Technical Information

Proline Promag 10P


Electromagnetic Flow Measuring System
Flow measurement of liquids in chemical or process applications

Application Your benefits


Electromagnetic flowmeter for bidirectional Promag measuring devices offer you cost-effective flow
measurement of liquids with a minimum measurement with a high degree of accuracy for a wide
conductivity of ≥ 50 μS/cm: range of process conditions.
• Acid, alkalis The uniform Proline transmitter concept comprises:
• Paints • High degree of reliability and measuring stability
• Pastes • Uniform operating concept
• Water, wastewater, etc.
The tried-and-tested Promag sensors offer:
• Flow measurement up to 9600 m³/h • No pressure loss
• Fluid temperature up to +130 °C • Not sensitive to vibrations
• Process pressures up to 40 bar • Simple installation and commissioning
• Lengths in accordance with DVGW/ISO
Application-specific lining material:
• PTFE

TI094D/06/en/08.09
71101431
Proline Promag 10 P

Table of contents

Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 18


Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring tube specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Material load diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Fitted electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Surface roughness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 27
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
C-tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Electrical connection, measuring unit . . . . . . . . . . . . . . . . . . . . . 4
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Electrical connection, terminal assignment . . . . . . . . . . . . . . . . . 5
Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Electrical connection, remote version . . . . . . . . . . . . . . . . . . . . . 5
Pressure measuring device approval . . . . . . . . . . . . . . . . . . . . . . 27
Supply voltage (power supply) . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 28
Remote version cable specifications . . . . . . . . . . . . . . . . . . . . . . . 6
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Performance characteristics. . . . . . . . . . . . . . . . . . . . . 9
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . 28
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Operating conditions: Installation . . . . . . . . . . . . . . . 10


Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Inlet and outlet run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Length of connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Operating conditions: Environment . . . . . . . . . . . . . 14


Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Shock and vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . 15

Operating conditions: Process . . . . . . . . . . . . . . . . . . 15


Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Medium pressure range
(nominal pressure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Pressure tightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Limiting flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

2 Endress + Hauser
Proline Promag 10 P

Function and system design


Measuring principle Following Faraday's law of magnetic induction, a voltage is induced in a conductor moving through a magnetic
field.
In the electromagnetic measuring principle, the flowing medium is the moving conductor.
The voltage induced is proportional to the flow velocity and is supplied to the amplifier by means
of two measuring electrodes. The flow volume is calculated by means of the pipe cross-sectional area.
The DC magnetic field is created through a switched direct current of alternating polarity.

Ue

V
B
I
L

a0003191

Ue = B · L · v
Q=A·v
Ue Induced voltage
B Magnetic induction (magnetic field)
L Electrode spacing
v Flow velocity
Q Volume flow
A Pipe cross-section
I Current strength

Measuring system The measuring system consists of a transmitter and a sensor.


Two versions are available:
• Compact version: Transmitter and sensor form a mechanical unit.
• Remote version: Sensor is mounted separate from the transmitter.
Transmitter:
• Promag 10 (key operation, two-line, unilluminated display)
Sensor:
• Promag P (DN 25...600)

Input
Measured variable Flow velocity (proportional to induced voltage)

Measuring ranges Measuring ranges for liquids


Typically v = 0.01...10 m/s with the specified accuracy

Operable flow range Over 1000 : 1

Endress + Hauser 3
Proline Promag 10 P

Output
Output signal Current output
• Galvanically isolated
• Active: 4...20 mA, RL < 700 Ω (for HART: RL ≥ 250 Ω)
• Full scale value adjustable
• Temperature coefficient: typ. 2 μA/°C, resolution: 1.5 μA
Pulse/status output
• Galvanically isolated
• Passive: 30 V DC / 250 mA
• Open collector
• Can be configured as:
– Pulse output: Pulse value and pulse polarity can be selected, max. pulse width adjustable (5...2000 ms),
pulse frequency max. 100 Hz
– Status output: for example, can be configured for error messages, empty pipe detection, flow recognition,
limit value

Signal on alarm • Current output → Failsafe mode can be selected


• Pulse output → Failsafe mode can be selected
• Status output → “Not conductive” in the event of fault or power supply failure

Load See “output signal”

Low flow cut off Low flow cut off →Switch-on point can be selected as required.

Galvanic isolation All circuits for inputs, outputs and power supply are galvanically isolated from each other

Power supply
Electrical connection,
measuring unit

e b
e b e b

f c
a
g d
i
h 24 25 26 27 1 2
+ – + – L1 N
(L+) (L-)

a0003192

Connecting the transmitter (aluminum field housing), cable cross-section max. 2.5 mm2
a Electronics compartment cover
b Power supply cable
c Ground terminal for power supply cable
d Terminal connector for power supply cable
e Signal cable
f Ground terminal for signal cable
g Terminal connector for signal cable
h Service connector
i Ground terminal for potential equalization

4 Endress + Hauser
Proline Promag 10 P

Electrical connection,
terminal assignment Order version Terminal No.

