kit Op. Op. Op. Op.
USSB-AI900A/EL based on the transducer position using magnetic sensor. UIFR-AI900A is required.
USMP-AI900A,
MicroPure kit This kit enables MicroPure, which supports visualization of small structures. Op. Op. Op. Op.
USMP-AI900A/EL
USPV-AI900A, B/W images can be obtained with a wider field of view by moving the transducer in
Panoramic View kit Op. Op. Op. Op.
USPV-AI900A/EL the lateral direction.
Multi-Reflection Canceller USRC-AI900A, This function is used to visualize images with fewer artifacts by cancelling multiple
Op. Op. Op. Op.
kit USRC-AI900A/EL reflections from the body.
USLP-AI900A
Liver Package kit Combination kit of USSW-AI900A, USDM-AI900A, and USAT-AI900A. NA NA Op. Op.
USLP-AI900A/EL
USLD-AI900A,
Doppler Luminance kit This kit allows display of the pseudo color doppler in three dimensions. Op. Op. Op. Op.
USLD-AI900A/EL
USGI-AI900A,
General Imaging kit This kit allows usage of General Imaging. NA. Op. NA NA
USGI-AI900A/EL
USSL-AI900A,
Slice thickness control kit This kit enables the slice thickness of the cross-sectional image to be changed. Op. Op. Op. Op.
USSL-AI900A/EL
USZS-AI900A,
Measurement Z score kit This kit enables Z-score analysis for the measurement results. Op. Op. Op. Op.
USZS-AI900A/EL
This kit consists of preset data suitable for OB examinations, an OB sticker, and a
OB kit UAOB-AI900A Op. NA Op. Op.
startup screen.
Limb Volume Measurement USTM-AI900A, This kit enables the calculation of Estimated Fetal Weight (EFW) from a Fetal thigh
Op. Op. Op. Op.
kit USTM-AI900A/EL volume data set acquired by mechanical 4D transducer. USMV-AI900A is required.
UZRI-AI900A Rack for mounting a B/W printer (For UP-D898MD/P95D) Op. Op. Op. Op.
UZRI-AI901A Rack for mounting a B/W printer (For UP-D711/AC or P95DW-DC) Op. Op. Op. Op.
Mounting kit for peripheral UZRI-AI902A Mounting kit for Color printer/DVD recorder. Compatible with UIFR-AI900A. Op. Op. Op. Op.
unit
Rack for mounting a B/W printer (For UP-D711/AC or P95DW-DC)
UZRI-AI903A Op. Op. Op. Op.
Compatible with UIFR-AI900A.
UZRI-AI904A Mounting rack for B/W printer or Color printer or DVD recorder. Op. Op. Op. Op.
6
MPDUS0212EAC
Main unit
28,1Foot switch UZFS-004A Switch used for freezing, printing, and some other operations by foot. Op. Op. Op. Op.
35,2,2Gel warmer UZGW-008A This unit warms the ultrasound gel to a suitable temperature. Op. Op. Op. Op.
Transducer cable hanger kit UZMK-AI900A Long hanger on which the transducer cable is hooked. Op. Op. Op. Op.
A basket to store transducer connectors is added to the side of the system main
Transducer Holder kit UZBK-AI900A Op. Op. Op. Op.
unit. (one basket is provided in the standard configuration.)
This kit used to establish connection to the DICOM network via wireless LAN.
UIWL-A500A Complied with the Radio Law of Japan and applicable laws and regulations of USA, Op. Op. Op. Op.
Canada, EU member states, Iceland, Norway, Liechtenstein, and Switzerland.
Wireless LAN kit
This kit enables connection to the DICOM network via wireless LAN. (For Singapore,
UIWL-AI900A Op. Op. Op. Op.
Malaysia, Saudi Arabia, Russia, Australia, and Turkey.)
EV/ER Transducer Holder kit UZPH-AI900A This kit is used to add a dedicated holder for endocavitary transducers. Op. Op. Op. Op.
This kit is used to add a hardware full keyboard for entering the patient ID and
Keyboard kit UIKB-AI900A Op. Op. Op. Op.
comments.
This kit adds a tablet terminal that communicates with the system main unit via
wireless LAN to allow viewing and easy operation of the examination screen.
2nd Console kit UITB-AI900A Complied with the Radio Law of Japan and applicable laws and regulations of USA, Op. Op. Op. Op.
Canada, EU member states, Iceland, Norway, Liechtenstein, and Switzerland. UIWL-
A500A is included.
OLED Monitor UIOM-001A 21.6-inch wide OLED Monitor to replace 23-inch LCD Monitor with LED back light. Op. Op. Op. Op.
Function for setting up the patient database in the USB HDD in order to perform
DataBase for External HDD USDB-AI900A,
examinations. Op. Op. Op. Op.
kit USDB-AI900A/EL
*UZHI-AI900A is required.
Mounting kit for External Box with lock for installing the external HDD. HDD itself is not included.
UZHI-AI900A Op. Op. Op. Op.
HDD *USDB-AI900A is required.
ECG Cable Hanger kit UZMK-AI902A Hook used to hang the ECG cables on the front of the operating panel. Op. Op. Op. Op.
Pointing device, dial on track ball allows easier Gain adjustment without reaching
UZPT-001A out for another buttons. Applicalbe with the system which was shipped as V5.0 or Op. Op. Op. Op.
V5.1. For white panel.
Palm Controller kit
Pointing device, dial on track ball allows easier Gain adjustment without reaching
UZPT-002A out for another buttons. Applicalbe with the system which was shipped as V6.0 or Op. Op. Op. Op.
later. For black panel.
USHE-AI900A,
Online Help kit Kit for displaying the operation manual on the viewing monitor. Op. Op. Op. Op.
USHE-AI900A/EL
7
Prism Edition
Main unit
USDL-AI900A, This kit enables a combined reportfor the following liver applications:
Multi Parametric Report kit Op. Op. Op. Op.
USDL-AI900A/EL Shear Wave Elastography, Dispersion Imaging and Attenuation Imaging.
This kit provides a system label, and a startup screen suitable for dedicated Woman’s
Women's Healthcare kit UAWH-AI900A Op. NA Op. Op.
Healthcare systems.
Security Management kit USSM-AI900A This kit provides McAfee® software for security management of the system. Op. Op. Op. Op.
USRA-AI900A,
RADS kit This kit enables the Reporting and Data System. Op. Op. Op. Op.
USRA-AI900A/EL
USIO-AI900A,
IOTA kit This kit enables International Ovarian Tumor Analysis. (*8) Op. Op. Op. Op.
USIO-AI900A/EL
USNA-AI900A, This kit enables large capacity RAW data management with network attached stor-
Network Storage kit Op. Op. Op. Op.
