Calibration and Verification SOP
Calibration and Verification SOP
1. Purpose
To establish a standard procedure for the calibration and verification of volumetric
glassware in compliance with ISO 8655 and ISO 4787 to ensure accuracy and reliability in
laboratory measurements.
2. Scope
This procedure applies to all volumetric glassware (pipettes, burettes, volumetric flasks,
graduated cylinders, dispensers) used in the laboratory for quantitative analysis.
3. Responsibility
- Laboratory personnel are responsible for performing calibration and recording data.
- The Quality Assurance (QA) department ensures adherence to this SOP and reviews
calibration records.
4. Definitions
- **Nominal Volume:** The volume indicated on the glassware.
- **Deviation:** The percentage difference between the measured and nominal volume.
- Calibration logbook.
- ISO 4787 water density table.
6. Procedure
3. **Dispensing Water:**
- Dispense the water into the pre-weighed collection vessel.
- Ensure complete drainage according to the pipette type (e.g., 'To Deliver' or 'To
Contain').
4. **Weighing:**
- Weigh the collection vessel with the dispensed water.
- Record the gross weight.
5. **Volume Calculation:**
- Use the formula:
- Refer to ISO 4787 for the density of water at the measured temperature.
6. **Deviation Calculation:**
- Calculate the percentage deviation using the formula:
7. **Acceptance Criteria:**
- The deviation must be within ±0.3% for pipettes or as per manufacturer specifications.
7. Documentation
1. Record the following data in the calibration logbook:
8. Frequency of Calibration
Perform calibration annually or more frequently if:
- Glassware is heavily used.
- Suspected damage or inaccuracy.
9. References
- ISO 8655: Piston-operated volumetric apparatus.
- Manufacturer specifications.
10. Appendices