24 (+) 25 (–) 26 (+) 27 (–) 1 (L1/L+) 2 (N/L–)

10***-***********A Pulse/status output HART current output Power supply

Functional values See “output signal” See “Supply voltage”

Electrical connection, remote


version

GND
S1
E1
E2
S2

E
S
d
a
c 6 5 7 8 4 37 36 42 41

2 1

b n.c. n.c. n.c.

5 7 4 37 42 41
E1
E2
GND
E

a0003193-en

Connecting the remote version


a Wall-mount housing connection compartment
b Sensor connection housing cover
c Signal cable
d Coil current cable
n.c. Not connected, insulated cable shields
Cable colors:
terminal numbers 5/6 = brown; 7/8 = white; 4 = green; 37/36 = yellow

Supply voltage (power supply) • 85...250 V AC, 45...65 Hz


• 20...28 V AC, 45...65 Hz, 11...40 V DC

Cable entry Power supply and signal cables (inputs/outputs):


• Cable entry M20 x 1.5 (8...12 mm)
• Thread for cable entries, ½" NPT, G ½"
Connecting cable for remote version:
• Cable entry M20 x 1.5 (8...12 mm)
• Thread for cable entries, ½" NPT, G ½"

Endress + Hauser 5
Proline Promag 10 P

Coil cable
Remote version cable • 2 x 0.75 mm2 PVC cable with common, braided copper shield (∅ ∼ 7 mm)
specifications • Conductor resistance: ≤ 37 Ω/km
• Capacitance core/core, shield grounded: ≤ 120 pF/m
• Operating temperature: –20...+8o °C
• Cable cross-section: max. 2.5 mm2
Signal cable
• 3 x 0.38 mm2 PVC cable with common, braided copper shield (∅ ∼ 7 mm)
and individual shielded cores
• With empty pipe detection (EPD): 4 x 0.38 mm2 PVC cable with common, braided copper shield (∅ ∼ 7
mm) and individual shielded cores
• Conductor resistance: ≤ 50 Ω/km
• Capacitance core/shield: ≤ 420 pF/m
• Operating temperature: –20...+8o °C
• Cable cross-section: max. 2.5 mm2

1
2
3
4
5
6
7
a b
a0003194

a Signal cable
b Coil current cable
1 Core
2 Core insulation
3 Core shield
4 Core jacket
5 Core reinforcement
6 Cable shield
7 Outer jacket

Operation in zones of severe electrical interference


The measuring device complies with the general safety requirements in accordance with EN 61010
and the EMC requirements of EN 61326/A1 (IEC 1326).
Caution!
Grounding is by means of the ground terminals provided for the purpose inside the connection housing.
Ensure that the stripped and twisted lengths of cable shield to the ground terminal are as short as possible.

Power consumption • 85...250 V AC: < 12 VA (incl. sensor)


• 20...28 V AC: < 8 VA (incl. sensor)
• 11...40 V DC: < 6 W (incl. sensor)
Switch-on current:
• 250 V AC → max 16 A (< 5 ms)
• 28 V AC → max. 5.5 A (< 5 ms)
• 24 V DC → max. 3.3 A (< 5 ms)

Power supply failure Lasting min. ½ cycle frequency: EEPROM saves measuring system data

6 Endress + Hauser
Proline Promag 10 P

Potential equalization Standard


Perfect measurement can only be guaranteed if the fluid and sensor are on the same electric potential. Most
Promag sensors have a built-in reference electrode as standard which guarantees the necessary connection. As
a result, the use of ground disks or other measures is generally not required.
Note!
When installing in metallic pipes, we recommend you connect the ground terminal of the transmitter housing
with the piping. Also, observe company-internal grounding guidelines.

a0003195

Potential equalization via the ground terminal of the transmitter

Caution!
For sensors without reference electrodes or without metallic process connections, potential equalization must
be carried out as described in the following special cases. These special measures apply especially where normal
grounding cannot be guaranteed or excessively strong equalizing currents are expected.

Metallic, not grounded piping


To prevent disturbing influences on the measurement, we recommend both sensor flanges be connected to the
pipe flange via a ground cable and grounded. The transmitter or sensor terminal compartment must be set to
ground potential via the ground terminal provided (see the diagram below).
Note!
The ground cable required for the flange-to-flange connection can be ordered separately from Endress+Hauser
as an accessory :
• DN ≤ 300: the ground cable is mounted directly on the conductive flange coating with the flange screws.
• DN ≥ 350: the ground cable is mounted directly on the transportation metal support.

DN  300 DN  350

6 mm² Cu
a0003201

Potential equalization in the case of equalization currents in metallic, not grounded piping

Plastic pipelines or pipelines with insulating lining


Normally, potential equalization takes place via the reference electrodes in the measuring tube. However, in
exceptional circumstances, large equalizing currents can flow via the reference electrodes due to the grounding
concept of a plant. This can destroy the sensor due to the electrochemical reduction of the electrodes, for
example. In such instances, e.g. in the case of fiberglass or PVC pipelines, we recommend the additional use
of ground disks for potential equalization.

