G7112 BUser
G7112 BUser
This manual covers the Agilent InfinityLab LC Series 1260 Infinity II Binary Pump
(G7112B).
1 Introduction
This chapter gives an introduction to the module and an instrument overview.
4 Optimizing Performance
This chapter gives information on how to optimize the performance of the Binary
Pump under special operational conditions.
6 Error Information
This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.
8 Maintenance
This chapter describes the maintenance of the module.
10 Identifying Cables
This chapter provides information on cables used with the Agilent InfinityLab LC
Series modules.
11 Hardware Information
This chapter provides detailed technical information about your binary pump.
12 LAN Configuration
This chapter provides information on connecting the module to the Agilent
ChemStation PC.
13 Appendix
This chapter provides addition information on safety, legal and web.
1 Introduction 10
Instrument and Operation 11
Overview of the Hydraulic Path 15
Leak and Waste Handling 20
4 Optimizing Performance 59
When to Use a Vacuum Degasser 60
When to Use the Active Seal Wash Option 61
When to Use the Low Volume Mixer 62
When to Remove Damper and Mixer 63
How to Optimize the Compressibility Compensation Setting 65
6 Error Information 73
What Are Error Messages 74
General Error Messages 75
Module Error Messages 81
13 Appendix 212
General Safety Information 213
Waste Electrical and Electronic Equipment (WEEE) Directive 219
Radio Interference 220
Sound Emission 221
Agilent Technologies on the Internet 222
Principle of Operation
The binary pump is based on a two-channel, dual-piston in-series design which
comprises all essential functions that a solvent delivery system has to fulfill.
Metering of solvent and delivery to the high-pressure side are performed by two
pump assemblies which can generate pressure up to 600 bar.
Each channel comprises a pump assembly including pump drive, pump head,
active inlet valve with replaceable cartridge, and outlet valve. The two channels
are fed into a low-volume mixing chamber which is connected via a restriction
capillary coil to a damping unit and a mixer. A pressure sensor monitors the
pump pressure. A purge valve with integrated PTFE frit is fitted to the pump
outlet for convenient priming of the pumping system.
to sampler
SSV
Purge valve
Seal wash pump to waste
Damper
Mixing chamber
Figure 2 The hydraulic path of the Binary Pump with damper and mixer
Damper and mixer can be bypassed for lowest delay volume of the binary pump.
This configuration is recommended for low flow rate applications with steep
gradients.
Figure 3 on page 13 illustrates the flow path in low delay volume mode. For
instructions on how to change between the two configurations, see “Convert the
Binary Pump to Low Delay Volume Mode” on page 64.
Bypassing the mixer while the damper remains in line is not a supported
NOTE configuration and may lead to undesired behavior of the binary pump.
to sampler
SSV
Purge valve
Seal wash pump to waste
Damper
Mixer
Mixing chamber
Figure 3 The hydraulic path of the Binary Pump with bypassed damper and mixer
Product Description
The Agilent 1260 Infinity II Binary Pump is suited for LC applications using
conventional or superficially porous particles columns, as well as STM column
technology. Gradient formation is based on a high-pressure mixing principle.
Standard or low delay volumes can be easily configured by the user. The Agilent
1260 Infinity II Binary Pump offers reproducible gradients and high-performance,
providing high-throughput and fast separations. The pump is suitable for routine
applications, with UV or MS detection, where high speed and resolution with
uncompromised data quality are required.
Status indicator
Purge valve
Degasser
Pump head A
Pump head B
Leak drain
Power switch
Features
• Configurable delay volume - down to 120 µL together with a flow range up to
5 mL/min provides universal applicability.
• Change from standard to low delay volume configuration is enabled.
• High gradient performance - even at low % B and narrow-bore flow rates.
• Integrated 2-channel-degasser.
• Fast and precise gradients using LC/MS, as well as UV-only systems.
• Fully exploits the speed and separation potential of Poroshell.
The solvent from the bottle in the solvent cabinet enters the pump through an
active inlet valve. Each side of the binary pump comprises two substantially
identical pump units. Both pump units comprise a ball-screw drive and a pump
head with two sapphire pistons for reciprocating movement.
Seal
Outlet valve
Piston
To mixing chamber
of 1 µL/min to 5 mL/min. The inlet of the first pumping unit is connected to the
active inlet valve which is processor-controlled opened or closed allowing solvent
to be drawn into the first pump unit.
The outlet of the first pump chamber is connected by a 500 µL absorber capillary
to the second pump chamber. The outlets of the second chambers of both pump
channels joined via a small mixing chamber. A coiled restriction capillary
connects the mixing chamber via a pressure pulse damper, a mixer and a
pressure sensor to the purge valve assembly. The outlet of the purge valve
assembly is then connected to the attached chromatographic system.
When turned on, the pump runs through an initialization procedure to determine
the upper dead center of the first piston of both pump channels. The first piston
moves slowly upwards to the mechanical stop of the pump head and from there
it moves back a predetermined path length. The controller stores this piston
position in memory. After this initialization the pump starts operation with the set
parameters for the two pump channels.
The active inlet valve is opened and the down moving piston draws solvent into
the first pump head. At the same time the second piston is moving upwards
delivering into the system. After a controller defined stroke length (depending on
the flow rate) the drive motors are stopped and the active inlet valve is closed.
The motor direction is reversed and moves the first piston up until it reaches the
stored upper limit and at the same time moving the second piston downwards.
Then the sequence starts again moving the pistons up and down between the
two limits. During the delivery stroke of the first piston the solvent in the pump
head is pressed through the outlet valve into the second pumping unit. The
second piston draws in half of the volume displaced by the first piston and the
remaining half volume is directly delivered into the system. During the drawing
stroke of the first piston, the second piston delivers the drawn volume into the
system.
For pump specifications, see “Performance Specifications” on page 30.
The Agilent InfinityLab LC Series has been designed for safe leak and waste
handling. It is important that all security concepts are understood and
instructions are carefully followed.
The solvent cabinet is designed to store a maximum volume of 8 L solvent. The
maximum volume for an individual bottle stored in the solvent cabinet should not
exceed 2 L. For details, see the usage guideline for the Agilent Infinity II Solvent
Cabinets (a printed copy of the guideline has been shipped with the solvent
cabinet, electronic copies are available on the Internet).
All leak plane outlets are situated in a consistent position so that all Infinity and
Infinity II modules can be stacked on top of each other. Waste tubes are guided
through a channel on the right hand side of the instrument, keeping the front
access clear from tubes.
The leak plane provides leak management by catching all internal liquid leaks,
guiding them to the leak sensor for leak detection, and passing them on to the
next module below, if the leak sensor fails. The leak sensor in the leak plane
stops the running system as soon as the leak detection level is reached.
Solvent and condensate is guided through the waste channel into the waste
container:
• from the detector's flow cell outlet
• from the Multisampler needle wash port
• from the Sample Cooler or Sample Thermostat (condensate)
• from the pump's Seal Wash Sensor (if applicable)
• from the pump's Purge Valve or Multipurpose Valve
The waste tube connected to the leak plane outlet on each of the bottom
instruments guides the solvent to a suitable waste container.
Leak Sensor
Solvent incompatibility
C AU T I O N
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor.
The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible
with DMF.
Do not use DMF as mobile phase.
Check the leak sensor regularly for corrosion.
Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use
your own waste solution, make sure that the tubes don't immerse in the liquid.
Site Requirements 26
Physical Specifications 29
Performance Specifications 30
This chapter provides information about site requirements and specifications for
the binary pump.
Site Requirements
Power Considerations
The module power supply has wide ranging capability. It accepts any line voltage
in the range described in Table 1 on page 29. Consequently there is no voltage
selector in the rear of the module. There are also no externally accessible fuses,
because automatic electronic fuses are implemented in the power supply.
Module is partially energized when switched off, as long as the power cord
WAR N IN G
is plugged in.
Repair work at the module can lead to personal injuries, e.g. shock hazard,
when the cover is opened and the module is connected to power.
Make sure that it is always possible to access the power plug.
Remove the power cable from the instrument before opening the cover.
Do not connect the power cable to the Instrument while the covers are
removed.
Power Cords
Country-specific power cords are available for the module. The female end of all
power cords is identical. It plugs into the power-input socket at the rear. The male
end of each power cord is different and designed to match the wall socket of a
particular country or region.
Agilent makes sure that your instrument is shipped with the power cord that is
suitable for your particular country or region.
Bench Space
The module dimensions and weight (see Table 1 on page 29) allow you to place
the module on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches)
in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is
designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex
NOTE Bench rack. This option helps to save bench space as all modules can be placed
into one single stack. It also allows to easily relocate the instrument to another
lab.
Condensation
Condensation within the module
C AU T I O N
Condensation can damage the system electronics.
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to
warm slowly to room temperature to avoid condensation.
Physical Specifications
Safety standards: Overvoltage category II, Pollution degree 2 For indoor use only
IEC, EN, CSA, UL
Performance Specifications
Flow precision ≤0.07 % RSD or < 0.02 min SD, whichever is Based on retention time at
greater constant temperature
Composition precision < 0.15 % RSD or < 0.04 min SD, whichever At 0.2 and 1 mL/min; based on
is greater retention time at constant
temperature
Instrument control LC & CE Drivers A.02.14 or above For details about supported
Instrument Control Framework (ICF) software versions refer to the
A.02.04 or above compatibility matrix of your
InfinityLab LC Companion (G7108AA) with version of the LC and CE
firmware D.07.25 or above Drivers
Instant Pilot (G4208A) with firmware
B.02.20 or above
Lab Advisor software B.02.08 or above
Magnets 33
Turn on/off 34
Status Indicators 36
Best Practices 37
Daily / Weekly tasks 37
Power up / Shut-down the pump 37
Prepare the pump 38
How to deal with solvents 38
Leak and Waste Handling 39
Hints for Successful Use of the Binary Pump 40
Setting up the Pump with the G4208A Instant Pilot 41
Setting up the Pump with the Instrument Control Interface 42
Overview 42
Setup of Basic Pump Parameters 43
Pump Control 44
Auxiliary Pump Parameters 44
Data Curves 45
Bottle Filling 46
Solvent Information 47
Algae Growth in HPLC Systems 54
How to Prevent and/or Reduce the Algae Problem 54
Prevent Blocking of Solvent Filters 55
Checking the Solvent Filters 56
Cleaning the Solvent Filters 56
Normal Phase Applications 57
Magnets
Turn on/off
1 2
Power switch: On
Status Indicators
Status indicators
1. Idle
2. Run mode
3. Not-ready. Waiting for a specific pre-run condition to be reached or
completed.
