0% found this document useful (0 votes)
22 views16 pages

06 Testing The Visual Field

Uploaded by

kk3457044
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
22 views16 pages

06 Testing The Visual Field

Uploaded by

kk3457044
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 16

TESTING THE VISUAL FIELD

AUTHOR

Luigi Bilotto: Brien Holden Vision Institute

PEER REVIEWERS

Timothy Wingert, University of the Incarnate Word Rosenberg School of Optometry

Maureen Hanley, The New England College of Optometry

INTRODUCTION

This chapter includes a review of:

 Testing the visual fields


o Purpose/indications
o General terminology
 Theory of visual field testing
o Units of measure
o Threshold
o Kinetic vs. Static
o Screening vs. Threshold
o Variables in VF testing
o Stimulus factors
o Response factors
o Clinical variable

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-1


Visual Fields

TESTING THE VISUAL FIELDS

Figure 5.8: Testing the visual field

Inspired by Anderson, D., Testing The Field of Vision, The C.V. Mosby Company, St-Louis, Mi, 1982.

PURPOSE/INDICATIONS

Certain characteristic VF variations, termed VF defects or VF losses, can occur as a result of abnormalities or
diseases affecting the visual pathway. Interpretation of these variations will allow the optometrist to detect, localize,
diagnose, and manage various conditions. The numerous conditions that may indicate the need for visual field
testing in an optometric setting are outlined in table 5.2.

VF testing is an integral part to the assessment of the visual, ocular and physical health of patients. A form of VF
screening test should be performed on every patient presenting for an eye examination. With any of the signs &
symptoms presented below, more elaborate VF testing procedures are needed. VF assessment is an absolutely
indispensable requirement to the practice of primary care optometry.

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-2


Visual Fields

General medical problems Ocular findings

 Vascular diseases (e.g. diabetes)  Unexplained loss of vision


 Demyelinating diseases (e.g. multiple sclerosis)  Proptosis
 Infectious diseases (e.g. syphilis, TB, etc.)  Abnormal confrontation fields
 Pupil abnormalities
Neurological problems  Motility abnormalities
 Pigment dispersion syndrome
 Unexplained headache  Exfoliation syndrome
 Unexplained orbital pain  Ocular hypertension (OHT)
 Paresthesias, numbness  Anomalous discs (FLD)
 Transient weakness of limbs  C/D asymmetry (OD vs. OS)
 Head injury  C/D asymmetry (vertical vs. horizontal)
 Notching
Ophthalmic problems  Disc hemorrhage
 Baring of the circumlinear vessel
 Diplopia  NFL defects
 Poor vision on one side of the body  Swollen ONH
 Bumping into objects on one side  Disc pallor
 Difficulty reading (inability to find or keep on line)  Retinal/choroidal disease
 Visual hallucinations
 Potentially toxic medications (e.g. chloroquine, Screening / Baseline information
ethambutol)

Table 5.2: Indications for visual field testing

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-3


Visual Fields

GENERAL TERMINOLOGY

A scotoma is an area of blindness (relative or absolute) surrounded by normal VF. An absolute scotoma is a
scotoma in which vision is entirely absent in the affected area; a relative scotoma is a VF loss in which the vision is
partially reduced or depressed with respect to the surrounding vision (Fig. 5.9). There is blindness to some stimuli
but not to others.

The term depth is used to refer to the severity of a relative scotoma. A deep scotoma indicates that the sensitivity
within it is low and that only stronger stimuli can be seen ; a shallow scotoma indicates that the sensitivity within the
scotoma is slightly reduced.

Figure 5.9: Relative scotomas and depth of relative scotomas

Figure 5.10: Characterizing scotomas

Figure 5.11: Characterizing scotomas

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-4


Visual Fields

A negative scotoma is a VF loss of which the person is unaware (e.g. the physiological blind spot); a positive
scotoma is one where the patient is aware of the visual loss.

Central VF refers to the central 30 visual area; Peripheral VF usually refers to eccentricities of greater than 30.

A central scotoma involves the fixation area (fovea) (Fig. 5.12a). A cecocentral scotoma is a VF defect that
extends from the blind spot (cecum) to the central area (fixation) (Fig. 5.12b). A paracentral scotoma involves the
area within 10 of fixation but does not include fixation (Fig. 5.12c). A pericentral scotoma surrounds fixation
without involving it (Fig. 5.12d).