Endress + Hauser 7
Proline Promag 10 P

Caution!
• Risk of damage by electrochemical corrosion. Please note the electrochemical series if ground disks and
measuring electrodes are made of different material.
• Also, observe company-internal grounding guidelines.

6 mm² Cu
a0003197

Potential equalization/ground disks for plastic or lined pipes

Pipe with cathodic protective units


In such instances, the device is to be installed potential-free in the piping:
• During installation, ensure that the sections of the piping are electrically connected with one another (copper
wire, 6 mm2).
• Make sure that no conductive connection to the device originates from the mounting material used and that
the mounting material withstands the torques used during installation.
• Please also note the applicable regulations regarding potential-free installation.

2 2

6 mm² Cu
a0003198

Potential equalization and cathodic protection


1 Power supply isolation transformer
2 Electrically isolated

8 Endress + Hauser
Proline Promag 10 P

Performance characteristics
Reference operating As per DIN EN 29104 and VDI/VDE 2641:
conditions • Fluid temperature: +28 C ± 2 K
• Ambient temperature: +22 C ± 2 K
• Warm-up period: 30 minutes
Installation:
• Inlet run >10 x DN
• Outlet run > 5 x DN
• Sensor and transmitter grounded.
• The sensor is centered in the pipe.

Maximum measured error • Current output: also typically ± 5 μA


• Pulse output: ± 0.5% o.r. ± 2 mm/s (o.r. = of reading)
Fluctuations in the supply voltage do not have any effect within the specified range.

[%]
2.5

2.0
0.5 %
1.5

1.0

0.5

0 v [m/s]
0 1 2 4 6 8 10
a0003200

Max. measured error in % of reading

Repeatability Max. ± 0.2% o.r. ± 2 mm/s (o.r. = of reading)

Endress + Hauser 9
Proline Promag 10 P

Operating conditions: Installation


Installation instructions Mounting location
Entrained air or gas bubble formation in the measuring tube can result in an increase in measuring errors.
Avoid the following installation locations in the pipe:
• Highest point of a pipeline. Risk of air accumulating!
• Directly upstream from a free pipe outlet in a vertical pipeline.

h ³ 2 x DN

a0003202

Mounting location

Installation of pumps
Sensors may not be installed on the pump suction side. This precaution is to avoid low pressure and the
consequent risk of damage to the lining of the measuring tube.
Information on the pressure tightness of the measuring tube lining can be found in the “Pressure tightness”
section in the “Operating conditions: Process” chapter.
Pulsation dampers may be needed when using piston pumps, piston diaphragm pumps or hose pumps.
Information on the shock and vibration resistance of the measuring system can be found in the “Shock and
vibration resistance” section in the “Operating conditions: Environment” chapter.

a0003203

Installation of pumps

Partially filled pipes


Partially filled pipes with gradients necessitate a drain-type configuration.
The empty pipe detection function (EPD ) provides additional security in detecting empty or partially filled
pipes.

³ 2 x DN

³ 5 x DN

a0003204

Installation with partially filled pipes

10 Endress + Hauser
Proline Promag 10 P

Down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes longer than 5 meters. This precaution
is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube. This measure
also prevents the liquid current stopping in the pipe which could cause air locks. Information on the pressure
tightness of the measuring tube lining can be found in the “Pressure tightness” section in the “Operating
conditions: Process” chapter.

>5m
a0003205

Installation measures for vertical pipes


1 Vent valve
2 Pipe siphon

Orientation
An optimum orientation helps avoid gas and air accumulations and deposits in the measuring tube. However,
the measuring device also offers the additional function of empty pipe detection (EPD) for detecting partially
filled measuring tubes or if outgassing fluids or fluctuating operating pressures are present.

Vertical orientation
This is the ideal orientation for self-emptying piping systems and for use in conjunction with empty pipe
detection.

a0003206

Vertical orientation

Horizontal orientation
The measuring electrode axis should be horizontal. This prevents brief insulation of the two measuring
electrodes by entrained air bubbles.
Caution!
Empty pipe detection only works correctly with horizontal orientation if the transmitter housing is facing
upwards. Otherwise there is no guarantee that empty pipe detection will respond if the measuring tube is only
partially filled or empty.

Endress + Hauser 11
Proline Promag 10 P

A
1

2 2

A
3
a0003207

Horizontal orientation
1 EPD electrode for empty pipe detection
2 Measuring electrodes for signal detection
3 Reference electrode for potential equalization

Vibrations
Secure the piping and the sensor if vibration is severe.
Caution!
If vibrations are too severe, we recommend the sensor and transmitter be mounted separately. Information on
the permitted shock and vibration resistance can be found in the
“Shock and vibration resistance” section in the “Operating conditions: Environment” chapter.