4. Error mode - interrupts the analysis and requires attention (for example, a
leak or defective internal components).
5. Resident mode (blinking) - for example, during update of main firmware.
6. Bootloader mode (fast blinking). Try to re-boot the module or try a
cold-start. Then try a firmware update.
Best Practices
Weekly Tasks
• Change seal wash solvent (10 % isopropanol in water) and bottle.
• If applications with salts were used, flush all channels with water and remove
possible salt deposits manually.
• Inspect solvent filters for dirt or blockages. Exchange if no flow is coming out
of the solvent line when removed from the degasser inlet.
Seal wash
Seal Wash guarantees a maximum seal life time. Use Seal Wash:
• When using buffers with elevated salt concentrations
• When using volatile solvents with non-volatile additives
For details, see the usage guideline for the solvent cabinet. A printed copy of the
NOTE guideline has been shipped with the solvent cabinet, electronic copies are
available in the Agilent Information Center or via the Internet.
• Place solvent cabinet with the solvent bottles always on top (or at a higher
level) of the pump.
• Flush the degasser with at least 5 mL per channel before operating the pump,
especially when the pumping system had been turned off for a certain length
of time (for example, overnight) and volatile solvent mixtures are used in the
channels.
• Prevent blocking of solvent inlet filters (never use the pump without solvent
inlet filters). Growth of algae should be avoided (see “Prevent Blocking of
Solvent Filters” on page 55).
• Check purge valve frit and column frit in regular time intervals. A blocked
purge valve frit can be identified by black, yellow or greenish layers on its
surface or by a pressure greater than 10 bar in low delay volume configuration
and 20 bar in standard configuration when pumping distilled water at a rate of
5 mL/min with an open purge valve.
• Whenever possible use a minimum flow rate of 5 µL/min per solvent channel
to avoid crossflow of solvent into the unused pump channel.
• Whenever exchanging the pump seals, the purge valve frit should be
exchanged, too.
• When using buffer solutions, flush the system with water before switching it
off. The seal wash option should be used when installed, especially when
buffer solutions with concentrations of 0.1 M or higher are being pumped for
long periods of time.
• Check the pump pistons for scratches, grooves and dents when changing the
piston seals. Damaged pistons cause micro leaks and will decrease the
lifetime of the seals.
• After changing the piston seals, apply the seal wear-in procedure (see “Seal
Wear-in Procedure” on page 133).
• Place the aqueous solvent on channel A and the organic solvent on channel B.
The default compressibility settings are set accordingly.
Generic operation of the G4208A Instant Pilot is covered in the Agilent Instant
Pilot G4208A User's Guide (G4208-90006). Details about setting up module
specific parameters can be found in the Instant Pilot online help.
The pump parameters are described in detail in “Overview” on page 42.
Overview
Parameters described in following sections are offered by the instrument control
interface and can usually be accessed through Agilent instrument control
software. For details, please refer to manuals and online help of respective user
interfaces.
• Flow 0.001 – 5 mL/min Total flow rate of the pump. See “When to Remove Damper and Mixer” on page 63
for pump hardware modifications to achieve lowest delay volume.
• Stop Time 0.01 min - no limit The stop time of the pump usually controls the run time of the whole LC system.
Use no limit to stop the run manually (useful for method development).
• Post Time off - 99999 min Time between the end of a run and the start of the next. Used for column
equilibration after a gradient.
• Pressure Max: 0 – 600 bar Max must be bigger than Min! Set max pressure to the maximum operating
Limits Min: 0 – 600 bar pressure of your column. A min pressure setting of e.g. 10 bar will turn off your
pump automatically when running out of solvent. A smarter way, however, is to use
the bottle fillings function (see “Bottle Filling” on page 46).
• Solvent A 0 – 100 % Although channel A can be set to 0 %, it cannot be turned off. This channel should
be used for the aqueous phase (water).
• Solvent B off - 100 % The percentage of channel B is automatically complemented by channel A to give
100 %.
• Solvent H2O, ACN, MeOH, IPA Select the solvent you are using in the respective solvent channel from the
type drop-down list. In case your solvent is not listed, perform a solvent compressibility
calibration (see “Running the Solvent Compressibility Calibration” on page 101).
For details on solvent compressibility see “Binary Pump Solvent Compressibility
Calibration” on page 100.
• Solvent Free text field for a description of the solvent. This description will show up in
Comment method printouts, etc.
• Timetable max. number of lines Use the timetable to build solvent gradients, flow gradients, or combinations of
depends on free space both. Gradients are always linear. Use multiple timetable entries to mimic
in pump memory exponential or parabolic gradients.
Pump Control
The pump can be switched between following states: On, Off or to Standby. In
Standby, the pump motor is still controlled. When the pump is switched on from
standby, it does not re-initialize.
Upon initialization, the pump ignores the Maximum Flow Gradient value.
C AU T I O N
This can result in a rapid and uncontrolled pressure increase.
To prevent harm to the column, open the purge valve until the initialization
is finished.
The optional seal wash pump can be controlled by either switching it off, using it
for a single time or specifying frequency and duration of periodic wash intervals.
Upon initialization, the pump ignores the Maximum Flow Gradient value.
C AU T I O N
This can result in a rapid and uncontrolled pressure increase.
To prevent harm to the column, open the purge valve until the initialization
is finished.
• Maximum Flow 0.1 – 100 mL/min² With this parameter flow rate changes can be ramped up and down slowly
Gradient default: 100 mL/min² to avoid pressure shocks to the column. The default value is 100 mL/min²
which in fact turns the function off.
• Minimum Stroke 20 – 100 µL The volume one pump piston delivers per stroke. In general, a smaller stroke
default: Auto volume results in lower pump ripple. The Auto setting adjusts the strokes
dynamically to the lowest possible value.
The strokes can be set individually for pump heads A and B.
• Compressibility 0 - 150·10-6 /bar or For best performance, check option Use enhanced compressibility
enhanced calibration. With this option, the pump will use solvent data libraries provided
compressibility by Agilent or data generated by using solvent compressibility calibrations
calibration (see “Running the Solvent Compressibility Calibration” on page 101).
default: use enhanced For details on solvent compressibility see “Binary Pump Solvent
comp. calibration Compressibility Calibration” on page 100.
For backward compatibility to 400 bar pumps, the solvent compressibility
can still be set manually for each channel when the box is unticked.
Data Curves
The binary pump provides the possibility to store the following operational data in
the data file of the Agilent data system:
• Solvent percentage for each channel,
• pump flow,
• pressure
The pressure data curve is generated from the pressure sensor readings, while
NOTE %A, %B and flow are calculated from the method settings of the pump.
For details, please refer to the online help or manual of your instrument control
software.
Bottle Filling
The pump offers a powerful feature to monitor the liquid level in the solvent
bottles. With total bottle volume and initial filling volume set correctly, the pump
subtracts the consumed volume continuously from the initial value and stops the
pump and method/sequence execution before the system runs dry or an analysis
is corrupted.
The bottle filling feature fails if multiple solvent inlets are put into one solvent
C AU T I O N bottle!
In that case implement a minimum pressure limit to avoid that the pump
runs dry when solvents are empty.
• Total Volume 0 – 1000 L This is the capacity (maximum possible volume) in liter of the solvent bottle. In
default: 0 L combination with the actual volume, this parameter is used for calculating and
displaying the relative liquid level.
• Actual 0 – 1000 L After filling the solvent bottles, enter the actual volumes into these boxes. The
Volume default: 0 L Actual Volume must not be larger than the Total Volume of the bottle.
• Prevent default: unchecked If this option is checked, the pump won’t start a new run if the solvent level in
analysis....... one or more bottles is below the minimum volume. Enter a minimum volume in
liter, which considers the position of the solvent inlet and size/shape of the
solvent bottle such that no air is drawn if the actual volume gets close to this
limit.
• Turn pump default: unchecked If this option is checked, the pump will turn off before air is aspirated. However,
off... the residual solvent volume has been calculated for 1 L solvent bottles and may
be too small for large bottles or other vessels.
Solvent Information
Part Materials
Degasser chamber TFE/PDD Copolymer, PFA (internal tubings), PEEK (inlets), FEP
(tubings), ETFE (fittings)
Pistons Sapphire
Piston seals/wash seals PTFE, SST (reversed phase) or UHMW-PE, SST (normal phase)
Capillaries/fittings SST
Tubings PTFE
Disclaimer
Subsequent data was collected from external resources and is meant as a
reference. Agilent cannot guarantee the correctness and completeness of such
information. Data is based on compatibility libraries, which are not specific for
estimating the long-term life time under specific but highly variable conditions of
UHPLC systems, solvents, solvent mixtures and samples. Information can also
not be generalized due to catalytic effects of impurities like metal ions,
complexing agents, oxygen etc. Apart from pure chemical corrosion, other
effects like electro corrosion, electrostatic charging (especially for
non-conductive organic solvents), swelling of polymer parts etc. need to be
considered. Most data available refers to room temperature (typically 20 – 25 °C,
68 – 77 °F). If corrosion is possible, it usually accelerates at higher temperatures.
If in doubt, please consult technical literature on chemical compatibility of
materials.
MP35N
MP35N is a nonmagnetic, nickel-cobalt-chromium-molybdenum alloy
demonstrating excellent corrosion resistance (for example, against nitric and
sulfuric acids, sodium hydroxide, and seawater) over a wide range of
concentrations and temperatures. In addition, this alloy shows exceptional
resistance to high-temperature oxidation. Due to excellent chemical resistance
and toughness, the alloy is used in diverse applications: dental products, medical
devices, nonmagnetic electrical components, chemical and food processing
equipment, marine equipment. Treatment of MP35N alloy samples with 10 %
NaCl in HCl (pH 2.0) does not reveal any detectable corrosion. MP35N also
demonstrates excellent corrosion resistance in a humid environment. Although
the influence of a broad variety of solvents and conditions has been tested, users
should keep in mind that multiple factors can affect corrosion rates, such as
temperature, concentration, pH, impurities, stress, surface finish, and dissimilar
metal contacts.
