Ultrasonic Examination Procedure
Ultrasonic Examination Procedure
BY : YUDHA ZEN
PROCEDURE FOR ULTRASONIC EXAMINATION OF WELDS
TABLE OF CONTENTS
1.0 SCOPE……………………………………………………………………………….
2.0 REFERENCE………………………………………………………………………..
IDENTIFICATION…………………………………………………………………..
10.0 COUPLANT…………………………………………………….…………………..
13.0 REPORTING………………………………………………………………………..
14.0 APPENDIX………………………………………………………………………….
1.0 SCOPE
This procedure establishes the requirement for pulse echo ultrasonic examination to
evaluate full penetration welds in steel plates, beams, turbulars and pipes. The general
requirement will be conditional to examination of all welds whilst specific changes may
be deemed necessary to suit particular dimension, geometry and material of specific
welds.
2.0 REFERENCE
All Personnel performing UT Shall be fully aware of the potential site working hazard
and take the necessary safety precaution during the time and inspection. E.g. Working
at height, confine spaces, etc.
5.1 The examination shall be conducted with an ultrasonic pulse reflection system
generating frequencies covering the nominal range of 1 MHz to 5 MHz with a
rectified ‘A’ scan display. Such equipment includes the Krautkramer USK 7.
5.2 The ultrasonic pulse-echo instrument shall provide linear vertical presentation
within +/- 5% of the full screen height for 20% to 80% of the full screen height
(base line to maximum calibrated screen height). The screen height linearity shall
be measured and recorded as per procedure set out in Appendix 1.
5.3 The ultrasonic pulse echo instrument shall utilize an amplitude control accurate
over its useful range to ± 20% of the nominal amplitude ratio, to allow
measurement of indication beyond the linear range of the vertical display on the
screen. The amplitude control linearity shall be measured and recorded as per
procedure set out in Appendix 2.
5.4 The pulse-echo instrument shall be calibrated weekly and the calibration results
filed using the format shown in Appendix 5. The equipment shall be serviced with
respect to its service manual at least once a year.
Equipment shall be checked at the beginning and end of each examination, when
examination personnel are exchanged or at anytime when malfunctioning
suspected.
5.6 For angle probes, frequencies of 2 MHz to 4 MHz with crystal size of
approximately 20mm shall normally be used. Exit point and exact angle of beam
propagation of each individual probe shall be checked and marked before
commencement of a test. Type of probes used shall include Krautkramer MWB
series.
The area to be tested shall be examined by moving the search unit over the
examination surface so as to scan the entire examination volume. As a
minimum, each pass of search unit shall overlap a minimum of 10% the
transducer dimension perpendicular to the direction of the scan.
The rate of search unit movement shall not exceed 6 in/sec (153mm/sec)
unless calibration is verified at scanning speed.
The examination volume shall be scanned with angle beam probes directed
both at right angles to the weld axis and along the weld axis. In order to
detect all possible defects, the following scanning pattern shall be utilized
as much as weld and test specimen configuration would permit.
During the movement both towards and away from the welded joint the
probe shall be done at an angle not exceeding 15 degrees from the
weld axis from both directions and sides of the weld.
Alternatively, the weld surface may be dressed flush with the parent
metal and a transverse scan be performed over the weld surface along
the weld axis.
The basic calibration block shall be a section of product form of the same
nominal size, schedule, heat treatment of material specification or
equivalent P-Number groupings as one of the materials being examined.
The block size and reflector locations shall be adequate to perform
calibration for the beam angles used. For calibration blocks for dissimilar
metal welds, the material selection shall be used on the side of the weld
from which the examination will be conducted.
7.2 Clad
Where the component materials are cladded, the block shall be cladded by
the same method as the production part.
The finish on the surface of the block shall be representative of the surface
on the components.
The basic calibration block shall be curved for materials with diameters 20”
and less. A single curved basic calibration block may be used to calibrate
the examination on surfaces in the range of curvature from 0.9 to 1.5 times
the basic calibration block diameter.
If a point on the DAC curve has moved on the sweep line more that 10% of the
sweep reading or 5% of full sweep, whichever is greater, the sweep range
calibration shall be noted and corrected. All recorded indications since the last
valid calibration check shall be re-examined with corrected calibration and their
values shall be changed on the data sheets.