> 10 m

a0003208

Measures to prevent vibration of the measuring device

12 Endress + Hauser
Proline Promag 10 P

Inlet and outlet run If possible, install the sensor well clear of assemblies such as valves, T-pieces, elbows, etc.
Note the following inlet and outlet runs to comply with measuring accuracy specifications:
• Inlet run: ≥ 5 x DN
• Outlet run: ≥ 2 x DN

 5 x DN  2 x DN

a0003210

Inlet and outlet run

Adapters Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-diameter
pipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving fluids.
The nomogram shown here can be used to calculate the pressure loss caused by reducers and expanders.
Note!
The nomogram only applies to liquids of viscosity similar to water.
1. Calculate the ratio of the diameters d/D.
2. From the nomogram read off the pressure loss as a function of flow velocity (downstream from the
reduction)
and the d/D ratio.

[mbar]
100

8 m/s

7 m/s

6 m/s

10
5 m/s

4 m/s
max. 8°
3 m/s
d D
2 m/s
1

1 m/s

0.5 0.6 0.7 0.8 0.9


d/D
a0003213

Pressure loss due to adapters

Endress + Hauser 13
Proline Promag 10 P

Length of connecting cable When mounting the remote version, please note the following to achieve correct measuring results:
• Fix cable run or lay in armored conduit. Cable movements can falsify the measuring signal especially in the
case of low fluid conductivities.
• Route the cable well clear of electrical machines and switching elements.
• If necessary, ensure potential equalization between sensor and transmitter.
• The permitted cable length Lmax is determined by the fluid conductivity.
A minimum conductivity of 50 μS/cm is needed for all fluids.
• When the empty pipe detection function is switched on (EPD), the maximum connecting cable length is
10 m.

[mS/cm]

200

L max

100

50

[m]
50 100 200

L max
a0003214

Permitted length of connecting cable for remote version


Area marked in gray = permitted range; Lmax = length of connecting cable in [m]; fluid conductivity in [μS/cm]

Operating conditions: Environment

Ambient temperature range • Sensor: -20...+60 °C


• Transmitter: -10...+60 °C
Caution!
The permitted temperature range of the measuring tube lining may not be undershot or overshot
(→ “Operating conditions: Process” → “Medium temperature range”).
Please note the following points:
• Install the device in a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• The transmitter must be mounted separate from the sensor if both the ambient and fluid temperatures are
high.

Storage temperature • The temperature range for storing the device corresponds to the permitted ambient temperature range of the
transmitter and the sensor (see “Ambient temperature range”).
• The measuring device must be protected against direct sunlight during storage in order to avoid unacceptably
high surface temperatures.
• A storage location must be selected where moisture does not collect in the measuring device. This will help
prevent fungus and bacteria infestation which can damage the liner.
• If protecting caps or protective covers are mounted, these must not be removed before mounting the device.

14 Endress + Hauser
Proline Promag 10 P

Degree of protection • Standard: IP 67 (NEMA 4X) for transmitter and sensor


• Optional: IP 68 (NEMA 6P) for sensor for remote version

Shock and vibration resistance Acceleration up to 2 g following IEC 600 68-2-6

Electromagnetic compatibility • As per EN 61326


(EMC) • Emission: to limit value for industry EN 55011

Operating conditions: Process

Medium temperature range –40...+130 °C for PTFE (DN 25...300), restrictions→ see diagrams

TA [°C]
60
50
40 -
Esc
+ E

30

PTFE
20
10
0
-10
-40 -20 0 20 40 60 80 100 120 140 160 180 T [°C]
F
a0003449

Compact version (TA = ambient temperature range, TF = fluid temperature)

TA [°C]
60
50
40
30
PTFE

20
10
0
-10
-40 -20 0 20 40 60 80 100 120 140 160 180 T [°C]
F
a0003446

Remote version (TA = ambient temperature range, TF = fluid temperature)

Conductivity The minimum conductivity is: ≥ 50 μS/cm


Note!
In the remote version, the necessary minimum conductivity also depends on the cable length
(→ “Operating conditions: Installation” → “Length of connecting cable”).

