Fluorospot Compact Software P2
Fluorospot Compact Software P2
Load application
Exit to DOS
Fig. 3:
• Click on ’OK’ for the next screen.
Fig. 4:
• Reenter the reboot password and click on ’OK’ to start the process.
After a few seconds, a new window comes up displaying the installation progress.
• Continue with (Copying the Image Contents to the System Partition / p. 10).
• A new window appears and displays the installation progress. This will take a few min-
utes.
Load application
Exit to DOS
NOTE With the COM1 port of the FL-Compact connected to the XCU
(X5 SPC), it is possible that the mouse cursor will sometimes
stick in the upper right corner of the display.
This is because the FL-C interprets the signals from the XCU
as a “mouse command”.
To resolve this, the serial mouse at the FL-C must be disabled by the following working
steps:
• Disconnect the the COM1 cable at the XCU (X5).
¹ Temporarily the mouse will work as usual.
• Enter FL-C service mode (Activating the Service Mode / p. 13).
• Select the System / System Boot tab.
- Select Exit to OS and enter the reboot password. For this, see the corresponding
entry in the AX Password List, which is published in the CS Knowledge Base.
- Open the Windows Start Menu and select Programs /Command Prompt. A com-
mand box will appear.
- Type devmgmt.msc ↵ in the DOS command box.
¹ The Device Manager window appears (Fig. 7 / p. 11).
Fig. 7: Device Manager, scan for HW changes Fig. 8: Device Manager, Disable Serial
Ballpoint
Fig. 10:
During the SW installation process, the video modes of the graphic adapter have already
been set to the requirements of the SIMOMED monitor. Check the following:
• In the service mode, select System / System Settings / Control Panel.
• Double-click on the Display icon.
• Select the Settings task card.
• Click on the left (first) monitor symbol and make sure that a screen resolution of
1280*1024 pixels and the color quality is set to Highest (32 bit).
The timing for the reference image monitor must be corrected manually as follows: