Biochemistry Laboratory Manual
Biochemistry Laboratory Manual
1. This will be due within the first 10 minutes of the following lab.
2. You are required to work on your lab reports individually (no copying others
work)
3. Each laboratory module will require the creation of a laboratory report that
contains the following sections;
a. An appropriate Title
b. Abstract/Summary of entire lab report (no more than 200 words)
c. Introduction – Gives a brief statement of about what is known
about the project (aka background), why this is important and the
approach taken.
d. Materials & Methods – This is an important section that
describes in detail the methods you used.
e. Results – This section presents your results as obtained in your
experiments during the lab. You should present all your results in
a professional manner using images, diagrams, plots and
graphs, etc. as appropriate to present a clear impression of what
was done. Note that even negative results are important. Present
your results here but Do Not Analyze Results (save that for
Discussion section).
f. Discussion – Here you should discuss the results obtained in the
previous section. Compare results to others and provide an
analysis of what they might mean.
g. Conclusions – What is the major outcome of this experiment? No
more than 2-3 sentences.
h. References Cited – Give credit where credit is due. Make sure
you cite everything you read that others did.
4. It is highly encouraged that you take photographs (with cell phones) of
experimental results so that you can include these in your reports.
5. Some data analysis may require you to use statistical methods in
spreadsheets such as Excel. UTEP My Apps (my.apps.utep.edu) gives you
access to this software.
Withdrawal Policy: The last day for you to withdraw from any course with an automatic "W" is listed in
the current academic calendar. Please note that it is the student's responsibility
to officially withdraw from a course. We will not administratively drop anyone
after the deadline.
Class Attendance: Lab attendance is required. Attendance will be routinely taken. Students are
responsible for attending lab regularly and knowing what takes place during the
lab. This includes not only the material covered in the class, but also all
announcements, handouts, changes in the syllabus, etc. If you must miss a lab,
you need to make a special effort to learn what occurred during your absence.
It is expected that the material be read over before the topic is presented in
class. With this background, the lectures and the lab will prove to be more
meaningful.
Disability: If you have or suspect a disability and need accommodations you should contact
Disabled Student Services Office (DSSO) at 747-5148 or at [email protected] or
come by Room 106 Union East Building.
Since the schedule for each laboratory is very tight, you are expected to come to each lab
prepared. You should read each lab before coming to class and familiarize yourselves with any
calculations that will be required and procedures that you will need to master that day. If your
absence is unavoidable, it is advised that you attend a different section so you don’t miss class.
1
Lab Safety Rules
1. Always wear eye protection in the lab.
4. Only authorized personnel are to be allowed into the lab (don’t bring your friends).
6. Dress appropriately (Closed toe shoes and long trousers must be worn in the lab. Sandals
and shorts are not allowed.)
7. Clean up after yourself. Wash all your glassware and clean (disinfect) your work area.
13. Always pour strong acids/bases into water and not the other way around. If you pour water
into acid, the heat of reaction will cause the water to explode into steam, sometimes
violently, and the acid will splatter.
14. If chemicals come into contact with your skin or eyes, flush immediately with copious
amounts of water and consult and report the incident your TA.
15. Do not place backpacks or other personal items on the lab benches.
16. Keep the lab CLEAN (it should be spotless).
18. Before using an instrument, make sure you are trained properly.
19. Keep an eye out for instruments that act up. Notify someone immediately.
20. Be as careful for the safety of others as for yourself. Think before you act.
2
Commonly used Biochemical Standards
1 mL water = 1 g/cm3
Standard Conditions
T = 25°C = 298K
P = 1 atm
Water 55.6 M
H+ conc. = 10-7 M (pH=7.0)
Common Units
molarity (M, mM, μM, nM, pM, etc.)
normality (N, whose use is discouraged by IUPAC)
weight/volume (g/L, mg/mL, μg/mL, μg/μL, etc.)
percent weight or volume (%)
dilution factor (1000x, 10x, 1x, etc.)
3
Laboratory #1, Basic Pipetting and Buffer Dilutions
Objectives
1. To learn how to properly measure solution volumes using serological pipets and
micro-pipettors.
2. Learn to measure solutions of varying viscosity accurately.
3. Learn to prepare the following solutions from stock solutions.
a. 50 mL of TE Buffer (for DNA) containing
i. 10 mM Tris-HCL pH 8.0
ii. 1 mM EDTA pH 8.0
b. 50 mL Borate buffer for DNA gels
Experimental procedure
1) You will work in groups of three so select your partners. Do not start on anything until
given instructions by the TA.
3) Your TA will give you instructions on the operation of the Micropipettors and serological
pipet bulbs. Please make notes on their proper use and ask plenty of questions.
4) You will now figure out how to pipet the following volumes for a dry run. Ask TA if unsure
a. 539 μl
b. 233 μl
c. 143 μl
d. 27 μl
5) Weigh your microcentrifuge tube and write down the value on a piece of paper (2
significant figures) or on the tube itself with a sharpie.
6) You will now pipet the following into the pre-weighed microcentrifuge tube to create a
mixture (you will mix solutions A through D in the microcentrifuge tube).
a. 539 μl of solution labeled A (water)
b. 233 μl of solution labeled B (glycerol)
c. 143 μl of solution labeled C (ethanol)
d. 27 μl of solution labeled D (acetonitrile)
4
7) When you are done, close the cap tightly and re-measure the weight. Now subtract this
weight from the initial weight determined earlier. The difference will be the weight of the
solution you just made.
8) Now you need to calculate how much of each component you need to make the 50 mL
of TE. You will need to use the serological pipet and pipet bulb for this solution (for
water). Please prepare 50 mL of TE in a 50 mL conical tube for use in next weeks’ lab.
9) You need to also prepare 50 mL of the 1 X Borate buffer for next week.
10) You must answer the questions on the next page and turn them in instead of the
lab report.
12) Finally, you should wipe down your bench with ethanol provided.
5
Lab Assignment
Since this is a short lab; In lieu of a formal laboratory report, you must answer the following
questions which are due at the beginning of the next lab period (within the first 10 minutes).
(Please use only the space provided)
1. The weight of the solution in the microcentrifuge tube has some experimental error in it.
Calculate what the solution in the microcentrifuge should have weighed and compare
this to your experimental value.
3. What is the appropriate Pipettor to use for measuring out 205 μl and why? What would
be an alternative way to measure out this same value and perhaps with more precision?