Endress + Hauser 15
Proline Promag 10 P

Medium pressure range • EN 1092-1 (DIN 2501)


(nominal pressure) – PN 10 (DN 200...600)
– PN 16 (DN 65...600)
– PN 25 (DN 200...600)
– PN 40 (DN 25...150)
• ANSI B 16.5
– Class 150 (DN 1"...24")
– Class 300 (DN 1"...6")
• JIS B2238
– 10 K (DN 50...300)
– 20 K (DN 25...300)
• AS 2129
– Table E (DN 25, 50)
• AS 4087
– Cl. 14 (DN 50)

Pressure tightness Diameter Measuring tube


Pressure tightness, measuring tube lining

[mm] [inch] Material Limit values for the absolute pressure [mbar] at fluid temperatures:

Lining 25 °C 80 °C 100 °C 130 °C

25 1" PTFE 0 0 0 100

32 - PTFE 0 0 0 100

40 1 ½" PTFE 0 0 0 100

50 2" PTFE 0 0 0 100

65 - PTFE 0 * 40 130

80 3" PTFE 0 * 40 130

100 4" PTFE 0 * 135 170

125 - PTFE 135 * 240 385

150 6" PTFE 135 * 240 385

200 8" PTFE 200 * 290 410

250 10" PTFE 330 * 400 530

300 12" PTFE 400 * 500 630

350 14" PTFE 470 * 600 730

400 16" PTFE 540 * 670 800

450 18" PTFE Partial vacuum is impermissible

500 20" PTFE

600 24" PTFE

* No value can be specified.

16 Endress + Hauser
Proline Promag 10 P

Limiting flow The diameter of the pipe and the flow rate determine the nominal diameter of the sensor.
The optimum flow velocity is between 2…3 m/s. The velocity of flow (v), moreover, has to be matched to the
physical properties of the fluid:
• v < 2 m/s: for abrasive fluids such as potter's clay, lime milk, ore slurry, etc.
• v > 2 m/s: for fluids causing build-up such as wastewater sludges, etc.

Flow characteristic values (SI units)

Diameter Recommended flow rate Factory settings

Min./max. full scale value Full scale value Pulse value Low flow cut off
Current output
[mm] [inch] (v ~ 0.3 or 10 m/s) (v ~ 2.5 m/s) (~ 2 pulses/s) (v ~ 0.04 m/s)
3 3 3
25 1" 9...300 dm /min 75 dm /min 0.50 dm 1 dm3/min

32 1 ¼" 15...500 dm3/min 125 dm3/min 1.00 dm3 2 dm3/min

40 1 ½" 25...700 dm3/min 200 dm3/min 1.50 dm3 3 dm3/min

50 2" 35...1100 dm3/min 300 dm3/min 2.50 dm3 5 dm3/min

65 2 ½" 60...2000 dm3/min 500 dm3/min 5.00 dm3 8 dm3/min

80 3" 90...3000 dm3/min 750 dm3/min 5.00 dm3 12 dm3/min

100 4" 145...4700 dm3/min 1200 dm3/min 10.00 dm3 20 dm3/min

125 5" 220...7500 dm3/min 1850 dm3/min 15.00 dm3 30 dm3/min

150 6" 20...600 m3/h 150 m3/h 0.025 m3 2.5 m3/h

200 8" 35...1100 m3/h 300 m3/h 0.05 m3 5.0 m3/h

250 10" 55...1700 m3/h 500 m3/h 0.05 m3 7.5 m3/h

300 12" 80...2400 m3/h 750 m3/h 0.10 m3 10 m3/h

350 14" 110…3300 m3/h 1000 m3/h 0.10 m3 15 m3/h

400 16" 140…4200 m3/h 1200 m3/h 0.15 m3 20 m3/h

450 18" 180…5400 m3/h 1500 m3/h 0.25 m3 25 m3/h

500 20" 220…6600 m3/h 2000 m3/h 0.25 m3 30 m3/h

600 24" 310...9600 m3/h 2500 m3/h 0.30 m3 40 m3/h

Flow characteristic values (US units)

Diameter Recommended flow rate Factory settings

Min./max. full scale value Full scale value Pulse value Low flow cut off
Current output
[inch] [mm] (v ~ 0.3 or 10 m/s) (v ~ 2.5 m/s) (~ 2 pulses/s) (v ~ 0.04 m/s)

1" 25 2.5...80 gal/min 18 gal/min 0.20 gal 0.25 gal/min

1 ¼" 32 4...130 gal/min 30 gal/min 0.20 gal 0.50 gal/min

1 ½" 40 7...190 gal/min 50 gal/min 0.50 gal 0.75 gal/min

2" 50 10...300 gal/min 75 gal/min 0.50 gal 1.25 gal/min

2 ½" 65 16...500 gal/min 130 gal/min 1 gal 2.0 gal/min

3" 80 24...800 gal/min 200 gal/min 2 gal 2.5 gal/min

4" 100 40...1250 gal/min 300 gal/min 2 gal 4.0 gal/min

5" 125 60...1950 gal/min 450 gal/min 5 gal 7.0 gal/min

6" 150 90...2650 gal/min 600 gal/min 5 gal 12 gal/min

8" 200 155...4850 gal/min 1200 gal/min 10 gal 15 gal/min

10" 250 250...7500 gal/min 1500 gal/min 15 gal 30 gal/min

12" 300 350...10600 gal/min 2400 gal/min 25 gal 45 gal/min

Endress + Hauser 17
Proline Promag 10 P

Flow characteristic values (US units)