USNA-AI900A/EL age (NAS) which is commercially vailable. (NAS itself is not included)
USTR-AI900A,
Tricefy Access kit This kit enables to access the Tricefy which is a cloud service for clinical images. Op. Op. Op. Op.
USTR-AI900A/EL
USUH-AI700A,
Ultra High Freq kit This kit enables use of 22-MHz, 24-MHz, 33-MHz transducers. Op. Op. Std. Std.
USUH-AI700A/EL
UZKF-AI900A
(French)
UZKG-AI900A
(German)
UZKI-AI900A
(Italian)
UZKS-AI900A
(Spanish) This kit is intended for changing the key tops of the full keyboard to support specific
Local Language Key-Top kit Op. Op. Op. Op.
UZKD-AI900A languages. UIKB-AI900A is required.
(Danish)
UZKN-AI900A
(Norwegian)
UZKW-AI900A
(Scandinavian)
UZKR-AI900A
(Russian)
8
MPDUS0212EAC
CW kit
USCW-AI900A
CV kit
UACV-AI900A
TEE Hanger
UAEH-AI900A, UAEH-AI901A
Foot Switch
UZFS-004A
A
9
Prism Edition
MVA 4D kit
USMA-AI901A 3D printer format export kit
USPF-AI900A
Note: In case of i800 and i700, USVM-AI800A, is required.
AVA kit
USAV-AI900A
Shadow Glass kit
USSG-AI900A
4D CHI kit
USVC-AI900A Note: USHI-AI900A is required.
10
MPDUS0212EAC
B C
Quad view for SWE kit Mounting kit Mounting kit Mounting kit Mounting kit
USQS-AI800A *Standard on i900 and i800. for a B/W printer for a B/W printer for Peripherals for Peripherals
(except UP-D711MD, (only for UP-D711 (for video recorder, (for B/W recorder,
*In case of USLP-AI900A is
P95D-DC) MD/AC, P95D-DC) color printer etc.) color printer etc.)
Dispersion Map for SWE kit installed, this kit is not UZRI-AI900A UZRI-AI901A, UZRI-AI903A UZRI-AI902A UZRI-AI904A
USDM-AI900A necessary.
*Not available on i700.
C D
11
Prism Edition
OLED Monitor
UIOM-001A
Gel warmer
UZGW-008A
Keyboard kit
UIKB-AI900A
Battery unit
UEBT-AI900A
12
MPDUS0212EAC
RADS kit
USRA-AI900A
IOTA kit
USIO-AI900A
*Only available in Europe
General Purpose
Electromagnetic Sensor Fly Thru. kit
610-1066 USFT-AI900A
General Purpose
Electromagnetic Sensor 3D printer format export kit
610-1066 USPF-AI900A
Virtu TRAX
610-1059
General Purpose
Electromagnetic Sensor
610-1066
13
Prism Edition
14
MPDUS0212EAC
15
Prism Edition
16
MPDUS0212EAC
17
Prism Edition
19
Prism Edition
· When PWD sampling volume is displayed, the Doppler – Phonocardiogram (PCG) (UJUR-AI902A is required.)
gate position, Doppler steering angle, and Doppler · PCG microphone: Acceleration type
angle are adjusted automatically. · Filter: Switching is possible.
• Slice Thickness Control (USSL-AI900A is required.) – Pulse (UJUR-AI902A is required.)
The ultrasound beam slice thickness can be adjusted. Pulse transducer: Air conduction type
– Respiration
Color Doppler M mode (MDF Mode) Impedance method using the reference signal cable
• Display mode • Heart Rate
– MCDI mode The heart mark blinks in synchronization with the heart
· Velocity display beat detected by the ECG.
· Velocity/variance display The heart rate is displayed.
· Power display • Reference Signal Sweep Speed
– M-TDI mode This changes the reference signal sweep speed.
• M Color Doppler Map (CDI MAP)
Color Doppler map can be selected for each mode. Other Diagnostic Function
• M Color Doppler Velocity Range Selection (C Scale) • CHI (Contrast Harmonic Imaging) (USHI-AI900A is required.)
The velocity range can be selected. The second-harmonic wave signals from the microbubbles
• M Color Doppler Baseline (C Baseline) in the contrast medium can be effectively visualized.
– The zero-velocity line on the M Color Doppler image can – The following image modes can be selected.
be shifted. · PS (Pulse Subtraction) -Low, PS-Low2
– The baseline shift setting can also be adjusted when · VRI (Vascular Recognition Imaging)
images that were frozen are displayed or when the · Fundamental
image in the image memory is played back. · CHI ADF
• Color Reverse Display · SMI (Superb Micro Vascular Imaging)
– The colors can be reversed. – The following functions can be selected.
– Changes can also be made when the image is frozen. · 2D TwinView
• Black and White/Color Balance · Micro Flow Imaging (MFI)
– Color weighting for B/W images can be set by compar- · Image Stabilizer
ing the M Color Doppler images and B/W images. · MI Constant function
– Changes can also be made when the image is frozen. · Quad display (USQC-AI800A is required for i700.)
• Color Gain • Parametric MFI
The display brightness of the M Color Doppler image can – Temporal information can be displayed as a color map
be changed. superimposed on images acquired by 2D mode (without
• M Color Doppler Multi-Frequency CHI starting up), ADF/SMI mode, and CHI mode (contrast
Doppler transmission frequency can be selected in M image for the period from the start of contrast medium
Color Doppler image acquisition. injection to the time when the contrast medium reaches
• M Color Doppler Filter the target region).
M Color Doppler low-cut filter can be changed. – The following functions are available.
· Micro Flow Imaging (MFI)
Reference Signals · Image Stabilizer
(USUR-AI900A and UJUR-AI900A, or USUR-AI900A and UJUR- • CVI (Contrast Vector Imaging)
AI901A are required.) (USVF-AI900A is required.)
• Type – This function allows color mapping of the flow of ultra-
– Electrocardiogram (ECG) sound contrast medium (bubbles) within the tissues for
· Lead I is the standard connection. contrast images acquired in CHI mode.
· External input is possible. – Color mapping for the following items is available.
· DC IN · Arrival time: Speed display (mm/s)
The connected device must comply with IEC 60601-1. · Direction: Direction display (deg)
· Top/bottom inversion is possible. · Arrival Time: Arrival time from the start posi-
· Lead switching tion for analysis (sec)
· Pacemaker – Measurement function for velocity and direction
· INST
20
MPDUS0212EAC
• 4D Magic Cut
Three-dimensional image data (volume data) can be gen- 4D ADF/SMI (USVS-AI900A is required.)
erated and displayed by using image data acquired for 4D CHI (USVC-AI900A and USHI-AI900A are required.)
three-dimensional image reconstruction. 4D Biopsy
– Mechanical 4D (USMV-AI900A is required.) FlyThru (USFT-AI900A is required.)
