Anesthesia-GE-Aisys CS2-Network Configuration Guide
Anesthesia-GE-Aisys CS2-Network Configuration Guide
Software Revision 11
User responsibility
This product will perform in conformity with the description thereof
contained in this User’s Reference manual and accompanying labels
and/or inserts, when assembled, operated, maintained, and repaired
in accordance with the instruction provided.
This product must be checked periodically. A defective product
should not be used. Parts that are broken, missing, plainly worn,
distorted, or contaminated should be replaced immediately. Should
repair or replacement become necessary, the manufacturer
recommends that a telephonic or written request for service advice
be made to the nearest manufacturer Customer Service Center. This
product or any of its parts should not be repaired other than in
accordance with written instructions provided by the manufacturer
and by the manufacturer trained personnel. The product must not be
altered without the prior written approval of the manufacturer. The
user of this product shall have the sole responsibility for any
malfunction which results from improper use, faulty maintenance,
improper repair, damage, or alteration by anyone other than the
manufacturer.
CAUTION U.S. Federal law restricts this device to sale by or on the
order of a licensed medical practitioner. Outside the
U.S.A., check the local laws for any restriction that may
apply.
Table of Contents
1 Introduction
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Intended audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Ordering manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Revision history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Customer/network installer . . . . . . . . . . . . . . . . . . . . . . . 1-3
Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Safety statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
3 HL7 integration
DOC1949974 i
Aisys™ CS² Network Configuration Guide
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
4 Sapphire integration
ii DOC1949974
Table of Contents
DOC1949974 iii
Aisys™ CS² Network Configuration Guide
iv DOC1949974
1 Introduction
Introduction
DOC1949974 1-1
Aisys™ CS² Network Configuration Guide
Overview
Intended audience
Intended use
This network manual is intended for hospital IT departments and
network administrators and designers. Its purpose is to aid the
hospital or hospital assigned entity in the design and configuration of
the Network to which Aisys CS² anesthesia machines and clients are
connected. Use this manual to achieve the GE recommended
network reliability, security, and performance that is required for
connectivity between the anesthesia machine and intended clients.
Ordering manuals
Contact your local GE representative for additional ordering
information.
Revision history
The part number, revision level and date code of this manual are
located on the back cover (lower-right corner) of this manual.
Example:
Part number: 2098870-001
Date code: 08 2016
Revision level: A
1-2 DOC1949974
1 Introduction
Responsibility
Customer/network installer
Manufacturer
DOC1949974 1-3
Aisys™ CS² Network Configuration Guide
Safety information
Safety statements
Warnings
1-4 DOC1949974
1 Introduction
Cautions
Symbols
Symbol Definition
Manufacturer name and address
DOC1949974 1-5
1-6 DOC1949974
2 Network configuration overview
DOC1949974 2-1
Aisys™ CS² Network Configuration Guide
Network architecture
The network architecture allows the anesthesia machine to share
hospital network resources. These resources include access
switches, distribution switches, routers, copper cabling and fiber
optic links for hospital-wide connectivity as well as network services
and security servers for unified network management.
The anesthesia machine network should extend only to hospital
areas that require anesthesia machine network traffic. The
anesthesia machine network should be routed to the hospital
network so that data may be sent to the EMR interface. Ensure the
router between the anesthesia machine network and the EMR does
not block any of the flows for Port 2.
This disclosure is intended to satisfy the requirements of
IEC60601-1:2005, clause 14.13 and IEC/ISO 80001-1:2010, clause
3.5 for disclosure of network-related specifications, requirements,
and residual risks in order to facilitate the responsible organization’s
risk management activities (pursuant of 80001-1) for their networks
incorporating Aisys CS² anesthesia machine.
Router EMR
Port 2 OR Network
Hospital Network
(example: GE IX Network)
Port 1
Bedside Device
AC.25.086
Figure 2-1 • Network configuration example
Client devices
Network communication
2-2 DOC1949974
2 Network configuration overview
Network communication
Data output
The network configuration of the anesthesia machine allows users to
obtain the following information:
Measurements, configurations, and patient and system information
Service information
• Service logs
Network management
• Network configurations
• Request for co-ordinated time
DOC1949974 2-3
Aisys™ CS² Network Configuration Guide
Network specifications
2-4 DOC1949974
2 Network configuration overview
DOC1949974 2-5
Aisys™ CS² Network Configuration Guide
2-6 DOC1949974
2 Network configuration overview
Encrypted: No
Traffic characterization: Periodic transfer of service logs and
upon user request or system failure
initiated from anesthesia device.
Bandwidth requirements: 1 Kbps
Latency max: 250 ms
DOC1949974 2-7
Aisys™ CS² Network Configuration Guide
Note: Use the Connectivity Test to check the status of each connected device with a network ping
command. Any device connected to Port 1 and Port 2 must be turned on before starting the test. The
acceptance of Ping commands initiated from external devices to the Aisys CS2 has been disabled for
security purposes.
2-8 DOC1949974
3 HL7 integration
HL7 integration
DOC1949974 3-1
Aisys™ CS² Network Configuration Guide
Intended use
Anesthesia machines can be configured to send data to external
devices, including Electronic Medical Record systems and other
Hospital Information Systems using the Health Level Seven (HL7)
protocol.
This section is intended to be a technical reference for GE customers
integrating the Aisys CS² anesthesia machine HL7 output with an
external system. It is specifically intended for HL7 integration
engineers who are responsible for configuring hospital information
technology systems to interface together. This document is not
intended to provide background information on the HL7 standard.
For detailed information regarding the HL7 standard, refer to HL7
Version 2.6 or above documentation at "www.hl7.org".
Overview
The HL7 standard defines a high level specification for exchanging
data between healthcare systems. HL7 includes a great deal of
flexibility which can make it very difficult to integrate specific
systems. Integrating systems with HL7 often requires developing
specific integration guidelines before sending and receiving systems
can exchange messages.
The Aisys CS² systems address this issue by conforming to the HL7
message content and communication protocol guidelines as defined
by the IHE PCD-01 Data Enterprise Communication profile. The
Aisys CS² anesthesia machine implements the Device Observation
Reporter (DOR) actor role in a PCD-01 compliant system.
It is assumed that the users of this document have a basic
understanding of the "PCD Technical Framework Volume 2, Revision
5".
3-2 DOC1949974
3 HL7 integration
DOC1949974 3-3
Aisys™ CS² Network Configuration Guide
Protocol
Messages are sent using HL7 MLLP (Minimal Lower Layer Protocol)
as identified in the PCD (Patient Care Device) Technical Framework.
Messages are sent over a single persistent TCP/IP connection for
each device. The connection is left open as long as the client
continues to read messages from the network. The Aisys CS²
system will not try to re-send data if the data is not acknowledged by
the receiving system. The protocol does not require the client to send
an acknowledgement, therefore there is no way to check if the client
is HL7.
Connections are initiated by the Aisys CS² anesthesia machine as
configured by the user. Refer to the Aisys CS² Software Revision 11
User’s Reference manual for detailed information. If a connection
cannot be made to an external client, the anesthesia machine may
attempt communication in less than 30 seconds, but could take
longer between connection attempts.
The anesthesia machine monitors the connection to the remote client
to determine if the remote client is operating properly. If the remote
client is not responding, the Aisys CS² system will close the
connection and attempt to reconnect. The amount of time it takes to
detect a failure depends on the type of problem and the frequency
for sending messages set by the user.