Diameter Recommended flow rate Factory settings

Min./max. full scale value Full scale value Pulse value Low flow cut off
Current output
[inch] [mm] (v ~ 0.3 or 10 m/s) (v ~ 2.5 m/s) (~ 2 pulses/s) (v ~ 0.04 m/s)

14" 350 500...15000 gal/min 3600 gal/min 30 gal 60 gal/min

16" 400 600...19000 gal/min 4800 gal/min 50 gal 60 gal/min

18" 450 800...24000 gal/min 6000 gal/min 50 gal 90 gal/min

20" 500 1000...30000 gal/min 7500 gal/min 75 gal 120 gal/min

24" 600 1400...44000 gal/min 10500 gal/min 100 gal 180 gal/min

Pressure loss • No pressure loss if the sensor is installed in a pipe with the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545
(→ “Operating conditions: Installation” → “Adapters”)

Mechanical construction
Measuring tube specifications Diameter Pressure rating Internal diameter

EN (DIN) AS 2129 AS 4087 ANSI JIS Measuring pipe [mm]

[mm] [inch] [bar] [lbs] PFA PTFE

25 1" PN 40 Table E Class 14 Cl.150 20K 23 26

32 – PN 40 Table E – – 20K 32 35

40 1 ½" PN 40 – – Cl.150 20K 36 41

50 2" PN 40 – – Cl.150 10K 48 52

65 – PN 16 – – – 10K 63 67

80 3" PN 16 – – Cl.150 10K 75 80

100 4" PN 16 – – Cl.150 10K 101 104

125 – PN 16 – – – 10K 126 129

150 6" PN 16 – – Cl.150 10K 154 156

200 8" PN 10 – – Cl.150 10K 201 202

250 10" PN 10 – – Cl.150 10K – 256

300 12" PN 10 – – Cl.150 10K – 306

350 14" PN 10 – – Cl.150 – – 337

400 16" PN 10 – – Cl.150 – – 387

450 18" PN 10 – – Cl.150 – – 432

500 20" PN 10 – – Cl.150 – – 487

600 24" PN 10 – – Cl.150 – – 593

18 Endress + Hauser
Proline Promag 10 P

Design, dimensions Transmitter, remote version

161...181
178
20...30 121

150
Esc
- + E

327.5
177.5
Ø 8.6 (M8)

100
133
ANSCHLUSSKLEMMEN - FIELD TERMINALS

113 135 100

25
123

a0003215

Transmitter dimensions, remote version

A B
248 ±2 238

Esc
Esc - + E
- + E

ANSCHLUSSKLEMMEN - FIELD TERMINALS


ANSCHLUSSKLEMMEN - FIELD TERMINALS

a0003216

Transmitter mounting, remote version


A Direct wall mounting
B Pipe mounting

Endress + Hauser 19
Proline Promag 10 P

Compact version

161...181
178 20...30 121

150
Esc
- + E

B
A
C
E
K L (DVGW)

a0003217

DN L A B C K E

EN (DIN) / JIS / ANSI [mm] [mm] [mm] [mm] [mm] [mm]


AS 1) [mm] [inch]

25 1" 200 341 257 84 120 94

32 - 200 341 257 84 120 94

40 1 ½" 200 341 257 84 120 94

50 2" 200 341 257 84 120 94

65 - 200 391 282 109 180 94

80 3" 200 391 282 109 180 94

100 4" 250 391 282 109 180 94

125 - 250 472 322 150 260 140

150 6" 300 472 322 150 260 140

200 8" 350 527 347 180 324 156

250 10" 450 577 372 205 400 156

300 12" 500 627 397 230 460 166

350 14" 550 738.5 456.5 282 564 276

400 16" 600 790.5 482.5 308 616 276

450 18" 650 840.5 507.5 333 666 292

500 20" 650 891.5 533.5 358.5 717 292

600 24" 780 995.5 585.5 410.5 821 402

The length (L) is always the same regardless of the pressure rating selected.
1)
For flanges to AS, only the nominal diameters DN 80, 100 and 150...300 are available

20 Endress + Hauser
Proline Promag 10 P

Remote version

129 163
143

102
B
A
C
E
K L (DVGW)

a0003219

DN L A B C K E

EN (DIN) / JIS / ANSI


AS 1) [mm] [inch] [mm] [mm] [mm] [mm] [mm] [mm]

25 1" 200 286 202 84 120 94

32 - 200 286 202 84 120 94

40 1 ½" 200 286 202 84 120 94

50 2" 200 286 202 84 120 94

65 - 200 336 227 109 180 94

80 3" 200 336 227 109 180 94

100 4" 250 336 227 109 180 94

125 - 250 417 267 150 260 140

150 6" 300 417 267 150 260 140

200 8" 350 472 292 180 324 156

250 10" 450 522 317 205 400 156

300 12" 500 572 342 230 460 166

350 14" 550 683.5 401.5 282 564 276

400 16" 600 735.5 427.5 308 616 276

450 18" 650 785.5 452.5 333 666 292

500 20" 650 836.5 478 358.5 717 292

600 24" 780 940.5 530 410.5 821 402

The length (L) is always the same regardless of the pressure rating selected.
1)
For flanges to AS, only the nominal diameters DN 80, 100 and 150...300 are available.