· The following functions can be used. Luminance (USLM-AI901A is required.)
Volume Color Shadow Glass (USLM-AI901A is required.)
Multi View Omni View
Volume View Smart Fetal Heart (USFP-AI900A is required.)
VolPure STL export (USPF-AI900A is required.)
Magic Cut · The following measurements can be performed.
STIC/STIC Color MPR
4D CHI (USHI-AI900A is required.) Multi Auto Volume Measurement (USOB-AI900A is
4D Biopsy required.) [ Table 2 ]
Fly Thru (USFT-AI900A is required.) – Cardiac 4D
Luminance (USLM-AI900A is required.) USVM-AI800A is required to use PSI-30VX, PSI-50VX and
Shadow Glass (USSG-AI900A is required.) PEI-512VX on i800 and i700.
OmniView (USVM-AI700A is required to use PSI-40VX on i700.)
Auto flexible cut line · The following functions can be used.
STL export (USPF-AI900A is required.) Multi plane
· The following measurements can be performed. Volume (including ECG synchronous scan)
MPR Volume Color (including ECG synchronous scan)
Multi Auto Volume measurement (USOB-AI900A is Shadow Glass (USLM-AI901A is required.)
required.) [ Table 1 ] Luminance
– General 4D (USVM-AI700A is required for i700.) Smart Fetal Heart (USFP-AI900A is required.)
· The following functions can be used. STL export (USPF-AI900A is required.)
Volume Color · The following measurements can be performed.
Multi View MPR [ Table 3 ]
Volume View
[ Table 1 ]
Single 4D STIC Volume 4D Fly Shadow Max sweep
Transducer name 4D Live STIC Luminance*1
Sweep CHI*1 Color Color Biopsy Thru*1 Glass*1 range (deg)
PVT-681MVL 150
[ Table 2 ]
4D Volume 4D Fly Shadow 4D Mitral Valve Aortic Valve
Transducer name 4D Live Luminance*1
CHI*1 Color Biopsy Thru*1 Glass*1 SMI*1 Analysis*1 Analysis
PSI-40VX
[ Table 3 ]
4D Volume 4D Fly Shadow 4D Mitral Valve
Transducer name 4D Live Luminance*1
CHI*1 Color Biopsy Thru*1 Glass*1 SMI*1 Analysis*1
PSI-30VX
PSI-50VX
PEI-512VX
21
Prism Edition
• Stress Echo (USUR-AI900A and UJUR-AI900A, or USUR- · Multivolume display (Quad display)
AI900A and UJUR-AI901A are required.) · 3D body mark display (Quad display)
Exercise and pharmacological stress echo examinations · Auto Registration (USFN-AI901A is required.)
can be performed. · Auto Track (UIOT-AI900A or UIOT-AI901A is required.)
– Data acquisition mode – The following measurements can be performed.
This mode is intended for selecting and creating protocols. · Basic measurement
– Review mode · Cardiac measurement (for ultrasound live image)
This mode provides playback function/data output func- • Protocol Assistant (USPA-AI900A is required.)
tion/Wall Motion Scoring (WMS) function. A series of operations (a protocol) that has been created
• Panoramic View (USPV-AI900A is required.) 13 for the intended examination can be executed automatically.
– A continuous image can be acquired by moving the Protocols can be created and edited.
transducer horizontally on the body surface. • Shear wave (USSW-AI900A is required.)
– Measurement using Panoramic View can be performed. Images representing the speed of propagation of tissue
• Elastography (USEL-AI901A or USEL-AI900A is required.) displacement (shear wave speed) can be visualized (shear
– Tissue stiffness can be visualized based on the changes wave scan) by locally displacing tissues by transmitting a
in velocity resulting from physical compression and burst wave with high acoustic pressure.
decompression of the target region. – The following Shear wave display modes are available.
– FLR measurement can be performed to calculate the · Speed: shear wave speed display (m/s)
strain within the set ROI. (Not available in the USA.) · Elasticity: elasticity display (kPa)
• VI (Vascularity Index) · Propagation: propagation display
The number of pixels in the Power/ SMI image and within · Dispersion: Frequency dispersion display
the ROI, the area of the Power/ SMI image and of the ROI, (USDM-AI900A is required.)
and the ratio of the number of pixels in the Power/ SMI – Quad display is possible. (USQS-AI800A is required for
image to the number of pixels within the ROI can be dis- i700.)
played for an image acquired in Power/ SMI mode. – ECG Sync Acquisition function can be used.
Available on all linear transducers and all convex transduc- – Shear wave measurement can be performed.
ers. – ATI (Attenuation Imaging)
• Histogram (USAT-AI900A is required.)
The number of pixels, average gradation, standard devia- The ultrasound wave attenuation can be displayed in
tion, and brightness distribution inside the ROI in the color parametric and measured.
image acquired with 2D mode are displayed. • Smart Navigation (USSN-AI900A etc. are required.)
• Fusion (Smart Fusion) (USFN-AI900A etc. are required.) A needle navigation line can be superimposed on the
Synchronization of ultrasound scanning with CT/MRI ultrasound image based on the positional relationship
image display adjusted according to the examination posi- between the magnetic sensor attached to the transducer
tion determined using a magnetic sensor attached to the and the magnetic sensor attached to the puncture needle.
transducer can be performed. • Smart Sensor 3D (USSS-AI900A, UIFR-AI900A, and UIFR-
Prostate biopsy report is available. AI501A are required.)
– The following video modes can be selected. Generation of 3D images can be performed based on the
· 2D mode (B mode) positional information obtained using the magnetic sensor
· Color Doppler attached to the transducer.
· PWD (pulsed-wave Doppler) – The following functions can be used.
· CWD (continuous-wave Doppler) · Volume Color
(USCW-Al900A is required.) · Multi View
· CHI (Contrast Harmonic Imaging) · Volume View
· Elastography (USEL-AI901A or USEL-AI900A is required.) · Magic Cut
– The following image data can be referred to. · CHI (USHI-AI900A is required.)
· CT · Fly Thru (USFT-AI900A is required.)
· MR · Luminance (USLM-AI900A is required.)
· PET · Shadow Glass (USSG-AI900A is required.)
· 3D US (USSS-AI900A is required.) · Shear Wave (USSW-AI900A is required.)
– The following functions can be used. · OmniView
· Blend display · ECG Sync Construction
· Segment display · STL export (USPF-AI900A is required.)
