Module 1
Module 1
ECT 414
1
MODULE 1
• Introduction to bio-medical engineering, bio-medical
instrumentation system.
• Overview of anatomy and physiological systems of the body.
• Sources of bio-electric potential: Resting and action potential,
propagation of action potentials.
• Bioelectric potentials examples (ECG, EEG, EMG, ERG, EOG, EGG
concept only.)
• Electrode theory: Nernst relation, Electrode skin interface
• Bio potential electrodes: Microelectrodes, skin surface electrodes,
needle electrodes
• Instrumentation for clinical laboratory: Bio potential
amplifiers-instrumentation amplifiers, carrier amplifiers, isolation
amplifiers, chopper amplifiers
2
Biomedical Engineering
3
Introduction to Biomedical Instrumentation System
4
5
Major Components of Medical Instrumentation System
1. Energy Source
2. Measurand
3. Sensor / Transducer
4. Signal Conditioning
5. Output Display
Auxiliary Components
1. A calibration signal
2. Control and feedback signal
6
1. Energy Source
• Used to energize the whole instrumentation system Examples:
Different sources used are electric, light, infrared, mechanical and
ultrasound
2. Measurand
• The physical quantity, property, or condition that the system
measures is called measurand.
Examples: Internal (Blood Pressure)
3. Sensor / Transducer
• The transducer is defined as a device that converts one form of
energy to another.
• A sensor converts a physical measurand to an electric output
7
4. Signal Conditioning
• Simple signal conditioners may only amplify and filter the signal or
merely match the impedance of the sensor to the display.
• Often sensor outputs are converted to digital form and then processed
by specialized digital circuits or a microcomputer.
• For example, signal filtering may reduce undesirable sensor signals.
• It may also average repetitive signals to reduce noise, or it may convert
information from the time domain to the frequency domain.
5. Output Display
• The results of the measurement process must be displayed in a form
that the human operator can perceive.
• The best form for the display may be:
• a. Numerical
8
• b. Graphical
• c. Displacement
• d. Discrete or continuous
• e. Visual / Hearing
The processed signal after conditioning passed through
1. Alarm System: Indicate when measurand goes beyond a preset
limit.
2. Data Storage: To maintain the data for future reference
3. Data Transmission: Used to transmit the information obtained
from one location to another.
9
Auxiliary Components
10
11
12
13
14
Bioelectric Potentials
15
Sources of Bioelectric potentials
16
Resting and Action potentials
17
Resting potentials
V -70 mV
Ground
Depolarization of cell
Na+
Na+
Cell Membrane
Na+
+
K +
K
+
+ K
Na +
K
Na+
+
K +
K
Na+ K
+
K
+
Na+
Na+
Action potentials
Cell Membrane
V 20 mV
Ground
Generation of Action Potentials
V -70 mV
Ground
Re polarization
43
Definition
• Electrodes are devices that convert ionic potentials into
electronic potentials.
44
Electrodes classifications
Electrode
s
Polarize Non
d polarized
electrode electrode
s s
45
A. Microelectrodes
46
Types
• 1. Metal microelectrode
• Metal microelectrodes are formed by electrolytically etching the tip of
fine tungsten to the desired size and dimension.
• Then the wire is coated almost to the tip with any type of insulating
material.
• The metal-ion interface takes place where the metal tip contacts
the electrolyte.
• The main features of metal microelectrodes are
1. Very good S/N ratio
2. Strong enough to penetrate
3. High biocompatibility
47
2.Micropipette
• The micropipette type of microelectrode is a glass
micropipette with its tip drawn out to the desired size.
• The micropipette is filled with an electrolyte which should be
compatible with the cellular fluids.
• A micropipette is a small and extremely fine pointed pipette
used in making microinjections.
B. Body Surface
Electrodes
49
1. Immersion electrodes:
50
2. Plate electrodes:
• These electrodes were
separated from subject’s skin
by cotton pads socked in a
strong saline solution.
• The plate electrodes have
generally smaller contact area
and they do not totally seal on
the patient.
• The electrode slippage and
displacement of plates were
the major difficulties faced by
these type of electrodes
• Since these types of electrodes
were very sensitive, it led to
measurement errors.
51
3. Floating electrodes
52
4.Disposable electrodes:
53
5. Suction
electrodes:
54
6. Ear clip & Scalp
electrodes:
55
C. Needle
Electrodes:
• Actually the needle electrodes are not inserted into the brain.
They nearly penetrate the skin.
56
• The needle electrodes for EMG measurement consist of
fine insulated wires placed in such a way that their tips
are in contact with the muscle, nerve or other tissues
from which the measurement is made.
• The needle creates the hole necessary for insertion and
the wires forming the electrodes are carried inside it.
57
• One of the main advantage of needle electrodes is that they are
less susceptible to movement errors than surface electrodes.
1. Instrumentation Amplifier
2. Isolation Amplifier
3. Chopper Amplifier
4. Carrier amplifier
1. Instrumentation Amplifier
• In biomedical applications, high gain and the high input
impedance are attained with an instrumentation
amplifier.
• A 3-amplifier setup forms the instrumentation amplifier
circuit.
• The output from the transducer is given as input to the
instrumentation amplifier.
• To each input of the differential amplifier, the
non-inverting amplifier is connected
• They are combined together to form the input stage of
the instrumentation amplifier.
• The third op-amp is the difference amplifier, and it is the
output of the instrumentation amplifier.
• The output from the difference amplifier Vout is the
difference between two input signals given at the input
points.
2. Isolation Amplifier
Input
• An isolation amplifier is a type of amplifier that provides
electrical isolation between its input and output circuits.
• The primary purpose of an isolation amplifier is to transmit
signals across a barrier, such as a physical barrier or isolation
barrier, while preventing any direct electrical connection
between the input and output sides.
• This isolation is crucial in applications where there is a need
to protect sensitive equipment, ensure safety, or avoid ground
loops.
66
Applications:
Operation:
• The electrical signals are obtained with electrodes.
• The signals received goes to the amplifier block, where
signals amplification occurs.
• After amplification, the signal enters the modulation
block.
• When it goes to the isolation barrier, optical cable or
transformer can be used.
• If in case of optical cable, modulator output travels to
LED.
• The LED converts electrical signals into light energy.
• If the transformer acts an isolation barrier, modulator
output connects the primary winding of the transformer.
• Energy from primary transfers to the secondary winding
based on the mutual induction principle.
• At the next stage, secondary output enters the
demodulation block.
• Finally, the amplified demodulated signal is obtained.
WORKING
• Chopping Technique: The chopper amplifier periodically
switches between two different input voltage sources.
• One source is the actual input signal, and the other is a
reference voltage.
• This switching occurs at a high frequency, typically in the kilo
hertz range.
71
• Offset Reduction: By rapidly switching between the input
signal and the reference voltage, any DC offset voltage
present in the amplifier gets modulated at the chopping
frequency.
• The output of the amplifier then contains both the desired AC
signal and the modulated offset.
72
• The chopper technique is particularly effective in minimizing
the offset voltage and offset drift, making chopper amplifiers
suitable for applications where high precision and stability
are essential.
• They are commonly used in instrumentation amplifiers,
data acquisition systems, and other applications where
accurate signal processing is crucial.
73
• The first block chopper accepts the DC input signal and
converts them to an AC signal. The AC amplifier block
amplifies the chopped AC signal.