Endress + Hauser 21
Proline Promag 10 P

Ground disk (DN 25...300)

ØB

6.5
Ø

15
Ø di
10

ØD
2

a0003221

DN 1) di B D H

EN (DIN) / JIS / ANSI


AS 4) [mm] [inch] [mm] [mm] [mm] [mm]

25 1" 26 62 77.5 87.5

32 - 35 80 87.5 94.5

40 1 ½" 41 82 101 103

50 2" 52 101 115.5 108

65 - 68 121 131.5 118

80 3" 80 131 154.5 135

100 4" 104 156 186.5 153

125 - 130 187 206.5 160

150 6" 158 217 256 184

200 8" 206 267 288 205

250 10" 260 328 359 240

300 2) 12" 2) 312 375 413 273

300 3) 12" 3) 310 375 404 268


1) Ground disks can be used for all flange standards/pressure ratings that can be delivered, except for DN 300.
2) PN 10/16, Cl. 150
3)
PN 25, JIS 10 K/20 K
4) For flanges to AS, DN 32, 40, 65 and 125 are not available.

22 Endress + Hauser
Proline Promag 10 P

Weight Weight data in kg

Remote version (without cable)

Diameter Compact version Sensor Transmitter

EN (DIN) ANSI / EN (DIN) ANSI /


[mm] [inch] / AS 1) JIS AWWA / AS 1) JIS AWWA Wall housing

25 1" 7.3 7.3 7.3 5.3 5.3 5.3 6.0

32 1 ¼" 8.0 7.3 – 6.0 5.3 – 6.0

PN 40

PN 40
40 1 ½" 9.4 8.3 9.4 7.4 6.3 7.4 6.0

50 2" 10.6 9.3 10.6 8.6 7.3 8.6 6.0

65 2 ½" 12.0 11.1 – 10.0 9.1 – 6.0

80 3" 14.0 12.5 14.0 12.0 10.5 12.0 6.0

10K

10K
PN 16

PN 16
100 4" 16.0 14.7 16.0 14.0 12.7 14.0 6.0

125 5" 21.5 21.0 – 19.5 19.0 – 6.0

Class 150

Class 150
150 6" 25.5 24.5 25.5 23.5 22.5 23.5 6.0

200 8" 45 41.9 45 43 39.9 43 6.0

250 10" 65 69.4 75 63 67.4 73 6.0

300 12" 70 72.3 110 68 70.3 108 6.0

350 14" 115 175 113 173 6.0


PN 10

PN 10
400 16" 135 205 133 203 6.0

450 18" 175 255 173 253 6.0

500 20" 175 285 173 283 6.0

600 24" 235 405 233 403 6.0


1)
For flanges to AS, only DN 25 and 50 are available.
• Transmitter (compact version): 1.8 kg
• Weight data valid for standard pressure ratings and without packaging material

Material • Housing: powder-coated die-cast aluminum


Sensor housing
• DN 25...300: powder-coated die-cast aluminum
• DN 350...2000: coated steel (Amerlock 400)
• Measuring tube
– DN < 350: stainless steel 1.4301 or 1.4306/304L; Flange material with Al/Zn protective coating
– DN > 300: stainless steel 1.4301 or 1.4306/304; Flange material with Amerlock 400 coating
• Flanges
– EN 1092-1 (DIN2501): RSt37-2 (S235JRG2) / C22 / Fe 410W B
(DN < 350: with Al/Zn protective coating; DN > 300 with Amerlock 400 coating)
– ANSI: A 105
(DN < 350: with Al/Zn protective coating; DN > 300 with Amerlock 400 coating)
– JIS: RSt37-2 (S235JRG2) / HII / 1.0425
(DN < 350: with Al/Zn protective coating; DN > 300 with Amerlock 400 coating)
– AS 2129
– (DN 25, 150, 200, 250, 300, 600) A105 or RSt37-2 (S235JRG2)
– (DN 50, 80, 100, 350, 400, 500) A105 or St44-2 (S275JR)
(DN < 350: with Al/Zn protective coating; DN > 300 with Amerlock 400 coating)
– AS 4087: A105 or St44-2 (S275JR)
(DN < 350: with Al/Zn protective coating; DN > 300 with Amerlock 400 coating)
• Ground disks: 1.4435/316L or Alloy C-22
• Electrodes: 1.4435/316L, Alloy C-22
• Seals: to DIN EN 1514-1

Endress + Hauser 23
Proline Promag 10 P

Material load diagram Caution!