Figure 5.12: Various types of scotomas

An arcuate scotoma is an arc-shaped defect that arches into the nasal field and follows the course of the retinal
nerve fiber bundles (Fig. 5.13a). Also known as Bjerrum’s, Seidel’s, “scimitar” or comet scotoma, the nerve
fiber bundle scotoma leaves the blind spot as a thin relative VF loss and becomes wider and deeper as it arches
around the fixation point usually between 10-20 and heads towards the nasal horizontal raphe. Arcuate scotomas
most often occur in glaucoma.

A nasal step is a step-like defect in the nasal field (typically associated to glaucoma), caused by asymmetrical
involvement of the retinal nerve fibers on either side of the horizontal raphe (Fig. 5.13b).

Figure 5.13: Various types of scotomas

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-5


Visual Fields

An enlarged blind spot as the name suggests indicates an apparently larger blindspot (Fig. 5.14a).

An altitudinal defect involve the 2 quadrants of the upper or lower halves of the VF (Fig. 5.14b). The defect can be
complete or incomplete (Fig. 5.14c).

Split fixation indicate a VF loss of either the upper or lower half of the VF across the horizontal through fixation.

Figure 5.14: Various types of scotomas

Hemianopsia (or hemianopia) is a loss of vision in 1/2 of the VF of one eye (unilateral hemianopsia) or of both
eyes (bilateral hemianopsia) with respect of the vertical midline.

Quadrantanopsia (or quadranopia) is a loss of vision in 1/4 of the VF of one eye (unilateral quadranopia) or of
both eyes (bilateral quadranopia) with respect of the vert. & horiz. Midlines (Fig. 5.15a).

Heteronymous refers to a hemianopic or quadrantanopic defect that occurs in opposite side of the VF. Hence a
heteronymous hemianopsia involves either both nasal (binasal hemianopsia) (Fig. 5.15e)or both temporal halves
(bitemporal hemianopsia) of the VF of the 2 eyes.

Homonymous refers to a hemianopic or quadrantanopic defect that occurs in the same side of the VF. Hence a
homonymous quadranopia involves the nasal side of the VF of one eye and the temporal side of the VF of the other
eye.

Complete and incomplete terms are used to define the extent of the hemianopic or quadranopic defect. Hence a
complete hemianopia implies a defect of the entire right or left hemifield; an incomplete hemianopia implies that a
portion of the right or left hemifield is affected.

Congruous and incongruous are terms used to denote the similarity of the defects in both eyes. A congruous
hemianopia or quadranopia is identical in size, shape and position in both eyes (Fig. 5.15c); it follows that
incongruous hemianopia or quadranopia are defects that are different in the 2 eyes (Fig. 5.15b).

A crossed defect refers particularly to a quadranopia that involves the upper field in one eye and the lower in the
other.

Macular sparing indicates that the central 5 to 15 is spared in both eyes in the presence of a homonymous
hemianopia (Fig. 5.15d).

A junctional scotoma indicates the loss of VA or VF in one eye with a superior temporal defect in the opposite eye
(Fig. 5.15e).

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-6


Visual Fields

Figure 5.15: Various types of scotomas

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-7


Visual Fields

THEORY OF VISUAL FIELD TESTING

UNITS OF MEASURE

VF testing is a measure the ability of the visual system to detect a difference in contrast between the background
and the target luminance. It is a measurement of the sensitivity of the visual system to brightness differences. VF are
therefore described in units of light brightness.

The Apostilb (asb) is the unit of luminance used to denote the amount of light reflected from a surface (1 asb =
1lumen/m2). The brightness of the target and background are measured in apostilbs. The higher the value, the
brighter the surface. To give an idea, a healthy human can under optimal conditions perceive a 1asb stimulus at the
macula.

The Log units (LU) is often used to denote the apostilb brightness of stimuli to reduce the large range of luminance
of testing instruments and the sensitivity of the visual system to a more usable form. For example, the range of the
Humphrey VF has a 10 000 - 0.08 asb. This simply reduces to a range of 5.1 LU (log 10000 - log 0.8 = 5.1).

The Decibels (dB) is used to denote sensitivity. The decibel scale is a logarithmic scale that is reciprocally related to
brightness. A 10 dB increase (1LU) indicates that the stimulus is 10X dimmer (10 dB = 1 LU  10 X). Thus the
higher the dB numeric value, the dimmer the target, the higher the sensitivity.