4. What formula do you use to calculate how to make a buffer from a more concentrated
stock solution? For example, if you want to make 30 mL of a 25 mM MOPS buffer from a
1.5 M stock solution. How can you calculate how much of the stock solution to add?
6
Laboratory #2, Polymerase Chain Reaction
Objectives
1. To learn how to perform Polymerase Chain Reaction (PCR) for the amplification of
DNA for future recombinant protein production.
2. To analyze the results of PCR using DNA gel electrophoresis.
Introduction
This semester, you will be cloning -galactosidase. Part of the cloning procedure is to
first amplify the gene of interest so that you can have a sufficient quantity to manipulate it further
(purification, digestion of restriction sites, etc.). PCR is very important and a key component of
modern molecular biology. For your lab report, you need to find the DNA sequence for -
galactosidase online. You also should design the forward and reverse primers (include Nde I
and Xho I restriction sites) to be used in the PCR reaction. For today’s experiment, you will be
given the PCR primers and the DNA template but you need to find those sequences yourself for
the report.
Experimental procedure (Protocol begins on page 13)
The above named -galactosidase primers were designed and purchased from IDT.
TA made 100 µL of a 10 µM stock of each for you as indicated below.
Primer # Concentration of Volume of DNA Volume of water Final concentration
Stock DNA of primer
1 100 µM 10 µL 90 µL 10 µM
2 100 µM 10 µL 90 µL 10 µM
Please note that protocols with Q5 High-Fidelity DNA Polymerase may differ from protocols with
other polymerases. Conditions recommended below should be used for optimal performance.
Reaction Setup:
We recommend assembling all reaction components on ice and quickly transferring the reactions to a
thermocycler preheated to the denaturation temperature (98°C). All components should be mixed prior to
use. See table below
7
Component 25 µl Reaction 50 µl Reaction Final Concentration
Nuclease-Free Water to 25 µl to 50 µl
Notes: Gently mix the reaction. Collect all liquid to the bottom of the tube by a quick spin if necessary.
Transfer PCR tubes to a PCR machine and begin thermocycling.
Hold 4–10°C
*Use of the NEB Tm Calculator is highly recommended.
General Guidelines:
Template:
Use of high quality, purified DNA templates greatly enhances the success of PCR. Recommended
amounts of DNA template for a 50 µl reaction are as follows:
DNA AMOUNT
1. Primers:
Oligonucleotide primers are generally 20–40 nucleotides in length and ideally have a GC content of 40–60%.
Computer programs such as Primer3 can be used to design or analyze primers. The best results are typically seen
when using each primer at a final concentration of 0.5 µM in the reaction.
8
2. Mg++ and additives:
The Q5 High-Fidelity Master Mix contains 2.0 mM Mg++ when used at a 1X concentration. This is optimal for most
PCR products generated with this master mix.
3. Deoxynucleotides:
The final concentration of dNTPs is 200 μM of each deoxynucleotide in the 1X Q5 High-Fidelity Master Mix. Q5
High-Fidelity DNA Polymerase cannot incorporate dUTP and is not recommended for use with uracil-containing
primers or templates.
5. Denaturation:
An initial denaturation of 30 seconds at 98°C is sufficient for most amplicons from pure DNA templates. Longer
denaturation times can be used (up to 3 minutes) for templates that require it.
During thermocycling, the denaturation step should be kept to a minimum. Typically, a 5–10 second denaturation
at 98°C is recommended for most templates.
6. Annealing:
Optimal annealing temperatures for Q5 High-Fidelity DNA Polymerase tend to be higher than for other PCR
polymerases. The NEB T m Calculator should be used to determine the annealing temperature when using this
enzyme. Typically use a 10–30 second annealing step at 3°C above the T m of the lower Tm primer. A temperature
gradient can also be used to optimize the annealing temperature for each primer pair. For high Tm primer pairs,
two-step cycling without a separate annealing step can be used (see note 10).
7. Extension:
The recommended extension temperature is 72°C. Extension times are generally 20–30 seconds per kb for
complex, genomic samples, but can be reduced to 10 seconds per kb for simple templates (plasmid, E. coli, etc.)
or complex templates < 1 kb. Extension time can be increased to 40 seconds per kb for cDNA or long, complex
templates, if necessary. A final extension of 2 minutes at 72°C is recommended.
8. Cycle number:
Generally, 25–35 cycles yield sufficient product. For genomic amplicons, 30-35 cycles are recommended.
9. 2-step PCR:
When primers with annealing temperatures ≥ 72°C are used, a 2-step thermocycling protocol (combining
annealing and extension into one step) is possible.
9
Protocol for Class
1. Get a thin walled PCR tube from TA and label it with your group name.
2. Set up PCR reactions as follows (added in order). Total volume is 50 µL
Reaction:
1. 19.0 µl deionized and nuclease free H20
2. 25.0 µl Q5 2X Master Mix
3. 2.5 µl 10 μM Forward Primer
4. 2.5 µl 10 μM Reverse Primer
5. 1.0 µl Template
10
Lane # Sample .
1. Markers
2. PCR Reaction from one group
3. PCR Reaction from another group
4. Blank (this adds asymmetry to the gel so that you can orient it)
5. Markers
c. Run the gel at 120 Volts for about 30 minutes or until the dye front is nearing the
bottom of the gel.
d. Carefully take the gel along with the gel tray to the blue-light box to observe your
results (you probably need to turn off the room lights).
e. Take a picture of the gel for your lab report (remember to label all lanes and
molecular weight markers).
a. You can get the -galactosidase DNA sequence by going to the NCBI website.
http://www.ncbi.nlm.nih.gov/nuccore
b. Search for the nucleotide sequence by typing “-galactosidase” in the search
area.
7. Please include the restriction sites Nde I at the 5’ end of the gene and Xho I at the 3’ end
of the gene.
8. You can look at YouTube videos of PCR and primer design to get a better understanding
of what we are trying to do.
11
Laboratory #3, PCR Product Purification and Restriction Enzyme
Digest.
Objectives
1. To learn how to purify DNA.
2. To learn how to use restriction enzymes to create “sticky ends” for subsequent
ligation reactions where the gene of interest is inserted into a plasmid.