· Multiplane display (Triple display)
22
MPDUS0212EAC
Measurement Functions
Basic Measurement Functions
• 2D-mode measurements
– Distance
· Distance
· Trace Length
· Mean-IMT 23,3
– Area
– Angle
· Angle
· Joint
– Volume
– Stenosis ratio
· %Stenosis (Distance)
· %Stenosis (Area)
• 4D-mode measurements (USMV-AI900A is required.)
– Distance
· Distance
· Trace Length
· Mean-IMT
– Area
– Angle
· Angle
· Joint
24
MPDUS0212EAC
25
Prism Edition
• TCA (Time Curve Analysis) (USHI-AI900A and USCQ-AI900A • RADS (Reporting And Data Systems)
are required.) (USRA-AI900A is required)
Quantitative analysis can be performed using a graph – BI-RADS™ (Breast Imaging Reporting And Data Systems)
showing the changes in intensity over time of images – TI-RADS™ (Thyroid Imaging Reporting And Data Systems)
acquired in CHI mode. – LI-RADS™ (Liver Imaging Reporting And Data Systems)
– Generation of a graph from the changes in intensity over • IOTA (International Ovarian Tumor Analysis)
time (USIO-AI900A is required)
– Motion Tracking function This function provides standardization for ovarian cancer
– Function for calculating characteristic value parameters characterization according to ADNEX (risk calculation algo-
by Curve Fitting (USCQ-AI901A is required.) rithm model) through workflow.
– Analytical data output to a file
• MVA (Mitral Valve Analysis) (USMA-AI900A is required.) Cine Memory (large-capacity image memory) 21.1.1
The mitral valve in the 3D image data acquired using the • Memory Capacity
ultrasound diagnostic system is extracted semi-automati- 960 MB 22
cally, and then it is displayed as a model so that the shape • Record/Playback Mode
of the mitral valve can be easily evaluated and analyzed to – Loop playback is possible.
display the parameters. – Frame advance playback is possible.
– Extraction of the mitral valve in the 3D image – Cine playback is possible in Doppler or M mode.
– Display of the model – Live images can be recorded. (Clips, Auto Store)
– Display of the parameters
– Extraction from mitral valve for multiple frames Video Recording
(USMA-AI901A is required.) The following DVD remote control operations are possible:
• AVA (Aortic Valve Analysis) (USAV-AI900A is required) Record, stop, play, fast-forward, rewind, forward search,
The aortic valve in the 3D image data acquired using the reverse search, and freeze (pause).
ultrasound diagnostic system is extracted semi-automati-
cally and the dimension of aortic valve complex is assessed Recording Function
with 3D model. • Printers (option)
– Black-and-white printer: USB connection
Measurement of Stored Image Data – Color printer: USB connection
The following measurements can be performed for the • Video Recording Units (Option)
DICOM data (DICOM, with raw data, and without raw data) – DVD Video
stored in HDD. • Electronic Filing
• Basic measurement – Hard disk drive
• Application measurement · Built-in HDD (SATA)
· External HDD (USB3.0)
* UZHI-AI900A and USDB-AI900A are required.
Report Function
• Worksheet functions The HDDs should meet the following specifications:
– The measurement and calculation items can be dis- Capacity: 500 GB to 6 TB
played for each application measurement. Speed: 7200 rpm or higher
– Data editing is possible (except for some items). Interface: USB 3.0 (bus power)
– Display of the following values can be set to ON or OFF. External dimension: 117 (W) × 81 (D) × 22.5 (H) mm
Mean value, latest value, maximum value, minimum value or less
– Trend graphs can be displayed (OB measurement work- Format: The following Windows® formats
sheet). (read/write compatible)
– Multi Parametric Report, which allows organization of · NTFS
results acquired using multiple abdominal measurement · FAT32 (up to 2 TB)
applications in a worksheet, can be displayed. (USDL- Other: Single partition
AI900A is required.) – DVD/CD drive
– Analysis results of RADS used during an examination can – USB flash drive
be displayed on the Worksheet. (USRA-AI900A is required.) – Network: DICOM connection
• Report function (On Board Report)
– Reports can be created on the system.
– The created reports can be printed.
– The created reports can be output as PDF files.
– The report template can be edited.
– Comment entry is possible.
26
MPDUS0212EAC
27
Prism Edition
287.7 (11.3)
Sony HVO-550MD/FHD (including
DVD video recorder 212 (8.4) 105.5 (4.2) 3.2 (7.1) 43.2 W
[NTSC/PAL]
the projection
section)
Sony UP-D711MD/AC 140 (5.5) 70 (2.8) 125 (4.0) 1.0 (2.2) 72 VA (printing)
Sony UP-D898MD 154 (6.1) 88 (3.5) 240 (9.4) 2.6 (5.7) 190 VA (printing)
B/W digital printer
Mitsubishi P95DW 154 (6.1) 84.5 (3.3) 130 (5.1) 2.6 (5.7) 190 VA (printing)
Mitsubishi P95DW-DC 154 (6.1) 84.5 (3.3) 130 (5.1) 1.6 (3.5) 68 VA (printing)
Mitsubishi CP30DW 212 (8.3) 125 (4.9) 425 (16.7) 7.3 (16.1) 180 VA (printing)
Color digital printer
Sony UP-D25MD 212 (8.3) 98 (3.9) 398 (15.7) 5.5 (12.1) 240 VA (printing)
*1: 580 mm in length when the provided transducer connector holder is removed.
*2: In Standby mode with internal battery UEBT-AI900A.