PEEK
PEEK (Polyether-Ether Ketones) combines excellent properties regarding
biocompatibility, chemical resistance, mechanical and thermal stability. PEEK is
therefore the material of choice for UHPLC and biochemical instrumentation.
It is stable in the specified pH range (for the Bio-Inert LC system: pH 1 – 13, see
bio-inert module manuals for details), and inert to many common solvents.
There is still a number of known incompatibilities with chemicals such as
chloroform, methylene chloride, THF, DMSO, strong acids (nitric acid > 10 %,
sulfuric acid > 10 %, sulfonic acids, trichloroacetic acid), halogens or aqueous
halogen solutions, phenol and derivatives (cresols, salicylic acid, and so on).
When used above room temperature, PEEK is sensitive to bases and various
organic solvents, which can cause it to swell. Under such conditions, normal
PEEK capillaries are sensitive to high pressure. Therefore, Agilent uses stainless
steel cladded PEEK capillaries in bio-inert systems. The use of stainless steel
cladded PEEK capillaries keeps the flow path free of steel and ensures pressure
stability up to 600 bar. If in doubt, consult the available literature about the
chemical compatibility of PEEK.
Polyimide
Agilent uses semi-crystalline polyimide for rotor seals in valves and needle seats
in autosamplers. One supplier of polyimide is DuPont, which brands polyimide as
Vespel, which is also used by Agilent.
Polyimide is stable in a pH range between 1 and 10 and in most organic solvents.
It is incompatible with concentrated mineral acids (e.g. sulphuric acid), glacial
acetic acid, DMSO and THF. It is also degraded by nucleophilic substances like
ammonia (e.g. ammonium salts in basic conditions) or acetates.
Polyethylene (PE)
Agilent uses UHMW (ultra-high molecular weight)-PE/PTFE blends for yellow
piston and wash seals, which are used in 1290 Infinity pumps, 1290 Infinity II
pumps, the G7104C and for normal phase applications in 1260 Infinity pumps.
Polyethylene has a good stability for most common inorganic solvents including
acids and bases in a pH range of 1 to 12.5. It is compatible with many organic
solvents used in chromatographic systems like methanol, acetonitrile and
isopropanol. It has limited stability with aliphatic, aromatic and halogenated
hydrocarbons, THF, phenol and derivatives, concentrated acids and bases. For
normal phase applications, the maximum pressure should be limited to 200 bar.
Tantalum (Ta)
Tantalum is inert to most common HPLC solvents and almost all acids except
fluoric acid and acids with free sulfur trioxide. It can be corroded by strong bases
(e.g. hydroxide solutions > 10 %, diethylamine). It is not recommended for the use
with fluoric acid and fluorides.
Titanium (Ti)
Titanium is highly resistant to oxidizing acids (for example, nitric, perchloric and
hypochlorous acid) over a wide range of concentrations and temperatures. This
is due to a thin oxide layer on the surface, which is stabilized by oxidizing
compounds. Non-oxidizing acids (for example, hydrochloric, sulfuric and
phosphoric acid) can cause slight corrosion, which increases with acid
concentration and temperature. For example, the corrosion rate with 3 % HCl
(about pH 0.1) at room temperature is about 13 μm/year. At room temperature,
titanium is resistant to concentrations of about 5 % sulfuric acid (about pH 0.3).
Addition of nitric acid to hydrochloric or sulfuric acids significantly reduces
corrosion rates. Titanium is sensitive to acidic metal chlorides like FeCl3 or CuCl2.
Titanium is subject to corrosion in anhydrous methanol, which can be avoided by
adding a small amount of water (about 3 %). Slight corrosion is possible with
ammonia > 10 %.
Gold
Gold is inert to all common HPLC solvents, acids and bases within the specified
pH range. It can be corroded by complexing cyanides and concentrated acids like
aqua regia.
Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases and solvents. There
are no documented incompatibilities for HPLC applications.
The presence of algae in HPLC systems can cause a variety of problems that
may be incorrectly diagnosed as instrument or application problems. Algae grow
in aqueous media, preferably in a pH range of 4-8. Their growth is accelerated by
buffers, for example phosphate or acetate. Since algae grow through
photosynthesis, light will also stimulate their growth. Even in distilled water
small-sized algae grow after some time.
Contaminated solvents or algae growth in the solvent bottle will reduce the
lifetime of the solvent filter and will influence the performance of the module.
This is especially true for aqueous solvents or phosphate buffers (pH 4 to 7). The
following suggestions will prolong lifetime of the solvent filter and will maintain
the performance of the module.
• Use a sterile, if possible amber, solvent bottle to slow down algae growth.
• Filter solvents through filters or membranes that remove algae.
• Exchange solvents every two days or refilter.
• If the application permits add 0.0001 – 0.001 M sodium azide to the solvent.
• Place a layer of argon on top of your solvent.
• Avoid exposure of the solvent bottle to direct sunlight.
Current passive inlet valves and outlet ball valves used with 1260 and 1290
Infinity pumps do not work well with applications using non-polar solvents as for
normal phase applications (e.g. hexane, heptane and CO2). With such
applications, pressure drops could be observed. They are a result of particles
electrostatically charging up in insulating solvents and sticking to the balls inside
the valves, such that the valves do not close properly any more after some time
of use (can be hours).
For normal phase applications, a second type of valves is available, which has a
design based on the existing one for 1260 and 1290 Infinity valves. These valves
use a new material for valve balls, which is a conductive ceramic and replaces
non-conductive ruby balls. The balls do not charge up electrostatically and show
good performance in normal phase.
The valves are marked with N for non-polar or normal phase.
Agilent recommends using these valves for (and only for) normal phase
applications.
No design change has been done for active inlet valves, which have already been
used successfully in the past for normal phase applications in 1200 Series and
1260 Infinity binary pumps.
Corrosion of valves
C AU T I O N
Normal phase balls/valves corrode quickly in aqueous solutions and acids (at
or below pH 7).
Do not use normal phase valves in applications running with aqueous
solutions.
The N-Valves have been tested successfully in using hexane at pressures below
100 bar; heptane can be used as a substitute for neurotoxic hexane.
Inlet valves 1260 Infinity Inlet Valve Type N 1290 Infinity Inlet Valve Type N (G4220-60122)
(G1312-60166) 1290 Infinity Quat Inlet Valve Type N
(G4204-60122)
Outlet valves Outlet Valve Type N/SFC 1290 Infinity Outlet Valve Type N (G4220-60128)
(G1312-60167)
This chapter gives information on how to optimize the performance of the Binary
Pump under special operational conditions.
A degasser removes air, which is dissolved in any solvent. When solvents are
heated or mixed with other solvents, air can leave the solvent and form small
bubbles. Over time, these bubbles accumulate and can cause pressure
fluctuations which may finally result in retention time shifts.
All Agilent 1200 Infinity II Series Pumps have a built-in degasser. While a
degasser is needed for low pressure mixing pumps like Agilent quaternary
pumps, high pressure mixing pumps like Agilent binary pumps are more robust
with respect to bubble formation. However, a degasser is recommended for best
performance.
Additionally, a degasser is highly recommended for the following applications:
• Your detector is used with maximum sensitivity in the low UV wavelength
range.
• Your application requires highest injection precision.
• Your application requires highest retention-time reproducibility (flow rates
below 0.5 mL/min).
• The binary pump is used with bypassed damper and mixer.
The external 1260 Infinity II Degasser G7122A is recommended for use with
applications using highly volatile solvents like Hexane or DCM, solvents with
special characteristics like THF, or applications using refractive index detection.
Concentrated buffer solutions will reduce the lifetime of the seals and pistons in
your binary pump. The active seal wash option allows to maintain the seal
lifetime by flushing the low pressure side of the seals with a wash solvent.
The seal wash option is strongly recommended if buffer concentrations of 0.1 M
or higher are used regularly with the pump.
The active seal wash option kit can be ordered by quoting Active Seal Wash
Upgrade Product including Service (G1399A).
The seal wash option comprises a peristaltic pump, secondary seals, gaskets,
seal holders and tubing for both pump heads. A bottle of premixed
water/isopropanol (90/10 vol%) is placed in the solvent cabinet and connected to
the peristaltic pump.
Always use a mixture of HPLC-grade water (90 %) and isopropanol (10 %) as
wash solvent. This mixture prevents bacteria growth in the wash bottle and
reduces the surface tension of the water.
The operation of the peristaltic pump can be controlled from the data system or
the Instant Pilot.
For adding a seal-wash option, please contact your local Agilent Technologies
service representative.
The Low volume mixer ( 200 µL) (5067-1565) is designed for use with the Agilent
InfinityLab LC Series 1260 Infinity II Binary LC System in low delay volume mode.
This configuration is typically used for 2.1 mm i.d., 1.8 µm particle size columns,
where emphasis is put on S/N ratio. The low volume mixer helps mixing
gradients starting with a low concentration of organic solvents, which can cause
noise on the baseline.
The binary pump is equipped with a pressure pulsation damper and a static
mixer. The total delay volume of the pump is 600 – 800 µL (depending on system
pressure). The mixer has a volume of 400 µL.
For applications that require lowest delay volume (for example, fast gradient
methods or gradient applications with low flow rates), damper and mixer can be
bypassed.
Wrench, 14 mm
Hex driver open, 1/4 inch
Preparations Flush the system (water if buffers were used, otherwise isopropanol).
Turn the flow off.
1 Open the doors. 2 Use the 1/4 inch hex driver to remove fitting B from port
2 of the pressure sensor. Fold capillary end B away. It
remains unconnected.
2
B
3 Disconnect fitting A from outlet 1 of the mixer. 4 Connect fitting A to port 2 of the pressure sensor. Seal
port 1 of the mixer with a plastic blank nut.
A
1 A 1
2
Acetone 126
Acetonitrile 115
Benzene 95
Chloroform 100
Cyclohexane 118
Ethanol 114
Heptane 120
Hexane 150
Isobutanol 100
Isopropanol 100
Methanol 120
1-Propanol 100
Toluene 87
Water 46
Status Indicators
The module is provided with two status indicators which indicate the operational
state of the module. The status indicators provide a quick visual check of the
operation of the module.
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the module
generates an error message in the user interface. For each message, a short
description of the failure, a list of probable causes of the problem, and a list of
suggested actions to fix the problem are provided (see chapter Error
Information).