3. To analyze the results of Restriction Enzyme digests using DNA gel electrophoresis.
Introduction
Nucleic acids have the property of being able to selectively bind to silica. An excellent
source of information about the process of purifying DNA using silica is the following website.
http://bitesizebio.com/13516/how-dna-extraction-rna-miniprep-kits-work . After purification of
your nucleic acid, your purified DNA (either plasmid or PCR product) is ready for restriction
enzyme digestion. Restriction enzymes are used because they leave overhangs in one of the
strands which can be used to ligate to another strand that had also been digested with the same
restriction enzyme. This overhang adds specificity to the ligation so that the gene of interest is
inserted into a plasmid in a specific direction only.
Experimental procedure
Monarch® PCR & DNA Cleanup Kit (5 μg) Protocol (NEB #T1030)
General Guidelines:
Input amount of DNA to be purified should not exceed the binding capacity of the column (5 μg). A
starting sample volume of 20–100 μl is recommended. For smaller samples, TE can be used to adjust the
volume to the recommended volume range. Centrifugation should be carried out at 16,000 x g in a
standard laboratory microcentrifuge at room temperature.
Buffer Preparation:
Always keep all buffer bottles tightly closed when not in use.
All centrifugation steps should be carried out at 16,000 x g. (~13K RPM in a typical microcentrifuge). This
ensures all traces of buffer are eluted at each step.
1. Dilute sample with DNA Cleanup Binding Buffer according to the table below. Mix well by pipetting up
and down or flicking the tube. Do not vortex. A starting sample volume of 20–100 μl is recommended.
For smaller samples, TE can be used to adjust the volume. For diluted samples larger than 800 μl, loada
portion of the sample, proceed with Step 2, and then repeat as necessary.
12
SAMPLE TYPE RATIO OF BINDING BUFFER: EXAMPLE
SAMPLE
2. You will do the 5:1 ratio for your PCR and Vector RE Digestions
3. Insert column into collection tube and load sample onto column and close the cap. Spin for 1 minute,
then discard flow-through.
4. Re-insert column into collection tube. Add 200 μl DNA Wash Buffer and spin for 1 minute. Discarding
flow-through is optional.
6. Transfer column to a clean 1.5 mL microfuge tube. Use care to ensure that the tip of the column does
not come into contact with the flow-through. If in doubt, re-spin for 1 minute to ensure traces of salt and
ethanol are not carried over to next step.
7. Add 35 μl of 1 mM Tris Buffer to the center of the matrix. Wait for 1 minute, then spin for 1 minute to
elute DNA.
Note: Typical elution volumes are 6–20 μl. Nuclease-free water (pH 7–8.5) can also be used to elute
the DNA. Yield may slightly increase if a larger volume of DNA Elution Buffer is used, but the DNA will
be less concentrated. For larger size DNA (≥ 10 kb), heating the elution buffer to 50°C prior to use can
improve yield.
Care should be used to ensure the elution buffer is delivered onto the matrix and not the wall of the
column to maximize elution efficiency.
Normally, you would determine the concentration of the purified DNA spectrophotometrically by
checking the absorbance of the DNA at a wavelength of 280nm. However, to save time you can
just digest your entire 35 μl in the restriction enzyme digest in the next step.
Restriction Enzyme Digest of PCR Product to create overhang ends
1. Restriction Enzyme Digest of PCR Reaction Product
Amount Component
L DNA _ _ µg/µl purified PCR product
L 10X Buffer
L Water
L Enzyme #1 Xho I of 1:10 dilution in 1x Buffer
L Enzyme #2 Nde I of 1:10 dilution in 1x Buffer
L Total Volume
13
2. pET-30a Plasmid Vector
Amount Component
L DNA 1.0 µg/µl pET30a vector
L 10X Buffer
L Water
L Enzyme #1 Xho I of 1:10 dilution in 1x Buffer
L Enzyme #2 Nde I of 1:10 dilution in 1x Buffer
L Total Volume
14
Laboratory #4, Ligation and Bacterial Transformation
Objectives
Absorbance 280 nm
Example
OD260 = 0.041 and you did a 1:20 dilution, then
0.041 x 50 x 20 = 41µg/mL or 0.041µg/µl 2.632 µg total yield
15
Now that you know the DNA concentrations you can set up the Ligation Reaction
Ligation Reaction
In order to set up a reaction where you have the same number of molecule ends to ligate
together, you first need to calculate the molar concentration of each
Calculation of mole ends;
Insert = 2 ( ng DNA) = nmole
( 618 base pairs) (650 Daltons/bp)
Amount Component
L water
L Total Volume
2. Gently mix the reaction by pipetting up and down and microfuge briefly.
3. For cohesive (sticky) ends, incubate at 16°C overnight or room temperature for 20 minutes.
4. Heat inactivate at 65°C for 10 minutes.
16
5. Chill on ice and transform 1-5 μl of the reaction into 50 μl competent cells.
Transformation Protocol for BL21(DE3) Competent Cells
Adapted from https://www.neb.com/protocols/1/01/01/transformation-protocol-for-bl21-de3-
competent-cells-c2527
1. Thaw a tube of BL21(DE3) Competent E. coli cells on ice for 10 minutes. It is critical that the
tube is never removed from the ice or warmed up!
2. Add 1–5 µl containing 1 pg–100 ng of plasmid DNA to the cell mixture. Carefully flick the tube
4–5 times to mix cells and DNA. Do not vortex.
3. Keep the mixture on ice for 30 minutes. Do not mix.
4. Heat shock at exactly 42°C for exactly 10 seconds. Do not mix.
5. IMMEDIATELY place on ice for exactly 5 minutes. Do not mix.
6. Pipette 950 µl of room temperature SOC media into the mixture.
7. Place at 37°C for 60 minutes. Shake vigorously (250 rpm) or rotate.
8. Warm two LB + 100 µg/mL Ampicillin plates to 37°C.
9. Mix the cells thoroughly by flicking the tube and inverting.
10. Spread 50 & 100 µl of cells onto each selection plate with a sterile spreader and incubate
overnight at 37°C.
Thawing: Cells are best thawed on ice and DNA added as soon as the last bit of ice in the tube
disappears. Cells can also be thawed by hand, but warming above 0°C will decrease the
transformation efficiency.
Incubation of DNA with Cells on Ice: For maximum transformation efficiency, cells and DNA
should be incubated together on ice for 30 minutes. Expect a 2-fold loss in transformation
efficiency for every 10 minutes you shorten this step.
Heat Shock: Both the temperature and the timing of the heat shock step are important and
specific to the transformation volume and vessel. Using the transformation tube provided, 10
seconds at 42°C is optimal.
