CDV 700
CDV 700
SECTION 1. SECTION 6.
PRECAUTIONS
Paragraph OPERATOR’S MAINTENANCE
Pag Paragraph
e Page
1.1 Precautions - -- 3
6.1 Battery Replacement 8
SECTION 2.
GENERAL DESCRIPTION SECTION 7.
2.1 Introduction 3
2.2 The Probe 3 PREVENTIVE MAINTENANCE
2.3 The Circuit Box - 3 7.1 Battery Life 9
2.4 The Head Phone 3 7..2 Storage 9
2.5 The Carrying Strap 3
SECTION 3. SECTION 8.
THEORY OF OPERATION CORRECTIVE MAINTENANCE
3.1 Introduction 4 8.1 Replacing The Batteries 9
3.2 The Geiger Tube 4 8.2 Replacing The
3.3 The High Voltage Supply. 4 Geiger Tube -- - 9
3.4 The Pulse Shaping and 8.3 Checking The High
Metering Circuit 4 Voltage Supply 9 -
3.5 Scale Ranges S
8.4 Checking The Pulse
3.6 The Headphone Circuit 5
Shaping and Integrating
SECTION 4. Circuit 10
INSTALLATION 8.5 Testing For Proper
4.1 Installing The Batteries 5 Operation Of The
Range Switch - 10
SECTION 5.
OPERATION
5.1 Operating The Circuit SECTION 9.
The First Time 7
5.2 Calibration 7 REPLACEABLE PARTS LIST
5.3 Scale Ranges 7 9.1 Electrical
5.4 Using The Headphones 7 Components 12-13
5.5 Normal Background 7 9.2 Mechanical
5.6 Checking Calibration 8 Components 14-15 -
LIST OF ILLUSTRATIONS
Figure Page
1. View of CD V-700, Model No. 6B, Showing Operating Controls 2
2. View of CD V-700, Model No. 6B, Showing Major Components 6
3. CD V-700, Model 6B, Schematic Circuit Diagram 11
1
1. PRECAUTIONS
1.1 PRECAUTIONS:
Although this instrument is operated from four “D” cells, the high voltage power supply
operates at voltages in excess of 900 volts which constitutes a shock hazard if not
handled carefully. The power supply high voltage components are located near the high
voltage transformer at the rear of the circuit hoard. These components should not be
touched, even when the instrument is “OFF”, until the high voltage capacitors are
discharged. The method of discharging these capacitors is to short-circuit the two leads
of the corona regulator tube V2 with a screwdriver which has an insulated handle.
2. GENERAL DESCRIPTION
2.1 INTRODUCTION:
This instrument is a portable survey meter using a geiger tube as the detector. The geiger tube is mounted in a probe
on the end of a thirty-six inch cable. The entire instrument and its accessories comprise a circuit box, a probe, a
headphone and a carrying strap; a radioactive sample is mounted on the side of the case.
2
3. THEORY OF OPERATION The blocking oscillator consists of components Qi and T1. Negative pulses from the
geiger tube appear across winding 3-4 of T1. These pulses are coupled into winding 2-5
and into the base circuit of Qi. When Qi is turned “ON” by a GM tube trigger, Qi
3.1 INTRODUCTION: saturates and nearly all of the battery voltage of BT3-4 appears across winding 1-6. The
winding current increases and a voltage is induced in winding 2-5. The induced voltage
Operationally this instrument consists of a geiger tube radiation detector, a regulated high is in a direction such that conduction of Qi is maintained. Winding 2-5 current increases
voltage supply, a pulse shaping and metering circuit, an indicating meter and headphone linearly until the transformer core saturates. At this time the circuit rapidly turns “OFF”
for audible detection of activity. and an inductive “fly-back” appears across both windings.
The metering circuit consists of an integrating capacitor Cl, and range multiplier resistors
Ri through R3. The multipliers determine the amount of charge that is placed on Cl
3.2 THE GEIGER TUBE:
during the pulse period of the blocking oscillator. The charge on the capacitor is
discharged by the meter and R4. R5B is used for calibration.
The geiger tube is a gas filled device which detects the presence of ionization within its
gaseous volume. The. ionization results from the passage of ionizing type radiation
through the gas. The primary type of radiation it detects is beta rays (high speed
3.5 SCALE RANGES:
electrons). These are produced as a primary breakdown product of radioactive substances
and in addition are produced within the geiger tube and within the walls of the probe by Three ranges of operation are provided. The first range Xl, requires 300 pulses per
gamma radiation. A shield is provided which stops external beta radiation, thus making minute for full scale indication; the second range X1O, 3000 pulses per minute; and
the detector sensitive to gamma radiation only, but a window in the probe may be opened the third range XlOO, 30,000 pulses per minute. These correspond respectively to 0.5
to make the system sensitive to beta radiation also. The geiger tube operates at 900 volts milliroentgens per hour, 5 milliroentgens per hour and 50 milliroentgens per hour of
which is essentially the center of a plateau extending from about 850 volts to about 920 radium-equivalent radiation. Scale changing is effected by switching meter range
volts. resistors.