If a point on the DAC curve has decreased 20% or 2 Db of its amplitude, all
data sheets since the last calibration check shall be marked void. A new
calibration shall be made and recorded and the area covered by the voided
data shall be re-examined. If any point of the DAC curve has increased more
than 20% or 2 Db of its amplitude all recorded indications since the last valid
calibration and their values shall be changed on the data sheets.
9.0 EXAMINATION OF WELDS
The base metal on each side of the weld shall be free of weld spatter, surface
irregularities or foreign matter that might interfere with the examination.
Scanning shall be performed at a gain setting at least two times the primary
reference level. Evaluation shall be performed with respect to the primary
reference level. Transducers angle include 45, 60, 70 degrees. A minimum
of three angle probes are to be used for each examination in addition to the
zero degree probe on parent metal. (where specimen thickness allows).
Scanning of butt welds should be done within half skip and full distance. In
the case of curve surfaces, the effect of curvature should be taken into
consideration in locating echo origin.
9.6 Evaluation
10.0 COUPLANT
Grease, light oil or cellulose paste shall be used as couplant. The couplant used
shall not be injurious to the material to be tested or disturb post scanning surface
treatment. The same couplant shall be used for sensitivity calibration and production
test.
11.0 DEFECT SIZING (LENGTH)
All reflections from weld defects will be maximized from all angles to find the best
angle which is compatible to the defect orientation. The defect echo will be adjusted
to the screen’s graticule. The defect echo will be adjusted to the screen’s graticule
height and sized in the following manner:
The 6 Db drop method shall be used for sizing large defects. Small defects
shall be expressed as a percentage amplitude of the back wall echo.
11.2 Shearwave
The 6 Db drop method shall be used for sizing large defects lengths whilst
the 20 Db drop method may be used for sizing small defects. For sizing
defect height, the 20 Db drop method or the maximum amplitude technique
may be used.
13.1 All tests performed shall be recorded in proper reports with the following
information included :
i. Project Title
viii. Couplant
x. Result of Test
13.2 Any defects found shall be recorded with the following information included :
i. Location of Defect
INSTRUMENT CALIBRATION
Position an angle beam search unit so that indications can be observed from both the ½ t
and ¾ t holes in a basic calibration block. Adjust the search unit position to give a 2 : 1 ratio
of amplitudes between the two indications, with the larger set at 80% of full screen height.
Without moving the search unit adjust sensitivity (gain) to successively set at the larger
indication from 100% to 20% of full screen height, in 10% increments (or 2 dB steps) and
read the smaller indication at each setting. The reading must be 50% of the larger amplitude
within 5% of full screen height. The settings and readings must be estimated to the nearest
1% of full screen. Alternatively, a straight beam search unit may be used on any calibration
block which will provide amplitude differences, with sufficient signal separation to prevent
overlapping of the signals.
APPENDIX 2
Position and angle beam search unit so that the indication from ½ t hole in basic calibration
block is peaked on the screen. With the increases and decreases in attenuation shown in
the following table, the indication must fall within the specific limits. Other convenient
reflectors from any calibration block may be used with angle or straight beam search units.
The settings and readings must be estimated to the nearest 1% of full screen.
Whichever is
applicable
NOT ACCESSIBLE
* Note :
As per specimen thickness
APPENDIX 4 SHT. 2 0F 8
0º Comp. Probe
70º,60º, 45º 70º,60º, 45º
Shear Probe Shear Probe
* Note :
As per specimen thickness
APPENDIX 4 SHT. 3 0F 8
70º,60º, 45º
Shear Probe
70º,60º, 45º
Shear Probe
* Note :
As per specimen thickness
APPENDIX 4 SHT. 4 0F 8
NOT
ACCESSIBLE
NOT
ACCESSIBLE
( NOT
ACCESSIBLE)
(NOT ACCESSIBLE)
* Note :
As per specimen thickness
APPENDIX 4 SHT. 6 OF 8
Tubular Internal
(Not Accessible)
Tubular Internal
(Not Accessible)
Tubular Internal
(Not Accessible)
Tubular Internal
(Not Accessible)
1) Calibrate the screen for the test range of 125mm using single crystal transducer
and IIW-VI Block.
2) Place the probe on 25mm thickness side and read the back wall echoes. Adjust
so that the 1st and 5th back wall echo coincide with graticule line the CRT
screen.
4) The percentage of error is calculated based on actual reading and the expected
reading.
Calculated Reading
Actual Reading
Deviation
APPENDIX 5
Name of Operator:________________________