Outgrowth: Outgrowth at 37°C for 1 hour is best for cell recovery and for expression of
antibiotic resistance. Expect a 2-fold loss in transformation efficiency for every 15 minutes you
shorten this step. SOC gives 2-fold higher transformation efficiency than LB medium; and
incubation with shaking or rotating the tube gives 2-fold higher transformation efficiency than
incubation without shaking.
Plating: Selection plates can be used warm or cold, wet or dry without significantly affecting the
transformation efficiency. However, warm, dry plates are easier to spread and allow for the most
rapid colony formation.
17
Laboratory #5, Protein Expression and Cell Lysis
Objectives
To save time, your TA selected a colony from your plates and inoculated a liquid culture in 250
mL of media called 2XTY (see recipe below). This must be done early in the morning because it
takes the cells about 8 hours to grow to log phage and to an Optical Density (OD) of 0.6 at 600
nm. You will begin todays experiment at Cell Lysis step below.
Prepare cells and Induce protein production:
1. Prepare a 50 mL of 2XTY containing 100 µg/mL ampicillin in a baffled flask for
increased aeration.
Tryptone 16 g
Yeast Extract 10 g
NaCl 5g
Add 2 drops of 10 M NaOH to adjust the pH to 7.0
2. Pick a single isolated colony from last week’s plate and inoculate the media.
3. Incubate the culture at 37 ºC with shaking at 220 rpm for ~ 8 hours until the optical
density at 600 nm (OD600) is between 0.6-0.8.
18
4. Induce protein production after the OD600 has reached the indicated OD by adding 1 M
IPTG to a final concentration of 1mM. This is just a simple 1:1000 dilution.
5. Protein induction is to proceed for 2 hours at a reduced temperature of 25 ºC. Keep
flask shaking 220 rpm for proper aeration.
Cell Lysis
10 mM HEPES pH 7.5 10 mL of 1 M
4. Add 10 μl of Lysozyme solution to the resuspended cells and gently mix end-over-end every
minute for 30 minutes at 4 ºC.
5. Freeze at -20 ⁰ C until frozen.
6. Thaw the tube by gently mixing end-over-end every minute at room temperature until
thawed.
7. Repeat step 4-5 for a total of 3 cycles of freeze/thaw.
8. Once sample has become thick and extremely viscous then the cells are lysed. Otherwise
you need to repeat freeze/thaw until sample has become viscous.
9. Add 10 μl of DNase I solution and 60 mM MgCl2.
10. Gently mix end-over-end every minute for 30 minutes at 4 ºC. Sample should go from
viscous to liquid again as DNA is degraded.
11. Transfer sample to a high speed centrifuge tube spin down at maximum speed for 30
minutes.
12. Carefully decant the supernatant into a new microcentrifuge tube. Discard cellular debris
pellet.
13. Freeze supernatant containing protein of interest at -80 ºC for purification next week.
14. Please return all supplies and equipment to your TA.
15. Finally, you should wipe down your bench with the ethanol provided.
19
Laboratory #6, Protein Purification using Column Chromatography
Objectives
Column chromatography is a very old, classical, and powerful technique for the
purification of proteins that is still very useful today. Columns are generally made of glass or
plastic that can withstand high pressures and are in the shape of a long cylinder that can be
filled with an appropriate matrix or stationary phase. The matrix of a column is generally in the
form of a microscopic bead that resembles a powder. The beads can be made to have very
specific characteristics such that they interact with proteins in various ways. The microscopic
beads can be used to pack a column when suspended into an aqueous liquid (generally water).
Because the beads are spherical, they can pack together quite tightly but leave empty space
between them the way a container full of ping-pong balls would pack. This way liquid can still
pass over the beads and through the matrix. As the liquid passes through the matrix, proteins in
the liquid can interact with the beads or perhaps even bind to the beads. Column
chromatography of various types is available which exploits different characteristics of a protein
to purify it. These different types include;
a. Size Exclusion Chromatography (SEC) – This technique separates proteins based
on their size and shape exclusively. The column matrix contains beads that have
microscopic pores in them. The pore size can be controlled very precisely and can
be of various sizes. As proteins pass through the matrix, smaller proteins can enter
the small pores on the bead surface whereas larger proteins cannot enter the pores.
The result is that larger proteins run past the beads faster while smaller proteins are
retarded by all their interactions with the beads. The result is that proteins are
separated by their size with large proteins eluting first and small proteins eluting
later.
20
high concentration. Because NaCl completely ionizes in solution, the ions can be
used to compete with the proteins for the charged binding sites on the matrix beads
which in turn helps elute the proteins from the column.
d. Isoelectric Focusing (IEF) – This is a less commonly used technique but it involves
exploiting the surface charges on the surface of the proteins just like in ion-exchange
chromatography with a major difference. In IEF, a pH gradient is established inside a
gel matrix using charged molecules called ampholytes (zwitterions). With the pH
gradient established in the column, proteins pass through the gel until they reach the
pH that matches their isoelectric point (where their net charge is zero). At that point,
they stop migrating and “focus” at that location. Isoelectric focusing can resolve
proteins that differ in as little as 0.01 pH units.
21
Most modern column chromatography is now computer controlled and proteins are monitored
using UV absorption as they elute from the column. In today’s lab, we will use an AKTA Start
chromatography system. The results of the protein purification will be analyzed next week.
Experimental procedure
Column Chromatography
1. Your TA has thawed out your crude extract from the -80 ºC freezer from last week and has
combined the samples for loading on to the chromatography column this week.
2. Before loading the sample on the column, we must spin down insoluble debris on micro-
centrifuge at max speed for 5 minutes.
3. Dilute the sample 20-fold in Chromatography Buffer A.
4. Load the total volume onto a High Trap Q HP 5 ml Column (a strong anion exchange
column).
5. The chromatography run is completely automated (apart from sample loading) and has
been programmed to wash the column with a large volume of Chromatography Buffer A so
that any unbound proteins are completely removed.
6. The computer will monitor the UV280nm and the conductivity of the buffer coming out of the
column. This data is recorded and graphed on the screen for you.
7. -galactosidase should elute at a conductivity of about 7.0- 8.0 mS/cm and so you should
look for a corresponding UV280nm peak at about this same area.
8. Save the fractions by capping the tubes of interest and make sure to label the tubes.
9. Please return all supplies and equipment to your TA.
10. Finally, you should wipe down your bench with the ethanol provided.
22
While you wait for the Chromatography run to finish, please work on the following questions.
