Man pr6024-64s Inst en 949905346400

Download as pdf or txt
Download as pdf or txt
You are on page 1of 26

Installation manual

Cable junction box for digital load cells PR 6024/64S

Translation of original installation manual 9499 053 46400 Edition 3.4.0 01/14/2019

Minebea Intec GmbH, Meiendorfer Str. 205 A, 22145 Hamburg, Germany


Phone: +49.40.67960.303 Fax: +49.40.67960.383
Foreword
Must be followed!
Any information in this document is subject to change without notice and does not represent a commitment on
the part of Minebea Intec unless legally prescribed. This product should be operated/installed only by trained
and qualiied personnel. In correspondence concerning this product, the type, name, and release number/serial
number as well as all license numbers relating to the product have to be cited.

Note
This document is partially protected by copyright. It may not be changed or copied, and it may not be used
without purchasing or written permission from the copyright owner (Minebea Intec). The use of this product
constitutes acceptance by you of the above-mentioned provisions.
Cable junction box for digital load cells PR 6024/64S Table of contents

Table of contents

1 Introduction......................................................................................................................................... 3
1.1 Read the manual......................................................................................................................................................... 3
1.2 This is what operating instructions look like .......................................................................................................... 3
1.3 This is what lists look like........................................................................................................................................... 3
1.4 This is what menu items and softkeys look like..................................................................................................... 3
1.5 This is what the safety instructions look like.......................................................................................................... 3
1.6 Hotline .......................................................................................................................................................................... 4

2 Safety instructions ............................................................................................................................... 5


2.1 General notes .............................................................................................................................................................. 5
2.2 Intended use................................................................................................................................................................ 5
2.3 Initial inspection.......................................................................................................................................................... 6
2.4 Before operational startup........................................................................................................................................ 6
2.5 Repairs and maintenance.......................................................................................................................................... 6
2.5.1 General information ........................................................................................................................................ 6
2.5.2 Changing fuses ................................................................................................................................................ 6

3 Speciications ...................................................................................................................................... 7
3.1 Equipment supplied................................................................................................................................................... 7
3.2 Technical Data.............................................................................................................................................................8
3.3 Electromagnetic Compatibility (EMC) ..................................................................................................................... 9
3.4 Possible marking for the Ex area.............................................................................................................................. 9
3.5 Dimensions ................................................................................................................................................................ 10

4 Installation and connection information ............................................................................................. 11


4.1 General information .................................................................................................................................................. 11
4.2 Terminals and jumpers ............................................................................................................................................. 11
4.3 LED...............................................................................................................................................................................12
4.4 Fuses............................................................................................................................................................................12
4.5 Cable gland.................................................................................................................................................................13
4.6 Installation of a cable ................................................................................................................................................13

5 Cable connections ...............................................................................................................................15


5.1 General information ..................................................................................................................................................15
5.2 Data and supply cables.............................................................................................................................................15
5.3 Cable connections .....................................................................................................................................................15
5.4 Equipotential bonding conductor...........................................................................................................................17

6 Maintenance/repairs/soldering work/cleaning ................................................................................. 18


6.1 Maintenance.............................................................................................................................................................. 18
6.2 Repairs........................................................................................................................................................................ 18
6.3 Soldering work .......................................................................................................................................................... 18

EN-1 Minebea Intec


Cable junction box for digital load cells PR 6024/64S Table of contents

6.4 Cleaning ..................................................................................................................................................................... 18

7 Disposal ............................................................................................................................................. 19

8 Appendix ...........................................................................................................................................20
8.1 Certiicates/safety instructions ..............................................................................................................................20

Minebea Intec EN-2


1 Introduction Cable junction box for digital load cells PR 6024/64S

1 Introduction
1.1 Read the manual
- Please read this manual carefully and completely before using the product.
- This manual is part of the product. Keep it in a safe and easily accessible location.

1.2 This is what operating instructions look like


1. - n. are placed before steps that must be done in sequence.
is placed before a step.
describes the result of a step.

1.3 This is what lists look like


- indicates an item in a list.

1.4 This is what menu items and softkeys look like


[ ] frame menu items and softkeys.
Example:
[Start]- [Applications]- [Excel]

1.5 This is what the safety instructions look like


Signal words indicate the severity of the danger involved when measures for preventing
hazards are not followed.

