UNIT-5 Instrumental Methods and Applications
UNIT-5 Instrumental Methods and Applications
Lecture-01
ELECTROMAGNETIC RADIATION
Electromagnetic radiation can be described in terms of a stream of mass-less particles called
photons each travelling in a wave like pattern at the speed of light.
Electromagnetic radiation consist of varying electric field and magnetic fields in mutually
perpendicular planes. The radiation propagate in the direction perpendicular to the plane of both
the fields.
ELECTROMAGNETIC SPECTRUM
If E is the total energy of a molecule, it can be expressed as the sum of translational, rotational,
vibrational and electronic contributions.
E = Etrans +Erot + Evib + Eelec
Absorption or emission of EMR causes a change in any of these types of energies.
In molecular spectroscopy, we measure the change in these energy states.
Translational energy – It is due to the overall movement of the molecule. Energy levels are not
quantized. Hence no spectroscopy.
Rotational energy – It is due to spinning of molecules about the axis passing through the centre
of gravity - Rotational Levels are quantized – Rotational spectroscopy (Microwave spectroscopy)
Vibrational energy – It is due to vibrations in molecules – Vibrational Levels are Quantized – IR
Spectroscopy (Vibrational spectroscopy)
Electronic energy – Consists of electronic levels which are quantized – UV/visible spectroscopy
(Electronic spectroscopy)
UNIT-V: INSTRUMENTAL METHODS AND APPLICATIONS
Lecture-3:
BEER-LAMBERT LAW
Lambert Law:
This law states that when monochromatic light passes through a transparent medium, the rate of
decrease in intensity with the thickness of the medium is proportional to the intensity of the light.
This is equivalent to stating that the intensity of the emitted light decreases exponentially as the
thickness of the absorbing medium increases arithmetically, or that any layer of given thickness of
the medium absorbs the same fraction of the light incident upon it.
Mathematically it can be expressed as
dI
− ∝ I
dl
dI
− =aI
dl
Where,
I=Intensity of the incident light
l=Thickness of the medium
a= Proportionality factor
dI
− = a dl
I
Now, integrating above equation,
𝐼 𝑙
dI
∫− = a ∫ 𝑑𝑙
I
𝐼𝑂 0
log 𝑒 𝐼 − log 𝑒 𝐼𝑜 = −𝑎 𝑙
𝐼
log 𝑒 = −𝑎𝑙
𝐼𝑜
𝐼
= 𝑒 −𝑎𝑙
𝐼0
So, 𝐼 = 𝐼𝑜 𝑒 −𝑎𝑙 --------------------- (1)
Beer Law:
Beer’s law states that the intensity of a beam of monochromatic light decreases exponentially as
the concentration of the absorbing substance increases arithmetically.
𝐼 = 𝐼𝑜 𝑒 −𝑘𝑐𝑙
𝐼
= 𝑒 −𝑘𝑐𝑙
𝐼0
𝐼
𝑙𝑜𝑔10 𝐼 = −𝜀𝑐𝑙
0
𝐼
A = log10 𝐼𝑜 = ɛ𝑐𝑙 ------------------------ (3)
Here ɛ= molar absorption coefficient Equation (3) often known as beer-lambert law.
Limitation of Beer-Lamber’s law:
The electromagnetic radiation should be monochromatic.
The light beam should not be scattered.
The solution should be diluted.
Applications of Beer-Lambert law:
In electromagnetic spectrum, which is falling in the region 100-400 nm are called UV rays.
In electromagnetic spectrum, which is falling in the region 400-700 nm are called visible region.
UV Visible spectroscopy is a type of absorption spectroscopy in which EMR of region (100-700
nm) is absorbed by the molecule which results in the excitation of the electrons from the ground
state to a higher energy state.
ELECTRONIC TRANSITIONS
σ – σ* transition: In this transition, the electrons present in σ bonding orbital (lower energy)
excited to σ* anti bonding orbital (higher energy). This transition requires higher energy. This
transition is only observed in compound containing σ bond / electrons and not lone pair electrons
on any atoms.
E.g. Transition in saturated compounds like CH4, C2H6 etc.
The absorption band occurs in the far ultraviolet region in the range of 100 to 135 nm Wavelength.
π – π* transition: In this transition, the π electrons in bonding orbital’s (lower energy) are excited
to π* anti bonding orbital (higher energy).
This transition is only observed in double bonded compounds containing π electrons.
The compound containing isolated double bond shows large absorption in the range of 160- 175
nm. This band is called as E band. The compound containing conjugated double bond shows very
large absorption in the range of 210 -280 nm. This band is called as K band.
The aromatic and hetero aromatic compounds absorbed at 220 – 270 nm Wavelength. This band
is called as B band.
n - π* transition: In this transition nonbonding or lone pair of electrons are excited from
nonbonding orbital to π* anti-bonding orbital.
This transition is observed in unsaturated compound containing double bond and lone pair of
electrons on one atom. e.g. C = O, C = N, C = S etc.