Activity is monitored by the anesthesia machine. If the Aisys CS²
system detects excessive CPU or network utilization it will close any
open connections and shut down the networking functionality until
the anesthesia machine is restarted.
The network monitoring strategy uses a very conservative approach
to protect the therapy functions of the anesthesia machine. It is
possible that a temporary issue may shut down communications for
extended periods in order to protect system resources.
Adherence to the configuration guidelines and using validated HL7
clients can help reduce, but not eliminate the chance of a
communication shutdown.
3-4 DOC1949974
3 HL7 integration
Messages
Aisys CS² anesthesia machine conforms to the message format
defined by the HL7 specification version 2.6 and with additional
guidelines given in the PCD-01 profile.
Nomenclature for the parameters is taken from the Ammended ISO
Nomenclature: IEEE standard 11073-10101a. Observation sub-IDs
use the virtual device numeric codes from IEEE 11073. Units of
measure for the parameters are given using the Unified Code for
Units of Measure.
Messages are always sent with UTF-8 character encoding.
The sending application namespace is GE_ANES_SYS and
universal ID type is EUI-64. GE Healthcare is sent as the sending
facility name.
Times are transmitted in universal coordinated time with a local time
zone offset.
When the message sending frequency is once every 10 seconds or
1 minute, the value sent is the median of the last 10 seconds of
sampled data. If the message sending frequency is lower, the value
sent is the median of the last 60 seconds of sampled data.
Each message contains the following parameters:
.
DOC1949974 3-5
Aisys™ CS² Network Configuration Guide
3-6 DOC1949974
3 HL7 integration
DOC1949974 3-7
Aisys™ CS² Network Configuration Guide
3-8 DOC1949974
3 HL7 integration
DOC1949974 3-9
Aisys™ CS² Network Configuration Guide
3-10 DOC1949974
Appendix 1 HL7
MSH|^~\&|GE_ANES_SYS^080019FFFE000000^EUI-64|GE Healthcare|||20160503150814-0600||ORU^R01^ORU_R01|2289C295_7ABC_47CD_8796_286099914FAB|P|2.6|||AL|NE||UNICODE
UTF-8|||IHE_PCD_001^IHE PCD^1.3.6.1.4.1.19376.1.6.1.1.1^ISO
PID|||^^^^PI||^^^^^^U
PV1||U
OBR|1|080019FFFE00000020160503150814^GE_ANES_SYS^080019FFFE000000^EUI-64|080019FFFE00000020160503150814^GE_ANES_SYS^080019FFFE000000^EUI-64|70040^MDC_DEV_SYS
_ANESTH^MDC|||20160503150814-0600
OBX|1|CWE|184352^MDC_VENT_MODE^MDC|1.29.0.26|^vent-mode-spont|{unitless}^{unitless}^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|2|NM|16929196^MDC_VENT_VOL_TIDAL_SETTING^MDC|1.29.0.21||mL^mL^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|3|NM|16928802^MDC_VENT_RESP_RATE_SETTING^MDC|1.29.0.22||{breath}/min^{breath}/min^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|4|NM|16929048^MDC_RATIO_IE_SETTING^MDC|1.29.0.23||{unitless}^{unitless}^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|5|NM|16929856^MDC_RATIO_INSP_SETTING^MDC|1.29.0.42||{unitless}^{unitless}^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|6|NM|16929192^MDC_VENT_PRESS_AWAY_END_EXP_POS_SETTING^MDC|1.29.0.25||cm[H2O]^cm[H2O]^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|7|NM|16929844^MDC_VENT_TIME_PD_INSP_PAUSE_PERCENT_SETTING^MDC|1.29.0.24||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|8|NM|16928882^MDC_RESP_SPONT_RATE_SETTING^MDC|1.29.0.30||{breath}/min^{breath}/min^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|9|NM|16930360^MDC_CONC_GASDLV_O2_INSP_SETTING^MDC|1.29.0.27|100.00|%^%^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|10|NM|16929872^MDC_VENT_FLOW_CONTINUOUS_SETTING^MDC|1.29.0.31|6.00|L/min^L/min^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|11|NM|16929656^MDC_CONC_AWAY_O2_ET_SETTING^MDC|1.29.0.28||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|12|NM|16929072^MDC_TIME_PD_APNEA_SETTING^MDC|1.29.0.29|30.00|s^s^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|13|NM|16929416^MDC_CONC_GASDLV_DESFL_INSP_SETTING^MDC|1.29.0.32|-1.00|%^%^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|14|NM|16929420^MDC_CONC_GASDLV_ENFL_INSP_SETTING^MDC|1.29.0.33||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|15|NM|16929424^MDC_CONC_GASDLV_HALOTH_INSP_SETTING^MDC|1.29.0.34||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|16|NM|16929428^MDC_CONC_GASDLV_SEVOFL_INSP_SETTING^MDC|1.29.0.36||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|17|NM|16929432^MDC_CONC_GASDLV_ISOFL_INSP_SETTING^MDC|1.29.0.35||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|18|NM|16929300^MDC_CONC_AWAY_DESFL_ET_SETTING^MDC|1.29.0.37||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|19|NM|16929304^MDC_CONC_AWAY_ENFL_ET_SETTING^MDC|1.29.0.38||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|20|NM|16929308^MDC_CONC_AWAY_HALOTH_ET_SETTING^MDC|1.29.0.39||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|21|NM|16929312^MDC_CONC_AWAY_SEVOFL_ET_SETTING^MDC|1.29.0.41||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|22|NM|16929316^MDC_CONC_AWAY_ISOFL_ET_SETTING^MDC|1.29.0.40||%^%^UCUM||OFF|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|23|NM|151793^MDC_PRESS_AWAY_MAX^MDC|1.29.0.1|60.00|cm[H2O]^cm[H2O]^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|24|NM|151972^MDC_VENT_PRESS_AWAY^MDC|1.29.0.2||cm[H2O]^cm[H2O]^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|25|NM|152016^MDC_VENT_VOL_MINUTE_AWAY_INSP^MDC|1.29.0.3|14.30|L/min^L/min^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|26|NM|152664^MDC_VOL_AWAY_TIDAL_EXP^MDC|1.29.0.5|600.00|mL^mL^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|27|NM|153144^MDC_CONC_GASDLV_O2_INSP^MDC|1.29.0.4|60.00|%^%^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|28|NM|151586^MDC_VENT_RESP_RATE^MDC|1.29.0.6|60.00|{breath}/min^{breath}/min^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|29|NM|151666^MDC_RESP_SPONT_RATE^MDC|1.29.0.7|60.00|{breath}/min^{breath}/min^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|30|NM|152776^MDC_PRESS_RESP_PLAT_DYNAMIC^MDC|1.20.0.2||cm[H2O]^cm[H2O]^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|31|NM|151792^MDC_PRESS_AWAY^MDC|1.20.0.1||cm[H2O]^cm[H2O]^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|32|NM|151804^MDC_PRESS_AWAY_END_EXP_POS^MDC|1.20.0.12||cm[H2O]^cm[H2O]^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|33|NM|151808^MDC_PRESS_AWAY_END_EXP_POS_INTRINSIC^MDC|1.20.0.11||cm[H2O]^cm[H2O]^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|34|NM|151692^MDC_COMPL_LUNG_DYN^MDC|1.20.0.10||mL/cm[H2O]^mL/cm[H2O]^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