DANGER
Warning of personal injury
DANGER indicates death or severe, irreversible personal injury which will occur if the
corresponding safety measures are not observed.
Take the corresponding safety precautions.

WARNING
Warning of hazardous area and/or personal injury
WARNING indicates that death or severe, irreversible injury may occur if appropriate
safety measures are not observed.
Take the corresponding safety precautions.

CAUTION
Warning of personal injury.
CAUTION indicates that minor, reversible injury may occur if appropriate safety
measures are not observed.
Take the corresponding safety precautions.

EN-3 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 1 Introduction

NOTICE
Warning of damage to property and/or the environment.
NOTICE indicates that damage to property and/or the environment may occur if
appropriate safety measures are not observed.
Take the corresponding safety precautions.

Note:
User tips, useful information, and notes.

1.6 Hotline
Phone: +49.40.67960.444
Fax: +49.40.67960.474
eMail: [email protected]

Minebea Intec EN-4


2 Safety instructions Cable junction box for digital load cells PR 6024/64S

2 Safety instructions
2.1 General notes

CAUTION
Warning of personal injury.
The product was in perfect condition with regard to safety features when it left the
factory.
To maintain this condition and to ensure safe operation, the user must follow the
instructions and observe the warnings in this manual.

2.2 Intended use


The junction box is designed for industrial weighing systems and weighing systems for
use in legal metrology which use a digital load cell.
It can also be operated in potentially explosive atmospheres.
The power supply PR 6024/62S is used to supply power to the load cells.
Product operation, commissioning and maintenance must be performed by trained and
qualiied personnel who are aware of and able to deal with the related hazards and take
suitable measures for self-protection.
The device relects the state of the art.
The manufacturer does not accept any liability for damage caused by third-party system
components or due to incorrect use of the product. The use of this product signiies
recognition of the stipulations listed above.
The following table shows the load cells that should and should not be used for diferent
applications.

Load cells to be used Load cells not to be used


PR 6201/.., all types
Inteco®/.., all types
PR 6204 Pendeo® Process
PR 6207/.., all types
PR 6211/.., all types
PR 6212/.., all types
PR 6221/.., all types
PR 6224 Pendeo® Truck
PR 6241/.., all types
Contego®/..D1, /..D1Ex, /..C3, /..C3Ex
PR 6246/.., all types
PR 6251/.., all types
Novego®/..D1, /..D1E, /..C3, /..C3E
MP load cells, all types

EN-5 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 2 Safety instructions

2.3 Initial inspection


Check the contents of the consignment for completeness. Check the contents visually to
determine whether any damage has occurred during transport. If there are grounds for
rejection of the goods, a claim must be iled with the carrier immediately. The
Minebea Intec sales or service organization must also be notiied.

2.4 Before operational startup

NOTICE
Perform visual inspection.
Before operational startup as well as after storage or transport, inspect the device
visually for signs of mechanical damage.

2.5 Repairs and maintenance


2.5.1 General information
Repairs are subject to inspection and must be carried out at Minebea Intec.
In case of defect or malfunction, please contact your local Minebea Intec dealer or service
center for repair.
When returning the device for repair, please include a precise and complete description of
the problem.
Maintenance work may only be carried out by a trained technician with expert knowledge
of the hazards involved and the required precautions.
2.5.2 Changing fuses

WARNING
Damage from overheating.
The use of repaired fuses and bypassing the fuse holder is prohibited.
Only the fuses listed in Chapter 3.2 are permissible.

Minebea Intec EN-6


3 Speciications Cable junction box for digital load cells PR 6024/64S

3 Speciications
3.1 Equipment supplied

No. Description
1 Cover
2 Box incl. electronics
3 Ex-zone locking pin 6 mm (3×)
4 Cable gland M16 (6×)
5 Gasket
6 Pressure compensation element
The following items are not shown:
7 Drilling template
8 Installation manual
9 Certiicates

