MY USG

Quick User Guide

1/16
IMPORTANT!
Read and understand this manual before operating
the equipment. After reading, keep this manual in an
easily accessible place.
Ver: Myusg-V1.0

Operation taboo:

Danger ※ Do not modify this equipment, including equipment components, software, cables

and so on. User modifications may result in security problems or reduced system performance. All
modifications must be completed by the personnel approved by KONTED.

2/16
Probe Components

1
6

2
3

8
5
9

10

1. Mid-line Mark 6. A probe


2. Orientation Mark 7. Wireless charging symbol
3. Power switch / freeze /Living button 8. Serial number
4. Type-C port 9. B probe
5. Battery Indicator Lights 10. Wireless charging pad

1 Download MY USG App


Note:
Search and install “MY USG” iOS version 11.0 or newer
on your smartphones/tablets/laptops. Android version 8.0 or
Please contact us to get the windows software newer
Windows10 Laptop,5G,
64-bit or newer

2 Turn on/off USG probe

Press the power button to turn on the probe, Note:


the indicator will show the battery icon. Press the power button
and hold for 5 seconds to turn off the probe and hold for 3 seconds to
switch the scanning probe
3/16
3 Connect USG probe

Method 1:Wi-Fi connection

Find SN of your USG probe Choose SSID “****BMH000” Enter password: uxccbmh000
Such as SN: UXCCBMH000

uxccbmh000

Note:
password is the SN of the
probe, but it is the small
letter not capital.

Method 2:USB connection

Connect the probe to your smart device by Type-C cable according to the picture below.

A B

Note:
The A-end must be fully inserted into the probe and the B-end must be fully inserted into the
smart device.
4/16
4 Scan now

Live
11 12 13 14 24 15 16 17

2 18
20
3 7 12
8
2
4 3

8
9

10

23 22 21 20 19

No Item Description
1 Serial number of probe 13 Preset selection
The Thermal Index (TI), Mechanical
2 Dynamic Range (DR) 14
Index (MI), and Hz values
DICOM Procedures, uploading images
3 Imaging Enhance (ENH) 15
to PACS
Mode: B⟺B/M, Color⟺PDI
4 16 Body Mark
Dual click to switch scanning mode
5 Freeze/Live button 17 APP version number
6 Mode:PW 18 Show patient information
7 Focus Position 19 Full screen
8 H: change scanning frequency(F) 20 B Gain (GN)
Biopsy Needle Guide line 21 Live/freeze display
9 To invert the image horizontally or vertically.
U/D flip, R/L flip 22 Mid-line
10 Depth Scale display Dual screen display mode
23 (only available for Apple ipad and android pad
11 8 TGC: adjust gains of different depth device)
Depth: adjust depth by swiping the 24 New patient & new report
12
button (D)
5/16
23 Dual screen display mode

Click button to switch the dual screen / single screen mode

Click the left/right screens to switch freely, and you can switch B/BM/Color/PW mode at the
same time

B+Color mode

B+B mode
23

Color+PW mode

24 New patient & new report

a. Enter new patient info, click OK

b. Scan now, then Freeze image

c. Edit the report, click to save the report to Data Table

24

6/16
Color/PDI Mode
1 GN Color Gain ±
Click Steer Adjust the deflection
2 Steer
angle of color ROI
5 Click PRF Adjust color pulse
3 PRF
repetition frequency
1 Click WF Adjusting the filtering
frequency of a pulse wave or
2 4 WF
continuous wave Doppler low
6
3 frequency signal
4 Tap and drag the green box to
5
move the Color ROI
Use the arrows provided to adjust
6
the angle and size
7 Invert Color Red⟺Blue invert
7
8 Slide the Color/PDI menu bar

B
+ First select Color mode
Color Then switch to PW mode
8 + (only available for convex probe of
PW dual head serial USG probe)
PW Mode Mode
9 GN PW Gain ±
Click Steer Adjust the deflection
10 Steer angle of sampling line of linear
array probe
Change the size of the sampling
11 Volume
volume
Used to change the spectrum
12 Angle
9 sampling line angle
14
Adjust color pulse repetition
10 13 PRF
frequency
11
14 Tap and drag the green line
12 Move the velocity scale
15Baseline
13
16 Slide the PW menu bar
Automatic blood flow measurement
Step1: select vessel flow preset
15 Step2:enter PW mode, produce a
17 uniform spectrum
Flow-mode Step3: click Flow mode ON
ON/OFF Step4: Freeze the PW image then
will show the result automatically
17
16 (only available for linear probe of
dual head serial USG probe)

7/16
Freeze

add annotations on any frozen


1
image

Save image, review the image in


2
Date Table
1

3 Measurements

2
Save video, review the video in Date
4
Table 3

4
5 Date Table

Auto cine review 5


6

6
7 Manual cine review 7

Image Browser

8 Review image/video/report

①Click 5 Date Table

②Click 8 check record


Watch
③Please swipe to the left to
browse the images/videos and
report.