The dB scale is a relative scale and the dB numbers on different VF instruments do not have comparable brightness
values (table. 5.3). On a given instrument, the maximal brightness of the target is assigned a value of 0 dB. With
each LU decrease in illumination from the brightest value, an increasingly higher decibel number is obtained. Hence
the decibel scale value will represent a different apostilb value on the two different testing units if the maximal
brightness is different. For example, the Humphrey VF has a maximum target brightness of 10,000 asb, while the
Goldmann VF has a maximum of only 1,000 asb. Hence 20 dB on the Humphrey unit is equal in brightness to 10 dB
on the Goldmann unit.

E.g. Humphrey VF Goldmann VF


Brightest target 10,000 asb = 0 dB 1000 asb = 0 dB
10X (1 UL) reduced 1000 asb = 10 dB 100 asb = 10 dB
100X (1 UL) reduced 100 asb = 20 dB 10 asb = 20 dB
1000X (1 UL) reduced 10 asb = 30 dB 1 asb = 30 dB
10000X (1 UL) reduced 1 asb = 40 dB 0.1 asb = 40 dB
100000X (1 UL) reduced 0.1 asb = 50 dB

Table 5.3: Relationship between dB scale and asb on different instruments of visual field testing

Note that a sensitivity of 0 dB indicates that the maximal stimulus for a given instrument is not perceived. The
sensitivity is very low, but it is not necessarily an absolute zero sensitivity (i.e. blind area). A stronger stimulus than
what the instrument can maximally produce may be perceptible at the given tested area.

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-8


Visual Fields

THRESHOLD

VF testing is a measure of the ability of the visual system to perceive targets against their background. The visibility
of a stimulus at a given location in the VF depends on a number of factors such as stimulus intensity, size, colour,
duration, etc. If a given stimulus is not visible, one of the factors (usually size or intensity) can be adjusted until it
becomes visible.

At the borderline between visibility and invisibility, the patient’s response will be uncertain or inconsistent, so a
borderline stimulus is sometimes seen and sometimes not. This borderline stimulus can be adjusted to change the
percentage of time that it is seen by the patient.

The frequency-of-seeing-curve (Fig. 5.16) is a plot of the stimulus visibility versus the percentage of times that it is
seen. The threshold is defined as the target that is perceived 50% of the times that it is presented at a given
discrete point.

Figure 5.16: Frequency-of-seeing curve

Hence the threshold is defined as the dimmest or smallest stimulus that a person can perceive 50 % of the times
that it is presented. Threshold is therefore inversely related to sensitivity: the lower the threshold, the higher
the sensitivity. The fovea at the top of the hill of vision would have the lowest threshold (highest sensitivity) and see
the dimmest & smallest stimuli.

KINETIC VS. STATIC

In a clinical setting, it is impractical (impossible!) to establish a frequency of seeing curve for each stimulus at every
point in space. Instead, one can reasonably approximate the threshold by using a static or a kinetic approach.

The static (not moving!) approach uses a “bracketing” strategy to determine the threshold of a fixed point. The
intensity or size of the stimulus is decreased or increased in a step-wise staircase fashion until it reaches a point
adequately close to threshold. Most commonly, a 4-2 dB algorithm is used. Suprathreshold stimuli are decreased in
4dB steps until they are no longer seen. The stimulus intensity is then increased in 2 dB steps until the stimulus is
seen again. The threshold estimate is recorded as the weakest stimulus seen or the average of the weakest seen
and the weakest not seen. When this is repeated at a number of locations in the visual field, the hill of vision can be
reasonably mapped out (Fig. 5.17).

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-9


Visual Fields

Figure 5.17: Determination of threshold using the static approach

The kinetic method involves the determination of isopters by moving a constant stimulus from non-seeing areas
to seeing areas (Fig. 5.18). The stimulus used to determine the isopter can roughly be taken as the threshold
stimulus along the edge of the isopter.

Figure 5.18: Determination of threshold using the kinetic approach

The static method is sensitive & more precise than the kinetic approach. However, static testing is very laborious in
manual VF testing and for this reason kinetic is the preferred approach. In automated perimetry, however, the kinetic
method is facilitated and preferred since it provides quantifiable and analyzable data.