29
Prism Edition
MASS
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
System main unit USVS-AI900A 4D ADF/SMI kit 0.1 (0.2)
TUS-AI900 Aplio900 110 (242.5) USVC-AI900A 4D CHI kit 0.1 (0.2)
TUS-AI800 Aplio800 110 (242.5) USPF-AI900A 3D printer format export kit 0.1 (0.2)
TUS-AI700 Aplio700 110 (242.5) USEL-AI901A Elastography-FLR kit 0.1 (0.2)
Options/Accessories for main unit USEL-AI900A Elastography kit 0.1 (0.2)
USCW-AI900A CW kit 0.1 (0.2) USSW-AI900A Shear Wave kit 0.1 (0.2)
USUR-AI900A Reference Signal kit 0.1 (0.2) USSW-AI901A Shear wave Hard kit 0.1 (0.2)
UJUR-AI900A Reference signal cable 0.3 (0.7) USQS-AI800A 0.1(0.2) 0.1 (0.2)
UJUR-AI901A Reference signal cable 0.1 (0.2) USDM-AI900A Dispersion Map for SWE kit 0.1 (0.2)
UJUR-AI902A Reference signal sensor unit 0.4 (0.9) USFN-AI900A Smart Fusion kit 0.1 (0.2)
USSE-AI900A Stress Echo kit 0.1 (0.2) USSN-AI900A Smart Navigation kit 0.1 (0.2)
USWT-AI900A 2D Wall Motion Tracking kit 0.1 (0.2) UIFR-AI900A Magnetic Generator kit 15.0 (33.0)
USWT-AI904A 2D Wall Motion Tracking Fetal kit 0.1 (0.2) UIFR-A501A Sensor kit for Fusion unit 0.1 (0.2)
USWT-AI906A 2D Wall Motion Tracking kit 0.1 (0.2) USFN-AI901A Auto Registration kit 0.1 (0.2)
USSD-AI900A SI-DI kit 0.1 (0.2) UZWT-A500A Fusion Pole Cart 26 (57.3)
USWT-AI901A 3D Wall Motion Tracking kit 0.1 (0.2) UAFS-001A Mounting kit for fusion sensor 0.1 (0.2)
USWT-AI902A 3D Wall Motion Tracking Advance kit 0.1 (0.2) UAFS-002A Mounting kit for fusion sensor 0.1 (0.2)
USQT-AI900A Quad Chamber Tracking kit 0.1 (0.2) UAFS-003A Mounting kit for fusion sensor 0.1 (0.2)
USMA-AI900A Mitral Valve Analysis kit 0.1 (0.2) UAFS-004A Mounting kit for fusion sensor 0.1 (0.2)
USMA-AI901A MVA 4D kit 0.1 (0.2) UAFS-005A Mounting kit for fusion sensor 0.1 (0.2)
USVM-AI800A Volume Matrix kit 0.1 (0.2) UAFS-006A Mounting kit for fusion sensor 0.1 (0.2)
USVM-AI700A Volume Matrix 2 kit 0.1 (0.2) UAFS-007A Mounting kit for fusion sensor 0.1 (0.2)
Luminance and Shadow Glass with UAFS-008A Mounting kit for fusion sensor 0.1 (0.2)
USLM-AI901A 0.1 (0.2)
Volume Matrix kit UAFS-009A Mounting kit for fusion sensor 0.1 (0.2)
USAV-AI900A AVA kit 0.1 (0.2) UAFS-010A Mounting kit for fusion sensor 0.1 (0.2)
USFH-AI600A Fetal Heart MPI Measurement kit 0.1 (0.2) UAFS-011A Mounting kit for fusion sensor 0.1 (0.2)
USFP-AI900A Smart Fetal Heart kit 0.1 (0.2) USMB-AI900A Breast Scan Guide kit 1.0 (2.2)
UIPC-AI900A Pencil Connector unit 0.2 (0.4) USSB-AI900A Smart Body Mark kit 1.0 (2.2)
UAEH-AI900A M-TEE hanger kit 1.5 (3.3) USSS-AI900A Smart Sensor 3D kit 0.1 (0.2)
UAEH-AI901A TEE hanger kit 1.2 (2.6) USMP-AI900A MicroPure kit 0.1 (0.2)
UACV-AI900A CV kit 1.0 (2.2) USPV-AI900A Panoramic View kit 0.1 (0.2)
UIST-AI900A STC kit 0.1 (0.2) USRC-AI900A Multi-reflection Canceller kit 0.1 (0.2)
UIST-AI901A STC kit 0.1(0.2) USLP-Al900A Liver Package kit 0.1 (0.2)
Ultrasound scanning condition USLD-AI900A Doppler Luminance kit 0.1 (0.2)
USVL-AI900A 0.1 (0.2)
export function Kit
USGI-AI900A General Imaging kit 0.1 (0.2)
USHI-AI900A CHI kit 0.1 (0.2)
USAT-AI900A Attenuation Imaging kit 0.1 (0.2)
USCQ-AI900A CHI-Q kit 0.1 (0.2)
USSL-AI900A Slice thickness control kit 0.1 (0.2)
USCQ-AI901A Fitting Curve kit 0.1 (0.2)
USZS-AI900A Measurement Z score kit 0.1 (0.2)
USQC-AI800A Quad view for CHI kit 0.1 (0.2)
UAOB-Al900A OB kit 0.1 (0.2)
USHI-AI901A High frame rate CHI kit 0.1 (0.2)
USTM-Al900A Limb Volume Measurement kit 0.1 (0.2)
USVF-AI900A Contrast Vector Imaging kit 0.1 (0.2)
UZRI-AI900A Mounting kit for peripheral unit 1.1 (2.4)
USMV-AI900A 4D kit 0.1 (0.2)
UZRI-AI901A Mounting kit for peripheral unit 0.6 (1.3)
USLM-AI900A Luminance kit 0.1 (0.2)
UZRI-AI902A Mounting kit for peripheral unit 3.1 (6.8)
USFT-AI900A Fly Thru. kit 0.1 (0.2)
UZRI-Al903A Mounting kit for peripheral unit 0.6 (1.3)
USSG-AI900A Shadow Glass kit 0.1 (0.2)
UZRI-Al904A Mounting kit for peripheral unit 4.6 (10.1)
USOB-AI900A Auto Volume Measurement kit 0.1 (0.2)
30
MPDUS0212EAC
Model name Name of component Mass [kg] (lb) Model name Name of component Mass [kg] (lb)
UZFS-004A Footswitch 0.6 (1.3) PSI-30VX Phased array transducer 0.8 (1.8)
UZGW-008A Gel warmer 1.0 (2.2) PSI-40VX Phased array transducer 0.85 (1.9)
UZMK-AI900A Transducer cable hanger kit 0.9 (2.0) PSI-50VX Phased array transducer 0.75 (1.7)
UZBK-AI900A Transducer Holder kit 0.3 (0.7) PSI-70BT Phased array transducer 0.65 (1.4)
UIWL-A500A Wireless LAN kit 0.5 (1.1) PST-25BT Phased array transducer 0.8 (1.8)
UIWL-AI900A Wireless LAN kit 0.5 (1.1) PST-28BT Phased array transducer 0.67 (1.5)