1. Predict the elution sequence of the listed compounds on anion exchange, cation
exchange and hydrophobic interaction chromatography at pH 7.0.
Arg Cys-Lys-Arg-Gly
Glu-Lys Leu-Phe-Ala
Phosphorylated Tyr Trp-Val-Phe
23
2. What type of column would you choose if your protein has the following
characteristics?
Characteristic Column Type to Use
The protein is rich in glutamate and has a
molecular weight of 30 kDa
The protein is rich in tryptophan and has a
molecular weight of 20 kDa
The protein is rich in lysine and has a molecular
weight of 50 kDa
The protein has a molecular weight of 150 kDa
and is isoelectric at pH 7
The protein is rich in cysteine and has a molecular
weight of 30 kDa
3. You find that your protein sample loses activity during storage but looks fine on an
SDS-PAGE. What can you do about this? Circle correct answer
4. Your protein is large and has all amino acids represented in its sequence. The
isoelectric point is exactly at the pH which you normally use to purify proteins (pH
7.5). Figure out a strategy to be able to use ion exchange chromatography with
this protein.
24
Laboratory #7, SDS-PAGE for Analytical Protein Gels
Objectives
the electric field, the SDS denatured proteins will migrate towards the anode (positively charged
electrode) and separated by the molecular sieving effect based solely on size. After the
visualization by a staining the proteins, the size of a protein can be calculated by comparing its
migration distance with that of a known molecular weight ladder (marker).
25
Approximate migration of proteins on various percentage gels.
Experimental procedure
Setting up the apparatus
1. Today’s lab will be including the use of precast gels and so you will not have to make
your own gel. First, get the following items from your TA
a. Precast 4-20% gel
b. Protein sample
c. Electrophoresis apparatus
d. Sample buffer
e. Running buffer
f. P20 Pipettor and tips
2. You need to prepare your protein for electrophoresis. You need to take your protein
and add sample buffer to it. Your TA will tell you the amounts to use.
3. Heat your sample to 100 ⁰C for 5 minutes and let it cool on your bench until needed.
4. Next you need to assemble your electrophoresis apparatus.
a. Remove the gel from the storage pouch and prepare it
for assembly
b. GENTLY remove the comb: Position thumb on the
indentation (middle of comb) and remove the comb by
pulling straight upward in one smooth motion.
c. Remove the tape: Pull gently to remove the green tape
from the bottom of the cassette. If necessary, use the
opening key or comb to help remove the tape at the
corners.
d. Rinse the wells: Use a syringe, wash bottle, or
disposable transfer pipet to rinse the wells with running
buffer. Leave buffer in the wells to prevent drying.
26
e. Set the electrode assembly to the open position on a clean, flat surface (A).
f. Place the gel cassettes into the electrode assembly. Use
the buffer dam (included with the cell) to complete the
assembly.
i. Place the first cassette with the short plate facing
inward and so the gel rests at a 30° angle away
from the center of the electrode assembly. Make
sure the electrode assembly remains balanced
and does not tip over.
ii. Place the second gel or buffer dam on the other
side of the electrode assembly, again by resting
the gel on the supports. The gels rest at 30°
angles, one on either side of the electrode
assembly, tilting away from the center of the frame
(B).
g. Gently push both gels toward each other, making sure
that they rest firmly and squarely against the green
gasket that is built into the electrode assembly. Align the short plates to ensure
the edge sits just below the notch at the top of the green gasket (C).
h. While gently squeezing the gel
cassettes (or cassette and buffer
dam) against the green gaskets
(maintaining constant pressure
and with both gels in place),
slide the green arms of the
clamping frame one at a time
over the gels, locking them into
place (D,E).
i. The wing clamps of the electrode assembly lift each gel cassette up against the
notch in the green gasket, forming a seal. Check again that the short plates sit
just below the notch at the top of the green gasket (C).
j. Place the electrophoresis module into the tank (F)
and fill the buffer chambers with 1x running buffer:
i. Fill the inner buffer chamber completely to
the top
ii. Add most if not all the remaining buffer to
the outer buffer chamber
5. Load samples and run the gels at a constant voltage
of 200V.
6. Stop the run when the dye front reaches 0.5 cm from
the bottom of the gel.
Removing and Staining the gel
1. After electrophoresis is complete, turn off the power supply and disconnect the
electrical leads.
2. Remove the lid from the tank and remove the gels from the cell. Pour off and discard
the running buffer.
27
3. To open the cassette, align the
arrow on the opening lever with
the arrows marked on the
cassette and insert the lever
between the cassette plates at
indicated locations. Apply
downward pressure to break
each seal. Do not twist the lever.
4. Pull the two plates apart from
the top of the cassette, and
gently remove the gel by teasing
it with the pipet tip and drop it into the staining solution.
a. Do not touch the gel with your hands because you will contaminate it!
b. Stain the gel for a minimum of 1 hour.
c. De-stain overnight or until desired intensity is achieved.
5. Please return all supplies and equipment to your TA.
6. Finally, you should wipe down your bench with the ethanol provided.
28
Laboratory #8, Protein Quantitation Methods (BCA Assay)
Objectives
The bicinchoninic acid (BCA) assay was developed by Paul Smith at the Pierce
Chemical Company in 1985 and is used to determine the total protein concentration in a
solution. Protein is determined by a color change from light green to purple depending on the
amount of protein present. There are two reactions that occur in the assay that result in the
color change. First, the peptide bonds of a protein will reduce Cu2+ ions to Cu+ ions at an
elevated temperature (between 40-60 ⁰C). The amount of Cu2+ ions reduced is directly
proportional to the amount of protein present. In the Second reaction, bicinchoninic acid will
chelate the resulting Cu+ ions which produces the purple colored complex BCA-Cu+. The purple
color that is developed can be quantitated using absorption of light at 562 nm in a
spectrophotometer. The unknown concentration of a protein must be compared to a known
concentration of a control protein and so a standard curve must first be generated. The
absorption of the unknown protein is then compared to the standard curve to estimate the
concentration of the unknown. The control protein that is usually chosen is bovine serum
albumin because it is a very stable and well known protein.