SCREENING VS. THRESHOLD

VF can be globally assessed using either a thresholding or screening approach. Thresholding implies that the
threshold is determined as precisely as possible as described above for a number of points in space. Each threshold
value can then be mapped on a 2 or 3-dimensional graph to numerically represent the hill of vision. Obviously this
method is more precise but more time consuming especially in manual VF testing.

Screening simply involves the use of a single stimulus to test the VF. Usually a suprathreshold target that is several
dB higher than the expected hill of vision is used. The target can be presented either statically at a number of
locations or kinetically to determine if it becomes invisible at some points. Obviously this method is much more
efficient than static. However, it is less precise and small relative defect can easily be missed (Fig. 5.19), especially if
the interval between the suprathreshold stimulus and the hill of vision is large.

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-10


Visual Fields

Figure 5.19: Determination of threshold using the kinetic approach

VARIABLES IN VF TESTING

VF results will depend considerably on a number of factors that may be related to stimulus, response or clinical
variables. The clinician must understand the effect that each element may have on VF, consider them during the
procedure and bear them in mind during VF interpretation. Table 5.4 summarizes the variables involved in VF testing
that are discussed more extensively below.

Stimulus Factors Response Factors Clinical Factors


Luminance Patient psyche Target blur
Stimulus size Instructions/examiner’s personality Spectacle correction
Duration Response criterion Pupil size
Kinetic/Static target Reaction time Media clarity
Speed of kinetic target Fixation Physical Limitation
Colour Learning curve Age
Fluctuation
Psychogenic factors
Figure 5.4: Variables involved in VF testing

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-11


Visual Fields

STIMULUS FACTORS

Luminance

The target brightness will affect its visibility. Obviously the brighter the target the more visible it is.

The background luminance, which is the brightness of the surface onto which the stimulus is projected, affects the
sensitivity by setting the level of retinal adaptation. In the photopic range (>3 asb), the cones are active and the
sensitivity is maximal (island of vision is high). In scotopic (0-0.003 asb) and mesopic (0.003-3 asb) levels, the cone
activity is reduced and the central sensitivity will be depressed (hill of vision will appear flattened). The low photopic
range is preferred in most perimeters (~31.5 asb).

The contrast, which truly denotes the light difference of a stimulus from the background, will naturally affect the VF
measurement. According to Weber’s law, the “just noticeable difference” in stimulus intensity (stimulus minus
background) divided by the background intensity is constant over a wide range of photopic conditions.

Constant = stimulus – background


background

Therefore, the stimulus will remain equally visible even if the absolute intensities of the target and background are
changed, as long as the constant in the above formula is maintained. In VF testing, the background is kept constant
and the contrast is altered by increasing the stimulus intensity.

Stimulus Size

For a given stimulus intensity, a large stimulus is more easily seen than a small one due to spatial summation.
Size, however, can be compensated for by increasing the light intensity of a stimulus. The relationship is expressed
by the following mathematical expression:

luminous intensity X (area of stimulus) k = threshold for a given location

One can therefore increase the intensity of a small stimulus to make it as visible as a bigger but dimmer stimulus.
“k” is a constant that varies with retinal position, adaptation level, individual characteristics, etc. so the relationship
can only be approximated. Although variable if necessary, the standard stimulus size used in most modern VF
testers is the Goldmann size III target (0.43 = 4mm2).

DURATION

Because of temporal summation, a static stimulus presented for 20 msec. is more visible than one presented for
10 msec. After a critical duration, however, temporal summation is complete and a stimulus presented for a longer
period of time will not become more visible. The critical duration depends on a number of target factors such as size
and luminance. In general, however, the temporal summation phenomenon begins to decline after 60msec and is
completed by 100 msec. Therefore a stimulus presented for 500 msec. will not be more visible than if presented 150
msec.

Below the critical duration, Block’s Law states:

contrast X time of presentation = constant threshold

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-12


Visual Fields

This implies that you can maintain the same threshold for a dimmer stimulus if you increase the time of presentation.
In order to avoid, this variable in VF testing, the time of presentation is best if kept above the critical duration.

The latency for saccadic movements, however, is 180-250msec. If the time of presentation exceeds the latency
period, tested subjects may attempt to fixate the presented targets. Hence the ideal time of presentation falls
between the critical duration and the latency of saccadic eye movements. Automated VF testers will have
presentation times between 100-200msec. Manual VF systems, unfortunately, have a presentation time between
0.5-1.0sec which is greater than the saccadic eye movement latency.

