En374-4 302020.4.20
En374-4 302020.4.20
En374-4 302020.4.20
7191235025-EEC20-WBH_CR1
(re-issue dated: 27 Apr 2020)
dated 22 Apr 2020
Note: This report is issued subject to the Testing and Certification Regulations of the TÜV SÜD Group and the
General Terms and Conditions of Business of TÜV SÜD PSB Pte Ltd. In addition, this report is governed by the
terms set out within this report.
SUBJECT:
Testing of Disposable Nitrile Gloves (NBR) submitted by
Anhui Intco Medical Products Co, Ltd on 30 Mar 2020.
TESTED FOR:
Anhui Intco Medical Products Co, Ltd
No.1 Haitang Road, Suixi District Economic Development Area,
Huaibei City, Anhui Province
TEST DATE:
16 Apr 2020
DESCRIPTION OF SAMPLES:
Sample received
S/N Product Description Colour Manufacturer
(pieces)
Anhui Intco
1 Disposable Nitrile Gloves (NBR) Blue 20 Medical Products
Co, Ltd
METHOD OF TEST:
BS EN ISO 374-4:2019 Protective gloves against dangerous chemicals and micro-organisms – Part 4:
Determination of resistance to degradation by chemicals
AMENDMENT:
The following amendment was made on 27 Apr 2020:
The “METHOD OF TEST” was amended to reflect the latest test standard.
RESULTS:
REMARKS:
1. For Clause 5.3 Degradation, the test specimens will be 3 gloves and 6 specimens will be cut from each glove.
For each glove, 3 specimens will be exposed to the challenge chemical (30% Hydroxide Peroxide) and 3
specimens will be unexposed. After prepare the specimens, and exposed to 30% Hydroxide Peroxide for 1
hour, puncture the specimen and record the peak force required.
APPENDIX:
Page 2 of 3
Test Report No. 7191235025-EEC20-WBH_CR1
(re-issue dated: 27 Apr 2020)
dated 22 Apr 2020
Please note that this Report is issued under the following terms :
1. This report applies to the sample of the specific product/equipment given at the time of its testing/calibration. The results are not used to
indicate or imply that they are applicable to other similar items. In addition, such results must not be used to indicate or imply that TÜV SÜD
PSB approves, recommends or endorses the manufacturer, supplier or user of such product/equipment, or that TÜV SÜD PSB in any way
“guarantees” the later performance of the product/equipment. Unless otherwise stated in this report, no tests were conducted to determine
long term effects of using the specific product/equipment.
2. The sample/s mentioned in this report is/are submitted/supplied/manufactured by the Client. TÜV SÜD PSB therefore assumes no
responsibility for the accuracy of information on the brand name, model number, origin of manufacture, consignment or any information
supplied.
3. Nothing in this report shall be interpreted to mean that TÜV SÜD PSB has verified or ascertained any endorsement or marks from any other
testing authority or bodies that may be found on that sample.
4. This report shall not be reproduced wholly or in parts and no reference shall be made by the Client to TÜV SÜD PSB or to the report or results
furnished by TÜV SÜD PSB in any advertisements or sales promotion.
5. Unless otherwise stated, the tests were carried out in TÜV SÜD PSB Pte Ltd, No.1 Science Park Drive Singapore 118221.
July 2011
Page 3 of 3