Calabration Procedure
Calabration Procedure
Procedure: Calibration
1.0 SUMMARY
1.1. The purpose of this procedure is to define the requirements for calibration or verification of
equipment used to determine the acceptability of product.
1.2. Typically, this is limited to inspection or test equipment used to “buy off” product prior to
movement to a subsequent process or prior to final delivery. However, at the discretion of
management, calibration or verification may also be applied to critical process equipment.
1.3. The [who?] is responsible for implementation and management of this procedure.
2.0 REVISION AND APPROVAL
Rev. Date Nature of Changes Approved By
[Rev [Procedure
[Date of Issue] Original issue.
Number] Approver Name]
Page 1 of 2
[Full Client Name Reg Caps]
Procedure: [Calibration Proc. Title]
Rev. [Rev Number]
the identification of any standards used by the calibration house, and their serial numbers,
allowing for traceability to NIST.
3.8. For tools calibrated in-house by [Short Client Name] staff, the results and standards used shall
be recorded on the Calibration Record and shall include any standards and/or procedures uses.
3.9. Calibrated devices will be identified with a calibration sticker that includes the current
calibration status, calibration due date, and device identification number. Where the device
cannot accommodate a calibration sticker due to size or frequency of use, the device shall be
numbered and the [who?] shall keep a log of those devices and their status. Employees may
only use devices for acceptance testing that are current on calibration.
3.10. Employees shall submit expired tools to [who?] for recalibration, and/or the [who?] shall
positively recall such expired tools for recalibration.
3.11. An “amnesty window” of 2 weeks [ adjust accordingly] is allowed beyond the due date
marked on the device, to accommodate arrangement of calibration or verification, or for
production capacity purposes. [Oxebridge recommends against this practice, and recommends
deleting this paragraph entirely.]
3.12. Devices in use for noncritical measurements are to be marked REFERENCE ONLY. [ this may
not be possible for all companies with large amounts of noncritical tools. Consider a different
approach in such cases.]
3.13. Any device failing to meet calibration standards will immediately be taken out of service. The
device may then be destroyed or sent out for repair. Repaired devices must be calibrated before
being returned to service.
3.14. When a measuring device is found to be out of tolerance, and/or reported on the calibration
certificate of having been found as “defective” or “out of tolerance” by the third-party provider,
the [who?] shall be notified immediately. The [who?] or designate shall oversee a study to
determine the impact of the out-of-tolerance device on product shipped; if deemed necessary, a
recall may be initiated. The customer possessing the material in question is immediately notified
of the problem. This study and the results shall be recorded and placed in the calibration file.
3.15. Measuring & monitoring devices must be stored and handled in a manner that does not
invalidate their calibration or ability to function without error.
4.0 VERIFICATION
4.1. Where a device cannot be calibrated against traceable standards, it must be verified against
some known-good object or method. This may be done by comparing the part against another
part or tool which has been evaluated and validated and proven as acceptable.
4.2. Known-good objects must be protected so their status is not altered, either by physical damage
or deterioration.
4.3. Known-good methods must be documented in procedures, with a rationale for their
acceptability being documented.
Page 2 of 2