GL-220
GL-220
GL-220
Technical specifications
General information Mixing system
Machine type Random access, open reagent & locked reagent (optional) Mixer paddle-type mixing, effective reducing of carry-over
Test speed Constant 150 tests/hour (Mono reagent or double reagent)
Test principle Colorimetric method, turbidimetry
Test method 1 point end, 2 point end, Fixed-Time, Kinetics
Calibration type Linear & nonlinear Optical system
Light source Halogen lamp
Wavelength 340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm;
Sample unit 700nm(4 more options)
Sample tray 40 sample positions Absorbance range 0-4.0 Abs
Sample tube Micro cup & test tube & blood collection tube can be used. Resolution 0.0001 Abs
( Φ 12 -13 )mm * ( 25 – 100 )mm
Sample volume 2-30μL, step by 0.1μL
Sample
" probe Liquid surface detection, timely tracking for liquid volume.
Collision avoidance
Sample barcode Sample barcode reader (optional) Operating system
Operation system Windows XP, Windows 7, Windows 8, Windows 10
Data processing Automatic calibration, test combination, effective time
management of reagents, serum index, whole process de-
Reagent unit tection, linear expansion of enzyme, blank deduction,dirty
Reagent tray 80 reagent positions (40 for R1 & 40 for R2), Non-stop cuvette memory , cross infection prevention, patient info-
cooling system with peltier pad inside, 24 hours 2℃-14℃, rmation memory and association input, automatic report
open system, could accept any reagent. audit, data fuzzy query, report statistics and printing, refe-
Reagent volume 20-300μl, step by 1μl
rence range grading, alarm information classification, user
Reagent needle Liquid surface detection, timely tracking for liquid volume.
operation privilege classification management.
Collision avoidance.
Report printing 6 formats optional, surpport custom
Regent barcode Regent barcode reader (optional)
pattern
interface TCP/IP network interface
Reaction unit
Reaction tray 44 reaction cuvettes, new amorphous optical plastics Working conditions
reaction cups, perfect transmittance. Power supply AC 220V, 50Hz;AC 110V,50/60Hz, 1000VA
Reaction volume 150μl-330μl Water consumption 6L/hour(maximum)
Reaction time 10 minutes Dimension 750mm*560mm*535mm(L*W*H)
GL-220
Temperature Peltier pad incubation system, reaction temperature Weight 50kg
(37℃ ± 0.1℃)
80 reagent positions
Excellent mixer design
Automatic liquid level sensing, anti-collision protection.
Teflon coating mixer. No water dropping (reducing carry-over)
Integrated full closed optical system, multi wavelength simultaneous detection. Excellent mixing effect with standard mixing procedure
Reaction tray
44 reaction cuvettes
Reaction volume as low as 150 μl
Stable & accurate temperature(37±0.1℃) for reaction