|35|NM|152664^MDC_VOL_AWAY_TIDAL_EXP^MDC|1.20.0.6||mL^mL^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|36|NM|151888^MDC_VOL_MINUTE_AWAY_INSP^MDC|1.20.0.3||L/min^L/min^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|37|NM|151884^MDC_VOL_MINUTE_AWAY_EXP^MDC|1.20.0.5||L/min^L/min^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|38|NM|151840^MDC_RES_AWAY^MDC|1.20.0.9||cm[H2O]/(L/s)^cm[H2O]/(L/s)^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|39|NM|151784^MDC_PRESS_RESP_PLAT_STATIC^MDC|1.20.0.4||cm[H2O]^cm[H2O]^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|40|NM|151562^MDC_RESP_RATE^MDC|1.20.0.7||{breath}/min^{breath}/min^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|41|NM|151832^MDC_RATIO_IE^MDC|1.20.0.14||{unitless}^{unitless}^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|42|NM|152608^MDC_TIME_PD_INSP^MDC|1.20.0.15||s^s^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|43|NM|152612^MDC_TIME_PD_EXP^MDC|1.20.0.16||s^s^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|44|NM|152836^MDC_PRESS_AIR_AMBIENT^MDC|1.20.0.13||mm[Hg]^mm[Hg]^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
DOC1949974 3-11
OBX|45|NM|152420^MDC_FLOW_O2_CONSUMP^MDC|1.20.0.18||mL/min^mL/min^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|46|NM|151776^MDC_FLOW_CO2_PROD_RESP^MDC|1.20.0.19||mL/min^mL/min^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|47|NM|152812^MDC_RESP_EXPENDED_ENERGY^MDC|1.20.0.24||kcal/d^kcal/d^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|48|NM|151828^MDC_QUO_RESP^MDC|1.20.0.25||{unitless}^{unitless}^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|49|NM|151594^MDC_CO2_RESP_RATE^MDC|1.20.0.17||{breath}/min^{breath}/min^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|50|NM|152196^MDC_CONC_AWAY_O2_INSP^MDC|1.20.0.20||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|51|NM|152440^MDC_CONC_AWAY_O2_ET^MDC|1.20.0.22||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|52|NM|152192^MDC_CONC_AWAY_N2O_INSP^MDC|1.30.0.1||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|53|NM|152108^MDC_CONC_AWAY_N2O_ET^MDC|1.30.0.2||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|54|NM|151716^MDC_CONC_AWAY_CO2_INSP^MDC|1.20.0.21||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|55|NM|151708^MDC_CONC_AWAY_CO2_ET^MDC|1.20.0.23||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|56|NM|152168^MDC_CONC_AWAY_DESFL_INSP^MDC|1.30.0.3||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|57|NM|152084^MDC_CONC_AWAY_DESFL_ET^MDC|1.30.0.8||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|58|NM|152172^MDC_CONC_AWAY_ENFL_INSP^MDC|1.30.0.4||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|59|NM|152088^MDC_CONC_AWAY_ENFL_ET^MDC|1.30.0.9||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|60|NM|152176^MDC_CONC_AWAY_HALOTH_INSP^MDC|1.30.0.5||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|61|NM|152092^MDC_CONC_AWAY_HALOTH_ET^MDC|1.30.0.10||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|62|NM|152180^MDC_CONC_AWAY_SEVOFL_INSP^MDC|1.30.0.6||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|63|NM|152096^MDC_CONC_AWAY_SEVOFL_ET^MDC|1.30.0.11||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|64|NM|152184^MDC_CONC_AWAY_ISOFL_INSP^MDC|1.30.0.7||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|65|NM|152100^MDC_CONC_AWAY_ISOFL_ET^MDC|1.30.0.12||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|66|NM|152168^MDC_CONC_AWAY_DESFL_INSP^MDC|1.30.0.3||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|67|NM|152084^MDC_CONC_AWAY_DESFL_ET^MDC|1.30.0.8||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|68|NM|152172^MDC_CONC_AWAY_ENFL_INSP^MDC|1.30.0.4||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|69|NM|152088^MDC_CONC_AWAY_ENFL_ET^MDC|1.30.0.9||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|70|NM|152176^MDC_CONC_AWAY_HALOTH_INSP^MDC|1.30.0.5||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|71|NM|152092^MDC_CONC_AWAY_HALOTH_ET^MDC|1.30.0.10||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|72|NM|152180^MDC_CONC_AWAY_SEVOFL_INSP^MDC|1.30.0.6||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|73|NM|152096^MDC_CONC_AWAY_SEVOFL_ET^MDC|1.30.0.11||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|74|NM|152184^MDC_CONC_AWAY_ISOFL_INSP^MDC|1.30.0.7||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|75|NM|152100^MDC_CONC_AWAY_ISOFL_ET^MDC|1.30.0.12||%^%^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|76|NM|152864^MDC_CONC_MAC_SUM^MDC|1.10.0.1||{unitless}^{unitless}^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|77|NM|153156^MDC_FLOW_O2_FG^MDC|1.10.0.10|2.40|L/min^L/min^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|78|NM|152876^MDC_FLOW_AIR_FG^MDC|1.10.0.11|2.90|L/min^L/min^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|79|NM|153092^MDC_FLOW_N2O_FG^MDC|1.10.0.12||L/min^L/min^UCUM||NAV|||X|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|80|NM|153148^MDC_VOL_DELIV_O2_CASE^MDC|1.10.0.2|22.92|L^L^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|81|NM|152880^MDC_VOL_DELIV_AIR_CASE^MDC|1.10.0.4|0.00|L^L^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|82|NM|152960^MDC_VOL_DELIV_N2O_CASE^MDC|1.10.0.3|0.00|L^L^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|83|NM|152900^MDC_VOL_DELIV_DESFL_LIQUID_CASE^MDC|1.10.0.5|0.00|mL^mL^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|84|NM|152932^MDC_VOL_DELIV_HALOTH_LIQUID_CASE^MDC|1.10.0.8|0.00|mL^mL^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|85|NM|152948^MDC_VOL_DELIV_ISOFL_LIQUID_CASE^MDC|1.10.0.7|0.00|mL^mL^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|86|NM|152916^MDC_VOL_DELIV_ENFL_LIQUID_CASE^MDC|1.10.0.6|0.00|mL^mL^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
OBX|87|NM|152980^MDC_VOL_DELIV_SEVOFL_LIQUID_CASE^MDC|1.10.0.9|0.00|mL^mL^UCUM|||||F|||20160503150814-0600||||080019FFFE000000^GE_ANES_SYS
DOC1949974 3-12
4 Sapphire integration
Sapphire integration
DOC1949974 4-1
Aisys™ CS² Network Configuration Guide
Introduction
Protocol
Profile
Interactions
4-2 DOC1949974
4 Sapphire integration
Data model
• name/given
• name/family
• identifier/type
• identifier/id
• identifier/authority
• gender
• visit/assignedLocation/bedName
• visit/patientData/patientAge
• visit/patientData/initialHeight
• visit/patientData/initialBodySurfaceArea
• visit/patientData/initialIdealBodyWeight
The anesthesia machine may receive any data, but only processes a
small subset needed to support anesthesia machine operation.