Kinetic vs. Static Target

Static VF are established with the use of non-moving stimuli presented a varying intensities to bracket the threshold
value. Kinetic VF are established by moving the stimuli from non-seeing areas into seeing areas until they hit the hill
of vision and are perceived. A moving target is always more visible than a static stimulus. For a similar stimulus,
therefore, the sensitivity to the static stimulus will appear lower than to the kinetic one (Fig. 5.20).

Figure 5.20: Threshold difference between static and kinetic targets

Speed of kinetic target

In kinetic VF testing, the target’s speed may introduce variability and error in the obtained isopters. A target that is
moved faster will have moved in a much further distance than one moving slower during the reaction time period.
For instance, with a 1sec. response delay, a target moving at 15/sec will be noted 13 further than a target moving
at 2/sec. The speed effect is more critical in the central 30 area. A high speed target of 15/sec and a reaction time
of 1 sec may reduce a peripheral isopter from 75 to 60, but it will reduce a central isopter from 20 to 5!

In performing kinetic VF in clinical practice, the ideal rate of a moving target is 2-4/sec.

Colour

VF can be performed using coloured targets and backgrounds. Naturally VF will vary due to a number of factors
such as selective photoreceptor testing and adaptation levels. Colour VF are not clinically widespread because of
the inconsistent and variable results obtained with the use of non-standardized parameters (colours, filters, light
sources, etc.). They are also controversial with regards to the exact nature of the psychophysical end points, the
question being whether the point of achromatic perception (target awareness) or the true chromatic recognition
(colour awareness) is established.

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-13


Visual Fields

With the advent of standardized automated VF testers, however, colour perimetry is gaining popularity and can be
used clinically. Red stimulus VF testing is typically used to measure threshold in central 10 to check for drug
toxicity (e.g. chloroquine, ethambutol). Current research may render blue-yellow VF testing useful in the early
detection of glaucoma as well.

RESPONSE FACTORS

Patient’s psyche

The state of mind is crucial to the performance of any difficult task. This is particularly true for VF testing which is in
general a very difficult procedure to complete for any individual. Factors such as mental status, anxiety/stress,
level of attentiveness, cooperation as well as intelligence may all play an important role in the results of the VF
test. Whenever possible, these variables should be qualitatively assessed, noted and kept in mind during VF
analysis.

Instructions/examiner’s personality

The instructions given to the patient and the examiner’s personality are extremely influential in the patient’s
state of mind and performance. As in any physical, psychological or psychophysical task, the quality of instructions
will play a significant role in motivating the subject to perform well, be attentive and answer the test as accurately as
possible. Additionally, proper instruction and guidance serve to establish the patient’s expectations towards the
performance by clarifying what exactly is required in performing the test and what the possible outcomes may be.

Response criterion

The response criterion, which directly influences the appearance of VF, is uncertain to many patients. A patient
with a strict criterion will be unwilling to answer unless the target is seen for sure. In comparison, another individual
with a more relaxed criterion may respond to a vague stimulus that is only possibly seen. Once again, the
examiner’s instruction can help minimize, but not eliminate completely, discrepancies between response criterion.

Reaction time

Reaction time is more relevant to kinetic VF testing than to static methods. An increased response time will allow the
stimulus to travel a further distance before the patient signals its presence. Hence, an isopter for a given stimulus
may appear smaller.

Reaction time will depend on the retinal location and the patient’s physical and mental status. In general, the
reaction time is longer for stimuli that are presented more peripherally. However, reaction time is more critical in the
central 30 area where small time delays make a more significant difference on the VF results. A 1 sec delay on a
target that is moving at 15/sec will reduce a peripheral isopter from 75 to 60, but it will reduce a central isopter
from 20 to 5!

Although not clinically controllable, one must factor the reaction time in VF results especially in older individual
where response time is increased.

Fixation

Steady fixation is crucial to the production of accurate VF. Unstable fixation leads to increased variability and
inaccurate results while shifted fixation results in displaced isopters. Even in healthy subjects, drifts and micro eye
movements tend to increase during VF testing due to fatigue, stress and the tendency to fixate peripheral stimuli.
Fixation needs to be constantly monitored while performing VF testing.