EN-7 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 3 Speciications

3.2 Technical Data

Protection classes per DIN EN 60529


IP68, IP69:
Dust-proof and leak-proof against water, with harmful
effects when immersed, (0.5 m water depth, 1,000 h)
and water jets (high pressure and temperature).
IP66:
Dust-proof and leak-proof against strong water jets.
Installation position Cable entry from below
Quantity of load cells 1 4
Clamping area screw connection M16 5 9 mm
Supply voltage 24 V DC
Overvoltage protection 1× 1 AT
Pressure equalization Stainless steel pressure equalization element
Material of the junction box High-grade stainless steel 1.4301 (X5CrNi 18-10) accor-
ding to EN 10088-3, AISI 304, JIS SUS304
Surface of the junction box 2R according to DIN EN 10088-2 (cold-rolled, bright
annealed, lat, blank, relective)
Net weight 1.1 kg
Shipping weight 1.3 kg
Service temperature area -30°C +80°C
Service temperature area, in explosion-prone area -20°C +60°C
Storage temperature range -30°C +80°C
Cable screw connections Metal Ex EMC cable screw connections
Ex-connection values See Chapter 8.1.
Seal EMC seal
Cable length Supply cable PR 6124/xxP for mains supply PR 6024/
62S: max. 200 m
Data cable PR 6124/xxD for device: max. 300 m
Service temperature area of cables Supply cable PR 6124/xxP: -30°C +90°C
Data cable PR 6124/xxD: -30°C +90°C

Minebea Intec EN-8


3 Speciications Cable junction box for digital load cells PR 6024/64S

3.3 Electromagnetic Compatibility (EMC)


All data in compliance EN 61326 industrial section

Housing High frequency electromagnetic ields EN 61000-4-3 10 V/m


(80 3000 MHz)
Electrostatic discharge (ESD) EN 61000-4-2 6/8 kV

Signal and control Fast transients (burst) EN 61000-4-4 1 kV


lines Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1 kV
Conducted disturbances by high fre- EN 61000-4-6 10 V
quency coupling (0.15 80 MHz)

3.4 Possible marking for the Ex area

Zone Marking Certiicate No.


2 and 22 II 3G Ex nA IIC T5 Gc SWT 12ATEX004 X
II 3D Ex tc IIIC T90 °C Dc

NOTICE
Installation in Ex Area
The Ex safety instructions in the appendices must be observed when installing in the Ex area.

EN-9 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 3 Speciications

3.5 Dimensions

all dimensions in mm

Minebea Intec EN-10


4 Installation and connection information Cable junction box for digital load cells PR 6024/64S

4 Installation and connection information


4.1 General information
- The safety instructions in Chapter 8.1 must be followed!
- Only use the metal EX EMC cable glands provided by the manufacturer.
- Install the junction box so that the cable glands are on the bottom.
- Do not open the junction box when connected to the voltage supply.

Note:
At ambient temperatures >45°C, cables that are suitable for at least 85°C must be used.

4.2 Terminals and jumpers

+ Red Terminal contacts + supply voltage (load cells)


LC1–4
Terminal contact + supply voltage (supply cable PR 6124/xxP for
PS the PR 6024/62S power supply)
- Blue Terminal contacts - supply voltage (load cells)
LC1–4
Terminal contact - supply voltage (supply cable PR 6124/xxP for
PS the PR 6024/62S power supply)
B green Terminal contacts RS-485 TxB (data transmission)
LC1–4
Terminal contact RS-485 TxB (data cable PR 6124/xxD for the device)
IND
A gray Terminal contacts RS-485 TxA (data transmission)
LC1–4
Terminal contact RS-485 TxA (data cable PR 6124/xxD for the device)
IND

EN-11 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 4 Installation and connection information

4.3 LED

The green LED (1) lights up when the supply voltage is 24 V DC.

4.4 Fuses

The power supply is protected by a time-delayed 1.0 A fuse (2) in a fuse holder with
protection against contact.
When a fuse has tripped:
- Disconnect the junction box from all power sources.
- Determine and eliminate the cause of the error.
- Replace the fuse.
Make sure that fuses are replaced only using fuses of the same type, see Chapter 3.2.
- Reconnect the junction box to all power sources.

Minebea Intec EN-12


4 Installation and connection information Cable junction box for digital load cells PR 6024/64S

4.5 Cable gland


The cables have to be fed into the device via glands to ensure leak-tightness. Cable
diameters of 5 9 mm are suitable for gland M16.
The cable wires are connected to the terminals inside the device.

