Comen S8 Service Manual
Comen S8 Service Manual
Service Manual
Copyright
Model: S8
Statement
Shenzhen Comen Medical Instruments Co., Ltd. (belowComen is the copyright of this
non-publicly published maintenance manual and has the right to treat it as confidential
information. This Service Manual is a reference for the maintenance of Comen products.
Copyright, no part of this Service Manual may be photocopied, copied or translated into
Comen is not liable for errors in the service manual or for incidental or
consequential damages resulting from the actual performance and use of this service
manual.Comen does not provide other parties with a franchise under the patent law.Comen
is not liable for legal consequences resulting from violations of patent law and any
Contact
Address:10-11F, Building 1A, 1A Building, Feiyada Clock Building, South Ring Road,
Tel:0755-26431236
2
Defibrillator Monitor S Series Service Manual
Fax:0755-26431232
Zip Code:518052
3
Defibrillator Monitor S Series Service Manual
Directory
目录
Chapter 1 Safety ................................................................................................................................................ 6
1.1 Safety Information ............................................................................................................................. 6
1.2 Symbol Description ........................................................................................................................ 11
Chapter2 Warranty ......................................................................................................................................... 12
2.1 Warranty Terms.............................................................................................................................. 12
2.2 Exclusion of warranty ....................................................................................................................... 12
2.3 Maintenance procedures ................................................................................................................. 13
2.3.1 Fill in the Customer Complaint Form (SCF) ............................................................................ 14
2.3.2 Submit a customer complaint form and choose a solution ................................................... 14
2.3.3 RMA Form............................................................................................................................... 15
2.3.4 Return ..................................................................................................................................... 15
Chapter 3 Introduction to Principles ............................................................................................................... 17
................................................................................................................................................................. 17
3.1 Overview ......................................................................................................................................... 17
3.1.1 Introduction to Defibrillator ................................................................................................... 17
3.1.2 System Composition ............................................................................................................... 17
3.1.3 system structure diagram....................................................................................................... 19
3.2 Module Introduction ....................................................................................................................... 19
3.2.1 Main control board................................................................................................................. 19
3.2.2 AC power board...................................................................................................................... 22
3.2.3 Keyboard ................................................................................................................................ 22
3.2.4 DC power board .................................................................................................................. 24
3.2.5 Treatment module ................................................................................................................. 26
3.2.6 Side panel ............................................................................................................................... 32
3.2.7 NIBP module ........................................................................................................................... 33
3.2.8 IBP module ............................................................................................................................. 36
3.2.9 ECG Module ............................................................................................................................ 36
3.2.10 SPO2 Module ........................................................................................................................ 37
Chapter 4 Troubleshooting ............................................................................................................................. 41
................................................................................................................................................................. 41
4.1 Introduction ...................................................................................................................................... 41
4.2 Replaceable parts ........................................................................................................................... 41
4.3 Checking the monitor ....................................................................................................................... 41
4.4 Checking the software ...................................................................................................................... 41
4.5 Check the error message prompt ................................................................................................ 42
4.6 Troubleshooting.............................................................................................................................. 42
4.6.1 Electricity failure..................................................................................................................... 42
4
Defibrillator Monitor S Series Service Manual
5
Defibrillator Monitor S Series Service Manual
Chapter 1 Safety
WARNING
adverse events Case. Failure to follow the warning will result in death or
CAUTION
property damage, and may result in more serious injury in the future.
PAY ATTENTION TO
WARNING
6
Defibrillator Monitor S Series Service Manual
● This defibrillator is intended for use in clinical patient care and only allows
● Before use, the user should check that the instrument and its accessories
● The alarm volume and alarm upper and lower limits should be set for
audible alarm system to monitor the patient. If the alarm volume is set too
small or completely turned off, the alarm may be invalidated and the
of the patient.
not use the socket and use a rechargeable battery device to supply power.
● Do not open the case of the instrument to avoid possible electric shock
accordance with local laws and regulations or the hospital's waste disposal
7
Defibrillator Monitor S Series Service Manual
● Carefully install the power cord and various accessory cables to avoid
interference.
