SuperMAX Electrometer

ADVANCED DOSIMETRY
A next generation, two channel,
reference grade electrometer
with an advanced, intuitive
touchscreen interface,
|

built-in chamber library


for real-time dose and
dose rate display, and
1 fA resolution

SUPERMAX ELECTROMETER

‡COMPREHENSIVE BUILT-IN CHAMBER LIBRARY


A versatile chamber library is built into the SuperMAX
interface, so no additional PC or software is necessary
to apply system factor and temperature/pressure cor-
Simple, intuitive rections. These factor applied measurements are even
measurement
displayed in real-time side by side with raw measurement
interface (top)
values for increased analysis capability. Over 100 different
Versatile Chamber
Library (bottom) chambers and calibrations can be stored within the
SuperMAX internal memory. Individual factors can
be applied to either or both measurement channels.

‡TWO INDEPENDENT MEASUREMENT CHANNELS ‡POWERFUL, YET EASY TO USE INTERFACE


The SuperMAX Electrometer features two measurement The 6.4” color TFT display has a built-in touchscreen to
channels with independent control over range, bias voltage, make measurement control and data entry intuitive and
and applied system factor. Both channels feature extensive enjoyable. Measurement values are shown with large
range (0.001 pA to 500.0 nA, 0.001 pC to 999.9 µC) for use type and corrected values are shown in a different color
in a wide range of applications such as cross calibration IRUHDV\YLHZLQJDQGFRQ½UPDWLRQRIVHWWLQJVHYHQIURP
between two chambers, isocenter versus off-axis and in-air a distance. An on-screen keypad and familiar pull down
versus in-water comparisons. The SuperMAX is designed menus makes entry of settings such as bias voltage,
WRPD[LPL]H¾H[LELOLW\HQFRXUDJLQJWKHIUHHGRPWR threshold detection settings, and chamber library
experiment with its measurement capabilities. data nearly effortless.

76
www.standardimaging.com
ELECTROMETERS

|
Features

SuperMAX Electrometer
Versatile Charge Collection Modes ‡ Measurement data can be exported to .csv format as easily
‡ Triggered collection with automatic start, stop, and as saving a word processing document. The included USB
saving of data ÁDVKGULYHFDQEHXVHGWRWUDQVIHUPHDVXUHPHQWGDWDWRD
‡ Threshold detection levels programmable for both high PC for further analysis.
and low range ‡ Create a database of sources and calculate projected
‡ Timed collection from 1-600 seconds, set in source strength with the Check Source Utility
1 second increments Superior Flexibility
‡ Continuous Collection with manual start and stop ‡ 8VHUXSGDWDEOHVRIWZDUHDQGÀUPZDUHDOORZVIRUIXWXUH
‡ Built-in speaker provides audible alert to completion of SuperMAX updates

|
charge collection and other operations or system events ‡ The adjustable handle accommodates comfortable operation
0HDVXUHZLWK&RQÀGHQFH and viewing angle if operator is sitting or standing
‡ Ready in less than 2 minutes after power on; The SuperMAX Applications
has been designed to minimize warm up time and make ‡ External Beam IMRT
a stable, repeatable measurement whenever called upon
‡ LDR or HDR Brachytherapy
‡ Designed to exceed AAPM, ADCL, and reference grade
‡ Two channel operations such as cross calibration between
LQVWUXPHQWVSHFLÀFDWLRQV two chambers, isocenter versus off-axis, and in-air versus
‡ Reference class electrometer according to IEC 60731 in-water comparisons
Powerful Data Management ‡ Any task where a high quality reference grade electrometer
‡ Store up to 100 different chamber/system factors in the is required
built-in chamber library, each with unique calibration

S U P E R M A X ( R E F 9 0 0 1 8 ) S P E C I F I C AT I O N S
DISPLAY RANGE CONFORMITY 93 /42 /EEC
Reference class according to IEC 60731
RATE: Low Range 0.001 pA – 500.0 pA, 1 fA resolution
High Range 0.001 nA – 500.0 nA, 1 pA resolution
ZERO DRIFT ± 0.25% of minimum effective scale reading
CHARGE: Low Range 0.001 pC – 999.9 µC, 1 fC resolution (low/high : < ± 0.001pA/< ± 0.001nA)
High Range 0.001 nC – 999.9 µC, 1 pC resolution (IEC 60731 requirement: ± 0.5%)

CHARGE COLLECTIONS ZERO SHIFT ± 0.25% of minimum effective scale reading


(low/high : < ± 0.001pA/< ± 0.001nA)
TRIGGER: $XWRPDWLFVWDUWVWRSUHVHWDQGVDYHGDWDEDVHGRQXVHUGHÀQHG (IEC 60731 requirement: ± 0.5%)
thresholds (Start: 0.2 – 9.9 pA; Stop: 0.1 – 9.8 pA)
TIMED: User set duration (Range: 1 – 600 seconds; DISPLAY 6.4" color TFT, touchscreen
Increment: 1 second)
INPUT (2) BNC two lug, triaxial connector
CONTINUOUS: Unlimited duration with manual stop
BIAS VOLTAGE Nominal ± 1000 volt bias
REAL TIME CLOCK Date and time stamp for all measurements
 IRUHDV\LGHQWLÀFDWLRQ USER SETTINGS: – 1000 to – 100, 0; 100 to 1000 (set in 1 volt increments)
ACCURACY: ± 0.3 volt
INTERNAL MEMORY Store preferences, >100 sources,
>100 chamber/system factors POWER 100-240 VAC, 0.5 A max, 50/60 Hz input to external
power supply, 9 VDC, 1.7 A power supply output to
RANGE SWITCHING User selectable — High or Low electrometer input, UL/TUL listed power supply

REPEATABILITY ± 0.1% (IEC 60731 requirement: ± 0.5%) ZEROING Automatic zero function, user activated

LONG-TERM STABILITY ± 0.5% (over one year) OUTPUT (2) USB ports

STABILIZATION TIME ± 0.1% (IEC 60731 requirement: ± 0.5% of DIMENSIONS Height: 8.1 cm, 3.2 in Width: 26.7 cm, 10.5 in
value at 1 hr for measurements taken at 15 min Length: 21.1 cm, 8.3 in Weight: 2.5 kg, 5.45 lbs
and 6 hrs)
OPTIONS
RESPONSE TIME < 3 s on high range (IEC 60731 requirement: < 3 s)
< 12 s on low range SuperMAX Accessory Kit (REF 72245)
,QFOXGHVH[WUDVW\OXVH[WUD86%ÁDVKGULYHDQGVHWRIH[WUDVFUHHQSURWHFWRUV
NON-LINEARITY ± 0.25% (IEC 60731 requirement: ± 1.0%) SuperMAX Electrometer with TNC connector (REF 90018-C)
6SHFLÀFDWLRQVVXEMHFWWRFKDQJHZLWKRXWQRWLFH

ORDERING
INFORMATION SUPERMAX ELECTROMETER REF 90018

77
800.261.4446 PH 608.831.0025