The absorption band occurs in region of 270 – 320 nm. This band is called as R band.
e.g. Aldehyde, ketone shows absorption in the range of 270 – 320 nm but unsaturated Aldehyde
and ketone are absorbs at 300 – 350 nm wavelength.
UNIT-V: INSTRUMENTAL METHODS AND APPLICATIONS
n - σ* transition: In this transition nonbonding or lone pair of electrons are excited from
nonbonding orbital to σ* anti bonding orbital. This transition is only observed in compound
containing σ bond and lone pair electrons on at least one atom. e.g. C – O, C – N, C – S
This band is observed in near UV region (180 – 225) nm.
σ- π* and π - σ*transition: These transitions are forbidden transitions and are only theoretically
possible. The relative energies of excitations are below as,
Selection Rules: The various electronic transitions which are governed by certain restrictions are
called selection rules. These are:
(i) The transitions which involve a change in the spin quantum number of an electron during the
transition do not occur. Thus, singlet-triplet transitions are forbidden.
Light Source:
Tungsten filament lamps and Hydrogen-Deuterium lamps are the most widely used and suitable
light sources as they cover the whole UV Visible region.
Hydrogen - Deuterium Discharge Lamp (200-400 nm)
Tungsten Filament lamp (400 - 750nm)
Monochromator:
It is a device used to resolve wide band of Polychromatic light radiation into narrow band of
Monochromatic radiations. Ex: prisms, Gratings etc
Beam splitter:
Splits the light from the source into two beams. One of the two divided beams is passed through
the sample solution and the second beam is passed through the reference solution.
UNIT-V: INSTRUMENTAL METHODS AND APPLICATIONS
Applications of UV Spectroscopy
• Detection of Impurities
• Structure elucidation of organic compounds
• Quantitative analysis
• Qualitative analysis
• Molecular weight determination
• Distinction between Cis & Trans isomerism
• Detection of functional group
• Kinetics of reaction
• Effect of Conjugation
• UV spectrophotometer may be used as a detector for HPLC.
UNIT-V: INSTRUMENTAL METHODS AND APPLICATIONS
Lecture-06
IR SPECTROSCOPY
In electromagnetic spectrum, which is falling in the region 12500 cm-1 to 20 cm-1 are called
Infrared region.
The range of infrared region is 12500 ~ 20 cm-1 and can be divided into
Near-infrared region (12500 ~ 4000 cm-1),
Mid-infrared region (4000 ~ 400 cm-1)
Far-infrared region (400 ~ 20 cm-1).
There are two types of fundamental molecular vibrations viz., stretching (change in bond
length) and bending (change in bond angle).
Stretching vibrations: In stretching vibrations the atoms move along the bond axis. As a
result, the bond length increases or decreases but bond angle remains unchanged.
Stretching vibrations are of two types’ viz., symmetrical Stretching & asymmetrical stretching.
Symmetric stretching: In this type the atoms of the molecule move in the same direction.
Asymmetrical stretching: In this type the atoms of the molecule move in the opposite
direction.
Bending vibrations: Bending vibrations involves a change in the bond angle whereas the bond
length remains unchanged
There are two types of bending vibrations:
a) In - plane bending vibrations:
Scissoring: In this type, the atoms move away and come close to each other in the same
plane just like the blades of a scissor.
Rocking: In this type the movement of atoms takes place in the same direction.
UNIT-V: INSTRUMENTAL METHODS AND APPLICATIONS
Homonuclear diatomic molecules are inactive in the infrared spectrum. They do not have a
dipole moment and during the vibration also the dipole moment is zero. eg: H2, O2, N2 etc.
Symmetrical polyatomic molecules such as CO2, the symmetric stretching vibration is infrared
inactive whereas the asymmetric stretching vibration is IR active ∆ν = ± 1, transition can take
place between Adjacent vibrational levels, 0 to 1, 1 to 2 etc.
UNIT-V: INSTRUMENTAL METHODS AND APPLICATIONS
Lecture-07
INSTRUMENTATION OF INFRARED SPECTROPHOTOMETER
Working Principle Irradiation of sample with IR radiation brings about vibrational changes
in molecules. The transition of molecule is from lower vibrational energy level to higher
vibrational energy level. The transition is induced by absorption of photon of the IR radiation
of appropriate frequency, which matches with energy gap between the two levels.
IR absorption by molecules happens only when there is a change in the dipole moment of the
molecule.
Radiation source: The Nernst glower and globar are the most common source of
radiation. Globar is a silicon carbide rod when heated electrically at 1200-2000°C, it
glows and produces IR radiations.
Monochromator: The radiation source emits radiations of various frequencies. As the
sample absorbs only at certain frequencies, it is therefore necessary to select desired
frequencies from the radiation source. This has been achieved by monochromators.
Prisms and gratings are commonly used for this purpose.
Sample holder (Cuvette): The sample holder made up of sodium chloride or
potassium bromide. It is used to contain sample solutions as well as reference solution
because they are transparent to IR radiation.