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-14


Visual Fields

Learning curve

As with many psychophysical tests, a learning curve is observable in VF testing. VF can change and improve
dramatically during the first few times that the test is repeated. The learning effect is greatest between the first and
second time but learning may be observable at the third or fourth times as well especially with advanced automated
VF tests. Therefore, first VF fields, and perhaps second as well, should always be interpreted with caution.

Fluctuation

Short term fluctuation is the variability of the patients responses during a short lasting exam (minutes). The short
term fluctuation can make results vary by as much as +/- 3dB over a few moments in normal individuals. Short term
fluctuation increases mildly with age but it may increase dramatically in disease processes.

Long term fluctuation is the normal variability of the patients responses over several exams (days or years). Like
many physiological functions, sensitivities are not fixed over time and individuals can manifest different VF results
from one test time to another.

Psychogenic factors

Psychogenic factors such as hysteria and malingering may also result in unusual VF results. In fact, some VF results
may be pathognomonic of psychogenic conditions. The clinician should always bear in mind this possibility in the
analysis of VF results.

CLINICAL VARIABLES

Target blur

When a spot is blurred or defocused, it becomes dimmer due to its apparent decreased density and edge contrast. A
blur can therefore yield an apparently reduced sensitivity and hill of vision. For each diopter of defocus the sensitivity
decreases by approximately 1.5dB. Uncorrected refractive error will therefore adversely affect VF results. The effect
is greater in the central field and is minimal with larger targets. Central VF must therefore always be performed
with the patient fully corrected.

A non-uniform retinal topography that may exist in certain conditions (e.g. staphyloma, macular edema) can induce
areas of relative defocus and result in refractive scotomas.

Media clarity

Light scattering effects by disturbed ocular media can decrease target contrast with the background. Media
opacities can also reduce the amount of light reaching the retina. As a result, retinal sensitivity will appear reduced
and depressions (focal or generalized) may be noted on VF testing.

Spectacle correction

High spectacle corrections can result in variable VF results due to the prismatic effects produced by spectacle
lenses. High plus prescriptions used for aphakes or high hyperopes can produce a compressed VF where the blind
spot is closer to fixation and smaller (Fig. 5.21). Alternatively high myopic corrections will show expanded VF
where the blind spot is further out and bigger. Contact lenses are recommended to perform VF when high refractive
errors (> +/- 10.00D) are present.

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-15


Visual Fields

Figure 5.21: Prismatic effect of high prescription lenses on the visual field

Pupil size

The amount of light entering the eye is proportional to the area of the pupil. Miosis produces an increased threshold
by decreasing the amount of illumination reaching the retina and by changing the level of retinal dark adaptation.
Remember that in mesopic conditions the retinal sensitivity is reduced.

Miosis can also induce diffraction effects especially if the pupil diameter is less than 2.4 mm. Diffraction produces a
target blur at the retinal level that increases threshold.

A small pupil will therefore produce a generalized depression of the VF. Pupils that are less than 3mm should be
dilated before performing VF testing.

Alternatively, large pupils can induce VF changes by producing aberrations that adversely affect the light entering
the eye.

The pupil diameter, therefore, should always be documented when performing VF testing in order to keep track of
“abnormal” results that may result from pupil size.

Physical limitations

A number of physical factors can affect the VF result. Overhanging eyebrows, ptosis of upper lid and large noses are
the most common ones. Droopy lids may need to be taped up should they limit the VF significantly. Trial lens frames
and glasses can also limit the extent of visual fields. Trial lenses and glasses need to be placed as close as possible
to the eye during central VF testing and need to be removed for peripheral evaluations. In cases of large (+/-10.00D)
refractive errors, contact lenses can be used to test the peripheral VF.

Age

Aging has dramatic effects on the VF. The overall hill of vision decreases with age at a rate of ~ 0.5 dB per decade,
as if the hill of vision is sinking in the sea of blindness. The reduction is due to nerve fiber loss normally associated
with aging. Other associated aging changes such as the mental health (delayed reaction time, reduced attention
span, increased variability, fatigue, etc.), the general status (physical fatigue, diseases, instability, etc.) and ocular
conditions (miosis, ptosis, media opacities) all contribute to adversely affect VF measurements.

October 2013, UPDATED Clinical Optometric Procedures 2, Chapter 5-16

You might also like