Experimental procedure
1. You will be working as usual in groups of three and you will need the following items
from your TA.
a. BSA standard
b. Reagent A and B for the BCA assay
c. Cuvettes for the spectrophotometer
d. A portable PASCO wireless spectrophotometer
i. Download the software from the following website (for phone laptop or
tablet)
ii. https://www.pasco.com/downloads/spectrometry/index.cfm
e. Pipettors and tips
f. 15 microcentrifuge tubes and a tube rack
g. A bottle of deionized water
h. One 50 ml conical tube.
i. Two spectrophotometer cuvettes (one for standards and one for unknowns)
2. Label your microcentrifuge tubes A-K and 1-4 as indicated in the Standards and
Unknowns table below.
3. Prepare a set of protein standards of known concentration by adding everything as listed
in the “Standards” table below and in the order listed here.
a. First pipet out the diluent (diH2O) into all the tubes.
b. Second, pipet the BSA standard (or unknown protein) into appropriate tubes.
c. Finally, prepare and pipette the working reagent.
29
4. For “Unknowns”, use 1l for tubes 1-2 and 2 l for tubes 3-4 instead of the BSA
standard.
5. Mix 180 l Reagent B with 8.8 ml Reagent A (1:50 solution immediately clears to a light
green color). Shake vigorously to mix very well.
6. Add the working reagent, cap the tubes and mix well.
7. Incubate at 60oC for 15-30 minutes.
8. After incubation, cool the tubes to room temperature, transfer to 1.5 mL disposable
cuvettes and read the absorbance at 562 nm vs. a water (or air) reference.
9. For unknowns, use 1-2 microliters (or more if you expect it to be dilute).
Standards
Tube # 2 mg/ml BSA Diluent Working A526 Protein (g)
Stock Reagent
(Y-axis) (X-axis)
A 0 l 500 l 500 l 0 g
B 1 499 500 l 2
C 2 498 500 l 4
D 3 497 500 l 6
E 4 496 500 l 8
F 5 495 500 l 10
G 6 494 500 l 12
H 7 493 500 l 14
I 8 492 500 l 16
J 9 491 500 l 18
K 10 490 500 l 20
Unknowns
Tube Sample Description Sample Diluent Working A526 Resulting
# Amount Reagent Concentration
(Bring up
to 1 mL)
g/sample amount
1 1 l 499 l 500 l
2 Duplicate of tube 1 1 l 499 l 500 l
3 2 l 498 l 500 l
4 Duplicate of tube 3 2 498 l 500 l
10. To analyze the results, download the excel spreadsheet from blackboard and input all of
your absorbance values. The spreadsheet automatically creates a plot of your standard
curve using linear regression to generate a best fit line through your data. It will also
automatically calculate your unknown protein concentration by automatically averaging
duplicates and adjusts for the amount of protein added.
11. Make sure you print out the spreadsheet results (entire page) for your lab report.
12. Please return all supplies and equipment to your TA.
13. Finally, you should wipe down your bench with the ethanol provided.
30
Pasco Wireless Spectrophotometer Instructions for Use
2. Calibration
a. Once you plug the spectrometer into your computer using the USB cable or
wirelessly associate the spectrometer using Bluetooth, then open the
spectrometry application.
b. Select ANALYZE SOLUTION from the menu at the top of the screen.
c. Select CALIBRATE DARK from the Menu at the bottom of the screen. The
Spectrometer will turn off all its lights and perform the calibration. A check mark
will appear when the calibration is finished.
d. Put distilled water into a cuvette. This should be the same distilled water that was
used as a diluent for the solutions being analyzed. Always handle the cuvette by
the ridged sides. Wipe off any fingerprints using a lint free wipe. Place the
cuvette into the spectrometer so that the ridged sides are facing the violet and
31
green light icons and clear sides face the white light and absorbance spectrum
icons.
e. Select CALIBRATE REFERENCE from the menu at the bottom of the screen. A
check mark will appear when the calibration is complete.
a. Place the cuvette into the spectrometer as you did in the calibration.
b. Select the red RECORD circle at the bottom left of the screen to start analyzing
the solution. (It changes into a square while data is being collected.)
c. Select the red STOP RECORDING square to stop data collection.
d. Select SCALE TO FIT to rescale your data.
e. Use the ADD COORDINATE to locate the 562 nm wavelength to analyze on the
curve.
f. A small hand will replace your cursor. Move it to the box that has appeared on
the graph and drag the box slowly toward the curve. As you get near the curve
an arrow will appear that indicates a specific wavelength on the curve. Releasing
the box will snap the box to the point on the curve the arrow is pointing to.
g. Drag the box along the curve to find the 562 nm wavelength to analyze.
4. Record the absorbance value in the table above and repeat the procedure for all
unknown and known solutions you prepared.
5. Please return all supplies and equipment to your TA.
6. Finally, you should wipe down your bench with the ethanol provided.
32
Laboratory #9, -Galactosidase Activity Assay
Objectives
1. To learn how to check that a protein you purified is biologically active.
2. To learn about the proper conditions that are required to get biological activity.
3. To analyze and interpret the results of the activity assay
Introduction
Enzymes are the biological version of catalysts in that they accelerate the rate of a
biological reaction by reducing the activation energy barrier that is required for the reaction to
proceed. Just like small molecule catalysts, enzymes are not consumed in the reaction and are
available to be reused repeatedly in many reactions. Enzymes like most other proteins are
globular in shape and can also be denatured which will render them inactive. To ensure that you
are working with a native and fully active protein, you can check for biological activity after the
enzyme has been purified. To check for activity, you must know certain characteristics about the
enzyme to create a suitable activity assay. First you must know what the natural substrate is for
this enzyme. You can utilize either the natural substrate or a derivative the closely mimics the
natural substrate but gives you an easy way to detect the product whereas the natural substrate
cannot be detected. For example, in today’s lab, we will be using ortho-Nitrophenyl-β-
galactoside (ONPG) instead of the natural substrate for -Galactosidase in E. coli, lactose.
Cleavage of the disaccharide lactose results in one D-galactose and one D-glucose. Both
cleavage products are colorless and difficult to detect as lactose cleavage products. However,
ONPG (is colorless) cleavage produces D-galactose and O-nitrophenyl. The O-nitrophenyl as a
bright yellow color that can easily be detected by a spectrophotometer at a wavelength of
420nm.
If you can detect the appearance of yellow ONP and the absence of yellow color in the
negative control, then you can be confident that your enzyme, in this case -Galactosidase, is
active. Keep in mind that different enzyme purifications will have different levels of activity,
indicating varying levels of native/denatured enzyme. Some preparations will be bad while
others will be excellent. This assay will only tell you if there is active enzyme present in your
purified protein and not the quality of the preparation. Next week, we will be doing kinetics which
is much more complex but is also more qualitative.