● Do not use a mobile phone near the monitor. The mobile phone generates
● For patients with pacemakers, the heart rate monitor may pulse
monitored. Refer to the monitor's instruction manual for the ability of the
● During defibrillation, the operator should not touch the patient, the table
and the instrument. Check that the function is normal before using these
cables again.
● When the monitor is shared with the electrosurgical device, the user
8
Defibrillator Monitor S Series Service Manual
information displayed by the monitor are for medical reference only and
the use of other Instrumentsin the vicinity of the instrument must meet
MRI devices can all be sources of interference because they emit high-
● After defibrillation, the recovery time of ECG is not more than 10s, and the
● To prevent burns to the patient when the monitor is used with a high
CAUTION
9
Defibrillator Monitor S Series Service Manual
● Before the device is powered on, please confirm that the power supply
● When the instrument and accessories are about to expire, they must be
● Disposable accessories can only be used once, and repeated use can result
● If the monitor will not be used for a long time, remove the battery and
PAY ATTENTION TO
maintain.
10
Defibrillator Monitor S Series Service Manual
manufacturing measuring
refer to the operation manual instrument license mark and
Guangdong
number
00000700
network connection
VGA interface
symbol
output port
table 1
11
Defibrillator Monitor S Series Service Manual
gauge2
Chapter2 Warranty
Comen provides a 24 month (host) warranty or a 6 month (accessory) warranty for all licensed products,
starting with the customer's purchase date. If your product is determined to be defective and promptly
notified to us during the warranty period, Comen will repair the product or replace it with a new product or
accessory.
12
Defibrillator Monitor S Series Service Manual
Repair request
Instrument collection
No need to Need to be
Instrument
status
Failed
Office maintenance
Product
testing
Return to office/client
figure 1
13
Defibrillator Monitor S Series Service Manual
Fill in the customer complaint form in detail: model number, serial number (SN) and problem description.
Without this information, Comen will not be obliged to accept this complaint. Customers can apply for a
complaint form from Comen's after-sales department if needed.
Serial number
(SN) see this
nameplate
Figure 1
Once the after-sales department receives the customer complaint form, the engineer will provide the
customer solution within three days. And deal with it according to the following two situations:
Warranty period:
1) After receiving the Return Authorization Form (RMA) from the After-sales Department, the customer
can return the defective part and inform the delivery number. We will send a new part with the loading
invoice to your address as soon as possible.
2) You can also send the signed statement to us by mail or fax. The statement has the legal effect to ensure
that the customer or the end customer will send the faulty part to Comen in time. In this case, we will
send the replacement parts and loading invoices in a timely manner.
14
Defibrillator Monitor S Series Service Manual
Caution
Once the customer complaint form has been confirmed, Comen engineers will provide you with
a return authorization and declaration form.
The customer is responsible for shipping, insurance and customs fees for the shipment of the
product to Comen.
After receiving the return authorization form, the customer will send the faulty part to Comen. Our
engineers will analyze the faulty parts and negotiate with the customer to repair or replace the faulty parts.
After the maintenance fee has been paid, we will send out new replacement parts as soon as possible.
Caution
The customer is responsible for all costs (including freight, insurance and customs fees) during
the transportation process.
Before returning the goods, the customer will receive an RMA Form (Return Authorization Form)
provided by our after-sales department, which includes the RMA number, a brief description of the returned
parts and shipping instructions. When shipping, please indicate the RMA number on the box.
Caution
In the absence of a notice to the Comen after-sales department, we will not be obligated to process
returns from the terminal or customer. And the shipper is responsible for all possible expenses.
2.3.4 Return
When disassembling the instrument, use an anti-static requirements. Do not touch the instrument
directly with your hands.
15
Defibrillator Monitor S Series Service Manual
When describing the product, please use the “*** sample” format to reflect the total value on the
invoice and mark “sample, no commercial value”. .
Before returning, please check with Comen (for example: total amount, address and other necessary
information on the invoice).