The anesthesia machine supports a subset of the parameters given
in the Sapphire Interface User's Reference Manual. The specific
DOC1949974 4-3
Aisys™ CS² Network Configuration Guide
Sapphire communication
4-4 DOC1949974
Appendix 1 Configuration Settings
<params>
<gasDeliv>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<cfg>
<gasControlMode BT="xs:unsignedByte" enum="0=unspecified, 1=freshGas, 2=e
mergencyAir, 3=O2ByPass, 4=expiredGas, 5=monitoringOnly"/>
<ecoFlow>
<fiO2Guard BT="xs:float" inv="NaN, -Inf=na" U="%"/>
</ecoFlow>
<respGas card="0,10">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,
1=Air,2=Ar,3=Balance,4=CO2,5=He,6=HeO2,7=N2,8=N2O,9=O2"/>
<freshGasConc BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
<endTidalConc BT="xs:float" inv="NaN, -Inf=na" U="%"/>
</respGas>
<aaGas card="0,6">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,
1=DES, 2=ENF, 3=HAL, 4=ISO, 5=SEV, 6=None"/>
<freshGasConc BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
<endTidalConc BT="xs:float" inv="NaN, -Inf=na" U="%"/>
</aaGas>
<totalGas>
<freshGasFlow BT="xs:float" inv="NaN, -Inf=na" U="L/min"/>
<freshGasFlowMin BT="xs:float" inv="NaN, -Inf=na" U="L/min" min="300"
/>
</totalGas>
<driveGas BT="xs:unsignedByte" enum="0=Unspecified, 1=Air, 2=O2"/>
</cfg>
</gasDeliv>
<gasMon>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<cfg>
<numLimit>
<respGas card="0,10">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecif
ied,1=Air,2=Ar,3=Balance,4=CO2,5=He,6=HeO2,7=N2,8=N2O,9=O2"/>
<fractionalInspiredConc>
<high BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
<low BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
</fractionalInspiredConc>
<endTidalConc>
<high BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
<low BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
</endTidalConc>
</respGas>
<aaGas card="0,10">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecif
ied, 1=DES, 2=ENF, 3=HAL, 4=ISO, 5=SEV"/>
<fractionalInspiredConc>
2098870-001 4-5
<high BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
<low BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
</fractionalInspiredConc>
<endTidalConc>
<high BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
<low BT="xs:float" inv="NaN, -Inf=na, -1.0=Off" U="%"/>
</endTidalConc>
</aaGas>
<macSumAgeCorrected>
<high BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="{unitless}"/>
<low BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="{unitless}"/>
</macSumAgeCorrected>
</numLimit>
</cfg>
</gasMon>
<resp>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<cfg>
<balanceGas BT="xs:unsignedByte" enum="0=unspecified, 1=air, 2=N2O"/>
<systemState BT="xs:unsignedByte" enum="0=Unspecified, 1=Off, 2=POST, 3=C
heckout, 4=StartCase, 5=General, 6=CalibrationProcedure, 7=CalibrationNoProcedure, 8=
Therapy, 9=SuperUser, 10=Service, 11=SoftwareUpdate, 12=Failed, 13=StandByShutdown, 1
4=StandBy, 15=MonitoringOnly"/>
<tidalVolume BT="xs:float" inv="NaN, -Inf=na" U="mL"/>
<respRate BT="xs:float" inv="NaN, -Inf=na" U="/min"/>
<mechRespRate BT="xs:float" inv="NaN, -Inf=na" U="/min"/>
<cardiacBypassMode BT="xs:unsignedByte" enum="0=Unspecified, 1=Off, 2=Pas
sive, 3=VCV"/>
<ventMode>
<type BT="xs:unsignedByte" card="1" enum="0=Unspecified, 1=CMV, 2=IMV
, 3=CSV"/>
<support BT="xs:unsignedByte" card="1" enum="0=Unspecified, 1=Pressur
e, 2=Volume"/>
<option BT="xs:unsignedByte" card="0,3" enum="0=Unspecified, 1=Assist
ed, 2=Supported, 3=VolumeTargeted"/>
</ventMode>
<IERatio>
<IERatioInsp BT="xs:float" inv="NaN, -Inf=na" U="{unitless}"/>
<IERatioExp BT="xs:float" inv="NaN, -Inf=na" U="{unitless}"/>
</IERatio>
<inspiratoryTime BT="xs:float" inv="NaN, -Inf=na" U="s"/>
<time>
<high BT="xs:float" inv="NaN, -Inf=na" U="ms"/>
<low BT="xs:float" inv="NaN, -Inf=na" U="ms"/>
</time>
<pause>
<inspiratoryTime BT="xs:float" inv="NaN, -Inf=na" U="%"/>
</pause>
<circuit>
<type BT="xs:unsignedByte" enum="0=Unspecified, 1=circle, 2=nonCircle
ACGO, 3=nonCircleSCGO"/>
</circuit>
<flowSensorType BT="xs:unsignedByte" enum="0=unspecified, 1=delite, 2=ped
4-6 2098870-001
ilite, 3=internal"/>
<patientType BT="xs:unsignedByte" enum="0=unspecified, 1=adult, 2=pediatr
ic"/>
<inspiredPressure BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<supportPressure BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<PEEP BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<biasFlow BT="xs:float" inv="NaN, -Inf=na" U="L/min"/>
<riseRate BT="xs:unsignedByte" enum="0=unspecified, 1=1, 2=2, 3=3, 4=4, 5
=5, 6=6, 7=7, 8=8, 9=9, 10=10, 11=auto"/>
<PSV>
<riseTime BT="xs:float" inv="NaN, -Inf=na" U="ms"/>
<riseRate BT="xs:unsignedByte" enum="0=unspecified, 1=1, 2=2, 3=3, 4=
4, 5=5, 6=6, 7=7, 8=8, 9=9, 10=10, 11=auto"/>
</PSV>
<trigger>
<flow BT="xs:float" inv="NaN, -Inf=na" U="L/min"/>
<pressure BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<window BT="xs:float" inv="NaN, -Inf=na" U="%{breath period}"/>
</trigger>
<endInspiratoryCycle>
<flow BT="xs:float" inv="NaN, -Inf=na" U="%{peak inspiratory flow}"/>
</endInspiratoryCycle>
<supportBackupDelayTime BT="xs:float" inv="NaN, -Inf=na" U="s"/>
<numLimit>
<airwayPressure>
<low BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="s"/>
</airwayPressure>
<apneaDelayTime>
<high BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="s"/>
</apneaDelayTime>
<expiredMinuteVolume>
<high BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="L/min"/>
<low BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="L/min"/>
</expiredMinuteVolume>
<expiredTidalVolume>
<high BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="mL"/>
<low BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="mL"/>
</expiredTidalVolume>
<pressure>
<high BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="cm[H2O]"/>
</pressure>