NOTICE
For protection against dust and moisture during transport and installation, the
cable glands are itted with a polyethylene cover.
For full IP protection, operation with the dust protection cover itted is not permitted.
Remove the dust protection cover.
If a cable gland is not used, it must be sealed with a supplied locking pin.

NOTICE
Property damage is possible.
Regularly check the itted cable gland for tightness and re-tighten it, if necessary.

4.6 Installation of a cable

NOTICE
Material damage is possible.
Do not guide the screen (4) into the device!
The cable shield (4) must be connected in the metal sleeve (6) of the cable gland.
Before, during and after installation, make sure that the sealing ring is seated
correctly.

EN-13 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 4 Installation and connection information

1. Unscrew the sleeve screw cap (1).


2. Slide the cap (1) and plastic cone (3) onto the cable (2).
3. Guide the cable (5) through the gland (6).
4. Fold the cable shield (4) over the lower part of the terminal insert (3) (approx. 10 mm).
5. Connect the cable conductors.
6. Tighten the sleeve screw cap (1).
7. Secure the gland (6) including the o-ring (7) using the counter nut (in the housing).

NOTICE
Material damage is possible.
Regularly check the cable gland for tightness and re-tighten it, if necessary.

Minebea Intec EN-14


5 Cable connections Cable junction box for digital load cells PR 6024/64S

5 Cable connections
5.1 General information
Cable entry must be from below.
Connect the wires to the terminals according to the color coding .
Connect all cable shields in the metal sleeve of the junction box cable glands, see
Chapter 4.6.
Establish potential equalization, see Chapter 5.4.

Note:
Chapter 5.3 shows an example of the cabling of 4 load cells via the PR 5510/04 interface
and the PR 6024/64S junction box.

5.2 Data and supply cables


The PR 6124/xxD data cable must be used to connect the PR 6024/64S or PR 6024/68S
cable junction box to the device, max. length: 300 m.
The PR 6124/xxP power cable must be used to connect the PR 6024/62S power supply to
the PR 6024/64S or PR 6024/68S cable junction box, max. length: 200 m

5.3 Cable connections

Note:
All components are only shown schematically.

Color Code

rd = red
gn = green
bu = blue
gy = gray

EN-15 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 5 Cable connections

Minebea Intec EN-16


5 Cable connections Cable junction box for digital load cells PR 6024/64S

5.4 Equipotential bonding conductor

NOTICE
It is especially important that the ground is connected correctly to the components
and the cable junction box.
You must also ground the device separately and ensure the power supply is properly
shielded against the efects of lightning. Simply connecting the protective grounding
conductor is not enough!
If the installation is not carried out according to our instructions, this voids the
warranty. In particular, the entire installation, including the power supply, must be
suiciently protected against lightning.
In Ex areas, equipotential bonding must be established to avoid any compensatory
currents between the individual conducting system components.

EN-17 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 6 Maintenance/repairs/soldering work/cleaning

6 Maintenance/repairs/soldering work/cleaning
6.1 Maintenance
Maintenance work may only be carried out by a trained technician with expert knowledge
of the hazards involved and the required precautions.

6.2 Repairs
Repairs are subject to inspection and must be carried out at Minebea Intec.
In case of defect or malfunction, please contact your local Minebea Intec dealer or service
center for repair.
When returning the device for repair, please include a precise and complete description of
the problem.

6.3 Soldering work


Soldering work on the device is neither required nor permitted.

6.4 Cleaning

NOTICE
Property damage caused by unsuitable cleaning utensils/agents.
Damage to the device.
Prevent moisture from penetrating the interior.
Do not use aggressive cleaning agents (solvents or similar agents).
For use in the food industry, use a cleaning agent suitable for that particular working
environment.
Use soft sponges, brushes and cloths.

1. Unplug device from mains supply, disconnect any data cables.


2. Clean the device with a cloth lightly moistened with a soap solution.
3. Wipe down the device with a soft, dry cloth after cleaning.

Minebea Intec EN-18


7 Disposal Cable junction box for digital load cells PR 6024/64S

7 Disposal
If the packaging is no longer required, please take it to your local waste disposal facility
and/or a reputable disposal company or collection point. The packaging largely consists
of environmentally friendly materials which can be used as secondary raw materials.
It is not permitted—even for small businesses—to dispose of this product with the regular
household waste or at collection points run by local public waste disposal companies.
EU legislation requires its Member States to collect electrical and electronic equipment
and dispose of it separately from other unsorted municipal waste so that it can then be
recycled.