Test Functions
A series of test functions are available for troubleshooting and operational
verification after exchanging internal components (see Tests and Calibrations).
Pressure Test
The Pressure Test is a quick test designed to determine the pressure tightness of
the system (i.e. the high pressure flow path between pump and column). After
exchanging flow path components (e.g. pump seals or injection seal), use this
test to verify the system is pressure tight, see “(System) Pressure Test” on
page 93.
User Interfaces
Depending on the user interface, the available tests vary. Some descriptions are
only available in the Service Manual.
The Agilent Lab Advisor Software (basic license, shipped with an Agilent LC
pump) is a standalone product that can be used with or without a
chromatographic data system. Agilent Lab Advisor helps to manage the lab for
high-quality chromatographic results by providing a detailed system overview of
all connected analytical instruments with instrument status, Early Maintenance
Feedback counters (EMF), instrument configuration information, and diagnostic
tests. With the push of a button, a detailed diagnostic report can be generated.
Upon request, the user can send this report to Agilent for a significantly improved
troubleshooting and repair process.
The Agilent Lab Advisor software is available in two versions:
• Lab Advisor Basic
• Lab Advisor Advanced
Lab Advisor Basic is included with every Agilent 1200 Infinity Series and Agilent
InfinityLab LC Series instrument.
The Lab Advisor Advanced features can be unlocked by purchasing a license key,
and include real-time monitoring of instrument actuals, all various instrument
signals, and state machines. In addition, all diagnostic test results, calibration
results, and acquired signal data can be uploaded to a shared network folder. The
Review Client included in Lab Advisor Advanced allows to load and examine the
uploaded data no matter on which instrument it was generated. This makes Data
Sharing an ideal tool for internal support groups and users who want to track the
instrument history of their analytical systems.
The optional Agilent Maintenance Wizard Add-on provides an easy-to-use,
step-by-step multimedia guide for performing preventive maintenance on Agilent
1200 Infinity LC Series instrument.
The tests and diagnostic features that are provided by the Agilent Lab Advisor
software may differ from the descriptions in this manual. For details, refer to the
Agilent Lab Advisor software help files.
This chapter describes the meaning of error messages, and provides information
on probable causes and suggested actions how to recover from error conditions.
Timeout
Error ID: 0062
The timeout threshold was exceeded.
1 The analysis was completed successfully, Check the logbook for the occurrence and source
and the timeout function switched off the of a not-ready condition. Restart the analysis
module as requested. where required.
2 A not-ready condition was present during a Check the logbook for the occurrence and source
sequence or multiple-injection run for a of a not-ready condition. Restart the analysis
period longer than the timeout threshold. where required.
Shutdown
Error ID: 0063
An external instrument has generated a shutdown signal on the remote line.
The module continually monitors the remote input connectors for status signals.
A LOW signal input on pin 6 of the Enhanced Remote Interface (ERI) connector
generates the error message.
1 Leak detected in another module with a CAN Fix the leak in the external instrument before
connection to the system. restarting the module.
2 Leak detected in an external instrument with Fix the leak in the external instrument before
a remote connection to the system. restarting the module.
4 The degasser failed to generate sufficient Check the vacuum degasser for an error
vacuum for solvent degassing. condition. Refer to the Service Manual for the
degasser or the pump that has the degasser
built-in.
Remote Timeout
Error ID: 0070
A not-ready condition is still present on the remote input. When an analysis is
started, the system expects all not-ready conditions (for example, a not-ready
condition during detector balance) to switch to run conditions within one minute
of starting the analysis. If a not-ready condition is still present on the remote line
after one minute the error message is generated.
1 Not-ready condition in one of the Ensure the instrument showing the not-ready
instruments connected to the remote line. condition is installed correctly, and is set up
correctly for analysis.
3 Defective components in the instrument Check the instrument for defects (refer to the
showing the not-ready condition. instrument’s documentation).
1 CAN cable disconnected. • Ensure all the CAN cables are connected
correctly.
• Ensure all CAN cables are installed correctly.
3 Defective mainboard in another module. Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.
Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak sensor circuit on the main board.
3 Loose or leaking purge valve, inlet valve, or Ensure pump components are seated correctly.
outlet valve. If there are still signs of a leak, exchange the
appropriate seal (purge valve, inlet valve, outlet
valve).
4 Loose or leaking purge valve, active inlet Ensure pump components are seated correctly.
valve, or outlet valve. If there are still signs of a leak, exchange the
appropriate seal (purge valve, active inlet valve,
outlet valve).
1 Leak sensor not connected to the power Please contact your Agilent service
switch board. representative.
3 Leak sensor incorrectly routed, being Please contact your Agilent service
pinched by a metal component. representative.
2 Leak sensor incorrectly routed, being Please contact your Agilent service
pinched by a metal component. representative.
1 Loose connection between the power switch Please contact your Agilent service
board and the mainboard representative.
2 Loose connection between the power switch Please contact your Agilent service
board and the mainboard representative.
Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the mainboard to monitor the fan
speed. If the fan speed falls below a certain limit for a certain length of time, the
error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are turned
off to assure that the module does not overheat inside.
1 Volume in bottle below specified volume. Refill bottles and reset solvent counters.
2 Incorrect setting. Make sure the set solvent volume matches the
actual bottle filling and set the shutoff limit to a
reasonable value (e.g. 100 mL for 1 L bottles)
1 Upper pressure limit set too low. Ensure the upper pressure limit is set to a value
suitable for the analysis.
2 Blockage in the flowpath (after the pressure Check for blockage in the flow path. The
sensor). following components are particularly subject to
blockage: inline filter frit, needle (autosampler),
seat capillary (autosampler), sample loop
(autosampler), column frits and capillaries with
small internal diameters (e.g. 50 µm ID).
1 Lower pressure limit set too high. Ensure the lower pressure limit is set to a value
suitable for the analysis.
2 Air bubbles in the mobile phase. • Make sure that the degasser is in flow path
and works correctly. Purge the module.
• Ensure solvent inlet filters are not blocked.
Valve Failed
Error ID: 2040
Valve 0 Failed: valve A1
Valve 1 Failed: valve A2
Valve 2 Failed: valve B2
Valve 3 Failed: valve B1
One of the solvent selection valves in the module failed to switch correctly.
The processor monitors the valve voltage before and after each switching cycle.
If the voltages are outside expected limits, the error message is generated.
2 Connection cable (inside instrument) not Please contact your Agilent service
connected. representative.
1 Defective solvent selection valve. Restart the pump. If the error message appears
again, exchange the solvent selection valve.
2 Defective connection cable (front panel to Please contact your Agilent service
main board). representative.
AIV Fuse
Error ID: 2044
Inlet-Valve Fuse 0: Pump channel A
Inlet-Valve Fuse 1: Pump channel B
One of the active-inlet valves in the module has drawn excessive current causing
the inlet-valve electronic fuse to open.
1 Defective active inlet valve. Restart the module. If the error message appears
again, exchange the active inlet valve.
2 Defective connection cable (front panel to Please contact your Agilent service
main board). representative.
Motor-Drive Power
Error ID: 2041, 2042
Motor-Drive Power: Pump channel A
B: Motor-Drive Power: Pump channel B
The current drawn by the pump motor exceeded the maximum limit.
Blockages in the flow path are usually detected by the pressure sensor, which
result in the pump switching off when the upper pressure limit is exceeded. If a
blockage occurs before the pressure sensor, the pressure increase cannot be
detected and the module will continue to pump. As pressure increases, the pump
drive draws more current. When the current reaches the maximum limit, the
module is switched off, and the error message is generated.
1 Flow path blockage in front of the pressure Ensure the capillaries and frits between the
sensor. pump head and pressure sensor inlet are free
from blockage.
2 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.
4 High friction (partial mechanical blockage) in Remove the pump-head assembly. Ensure there
the pump drive assembly. is no mechanical blockage of the pump-head
assembly or pump drive assembly.
Encoder Missing
Error ID: 2046, 2050, 2510
Encoder Missing: Pump channel A
B: Encoder Missing: Pump channel B
The optical encoder on the pump motor in the module is missing or defective.
The processor checks the presence of the pump encoder connector every 2 s. If
the connector is not detected by the processor, the error message is generated.
1 Mechanical blockage of the module. Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
3 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.
1 Pump head not installed correctly (screws Install the pump head correctly. Ensure nothing
not secured, or pump head not seated (e.g. capillary) is trapped between the pump head
correctly). and body.
Index Limit
Error ID: 2203, 2213
Index Limit: Pump channel A
B: Index Limit: Pump channel B
The time required by the piston to reach the encoder index position was too short
(pump).
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is reached too fast, the error
message is generated.
1 Irregular or sticking drive movement. Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
Index Adjustment
Error ID: 2204, 2214
Index Adjustment: Pump channel A
B: Index Adjustment: Pump channel B
The encoder index position in the module is out of adjustment.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the time to reach the index position is too long, the
error message is generated.
1 Irregular or sticking drive movement. Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
Index Missing
Error ID: 2205, 2215, 2505
Index Missing: Pump channel A
B: Index Missing: Pump channel B
The encoder index position in the module was not found during initialization.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is not recognized within a defined
time, the error message is generated.
Initialization Failed
Error ID: 2207, 2217
Initialization Failed: Pump channel A
B: Initialization Failed: Pump channel B
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
1 Blocked (passive or active) inlet valve. Exchange the (passive or active) inlet valve.
This chapter explains all test functions that are available for the binary pump.
This Lab Advisor test uses different names depending on the firmware revision
NOTE used: FW revision > A.06.50: System Pressure Test
Description
The system pressure test is a quick built-in test designed to demonstrate the leak
tightness of the system. The test involves monitoring the flow profile while the
pump delivers against a blank nut. The result is presented as the leak rate of the
module and provides information about the leak tightness of the system between
the outlet valves of the pump and the blank nut.
The blank nut can be positioned anywhere between the purge valve of the pump
NOTE and the detector inlet to pressure test the desired part of the system.
Step 1
The test begins with the initialization of both pump heads. After initialization, the
pump is starting the compression phase and the required flow rate is constantly
monitored and adjusted. The pump continues to pump until a system pressure of
around 600 bar is reached.
Step 2
When the system pressure reaches 600 bar, the pump continues to pump at a
flow rate that keeps the pressure constant. The flow that is needed to keep the
pressure constant is directly translated into a leak rate.