UZPH-AI900A EV/ER Transducer Holder kit 1.6 (3.5) PST-30BT Phased array transducer 0.8 (1.8)
UIKB-AI900A Keyboard kit 1.1 (2.4) PST-50BT Phased array transducer 0.8 (1.8)
UITB-AI900A 2nd Console kit 1.1 (2.4) PST-65BT Phased array transducer 0.73 (1.61)
UIOM-001A OLED Monitor 4.1 (9.0) PVI-482BX Convex array transducer 0.7 (1.5)
USDB-AI900A DataBase for External HDD kit 0.1 (0.2) PVI-475BT Convex array transducer 0.75 (1.7)
UZHI-AI900A Mounting kit for External HDD kit 1.0 (2.2) PVI-450BXP Convex array transducer 0.8 (1.76)
UZMK-AI902A ECG Cable Hanger kit 0.1 (0.2) PVI-475BX Convex array transducer 0.8 (1.8)
UIUB-AI900A Panel USB Port kit 0.1 (0.2) PVI-574BX Convex array transducer 0.75 (1.7)
UZPT-001A Palm Controller kit 0.2 (0.4) PVT-350BTP Convex array biopsy transducer 0.95 (2.1)
UZPT-002A Palm Controller kit 0.2(0.4) PVT-375BT Convex array transducer 0.95 (2.1)
UEBT-AI900A Battery unit 13 (28.7) PVT-375SC Convex array transducer 0.95 (2.1)
USHE-AI900A Online Help kit 0.1 (0.2) PVT-382BT Convex array transducer 0.8 (1.8)
USPA-AI900A Protocol Assistant kit 0.1 (0.2) PVT-482BT Convex array transducer 0.75 (1.7)
USPA-AI901A MSK Protocol Movie kit 0.1(0.2) PVT-574BT Convex array transducer 0.75 (1.7)
USDL-AI900A Multi Parametric Report kit 0.1 (0.2) PVT-674BT Convex array transducer 0.9 (2.0)
UAWH-AI900A Woman Helthcare kit 0.1 (0.2) PVT-675MVL Convex array transducer 1.1 (2.4)
USSM-AI900A Security Management kit 0.1 (0.2) PVT-675MVS Convex array transducer 1.05 (2.3)
USRA-AI900A RADS kit 0.1 (0.2) PVT-681MVL Endocavitary transducer 1.15 (2.5)
USIO-AI900A IOTA kit 0.1 (0.2) PVT-712BT Convex array transducer 0.8 (1.8)
USUH-AI700A Ultra High Freq kit 0.1 (0.2) PVT-745BTF Convex array transducer 0.8 (1.8)
UZTB-Al900A Track Ball kit 0.2 (0.4) PVT-745BTH Convex array transducer 0.8 (1.8)
UZTB-AI901A Track Ball kit 0.2(0.4) PVT-745BTV Convex array transducer 0.78 (1.7)
UZKF-AI900A Local Language Key-Top kit 0.5 (1.1) PVT-770RT Endocavitary transducer 2.0 (4.4)
UZKG-AI900A Local Language Key-Top kit 0.5 (1.1) PVT-781VT Endocavitary transducer 0.97 (2.1)
UZKI-AI900A Local Language Key-Top kit 0.5 (1.1) PVT-781VTE Endocavitary transducer 0.97 (2.1)
UZKS-AI900A Local Language Key-Top kit 0.5 (1.1) PVL-715RST Endocavitary transducer 1.04 (2.3)
UZKD-AI900A Local Language Key-Top kit 0.5 (1.1) PLI-705BX Linear array transducer 0.83 (1.8)
UZKN-AI900A Local Language Key-Top kit 0.5 (1.1) PLI-605BX Linear array transducer 0.83(1.8)
UZKW-AI900A Local Language Key-Top kit 0.5 (1.1) PLI-1205BX Linear array transducer 0.78 (1.7)
UZKR-AI900A Local Language Key-Top kit 0.5 (1.1) PLI-2002BT Linear array transducer 0.75 (1.7)
HVO-550MD/ PLI-2004BX Linear array transducer 0.78 (1.7)
DVD video recorder 3.2 (7.1)
FHD PLI-3003BX Linear array transducer 0.81 (1.8)
UP-D711MD/AC B/W printer 1.0 (2.2) PLT-308BTP Linear array biopsy transducer 0.8 (1.8)
UP-D898MD B/W printer 2.6 (5.7) PLT-704SBT Linear array transducer 0.9 (2.0)
P95DW B/W printer 2.6 (5.7) PLT-705BT Linear array transducer 0.85 (1.9)
P95DW-DC B/W printer 1.6 (3.5) PLT-705BTF Linear array transducer 0.78 (1.7)
CP30DW Color printer 7.3 (16.1) PLT-705BTH Linear array transducer 0.78 (1.7)
UP-D25MD Color printer 5.5 (12.1) PLT-1005BT Linear array transducer 0.85 (1.9)
Transducers PLT-1202BT Linear array transducer 0.75 (1.7)
PSI-30BX Phased array transducer 0.67 (1.5) PLT-1204BT Linear array transducer 0.85 (1.9)
31
MPDUS0212EAC
Prism Edition
https://global.medical.canon
©Canon Medical Systems Corporation 2020-2021. All rights reserved.
Design and specifications are subject to change without notice.
Model number: TUS-AI900/AI800/AI700 MPDUS0212EAC V6.0 2021-05 CMSC/Produced in Japan
Canon Medical Systems Corporation meets internationally recognized standards for Quality Management System ISO 9001, ISO 13485.
Canon Medical Systems Corporation meets the Environmental Management System standard ISO 14001.
Aplio, Dynamic Flow, ApliPure, MicroPure, Pulse Subtraction, VolPure, TwinView and Made for Life are trademarks of Canon Medical Sys-
tems Corporation.
Windows and AppLocker are registered trademarks of Microsoft Corporation in the United States and / or other countries.
McAfee is a registered trademark of McAfee Associates, Inc.
omniTRAX and VirtuTRAX are trademarks of CIVCO Medical Solutions.
BI-RADS, TI-RADS and LI-RADS are trademarks of the American College of Radiology.
This document may include trademarks or registered trademarks of their respective owners.