Before disposing of or scrapping the product, any batteries should be removed and taken
to a suitable collection point.
Please see our T&Cs for further information.

Service addresses for repairs are listed in the product information supplied with the
product and on our website (www.minebea-intec.com).
We reserve the right not to accept products that are contaminated with hazardous
substances (ABC contamination) for repair.
Should you have any further questions, please contact your local service representative or
our service center.
Minebea Intec GmbH
Repair center
Meiendorfer Strasse 205 A
22145 Hamburg, Germany
Phone: +49.40.67960.666
[email protected]

EN-19 Minebea Intec


Cable junction box for digital load cells PR 6024/64S 8 Appendix

8 Appendix
8.1 Certiicates/safety instructions

Ser. no. Description Document no.


1 EU-Declaration of Conformity MEU17054
2 Declaration of Conformity MDC17005
3 Manufacturer's Certiicate SWT 12 ATEX004 X
4 Safety instructions 65835-790-16

Minebea Intec EN-20


Explosion-safe area Explosion-risk area Zone 2 or Zone 22

Gas: Group II, temperature class T5


Dust: Maximum device surface temperature 90ᄚC

1. Main connection

Connector
Use explosionprotected connector

Alternative ways of
connecting device

Protect connector from being 4


disconnected. Make sure the
seines Inhaltes sind verboten, soweit nicht ausdrücklich gestattet.
Zuwiderhandlungen verpflichten zu Schadensersatz. Alle Rechte für den Fall der
Weitergabe sowie Vervielfältigung dieses Dokuments, Verwertung und Mitteilung

emergency stop works


Patent-, Gebrauchsmuster oder Geschmacksmustereintragung vorbehalten.

7
9
Power supply
PR6024/62S 1
Ta: -20°C ... +50°C

2. Data transfer only for data transfer, there is 8


no supply voltage going from 6
the connected device to the
cable junction box. 5
Indicator
8 8

9 2
This device is also suitable
Cable junction box
for installation in Zone 2
PR6024/64S or PR6024/68S
and Zone 22 if it is rated for
category 3G or 3D in Ta: -20°C ... +60°C
accordance with the ATEX
directive such as Combics
Pro with option Y2
PR6024/64S: 8
held liable for the payment of damages. All rights reserved in the event of the grant of a
of its contents to others without expressed authorization is prohibited. Offenders will be
The copying, distribution and utilization of this document as well as the communication

up to 4 digital load cells


PR6024/68S:
up to 8 digital load cells 6
patent, utility model or ornamental design registration.

9 9
Digital load cell Digital load cell 3
PR6204/..t C. or PR6204/..t C. or
PR6224/..t C. PR6224/..t C.
Ta: -20°C ... +70°C Ta: -20°C ... +70°C

Datum Material Maßstab /


Name
Date . Scale
Erstellt
29.11.18 Schallhorn Benennung / Title 1:1
Written by
Geprüft
Safety Precautions Blatt
Reviewed by
03.12.18 Hiller Sheet 1
Ausgabe / Revision Änderung / Alteration Zeichnungs-Nr. / Drawing number Teil Dok Nr./
Freigabe von
Released by
03.12.18 Schallhorn 02 PA50181110 65835-790-16 Part Doc No. --- of 2
These safety instructions apply to the installation, use, maintenance and repair of the appliance.
The numbers shown in brackets refer to the numbered positions on page 1.

1. The power supply unit (1), cable junction box (2) and the digital load cell (3) may only be used in this constellation in
Zone 2 and Zone 22 and not as portable devices.

2. Installation should be performed by an authorized specialist, in accordance with applicable laws, rules and
regulations, ordinances and standards. Particular attention should be paid to the European Standards EN 60079-14

3. It is essential that the recommendations on the installation, use, maintenance and repair contained in the manuals
supplied are complied with for all equipment.

4. Only use equipment in the temperature ranges indicated. Avoid exposure to inadmissible sources of heat or cold,
solar radiation or vibrations. Install the devices in such a way that heat can be sufficiently dissipated on all sides and
that external heat sources are located far enough away from the power supply.