9
Sharing scanned files by other
app
Share
Printing image/report

Delete local files


Delete
The data table can store 500 files

8/16
5 Settings

Adjust cine loop frame number


1.Cine loop 100-200-500-1000
(100=10s)

Hide the patient information and


2.Show
parameter information on the
information
screen
1

3 Measuring magnify function can


4 3.Measure help you measure the boundary of
Magnify smaller target.
5

When scanning image


transmission is delayed or stuck,
4.WiFi Please contact us firstly then
channel switch one channel, press
setting “select” ,restart the USG probe
and reconnect the WiFi with your
phone.

a.Connect mobile phone with


Workplace WiFi
b. Fill in the
DICOM1-DICOM2-Worklist
5.DICOM information and tap Ping buttons
setting to test run the set ups.
c. If success messages appear,
then the DICOM/PACS
connections are completed.

9/16
6 Charging

Charging by USB cable

Charging by wireless charging pad:


The probe supports wireless charging.
1. Disconnect the probe from your mobile device.
2. Connect the Micro USB cable to the wireless charging pad.
3. Connect the USB end of the cable to the wall adapter.
4. Plug the wall adapter into a power outlet.
5. Place the probe onto the white wireless charging pad.

Power outlet

10/16
Manufacturer

Manufacturer: Beijing Konted Medical Technology Co., Ltd.


Registered address:1F,Building 3,No.27 Yongwang Road, Daxing Biological Pharmaceutical
Industry Base, Daxing District,Beijing,China
Production address: 1F,Building 3,No.27 Yongwang Road, Daxing Biological Pharmaceutical
Industry Base, Daxing District,Beijing,China
Zip code:102629 Tel: 8610-60219113 Fax:8610-60219213 Web:www.konted.cn
Product: Pocket Ultrasound System
Model: C10
Authorized European Representative:
Shanghai International Holding Corp. GmbH(Europe) Eiffestraβe 80, 20537 Hamburg Germany
CE certificate No. G2 003973 0002 Rev.01

11/16
Manual Cleaning

Manual Cleaning
The operation process is as follows:
The following cleaning methods shall be taken immediately after each complete test on the
patient to avoid drying of the coupling remaining on the probe, which is not conducive to the cleaning
of the probe and to prevent the body and sound head from being corroded by coupling or other
reagents . Before the cleaning or disinfection work begins, the operator should wear professional
protective equipment such as goggles and gloves, and check whether the instruments, equipment,
materials and environment used meet the requirements in advance.

1、 Turn off the power of the ultrasound probe, press and hold the button for more than 3s until the
light goes out. For the specific operation process, please refer to the product instruction manual.
At the same time, check whether the waterproof cap of the charging port of the machine is
installed and returned. If it is opens, the waterproof cap should be installed and returned.

2、 Use clean special wiping paper for ultrasonic instrument (the wiping paper shall be soft non
abrasive disposable paper towel) or clean lint free soft cloth to wipe the acoustic head. Gently
wipe the paper or soft lint-free cloth along the wide side of the vocal head to the other end ( if the
wiping force is too large, it may cause the depression of the vocal head and make the machine
unusable) . The wiping direction should be in accordance with the Repeated wiping is prohibited
in the same direction. If the sound head is also stained with coupling or other reagents, it should
also be wiped in the same way.

3、 During the wiping process, after each wiping from the beginning to the end, fold the special wipe
paper for ultrasonic instruments in half, and fold the wipe paper in half towards the side stained
with coupling or other ultrasonic special colloid residues (Fold the stained side inside the wiping
paper).

4、 Ultrasound instrument special wipes or lint-free soft cloth should not be folded in half at most for
more than 4 times. If there is still coupling or other ultrasonic special colloids remaining on the
ultrasonic probe part after wiping for the 4th time, you need to replace with new ultrasonic
instrument special wipes, and repeat after replacement. The above steps are performed until the
coupling remaining in the vocal head is completely wiped.

5、 Use ultrasonic special wiping paper or lint free cloth dipped with enzyme containing
detergent(the usage of specific cleaners should refer to the usage method of cleaner labels and
instructions, and pay attention to whether the use time of detergent is within the warranty period
of the product) to clean until there is no visible stain on the acoustic head and the acoustic
head and its surroundings have been wiped.

6、 Rinse the probe with drinking water or softened filtered water ( the water quality of tap water is
too hard, there may be residual salt and alkali substances, which will cause different degrees of
corrosion to the ultrasonic probe. For specific water quality requirements, please refer to AAMI
TIR 34 ) . During the process, keep wiping the vocal head with your hands until all the cleaning
agents are rinsed. In principle, the rinsing time should not be less than 30s. During the rinsing
process, water should be prevented from splashing on the fuselage. If any liquid enters the
fuselage,it will cause irreversible damage to the electronic components of the machine.
12/16
7、 Use the special wipe paper for ultrasonic instruments or a clean lint-free soft cloth (The same lint
free soft cloth used before is no longer used) to wipe the entire body and the vocal head, and
wipe the residual water stains on the acoustic head until all the residual water stains on the body
are wiped clean.