Volume Matrix
Volume Matrix
intelligent
DMS
2 3
V 4.0 V 4.0 V 4.0 V 4.0
PST- 25BT 2 27,7 PVT - 375BT PVT - 482BT PVT - 712BT 27,6
*
PSI-40VX PLT-1202BT
Sector
Sector
Convex
Convex
Convex
Linear
1 2.5 1.7~4.7 Volume 1 4.0 1.8~6.0 1 3.5 1.5~6.0 1 4.0 1.8~6.0 1 7.0 4.3~11.0 1 12.0 4.5~17.0
2 Approx. 90° - Matrix 2 Approx. 120° - 2 Approx. 70° Approx. 50mm 2 Approx. 80° Approx. 20mm 2 Approx. 100° Approx. 15mm 2 Approx. 25.6mm -
Cardiac (adult heart), Neonatal head, abdominal, Small parts (superficial), MSK,
3 3 3D Abdominal, fetal, pediatric 3 Abdominal, fetal, pediatric 3 Abdominal, pediatric, fetal 3 3
Transcranial Doppler pediatric intraoperative
4 680-106 680-086 610-608 4 680-156 680-157 610-1500/1502 4 680-105 680-085 610-608 4 - - 610-1500/1502 4 680-127 680-126 610-608 4 - - -
5 - 5 - 5 - 5 - 5 - 5 -
6 - 6 UAFS-009A 6 Included with UIFR-AI900A 6 UAFS-001A 6 - 6 UAFS-010A
7 - Single Crystal 7 - 7 - Single Crystal 7 - 7 - 7 -
*
PVT-375SC PVI-482BX
Linear
1 3.0 1.5~6.0 1 5.0 3.0~8.2 1 3.5 1.5~6.0 intelligent 1 4.0 1.8~6.0 1 7.5 4.0~10.5 1 12.0 4.5~18.0
2 Approx. 120° - 2 Approx. 90° - 2 Approx. 70° Approx. 50mm DMS 2 Approx. 80° Approx. 20mm 2 Approx. 38mm - 2 Approx. 38mm -
Cardiac (adult heart), Cardiac Peripheral vascular (carotid, Small parts
3 3 3 Abdominal, fetal, pediatric 3 Abdominal, fetal, pediatric 3
Transcranial Doppler (pediatric, neonatal heart) lower leg), small parts, MSK 3 (superficial, breast), MSK,
peripheral vascular
4 - - - 4 - - - 4 680-105 680-085 610-608 4 - - 610-1500/1502 4 680-107 680-087 610-608
680-102 680-088 610-608
5 - 5 - 5 - 5 - 5 -
4 610-1231
6 UAFS-009A 6 - 6 Included with UIFR-AI900A 6 UAFS-001A 6 - 680-123 680-122
610-1235
Single Crystal 7 - 7 - Single Crystal 7 - Single Crystal 7 - 7 - 5 -
7 UACK-004A
*
1 3.0 1.7~5.2 Volume 1 5.0 2.2~6.0 1 3.5 2.0~5.0 1 5.0 2.0~9.5 1 7.0 3.0~8.5 intelligent 1 12.0 4.0~18.2
2 Approx. 90° - Matrix 2 Approx. 90° - 2 Approx. 80° Approx. 20mm 2 Approx. 65° Approx. 40mm 2 Approx. 45mm - DMS 2 Approx. 46mm -
Cardiac (adult heart), 3D Cardiac Peripheral vascular (lower leg, Small parts (breast), MSK,
3 3 3 Abdominal, pediatric, fetal 3 Abdominal, fetal, pediatric 3 3
Transcranial Doppler (pediatric, neonatal heart) carotid), small parts, MSK peripheral vascular
4 680-106 680-086 610-608 4 - - - 4 680-111 680-110 680-112 4 680-168 680-169 610-1500 4 680-129 680-128 610-608 680-139 680-138 610-608
5 - 5 - 5 UAGV-027A 5 - 5 - 680-149 680-148 610-1500
4
6 - 6 UAFS-009A 6 UAFS-001A 6 UAFS-008A 6 - 610-1017
680-132 680-133 610-1018
7 - Single Crystal 7 7 - Single Crystal 7 - 7 - 610-1019
6 UAFS-008A
7 UACK-004A
PSI- 30BX 2
PVI - 475BT 2
PLI- 705BX 2
*
*
PST-65BT PVI-574BX
2
PLI- 2002BT
*
intelligent 1 3.0 1.5~6.0 1 7.0 3.5~12.0 1 4.0 1.8~6.4 intelligent 1 5.0 2.0~9.5 intelligent 1 7.0 3.5~11.5
DMS 2 Approx. 120° - 2 Approx. 120° - 2 Approx. 70° Approx. 50mm DMS 2 Approx. 65° Approx. 40mm DMS 2 Approx. 46mm - 1 20.0 8.8~22.0
Cardiac (adult heart), 3 Abdominal, fetal, pediatric Peripheral vascular (lower leg, 2 Approx. 25.6mm -
3 3 Cardiac (pediatric heart) 3 Abdominal, fetal, pediatric 3
Transcranial Doppler carotid), small parts, MSK
680-145 680-144 610-608 Small parts (superficial), MSK,
4 - - - 4 - - - 4 4 680-166 680-167 610-1500 680-150 680-151 610-1500 3
intraoperative
680-155 680-154 610-1500
5 - 5 - 5 - 4 610-1017 4 - - -
5 - 680-164 680-165 610-1018
6 UAFS-009A 6 - 6 UAFS-008A 610-1019 5 -
UAFS-007A
Single Crystal 7 - 7 - Single Crystal 6 Single Crystal 7 - 5 -
(included with UIFR-AI900A) 6 UAFS-010A
Volume 1 3.0 1.8~5.2 1 7.0 3.5~9.0 intelligent 1 4.0 1.8~6.4 1 6.0 3.5~9.7 1 10.0 3.8~14.0 1 Frequency (MHz) Displayed range of freq. (MHz)
Matrix 2 Approx. 120° - 2 Approx. 90° - DMS 2 Approx. 70° Approx. 50mm 2 Approx. 65° Approx. 40mm 2 Approx. 58mm - 2 Field of view / angle Curvature
3 3D Cardiac (adult heart) 3 Cardiac (neonatal heart) 3 Abdominal, fetal, pediatric 3 Abdominal, fetal, pediatric Small parts (breast), MSK, 3 Indication for use
3
peripheral vascular 1
*
680-143 680-142 610-608 4 Starter Ki t Bracket Needle guide
4 680-156 680-157 610-1500/1502 4 - - - 4 4 - - -
680-109 680-090 610-608
680-153 680-152 610-1500 5 Reusable biopsy adapter
5 - 5 - 5 UAGV-028A 4
5 - 610-1231 6 Fusion adapter
680-125 680-124
6 UAFS-009A 6 - 6 - 610-1235
UAFS-007A 7 Water bag / Sono coupler
Single Crystal 7 - 7 - Single Crystal 6 7 - 5 -
(included with UIFR-AI900A)
