8 Software Installation

FLUOROSPOT COMPACT Software Installation

Prepare hard disk + Load application

Load application

Exit to DOS

Fig. 2: FLUOROSPOT COMPACT Software Installation

NOTE - Prepare the hard disk + load application


This option is used if you want to install the entire applica-
tion, the operating system and an empty image partition. If
you choose this option, you will lose your medical studies.
Continue at (New Installation of Software, including Partition-
ing of the HD / p. 9)
- Load application
This option is used if you need to install only the operating
system and the application. The second partition of the HD
with the medical studies is not affected. The existing medical
studies are retained. Continue at (New Installation of Soft-
ware without Partitioning of the HD / p. 9).

FLUOROSPOT Compact AX41-060.816.09.05.02 Page 8 of 36 © Siemens, 2009


11.09 CS PS AX R/F
Software Installation 9

New Installation of Software, including Partitioning of the HD 0

• Click on "Prepare hard disk + load application"


The following screen appears with a warning about the consequences:

Fig. 3:
• Click on ’OK’ for the next screen.

Fig. 4:
• Reenter the reboot password and click on ’OK’ to start the process.
After a few seconds, a new window comes up displaying the installation progress.
• Continue with (Copying the Image Contents to the System Partition / p. 10).

New Installation of Software without Partitioning of the HD 0

The patient studies will not be deleted!


• Click on ’Load application’.
• Continue with (Copying the Image Contents to the System Partition / p. 10).

© Siemens, 2009 AX41-060.816.09.05.02 Page 9 of 36 FLUOROSPOT Compact


11.09 CS PS AX R/F
10 Software Installation

Copying the Image Contents to the System Partition 0

• A new window appears and displays the installation progress. This will take a few min-
utes.

Fig. 5: Software installation progress window


After the installation, the ’FLUOROSPOT COMPACT Software Installation’ menu will
appear again.

FLUOROSPOT COMPACT Software Installation

Prepare hard disk + Load application

Load application

Exit to DOS

Fig. 6: FLUOROSPOT COMPACT Software Installation


• Click on ’Exit to DOS’
The DOS prompt is displayed.
• Remove the ’Software Installation Disk’ from the drive and press the key combination
’Ctrl + Alt + Del’ to reboot.
• In the reboot that follows, Windows setup will be performed automatically (certain
pop-ups will appear). No user action is required.
• Then the system reboots again automatically and comes up with the FL-C patient list.

FLUOROSPOT Compact AX41-060.816.09.05.02 Page 10 of 36 © Siemens, 2009


11.09 CS PS AX R/F
Software Installation 11

Reestablishing the FL-C Configuration 1.3

Deactivating the Serial Mouse at COM1 port 0

NOTE With the COM1 port of the FL-Compact connected to the XCU
(X5 SPC), it is possible that the mouse cursor will sometimes
stick in the upper right corner of the display.
This is because the FL-C interprets the signals from the XCU
as a “mouse command”.

To resolve this, the serial mouse at the FL-C must be disabled by the following working
steps:
• Disconnect the the COM1 cable at the XCU (X5).
¹ Temporarily the mouse will work as usual.
• Enter FL-C service mode (Activating the Service Mode / p. 13).
• Select the System / System Boot tab.
- Select Exit to OS and enter the reboot password. For this, see the corresponding
entry in the AX Password List, which is published in the CS Knowledge Base.
- Open the Windows Start Menu and select Programs /Command Prompt. A com-
mand box will appear.
- Type devmgmt.msc ↵ in the DOS command box.
¹ The Device Manager window appears (Fig. 7 / p. 11).

Fig. 7: Device Manager, scan for HW changes Fig. 8: Device Manager, Disable Serial
Ballpoint

© Siemens, 2009 AX41-060.816.09.05.02 Page 11 of 36 FLUOROSPOT Compact


11.09 CS PS AX R/F
12 Software Installation

Fig. 9: Serial ballpoint properties


- After starting the Device Manager, select Mice and other pointing devices.
- Select the context menu with the right mouse button.
- Now reconnect the service PC cable at the XCU (X5).
- Select Scan for hardware changes (Fig. 7 / p. 11).
¹ With the cable connected, the serial mouse driver will be found and the cursor will
begin to move by itself.
• Use the up / down arrow keys to navigate to Mice and other pointing devices, if it is
not already highlighted.
- Use left / right arrow keys to open the submenu
• Press the Alt + Tab keys to ensure that the Device Manager window is active.
• With the arrow down key, select Microsoft Serial BallPoint.
- Press Enter.
¹ The Microsoft Serial BallPoint Properties window appears (Fig. 9 / p. 12), with
the General tab being displayed.
- Press Tab until Device usage is selected.
- Use the up / down arrow keys to select Do not use this device (disable).
- Use Tab until OK is activated.
- Press Enter.
¹ The mouse for this hardware profile will be deactivated for COM1 and the Device
Manager will display a red cross in front of the deactivated component.
(Fig. 8 / p. 11).
• Now the mouse will work properly and you can close the Device Manager window.
- Close the DOS command window.
- Select Start / Programs / startup / FLC_V5.exe and wait until the FL-C has booted
up.
- Acknowledge error messages, if applicable.

FLUOROSPOT Compact AX41-060.816.09.05.02 Page 12 of 36 © Siemens, 2009


11.09 CS PS AX R/F
Software Installation 13

Activating the Service Mode 0

NOTE English is set as the default layout for the keyboard.


(Fig. 10 / p. 13) shows the English layout for the keyboard.
This is intended to serve as an aid in finding keys if the exist-
ing keyboard does not have the English layout. Enter the last
six places of the variable XCU password! If the password is
not accepted, switch the system off and on.

Fig. 10:

VGA Settings for the Live Monitor 0

During the SW installation process, the video modes of the graphic adapter have already
been set to the requirements of the SIMOMED monitor. Check the following:
• In the service mode, select System / System Settings / Control Panel.
• Double-click on the Display icon.
• Select the Settings task card.
• Click on the left (first) monitor symbol and make sure that a screen resolution of
1280*1024 pixels and the color quality is set to Highest (32 bit).

VGA Settings for the Reference Image Monitor (option) 0

Activating the Reference Image Monitor


Activate and check the settings for the reference image monitor as follows:
• Click on the right (second) monitor symbol (1/Fig. 11 / p. 14) and make sure that the
checkbox Extend my Windows desktop on this monitor (2/Fig. 11 / p. 14) is
checked.

© Siemens, 2009 AX41-060.816.09.05.02 Page 13 of 36 FLUOROSPOT Compact


11.09 CS PS AX R/F
14 Software Installation

Fig. 11: Display Properties for Reference Image Monitor


• Make sure that a screen resolution of 1280*1024 pixels (3/Fig. 11 / p. 14) is selected
and the color quality is set to Highest (32 bit) (4/Fig. 11 / p. 14).

Correcting the timing for the reference image monitor 0

The timing for the reference image monitor must be corrected manually as follows:

FLUOROSPOT Compact AX41-060.816.09.05.02 Page 14 of 36 © Siemens, 2009


11.09 CS PS AX R/F