Vaisala Oil MMT330

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

MMT330 Moisture and Temperature

Transmitter Series for Oil


Features/Benefits
• Continuous on-line measurement of
moisture in oil
• Ball valve installation
- no need to shut down the process
• Incorporates Vaisala HUMICAP®
Sensor - more than 30 years of field
performance
• Ten years of experience in measuring
moisture in oil
• Excellent long-term stability
• Easy to calibrate and maintain in
the field - Compatible with Vaisala
HUMICAP® Hand-Held Moisture for
Oil Meter MM70
• NIST traceable calibration
(certificate included)
The MMT330 transmitter family offers a range of solutions for demanding moisture in oil • Analog outputs, WLAN/LAN
measurements

The Vaisala HUMICAP® Moisture MMT330 can output ppm, the average
and Temperature Transmitter Series mass concentration of water in oil.
for Oil MMT330 enables fast and Vaisala has this conversion readily
reliable detection of moisture in oil. availabe for mineral transformer oil.
The MMT330 can be used in on-line For other oils, the oil specific
moisture monitoring and as a control conversion coefficients can be
device, allowing separators and oil programmed to the transmitter if the
driers to be started only when needed. water solubility of the oil is known.
Proper monitoring saves both oil and
the environment. With the MMT330 it Graphical measurement
The display shows measurement trends,
is easy and economical to monitor the real time data and history.
trend and historical display
changes of moisture in oil. The MMT330 can be ordered with a
large numerical and graphical display
Reliable Vaisala HUMICAP® lubrication systems, hydraulic systems, with a multilingual menu. It allows
technology and transformers. the user to monitor operational data,
The MMT330 incorporates the latest measurement trends and up to 1-year
generation of the Vaisala HUMICAP® Indicates the margin to measurement history. The optional
Sensor, which is the result of ten years water saturation data logger with real-time clock makes
of field experience. It was developed The MMT330 measures moisture in it possible to generate over four years
for demanding moisture measurement oil in terms of the water activity (aw) of measured history, and zoom in on
in liquid hydrocarbons. The sensor´s and temperature (T). Water activity any desired time or time frame.
excellent chemical tolerance provides indicates directly whether there is The display alarm allows tracking of
accurate and reliable measurement a risk of free water formation. The any measured parameter, with a freely
over a wide measurement range. measurement is also independent of oil configurable low and high limit.
type and age.
For diverse applications Data collection and
and demanding conditions Water content as ppm (wireless) transfer to PC
Because of the variety of probes, the conversion The recorded measurement data can be
transmitter can be used in In addition to water activity, the viewed on the display or transferred to
MMT330 MOISTURE IN OIL

a PC with Microsoft Windows® software. connection, RS232 and RS485 can be


The transmitter can also be connected used. In addition, alarm relay option is
to a network with an optional (W)LAN available.
interface, which enables a (wireless) The MMT330 has several options for
Ethernet connection. transmitter mounting. Transmitters
are delivered pre-configured with all
Versatile outputs and easy settings installation ready.
installation
The MMT330 provides up to three
analog outputs. Galvanic isolation of The Vaisala HUMICAP® Hand-Held
supply power and analog outputs is also Moisture for Oil Meter MM70 is designed
available. For serial interface the USB for field checking MMT330 transmitters.

MMT332 For High Pressure Dimensions


Installations Dimensions in mm (inches).
Pressure range 0 ... 250 bar / 0 ... 3625 psia  


Probe diameter 12 mm / 0.5 inch 

Œ  

Installation

²
Flange 36 mm / 1.4 inch
Temperature
Measurement range -40 ... +180 °C

-
 

(-40 ... +356 °F) Œ    

The MMT332 probe is installed using a    

flange. It is for high pressure applications.

MMT337 with Small Sized


Probe
Pressure range 0 ... 10 bar / 0 ... 145 psia
)3/  OR

Hex 22m
.04

Probe diameter 12 mm / 0.5 inch (0.87)


Installation
Fitting Body R 3/8” ISO
Œ  

Fitting Body 1/2” ISO


Fitting Body NPT 1/2”
The MMT337 probe, with optional Temperature  
Swagelok connector, is ideal for tight spaces Measurement range -40 ... +180 °C
with a thread connection. The small probe (-40 ... +356 °F)  

is designed for integrating into confined


spaces with small diameter lines.