When The test should be used when problems with small leaks are suspected, or after maintenance of flow
path components (e.g., pump seals, injection seal) to prove pressure tightness up to 600 bar
Preparations Place two bottles of HPLC-grade water in channels A and B (A1 and B1 if the pump is equipped with
a solvent selection valve)
Make absolutely sure that all parts of the flow path that are part of the test are
NOTE very thoroughly flushed with water before starting to pressurize the system! Any
trace of other solvents or the smallest air bubble inside the flow path definitely
will cause the test to fail!
1 Select the system pressure test from the test selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by opening the purge valve when the test has
NOTE completed. Otherwise the pump may generate an overpressure error.
Please notice the difference between an error in the test and a failure of the test!
NOTE An error is caused by the abnormal termination during the operation of the test
whereas a failure of a test indicates that the test results were not within the
specified limits.
Often it is only a damaged blank nut itself (poorly shaped from overtightening)
NOTE that causes the test to fail. Before investigating on any other possible sources of
failure make sure that the blank nut you are using is in good condition and
properly tightened!
• If the test fails again, insert the blank nut at the outlet of the previous module
in the stack (e.g. autosampler, port 6 of the injection valve), and repeat the
pressure test. Exclude each module one by one to determine which module is
leaking.
• If the pump is determined to be the source of the leak, run the valve test to
identify the defective pump component.
Damaged pump seals or pistons. Run the valve test to identify the defective
component.
Loose purge valve. Tighten the purge valve nut (14 mm wrench).
Damaged metering seal or piston. Exchange the metering seal. Check the piston for
scratches. Exchange the piston if required.
Valve Test
Description
The Valve Test provides a fast and accurate way to verify proper hydraulic
operation of the binary pump. Problems that are related to defective valves, seals
or pistons can be diagnosed and usually the defective part is identified.
Step 1
The system is setup with water on both channels and a restriction capillary is
attached to the outlet of the pump. Pump head A is delivering at 1 mL/min. The
pressure signal is monitored and overlaid with the piston movement plot. The
pressure pattern and the slope of the pressure signal are evaluated for the
delivery strokes of both pistons.
Step 2
The procedure from step 1 is repeated on pump head B.
Step 3
The data from step 1 and 2 are evaluated. In case test failed, a conclusion about
the defective part is made.
When The test should be used to prove proper operation of the binary pump after repairs or when the
pressure test (see “(System) Pressure Test” on page 93) determined a problem with the pump.
Preparations Place two bottles of HPLC-grade water in channels A and B (A1 and B1 if the pump is equipped with
a solvent selection valve)
The Valve Test should only be performed when firmware revisions A.06.50 or
NOTE below are installed.
For firmware revisions above A.06.50, perform the Leak Rate Test instead.
Make absolutely sure that the pump is very thoroughly flushed with water before
NOTE starting the test! Any trace of other solvents or the smallest air bubble inside the
flow path definitely will cause the test to generate misleading results!
Make sure to release the pressure by opening the purge valve when the test has
NOTE completed. Otherwise the pump may generate an overpressure error.
Description
Each solvent or solvent mixture has unique compressibility at different
pressures. In order to deliver accurate flow with minimal pressure- and
composition ripple over the full operational pressure range, it is necessary that
the pump compensates precisely for the compressibility of the solvents in use.
The binary pump comes with compressibility parameters for the most common
HPLC solvents and solvent mixtures. If a solvent is not available in the list of
pre-calibrated solvents, the solvent compressibility calibration allows the
appropriate compressibility data to be generated.
Technical background
The solvent compressibility calibration relies on an accurate elasticity calibration
of the pump. With a proper elasticity calibration in place, the pump is switched
into pressure control mode. A restriction capillary is connected to the purge valve
outlet. By varying the flow rate, the pump maintains a certain pressure. The pump
optimizes the compressibility value of the solvent until the lowest possible pump
ripple is reached. The pump increases the flow rate and adjusts the pressure to
the next calibration step where the pump ripple is minimized again. This process
is repeated until solvent compressibility data for the whole operating pressure
range of the pump are available.
The compressibility data set for this solvent is stored in an XML-file on the
desktop. It can be shared with other InfinityLab LC Series binary pumps via the
controlling data system.
When If a solvent is not available in the list of pre-calibrated solvents, the solvent compressibility calibration
allows to generate appropriate compressibility data.
Preparations Place a bottles with solvent to be calibrated in channel A (resp. A1 if a solvent selection valve is
installed).
Make absolutely sure that the pump is very thoroughly flushed with the solvent to
NOTE be calibrated before starting the procedure! Any trace of other solvents or the
smallest air bubble inside the flow path definitely will cause the calibration to fail!
Make sure to release the pressure by opening the purge valve when the test has
NOTE completed. Otherwise the pump may generate an overpressure error.
Description
The flow path components of the binary pump have an inherent and pressure
dependent elasticity which differs from pump to pump. When this
elasticity/pressure function is known, a correction algorithm can be applied. This
results in significantly improved pump performance in low delay volume mode
(damper and mixer bypassed).
The pump elasticity calibration uses a solvent with well known properties
(HPLC-grade water) to determine the pump elasticity over the entire operating
pressure range and stores the calibration values in the non-volatile RAM of the
pump mainboard.
The initial calibration is done at the factory. It only needs to be repeated after
replacement of major pump parts (mainboard, pump drive). The test allows to
define which pump head will be calibrated.
When The initial calibration is done at the factory. It only needs to be repeated after replacement of major
pump parts (mainboard, pump drive).
Make absolutely sure that the pump is very thoroughly flushed with the solvent to
NOTE be calibrated before starting the procedure! Any trace of other solvents or the
smallest air bubble inside the flow path definitely will cause the calibration to fail!
If a solvent selection valve is installed flush all four solvent channels to avoid that
NOTE air from a dry solvent intake tube is drawn into the flow path upon initialization.
1 Select the pump elasticity calibration from the test selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by opening the purge valve when the test has
NOTE completed. Otherwise the pump may generate an overpressure error.
Introduction
The Pump Leak Rate Test is used for verifying the internal tightness of the pump
and helps identifying parts which may have caused a leak.
Minimum firmware revisions:
• D.07.01
1 Select the Pump Leak Rate Test from the Test Selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by slowly opening the purge valve when the
NOTE test has finished.
to column
Damper
Outlet valve
Purge valve
Seals
Piston 2
Secondary Leak
If a leak is found for movement of piston 2 (secondary leak), the following
reasons are possible:
6 Leak at piston 2 or seal in chamber 2 Inspect piston, replace piston and/or seal
Primary Leak
If a leak is found for movement of piston 1 (primary leak), any leak described for
piston movement 2 will cause a failure for piston 1 as well, as the liquid can move
through the outlet valve to chamber 2. Such cases need to be identified as
described before. Additionally, following causes are possible:
1 Leak at piston 1 or seal in chamber 1 Inspect piston, replace piston and/or seal
2 Leak at inlet valve Replace inlet valve or inlet valve cartridge (AIV
only)
1 Leak at outlet valve Replace the part which has failed and re-run the
test.
Introduction to Maintenance
The pump is designed for easy maintenance. The most frequent maintenance
procedures such as piston seal replacement and purge valve frit exchange can
be done from the front side without removing the pump from the system stack.
These procedures are described in “Overview of Maintenance and Simple
Repairs” on page 113.
The module is partially energized when switched off, as long as the power
WAR N IN G
cord is plugged in.
Repair work at the module can lead to personal injuries, e.g. electrical
shock, when the cover is opened and the module is connected to power.
Always unplug the power cable before opening the cover.
Do not connect the power cable to the instrument while the covers are
removed.
To keep the module case clean, use a soft cloth slightly dampened with water, or
a solution of water and mild detergent. Avoid using organic solvents for cleaning
purposes. They can cause damage to plastic parts.
Liquid dripping into the electronic compartment of your module can cause
WAR N IN G
shock hazard and damage the module
Do not use an excessively damp cloth during cleaning.
Drain all solvent lines before opening any connections in the flow path.
Figure 10 on page 113 shows the main user accessible assemblies of the binary
pump. The pump heads and its parts require normal maintenance (for example,
seal exchange) and can be accessed from the front (simple repairs).
Replacement of valve cartridges or filters don’t require to remove the pump from
the system stack.
Pressure sensor
Purge valve
Outlet valve
Pump head
1 Purge valve, see “Exchange the Purge Valve Frit or the Purge Valve” on page 117
3 Active inlet valve, see “Exchange the Active Inlet Valve (AIV) or its Cartridge” on page 134
4 Pump head, see “Maintenance of a Pump Head with Seal Wash” on page 127
5 Solvent selection valve, see “Exchange the Solvent Selection Valve” on page 145
Maintenance Procedures
The procedures described in this section can be done with the binary pump in
place in the system stack.
Table 12 Maintenance procedures
“Exchange the Purge Valve Frit or the Purge Yearly, or if the frit shows A pressure drop of > 10 bar in low delay volume
Valve” on page 117 indication of configuration and > 20 bar in standard
contamination or configuration across the frit (5 mL/min H 2O with
blockage purge valve open) indicates blockage
If internally leaking Solvent dripping out of waste outlet when valve is
closed
“Remove the Pump Head Assembly” on During yearly Necessary to get access to pump seals and
page 122 maintenance pistons
“Maintenance of a Pump Head without Seal Yearly, or if pump Leaks at lower pump head side, unstable
Wash” on page 124 performance indicates retention times, pressure ripple unstable — run
seal wear Pump Leak Rate Test for verification
Seal life time shorter than normally expected —
check pistons while changing the seals
“Maintenance of a Pump Head with Seal Yearly, or if pump Only necessary when Seal Wash Option is
Wash” on page 127 performance indicates installed. Leaks at lower pump head side, loss of
seal wear wash solvent
“Exchange the Active Inlet Valve (AIV) or its If leaking externally Error messages “Inlet Valve Fuse” or “Inlet Valve
Cartridge” on page 134 If solenoid is defective Missing”
“Exchange the Outlet Valve” on page 141 If internally leaking Pressure ripple unstable, run Pump Leak Rate
Test for verification
“Exchange the Solvent Selection Valve” on If internally leaking Cross port flow
page 145 If solenoid is defective Error message “Valve Failed”
The figures shown in this procedure exemplarily show the Infinity II Multisampler
NOTE module.