After receiving confirmation from Comen, please return the faulty part.
16
Defibrillator Monitor S Series Service Manual
3.1 Overview
S-Series defibrillation monitors can monitor ECG, RESP, SpO2, NIBP, Invasive Blood Pressure (IBP), Carbon
Dioxide (CO2), Body Temperature (TEMP), etc. Vital sign parameters. This product is suitable for manual
defibrillation, AED defibrillation, pacing and vital signs monitoring.
The defibrillator monitor consists of a host, a battery, a defibrillation electrode plate (defibrillation
external electrode plate, and a defibrillation internal electrode plate) and corresponding functional
18
Defibrillator Monitor S Series Service Manual
Figure 3
The main board is the core of the entire monitor, responsible for processing: input and output
control, data storage and processing, display, system control, communication management and more.
19
Defibrillator Monitor S Series Service Manual
Figure 2
20
Defibrillator Monitor S Series Service Manual
J33
J3
J4
J32
P1
J8
J7
Figure 3
5V, 12V, 18V connected to the LCD screen 5V, 12V, 3.3V
P1 (8000F
table 1
21
Defibrillator Monitor S Series Service Manual
P1
P2
Figure 6
table 2
3.2.3 Keyboard
The keypad is located in the front of the monitor. Through the MCU processing on the keypad, the
signal can be transmitted to the main board, as well as the display of the power indicator, AC indicator and
charging indicator.
22
Defibrillator Monitor S Series Service Manual
J5
J10
J14
J13 J9
J8
Figure 7
Table 3
23
Defibrillator Monitor S Series Service Manual
J9
J7
J6
J3 Figure 8
24
Defibrillator Monitor S Series Service Manual
J15
J1 J14
J12
J2
J13
Figure 9
J1 10pin, battery A /
J2 10pin, battery B /
25
Defibrillator Monitor S Series Service Manual
table 4
J2
J13
Figure 10
J4 J8
26
Defibrillator Monitor S Series Service Manual
J12
J25
J26
Figure 11
5pin,connection Charging
J1 /
capacitor
27
Defibrillator Monitor S Series Service Manual
table 5
J1
J3
Figure 12
28
Defibrillator Monitor S Series Service Manual
J23
J7
Figure 13
29
Defibrillator Monitor S Series Service Manual
J3
J2 J7
Figure 14
30
Defibrillator Monitor S Series Service Manual
J2
J10
J11
J9
J8
J13
J6 J5
J7
Figure 15
31
Defibrillator Monitor S Series Service Manual
Figure 16
32
Defibrillator Monitor S Series Service Manual
table 6
The NIBP module is responsible for blood pressure pumping, pressure detection, and
overvoltage protection.
EEPROM
CPU
COM
NIBP通讯接口
Sensor CUFF
NIBP board
Figure 17
J7
J4 J3 J6
33
Defibrillator Monitor S Series Service Manual
J11
J5
J1
J2
J10
J8
J9
Figure 18
J12
Figure 18
34
Defibrillator Monitor S Series Service Manual
J7 10pin, docking /
analog/digital/Nellcor oximeter
table 7
35
Defibrillator Monitor S Series Service Manual
J2
J1
Figure19
table 8
36
Defibrillator Monitor S Series Service Manual
J2
J3
Figure 20
table9
Three Blood Oxygen Modules: COMEN Blood Oxygen, Masimo Blood Oxygen and Nellcor
Blood Oxygen
The oximeter is mainly used for the collection of blood oxygen signals,
amplification simulation, and conversion and processing of electrical signals.
During the monitoring process, the pulse amplitude of the optical signal also
changes. By calculation, blood oxygen parameters, pulse rate signals and volume waves
can be obtained. This data is then transferred to the main board via a special
communication protocol.