<respRate>
<high BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="/min"/>
<low BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="/min"/>
</respRate>
<peakPressure>
<high BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="cm[H2O]"/>
</peakPressure>
<circuitLeak>
<high BT="xs:float" inv="NaN, -Inf=na, 0.0=Off" U="L/min"/>
</circuitLeak>
</numLimit>
</cfg>
</resp>
<system>
2098870-001 4-7
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<cfg>
<language BT="xs:unsignedByte" enum="0=Unspecified, 1=English, 2=French,
3=German, 4=Spanish, 5=Italian, 6=Japanese, 7=Portuguese, 8=Dutch, 9=Chinese, 10=Fin
nish, 11=Norwegian, 12=Danish, 13=Polish, 14=Greek, 15=Czech, 16=Turkish, 17=Russian,
18=Swedish, 19=Hungarian"/>
<alarmVolume BT="xs:unsignedByte" U="%" S="10"/>
<altitude BT="xs:short" inv="-32768" U="m"/>
<configuration BT="xs:byte" U="{unitless}" enum="0=Unspecified, 1=Clinica
l, 2=Bench"/>
<ge_anesthesiaDevice>
<display>
<CO2>
<units BT="xs:byte" U="{unitless}" enum="0=Unspecified, 1=%,
2=kPa, 3=mm[Hg]"/>
</CO2>
<altitude>
<units BT="xs:byte" U="{unitless}" enum="0=Unspecified, 1=ft,
2=m"/>
</altitude>
<flow>
<units BT="xs:byte" U="{unitless}" enum="0=Unspecified, 1=L/m
in, 2=L/s, 3=mL/s"/>
</flow>
<gasSupply>
<units BT="xs:byte" U="{unitless}" enum="0=Unspecified, 1=psi
, 2=kPa, 3=bar"/>
</gasSupply>
<height>
<units BT="xs:byte" U="{unitless}" enum="0=Unspecified, 1=in,
2=cm"/>
</height>
<pressure>
<units BT="xs:byte" U="{unitless}" enum="0=Unspecified, 1=mba
r, 2=cm[H2O], 3=kPa, 4=hPa, 5=mm[Hg]"/>
</pressure>
<weight>
<units BT="xs:byte" U="{unitless}" enum="0=Unspecified, 1=lb,
2=kg"/>
</weight>
<spirometry>
<dataSource BT="xs:unsignedByte" enum="0=unspecified, 1=Vent,
2=ge_MGas"/>
</spirometry>
</display>
</ge_anesthesiaDevice>
</cfg>
</system>
</params>
4-8 2098870-001
Appendix 2 Events
<types card="0">
<baseEvt typeid="baseEvt">
<evtSN BT="xs:unsignedInt" inv="0" card="1"/>
<transitionSN BT="xs:unsignedByte" inv="0" card="1"/>
<transitionType BT="xs:unsignedByte" card="1" enum="0=unknown, 1=start, 2=upd
ate, 3=end, 4=tpoint, 5=continue, 6=reset"/>
<dttm BT="xb:NTP-64" inv="0" card="1"/>
<evtType BT="xs:unsignedByte" card="1" enum="0=unknown, 1=limitLow, 2=limitHi
gh, 3=rhythm, 4=technical, 5=measurement"/>
</baseEvt>
<alarmEvt typeref="baseEvt" typeid="alarmEvt">
<sysAlmInactState BT="xs:unsignedByte" card="1" enum="0=unknown, 1=enabled, 2
=audioPaused, 3=audioOff, 4=alarmPaused, 5=alarmOff"/>
<almInactState BT="xs:unsignedByte" card="1" enum="0=unknown, 1=enabled, 2=au
dioPaused, 3=audioOff, 4=alarmPaused, 5=alarmOff"/>
<alState BT="xs:unsignedByte" card="1" enum="0=unknown, 1=inactive, 2=active,
3=latched, 4=off"/>
<priority BT="xs:unsignedByte" card="1" enum="0=unknown, 1=high, 2=medium, 3=
low, 4=none, 5=notInAlarm, 6=info, 7=notAssigned"/>
<physioTech BT="xs:unsignedByte" card="1" enum="0=unknown, 1=physiological, 2
=technical, 3=technicalPhysiological"/>
<audioPriority BT="xs:unsignedByte" enum="0=unknown, 1=high, 2=medium, 3=low,
4=none, 5=notInAlarm, 6=info, 7=notAssigned"/>
<audioPhysioTech BT="xs:unsignedByte" enum="0=unknown, 1=physiological, 2=tec
hnical, 3=technicalPhysiological"/>
</alarmEvt>
<respAlarm typeref="alarmEvt" typeid="respAlarm">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,1=Air,2=
Ar,3=Balance,4=CO2,5=He,6=HeO2,7=N2,8=N2O,9=O2"/>
</respAlarm>
<aaAlarm typeref="alarmEvt" typeid="aaAlarm">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified, 1=DES,
2=ENF, 3=HAL, 4=ISO, 5=SEV"/>
</aaAlarm>
<deviceTestAlarm typeref="alarmEvt" typeid="deviceTestAlarm">
<result BT="xs:unsignedByte" card="1" enum="0=unspecified, 1=pass, 2=fail"/>
</deviceTestAlarm>
<respCylinderAlarm typeref="respAlarm" typeid="respCylinderAlarm">
<index BT="xs:unsignedByte" card="1" isattr="true"/>
</respCylinderAlarm>
</types>
<params>
<gasDeliv>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<evt>
<tech>
<agentLevelLost typeref="alarmEvt"/>
<airOnly typeref="alarmEvt"/>
<airPipeInvalid typeref="alarmEvt"/>
<alternateO2 typeref="alarmEvt"/>
<cassettePressureError typeref="alarmEvt"/>
2098870-001 4-9
<expiredGasControlExited typeref="alarmEvt"/>
<expiredGasControlSampleLineLeak typeref="alarmEvt"/>
<expiredGasControlSuspended typeref="alarmEvt"/>
<highEtO2Egc typeref="alarmEvt"/>
<inflowOutflowCrossCheckFailRecoverable typeref="alarmEvt"/>
<lowSupplyPressureAir typeref="alarmEvt"/>
<lowSupplyPressureN2O typeref="alarmEvt"/>
<lowSupplyPressureO2 typeref="alarmEvt"/>
<manifoldOverTemp typeref="alarmEvt"/>
<mixerRamFail typeref="alarmEvt"/>
<noVaporizorCassette typeref="alarmEvt"/>
<o2CellCalibrationError typeref="alarmEvt"/>
<o2PipeInvalid typeref="alarmEvt"/>
<outflowLimitReached typeref="alarmEvt"/>
<replaceO2Cell typeref="alarmEvt"/>
<vaporizorCassetteFailure typeref="alarmEvt"/>
<vaporizorCassetteLevelLow typeref="alarmEvt"/>
<vaporizorFailure typeref="alarmEvt"/>
</tech>
<numLimit>
<circuitO2 typeref="alarmEvt" card="0,8"/>
</numLimit>
<measure>
<agentDeliveryTest typeref="deviceTestAlarm"/>
<mixerTest typeref="deviceTestAlarm"/>
</measure>
</evt>
</gasDeliv>
<gasMon>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<evt>
<tech>
<co2MonitoringNotConnected typeref="alarmEvt"/>
<aaMonitoringNotConnected typeref="alarmEvt"/>
<o2SensorFailure typeref="alarmEvt"/>
<mGasSensorInop typeref="alarmEvt"/>
<mGasAneWarmingUp typeref="alarmEvt"/>
<mGasWarmingUp typeref="alarmEvt"/>
<mGasFailure typeref="alarmEvt"/>
<mGasOutletOccluded typeref="alarmEvt"/>
<mGasFilterBlocked typeref="alarmEvt"/>
<mGasSampleLineBlocked typeref="alarmEvt"/>
<mGasNoSampleLine typeref="alarmEvt"/>
<moduleNotCompatible typeref="alarmEvt"/>
<mGasSampleFlowDeviation typeref="alarmEvt"/>
<gasZeroing typeref="alarmEvt"/>
<spirometryZeroing typeref="alarmEvt"/>
<gasZeroError typeref="alarmEvt"/>