The principle of how to remove and/or install doors works in the same way for all
Infinity II modules.
1 Press the release buttons and pull the front door out. 2 For the Installation of the front door. Insert the hinges
into their guides and move the door in until the release
buttons click into their final position.
When • Frit – when piston seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar in low delay volume configuration and > 20 bar in standard configuration across the frit
at a flow rate of 5 mL/min of water with purge valve opened)
• Purge valve – if internally leaking
Pair of tweezers
OR Toothpick
G7111-60061
A Purge valve
5067-4728
A Seal cap
1 Using a 1/4 inch wrench disconnect the pump outlet 2 Using the 14 mm wrench, unscrew the purge valve and
capillary from the purge valve. Disconnect the waste remove it from the purge valve holder.
tube. Beware of leaking solvents due to hydrostatic
pressure.
3 Remove the seal cap from the purge valve. 4 Using a pair of tweezers or a toothpick remove the frit.
Seal cap
Frit
5 Place a new frit into the purge valve with the orientation 6 Reinstall the seal cap including the gold seal.
of the frit as shown below (slit in frit points to the front).
8710-1924
A Open-end wrench 14 mm
Pair of tweezers
OR Toothpick
1 Disassemble the purge valve. 2 Remove the old o-ring from the purge valve.
3 Clean the purge valve parts.
4 Place the new o-ring on the Screw Purge Valve. 5 Place the screw with o-ring on the Purge Valve Body.
6 Place the mounting ring on the screw and push down 7 Push the screw up and guide the o-ring into the gap.
the o-ring.
8 Place a new frit into the purge valve with the orientation 9 Install the purge valve to the pump. Make sure not to
of the frit as shown below (slit in frit points to the front). turn the purge valve body when the screw is fixed to the
Reinstall the seal cap including the gold seal. pump. The o-ring will take damage.
NOTE
Before reinstallation always check the gold seal in the seal
cap. A deformed seal cap should be exchanged.
Hexagonal key, 4 mm
5023-0240
A Hex driver, ¼", slitted
Both pump head assemblies use the same internal components. In addition,
NOTE pump head A is fitted with the purge valve. The following procedure describes
the removal and disassembly of pump head A (left). For pump head B (right)
proceed in the same way and skip steps that deal with the purge valve.
1 Open the doors. 2 Disconnect the capillaries at the back of the purge valve
holder, the pump head adapter and the tube at the
active inlet valve. Beware of leaking solvents.
3 Using a 4 mm hexagonal key stepwise loosen and 4 Unplug the active inlet valve cable from the connector.
remove the two pump head screws. Remove the pump head assembly from the module.
Hexagonal key, 4 mm
01018-23702
A Insert tool
OR 1 0905-1420
A PE seals (pack of 2)
1 5063-6586
A Sapphire piston
Both pump head assemblies use the same internal components. In addition,
NOTE pump head A is fitted with the purge valve. The following procedure describes
the removal and disassembly of pump head A (left). For pump head B (right)
proceed in the same way and skip steps that deal with the purge valve.
1 Place the pump head on a flat surface. Loosen the lock 2 While holding the lower half of the assembly (piston
screw (two turns). housing), carefully pull the pump chamber housing
away from the piston housing.
3 Remove the support rings from the piston housing. 4 Lift the housing away from the pistons.
5 Check the piston surface and remove any deposits or 6 Using the steel side of the insert tool, carefully remove
layers: clean the piston surface with abrasive paper and the seal from the pump housing.
rinse with 2-propanol. Replace piston if scratched.
Piston Surface
7 Using the plastic side of the insert tool, insert new seals 8 Place the support rings on the piston housing. Note the
into the pump head. correct position of the pins.
9 Place the pump housing onto the piston housing. 10 Insert the pistons and carefully press them into the
seals.
11 Tighten the lock screw. 12 Install the pump head, see “Reinstall the Pump Head
Assembly” on page 131.
Both pump head assemblies use the same internal components. In addition,
NOTE pump head A is fitted with the purge valve. The following procedure describes
the removal and disassembly of pump head A (left). For pump head B (right)
proceed in the same way and skip steps that deal with the purge valve.
1 Place the pump head on a flat surface. Loosen the lock 2 While holding the lower half of the assembly (piston
screw (two turns). housing), carefully pull the pump housing away from the
piston housing.
3 Remove the seal holder and the seal wash support rings 4 Remove the seal holder from the support ring
from the piston housing. assembly.
5 Lift the housing away from the pistons. 6 Check the piston surface and remove any deposits or
layers: clean the piston surface with abrasive paper and
rinse with 2-propanol. Replace piston if scratched.
Piston Surface
7 Using the steel side of the insert tool, carefully remove 8 Using the plastic side of the insert tool, insert new seals
the seal from the pump housing. into the pump head.
9 Using the steel side of the insert tool, remove the seal 10 Using the plastic side of the insert tool, press the new
wash gasket and the wash seal from the support ring. wash seal (spring pointing upwards) into the recess of
the support ring.
11 Place a seal wash gasket in the recess of the support 12 Place the support rings on the piston housing. Note the
ring. Use a matching orientation of gasket and support correct position of the pins on the support ring.
ring. Put the seal holder on top of the gasket.
13 Place the pump chamber housing onto the piston 14 Insert the pistons and carefully press them into the
housing. seals.
15 Tighten the lock screw. 16 Install the pump head, see “Reinstall the Pump Head
Assembly” on page 131.
Hexagonal key, 4 mm
5023-0240
A Hex driver, ¼", slitted
1 Slide the pump head assembly onto the pump drive. 2 Using a 4 mm hexagonal key, tighten the pump head
Reconnect the active inlet valve cable to the connector. screws stepwise with increasing torque.
1 Put a bottle with 100 ml of isopropanol in the solvent cabinet and place the
solvent intake filter of the pump head you want to wear in into this bottle.
2 Screw the PEEK adapter 1/4-28 to 10-32 (0100-1847) onto the active inlet
valve and connect the inlet tube from the bottle head directly to it.
3 Connect the Restriction capillary (5022-2159) to the purge valve. Connect its
other end to a waste container.
4 Open the purge valve and purge the system for 5 min with isopropanol at a
flow rate of 2 mL/min.
5 Close the purge valve, set the flow to a value that gives a pressure of 580 bar.
Pump 15 min at this pressure to wear the seals in. The pressure can be
monitored with the Instant Pilot, chromatographic data system or any other
controlling device connected to your pump.
6 Turn OFF the pump, slowly open the purge valve to release the pressure from
the system, disconnect the restriction capillary and reconnect the outlet
capillary to the purge valve. Reconnect the intake tubing to the solvent
selection valve and the connecting tube from the solvent selection valve
(if installed) to the AIV.
7 Purge your system with the solvent used for your next application.
1 Open the doors. 2 Unplug the active inlet valve cable from the connector.
3 Disconnect the solvent inlet tube at the active inlet valve 4 Using a 14 mm wrench, loosen the active inlet valve and
(beware of leaking solvents). remove the valve from the pump head.
5 Using a pair of tweezers, remove the valve cartridge 6 Push the cartridge into the new active inlet valve.
from the defective active inlet valve.
7 Screw the new valve into the pump head. With the 8 Position the valve so that the solvent inlet tube
14 mm wrench, turn the nut until it is hand tight. connection points towards the front.
9 Using the 14 mm wrench, tighten the nut by turning the 10 Reconnect the inlet tube to the valve.
valve in its final position (not more than a quarter turn).
Do not overtighten the valve.
11 Reconnect the Active Inlet Valve cable to the connector 12 Close the doors.
in the Z-panel.
1 Remove the wash solvent tubings from the support ring 2 Unclip the peristaltic pump cartridge from the module
outlet and from the adapter leading to the waste bottle. housing and remove it.
3 Put the new peristaltic pump cartridge onto the rod of 4 Connect the peristaltic pump tubes to the support rings
the pump motor and push the plastic clips into the outlet and to the adapter leading to the waste bottle.
module housing.
1 Open the doors. 2 Press the Leak Adapter down (1.) and remove it
together with the tubing (2.).
1.
2.
3 Install the Leak Adapter by pressing it into the Main 4 Insert the Tubing (approximately 85 mm required for
Cover. replacement) between Leak Adapter outlet and Leak
Panel.
8710-0510
A Open-end wrench 1/4 — 5/16 inch
5067-5688
A Torque wrench 1 – 25 Nm with 14 mm wrench
OR 1 G1312-60167
A Outlet valve (type N/SFC)
1 Using a ¼ inch wrench disconnect the absorber 2 Unscrew the valve with the 14 mm wrench and remove
capillary from the outlet valve. it from the pump body.
NOTE
Do not disassemble the outlet valve, as this can damage the
valve.
3 Reinstall the outlet valve and tighten it using a torque 4 Reconnect the capillary.
wrench (approx.12 Nm).
approx.
12 Nm
A solvent selection valve allows you to choose between four different solvents
that can be used with a binary pump. The valve switches between two solvents
A1 and A2 for channel A of the left pump head and two solvents B1 and B2 for
channel B of the right pump head.
1 Break out the metal sheet piece to get access to the 2 Guide the SSV cable through the slots in the distance
connection socket inside (behind the pump front metal sheet and route it through the slot in the pump front
panel). metal panel to the internal connection socket.
3 Carefully plug in the connector of the solvent selection 4 Install the solvent selection assembly by fixing the
valve into the internal connection socket on the left side indicated screws with distance sheet under the SSV
of the slot behind the pump front metal panel. body.
Block unused channels of the SSV using a plug (Blank plug (5041-8365)) to avoid
NOTE leaks or air entering the solvent channels.
Before using of the system with a new installed valve it may be required to pump
NOTE several mL of solvent to get the flow stabilized at a pressure ripple as low as it
used to be when the system was still working properly.
When If leaking internally (crossflow between the ports), or if one of the channels is blocked
1 Lift solvent bottles out of the solvent cabinet and place 2 Disconnect all tubings from the solvent selection valve.
them on the table. Disconnect the solvent tubes from
the solvent selection valve and empty the tubes into the
bottles. Place the bottles back into the solvent cabinet.
3 Using a screwdriver, loosen the holding screws of the 4 Carefully pull the valve holder out.
valve holder.