37
Defibrillator Monitor S Series Service Manual
J2
J1
Figure 21
J1 2*5pin,SPO2Module 3.3V
Communication Interface
J2 2*4pin,SPO2Module /
table 10
38
Defibrillator Monitor S Series Service Manual
J1
J2
Figure 22
J2 2*5pin , SPO2Module /
Measurement signal input
interface
table 11
J1
J2
39
Defibrillator Monitor S Series Service Manual
Figure
J1 2*5pin,SPO2Module 3.3V
Communication Interface
J2 2*4pin , SPO2Module /
Measurement signal input
interface
table 12
40
Defibrillator Monitor S Series Service Manual
Chapter 4 Troubleshooting
4.1 Introduction
In this chapter, some common faults and troubleshooting methods for defibrillation monitors are
introduced. If your fault is not listed, please contact Comen's after-sales service department.
The main components and components in the monitor PCB are replaceable. If a separate PCB board is
faulty, follow the instructions for “Assemble and Disassemble”, replace it with a brand new PCB board, and
finally check if the monitor's performance is restored. If the problem persists, reinstall the faulty PCB and
continue troubleshooting according to the instructions in this section. You can also send the PCB back to
Comen for repair.
In troubleshooting, some require you to check the hardware version of the monitor and the status of
the instrument.
In troubleshooting, some require the customer to check the monitor's configuration and software
version.
View the system configuration and software version: After booting the self-test (heep beep), press
Main Menu -> Monitor Maintenance -> Monitor Information to view the system running time, system
software version and hardware version.
41
Defibrillator Monitor S Series Service Manual
Figure 4
Check the error message prompt area of the monitor before troubleshooting. If there is an error
message, remove the alarm first according to the instruction manual.
4.6 Troubleshooting
The monitor does not The power cable is not Check the connection of the
42
Defibrillator Monitor S Series Service Manual
AC power module is
Replace the AC power module
damaged
table 1
Broken wiring or poor and the main board; check the LED
The monitor works
normally, the display is contact Screen and main control board
lacking color or the screen is wiring
43
Defibrillator Monitor S Series Service Manual
Display module
Replace the display module
damage
Display module
Replace the display module
damage
table 2
44
Defibrillator Monitor S Series Service Manual
table 3
phenomen
reason measure
on
SPO2 Module is
ReplaceSPO2Module
damaged
45
Defibrillator Monitor S Series Service Manual
blood pressure
Measuring SPO2 under normal
measurement or other
conditions
conditions such as shock or
low temperature
SPO2 Moduleis
Replace SPO2 Module
damaged
table 4
46
Defibrillator Monitor S Series Service Manual
pressure)
The inflation has a numerical value, but
Blood pressure pump leaks the value is very unstable. It can be seen
Patient type error with the cuff in the patient type display
area
Wrong blood
The cuff is too tight or too loose; the
pressure Blood pressure accessories
cuff is leaking or is damaged; the blood
leak
pressure trachea is leaking
table 5
47
Defibrillator Monitor S Series Service Manual
value display Host display failure Please press 4.6.2 to check and replace
table 6
table 7
48
Defibrillator Monitor S Series Service Manual
table 8
table 22
49
Defibrillator Monitor S Series Service Manual
treatment Module
50
Defibrillator Monitor S Series Service Manual
Module
51
Defibrillator Monitor S Series Service Manual
table 23
1.Before upgrading, please check the version of the software (for details, please refer to section 4.4).
2.Put the folder on the computer (the folder name is SOFTUPDATE, you can't
change the file name, otherwise you can't complete the upgrade). The whole copy is in the root directory
of the U disk (that is, open the U disk directly)
After preparing for the work, perform the following software upgrades:
1) Press the power button and the machine starts:
2) The machine will automatically enter the upgrade interface, please wait for the machine
upgrade to complete:
3) Wait a few minutes to upgrade, the screen appears "system update successed, please
restart", you can reboot.
Caution
52
Defibrillator Monitor S Series Service Manual
Please check with Comen after-sales personnel to see if the upgrade software meets your
instrument.
The upgrade process will last for 30 seconds - 2 minutes. Please ensure that there is sufficient
battery power or AC power.
The upgrade of the program can only be operated by professional after-sales personnel.
Do not remove or insert a USB flash drive while the software is being upgraded.