8、 Place the whole machine in the air with humidity ≤ 60 % , temperature ≤ 40 ℃ and ≥ 15 ℃, fine
particles ≤ 3 5ug/ m³ Ventilation and drying shall be carried out in a ventilated indoor
environment,, and the drying time should be ≥ 3 0 s. Meanwhile, direct exposure to sunlight shall
be avoided.

9、 Check whether the whole part of the fuselage is damaged, such as cracks, cracks, cracks or
protruding sound head.In case of the above situations, stop using immediately and contact the
local dealer or sales representative.

Manual Disinfection
Manual Disinfection
The operation process is as follows:
Before the machine is sterilized, it should be ensured that the cleaning work has been carried
out and the cleaning process is correctly followed. Operators should wear protective equipment such
as professional goggles and gloves, and check whether the instruments, equipment, materials and
environment used meet the requirements in advance.
1、 Use alcohol disinfectants (For example: The compound double- chain quaternary ammonium
salt with free safety content ≤ 2% is mixed with purified water or softened filtered water. The
mixed quaternary ammonium salt content is 200mg / L-1000mg / L, which conforms to the
alcohol disinfection solution of double chain quaternary ammonium salt ) for spraying and
disinfection . Install the mixed disinfectant reagent into a spray pot with spraying function, and
use the spray pot to evenly spray the disinfectant on the surface of the ultrasonic probe and
around the ultrasonic probe shell. ( Quaternary ammonium salts should not be used in
combination with soap or anionic surfactants , citrates, iodides, nitrates, potassium
permanganate, salicylates, silver salts, tartrates and alkaloids, aluminum, sodium fluorescein,
permanganate Hydrogen oxide, kaolin, water-containing lanolin, etc. ) Put the machine flat on
the workbench for static disinfection, and the static time is 2min- 5min .
2、 Take a piece of special wiping paper for ultrasonic instrument, and spray it around the central
part of the wiping paper for 3-5 times with a spray pot equipped with quaternary ammonium salt
disinfectant to ensure that the area of the wiping paper containing disinfectant is ≥ 80% . Use
wipes sprayed with disinfectant to wipe the machine, and ensure that all surfaces and gaps of
the machine body are completely wiped and disinfected.
3、 Take a new piece of special wiping paper or clean lint free soft cloth for ultrasonic instrument,
wipe the whole body and the acoustic head again, and wipe it from the acoustic head to the body
until all the disinfection reagent residues on the body are wiped clean.
4、 Check whether there are residual stains or water stains on the body and sound head. If there are
still stains or water stains, repeat steps 2 and 3 until the stains on the body and sound head are
wiped off. If not, proceed to the next step directly.
5、 Place the wiped machine in the air with humidity ≤ 60%, temperature ≤ 40 ℃ and ≥ 15 ℃, and
fine particles ≤ 35ug / m ³ The secondary ventilation and drying shall be carried out in a
ventilated indoor environment, and the drying time shall be ≥ 2min. Meanwhile,direct exposure
to sunlight shall be avoided.

13/16
Troubleshooting
Connection issues

Display the error password 1. The SN number of the probe is the wifi password, re-enter the
password, The password’s letters must be input as small letters, not
capital.
2. Try to connect to the probe with type-C cable.
The probe can not connected to Mobile 1. Please change the Wi-Fi channel with your laptop.
phone/tablet, but can work with laptop 2. Try to connect the probe to your mobile phone again.
The probe can work by Wi-Fi, but can not work with 1. The A-end and B-end of the cable can not be inserted reversely.
Type-C The A-end must be fully inserted into the probe and the B-end must
be fully inserted into the smart device.
2. Try to connect the probe with the other side of the type-C A port
interface
Probe issues
Probe can not be charged by cable 1. The A-end must be fully inserted into the probe and the B-end
must be fully inserted into the smart device.
2. Charge the probe for 1h with wireless charger.
3. If not work, pls contact support!
Can not turn on the probe 1. Charge the probe for 30 minutes firstly
2. Try to turn on the probe again
3. If not work, pls contact support!
Can not turn off the probe 1. Press and hold the probe’s power Button for 15-20 seconds.
2. Charge the probe
App issues
App can not turn on 1. Delete and re-install the App
2. Update the App
App crashes 3. Try to install the app to other mobile device
App opens but will not scan images 1. Make sure the probe is connected successfully
2. Try to press the probe power button
3. Re-install and update the App
4. Charge the probe
Black screen or screen no longer updates 1. Close the App and restart the App.
2. Unplug the probe from the mobile platform (mobile device) and
reconnect.
Imaging issues
Image degradation or occurrence of image artifacts Make sure you are using the appropriate preset and the depth is
appropriate for the anatomy being scanned.
Image quality degraded 1. Make sure you are using enough approved ultrasound gel. If
quality does not improve.
2. If not work, contact Support
Image is nor clear 1. Adjust the image parameters follow the page-24
2. Use enough ultrasound Gel.

14/16
Copyright Ownership and Usage Limitation
· All copyrights and intellectual property rights of the contents created are owned by
KONTED.