5 Disconnect the valve cable from the internal connection 6 Completely remove the old valve.
socket by pressing the fixing clip through the left side
middle hole with the screwdriver.
7 Bend the cable at the connector of the new valve. 8 Guide the cable and connector into the hole and push
the connector into the socket.
9 Exchange the defective solvent selection valve. 10 Tighten the screws of the valve holder.
Preparations Read update documentation provided with the Firmware Update Tool.
This chapter lists all parts and tools that are required for maintenance and simple
repairs.
15 1 G1312-87330 Mixer
12 11 13 15 16
18
1
19
8 14
7 20
5 6 17 10 9
Figure 11 Hydraulic Path with Solvent Selection Valve
13 G1312-87330 Mixer
10 9 11 13 14
17
16
12
6
18
5
4 15 8 7
Figure 12 Hydraulic Path without Solvent Selection Valve, with Active Seal Wash
5 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
7 0515-0175 Mounting screw for manual purge valve holder, M4, 20 mm long
13 G1312-23201 Adapter
7
8 1
9
2
10 3a
3b
11 3c
4
5
14
13 6
12 a,b
9 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
OR 0905-1420 PE seals (pack of 2)
10 0515-0175 Mounting screw for manual purge valve holder, M4, 20 mm long
17 G1312-23201 Adapter
10 2
11
12 3
5 4
13 6
14 7
8
9
18
17
15
16 a,b
Outlet Valve
5023-2504
(Hex Driver SW-4 slitted)
5023-3138 5023-2503
(Reversible Screwdriver) (Hex Driver
SW-5 slitted)
5023-2502
(Hex Driver SW-6,
35/ ¼” slitted)
5023-3088
8710-1924
8710-2409
8710-0510
8710-0510
5023-2500
8710-1534
8710-2394
(Hex Key 9/64“, 15cm)
Box with:
9301-0411 (Syringe, Plastic) 5023-3089
9301-1337 (Syringe Adapter) Torx Key Set (T8,T9,T10,T15,T20,T25)
0100-1710 (Mounting tool for flangeless nut)
0100-1681 (Adapter luer/barb) 5023-2499
01018-23702 (Seal Insert tool) (Hex Key Set)
5067-6127 (Blank Nut V) 5043-1361
5023-2653 (Hex Key 3/32“) (Hex Key Set Driver)
The Active Seal Wash Option kit (G1399B) contains the following parts:
# p/n Description
1 5062-2484 Gasket, seal wash (pack of 6)
1 5063-6589 Piston seal PTFE, carbon filled, black (pack of 2), default
Solvent Cabinet
p/n Description
5067-6871 Solvent Cabinet Kit
For details refer to: Usage Guideline for the Solvent Cabinet (01200-90150)
p/n Description
5063-6598 Ferrules with lock ring (10/Pk)
Wire marker
Accessory Kit
Cover Parts
This chapter provides information on cables used with the Agilent InfinityLab LC
Series modules.
Cable Overview
Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE proper functionality and compliance with safety or EMC regulations.
Analog cables
p/n Description
35900-60750 Agilent 35900A A/D converter
Remote cables
p/n Description
5188-8029 ERI
to general purpose
CAN cables
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
LAN cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
RS-232 cables
(not for p/n Description
FUSION board)
RS232-61601 RS-232 cable, 2.5 m
Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It is also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
USB cables
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)
Analog Cables
1 Not connected
2 Shield Analog -
3 Center Analog +
1 Not connected
2 Black Analog -
3 Red Analog +
Remote Cables
10 violet DGND
12 red-blue PGND
13 white-green PGND
NC yellow-brown
10 GND 1
2 Stop 8 Low
3 Ready 7 High
5 Power on 6 High
4 Future 5
7 Start 3 Low
8 Prepare 2 Low
p/n 5188-8057 Pin (ERI) Signal Pin (APG) Active (TTL) Pin (RJ45)
10 GND 1 5
1 Start 9 High
Request
2 Stop 8 High
3 Ready 7 High
4 Fraction 5 High 4
Trigger
5 Power on 6 High
7 Start 3 High
8 Prepare 2 High
Ground Cable NC
Shielding
p/n 5061-3378 Pin 35900 A/D Pin Agilent Signal Name Active
module (TTL)
CAN/LAN Cables
p/n Description
5181-1516 CAN cable, Agilent module to module, 0.5 m
LAN Cables
p/n Description
5023-0203 Cross-over network cable, shielded, 3 m (for point to point connection)
p/n Description
RS232-61601 RS-232 cable, 2.5 m
Instrument to PC, 9-to-9 pin (female). This cable has special pin-out,
and is not compatible with connecting printers and plotters. It is also
called "Null Modem Cable" with full handshaking where the wiring is
made between pins 1-1, 2-3, 3-2, 4-6, 5-5, 6-4, 7-8, 8-7, 9-9.
To connect a USB Flash Drive use a USB OTG cable with Mini-B plug and A
socket.
p/n Description
5188-8050 USB A M-USB Mini B 3 m (PC-Module)
This chapter provides detailed technical information about your binary pump.
Firmware Description
Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232)
• memory management
• ability to update the firmware of the 'main system'
Main System
Its properties are:
• the complete communication capabilities (CAN, LAN, USB and RS- 232)
• memory management
• ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are divided
into common functions like
• run synchronization through APG/ERI remote,
• error handling,
• diagnostic functions,
• or module specific functions like
• internal events such as lamp control, filter movements,
• raw data collection and conversion to absorbance.
Firmware Updates
Firmware updates can be done with the Agilent Lab Advisor software with files
on the hard disk (latest version should be used).
Required tools, firmware and documentation are available from the Agilent web:
http://www.agilent.com/en-us/firmwareDownload?whid=69761
The file naming conventions are:
PPPP_RVVV_XXX.dlb, where
• PPPP is the product number, for example, 1315B for the G1315B DAD,
• R the firmware revision, for example, A for G1315B or B for the G1315C DAD,
• VVV is the revision number, for example 650 is revision 6.50,
• XXX is the build number of the firmware.
For instructions on firmware updates refer to section Replacing Firmware in
chapter "Maintenance" or use the documentation provided with the Firmware
Update Tools.
Update of main system can be done in the resident system only. Update of the
NOTE resident system can be done in the main system only.
Main and resident firmware must be from the same set.
Main FW update
Resident System Main System
Resident FW Update
The firmware update tools, firmware and documentation are available from the
Agilent web.
• http://www.agilent.com/en-us/firmwareDownload?whid=69761
Electrical Connections
• The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal module data transfer and
synchronization.
• The ERI/REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shutdown, prepare, and so on.
• With the appropriate software, the LAN connector may be used to control the
module from a computer through a LAN connection. This connector is
activated and can be configured with the configuration switch.
• With the appropriate software, the USB connector may be used to control the
module from a computer through a USB connection.
• The power input socket accepts a line voltage of 100 – 240 VAC ± 10 % with a
line frequency of 50 or 60 Hz. Maximum power consumption varies by
module. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible fuses
because automatic electronic fuses are implemented in the power supply.
Never use cables other than the ones supplied by Agilent Technologies to ensure
NOTE proper functionality and compliance with safety or EMC regulations.
CCXZZ00000 Format
CC Country of manufacturing
• DE = Germany
• JP = Japan
• CN = China
Configuration switch
USB-Mini-Port
LAN
ERI
CAN
Power socket
Interfaces
Pumps
Samplers
Detectors
Fraction Collectors
Others
G1170A 2 No No No No No
G7122A No No No Yes No A
Overview Interfaces
CAN
The CAN is inter-module communication interface. It is a 2-wire serial bus
system supporting high speed data communication and real-time requirement.
LAN
The modules have either an interface slot for a LAN card (e.g. Agilent G1369B/C
LAN Interface) or they have an on-board LAN interface (e.g. detectors G1315C/D
DAD and G1365C/D MWD). This interface allows the control of the
module/system via a PC with the appropriate control software. Some modules
have neither on-board LAN nor an interface slot for a LAN card (e.g. G1170A
Valve Drive or G4227A Flexible Cube). These are hosted modules and require a
Host module with firmware B.06.40 or later or with additional G1369C LAN Card.
USB
The USB interface replaces the RS-232 Serial interface in new FUSION generation
modules. For details on USB refer to “USB (Universal Serial Bus)” on page 192.
Remote (ERI)
The ERI (Enhanced Remote Interface) connector may be used in combination
with other analytical instruments from Agilent Technologies if you want to use
features as common shut down, prepare, and so on.
It allows easy connection between single instruments or systems to ensure
coordinated analysis with simple coupling requirements.
The subminiature D connector is used. The module provides one remote
connector which is inputs/outputs (wired- or technique).
To provide maximum safety within a distributed analysis system, one line is
dedicated to SHUT DOWN the system’s critical parts in case any module detects
a serious problem. To detect whether all participating modules are switched on
or properly powered, one line is defined to summarize the POWER ON state of all
connected modules. Control of analysis is maintained by signal readiness READY
for next analysis, followed by START of run and optional STOP of run triggered on
the respective lines. In addition PREPARE and START REQUEST may be issued.
The signal levels are defined as:
• standard TTL levels (0 V is logic true, + 5.0 V is false),
• fan-out is 10,
• input load is 2.2 kOhm against + 5.0 V, and
• output are open collector type, inputs/outputs (wired- or technique).
All common TTL circuits operate with a 5 V power supply. A TTL signal is defined
NOTE as "low" or L when between 0 V and 0.8 V and "high" or H when between 2.0 V
and 5.0 V (with respect to the ground terminal).
1 START REQUEST (L) Request to start injection cycle (for example, by start key on any
module). Receiver is the autosampler.
2 STOP (L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is any
module performing run-time controlled activities.
3 READY (H) System is ready for next analysis. Receiver is any sequence
controller.
4 POWER ON (H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.
5 Not used
6 SHUT DOWN (L) System has serious problem (for example, leak: stops pump).
Receiver is any module capable to reduce safety risk.
Special Interfaces
There is no special interface for this module.
ERI Description
The ERI interface contains eight individual programmable input/output pins. In
addition, it provides 24 V power and 5 V power and a serial data line to detect and
recognize further add-ons that could be connected to this interface. This way the
interface can support various additional devices like sensors, triggers (in and out)
and small controllers, etc.