53
Defibrillator Monitor S Series Service Manual
requirements:
Suitable catheter
Cylinder
3) Select [NIBP] in the parameter area, and select [Leakage Detection] in the [NIBP
Settings] menu that pops up.
4) At this time, “leakage detection...” will be displayed below the NIBP parameter area
on the screen, and the table system will start to perform air leak detection.
6) After about 20 seconds, the system will automatically open the bleed valve to mark
the completion of the leak measurement.
If there is no prompt information in the NIBP parameter area, the table shows that there
is no air leakage. If “Pump leak...” is displayed, there may be a leak in the gas path. At
this point, the operator should check whether the entire connection is loose. After confirming
that the connection is correct, perform another leak detection. If there is still a fault
indication, please contact the manufacturer for repair.
54
Defibrillator Monitor S Series Service Manual
Figure 5
requirements:
T-type interface
suitable trachea
1. Connect the monitor, pressure gauge, spherical air pump and metal container
(instead of the cuff) as shown in the figure below。
55
Defibrillator Monitor S Series Service Manual
Figure 6
2. The pressure gauge reading should be zero before inflation. If it is not zero, open the ball pump valve
so that the entire gas path leads to the atmosphere, and then close the valve after the standard
pressure gauge reads zero.
3. In the [Monitor Maintenance] menu, select [User Maintenance], enter the password and enter the
[User Maintenance] menu, select [NIBP Check], and set the monitor to "Check" mode.。
4. The pressure in the metal container was inflated to 0, 50 and 200 mmHg by a ball type air pump.
The difference between the value of the standard pressure gauge and the pressure indicated by the
monitor should be within 3 mmHg. Otherwise, please contact the Comen service engineer.。
Test requirements:
1. Connect the invasive pressure simulator and monitor through the accessory, set the
output static pressure of the invasive pressure simulator to 0mmHg (0kPa), change
the current IBP channel pressure name in the monitor to ART and zero, the system
should prompt the school Zero success, and the average pressure value is also shown
as 0 mmHg (0 kPa).
56
Defibrillator Monitor S Series Service Manual
2. Set the static pressure value of the artificial pressure simulator to 100mmHg, 200mmHg,
dynamic pressure value 120/80mmHg, the pressure display range of the current channel
should be (0~1mmHg, 98~102mmHg, 196~204mmHg, (118~122) / (79 ~ 81) mmHg).
Test requirements:
Use ECG lead and ECG simulator to connect one by one. After the ECG simulator is powered
on, press the “NSR” button of the ECG simulator to observe the HR value of the simulator
screen and the ECG waveform of the monitor. After stabilization, the HR value of the
instrument screen is ±1% or ±1bmp.
requirements requirements:
SPO2 probe
1. Connect the SPO2 probe to the monitor. Set the patient type to adult.
3. Check the volumetric and pulse rate readings to ensure that the SPO2 value is in the
range of 95%-100%.
4. Remove the SPO2 sensor from your finger and the value and waveform gradually become
zero.
57
Defibrillator Monitor S Series Service Manual
requirements requirements:
Constant temperature water bath, standard mercury thermometer with a graduation value of 0.05 °C.
The standard mercury thermometer with a graduation value of 0.05 ° C and the monitor
body temperature sensor are closely attached together, and the thermostatic water bath
is placed at the same time, and the water temperature of the constant temperature water
bath is adjusted. Compare the measured value of the monitor with the standard thermometer
degree. In order to ensure the reliability of the test, the volume of water is required
to be not less than 2L, the water depth is not less than 10cm, and the water temperature
is uniform after being fully stirred. Using a constant temperature water bath and a
standard thermometer, select the temperature point of 0 ° C, 37 ° C, 50 ° C, the
monitor body temperature probe and thermometer are simultaneously put into the water bath
for testing, the test results should be displayed in the following range: 0 ~ 0.1 ° C,
36.9 ~37.1 ° C, 49.9 ~ 50.1 ° C
58
Defibrillator Monitor S Series Service Manual
59