5. Protect the mains connecting cable (4) from damage and connect it correctly to the mains power supply (100 – 240
VAC). The power supply is approved for circuits up to 1,500 A. Only use the mains connecting cable in the hazardous area
seines Inhaltes sind verboten, soweit nicht ausdrücklich gestattet.
Zuwiderhandlungen verpflichten zu Schadensersatz. Alle Rechte für den Fall der
Weitergabe sowie Vervielfältigung dieses Dokuments, Verwertung und Mitteilung

with a suitable and approved explosion-protected plug. Alternatively: Protect connector from being pulled out or directly
wire in the cable; if directly wired-in, provide adequate emergency stop option.
Patent-, Gebrauchsmuster oder Geschmacksmustereintragung vorbehalten.

6. If not supplied, the mains connection cable (4), supply cable (5) and data cable (6) used must be suitable with
respect to the relevant zones they are used in, the manner in which they are laid and their relevant mechanical loads.
Cables sourced from other manufacturers are the operator's responsibility.

7. The ATEX-certified cable glands (7) and (8) are only suitable for fixed installation when used in potentially explosive
atmospheres, meaning that the equipment installer has to fit the cable behind the cable gland (e.g. using a cable clip).
These cable glands can be replaced by those which are ATEX-certified and are suitable for flexibly installed cable.

8. Any connections that are not being used must be sealed in such a way as to assure ingress protection (IP)
(e.g. using ATEX-certified locking screws).

9. Any metal parts (casing) must be DC coupled to the potential equalization. To do so, the operator of the system must
connect a conductor with a cross-section of at least 4 mm2 in a suitable place using adequate mounting material (screws).
During on-site installation, check if this connection to the equipotential bonding conductor is of low
resistance. The equipment operator shall be responsible for preventing any risks caused by static electricity.

10. Before opening the instruments, switch off the supply voltage, or make sure that the area is not potentially explosive.
Do not connect or disconnect any live cables inside an explosion-risk area.

11. If the installation does not operate properly, disconnect it from the supply voltage immediately.

12. During installation, take suitable measures to prevent stray electrical interference (e.g. due to magnetic fields).

13. Chemicals that can attack housing gaskets and cable sheathings must be kept away from the instrument.
These include oil, vegetable and animal fats, petrol, chlorinated and aromatic solvents, lye and acids, acetone
and ozone. If you are uncertain, contact the manufacturer.
held liable for the payment of damages. All rights reserved in the event of the grant of a
of its contents to others without expressed authorization is prohibited. Offenders will be
The copying, distribution and utilization of this document as well as the communication

14. The installation must be checked for correct function and safety by a trained and qualified person at appropriate
intervals.

15. In the event of repair, use only original spare parts supplied by the manufacturer.

16. Any modifications to the instrument (except by persons authorized by Minebea Intec) cause loss of conformity for use
in Zone 2 and Zone 22 explosion-risk areas and invalidate all guarantee claims. Similarly, the instrument may only be
opened by qualified and authorized persons.
patent, utility model or ornamental design registration.

17. Modifications (including by Minebea Intec personnel) are subject to written approval.

18. Any devices used in Zone 22 must be regularly cleared of dust. Dust layers of more than 5 mm are not permitted.

19. These instructions are given in addition to those in the instruction manuals and do not release the operator from his
responsibilities for the installation, operation and inspection of the equipment assembly in compliance with the valid
standards, directives, ordinances and laws.
Datum Material Maßstab /
Name
Date . Scale
Erstellt
29.11.18 Schallhorn Benennung / Title 1:1
Written by
Geprüft
Safety Precautions Blatt
Reviewed by
03.12.18 Hiller Sheet 2
Ausgabe / Revision Änderung / Alteration Zeichnungs-Nr. / Drawing number Teil Dok Nr./
Freigabe von
Released by
03.12.18 Schallhorn 02 PA50181110 65835-790-16 Part Doc No. --- of 2
Published by
Minebea Intec GmbH | Meiendorfer Strasse 205 A | 22145 Hamburg, Germany
Phone: +49.40.67960.303 | Email: [email protected]
www.minebea-intec.com