<spirometryZeroError typeref="alarmEvt"/>
</tech>
<numLimit>
<respGas>
<fractionalInspiredConc typeref="respAlarm" card="0,8"/>
4-10 2098870-001
<endTidalConc typeref="respAlarm" card="0,8"/>
</respGas>
<aaGas>
<fractionalInspiredConc typeref="aaAlarm" card="0,8"/>
<endTidalConc typeref="aaAlarm" card="0,8"/>
</aaGas>
<agentMixture>
<lowerThanMAC typeref="alarmEvt" card="0,8"/>
<greaterThanEqMAC typeref="alarmEvt" card="0,8"/>
</agentMixture>
<macSumAgeCorrected typeref="alarmEvt" card="0,8"/>
</numLimit>
<CO2Apnea typeref="alarmEvt" card="0,8"/>
<mGasAPNEA typeref="alarmEvt" card="0,8"/>
</evt>
</gasMon>
<resp>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<evt>
<tech>
<apneaDetectOn typeref="alarmEvt"/>
<auxOutletFail typeref="alarmEvt"/>
<backupVentilationMode typeref="alarmEvt"/>
<bagVentSwitchWrongPosition typeref="alarmEvt"/>
<bellowsEmpty typeref="alarmEvt"/>
<breathingSystemNotLatched typeref="alarmEvt"/>
<checkFlowSensors typeref="alarmEvt"/>
<controlSettingsFailure typeref="alarmEvt"/>
<deliveredVolumeMismatch typeref="alarmEvt"/>
<driveGasLost typeref="alarmEvt"/>
<expFlowSensorNotConnected typeref="alarmEvt"/>
<expFlowWithNonCircle typeref="alarmEvt"/>
<fanFailure typeref="alarmEvt"/>
<fiveVoltHourAmpsGasSupplyTransducers typeref="alarmEvt"/>
<flowSensorCalDataCorrupt typeref="alarmEvt"/>
<gasInletValve typeref="alarmEvt"/>
<inspFlowSensorNotConnected typeref="alarmEvt"/>
<inspirationStopped typeref="alarmEvt"/>
<mechanicalVentDisabled typeref="alarmEvt"/>
<mechanicalVentilationOn typeref="alarmEvt"/>
<mixerFailure typeref="alarmEvt"/>
<noPressureControlPeep typeref="alarmEvt"/>
<nonCircleSelected typeref="alarmEvt"/>
<o2FlushFailure typeref="alarmEvt"/>
<o2SensorNotConnected typeref="alarmEvt"/>
<onStandbySwitchToStandby typeref="alarmEvt"/>
<patientCircuitLeak typeref="alarmEvt"/>
<patientCircuitLeakSilenced typeref="alarmEvt"/>
<patientDisconnected typeref="alarmEvt"/>
<patientVolumeMismatchOccurred typeref="alarmEvt"/>
<peepPcvNotAvailable typeref="alarmEvt"/>
<powerControllerDcFail typeref="alarmEvt"/>
<powerSupply75C typeref="alarmEvt"/>
2098870-001 4-11
<powerSupplyFailure typeref="alarmEvt"/>
<reverseExpiratoryFlow typeref="alarmEvt"/>
<serviceCalibrationsDue typeref="alarmEvt"/>
<standbyPatientDetected typeref="alarmEvt"/>
<standbyPatientDetectionMOMode typeref="alarmEvt"/>
<subAtmosphericAirwayPressure typeref="alarmEvt"/>
<sustainedAirwayPressure typeref="alarmEvt"/>
<therapyComputerFailure typeref="alarmEvt"/>
<twelveVoltHourAmpsMgas typeref="alarmEvt"/>
<twelveVoltHourAmpsVentSiBoard typeref="alarmEvt"/>
<ventFailure typeref="alarmEvt"/>
<verifyLowVeLimit typeref="alarmEvt"/>
<volumeCompensationLocked typeref="alarmEvt"/>
<vteGreaterThanInspVte typeref="alarmEvt"/>
</tech>
<numLimit>
<airwayPressure typeref="alarmEvt" card="0,8"/>
<peakPressure typeref="alarmEvt" card="0,8"/>
<respRate typeref="alarmEvt" card="0,8"/>
<expiredTidalVolume typeref="alarmEvt" card="0,8"/>
<expiredMinuteVolume typeref="alarmEvt" card="0,8"/>
</numLimit>
<measure>
<complianceTest typeref="deviceTestAlarm"/>
<ventTest typeref="deviceTestAlarm"/>
</measure>
<expFlowWithNonCircle typeref="alarmEvt" card="0,8"/>
<pLimitReached typeref="alarmEvt"/>
<pressureSupplyAdjustMVAlarmLimit typeref="alarmEvt" card="0,8"/>
<reverseInspiratoryFlow typeref="alarmEvt" card="0,8"/>
<tidalVolumeNotAchieved typeref="alarmEvt" card="0,8"/>
<volumeApnea typeref="alarmEvt" card="0,8"/>
<volumeApneaMoreThanTwoMin typeref="alarmEvt" card="0,8"/>
<ge_anesthesiaDevice>
<systemApnea typeref="alarmEvt" card="0,8"/>
</ge_anesthesiaDevice>
</evt>
</resp>
<system>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<evt>
<tech>
<batteryChargeFailure typeref="alarmEvt"/>
<batteryFailure typeref="alarmEvt"/>
<caseFanSpeedFail typeref="alarmEvt"/>
<co2Absorber typeref="alarmEvt"/>
<cpuFanSpeedFail typeref="alarmEvt"/>
<cpuOverheat typeref="alarmEvt"/>
<lowBattery typeref="alarmEvt"/>
<lowBatteryCharge typeref="alarmEvt"/>
<noBattery typeref="alarmEvt"/>
<onBattery typeref="alarmEvt"/>
<screenLocked typeref="alarmEvt"/>
<standbySwitchToStandby typeref="alarmEvt"/>
4-12 2098870-001
<touchscreenFail typeref="alarmEvt"/>
</tech>
<ge_anesthesiaDevice>
<tech>
<aaCassetteToContinueEgc typeref="alarmEvt"/>
<acb4_096VoltAdcRef typeref="alarmEvt"/>
<acbAdcFail typeref="alarmEvt"/>
<acbClockSpeed typeref="alarmEvt"/>
<acbComFailure typeref="alarmEvt"/>
<acbDcbComFailure typeref="alarmEvt"/>
<acbProcedureComFail typeref="alarmEvt"/>
<acbSoftwareError typeref="alarmEvt"/>
<acbUnexpectedReset typeref="alarmEvt"/>
<airPressLowDuring21PercentO2 typeref="alarmEvt"/>
<bagVentSwitchWrongAcgo typeref="alarmEvt"/>
<bagVentSwitchWrongPosition typeref="alarmEvt"/>
<batteryReversedConnections typeref="alarmEvt"/>
<cardiacBypass typeref="alarmEvt"/>
<carrierBoardOverheat typeref="alarmEvt"/>
<cycling typeref="alarmEvt"/>
<cyclingAborted typeref="alarmEvt"/>
<dcbProcedureComFail typeref="alarmEvt"/>
<displayComputerRestarted typeref="alarmEvt"/>
<eepromFailure typeref="alarmEvt"/>
<flowPaused typeref="alarmEvt"/>
<monitoringOnly typeref="alarmEvt"/>
<passwordReset typeref="alarmEvt"/>
<pleaseDoCheckout typeref="alarmEvt"/>
<powerControllerComFail typeref="alarmEvt"/>
<standbyPatientDetection typeref="alarmEvt"/>
<systemPowerCycleOverdue typeref="alarmEvt"/>
<twelveHourTest typeref="alarmEvt"/>
<twelveVoltsToAcb typeref="alarmEvt"/>
<vitalCapacity typeref="alarmEvt"/>
</tech>
<measure>
<quickCheckTest typeref="deviceTestAlarm"/>
</measure>
</ge_anesthesiaDevice>
</evt>
</system>
</params>
2098870-001 4-13
Appendix 3 Numerical Measurements
<params>
<gasDeliv>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<num>
<respGas card="0,9">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,
1=Air,2=Ar,3=Balance,4=CO2,5=He,6=HeO2,7=N2,8=N2O,9=O2"/>