The following diagrams contain material load diagrams (reference curves) for flange materials with regard to
the medium temperature. However, the maximum medium temperatures permitted always depend on the
lining material of the sensor and/or the sealing material.
Flange connection to EN 1092-1 (DIN 2501)
Material: RSt37-2 (S235JRG2) / C22 / Fe 410W B

[bar]
40
PN 40
35

30

25
PN 25
20

15
PN 16
10
PN 10
5 PN 6

0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

F06-xxFxxxxx-05-xx-xx-xx-000

Flange connection to EN 1092-1 (DIN 2501)


Material: 316L / 1.4571

[bar]

40
PN 40
35

30

25
PN 25
20

15 PN 16

10
PN 10

5 PN 6

0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

A0005304-en

24 Endress + Hauser
Proline Promag 10 P

Flange connection to ANSI B16.5


Material: A 105

[bar] [psi]

60 870

50 725
Class 300
40 580

30 435

20 290
Class 150
10 145

0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

a0003226

Flange connection to ANSI B16.5


Material: F316L

[bar]

60

50

40 Class 300

30

20
Class 150
10
0
-40 -20 0 20 40 60 80 100 120 140 160 180 [°C]
A0005307-en

Flange connection to JIS B2238


Material: RSt37-2 (S235JRG2) / HII / 1.0425

[bar]

30

20K
20

10K
10

0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C]

a0003228

Endress + Hauser 25
Proline Promag 10 P

Flange connection to AS 2129 Table E or AS 4087 Cl. 14


Material: A105 / RSt37-2 (S235JRG2) / St44-2 (S275JR)

[bar]
25

20

15

10

0
-60 -40 -20 0 20 40 60 80 100 120 140 160 [°C]
F06-xxFxxxxx-05-xx-xx-xx-010

Fitted electrodes Measuring electrodes, reference electrodes and empty pipe detection electrodes available as standard with:
• 1.4435
• Alloy C-22

Process connections Flange connection:


• EN 1092-1 (DIN 2501), < DN 350 form A, > DN 300 form B
(Dimensions to DIN 2501, DN 65 PN 16 and DN 600 PN 16 exclusively to EN 10921)
• ANSI B16.5
• JIS B2238
• AS 2129 Table E
• AS 4087 Cl. 14

Surface roughness Electrodes with 1.4435 (AISI 316L), Alloy C-22: ≤ 0.3...0.5 μm
(All data refer to parts in contact with medium)

Human interface
Display elements • Liquid crystal display: unilluminated, two-line, 16 characters per line
• Display (operating mode) preconfigured: volume flow and totalizer status
• 1 totalizer

Operating elements Local operation via three keys (S, O, F)

Remote operation Operation via HART protocol and ToF Tool - Fieldtool Package

26 Endress + Hauser
Proline Promag 10 P

Certificates and approvals


CE mark The measuring system is in conformity with the statutory requirements of the EC Directives.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.

C-tick mark The measuring system meets the EMC requirements of the Australian Communications Authority
(ACA)

Ex approval Information about currently available Ex versions (ATEX, FM, CSA, etc.) can be supplied by your Endress
+Hauser Sales Center on request. All explosion protection data are given in a separate documentation which
is available upon request.

Other standards and • EN 60529


guidelines Degrees of protection by housing (IP code)
• EN 61010
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory
Procedures.
• EN 61326/A1 (IEC 1326)
“Emission in accordance with requirements for Class A”.
Electromagnetic compatibility (EMC requirements)
• ANSI/ISA-S82.01
Safety Standard for Electrical and Electronic Test, Measuring, Controlling and related Equipment - General
Requirements Pollution degree 2, Installation Category II.
• CAN/CSA-C22.2 No. 1010.1-92
Safety requirements for Electrical Equipment for Measurement and Control and Laboratory Use.
Pollution degree 2, Installation Category II

Pressure measuring device Measuring devices with a nominal diameter smaller than or equal to DN 25 correspond to Article 3(3)
approval of the EC Directive 97/23/EC (Pressure Equipment Directive) and have been designed and manufactured
according to good engineering practice. Where necessary (depending on the medium and process pressure),
there are additional optional approvals to Category II/III for larger nominal diameters.

Endress + Hauser 27
Proline Promag 10 P

Ordering information
Your Endress+Hauser service organization can provide detailed ordering information and information on the
order codes on request.

Accessories
Various accessories, which can be ordered separately from Endress
+Hauser, are available for the transmitter and the sensor. Your Endress+Hauser service organization can
provide detailed information on the order codes in question

Documentation
• System Information Promag 10 (SI042D/06/en)
• Operating Instructions Promag 10 (BA082D/06/en)

Registered trademarks
HART®
Registered trademark of the HART Communication Foundation, Austin, USA
ToF Tool - Fieldtool® Package
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH

Instruments International

Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland

Tel. +41 61 715 81 00


Fax +41 61 715 25 00
www.endress.com
[email protected]

TI094D/06/en/08.09
71101431
FM+SGML6.0 ProMoDo

You might also like