MMT338 with Probe for


Pipeline Installations
Pressure range 0 ... 40 bar / 0 ... 580 psia
Œ  

with ball valve up to 120 °C (248 °F) and R1/2" ISO7/1


Œ  

40 bar  

Adjustable length 35 ... 157/379 mm /


1.37 ... 6.2/14.9 inch  
Installation   

Fitting Body R1/2” ISO  

The MMT338 is ideal for installations in Fitting Body NPT 1/2”  

pressurized processes where the probe Ball Valve Set BALLVALVE-1


needs to be removed while the process is Sampling Cell DMT242SC2
running. The probe depth is adjustable. Temperature
Measurement range -40 ... +180 °C
(-40 ... +356 °F)
MMT330 SERIES MOISTURE
P.O. Box 26, FI-00421 IN OIL
Helsinki, FINLAND
Tel: +358 9 894 91
Fax: +358 9 8949 2485
Email: [email protected]

Technical Data www.vaisala.com

Measured values Display LCD with backlight, graphic


Water activity trend display of any parameter
Measurement range aw 0 ... 1 Display menu languages English, Chinese, Spanish, Japanese,
Accuracy (including nonlinearity, hysteresis and repeatability) French, German, Russian, Swedish, Finnish
0 ... 0.9 ±0.02
0.9 ... 1.0 ±0.03 Mechanics
Response time (90%) at +20 °C Cable bushing M20x1.5 for cable diameter 8 ... 11mm/0.31 ... 0.43”
in still oil (with stainless steel filter) 10 min. Conduit fitting 1/2”NPT
Sensor HUMICAP® Interface cable connector (optional) M12 series 8 pin (male)
option 1 with plug ( female) with 5 m / 16.4 ft black cable
Performance option 2 with plug ( female) with screw terminals
Temperature USB-RJ45 Serial Connection Cable (incl. Mi70 Link software) 219685
Measurement range Probe cable diameter 5.5 mm
MMT332 -40 ... +180 °C (-40 ... +356 °F) Probe cable lengths 2 m, 5 m or 10 m
MMT337 -40 ... +180 °C (-40 ... +356 °F) Housing material G-AlSi 10 Mg (DIN 1725)
MMT338 -40 ... +180 °C (-40 ... +356 °F) Housing classification IP 65 (NEMA 4X)
Accuracy at +20 °C (+68 °F) ± 0.2 °C (0.36 °F)
Mounting options
Operating environment
Operating temperature
for probes same as measurement ranges
for transmitter body -40 ... +60 °C (-40 ... +140 °F)
with display 0 ... +60 °C (+32 ... +140 °F)
Pressure range for probes See probe specifications

Complies with EMC standard EN61326-1, Electrical equipment for


measurement, control and laboratory use - EMC requirements;
Industrial environment. Mounting with Mounting with DIN Rail
Wall Mounting Kit Installation Kit
Inputs and outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC 50/60 Hz
Power consumption @ 20 °C (Uin 24VDC)
RS-232 max 25 mA
Uout 2 x 0 ... 1V / 0 ... 5V / 0 ... 10V max 25 mA
Iout 2 x 0...20 mA max 60 mA
display and backlight + 20 mA
Analog outputs (2 standard, 3rd optional)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0 ... 1 V, 0 ... 5 V, 0 ... 10 V Pole Installation with Mounting Rain Shield
Accuracy of analog outputs at 20 °C ± 0.05 % full scale Installation Kit with Installation Kit
Temperature dependence of the for Pole or Pipeline
analog outputs ± 0.005 %/°C full scale
External loads Dimensions
current ouputs RL < 500 ohm
0 ... 1V output RL > 2 kohm
0 ... 5V and 0 ... 10V outputs RL > 10 kohm
Max wire size 0.5 mm2 (AWG 20) stranded wires recommended
 
 

Digital outputs RS-232, RS-485 (optional)


Service connection RS-232, USB
Relay outputs 0.5 A, 250 VAC, SPDT, Potential Free (optional)
Ethernet interface (optional)
Supported standards 10/100Base-T
Connector RJ45  
Protocols Telnet
Software support Vaisala MI70 link  
WLAN interface (optional)
Ref. B210724EN rev. B 2008 - 6

Supported standard 802.11b


Antenna connector type RP-SMA
Protocols Telnet
Security WEP 64/128,WPA
Software support Vaisala MI70 link
 

 

Authentication / Encryption
Open / no encryption
Œ  

Open / WEP
WPA Pre shared key / TKIP
WPA Pre shared key / CCMP (a.k.a. WPA2)
Optional data logger with real-time clock
Logged parameters max. three with trend/min/max values
Logging interval 10 sec ( fixed)
Max. logging period 4 years 5 months HUMICAP® is a registered trademark of Vaisala.
Logged points 13,7 million points per parameter Specifications subject to change without prior notice.
Battery lifetime min. 5 years ©Vaisala Oyj