<flow BT="xs:float" inv="NaN, -Inf=na" U="L/min"/>
<caseConsumption BT="xs:float" inv="NaN, -Inf=na" U="L"/>
<totalConsumption BT="xs:float" inv="NaN, -Inf=na" U="L"/>
<pipelinePressure BT="xs:float" inv="NaN, -Inf=na" U="kPa"/>
<cylinder card="0,2">
<index BT="xs:unsignedByte" card="1" isattr="true"/>
<pressure BT="xs:float" inv="NaN, -Inf=na" U="kPa"/>
</cylinder>
</respGas>
<aaGas card="0,5">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,
1=DES, 2=ENF, 3=HAL, 4=ISO, 5=SEV"/>
<flow BT="xs:float" inv="NaN, -Inf=na" U="mL/h"/>
<caseConsumption BT="xs:float" inv="NaN, -Inf=na" U="mL"/>
<totalConsumption BT="xs:float" inv="NaN, -Inf=na" U="mL"/>
</aaGas>
<ecoFlow>
<recommendedO2Flow BT="xs:float" inv="NaN, -Inf=na" U="L/min"/>
</ecoFlow>
</num>
</gasDeliv>
<gasMon>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<num>
<respGas card="0,9">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,
1=Air,2=Ar,3=Balance,4=CO2,5=He,6=HeO2,7=N2,8=N2O,9=O2"/>
<fractionalInspiredConc BT="xs:float" inv="NaN, -Inf=na" U="%"/>
<endTidalConc BT="xs:float" inv="NaN, -Inf=na" U="%"/>
<inspiredEndTidalDifference BT="xs:float" inv="NaN, -Inf=na" U="%"/>
</respGas>
<aaGas card="0,5">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,
1=DES, 2=ENF, 3=HAL, 4=ISO, 5=SEV"/>
<fractionalInspiredConc BT="xs:float" inv="NaN, -Inf=na" U="%"/>
<endTidalConc BT="xs:float" inv="NaN, -Inf=na" U="%"/>
</aaGas>
<metabolicConsumption>
<volO2 BT="xs:float" inv="NaN, -Inf=na" U="mL/min"/>
<volO2PerBodyWeight BT="xs:float" inv="NaN, -Inf=na" U="mL/min/kg"/>
<volO2PerBSA BT="xs:float" inv="NaN, -Inf=na" U="mL/min/m2"/>
</metabolicConsumption>
4-14 2098870-001
<metabolicProduction>
<volCO2 BT="xs:float" inv="NaN, -Inf=na" U="mL/min"/>
<volCO2PerBodyWeight BT="xs:float" inv="NaN, -Inf=na" U="mL/min/kg"/>
<volCO2PerBSA BT="xs:float" inv="NaN, -Inf=na" U="mL/min/m2"/>
</metabolicProduction>
<expendedEnergy BT="xs:float" inv="NaN, -Inf=na" U="kcal/day"/>
<respiratoryQuotient BT="xs:float" inv="NaN, -Inf=na" U="{unitless}"/>
<macSum BT="xs:float" inv="NaN, -Inf=na" U="{unitless}"/>
<macSumAgeCorrected BT="xs:float" inv="NaN, -Inf=na" U="{unitless}"/>
</num>
</gasMon>
<resp>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<num>
<circuitO2 BT="xs:float" inv="NaN, -Inf=na" U="%"/>
<inspiredTidalVolume>
<total BT="xs:float" inv="NaN, -Inf=na" U="mL"/>
<breathTrigger BT="xs:unsignedByte" enum="0=unspecified, 1=vent, 2=pa
tient"/>
</inspiredTidalVolume>
<expiredTidalVolume>
<total BT="xs:float" inv="NaN, -Inf=na" U="mL"/>
<spontaneous BT="xs:float" inv="NaN, -Inf=na" U="mL"/>
</expiredTidalVolume>
<inspiredMinuteVolume>
<total BT="xs:float" inv="NaN, -Inf=na" U="L/min"/>
</inspiredMinuteVolume>
<expiredMinuteVolume>
<total BT="xs:float" inv="NaN, -Inf=na" U="L/min"/>
<spontaneous BT="xs:float" inv="NaN, -Inf=na" U="L/min"/>
</expiredMinuteVolume>
<respRate>
<total BT="xs:float" inv="NaN, -Inf=na" U="{breath}/min"/>
<spontaneous BT="xs:float" inv="NaN, -Inf=na" U="{breath}/min"/>
</respRate>
<respRateCO2 BT="xs:float" inv="NaN, -Inf=na" U="/min"/>
<airwayPressure BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<peakPressure>
<measured BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
</peakPressure>
<plateauPressure>
<static BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<measured BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
</plateauPressure>
<PEEP>
<dynamic>
<total BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<extrinsic BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<intrinsic BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
</dynamic>
<static>
<intrinsic BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
<extrinsic BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]"/>
2098870-001 4-15
</static>
</PEEP>
<ambientPressure BT="xs:float" inv="NaN, -Inf=na" U="mm[Hg]"/>
<resistance>
<dynamic BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]/L/s"/>
</resistance>
<compliance>
<static BT="xs:float" inv="NaN, -Inf=na" U="mL/cm[H2O]"/>
<dynamic BT="xs:float" inv="NaN, -Inf=na" U="mL/cm[H2O]"/>
</compliance>
<inspiratoryTime BT="xs:float" inv="NaN, -Inf=na" U="s"/>
<expiratoryTime BT="xs:float" inv="NaN, -Inf=na" U="s"/>
<IERatio>
<IERatioExp BT="xs:float" inv="NaN, -Inf=na" U="{unitless}" S="100.0"
/>
<IERatioInsp BT="xs:float" inv="NaN, -Inf=na" U="{unitless}" S="10.0"
/>
</IERatio>
</num>
</resp>
</params>
4-16 2098870-001
Appendix 4 Waveforms
<params>
<gasMon>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<wav>
<respGas card="0,10">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,
1=Air, 2=Ar, 3=Balance, 4=CO2, 5=He, 6=HeO2, 7=N2, 8=N2O, 9=O2"/>
<conc BT="xs:float" inv="NaN, -Inf=na" U="%" vsizeBT="10"/>
</respGas>
<aaGas card="0,6">
<gas BT="xs:unsignedByte" card="1" isattr="true" enum="0=Unspecified,
1=DES, 2=ENF, 3=HAL, 4=ISO, 5=SEV"/>
<conc BT="xs:float" inv="NaN, -Inf=na" U="%" vsizeBT="10"/>
</aaGas>
<inspExpMarker BT="xs:unsignedByte" vsizeBT="10" enum="0=none, 1=breath"/
>
</wav>
</gasMon>
<resp>
<source BT="xs:unsignedByte" card="1" enum="0=unknown, 9=ge_anesthesiaDevice,
10=ge_gasAnalyzer, 13=ge_ventilator"/>
<index BT="xs:unsignedByte"/>
<wav>
<flow BT="xs:float" inv="NaN, -Inf=na" U="L/min" vsizeBT="10"/>
<volume BT="xs:float" inv="NaN, -Inf=na" U="L" vsizeBT="10"/>
<pressure BT="xs:float" inv="NaN, -Inf=na" U="cm[H2O]" vsizeBT="10"/>
<inspExpMarker BT="xs:unsignedByte" vsizeBT="10" enum="0=none, 1=insp, 2=
exp"/>
</wav>
</resp>
</params>
2098870-001 4-17
4-18 2098870-001
5 Pre-installation Network Checklist
DOC1949974 5-1
Aisys and Aisys CS2™
Pre-installation Network Checklist
Network connections
Machines with System Software 11 can be set up to directly connect
to one device through Port 1 and up to three devices through Port 2
using the hospital network.