Clamp the circuit box back together and turn the switch to the X10 scale. Make certain that the sliding beta window of
the probe is closed. Wait thirty seconds for the system to reach stability. The indicator should remain substantially at
zero.
Open the window on the probe and present it to the center of the calibration source which is a beta radiation sample.
The indicator should fall between 1.5 mr/hr and 2.5 mr/hr, averaging about 2 mr/hr.
5.2 CALIBRATION:
NOTE:
The beta source must constitute the sole source of radiation when calibration is performed. Calibration must not be
under. taken when the background is above normal or when the probe is in a radiation field other than that produced by
the known beta source supplied with the instrument.
If the indication falls above or below this range, it may be corrected by the
screw.driver adjustment inside the box which is marked “CAL”. Advancing
the adjustment clockwise increases the reading.
Since normal background of radioactivity is of the order of 0.01 to 0.02 milliroentgens per hour, little activity will
normally be observed. Under back. ground conditions only about 20 per minute of these clicks occur and they are
randomly spaced so that one may wait for several seconds before any
6
click is observed and then there may be two or three in rapid succession. Very accurate
measurements of background and other low level radiation can be made by counting 7. PREVENTIVE MAINTENANCE
headphone clicks and timing with a watch which has a second hand. The procedure is to count 7.1 BATTERY LIFE:
a given number of counts and observe the time required to obtain these counts. The radiation
rate in counts per minute is the number of counts divided by the time in minutes. Table 1 CAUTION: Make certain the instrument is turned oIl at all times when not in use,
gives the number of counts that are required to provide a given percentage error where percent otherwise the batteries will certainly be depleted and the instrument rendered ineffective. The
standard error is defined as that error for which in 63 cases out of 100 the true error will not life of the batteries in the front battery box is about 100 hours under continuous operation and
exceed the given percentage error. The nine-tenths error is that error for which the true about 175 hours when operated four hours a day. The life of the other two cells is
reading is no different from the observed reading within the given percentage limits for 90 considerably longer but it is recommended that all batteries be replaced at one time. It is also
cases out of 100. recommended that the operator become accustomed to noting that the operating switch is in
the “OFF” position when the instrument is set aside.
The operator should periodically check the calibration of the instrument to verify that it is 8.2 REPLACING THE GEIGER TUBE:
correct. This operation is described in paragraph 5.2. The chief maintenance on this instrument is replacing the batteries. However, the geiger tube
also expends itself with use and must be replaced occasionally but one cannot predict
5.7 USING THE CARRYING STRAP: precisely the life of a geiger tube since the total number of counts it has accumulated and the
operating conditions of temperature, voltage and load characteristics are very important.
The instrument may be carried in the hand or by a strap over the shoulder. The strap Whenever fresh batteries are installed into the instrument and the instrument does not work
anchors are arranged in such a way that the meter is visible when carried over the
correctly, it is wise first to try replacing the geiger tube before making any further attempts at
right shoulder.
circuit checking.
Battery replacement is easily accomplished by removing the circuit box bottom and
the battery retaining clips on the two battery boxes. The old cells may be pulled out
without tools and the new ones inserted.
S 9
and the setting of R5A should be checked. This screw-driver adjustment is set properly
by the following procedure:
a. Inset a 0-100 millampere meter in series with the power supply batteries.
The high voltage output of the power supply should be tested with an electrostatic
voltmeter. The voltage should read 910 ±10 volts. This voltage may also be measured
with a 20,000 ohms-per-volt meter on the 5,000 volt range with new batteries in the
instrument.
The nominal resistance values for the transformer are: winding 1-2, 4500 ohms; winding
3-4, 11 ohms; and winding 5-6, 5 ohms (all values ±20%).
After continuity tests have been made, the integrating circuit should be tested further
using an oscilloscope. The proper pulse waveshape on the collector of Q i is a positive 3
volt square wave of a nominal 150 microseconds in duration followed by a “fly-back” of
about —20 volts in amplitude. Diodes CR1 and CR2 may be tested with an ohmmeter on
the RX1O,000 range. One end of the diode under test should be disconnected from the
circuit for this test. (Diode CR3 in the high voltage power supply will not respond to this
test.) The indication of a good diode is very low resistance with the ohmmeter leads
connected a given way. With the leads reversed the resistance reading should be 100k
ohms or greater for CR2, and 1 megohm or better for CR1.
X100 range reads 1900 ohms, X1O range reads 200 ohms, and Xl range reads 16.5 ohms.
(All values ±5%). The batteries should be removed for all tests in this section. VOLTAGE
V2 RANGE
GV3A VOLTAGES MEASURED IN XI POSITION
USE 20,000 OHMS-PER-VOLT VOLTMETER
CAPACITORS IN MFD
Figure 3. CD V-TOO. Model Au. 6B, Schematic
Circuit Diagram
10 11
9.3 List of Manufacturers
HARRY DAVIES MOLDING COMPANY, 1428 North Wells Street, Chicago 10, Illinois
UNION CARBIDE CONSUMERS PRODUCTS COMPANY, 30 East 42nd Street, New York, New York
WATERBURY BUCKLE COMPANY, 862 South Main Street, Waterbury 20, Connecticut