Detector: The detectors generally convert photo signals into photo electric signals.
Thermocouples, Bolometers, thermisters, Golay cell, and pyro-electric detectors are
used for this purpose.
Amplifier: Amplifier will amplify the photoelectric signals received from the detector.
The amplified photo signals are then sent to recorder.
Recorders: This records the amplified photo electric signals received from the
amplifier and the results are given in the form of a spectra. (Functional group region &
Finger print region)
Applications of Infrared (IR) Spectroscopy
Most of the bands that indicate what functional group is present are found in the region from
4000 cm-1 to 1300 cm-1. Their bands can be identified and used to determine the functional
group of an unknown compound.
Bands that are unique to each molecule, similar to a fingerprint, are found in the fingerprint
region, from 1300 cm-1 to 400 cm-1. These bands are only used to compare the spectra of one
compound to another.
Lecture-08
CHROMATOGRAPHY
The Russian botanist Mikhail Tswett coined the term chromatography in 1906.
Definition: Chromatography is a technique for separating mixtures into their components in
order to analyze, identify, purify, and/or quantify the mixture or components is known as
chromatography.
Principle: chromatography is based on a principle of selective distribution of the different
components of a mixture between two phases, namely stationary phase and mobile phase. The
stationary phase can be a solid or liquid; while the mobile phase is a liquid or gas. When the
stationary phase is solid, the selective distribution is based on adsorption; while it is a liquid
the basis of selective distribution is partition.
Mobile phase: Mobile Phase in Chromatography is the component that moves with the
sample. It is either a gas or a liquid and is passed through the column where the components
of the mixture are absorbed.
Separated molecules The type of interaction between the stationary phase, mobile phase, and
substances contained in the mixture is the basic component effective on the separation of
molecules from each other.
Analyte:- Analyte is the substance that is to be separated from the mixture during
chromatography.
Elution :- is a process of removing adsorbed material from stationary phase by the movement
of mobile phase.
Eluent :- It is a solvent that used for separation of absorbed material from stationary phase
Elute: Elute is the fluid containing the sample that exits the chromatographic column
Chromatogram: A chromatogram is the trace generated by the detector signal and requires a
carefully controlled flow rate of the carrier gas (mobile phase) and a carefully controlled
temperature of the column (stationary phase) to yield repeatable results.
Retention time : Time takes for a particular analyte to pass through the system (from the column
inlet to the detector) under set conditions.
Principle of HPLC:
Lecture-10
INSTRUMENTATION OF HPLC
Solvent reservoir
The solvent reservoir store the solvent or mobile phase to supply to the column as necessary.
Degasser
The eluent used for LC analysis may contain gases such as oxygen that are non-visible to our
eyes. When gas is present in the eluent, this is detected as noise and causes an unstable baseline.
Degasser uses special polymer membrane tubing to remove gases. The numerous very small
pores on the surface of the polymer tube allow the air to go through while preventing any liquid
to go through the pore.
Pump:
The pump is used to deliver the mobile phase at a constant flow rate through the stationary
phase in the column.
Injector:
The injector is used to introduce the sample into the mobile phase. The sample can be injected
manually or automatically. The simplest method is to use a syringe, and the sample is introduced
to the flow of eluent.
Column:
The column contains the stationary phase, which is typically a solid material that interacts with
the sample components based on their physicochemical properties. The separation is performed
inside the column.
Column Heater:
The LC separation is often largely influenced by the column temperature. In order to obtain
repeatable results, it is important to keep consistent temperature conditions. Also for some
analysis, such as sugar and organic acid, better resolutions can be obtained at elevated
temperatures (50 to 80°C). Thus columns are generally kept inside the column oven (column
heater).
Detector:
The detector monitors the eluent from the column and produces a signal that is proportional to
the concentration of the eluted components. Separation of analytes is performed inside the
column, whereas a detector is used to observe the obtained separation.
The composition of the eluent is consistent when no analyte is present. While the presence of
analyte changes the composition of the eluent. What detector does is to measure these
differences. This difference is monitored as a form of an electronic signal. There are different
types of detectors available.
Record: The data processing unit records and analyzes the output of the detector, and produces
a chromatogram that represents the separation of the sample components.
UNIT-V: INSTRUMENTAL METHODS AND APPLICATIONS
Lecture-11
Applications of HPLC:
The HPLC has developed into a universally applicable method so that it finds its use in almost
all areas of chemistry, biochemistry, and pharmacy.
Analysis of drugs
Analysis of synthetic polymers
Analysis of pollutants in environmental analytics
Determination of drugs in biological matrices
Isolation of valuable products
Product purity and quality control of industrial products and fine chemicals
Separation and purification of biopolymers such as enzymes or nucleic acids
Advantages of HPLC
Limitations of HPLC
Cost: Despite its advantages, HPLC can be costly, requiring large quantities of expensive
organics.
HPLC does have low sensitivity for certain compounds, and some cannot be detected as they
are irreversibly adsorbed.
Volatile substances are better separated by gas chromatography.