33
Experimental procedure
We will again be using the PASCO spectrophotometers for this week’s lab. Please refer
to the set up and use in last week’s lab write-up (Section labeled “Pasco Wireless
Spectrophotometer Instructions for Use”).
1. You will be working as usual in groups of three and you will need the following items
from your TA.
2. First, please connect your spectrophotometer to your computer. You will not have time to
set this up after you mix your reaction components so do it first.
a. It is going to be critical that you begin collecting data as soon as you mix your
components because the reactions occur remarkably fast (E. coli -
Galactosidase can carry out as many as 400 reactions per second at room
temperature). If you don’t start collecting data immediately, you will miss the
reaction!
b. You will be using the timed run for your spectrophotometer readings. Set it up to
collected data for 10-15 minutes with 10 second intervals.
c. In the software, go to Analyze solution. Collect the dark reference and a light
reference (just use air). Then go to TIME and start recoding when ready.
3. You will do your negative control first (it will also serve as practice).
a. Add 950 μl Reaction buffer to a labeled cuvette.
b. Add 0 μl Purified -Galactosidase enzyme and mix well.
c. Place cuvette in spectrophotometer and begin reading as soon as you add the
ONPG in step d below.
d. Add 50 μl ONPG (10 mg/ml) supplied as a 10 X stock and mix well by pipetting
up and down several times.
e. Once the run has started, do NOT touch the spectrophotometer. Even bumping it
will give you altered readings.
f. Once the run is finished hit the stop record button and export the file to a csv file
which you can open with excel. Please create a plot of your data for your report.
34
a. Add 948 μl Reaction buffer to a labeled cuvette.
b. Add 2 μl Purified -Galactosidase enzyme and mix well.
c. Place cuvette in spectrophotometer and begin reading as soon as you add the
ONPG in step d below.
d. Add 50 μl ONPG (10 mg/ml) supplied as a 20 X stock and mix well by pipetting
up and down several times to mix very well. Mixing is critical!!!
e. Once the run has started, do NOT touch the spectrophotometer and do not take
out the cuvette. Even bumping it will give you altered readings.
f. Once the run is finished hit the stop record button and export the file to a csv file
which you can open with excel. Please create a plot of your data for your report.
35
Laboratory #10, -Galactosidase Enzyme Kinetics
Objectives
1. To learn how to determine the Km and Vmax and type of enzyme inhibition present.
2. To learn about reaction rates and enzyme inhibitors.
3. To analyze and interpret complex experimental kinetics data using an excel
spreadsheet.
Introduction
This week’s lab is an extension of last week where you looked at the activity assay to
answer the simple question “Is my protein active”. This week you will be asking “How active is
my protein compared to other enzymes and what is the effect of an inhibitor”. Kinetics data is
calculated differently in that you keep the enzyme concentration constant while you modify the
substrate concentration. You then determine the initial velocity (V o) of the reaction at those
different substrate concentrations. The initial velocities are then plotted versus substrate
concentration to calculate the Km and Vmax. To calculate the V o you simply calculate the slope of
the line when you plot the absorbance vs. time. For V o only use the time from zero to about 5-10
minutes when the velocity is still linear.
Experimental procedure
The first part of the set-up this week is identical to last week.
1. You will be working as usual in groups of three and you will need the following items
from your TA.
2. First, please connect your spectrophotometer to your computer. You will not have time to
set this up after you mix your reaction components so do it first.
a. It is going to be critical that you begin collecting data as soon as you mix your
components because the reactions occur remarkably fast (E. coli -
Galactosidase can carry out as many as 400 reactions per second at room
temperature). If you don’t start collecting data immediately, you will miss the
reaction!
36
b. You will be using the timed run for your spectrophotometer readings. Set it up to
collected data for 10-15 minutes with 10 second intervals.
c. In the software, go to Analyze solution. Collect the dark reference and a light
reference (just use air). Then go to TIME and start recoding when ready.
3. To calculate initial velocities (V o), each group will run 6 reactions as listed in the table
below (if there are not enough groups then only do the first sets of reactions listed).
a. Results from the entire class will be pooled together to calculate the kinetics
values. Be sure to ask the TA about which reactions you are responsible for
before you start so that you do not replicate date that someone else is also
doing.
i. Add the reaction buffer to the cuvette first
ii. Next add the -Galactosidase enzyme and mix well
iii. The substrate will initiate the reaction and so you must do this as quickly
as possible.
1. Place the cuvette in the spectrophotometer
2. Take the appropriate amount of ONPG and put it into a
microcentrifuge tube (NOT the cuvette with the enzyme!).
3. Add the inhibitor to the ONPG and mix it well.
4. Now pipet the ONPG/inhibitor mixture with a P200 pipettor set to
200 μl and quickly mix it into the reaction mixture by pipetting up
and down three times. DO NOT MAKE BUBBLES/FOAM
5. Collect data for about 10 minutes and do not touch the cuvette or
spectrophotometer because it may influence your results.
6. After you stop the recording, export your data as a tube#.csv file
(which can be read by excel).
7. Move on to the next tube and proceed as above.
37
Experimental Reaction Setup
Reaction Galactosidase Inhibitor (IPTG) Resulting
Tube # ONPG (μl)
Buffer (μl) (μl) (μl) Vo
1 993 2 5 0
2 988 2 10 0
3 978 2 20 0
4 968 2 30 0
5 958 2 40 0
6 948 2 50 0
7 983 2 5 10
8 978 2 10 10
9 968 2 20 10
10 958 2 30 10
11 948 2 40 10
12 938 2 50 10
13 973 2 5 20
14 968 2 10 20
15 958 2 20 20
16 948 2 30 20
17 938 2 40 20
18 928 2 50 20
19 963 2 5 30
20 958 2 10 30
21 948 2 20 30
22 938 2 30 30
23 928 2 40 30
24 918 2 50 30
25 953 2 5 40
26 948 2 10 40
27 938 2 20 40
28 928 2 30 40
29 918 2 40 40
30 908 2 50 40
31 943 2 5 50
32 938 2 10 50
33 928 2 20 50
34 918 2 30 50
35 908 2 40 50
36 898 2 50 50
4. For data analysis, you need to plot your data in excel and then calculate a best fit line
through the data (I would recommend that you use linear regression). The slope of that
line is your initial velocity. We will need an initial velocity data point from each one of the
experimental runs in the tables above to calculate the kinetics data.