ERI
1 IO 1 (START
REQUEST)
2 IO 2 (STOP)
3 IO 3 (READY)
4 IO 4 (POWER ON)
5 IO 5 (NOT USED)
6 IO 6 (SHUT DOWN)
7 IO 7 (START)
8 IO 8 (PREPARE)
9 1 wire DATA
10 DGND
11 +5 V ERI out
12 PGND
13 PGND
IO (Input/Output) Lines
• Eight generic bi-directional channels (input or output).
• Same as the APG Remote.
• Devices like valves, relays, ADCs, DACs, controllers can be
supported/controlled.
The 6-bit configuration switch is located at the rear of the module with FUSION
electronics. Switch settings provide configuration parameters for LAN and
instrument specific initialization procedures.
All modules with FUSION electronics:
• Default is ALL switches DOWN (best settings).
• Default IP address for LAN 192.168.254.11
• For specific LAN modes switches 4-5 must be set as required.
• For boot resident/cold start modes switches 1+2 or 6 must be UP.
Configuration switch
Mode Function/Setting
1 When selecting mode COM, settings are stored to non-volatile memory. When selecting mode TEST, COM settings are taken from
non-volatile memory.
Special Settings
Boot-Resident/Main
Firmware update procedures may require this mode in case of firmware loading
errors (main/resident firmware part).
If you use the following switch settings and power the instrument up again, the
instrument firmware stays in the resident/main mode. In resident mode, it is not
operable as a module. It only uses basic functions of the operating system for
example, for communication. In this mode the main firmware can be loaded
(using update utilities).
Loss of data
C AU T I O N
Forced cold start erases all methods and data stored in the non-volatile
memory. Exceptions are calibration settings, diagnosis and repair log books
which will not be erased.
Save your methods and data before executing a forced cold start.
If you use the following switch settings and power the instrument up again, it will
start as described above.
EMF Counters
EMF counters increment with use and can be assigned a maximum limit which
provides visual feedback in the user interface when the limit is exceeded. Some
counters can be reset to zero after the required maintenance procedure.
Instrument Layout
This chapter is generic and may show figures that differ from your module. The
NOTE functionality is the same.
1 Note the MAC (Media Access Control) address for further reference. The MAC
or hardware address of the LAN interfaces is a world wide unique identifier.
No other network device will have the same hardware address. The MAC
address can be found on a label at the rear of the module underneath the
configuration switch (see Figure 24 on page 200).
Figure 23 MAC-Label
LAN interface
MAC label
Configuration Switches
Configuration switch
The module is shipped with all switches set to OFF, as shown above.
To perform any LAN configuration, SW1 and SW2 must be set to OFF.
NOTE
0 0 0 1 0 0 Use DHCP
Using Stored
When initialization mode Using Stored is selected, the parameters are taken from
the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured previously
by one of the described methods.
Non-Volatile
RAM
Active
Parameter
Stored
Parameter
Using Default
When Using Default is selected, the factory default parameters are taken instead.
These parameters enable a TCP/IP connection to the LAN interface without
further configuration, see Table 19 on page 204.
Default Active
Parameter Parameter
Using the default address in your local area network may result in network
NOTE problems. Take care and change it to a valid address immediately.
IP address: 192.168.254.11
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the module must reside in the same
subnet.
The user may open a Telnet session using the default IP address and change the
parameters stored in the non-volatile memory of the module. He may then close
the session, select the initialization mode Using Stored, power-on again and
establish the TCP/IP connection using the new parameters.
When the module is wired to the PC directly (e.g. using a cross-over cable or a
local hub), separated from the local area network, the user may simply keep the
default parameters to establish the TCP/IP connection.
In the Using Default mode, the parameters stored in the memory of the module
NOTE are not cleared automatically. If not changed by the user, they are still available,
when switching back to the mode Using Stored.
DHCP Active
Server Parameter
1 It may take some time until the DHCP server has updated the DNS server with
NOTE the hostname information.
2 It may be necessary to fully qualify the hostname with the DNS suffix, e.g.
0030d3177321.country.company.com.
3 The DHCP server may reject the hostname proposed by the card and assign a
name following local naming conventions.
Setup (DHCP)
The DHCP functionality is available on all Agilent HPLC modules with on-board LAN Interface or LAN
Interface Card G1369C, and “B”-firmware (B.06.40 or above) or modules with "D"-firmware. All
modules should use latest firmware from the same set.
1 Note the MAC address of the LAN interface (provided with G1369C LAN
Interface Card or mainboard). This MAC address is on a label on the card or at
the rear of the mainboard, for example, 0030d3177321.
On the Local Controller the MAC address can be found under Details in the
LAN section.
2 Set the configuration switch to DHCP either on the G1369C LAN Interface
Card or the mainboard of above mentioned modules.
Table 20 G1369C LAN Interface Card (configuration switch on the card)
SW 4 SW 5 SW 6 SW 7 SW 8 Initialization Mode
Table 21 LC Modules with 8-bit configuration switch (B-firmware) (configuration switch at rear
of the instrument)
SW 6 SW 7 SW 8 Initialization Mode
Manual Configuration
Manual configuration only alters the set of parameters stored in the non-volatile
memory of the module. It never affects the currently active parameters.
Therefore, manual configuration can be done at any time. A power cycle is
mandatory to make the stored parameters become the active parameters, given
that the initialization mode selection switches are allowing it.
TELNET
Session
Non-Volatile
RAM
Stored
Parameter
Control
Module
With Telnet
Whenever a TCP/IP connection to the module is possible (TCP/IP parameters set
by any method), the parameters may be altered by opening a Telnet session.
1 Open the system (DOS) prompt window by clicking on Windows START
button and select “Run...”. Type “cmd” and press OK.
2 Type the following at the system (DOS) prompt:
• c:\>telnet <IP address> or
• c:\>telnet <host name>
where <IP address> may be the assigned address from a Bootp cycle, a
configuration session with the Handheld Controller, or the default IP address
(see “Configuration Switches” on page 202).
When the connection was established successfully, the module responds
with the following:
3 Type
? and press enter to see the available commands.
Value Description
6 Change the IP address (in this example 192.168.254.12) and type “/” to list
current settings.
change of IP setting to
Initialization mode is Using Stored
If the Initialization Mode Switch is changed now to “Using Stored” mode, the
NOTE instrument will take the stored settings when the module is re-booted. In the
example above it would be 192.168.254.12.
Safety Standards
This is a Safety Class I instrument (provided with terminal for protective earthing)
and has been manufactured and tested according to international safety
standards.
General
Do not use this product in any manner not specified by the manufacturer. The
protective features of this product may be impaired if it is used in a manner not
specified in the operation instructions.
In Case of Damage
Damage to the module
WAR N IN G
Personal injury (for example electrical shock, intoxication)
Instruments that appear damaged or defective should be made
inoperative and secured against unintended operation until they can be
repaired by qualified service personnel.
Solvents
Toxic, flammable and hazardous solvents, samples and reagents
WAR N IN G
The handling of solvents, samples and reagents can hold health and safety
risks.
When working with these substances observe appropriate safety
procedures (for example by wearing goggles, safety gloves and
protective clothing) as described in the material handling and safety
data sheet supplied by the vendor, and follow good laboratory practice.
Do not use solvents with an auto-ignition temperature below 200 °C
(392 °F). Do not use solvents with a boiling point below 56 °C (133 °F).
Avoid high vapor concentrations. Keep the solvent temperature at least
40 °C (72 °F) below the boiling point of the solvent used. This includes
the solvent temperature in the sample compartment. For the solvents
methanol and ethanol keep the solvent temperature at least 25 °C
(45 °F) below the boiling point.
Do not operate the instrument in an explosive atmosphere.
Do not use solvents of ignition Class IIC according IEC 60079-20-1
(for example, carbon disulfide).
Reduce the volume of substances to the minimum required for the
analysis.
Never exceed the maximum permissible volume of solvents (8 L) in the
solvent cabinet. Do not use bottles that exceed the maximum
permissible volume as specified in the usage guideline for solvent
cabinet.
Ground the waste container.
Regularly check the filling level of the waste container. The residual free
volume in the waste container must be large enough to collect the waste
liquid.
To achieve maximal safety, regularly check the tubing for correct
installation.
For details, see the usage guideline for the solvent cabinet. A printed copy of the
NOTE guideline has been shipped with the solvent cabinet, electronic copies are
available in the Agilent Information Center or via the Internet.
Safety Symbols
Table 23 Symbols
The apparatus is marked with this symbol when the user shall refer to the
instruction manual in order to protect risk of harm to the operator and to
protect the apparatus against damage.
The apparatus is marked with this symbol when hot surfaces are available
and the user should not touch it when heated up.
Flammable Material
For Sample Thermostat which uses flammable refrigerant consult Agilent
Information Center / User Manual before attempting to install or service this
equipment. All safety precautions must be followed.
Manufacturing date.
Pacemaker
Magnets could affect the functioning of pacemakers and implanted heart
defibrillators.
A pacemaker could switch into test mode and cause illness. A heart
defibrillator may stop working. If you wear these devices keep at least
55 mm distance to magnets. Warn others who wear these devices from
getting too close to magnets.
Table 23 Symbols
Magnetic field
Magnets produce a far-reaching, strong magnetic field. They could damage
TVs and laptops, computer hard drives, credit and ATM cards, data storage
media, mechanical watches, hearing aids and speakers. Keep magnets at
least 25 mm away from devices and objects that could be damaged by
strong magnetic fields.
A WARNING
WAR N IN G
alerts you to situations that could cause physical injury or death.
Do not proceed beyond a warning until you have fully understood and
met the indicated conditions.
A CAUTION
C AU T I O N
alerts you to situations that could cause loss of data, or damage of equipment.
Do not proceed beyond a caution until you have fully understood and met
the indicated conditions.
This product complies with the European WEEE Directive marking requirements.
The affixed label indicates that you must not discard this electrical/electronic
product in domestic household waste.
Radio Interference
Sound Emission
Sound pressure
Sound pressure Lp <70 db(A) according to DIN EN ISO 7779
Schalldruckpegel
Schalldruckpegel Lp <70 db(A) nach DIN EN ISO 7779
For the latest information on products and services visit our worldwide web site
on the Internet at:
http://www.agilent.com
www.agilent.com
Agilent Technologies Inc. 2016-2021
Edition: 10/2021
Document No: SD-29000214 Rev. C