2 1 1
AC.25.091
1
Pre-installation Network Checklist
This form is intended to be filled out by Hospital IT personnel. This
information is to be used by an authorized GE representative during
system installation and setup.
.
□ 100 MB Full
□ 100 MB Half
□ 10 MB Full
□ 10 MB Half
□ 1 min
□ 5 min
□ 15 min
□ 100 MB Full
□ 100 MB Half
□ 10 MB Full
□ 10 MB Half
□ 1 min
□ 5 min
□ 15 min
□ 1 min
□ 5 min
□ 15 min
□ 1 min
□ 5 min
□ 15 min
Log Transfer
.
DOC1949974 6-1
Aisys™ CS² Network Configuration Guide
6-2 DOC1949974
6 Privacy and Security
Access Controls
DOC1949974 6-3
Aisys™ CS² Network Configuration Guide
Note The factory default Super User password and Service passwords
can be changed. Users are advised to change the passwords after
installation and to share the passwords with authorized and trained
personnel only.
6-4 DOC1949974
6 Privacy and Security
DOC1949974 6-5
Aisys™ CS² Network Configuration Guide
Information Protection
This section focuses on privacy and security operations and contains
information to guide in the preparation of a secure environment for
the Aisys CS2 11.00
Security operations are best implemented as part of an overall
“defense in depth” information assurance strategy and are used
throughout an Information Technology system that addresses
personnel, physical security, and technology. The layered approach
of “defense in depth” limits the risk that the failure of a single
security safeguard will allow compromise of the system.
Network Security
Serial Communication
The Aisys CS2 11.00 supports USB removable media for the
following operations:
• System software upgrade.
• Import and export of system settings.
• Password reset to factory default.
• Exporting of system logs.
The Aisys CS2 11.00 supports booting from a USB portable
media, which is a specially configured tool available only to
Service users. This tool is used for system software upgrades.
6-6 DOC1949974
6 Privacy and Security
The Aisys CS2 11.00 does not encrypt the removable media, as
only the PHI of patient age and weight is captured in the log
files.
The Aisys CS2 11.00 does not encrypt its storage, as only service
logs are captured in the log files.
The Aisys CS2 11.00 does not encrypt its storage memory, as the
system must be disassembled to access the storage memory (Flash
Card).
The Aisys CS2 11.00 contains capabilities to assure that data is not
inappropriately modified accidently or maliciously.
• System software and settings are integrity checked on every
boot using CRC. If the system integrity check fails on boot the
system enters a failed state.
• System software and settings are monitored for their integrity
using CRC over run-time every 30 seconds. If the system
software integrity fails over run-time, the system enters a failed
state.
• System software upgrade employs an integrity check
mechanism of the new software before loading it on the system.
If the integrity check of the new software fails, the system
software upgrade is aborted.
• System settings are integrity checked before being loaded on
the system. System settings are also validated for the ranges
before being loaded on the system.
De-identification Capabilities
The Aisys CS2 11.00 does not store any data on its system that can
be used to identify an individual. The Personal Information listed in
"Personal Information Collected/Disclosed by the Product" cannot be
used to identify an individual without other information.
The Personal Identification Information listed in the "Optional Patient
Identification Capabilities" can be used to identify an individual.
However, it is disabled by default and it is at the discretion of the
customer to enable Patient Identification capabilities in the system.
If the customer wants to de-identify the information sent over
network, the optional Patient Identification Information can be
DOC1949974 6-7
Aisys™ CS² Network Configuration Guide
Business Continuity
The Aisys CS2 11.00 supports the following mechanism for disaster
recovery and business continuity planning:
• System supports exporting and importing of the facility
configuration of the system.
6-8 DOC1949974
6 Privacy and Security
System Protection
The Aisys CS2 11.00 must be configured and maintained in a way
that continually protects privacy and security.
In addition to the threat mitigations built into the Aisys CS2 11.00
anesthesia machine, the customer plays a significant role in ensuring
the safe and effective operation of the device. A holistic security
defense includes the following components:
Antivirus Programs
DOC1949974 6-9
Aisys™ CS² Network Configuration Guide
Due to the cited risks, the use of anti-virus software is not part of this
product security strategy.
6-10 DOC1949974
6 Privacy and Security
DOC1949974 6-11
Aisys™ CS² Network Configuration Guide
6-12 DOC1949974
6 Privacy and Security
The Aisys CS2 11.00 has been designed with privacy and security
functionality integrated into the core design. However, privacy and
security residual risks exist that must be mitigated once the Aisys
CS2 11.00 is integrated into the intended use environment. This
section contains some risks that should be imported into the Risk
Assessment of the deployment of the Aisys CS2 11.00 for proper
mitigation.
DOC1949974 6-13
Aisys™ CS² Network Configuration Guide
6-14 DOC1949974
Index
A Network specifications
network interface flow 2-4
Access controls 6-3 network interface techncial specifications
Additional privacy and security considerations 2-4
6-13
Antivirus programs 6-9
Appendix
O
sample code 3-11, 4-5 Optional patient identification capabilities 6-12
Audit logging; Accountability 6-4 Overview
HL7 3-2
B intended audience 1-2
intended use 1-2
Business continutiy 6-8
P
D
Patient privacy consent management 6-4
Data at rest security 6-7 Personal information collected/disclosed by
Data integrity capabilities 6-7 the product 6-12
De-identification capabilities 6-7 Pre-installation network checklist 5-1
Privacy and security capabilities 6-3
H Privacy and security environment 6-2
HL7 integration 3-1 Privacy and security overview 6-2
HL7 intended use 3-2 Protocol 3-4
I R
Information protection 6-6 Removable media security 6-6
Intended audience 1-2
intended use 1-2 S
Introduction 6-1 Safety information
Cautions 1-4
M safety statements 1-4
Malicious software protection customer Warnings 1-4
responsibilities 6-9 Safety statements 1-4
Messages 3-5 Serial communication 6-6
status delay 3-3
N Symbols 1-5
System (Product) change management 6-11
network cable disconnection 3-3 System protection 6-9
network communication 2-3
Network configuration U
client devices 1-3
data output 1-3 User accounts and authentication 6-3
overview 2-2
Network security 6-6
DOC1949974 I-1
Aisys™ CS² Network Configuration Guide
I-2 DOC1949974
Aisys™ CS² Network Configuration Guide
User's Reference Manual
English
DOC1949974
10 2017 Rev B