5. The lab for next week will include a tutorial on the data analysis for Km and Vmax
calculations. Therefore, you will not need to turn in a lab report until the end of the next
lab. In other words, we are combining the lab report from this week with next week’s lab.
a. If you want to get ahead for next week, you can download and try out the excel
spreadsheet called “Lineweaver_Burk_Template.xlsx” from Blackboard.
b. Make sure you bring a laptop with excel with you next week!
6. Please return all supplies and equipment to your TA.
7. Finally, you should wipe down your bench with the ethanol provided.
38
Laboratory #11, Kinetics Calculations, continuation from last week.
Objectives
1. To learn how to calculate the initial velocity from the absorption data collected last
week.
2. To learn how to calculate the Km and Vmax using excel
3. To analyze the results and determine the effect of the inhibitor IPTG.
Introduction
This week’s lab is a continuation of last week. We will use the data collected last week to
calculate the kinetics values Km and Vmax. This is a relatively simple idea but it is data heavy
which can lead to some confusion. Therefore, a spreadsheet has been created for you to help
guide you along the data analysis pathway. You’ll be walked through the data analysis in class.
Experimental procedure
1. There is no wet lab this week. You need a laptop computer (or take very good notes)
that has excel installed on it so you can follow along as everything is explained to you.
2. Please download the spreadsheet named “Lineweaver_Burk_Template.xlsx” from
Blackboard.
3. There are two spreadsheets in the excel file. The first one is for calculating the V o and
the second brings all the data from the entire class to plot a Lineweaver-Burk plot.
DO NOT MAKE CHANGES TO ANYTHING OTHER THAN THE YELLOW CELLS!
4. First, use your data from last week to calculate the six Vo values for your group.
a. Open the tube-1.csv file in excel for your first reaction from last week.
b. Copy the absorbance values and paste them into the yellow column on the V o
calculation sheet.
c. Use only the linear part at the beginning of the reaction. You will therefore have
to adjust the formula to use fewer cells. Adjust the formula by putting the cursor
on cell F23. Then modify the values underlined here [=INDEX(LINEST(C2:C14,A2:A14),1)].
Leave the rest of the formula alone!
d. The slope of the line is equal to the initial velocity V o and shows up automatically
at F23.
e. Enter the Vo value you get in the table below and then repeat the same
procedure for your other reactions until you have calculated all the V o values for
your group.
5. Once all the Vo values have been determined, switch to the Lineweaver-Burk
spreadsheet by clicking on the second tab at the bottom of the Excel program window.
6. Now fill in the yellow cells with the initial velocity Vo values you just calculated. Fill in all
others with the appropriate Vo values from all the other groups in you class.
7. The Lineweaver-Burk plot is plotted automatically. The Km and Vmax values are also
calculated automatically and show up in the cells L22 and L23.
8. You can also determine the type of IPTG inhibition by checking the Lineweaver-Burk plot
and how the lines behave relative to each other.
39
Reaction Galactosidase Inhibitor (IPTG) Resulting
Tube # ONPG (μl)
Buffer (μl) (μl) (μl) Vo
1 993 2 5 0
2 988 2 10 0
3 978 2 20 0
4 968 2 30 0
5 958 2 40 0
6 948 2 50 0
7 983 2 5 10
8 978 2 10 10
9 968 2 20 10
10 958 2 30 10
11 948 2 40 10
12 938 2 50 10
13 973 2 5 20
14 968 2 10 20
15 958 2 20 20
16 948 2 30 20
17 938 2 40 20
18 928 2 50 20
19 963 2 5 30
20 958 2 10 30
21 948 2 20 30
22 938 2 30 30
23 928 2 40 30
24 918 2 50 30
25 953 2 5 40
26 948 2 10 40
27 938 2 20 40
28 928 2 30 40
29 918 2 40 40
30 908 2 50 40
31 943 2 5 50
32 938 2 10 50
33 928 2 20 50
34 918 2 30 50
35 908 2 40 50
36 898 2 50 50
40
Appendix
List of supplies for TAs
Laboratory #1 Required Supplies and Equipment
For TA: Supplies needed per lab section of 18 students. 6 groups of 3
1. Spray bottle with Ethanol
2. A bag of 50 mL conical tubes (2 for each group)
3. 5 mL serological pipets (6)
4. 10 mL serological pipets. (6)
5. 6 pipet bulbs.
6. Waste containers for tips etc. (6)
7. 50 mL conical tube racks (6)
8. 1.5 mL tubes racks (6)
9. Microcentrifuge tubes (1 per group)
10. 6 sets of micro pipettors (P1000, P200, P20). 7th set of micro pipettors is for TA
11. 6 sets of micropipettor tips (P1000 and P200) Preferably the ART tips to avoid pipet
contamination.
12. 6 each 25 mL of 1 M Tris stock solutions (for all 3 labs)
13. 6 each 25 mL of 0.5M EDTA stock solutions (for all 3 labs)
14. 6 each 25 mL of 20X Borate stock solutions (for all 3 labs)
15. 6 sets of microfuge tubes with 1 mL water (labeled A)
16. 6 sets of microfuge tubes with 1 mL glycerol (labeled B)
17. 6 sets of microfuge tubes with 1 mL ethanol (labeled C)
18. 6 sets of microfuge tubes with 1 mL acetonitrile (labeled D)
19. You will use balance already in analytical labs
41
Laboratory #3 Required Supplies and Equipment
For TA: Supplies needed per lab section of 18 students. 6 groups of 3
42
Laboratory #5 Required Supplies and Equipment
For TA: Supplies needed per lab section of 18 students. 6 groups of 3
3. Lysis Buffer
10 mM HEPES pH 7.5 10 mL of 1 M
43
Laboratory #7 Required Supplies and Equipment
For TA: Supplies needed per lab section of 18 students. 6 groups of 3
44
Laboratory #10 Required Supplies and Equipment
For TA: Supplies needed per lab section of 18 students. 6 groups of 3
REFERENCE:
Ricardo Bernal, Department of Chemistry University of El Paso. Retrieved:
https://digitalmeasures.utep.edu/ai/rbernal/schteach/CHEM_3131_Biochemistry_Lab_Manual_S
econd_Edition-1.pdf; August 5, 2021.
45