1 Includes the bracket and the sterile procedure kit.
* *
i6SVX1 i10S4 i8CX1 7 - 10C3 14L5 6 UAFS-004A 2 Not compatible with Aplio i600
4 5
V 4.0 V 4.0 V 4.0 V 4.0
PLI- 2004BX 2
PVL- 715RST PVT- 745BTV PET - 805LA
*
Linear
Endorectal
Intraoperative
Laparoscopic
intelligent 1 20.0 8.8~24.0 1 7.5 4.5~9.0 1 7.0 3.2~10.0 1 8.0 3.5~9.0
DMS 2 Approx. 41mm - Approx. 170° 2 Approx. 40° Approx. 45mm 2 Approx. 45mm -
2 Approx. 10mm
Approx. 56mm
Intraoperative, abdominal,
3 Small parts (superficial), MSK 3 3 Laparoscopic
3 Prostate, bladder small parts
680-137 680-136 610-608 4 - - - 4 - - -
4 - - -
680-147 680-146 610-1500 5 UAGV-029A 5 -
4 UAGL-001AHA
5
610-1017 UAGL-002AHA 6 - 6 -
680-134 680-135 610-1018
610-1019 6 UAFS-005A 7 - 7 -
*
Transesophagus
Transesophagus
Mechanical 4D
1 7.0 4.7~10.0 1 7.0 3.0~10.0 1 5.0 2.5~7.0 Volume 1 5.0 1.8~6.0 1 6.0 3.6~11.0
2 Approx. 170° Approx. 10mm 2 Approx. 40° Approx. 45mm 2 Approx. 90° - Matrix 2 Approx. 90° - 2 Approx. 180° Approx. 10mm
2
PLI- 3003BX
*
*
PET-609MA
Pencil
PLT- 308BTP 3
PVT - 745BTH PET - 512MA 3
*
*
i33LX9
Biopsy
UAGL-004AHA 5 - 5 -
5 5 UAGV-034A 5 -
1 7.0 3.6~11.0 (included with transducer)
6 - 6 -
6 - 6 - 6 -
2 Approx. 180° Approx. 10mm
7 - 7 -
7 - 7 - 7 -
3 Transvaginal uterus, ovary 8S3 P2
6LP3 11CI4 8S2
4 - - 680-121
5 UAGV-035A
PVT- 675MVL PC - 50M
Mechanical 4D
6 UAFS-003A
PVT- 350BTP 3
PLT - 705BTF PET- 512MC
*
*
4 - - - 4 Starter Ki t Bracket Needle guide
5 UAGV-035A 4 - - - 4 - - - 4 - - -
5 - 5 Reusable biopsy adapter
6 UA FS -006A 5 UAGV020A 5 - 5 -
6 UAFS-011A 6 Fusion adapter
11C3 7 - 6 - 6 - 6 -
7 - 7 Water bag / Sono coupler
Single Crystal 7 - 7 - 7 -
1 Includes the bracket and the sterile procedure kit.
* * *
i8CXP1 11LI4 8SM2 9CV2 2
3
Not compatible with Aplio i600
Not available in the USA and Canada
6 7
Tender Specs
Caution: This is only for tender purpose. Please don't disclose this information for any other purpose.
Luminance (cd/m2) typ. 300 / typ. 200 (7500 K), typ. 145 (13000 K)
M/D M/D mode sweep speed 2.381 - 19.05 sec / full screen, 8 steps (NTSC)
2.381 - 19.05 sec / full screen, 8 steps (PAL)
1018.5cm/s (PLT-705BT)
667.0cm/s (PLT-1204BT)
342.8.0cm/s (PLI-2004BX)
(i600 does not support PLI-2004BX)
Sample gate size for PWD 0.3-20 mm, 15 steps
PVI-475BX
Lowest Velocity 0.2 cm/s
Highest Velocity 10.0 cm/s
PLI-1205BX
Lowest Velocity 0.3 cm/s
Highest Velocity 11.2 cm/s
PLT-1204BT
Lowest Velocity 0.3 cm/s
Highest Velocity 9.3 cm/s
PLI-2004BX
Lowest Velocity 0.1 cm/s
Highest Velocity 4.3 cm/s
PLI-3003X
Lowest Velocity 0.1 cm/s
Highest Velocity 10.6 cm/s
PVl-475BX
POWERTECH
TX9
El sistema de potencia ininterrumpida (UPS)
Powertech TX9 1kVA / 1.5K kVA / 2kVA / 3kVA modelo TX9 de Powertech, es un sistema altamente
confiable, con tecnología de doble conversión
en línea, que garantiza una onda senoidal pura,
ideal para equipo de misión crítica (redes,
telecomunicaciones, servidores, y aplicaciones
industriales), con certificación UL Listed / UL 1778.
3
UPS EN LÍNEA
Dimensión
(DxWxH) 380 x 438 x88 480 x 438 x88 600 x 438 x88
ESPECIFICACIONES TÉCNICAS
POWERTECH TX9
T X9 1 kVA TX9 1.5 kVA TX9 2 kVA TX9 3 kVA
Fases Monofásico con tierra (fase, neutro, tierra)
ENTRADA
Voltaje 80VAC / 70VAC / 60VAC / 55VAC ± 5%
transferencia baja 5% (basado en el porcentaje de carga)100%-80% / 80%-70% / 70%-60% / 60%-0)
Voltaje de regreso 85VAC / 75VAC / 65VAC / 60VAC ± 5%
bajo 5% (basado en el porcentaje de carga) 100%-80% / 80%-70% / 70%-60% / 60%-0)
Rango de voltaje
Voltaje de 150 VAC ± 5%
transferencia alto
Voltaje de regreso
alto 140 VAC ± 5%
SALIDA
Voltaje de salida 110/115/120/127 VAC
Regulación de voltaje AC (Modo Batería) (±1)%
EFICIENCIA
Modo AC 86% típico, 88% a pico 88% típico, 90% a pico
Modo Emergencia 83% típico, 88% a pico 85% típico, 88% a pico
BATERÍA
Tipo de batería / Autonomía 12V/9AH / 10 min aprox
Cantidad 2 3 4 6
Voltaje de carga 27.4 VDC ± 1% 41.1 VDC ± 1% 54.7 VDC ± 1% 82.1 VDC ± 1%
INDICADORES
Pantalla LCD Nivel de carga, nivel de falla o alarma
ALARMA
Modo batería Emite sonido cada 4 segundos
CARACTERÍSTICAS FÍSICAS
Dimensiones DxWxH (mm) 380 x 438 x 88 (2U) 480 x 438 x 88 (2U) 380 x 438 x 88 (2U) 600 x 438 x 88 (2U)
Conectores entrada Nema 5-15P Nema 5-15P Nema 5-20P L5- 30P
Conectores salida Nema 5-15R Nema 5-15R Nema 5-15R L5-30R / Nema 5-15R
Número de
Receptáculos 8 protegidos por UPS y contra sobretensiones
CONDICIONES AMBIENTALES
Humedad 20-90 %RH @ 0-40°C (Sin condensación)
GESTIÓN
Smart RS - 232/USB Windows® 2000/2003/XP/Vista/2008, Windows® 7/8, Linux, Unix, and Mac
5
Powertech TX 1 kVA a 230VAC
9
Powertech TX9 6-10KVA a 230VAC
7
11